Medrad Continuum: troubleshooting and error codes

The ServiceForge knowledge base holds 15 documented symptoms for the Medrad Continuum, with 26 ranked causes, 17 diagnostic checks with test points and expected values, and 29 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Down occlusion alarm patient safety

Infusion stops; downstream occlusion detected.

Ranked causes

  1. 1. Administration set is kinked (common)
  2. 2. Cannula is blocked (common)
  3. 3. Administration set loaded incorrectly (occasional)
  4. 4. Clip clamp on patient extension set is closed (occasional)
First check

Visually inspect administration set for kinks

Test point
Full length of administration set tubing
Expected
No kinks in tubing

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air in line alarm patient safety

Infusion stops; air detected in administration set.

Ranked causes

  1. 1. Air present in administration set (common)
  2. 2. Roller or clip clamp closed (common)
  3. 3. IV line loaded incorrectly (common)
First check

Visually inspect administration set for air bubbles

Test point
Administration set tubing
Expected
No visible air bubbles in set

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air sensor malfunction or disconnection patient safety

Air sensor not functioning properly, may be disconnected or misaligned in front housing

Ranked causes

  1. 1. Air sensor disconnected at JP4 (occasional)
  2. 2. Air sensor glue failure or misalignment (occasional)
First check

Verify air sensor connector is seated at JP4

Test point
Connector JP4
Expected
Connector fully seated and secure

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery charging or charged indicators do not illuminate

Red battery charging LED and green battery charged LED remain dark when pump is in charger; yellow power light is illuminated.

Ranked causes

  1. 1. F1 fuse on Continuum motor PCB is open (common)
  2. 2. Battery pack not functional (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery fully charged indicator delayed caution

Red battery LED stays lit longer than expected before green LED illuminates

Ranked causes

  1. 1. Normal charging cycle in progress (common)
  2. 2. Battery left discharged for extended period (occasional)
First check

Monitor charge progress

Test point
Charger LEDs
Expected
Red LED lit during charging; green LED when fully charged

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery not charging

Red battery LED on charger does not light when pump is inserted

Ranked causes

  1. 1. Charger not connected to AC power (common)
  2. 2. Pump not fully seated in charger (common)
First check

Verify AC power connection

Test point
AC indicator on charger
Expected
AC indicator is lit

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery output less than 7.2 VDC

Measured battery voltage is below 7.2 volts; pump power output insufficient.

Ranked causes

  1. 1. Battery is discharged (common)
  2. 2. Battery pack not functional (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pump unattended alarm

Pump operation stops after 2 minutes without button press.

Ranked causes

  1. 1. Two minutes elapsed without button press (common)
  2. 2. Pump left in stop state more than 2 minutes (common)
First check

Check if pump stopped due to inactivity

Test point
Pump display and operation status
Expected
Pump in standby/stop mode

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

End battery alarm

Pump operation stops; battery completely depleted.

Ranked causes

  1. 1. Battery is depleted (common)
First check

Confirm pump stopped due to battery depletion

Test point
Pump power and display
Expected
Pump inoperative; no power

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fatal error alarm patient safety

Pump operation stops; fatal internal error detected.

Ranked causes

  1. 1. Fatal internal error in pump electronics or firmware (rare)
First check

Record error number and attempt restart

Test point
Pump display and error code
Expected
Error number visible on display

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Green battery charged indicator does not light after five hours on charge

After battery has been on charger for more than five hours, green 'charged' LED does not illuminate.

Ranked causes

  1. 1. Battery pack not functional (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Lock mode enabled

Settings cannot be changed; pump locked.

Ranked causes

  1. 1. Lock mode is turned on (common)
First check

Confirm lock mode status

Test point
Pump display or menu
Expected
Lock mode disabled

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low battery alarm

Pump will run for only 15 more minutes; LED blinks red; alarm sounds repeatedly.

Ranked causes

  1. 1. Battery charge depleted to 15 minutes capacity at max flow (common)
First check

Confirm low battery indication

Test point
LED color and display message
Expected
Red LED blinking and 'Low Battery' displayed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Missing key alarm

Pump will not start; key not detected in keyway.

Ranked causes

  1. 1. Administration set not loaded correctly with key in keyway (common)
First check

Verify administration set key in keyway

Test point
Set loading slot and keyway
Expected
Administration set key fully seated in keyway

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

No LEDs illuminated when pump is in charger

No LED indicators light up when pump is positioned in charger unit.

Ranked causes

  1. 1. No AC power at the charger (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
10Mechanical malfunctionPump mechanical failureerror
1, 11, 12, 13, 26Motor control or encoder malfunctionMotor unable to turn due to mechanical obstruction, encoder damage, or electronic control failureerror
14Microprocessor malfunctionMain processor failurecritical
15Microprocessor malfunctionMain processor failurecritical
16Electronic malfunctionElectronic control circuit failureerror
17Electronic malfunctionElectronic control circuit failureerror
19Electronic malfunction; motor failure during on/off testMotor fails to respond during power cycle testerror
20Microprocessor malfunctionMain processor failurecritical
21Microprocessor malfunctionMain processor failurecritical
23External Timer Battery depletedBackup battery voltage lowwarning
24, 28, 32External interrupt or electronic malfunctionElectrostatic discharge or system faulterror
25Battery malfunctionMultiple battery-related issueserror
27Electronic malfunctionElectronic control circuit failureerror
28Encoder detects external lightPMT sensor exposed to external light or device case improperly sealederror
30Memory malfunctionData storage or retrieval errorwarning
33Microprocessor malfunctionMain processor failurecritical
34Microprocessor malfunctionMain processor failurecritical
35Wrong Pressure CalibrationPressure sensor calibration out of specificationerror
39Microprocessor malfunction or software code corruptedProcessor failure or firmware corruptioncritical
7Very low reading during operationSystem reading abnormally lowwarning
Air in LineInfusion stopped due to air in administration setAir bubble detected in fluid pathwarning
Down OcclusionInfusion stopped due to downstream occlusionPressure buildup in downstream tubing or cannulawarning
End BatteryBattery completely depleted; pump inoperativeBattery charge completely exhaustederror
End ProgramProgrammed infusion completed normallyPump delivered programmed volume or ran programmed durationinfo
Fatal ErrorPump internal system failure; operation unsafeCritical firmware, electronic, or hardware malfunction detectedcritical
Lock ModePump settings locked; no parameter changes allowedLock mode activated on pump control panelinfo
Low BatteryBattery capacity reduced to 15 minutes runtime at maximum flowBattery charge level drops to low thresholdwarning
Missing KeyAdministration set key not detected in keywayAdministration set not loaded or key not in keyway positionwarning
Pump unattendedPump operation stopped due to inactivity timeoutNo button press for 2 minutesinfo

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI