Medivators Advantage Plus 2.0: troubleshooting and error codes
The ServiceForge knowledge base holds 119 documented symptoms for the Medivators Advantage Plus 2.0, with 183 ranked causes, 55 diagnostic checks with test points and expected values, and 140 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
No flow: channel
Disinfection cycle detects no flow in Air, Suction, Elevator, Biopsy, Jet, Extra, or Water channel during program
Ranked causes
- 1. Channel connector not connected (common)
- 2. Terminal block handle not fastened or misaligned (common)
- 3. Incorrect program selected for instrument (occasional)
- 4. Wrong block installed (occasional)
- 5. Insufficient pump pressure or faulty channel sensor (occasional)
- 6. Sprayer worn, causing excessive pressure loss (occasional)
- 7. Channel sensor out of alignment (occasional)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Basin not emptying
Rinse basin fails to drain properly during cycle
Ranked causes
- 1. Drain is blocked (common)
- 2. Drain valve is faulty (occasional)
- 3. Drainage hose sags (occasional)
- 4. Basin empty sensor is faulty (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Minor endoscope leak, cycle continuing safely patient safety
Very tiny leak detected in endoscope, leak test system, terminal connector, or connection block; cycle continues with warning
Ranked causes
- 1. Very tiny leak in endoscope (occasional)
- 2. Tiny leak in leak test system, terminal connector, or connection block (occasional)
- 3. Power failure vented endoscope (occasional)
- 4. Endoscope stays pressurized, faulty sensor (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Previous run ended incorrectly, program aborted to safe state patient safety
System error occurred during run, endoscope rinsed with water only (not disinfected), recovery program runs automatically and cannot be stopped
Ranked causes
- 1. PC powered off or defective (occasional)
- 2. Communication problem: cable fault, loose cable, or program error (occasional)
- 3. SCU interface communication problem with warning W-701 (occasional)
- 4. LIO program interrupted with warning W-703, cover locked (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Basin level dropped
Liquid level in rinse basin drops unexpectedly during operation
Ranked causes
- 1. Valve seal contamination causes leakage (common)
- 2. Drainage valve opened due to control error (occasional)
- 3. Rinse basin level sensor is defective or contaminated (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blocked channel caution
Disinfection cycle detects blockage in Air, Suction, Elevator, Biopsy, Jet, Extra, or Water channel; message persists after restart attempt
Ranked causes
- 1. Connection hose kinked or pinched (common)
- 2. Endoscope channel blocked (common)
- 3. Connection block obstructed (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leak test venting blocked, pressure remains high patient safety
Air valve defective or leak test sensor faulty, preventing proper endoscope ventilation after leak testing
Ranked causes
- 1. Air valve defect, air not venting from endoscope (occasional)
- 2. Leak test sensor defective or misaligned (occasional)
- 3. Flow control air valve contaminated or misaligned (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No pressure: channel
Disinfection cycle detects insufficient pressure to blow moisture from Air, Suction, Elevator, Biopsy, Jet, Extra, or Water channel
Ranked causes
- 1. Insufficient air pressure (common)
- 2. Hose detached (common)
- 3. Incorrect parameter set (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SCU not connected
System reports S304 or W701 error indicating sterilization control unit communication failure
Ranked causes
- 1. Computer or main board not properly initialized (common)
- 2. Main board to PC communication interrupted (occasional)
- 3. Peripheral board not communicating (occasional)
Computer has fully shut down and restarted with LIO initialized
- Test point
- System startup completion
- Expected
- LIO finished starting without errors
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Start not allowed, vent hook-up block pressurized patient safety
No disinfection program active, leak test sensor on, endoscope remains pressurized after venting too quickly
Ranked causes
- 1. Endoscope vented too quickly, remains pressurized (common)
- 2. Faulty leak test sensor reading on (occasional)
- 3. Air valve leaks (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Start not allowed: Basin not empty
Rinse basin contains water or basin empty sensor faulty; no disinfection program can start until basin is empty and sensor releases
Ranked causes
- 1. Water in rinse basin (common)
- 2. System not released due to non-zero volume reading (occasional)
- 3. Basin empty sensor faulty when basin is empty (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Start not allowed: Vent hook-up block
Disinfection program cannot be started when channel sensor is on (Air, Suction, Elevator, Biopsy, Jet, Extra, or Water channels affected)
Ranked causes
- 1. Endoscope channel is blocked (common)
- 2. Connection block is blocked (common)
- 3. Channel sensor is defective or misaligned (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
W702 or S302 error displayed
Error code indicating a peripheral board communication failure
Ranked causes
- 1. SSG board not communicating (unknown)
- 2. Channel board not communicating (left or right) (unknown)
- 3. POD board not communicating (left or right) (unknown)
Access main board diagnostic CAN errors menu
- Test point
- Main board diagnostic display
- Expected
- Second line shows which board is not communicating
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Wrong connection block
Machine disinfection started but wrong block type installed (expects hot water block, found endoscope block, or no block)
Ranked causes
- 1. No connection block present (common)
- 2. Endoscope block installed instead of disinfection block (common)
- 3. Machine disinfection block present but error persists (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Automatic program start not possible, warning messages present
Auto start blocked due to warning messages in report field or pre-existing auto start
Ranked causes
- 1. Warning text in report field (occasional)
- 2. Auto start for endoscope disinfection already set (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Basin minimum level not reached
Water level in rinse basin fails to reach minimum required level
Ranked causes
- 1. Drainage valve leaks (common)
- 2. Basin empty sensor is defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Basin suddenly empty
Rinse basin empties unexpectedly during rinsing cycle
Ranked causes
- 1. Drain valve leaks due to contamination (common)
- 2. Basin empty sensor is faulty (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Basin temperature too high caution
Rinse basin or incoming water temperature exceeds acceptable level during cycle
Ranked causes
- 1. Drainage block too warm after disinfection (common)
- 2. Mixing valve setting is incorrect (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Basin temperature too low
Rinse basin or incoming water temperature is below acceptable level during cycle
Ranked causes
- 1. Mixing valve is out of order or incorrectly set (common)
- 2. Warm water supply is stagnated (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Block not pressurized or not connected patient safety
Endoscope not properly connected to leak test block, connections loose or disconnected
Ranked causes
- 1. Endoscope not properly connected (common)
- 2. Machine fault (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Channel disconnected
Control detects loose connection on interconnected endoscope channels (Air, Suction, Elevator, Biopsy, Jet, Extra, or Water)
Ranked causes
- 1. Loose connector on endoscope (common)
- 2. Uncertain if all channels connected (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Channel not disconnected
Control detects small endoscope channel that should be disconnected but is still connected (Air, Suction, Elevator, Biopsy, Jet, Extra, or Water)
Ranked causes
- 1. Loose connector on endoscope (common)
- 2. Uncertain if all channels connected (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control error F041: Air channel error
Air channel sensor malfunction or channel blockage detected
Ranked causes
- 1. Air channel sensor error or pump/valve broken (occasional)
- 2. Partially blocked air channel or terminal block (occasional)
Control error F051: Basin empty sensor error
Basin empty sensor does not detect rinse basin draining or does not sense disinfectant flow
Ranked causes
- 1. Basin empty sensor defective or drainage blocked (occasional)
- 2. Sensor out of alignment or endoscope blocks disinfectant flow (occasional)
Control error F052: Basin full sensor error
Basin full sensor not activated when basin fills during refill operation
Ranked causes
- 1. Basin full sensor contaminated or water supply too slow (occasional)
- 2. Valve leak causes incorrect sensor reading during rinsing (occasional)
Delayed program start not possible, warning messages present
Delayed start blocked due to warning messages during auto startup or pre-existing auto start for another program
Ranked causes
- 1. Warning messages W700-W799 present on screen (occasional)
- 2. Auto start for another program already set (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Endoscope leak or connected incorrectly patient safety
Endoscope or connection hose leaking seriously, leak test connection faulty, or block misaligned to connection plate
Ranked causes
- 1. Endoscope or connection hose leaking seriously (occasional)
- 2. Block out of alignment to connection plate (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal machine fluid leak, check drip tray caution
Liquid detected in front of machine; program cannot be stopped until sensor released after drying (may take 5–10 minutes)
Ranked causes
- 1. Leak present in machine (common)
- 2. Sensor misaligned while tray is dry (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lid open patient safety
Lid open sensor failure during program indicates cover is open when it is actually closed, or lid was physically opened during cycle
Ranked causes
- 1. Lid sensor failed during program (occasional)
- 2. Lid opened during program (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measured contact time is too short patient safety
Prescribed contact time for disinfectant is not achieved during cycle
Ranked causes
- 1. Drain and fill time for basin changed (occasional)
- 2. Program parameters have been changed (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No alcohol flow
Disinfection cycle detects no alcohol flow from reservoir or pump during program
Ranked causes
- 1. Alcohol reservoir empty (common)
- 2. Alcohol sensor or pump defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No detergent flow
Disinfection cycle detects no detergent flow from reservoir or pump during program
Ranked causes
- 1. Detergent reservoir empty (common)
- 2. Detergent sensor or pump defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No water supply
No water available to the machine
Ranked causes
- 1. Water tap is closed (common)
- 2. Water supply valve is defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Peripheral board not responding - blue indicator light abnormal
Board blue indicator light is erratic, solid on, or not flashing; indicates board is not communicating with main board or has failed
Ranked causes
- 1. Board communication failure, connector loose, or board defective (unknown)
- 2. Board defective or main board failure (unknown)
Observe blue indicator light status during normal operation
- Test point
- Board indicator LED
- Expected
- Light flashes normally at regular intervals
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Program aborted with incorrect parameter set
Program aborted to safe situation but wrong parameters were selected for disinfection or recovery program
Ranked causes
- 1. Incorrect parameter set selected (occasional)
- 2. LIO program interrupted during run with warning W-703 (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Running test cycle with Test Block
Test Block inserted in basin does not initiate or complete test cycle
Ranked causes
- 1. Test Block not properly positioned in basin (common)
- 2. Incorrect parameter set selected (common)
Verify Test Block is placed in basin
- Test point
- Basin
- Expected
- Test Block seated in basin
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SCU failure during run execution, program aborted to safe state patient safety
SCU error occurred, endoscope rinsed with water only (not disinfected), recovery program runs automatically. Endoscope must complete full disinfection before use.
Ranked causes
- 1. Internal PC communication problem: cable fault or loose cable (occasional)
- 2. New program started while another active (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SCU not connected error (S304 or W701)
Error displayed indicating communication failure between main board and peripherals
Ranked causes
- 1. Main board or PC communication reset needed (common)
- 2. Peripheral board not communicating (channel, SSG, or POD) (occasional)
Check if error clears after power cycle
- Test point
- LIO display
- Expected
- Error S304 or W701 no longer shown
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Start not allowed: Alcohol level low
Alcohol reservoir empty or sensor defective; program cannot start
Ranked causes
- 1. Alcohol reservoir empty (common)
- 2. Alcohol reservoir sensor defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Start not allowed: Detergent level low
Detergent reservoir empty or sensor defective; program cannot start
Ranked causes
- 1. Detergent reservoir empty (common)
- 2. Detergent reservoir sensor defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Start not allowed: Fluid in basin
Rinse basin contains water or basin level sensor faulty; level sensor signal is on and no disinfection program can start
Ranked causes
- 1. Water in rinse basin (common)
- 2. Basin level sensor faulty (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unknown barcode scanned
Barcode reader activated but scanned code not recognized or not in system database
Ranked causes
- 1. Barcode belongs to different item or wrong type (occasional)
- 2. Barcode not entered in management system (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Water inlet flow too low
Water flows into basin slower than expected
Ranked causes
- 1. Water supply pressure is too low or filter is contaminated (common)
- 2. Volume meter is contaminated or defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air cylinder does not charge
Cylinder does not fill with air to check endoscope connectivity
Ranked causes
- 1. Air cylinder not filled (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air cylinder not discharged
Air cylinder remains pressurized and cannot be vented after connectivity check
Ranked causes
- 1. Cylinder cannot be vented (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alcohol sensor defective
Alcohol sensor out of alignment or faulty; detected during startup check
Ranked causes
- 1. Alcohol sensor out of alignment or faulty (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Another program is active, cannot start machine disinfection
Machine disinfection cannot start because an active cycle is currently running
Ranked causes
- 1. Active cycle running (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode not found
Barcode was read but OK button not pressed to confirm
Ranked causes
- 1. OK button not enabled after barcode read (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Basin temperature sensor defective high voltage
Temperature reading exceeds 100°C on display
Ranked causes
- 1. Temperature sensor is defective (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Basin temperature too high (extreme) high voltage
Water temperature in rinse basin exceeds safe limits (>50°C disinfection, >90°C machine disinfection)
Ranked causes
- 1. Water temperature exceeds program limits (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Basin water temperature out of range
Basin temperature displayed on LIO screen does not read 35°C +/- 2°C during water circulation
Ranked causes
- 1. External mixing valve not properly adjusted (common)
Monitor basin water temperature on LIO screen
- Test point
- LIO screen display
- Expected
- 35ºC +/- 2ºC °C, +/- 2°C
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Air compressor preventive maintenance caution
Annual PM for air compressor including filter replacement, check valve service, and leak inspection
Ranked causes
- 1. Scheduled preventive maintenance (common)
Unplug power from compressor
- Test point
- Power cord
- Expected
- Unplugged
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Air tank sensor S33 caution
In-place calibration of the Air Cylinder analog sensor S33 to convert analog sensor values to mmHg. Sensor configured to turn on at 20 psi and off at 15 psi.
Ranked causes
- 1. Sensor calibration required using low and high pressure settings (unknown)
Set regulator to low pressure setting for calibration
- Test point
- 1.7 bar regulator
- Expected
- 1.0 Bar
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Channel regulators
Annual PM calibration of Channel pressure regulators to 24 psi specification on both sides
Ranked causes
- 1. Channel regulators out of specification (unknown)
Measure Channel pressure
- Test point
- Channel regulator gauge
- Expected
- 24 psi, ± 1 psi
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Connectivity regulator
Annual PM calibration of Connectivity pressure regulator to 1.7 bar specification
Ranked causes
- 1. Connectivity regulator out of specification (unknown)
Measure Connectivity pressure
- Test point
- Connectivity regulator gauge
- Expected
- 1.7 bar, ± 0.1 bar
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Endoscope reprocessor post-restart operation
Verification of normal operation after pagefile configuration and system restart
Ranked causes
- 1. System restart required for changes to take effect (common)
Verify Windows restart completion
- Test point
- Windows boot sequence
- Expected
- Windows boots to desktop without errors
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Leak Test regulator
Annual PM calibration of Leak Test pressure regulator to 0.25 bar specification
Ranked causes
- 1. Leak Test regulator out of specification (unknown)
Measure Leak Test pressure
- Test point
- Leak Test regulator gauge
- Expected
- 0.25 bar, ± 0.1 bar
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Lid sensor adjustment
Lid position not registering correctly; sensor misaligned with piston
Ranked causes
- 1. Piston unscrewed from turnbuckle or sensor misaligned (common)
Verify piston-turnbuckle connection
- Test point
- Connection at turnbuckle
- Expected
- Tight, no play
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Main Air regulator
Annual PM calibration of Main Air pressure regulator to 3.0 bar specification
Ranked causes
- 1. Main Air regulator out of specification (unknown)
Measure Main Air pressure
- Test point
- Main Air regulator gauge
- Expected
- 3.0 bar, ± 0.1 bar
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pre-filter regulator
Annual PM calibration of Pre-filter pressure regulator to 35 psi specification
Ranked causes
- 1. Pre-filter regulator out of specification (unknown)
Measure Pre-filter pressure
- Test point
- Pre-filter regulator gauge
- Expected
- 35 psi, ± 1 psi
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Clock synchronized with server
- Control error F012: Incorrect step
- Control error F013: Wrong step sequence
- Control error F014: S14 outside calibration limits
- Control error F015: S16 outside calibration limits
- Control error F021: Minimum time violation
- Control error F022: Maximum time violation
- Control error F031: Lid open during program
- Control error F032: Lid open
- Control error F042: Biopsy channel error
- Control error F043: Jet channel error
- Control error F044: Suction channel error
- Control error F045: Water channel error
- Control error F046: Extra channel error
- Control error F047: Elevator channel error
- Control error F053: Dosing reservoir A not empty
- Control error F054: Dosing reservoir B not empty
- Control error F055: Dosing reservoir A not full
- Control error F056: Dosing reservoir B not full
- Control error F057: Dosing pump A malfunction
- Control error F058: Dosing pump B malfunction
- Control error F062: Detergent sensor error
- Control error F072: Detergent pump malfunction
- Control error F073: Channel pump malfunction
- Control error F074: Alcohol pump malfunction
- Control error F075: Minimum spray arm speed violation
- Control error F076: Maximum spray arm speed violation
- Control error F081: Drainage valve malfunction
- Expert control timeout, re-login required
- External water supply valve failed
- Firmware version prior to 30.s
- Hardware definition contains invalid signal
- Incorrect code entered for fluid replacement
- Internal LIO failure, wrong parameter selection
- Machine Control locked by other side
- Main board firmware outdated
- Main board not communicating with computer
- MRC validation failed
- No incoming water flow
- Peripheral board not communicating with main board
- Pressure sensor activation threshold adjustment
- Pressure sensor deactivation hysteresis adjustment
- Pressure sensor switching points out of specification
- Run could not be configured, SQL-server connection error
- Selected hardware definition incorrect for unit type
- Server disconnected, selection aborted
- Start not allowed: Check alcohol pressure sensor
- Start not allowed: Check detergent pressure sensor
- Tag-reader not available, identification system defective
- Test cycle with Scope recommended
- Unknown endoscope tag scanned
- Unknown operator tag scanned
- Unrequested inlet water detected
- User data could not be retrieved, SQL-server connection error
- Volume measurement V1 > V2
- Water inlet flow too high
- Windows registry maximum size error
- Windows XP registry maximum size error
- Wrong system side
59 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
C-015 | Server disconnected, selection aborted | Connection to database interrupted | error |
C-016 | Run could not be configured, check SQL-server connection | System unable to retrieve database information for run setup | error |
C-017 | User data could not be retrieved, check SQL-server connection | System unable to access user information from database | error |
C-019 | The clock has been synchronized with the server computer | Time difference between system and server clock exceeds 30 seconds | info |
C-021 | The Expert control has been automatically closed, you have to log on again | Expert screen inactivity timeout | warning |
C-030 | The previous run has ended incorrectly. Program aborted to a safe situation. | System error occurred during run execution, recovery mode activated | error |
C-032 | A failure occurred during the execution of a run. Program aborted to a safe situation. | SCU error during run execution, recovery mode initiated | error |
C-033 | Program aborted to a safe situation, but incorrect parameter set | Wrong parameters selected for disinfection or recovery program, or LIO program interrupted during run | warning |
C-040 | Hardware definition contains an invalid signal %d | Incorrect address entering controller from SCU unit | error |
C-042 | Selected hardware definition not correct for "%s" type | Unit type parameters not properly configured in system settings | error |
C-052 | Automatic start of program is not possible in the current state | Warning messages in report field or auto start already set for another program | warning |
C-053 | Another program is active | Machine disinfection attempted while active cycle running | warning |
C-060 | Delayed start of program is not possible in the current state. | Warning messages present during auto startup or auto start already set for another program | warning |
C-080 | Tag-reader not available | Automatic identification system (Tag chip reader) not responding or malfunctioning | warning |
C-099 | Internal LIO failure : %s | Wrong selection executed when setting machine system parameters | error |
F012 | Incorrect step | Active program step not programmed into system control unit | error |
F013 | Wrong step sequence | Stop button pressed during run failure; program skipped steps | error |
F014 | S14 outside calibration limits | Volume measurement out of alignment in step 14 | error |
F015 | S16 outside calibration limits | Volume measurement out of alignment in step 16 | error |
F021 | Minimum time violation | LIO program step time deviates from SCU control time or error report lasted too long | warning |
F022 | Maximum time violation | Program step exceeded maximum allowed duration | warning |
F031 | Lid open during program | Cover sensor off or air valve operated manually during disinfection | critical |
F032 | Lid open | Lid open detected during disinfection program | critical |
F041 | Air channel error | Air channel sensor reads high (pump/valve broken) or low (blockage) | error |
F042 | Biopsy channel error | Biopsy channel sensor malfunction or blockage | error |
F043 | Jet channel error | Jet channel sensor malfunction or blockage | error |
F044 | Suction channel error | Suction channel sensor malfunction or blockage | error |
F045 | Water channel error | Water channel sensor malfunction or blockage | error |
F046 | Extra channel error | Extra channel sensor malfunction or blockage | error |
F047 | Elevator channel error | Elevator channel sensor malfunction or blockage | error |
F051 | Basin empty sensor error | Sensor does not detect basin draining or does not sense disinfectant flow | error |
F052 | Basin full sensor error | Basin full sensor not activated during basin fill; reads incorrectly during rinsing | error |
F053 | Dosing reservoir A not empty | Part A reservoir not emptied at end of delivery cycle | critical |
F054 | Dosing reservoir B not empty | Part B reservoir not emptied at end of delivery cycle | critical |
F055 | Dosing reservoir A not full | Part A reservoir not filled at start of delivery cycle | critical |
F056 | Dosing reservoir B not full | Part B reservoir not filled at start of delivery cycle | critical |
F057 | Dosing pump A malfunction | Part A fill pump energized during delivery when it should be off | error |
F058 | Dosing pump B malfunction | Part B fill pump energized during delivery when it should be off | error |
F062 | Detergent sensor error | Sensor activated at wrong time or dosage time out of limits | warning |
F072 | Detergent pump malfunction | Detergent pump runs at incorrect time due to software or electronics error | error |
F073 | Channel pump malfunction | Channel pump runs at incorrect time due to software or electronics error | error |
F074 | Alcohol pump malfunction | Alcohol pump high (energized) during step when it should be off | error |
F075 | Minimum spray arm speed violation | Spray arm spinning below minimum speed of 15 rpm | warning |
F076 | Maximum spray arm speed violation | Spray arm spinning above maximum speed of 250 rpm | warning |
F081 | Drainage valve malfunction | Drainage valve opens at wrong step due to software or electronics error | error |
F082 | Water inlet valve malfunction | Water inlet valve failure | error |
F091 | Volume meter counting pulses at wrong step | During volume measurement | error |
F092 | Volume minimum or maximum exceeded during rinse basin filling | When filling the rinse basin | error |
F094 | V1 and V2 meter deviation exceeds 15 percent | During filling verification between V1 and V2 meters | error |
F095 | V2 meter reading exceeds V1 meter reading | During filling verification | error |
F101 | Water temperature outside acceptable range | During water temperature monitoring | error |
F102 | Water temperature below minimum threshold | During temperature monitoring | error |
F103 | Basin temperature sensor readings differ by more than 3 degrees Celsius | During temperature monitoring | error |
M-910 | Unknown barcode | Barcode scanned but not recognized by system | warning |
M-920 | Unknown operator tag | Operator identification tag scanned but not recognized | warning |
M-922 | The scanned tag is an unknown endoscope code | Endoscope identification tag scanned but not recognized | warning |
M-930 | Barcode not found | Barcode read but OK button not pressed to confirm | info |
M-940 | The code "%s" is not the correct %s code. | Invalid code or lot number entered during fluid replacement utility | warning |
S-102 | Start not allowed: Vent hook-up block | No disinfection program active and leak test sensor on, block pressurized or faulty sensor | warning |
S-103 | Endoscope leak or connected incorrectly | Leak test failure or block misalignment detected during connection check | error |
S-104 | Minor endoscope leak, cycle continuing safely | Tiny leak detected but cycle allowed to continue with warning | warning |
S-105 | Leak test vented by pressure remains high | Ventilation process failed to relieve endoscope pressure | error |
S-106 | Block not pressurized or not connected | Leak test block not receiving pressure or not properly connected to endoscope | error |
S-112, S-122, S-132, S-142, S-152, S-162, S-172 | Start not allowed: Vent hook-up block (Air, Suction, Elevator, Biopsy, Jet, Extra, Water channels) | Disinfection program selected and specific channel sensor is on | error |
S-113, S-123, S-133, S-143, S-153, S-163, S-173 | No flow: channel (Air, Suction, Elevator, Biopsy, Jet, Extra, Water channels) | Flow not detected in channel during disinfection program | error |
S-114, S-124, S-134, S-144, S-154, S-164, S-174 | Blocked channel (Air, Suction, Elevator, Biopsy, Jet, Extra, Water channels) | Channel blockage detected during program; persists after one restart attempt | error |
S-115, S-125, S-135, S-145, S-155, S-165, S-175 | No pressure: channel (Air, Suction, Elevator, Biopsy, Jet, Extra, Water channels) | Insufficient pressure to blow moisture from channel during final dry cycle | warning |
S-116, S-126, S-136, S-146, S-156, S-166, S-176 | Channel disconnected (Air, Suction, Elevator, Biopsy, Jet, Extra, Water channels) | Interconnected channel control detects loose connection during program | error |
S-117, S-127, S-137, S-147, S-157, S-167, S-177 | Channel not disconnected (Air, Suction, Elevator, Biopsy, Jet, Extra, Water channels) | Small channel control detects connection that should be disconnected during program | error |
S-201 | Internal machine fluid leak, check drip tray | Leak sensor detects liquid in front of machine | error |
S-203 | Wrong connection block | Machine disinfection startup automatic check detects incorrect block installed | error |
S-211 | Lid open | Lid sensor indicates cover is open during program, or lid physically opened | error |
S-222 | Start not allowed: Check detergent pressure sensor | System startup check detects detergent sensor fault | error |
S-223 | No detergent flow | Detergent flow not detected during program | error |
S-224 | Start not allowed: Detergent level low | Detergent reservoir level sensor indicates low or empty level at startup | warning |
S-241 | Alcohol sensor defective | Alcohol sensor fault detected during startup check | error |
S-242 | Start not allowed: Check alcohol pressure sensor | Alcohol pressure sensor fault detected at startup | error |
S-243 | No alcohol flow | Alcohol flow not detected during program | error |
S-244 | Start not allowed: Alcohol level low | Alcohol reservoir level sensor indicates low or empty level at startup | warning |
S-271 | Start not allowed: Fluid in basin | Basin level sensor on when no disinfection program active | warning |
S-272 | Start not allowed: Basin not empty | Basin empty sensor on when no disinfection program active; volume reading not zero | warning |
S-273 | Basin suddenly empty | Drain valve leaks or basin empty sensor fails during rinsing | warning |
S-274 | Basin not emptying | Drain blockage, faulty drain valve, sagging hose, or sensor failure | warning |
S-275 | Basin level dropped | Valve seal contamination, control error, or defective sensor | warning |
S-276 | Basin minimum level not reached | Drainage valve leak or defective basin empty sensor | warning |
S-281 | No incoming water flow | Water volume not measured by sensor | error |
S-282 | Water inlet flow too high | Water supply pressure too high | warning |
S-283 | Water inlet flow too low | Low water supply pressure or contaminated filter | warning |
S-285 | No water supply | Water tap closed or supply valve defective | error |
S-286 | Volume measurement V1 > V2 | Flow meter measurement deviation | error |
S-287 | Volume measurement V1 < V2 | Flow meter measurement deviation | error |
S-288 | Unrequested inlet water detected | Water supply valve leaking | critical |
S-289 | External water supply valve failed | Filter unit supply valve leaking | critical |
S-292 | Basin temperature sensor defective | Temperature reading exceeds 100°C on display | critical |
S-293 | Basin temperature too high | Excessive drainage block warmth or incorrect mixing valve setting | warning |
S-294 | Basin temperature too low | Incorrect mixing valve setting or stagnant warm water supply | warning |
S-297 | Basin temperature too high (extreme) | Water temperature exceeds limits (>50°C disinfection, >90°C machine disinfection) | critical |
S-301 | SCU in OK state | Safety Control Unit status check | info |
S-302 | SCU wrong configuration detected | PC communication with SCU detects configuration mismatch | error |
S302 | Peripheral board communication error | One or more peripheral boards (channel, SSG, POD) not responding to main board | error |
S304 | SCU not connected | Main board or peripheral board communication failure | error |
S-304 | SCU no communication with PC | Communication loss between PC and electronics drive controller | error |
S-350 | Manual control in Expert screen is active | Manual control functions engaged in Expert screen | warning |
S-360 | Program stopped by operator during active cycle | Stop button pressed during active program execution | info |
S-511 | Measured contact time is too short | Drain/fill timing changed or program parameters modified | critical |
S-660 | MRC validation failed | Disinfectant sample failed MRC validation test at end of cycle | critical |
S-671 | Machine Control locked by other side | Machine disinfection program active on opposite side | info |
S-672 | Wrong system side | Machine disinfection started from right side instead of left | warning |
S-831 | Air cylinder does not charge | Cylinder not filling with air for endoscope connectivity check | warning |
S-832 | Air cylinder not discharged (pressurized) | Cylinder cannot be vented after connectivity check | warning |
W701 | SCU not connected warning | Main board or peripheral board communication failure | warning |
W702 | Peripheral board communication error | One or more peripheral boards (channel, SSG, POD) not responding to main board | warning |
W702 | SCU Wrong Configuration | Peripheral board has lost communication with main board | error |
W-703 | SCU incorrect status with lid locked and inactive | Safety control rejected previous run and blocked cover | warning |
W-704 | SCU incorrect status with red LED indicator | Safety control program rejects run | warning |
W-705 | SCU in OK status operational state | Normal SCU status monitoring | info |
W-711 | Cover lid is open | Cover sensor active when no program running | warning |
W-712 | Lid will not open | Attempt to open cover during or after cycle | warning |
W-713 | Lid will not close | Attempt to close cover during cycle | warning |
W-715 | Basin not empty after program | Basin liquid detected after program abort or draining | warning |
W-716 | Fluid leak detected | Liquid detected during standby or reservoir change | warning |
W-721 | Water temperature sensor defective | Temperature reading exceeds 100 degrees Celsius | warning |
W-743 | Detergent level sensor high | During sensor monitoring in standby | warning |
W-744 | Rinse basin minimum level sensor high | During sensor monitoring | warning |
W-749 | Alcohol level sensor high | During sensor monitoring | warning |
W-751 | Leak test sensor S11 high pressure detected | During disinfection cycle or after user cancel | warning |
W-753 | Part A dosing reservoir fill sensor S25 high | During sensor monitoring | warning |
W-754 | Part A dosing reservoir fill sensor S26 high | During sensor monitoring | warning |
W-757 | Air cylinder sensor S33 high | When machine in standby without active disinfection program | warning |
W-764 | Air channel sensor high | Machine in standby; channel sensor detects blockage | warning |
W-765 | Suction channel sensor high | Machine in standby; channel sensor detects blockage | warning |
W-766 | Biopsy channel sensor high | Machine in standby; channel sensor detects blockage | warning |
W-767 | Jet channel sensor high | Machine in standby; channel sensor detects blockage | warning |
W-768 | Extra channel sensor high | Machine in standby; channel sensor detects blockage | warning |
W-769 | Elevator wire channel sensor high | Machine in standby; channel sensor detects blockage | warning |
W-770 | Water channel sensor high | Machine in standby; channel sensor detects blockage | warning |
W-800 | Manual control active in Expert mode | One or more functions manually set in Expert mode | warning |
W-801 | Purge in progress (leak test system) | Machine in standby running automatic leak test | info |
W-802 | Venting alcohol line with lid open | Machine idle with pressurized alcohol lines | warning |
W-803 | Discharging air cylinder with lid closed | Machine idle with pressurized air cylinder | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.