Medivators DSD: troubleshooting and error codes

The ServiceForge knowledge base holds 12 documented symptoms for the Medivators DSD, with 57 ranked causes, 0 diagnostic checks with test points and expected values, and 11 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Slow water fill into basin

Water enters the basin at a reduced rate during the fill phase

Ranked causes

  1. 1. Air pressure insufficient to open valves (common)
  2. 2. Insufficient water supply pressure or flow rate (common)
  3. 3. External water pre-filter plugged (common)
  4. 4. Internal water filter is plugged (common)
  5. 5. Incoming water regulator setting is incorrect (occasional)
  6. 6. Air pressure switch harness disconnected (occasional)
  7. 7. Water line valve not opening (occasional)
  8. 8. Chamber valve not opening (occasional)
  9. 9. Air manifold not operating (occasional)
  10. 10. Valve drive board defective (rare)
  11. 11. Harness disconnected (occasional)
  12. 12. Air trapped in filter housing (occasional)

12 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Slow disinfectant fill into basin

Disinfectant solution enters the basin at a reduced rate during the disinfection phase

Ranked causes

  1. 1. Air pressure insufficient to open valves (common)
  2. 2. Disinfectant screen plugged (common)
  3. 3. Air pressure switch harness disconnected (occasional)
  4. 4. Disinfectant supply valve not opening (occasional)
  5. 5. Chamber valve not opening (occasional)
  6. 6. Air manifold not operating (occasional)
  7. 7. Valve drive board defective (rare)
  8. 8. Harness disconnected (occasional)

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printer does not operate

Printer is not functioning or producing output

Ranked causes

  1. 1. No power to reprocessor (common)
  2. 2. Power supply unplugged (occasional)
  3. 3. Printer ribbon is jammed (occasional)
  4. 4. Printer paper is jammed (occasional)
  5. 5. Printer connection is disconnected (occasional)
  6. 6. Printer damaged or defective (occasional)
  7. 7. Defective motherboard (rare)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Reprocessor does not start

Unit is not powered on or responding to control inputs

Ranked causes

  1. 1. No power to the reprocessor (common)
  2. 2. GFCI is tripped (common)
  3. 3. Main circuit breaker is tripped (common)
  4. 4. Defective connection to reprocessor (occasional)
  5. 5. Defective connection to motherboard (occasional)
  6. 6. Blown power supply fuse (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Reprocessor lid does not open on units with optional lidlocks

The reprocessor lid remains locked and cannot be opened manually

Ranked causes

  1. 1. Lidlock remains activated (common)
  2. 2. Lidlock defective (occasional)
  3. 3. Harness disconnected or defective (occasional)
  4. 4. Digital I/O board defective (rare)
  5. 5. 24/28V power supply defective (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Scopes still wet at end of cycle

Endoscopes remain wet after the drying/air phase concludes

Ranked causes

  1. 1. Insufficient air flow (common)
  2. 2. Air valve not opening (occasional)
  3. 3. Compressor vent valve open or leaking (occasional)
  4. 4. Chamber valves leaking (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Water does not drain from basin during flush or rinse

Water remains in the basin after flush or rinse phase instead of draining

Ranked causes

  1. 1. Drain valve is defective (common)
  2. 2. External drain line kinked (common)
  3. 3. Drain line plugged (common)
  4. 4. Basin drain screen plugged (common)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Disinfection cycle does not start

Unit powers on but does not begin the disinfection cycle

Ranked causes

  1. 1. Software/hardware locked (occasional)
  2. 2. Keypad not responding (occasional)
  3. 3. Control panel defective (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

LCD screen unreadable

LCD display is difficult to see or shows no visible content

Ranked causes

  1. 1. Contrast out of adjustment (common)
  2. 2. Ribbon cable disconnected or damaged (occasional)
  3. 3. LCD screen failure (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No air after cycle

Compressed air is not available or exits the system after the cycle completes

Ranked causes

  1. 1. Incorrect air program setting (common)
  2. 2. Excessive fluid remaining in endoscope channels (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printing unreadable

Printer output is faint, unclear, or illegible

Ranked causes

  1. 1. Printer ribbon is infrequently used (common)
  2. 2. Printer ribbon worn (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Water flows constantly from hookups

Water continuously exits the water hookup/connection without stopping

Ranked causes

  1. 1. Water valve not sealing correctly (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
Bas Sen ErrBasin sensor error - the basin level sensor has detected a problemDetected during basin fill or drainage monitoringerror
Flow Sen ErrFlow sensor error - the water flow sensor is not functioning correctlyDetected during initial starterror
Lid AjarReprocessor lid is open - safety interlock detected open lid during operationDetected during initial startwarning
Low ChamberChamber pressure low - insufficient pressure in the disinfectant or rinse chamberDetected during disinfectant phase or disinfectant/rinse phaseserror
No Air FlowNo air flow detected - compressed air system is not functioningDetected when air flow is needed for operationerror
No Fluid FI2No fluid flow indicator 2 - fluid flow failure in secondary circuit or drain pathDetected during disinfection or drainage phaseserror
No Fluid FlowNo fluid flow detected - water or disinfectant is not flowing as expectedDetected during disinfect, flush, or rinse phaseserror
Res T HighReservoir temperature high - water temperature exceeds acceptable limitsDetected during heating phase or cycle operationwarning
Res T LowReservoir temperature low - water temperature is below acceptable limitsDetected during heating phase or cycle operationwarning
Sheath FailLeak test failure - the optional leak tester detected a leak in the endoscopeDetected during leak test phaseserror
Shth Sen ErrSheath sensor error - the leak tester sensor has detected a problemDetected during initial starterror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.