Medivators ADV: troubleshooting and error codes
The ServiceForge knowledge base holds 4 documented symptoms for the Medivators ADV, with 5 ranked causes, 22 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Calibration: Pressure sensor set points
Set pressure sensor calibration points for leak test, jet, biopsy, air, elevator, water, suction, extra, detergent and alcohol channels using LIO calibration bits and mmHg thresholds.
Ranked causes
- 1. Firmware revision 30.U and higher requires Table 2 calibration values (common)
- 2. Firmware revision 30.T and below requires Table 3 calibration values (common)
Verify main board firmware revision level
- Test point
- Main board
- Expected
- 30.U or higher for Table 2, 30.T or below for Table 3
10 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Air Cylinder Sensor S33
In-place calibration of analog sensor S33 to convert pressure readings to mmHg. Sensor set to turn on at 20 psi and off at 15 psi.
Ranked causes
- 1. Sensor calibration required using low and high pressure settings (common)
Remove right side panel to access regulator and pressure gauge
- Test point
- Cabinet right rear
- Expected
- Panel removed, 1.7 bar regulator and gauge visible
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Air Tank Sensor S33 caution
In-place calibration of the Air Cylinder analog sensor S33 using low and high pressure settings to calculate offset and slope for mmHg conversion
Ranked causes
- 1. Sensor requires calibration using two pressure points (common)
Remove right side panel to access regulator and pressure gauge
- Test point
- Right rear of cabinet
- Expected
- 1.7 bar regulator and pressure gauge accessible
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Regulator pressure set to normal operation
Verify 1.7 bar regulator is set to normal operating pressure of 1.7 bar (25 psi)
Ranked causes
- 1. Regulator adjustment for normal operation (common)
Verify regulator pressure setting
- Test point
- 1.7 bar pressure gauge located in cabinet right rear
- Expected
- 1.7 bar (25 psi), 25 psi
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI