Medrad Continuum: troubleshooting and error codes
The ServiceForge knowledge base holds 15 documented symptoms for the Medrad Continuum, with 26 ranked causes, 17 diagnostic checks with test points and expected values, and 29 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Down occlusion alarm patient safety
Infusion stops; downstream occlusion detected.
Ranked causes
- 1. Administration set is kinked (common)
- 2. Cannula is blocked (common)
- 3. Administration set loaded incorrectly (occasional)
- 4. Clip clamp on patient extension set is closed (occasional)
Visually inspect administration set for kinks
- Test point
- Full length of administration set tubing
- Expected
- No kinks in tubing
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air in line alarm patient safety
Infusion stops; air detected in administration set.
Ranked causes
- 1. Air present in administration set (common)
- 2. Roller or clip clamp closed (common)
- 3. IV line loaded incorrectly (common)
Visually inspect administration set for air bubbles
- Test point
- Administration set tubing
- Expected
- No visible air bubbles in set
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air sensor malfunction or disconnection patient safety
Air sensor not functioning properly, may be disconnected or misaligned in front housing
Ranked causes
- 1. Air sensor disconnected at JP4 (occasional)
- 2. Air sensor glue failure or misalignment (occasional)
Verify air sensor connector is seated at JP4
- Test point
- Connector JP4
- Expected
- Connector fully seated and secure
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery charging or charged indicators do not illuminate
Red battery charging LED and green battery charged LED remain dark when pump is in charger; yellow power light is illuminated.
Ranked causes
- 1. F1 fuse on Continuum motor PCB is open (common)
- 2. Battery pack not functional (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery fully charged indicator delayed caution
Red battery LED stays lit longer than expected before green LED illuminates
Ranked causes
- 1. Normal charging cycle in progress (common)
- 2. Battery left discharged for extended period (occasional)
Monitor charge progress
- Test point
- Charger LEDs
- Expected
- Red LED lit during charging; green LED when fully charged
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery not charging
Red battery LED on charger does not light when pump is inserted
Ranked causes
- 1. Charger not connected to AC power (common)
- 2. Pump not fully seated in charger (common)
Verify AC power connection
- Test point
- AC indicator on charger
- Expected
- AC indicator is lit
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery output less than 7.2 VDC
Measured battery voltage is below 7.2 volts; pump power output insufficient.
Ranked causes
- 1. Battery is discharged (common)
- 2. Battery pack not functional (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump unattended alarm
Pump operation stops after 2 minutes without button press.
Ranked causes
- 1. Two minutes elapsed without button press (common)
- 2. Pump left in stop state more than 2 minutes (common)
Check if pump stopped due to inactivity
- Test point
- Pump display and operation status
- Expected
- Pump in standby/stop mode
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
End battery alarm
Pump operation stops; battery completely depleted.
Ranked causes
- 1. Battery is depleted (common)
Confirm pump stopped due to battery depletion
- Test point
- Pump power and display
- Expected
- Pump inoperative; no power
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fatal error alarm patient safety
Pump operation stops; fatal internal error detected.
Ranked causes
- 1. Fatal internal error in pump electronics or firmware (rare)
Record error number and attempt restart
- Test point
- Pump display and error code
- Expected
- Error number visible on display
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Green battery charged indicator does not light after five hours on charge
After battery has been on charger for more than five hours, green 'charged' LED does not illuminate.
Ranked causes
- 1. Battery pack not functional (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lock mode enabled
Settings cannot be changed; pump locked.
Ranked causes
- 1. Lock mode is turned on (common)
Confirm lock mode status
- Test point
- Pump display or menu
- Expected
- Lock mode disabled
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery alarm
Pump will run for only 15 more minutes; LED blinks red; alarm sounds repeatedly.
Ranked causes
- 1. Battery charge depleted to 15 minutes capacity at max flow (common)
Confirm low battery indication
- Test point
- LED color and display message
- Expected
- Red LED blinking and 'Low Battery' displayed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Missing key alarm
Pump will not start; key not detected in keyway.
Ranked causes
- 1. Administration set not loaded correctly with key in keyway (common)
Verify administration set key in keyway
- Test point
- Set loading slot and keyway
- Expected
- Administration set key fully seated in keyway
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No LEDs illuminated when pump is in charger
No LED indicators light up when pump is positioned in charger unit.
Ranked causes
- 1. No AC power at the charger (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
10 | Mechanical malfunction | Pump mechanical failure | error |
1, 11, 12, 13, 26 | Motor control or encoder malfunction | Motor unable to turn due to mechanical obstruction, encoder damage, or electronic control failure | error |
14 | Microprocessor malfunction | Main processor failure | critical |
15 | Microprocessor malfunction | Main processor failure | critical |
16 | Electronic malfunction | Electronic control circuit failure | error |
17 | Electronic malfunction | Electronic control circuit failure | error |
19 | Electronic malfunction; motor failure during on/off test | Motor fails to respond during power cycle test | error |
20 | Microprocessor malfunction | Main processor failure | critical |
21 | Microprocessor malfunction | Main processor failure | critical |
23 | External Timer Battery depleted | Backup battery voltage low | warning |
24, 28, 32 | External interrupt or electronic malfunction | Electrostatic discharge or system fault | error |
25 | Battery malfunction | Multiple battery-related issues | error |
27 | Electronic malfunction | Electronic control circuit failure | error |
28 | Encoder detects external light | PMT sensor exposed to external light or device case improperly sealed | error |
30 | Memory malfunction | Data storage or retrieval error | warning |
33 | Microprocessor malfunction | Main processor failure | critical |
34 | Microprocessor malfunction | Main processor failure | critical |
35 | Wrong Pressure Calibration | Pressure sensor calibration out of specification | error |
39 | Microprocessor malfunction or software code corrupted | Processor failure or firmware corruption | critical |
7 | Very low reading during operation | System reading abnormally low | warning |
Air in Line | Infusion stopped due to air in administration set | Air bubble detected in fluid path | warning |
Down Occlusion | Infusion stopped due to downstream occlusion | Pressure buildup in downstream tubing or cannula | warning |
End Battery | Battery completely depleted; pump inoperative | Battery charge completely exhausted | error |
End Program | Programmed infusion completed normally | Pump delivered programmed volume or ran programmed duration | info |
Fatal Error | Pump internal system failure; operation unsafe | Critical firmware, electronic, or hardware malfunction detected | critical |
Lock Mode | Pump settings locked; no parameter changes allowed | Lock mode activated on pump control panel | info |
Low Battery | Battery capacity reduced to 15 minutes runtime at maximum flow | Battery charge level drops to low threshold | warning |
Missing Key | Administration set key not detected in keyway | Administration set not loaded or key not in keyway position | warning |
Pump unattended | Pump operation stopped due to inactivity timeout | No button press for 2 minutes | info |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI