Medivators Advantage Plus 2.0: troubleshooting and error codes

The ServiceForge knowledge base holds 119 documented symptoms for the Medivators Advantage Plus 2.0, with 183 ranked causes, 55 diagnostic checks with test points and expected values, and 140 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

No flow: channel

Disinfection cycle detects no flow in Air, Suction, Elevator, Biopsy, Jet, Extra, or Water channel during program

Ranked causes

  1. 1. Channel connector not connected (common)
  2. 2. Terminal block handle not fastened or misaligned (common)
  3. 3. Incorrect program selected for instrument (occasional)
  4. 4. Wrong block installed (occasional)
  5. 5. Insufficient pump pressure or faulty channel sensor (occasional)
  6. 6. Sprayer worn, causing excessive pressure loss (occasional)
  7. 7. Channel sensor out of alignment (occasional)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Basin not emptying

Rinse basin fails to drain properly during cycle

Ranked causes

  1. 1. Drain is blocked (common)
  2. 2. Drain valve is faulty (occasional)
  3. 3. Drainage hose sags (occasional)
  4. 4. Basin empty sensor is faulty (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Minor endoscope leak, cycle continuing safely patient safety

Very tiny leak detected in endoscope, leak test system, terminal connector, or connection block; cycle continues with warning

Ranked causes

  1. 1. Very tiny leak in endoscope (occasional)
  2. 2. Tiny leak in leak test system, terminal connector, or connection block (occasional)
  3. 3. Power failure vented endoscope (occasional)
  4. 4. Endoscope stays pressurized, faulty sensor (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Previous run ended incorrectly, program aborted to safe state patient safety

System error occurred during run, endoscope rinsed with water only (not disinfected), recovery program runs automatically and cannot be stopped

Ranked causes

  1. 1. PC powered off or defective (occasional)
  2. 2. Communication problem: cable fault, loose cable, or program error (occasional)
  3. 3. SCU interface communication problem with warning W-701 (occasional)
  4. 4. LIO program interrupted with warning W-703, cover locked (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Basin level dropped

Liquid level in rinse basin drops unexpectedly during operation

Ranked causes

  1. 1. Valve seal contamination causes leakage (common)
  2. 2. Drainage valve opened due to control error (occasional)
  3. 3. Rinse basin level sensor is defective or contaminated (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Blocked channel caution

Disinfection cycle detects blockage in Air, Suction, Elevator, Biopsy, Jet, Extra, or Water channel; message persists after restart attempt

Ranked causes

  1. 1. Connection hose kinked or pinched (common)
  2. 2. Endoscope channel blocked (common)
  3. 3. Connection block obstructed (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Leak test venting blocked, pressure remains high patient safety

Air valve defective or leak test sensor faulty, preventing proper endoscope ventilation after leak testing

Ranked causes

  1. 1. Air valve defect, air not venting from endoscope (occasional)
  2. 2. Leak test sensor defective or misaligned (occasional)
  3. 3. Flow control air valve contaminated or misaligned (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No pressure: channel

Disinfection cycle detects insufficient pressure to blow moisture from Air, Suction, Elevator, Biopsy, Jet, Extra, or Water channel

Ranked causes

  1. 1. Insufficient air pressure (common)
  2. 2. Hose detached (common)
  3. 3. Incorrect parameter set (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

SCU not connected

System reports S304 or W701 error indicating sterilization control unit communication failure

Ranked causes

  1. 1. Computer or main board not properly initialized (common)
  2. 2. Main board to PC communication interrupted (occasional)
  3. 3. Peripheral board not communicating (occasional)
First check

Computer has fully shut down and restarted with LIO initialized

Test point
System startup completion
Expected
LIO finished starting without errors

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Start not allowed, vent hook-up block pressurized patient safety

No disinfection program active, leak test sensor on, endoscope remains pressurized after venting too quickly

Ranked causes

  1. 1. Endoscope vented too quickly, remains pressurized (common)
  2. 2. Faulty leak test sensor reading on (occasional)
  3. 3. Air valve leaks (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Start not allowed: Basin not empty

Rinse basin contains water or basin empty sensor faulty; no disinfection program can start until basin is empty and sensor releases

Ranked causes

  1. 1. Water in rinse basin (common)
  2. 2. System not released due to non-zero volume reading (occasional)
  3. 3. Basin empty sensor faulty when basin is empty (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Start not allowed: Vent hook-up block

Disinfection program cannot be started when channel sensor is on (Air, Suction, Elevator, Biopsy, Jet, Extra, or Water channels affected)

Ranked causes

  1. 1. Endoscope channel is blocked (common)
  2. 2. Connection block is blocked (common)
  3. 3. Channel sensor is defective or misaligned (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

W702 or S302 error displayed

Error code indicating a peripheral board communication failure

Ranked causes

  1. 1. SSG board not communicating (unknown)
  2. 2. Channel board not communicating (left or right) (unknown)
  3. 3. POD board not communicating (left or right) (unknown)
First check

Access main board diagnostic CAN errors menu

Test point
Main board diagnostic display
Expected
Second line shows which board is not communicating

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Wrong connection block

Machine disinfection started but wrong block type installed (expects hot water block, found endoscope block, or no block)

Ranked causes

  1. 1. No connection block present (common)
  2. 2. Endoscope block installed instead of disinfection block (common)
  3. 3. Machine disinfection block present but error persists (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Automatic program start not possible, warning messages present

Auto start blocked due to warning messages in report field or pre-existing auto start

Ranked causes

  1. 1. Warning text in report field (occasional)
  2. 2. Auto start for endoscope disinfection already set (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Basin minimum level not reached

Water level in rinse basin fails to reach minimum required level

Ranked causes

  1. 1. Drainage valve leaks (common)
  2. 2. Basin empty sensor is defective (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Basin suddenly empty

Rinse basin empties unexpectedly during rinsing cycle

Ranked causes

  1. 1. Drain valve leaks due to contamination (common)
  2. 2. Basin empty sensor is faulty (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Basin temperature too high caution

Rinse basin or incoming water temperature exceeds acceptable level during cycle

Ranked causes

  1. 1. Drainage block too warm after disinfection (common)
  2. 2. Mixing valve setting is incorrect (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Basin temperature too low

Rinse basin or incoming water temperature is below acceptable level during cycle

Ranked causes

  1. 1. Mixing valve is out of order or incorrectly set (common)
  2. 2. Warm water supply is stagnated (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Block not pressurized or not connected patient safety

Endoscope not properly connected to leak test block, connections loose or disconnected

Ranked causes

  1. 1. Endoscope not properly connected (common)
  2. 2. Machine fault (rare)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Channel disconnected

Control detects loose connection on interconnected endoscope channels (Air, Suction, Elevator, Biopsy, Jet, Extra, or Water)

Ranked causes

  1. 1. Loose connector on endoscope (common)
  2. 2. Uncertain if all channels connected (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Channel not disconnected

Control detects small endoscope channel that should be disconnected but is still connected (Air, Suction, Elevator, Biopsy, Jet, Extra, or Water)

Ranked causes

  1. 1. Loose connector on endoscope (common)
  2. 2. Uncertain if all channels connected (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Control error F041: Air channel error

Air channel sensor malfunction or channel blockage detected

Ranked causes

  1. 1. Air channel sensor error or pump/valve broken (occasional)
  2. 2. Partially blocked air channel or terminal block (occasional)

Control error F051: Basin empty sensor error

Basin empty sensor does not detect rinse basin draining or does not sense disinfectant flow

Ranked causes

  1. 1. Basin empty sensor defective or drainage blocked (occasional)
  2. 2. Sensor out of alignment or endoscope blocks disinfectant flow (occasional)

Control error F052: Basin full sensor error

Basin full sensor not activated when basin fills during refill operation

Ranked causes

  1. 1. Basin full sensor contaminated or water supply too slow (occasional)
  2. 2. Valve leak causes incorrect sensor reading during rinsing (occasional)

Delayed program start not possible, warning messages present

Delayed start blocked due to warning messages during auto startup or pre-existing auto start for another program

Ranked causes

  1. 1. Warning messages W700-W799 present on screen (occasional)
  2. 2. Auto start for another program already set (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Endoscope leak or connected incorrectly patient safety

Endoscope or connection hose leaking seriously, leak test connection faulty, or block misaligned to connection plate

Ranked causes

  1. 1. Endoscope or connection hose leaking seriously (occasional)
  2. 2. Block out of alignment to connection plate (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Internal machine fluid leak, check drip tray caution

Liquid detected in front of machine; program cannot be stopped until sensor released after drying (may take 5–10 minutes)

Ranked causes

  1. 1. Leak present in machine (common)
  2. 2. Sensor misaligned while tray is dry (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Lid open patient safety

Lid open sensor failure during program indicates cover is open when it is actually closed, or lid was physically opened during cycle

Ranked causes

  1. 1. Lid sensor failed during program (occasional)
  2. 2. Lid opened during program (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Measured contact time is too short patient safety

Prescribed contact time for disinfectant is not achieved during cycle

Ranked causes

  1. 1. Drain and fill time for basin changed (occasional)
  2. 2. Program parameters have been changed (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No alcohol flow

Disinfection cycle detects no alcohol flow from reservoir or pump during program

Ranked causes

  1. 1. Alcohol reservoir empty (common)
  2. 2. Alcohol sensor or pump defective (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No detergent flow

Disinfection cycle detects no detergent flow from reservoir or pump during program

Ranked causes

  1. 1. Detergent reservoir empty (common)
  2. 2. Detergent sensor or pump defective (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No water supply

No water available to the machine

Ranked causes

  1. 1. Water tap is closed (common)
  2. 2. Water supply valve is defective (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Peripheral board not responding - blue indicator light abnormal

Board blue indicator light is erratic, solid on, or not flashing; indicates board is not communicating with main board or has failed

Ranked causes

  1. 1. Board communication failure, connector loose, or board defective (unknown)
  2. 2. Board defective or main board failure (unknown)
First check

Observe blue indicator light status during normal operation

Test point
Board indicator LED
Expected
Light flashes normally at regular intervals

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Program aborted with incorrect parameter set

Program aborted to safe situation but wrong parameters were selected for disinfection or recovery program

Ranked causes

  1. 1. Incorrect parameter set selected (occasional)
  2. 2. LIO program interrupted during run with warning W-703 (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Running test cycle with Test Block

Test Block inserted in basin does not initiate or complete test cycle

Ranked causes

  1. 1. Test Block not properly positioned in basin (common)
  2. 2. Incorrect parameter set selected (common)
First check

Verify Test Block is placed in basin

Test point
Basin
Expected
Test Block seated in basin

4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SCU failure during run execution, program aborted to safe state patient safety

SCU error occurred, endoscope rinsed with water only (not disinfected), recovery program runs automatically. Endoscope must complete full disinfection before use.

Ranked causes

  1. 1. Internal PC communication problem: cable fault or loose cable (occasional)
  2. 2. New program started while another active (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

SCU not connected error (S304 or W701)

Error displayed indicating communication failure between main board and peripherals

Ranked causes

  1. 1. Main board or PC communication reset needed (common)
  2. 2. Peripheral board not communicating (channel, SSG, or POD) (occasional)
First check

Check if error clears after power cycle

Test point
LIO display
Expected
Error S304 or W701 no longer shown

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Start not allowed: Alcohol level low

Alcohol reservoir empty or sensor defective; program cannot start

Ranked causes

  1. 1. Alcohol reservoir empty (common)
  2. 2. Alcohol reservoir sensor defective (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Start not allowed: Detergent level low

Detergent reservoir empty or sensor defective; program cannot start

Ranked causes

  1. 1. Detergent reservoir empty (common)
  2. 2. Detergent reservoir sensor defective (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Start not allowed: Fluid in basin

Rinse basin contains water or basin level sensor faulty; level sensor signal is on and no disinfection program can start

Ranked causes

  1. 1. Water in rinse basin (common)
  2. 2. Basin level sensor faulty (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unknown barcode scanned

Barcode reader activated but scanned code not recognized or not in system database

Ranked causes

  1. 1. Barcode belongs to different item or wrong type (occasional)
  2. 2. Barcode not entered in management system (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Water inlet flow too low

Water flows into basin slower than expected

Ranked causes

  1. 1. Water supply pressure is too low or filter is contaminated (common)
  2. 2. Volume meter is contaminated or defective (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air cylinder does not charge

Cylinder does not fill with air to check endoscope connectivity

Ranked causes

  1. 1. Air cylinder not filled (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air cylinder not discharged

Air cylinder remains pressurized and cannot be vented after connectivity check

Ranked causes

  1. 1. Cylinder cannot be vented (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alcohol sensor defective

Alcohol sensor out of alignment or faulty; detected during startup check

Ranked causes

  1. 1. Alcohol sensor out of alignment or faulty (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Another program is active, cannot start machine disinfection

Machine disinfection cannot start because an active cycle is currently running

Ranked causes

  1. 1. Active cycle running (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Barcode not found

Barcode was read but OK button not pressed to confirm

Ranked causes

  1. 1. OK button not enabled after barcode read (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Basin temperature sensor defective high voltage

Temperature reading exceeds 100°C on display

Ranked causes

  1. 1. Temperature sensor is defective (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Basin temperature too high (extreme) high voltage

Water temperature in rinse basin exceeds safe limits (>50°C disinfection, >90°C machine disinfection)

Ranked causes

  1. 1. Water temperature exceeds program limits (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Basin water temperature out of range

Basin temperature displayed on LIO screen does not read 35°C +/- 2°C during water circulation

Ranked causes

  1. 1. External mixing valve not properly adjusted (common)
First check

Monitor basin water temperature on LIO screen

Test point
LIO screen display
Expected
35ºC +/- 2ºC °C, +/- 2°C

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Air compressor preventive maintenance caution

Annual PM for air compressor including filter replacement, check valve service, and leak inspection

Ranked causes

  1. 1. Scheduled preventive maintenance (common)
First check

Unplug power from compressor

Test point
Power cord
Expected
Unplugged

5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Air tank sensor S33 caution

In-place calibration of the Air Cylinder analog sensor S33 to convert analog sensor values to mmHg. Sensor configured to turn on at 20 psi and off at 15 psi.

Ranked causes

  1. 1. Sensor calibration required using low and high pressure settings (unknown)
First check

Set regulator to low pressure setting for calibration

Test point
1.7 bar regulator
Expected
1.0 Bar

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Channel regulators

Annual PM calibration of Channel pressure regulators to 24 psi specification on both sides

Ranked causes

  1. 1. Channel regulators out of specification (unknown)
First check

Measure Channel pressure

Test point
Channel regulator gauge
Expected
24 psi, ± 1 psi

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Connectivity regulator

Annual PM calibration of Connectivity pressure regulator to 1.7 bar specification

Ranked causes

  1. 1. Connectivity regulator out of specification (unknown)
First check

Measure Connectivity pressure

Test point
Connectivity regulator gauge
Expected
1.7 bar, ± 0.1 bar

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Endoscope reprocessor post-restart operation

Verification of normal operation after pagefile configuration and system restart

Ranked causes

  1. 1. System restart required for changes to take effect (common)
First check

Verify Windows restart completion

Test point
Windows boot sequence
Expected
Windows boots to desktop without errors

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Leak Test regulator

Annual PM calibration of Leak Test pressure regulator to 0.25 bar specification

Ranked causes

  1. 1. Leak Test regulator out of specification (unknown)
First check

Measure Leak Test pressure

Test point
Leak Test regulator gauge
Expected
0.25 bar, ± 0.1 bar

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Lid sensor adjustment

Lid position not registering correctly; sensor misaligned with piston

Ranked causes

  1. 1. Piston unscrewed from turnbuckle or sensor misaligned (common)
First check

Verify piston-turnbuckle connection

Test point
Connection at turnbuckle
Expected
Tight, no play

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Main Air regulator

Annual PM calibration of Main Air pressure regulator to 3.0 bar specification

Ranked causes

  1. 1. Main Air regulator out of specification (unknown)
First check

Measure Main Air pressure

Test point
Main Air regulator gauge
Expected
3.0 bar, ± 0.1 bar

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pre-filter regulator

Annual PM calibration of Pre-filter pressure regulator to 35 psi specification

Ranked causes

  1. 1. Pre-filter regulator out of specification (unknown)
First check

Measure Pre-filter pressure

Test point
Pre-filter regulator gauge
Expected
35 psi, ± 1 psi

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Clock synchronized with server
  • Control error F012: Incorrect step
  • Control error F013: Wrong step sequence
  • Control error F014: S14 outside calibration limits
  • Control error F015: S16 outside calibration limits
  • Control error F021: Minimum time violation
  • Control error F022: Maximum time violation
  • Control error F031: Lid open during program
  • Control error F032: Lid open
  • Control error F042: Biopsy channel error
  • Control error F043: Jet channel error
  • Control error F044: Suction channel error
  • Control error F045: Water channel error
  • Control error F046: Extra channel error
  • Control error F047: Elevator channel error
  • Control error F053: Dosing reservoir A not empty
  • Control error F054: Dosing reservoir B not empty
  • Control error F055: Dosing reservoir A not full
  • Control error F056: Dosing reservoir B not full
  • Control error F057: Dosing pump A malfunction
  • Control error F058: Dosing pump B malfunction
  • Control error F062: Detergent sensor error
  • Control error F072: Detergent pump malfunction
  • Control error F073: Channel pump malfunction
  • Control error F074: Alcohol pump malfunction
  • Control error F075: Minimum spray arm speed violation
  • Control error F076: Maximum spray arm speed violation
  • Control error F081: Drainage valve malfunction
  • Expert control timeout, re-login required
  • External water supply valve failed
  • Firmware version prior to 30.s
  • Hardware definition contains invalid signal
  • Incorrect code entered for fluid replacement
  • Internal LIO failure, wrong parameter selection
  • Machine Control locked by other side
  • Main board firmware outdated
  • Main board not communicating with computer
  • MRC validation failed
  • No incoming water flow
  • Peripheral board not communicating with main board
  • Pressure sensor activation threshold adjustment
  • Pressure sensor deactivation hysteresis adjustment
  • Pressure sensor switching points out of specification
  • Run could not be configured, SQL-server connection error
  • Selected hardware definition incorrect for unit type
  • Server disconnected, selection aborted
  • Start not allowed: Check alcohol pressure sensor
  • Start not allowed: Check detergent pressure sensor
  • Tag-reader not available, identification system defective
  • Test cycle with Scope recommended
  • Unknown endoscope tag scanned
  • Unknown operator tag scanned
  • Unrequested inlet water detected
  • User data could not be retrieved, SQL-server connection error
  • Volume measurement V1 > V2
  • Water inlet flow too high
  • Windows registry maximum size error
  • Windows XP registry maximum size error
  • Wrong system side

59 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
C-015Server disconnected, selection abortedConnection to database interruptederror
C-016Run could not be configured, check SQL-server connectionSystem unable to retrieve database information for run setuperror
C-017User data could not be retrieved, check SQL-server connectionSystem unable to access user information from databaseerror
C-019The clock has been synchronized with the server computerTime difference between system and server clock exceeds 30 secondsinfo
C-021The Expert control has been automatically closed, you have to log on againExpert screen inactivity timeoutwarning
C-030The previous run has ended incorrectly. Program aborted to a safe situation.System error occurred during run execution, recovery mode activatederror
C-032A failure occurred during the execution of a run. Program aborted to a safe situation.SCU error during run execution, recovery mode initiatederror
C-033Program aborted to a safe situation, but incorrect parameter setWrong parameters selected for disinfection or recovery program, or LIO program interrupted during runwarning
C-040Hardware definition contains an invalid signal %dIncorrect address entering controller from SCU uniterror
C-042Selected hardware definition not correct for "%s" typeUnit type parameters not properly configured in system settingserror
C-052Automatic start of program is not possible in the current stateWarning messages in report field or auto start already set for another programwarning
C-053Another program is activeMachine disinfection attempted while active cycle runningwarning
C-060Delayed start of program is not possible in the current state.Warning messages present during auto startup or auto start already set for another programwarning
C-080Tag-reader not availableAutomatic identification system (Tag chip reader) not responding or malfunctioningwarning
C-099Internal LIO failure : %sWrong selection executed when setting machine system parameterserror
F012Incorrect stepActive program step not programmed into system control uniterror
F013Wrong step sequenceStop button pressed during run failure; program skipped stepserror
F014S14 outside calibration limitsVolume measurement out of alignment in step 14error
F015S16 outside calibration limitsVolume measurement out of alignment in step 16error
F021Minimum time violationLIO program step time deviates from SCU control time or error report lasted too longwarning
F022Maximum time violationProgram step exceeded maximum allowed durationwarning
F031Lid open during programCover sensor off or air valve operated manually during disinfectioncritical
F032Lid openLid open detected during disinfection programcritical
F041Air channel errorAir channel sensor reads high (pump/valve broken) or low (blockage)error
F042Biopsy channel errorBiopsy channel sensor malfunction or blockageerror
F043Jet channel errorJet channel sensor malfunction or blockageerror
F044Suction channel errorSuction channel sensor malfunction or blockageerror
F045Water channel errorWater channel sensor malfunction or blockageerror
F046Extra channel errorExtra channel sensor malfunction or blockageerror
F047Elevator channel errorElevator channel sensor malfunction or blockageerror
F051Basin empty sensor errorSensor does not detect basin draining or does not sense disinfectant flowerror
F052Basin full sensor errorBasin full sensor not activated during basin fill; reads incorrectly during rinsingerror
F053Dosing reservoir A not emptyPart A reservoir not emptied at end of delivery cyclecritical
F054Dosing reservoir B not emptyPart B reservoir not emptied at end of delivery cyclecritical
F055Dosing reservoir A not fullPart A reservoir not filled at start of delivery cyclecritical
F056Dosing reservoir B not fullPart B reservoir not filled at start of delivery cyclecritical
F057Dosing pump A malfunctionPart A fill pump energized during delivery when it should be offerror
F058Dosing pump B malfunctionPart B fill pump energized during delivery when it should be offerror
F062Detergent sensor errorSensor activated at wrong time or dosage time out of limitswarning
F072Detergent pump malfunctionDetergent pump runs at incorrect time due to software or electronics errorerror
F073Channel pump malfunctionChannel pump runs at incorrect time due to software or electronics errorerror
F074Alcohol pump malfunctionAlcohol pump high (energized) during step when it should be offerror
F075Minimum spray arm speed violationSpray arm spinning below minimum speed of 15 rpmwarning
F076Maximum spray arm speed violationSpray arm spinning above maximum speed of 250 rpmwarning
F081Drainage valve malfunctionDrainage valve opens at wrong step due to software or electronics errorerror
F082Water inlet valve malfunctionWater inlet valve failureerror
F091Volume meter counting pulses at wrong stepDuring volume measurementerror
F092Volume minimum or maximum exceeded during rinse basin fillingWhen filling the rinse basinerror
F094V1 and V2 meter deviation exceeds 15 percentDuring filling verification between V1 and V2 meterserror
F095V2 meter reading exceeds V1 meter readingDuring filling verificationerror
F101Water temperature outside acceptable rangeDuring water temperature monitoringerror
F102Water temperature below minimum thresholdDuring temperature monitoringerror
F103Basin temperature sensor readings differ by more than 3 degrees CelsiusDuring temperature monitoringerror
M-910Unknown barcodeBarcode scanned but not recognized by systemwarning
M-920Unknown operator tagOperator identification tag scanned but not recognizedwarning
M-922The scanned tag is an unknown endoscope codeEndoscope identification tag scanned but not recognizedwarning
M-930Barcode not foundBarcode read but OK button not pressed to confirminfo
M-940The code "%s" is not the correct %s code.Invalid code or lot number entered during fluid replacement utilitywarning
S-102Start not allowed: Vent hook-up blockNo disinfection program active and leak test sensor on, block pressurized or faulty sensorwarning
S-103Endoscope leak or connected incorrectlyLeak test failure or block misalignment detected during connection checkerror
S-104Minor endoscope leak, cycle continuing safelyTiny leak detected but cycle allowed to continue with warningwarning
S-105Leak test vented by pressure remains highVentilation process failed to relieve endoscope pressureerror
S-106Block not pressurized or not connectedLeak test block not receiving pressure or not properly connected to endoscopeerror
S-112, S-122, S-132, S-142, S-152, S-162, S-172Start not allowed: Vent hook-up block (Air, Suction, Elevator, Biopsy, Jet, Extra, Water channels)Disinfection program selected and specific channel sensor is onerror
S-113, S-123, S-133, S-143, S-153, S-163, S-173No flow: channel (Air, Suction, Elevator, Biopsy, Jet, Extra, Water channels)Flow not detected in channel during disinfection programerror
S-114, S-124, S-134, S-144, S-154, S-164, S-174Blocked channel (Air, Suction, Elevator, Biopsy, Jet, Extra, Water channels)Channel blockage detected during program; persists after one restart attempterror
S-115, S-125, S-135, S-145, S-155, S-165, S-175No pressure: channel (Air, Suction, Elevator, Biopsy, Jet, Extra, Water channels)Insufficient pressure to blow moisture from channel during final dry cyclewarning
S-116, S-126, S-136, S-146, S-156, S-166, S-176Channel disconnected (Air, Suction, Elevator, Biopsy, Jet, Extra, Water channels)Interconnected channel control detects loose connection during programerror
S-117, S-127, S-137, S-147, S-157, S-167, S-177Channel not disconnected (Air, Suction, Elevator, Biopsy, Jet, Extra, Water channels)Small channel control detects connection that should be disconnected during programerror
S-201Internal machine fluid leak, check drip trayLeak sensor detects liquid in front of machineerror
S-203Wrong connection blockMachine disinfection startup automatic check detects incorrect block installederror
S-211Lid openLid sensor indicates cover is open during program, or lid physically openederror
S-222Start not allowed: Check detergent pressure sensorSystem startup check detects detergent sensor faulterror
S-223No detergent flowDetergent flow not detected during programerror
S-224Start not allowed: Detergent level lowDetergent reservoir level sensor indicates low or empty level at startupwarning
S-241Alcohol sensor defectiveAlcohol sensor fault detected during startup checkerror
S-242Start not allowed: Check alcohol pressure sensorAlcohol pressure sensor fault detected at startuperror
S-243No alcohol flowAlcohol flow not detected during programerror
S-244Start not allowed: Alcohol level lowAlcohol reservoir level sensor indicates low or empty level at startupwarning
S-271Start not allowed: Fluid in basinBasin level sensor on when no disinfection program activewarning
S-272Start not allowed: Basin not emptyBasin empty sensor on when no disinfection program active; volume reading not zerowarning
S-273Basin suddenly emptyDrain valve leaks or basin empty sensor fails during rinsingwarning
S-274Basin not emptyingDrain blockage, faulty drain valve, sagging hose, or sensor failurewarning
S-275Basin level droppedValve seal contamination, control error, or defective sensorwarning
S-276Basin minimum level not reachedDrainage valve leak or defective basin empty sensorwarning
S-281No incoming water flowWater volume not measured by sensorerror
S-282Water inlet flow too highWater supply pressure too highwarning
S-283Water inlet flow too lowLow water supply pressure or contaminated filterwarning
S-285No water supplyWater tap closed or supply valve defectiveerror
S-286Volume measurement V1 > V2Flow meter measurement deviationerror
S-287Volume measurement V1 < V2Flow meter measurement deviationerror
S-288Unrequested inlet water detectedWater supply valve leakingcritical
S-289External water supply valve failedFilter unit supply valve leakingcritical
S-292Basin temperature sensor defectiveTemperature reading exceeds 100°C on displaycritical
S-293Basin temperature too highExcessive drainage block warmth or incorrect mixing valve settingwarning
S-294Basin temperature too lowIncorrect mixing valve setting or stagnant warm water supplywarning
S-297Basin temperature too high (extreme)Water temperature exceeds limits (>50°C disinfection, >90°C machine disinfection)critical
S-301SCU in OK stateSafety Control Unit status checkinfo
S-302SCU wrong configuration detectedPC communication with SCU detects configuration mismatcherror
S302Peripheral board communication errorOne or more peripheral boards (channel, SSG, POD) not responding to main boarderror
S304SCU not connectedMain board or peripheral board communication failureerror
S-304SCU no communication with PCCommunication loss between PC and electronics drive controllererror
S-350Manual control in Expert screen is activeManual control functions engaged in Expert screenwarning
S-360Program stopped by operator during active cycleStop button pressed during active program executioninfo
S-511Measured contact time is too shortDrain/fill timing changed or program parameters modifiedcritical
S-660MRC validation failedDisinfectant sample failed MRC validation test at end of cyclecritical
S-671Machine Control locked by other sideMachine disinfection program active on opposite sideinfo
S-672Wrong system sideMachine disinfection started from right side instead of leftwarning
S-831Air cylinder does not chargeCylinder not filling with air for endoscope connectivity checkwarning
S-832Air cylinder not discharged (pressurized)Cylinder cannot be vented after connectivity checkwarning
W701SCU not connected warningMain board or peripheral board communication failurewarning
W702Peripheral board communication errorOne or more peripheral boards (channel, SSG, POD) not responding to main boardwarning
W702SCU Wrong ConfigurationPeripheral board has lost communication with main boarderror
W-703SCU incorrect status with lid locked and inactiveSafety control rejected previous run and blocked coverwarning
W-704SCU incorrect status with red LED indicatorSafety control program rejects runwarning
W-705SCU in OK status operational stateNormal SCU status monitoringinfo
W-711Cover lid is openCover sensor active when no program runningwarning
W-712Lid will not openAttempt to open cover during or after cyclewarning
W-713Lid will not closeAttempt to close cover during cyclewarning
W-715Basin not empty after programBasin liquid detected after program abort or drainingwarning
W-716Fluid leak detectedLiquid detected during standby or reservoir changewarning
W-721Water temperature sensor defectiveTemperature reading exceeds 100 degrees Celsiuswarning
W-743Detergent level sensor highDuring sensor monitoring in standbywarning
W-744Rinse basin minimum level sensor highDuring sensor monitoringwarning
W-749Alcohol level sensor highDuring sensor monitoringwarning
W-751Leak test sensor S11 high pressure detectedDuring disinfection cycle or after user cancelwarning
W-753Part A dosing reservoir fill sensor S25 highDuring sensor monitoringwarning
W-754Part A dosing reservoir fill sensor S26 highDuring sensor monitoringwarning
W-757Air cylinder sensor S33 highWhen machine in standby without active disinfection programwarning
W-764Air channel sensor highMachine in standby; channel sensor detects blockagewarning
W-765Suction channel sensor highMachine in standby; channel sensor detects blockagewarning
W-766Biopsy channel sensor highMachine in standby; channel sensor detects blockagewarning
W-767Jet channel sensor highMachine in standby; channel sensor detects blockagewarning
W-768Extra channel sensor highMachine in standby; channel sensor detects blockagewarning
W-769Elevator wire channel sensor highMachine in standby; channel sensor detects blockagewarning
W-770Water channel sensor highMachine in standby; channel sensor detects blockagewarning
W-800Manual control active in Expert modeOne or more functions manually set in Expert modewarning
W-801Purge in progress (leak test system)Machine in standby running automatic leak testinfo
W-802Venting alcohol line with lid openMachine idle with pressurized alcohol lineswarning
W-803Discharging air cylinder with lid closedMachine idle with pressurized air cylinderwarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI