Medivance Arctic Sun 2000: troubleshooting and error codes
The ServiceForge knowledge base holds 83 documented symptoms for the Medivance Arctic Sun 2000, with 158 ranked causes, 141 diagnostic checks with test points and expected values, and 80 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Low water flow or air in pads caution
Low flow rate or continuous air streaming from pads while pump is audible; may follow pad refill cycle
Ranked causes
- 1. Trapped air bubbles in pads (common)
- 2. Pad connectors not fully seated in manifold (common)
- 3. Pad lines kinked or occluded (common)
- 4. Leak in pad (occasional)
- 5. Leak at seal between pad connector and manifold (occasional)
- 6. One pad not functioning properly (occasional)
- 7. Air leak in manifold (occasional)
- 8. External or internal fluid system leak (occasional)
- 9. Pad folded or creased during placement (occasional)
- 10. Clogged external filter or loose fitting (occasional)
- 11. Debris in pump valve or internal leak (occasional)
- 12. Clogged circulating pump internal filter (occasional)
- 13. Ineffective pump head due to internal debris (rare)
Allow time for air to exit system
- Test point
- Pad outlet tubes
- Expected
- No continuous air stream after 20 minutes
12 more checks with pass and fail actions, 13 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control Module is noisy
System produces abnormal audible noise during operation.
Ranked causes
- 1. Cooling fan failing in Power Module (common)
- 2. Cooling fan may be failing in Card Cage (occasional)
- 3. Chiller fan may be failing (occasional)
- 4. Air trapped in Circulating Pump head (common)
- 5. Circulating Pump failing (occasional)
- 6. Mixing Pump failing (occasional)
- 7. Electronics fault (rare)
Determine when noise occurs
- Test point
- System audible output during different modes
- Expected
- Normal operation is quiet or minimal fan hum
6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control Module will not fill
System unable to ingest water when fill cycle is initiated; tank remains empty or below normal level.
Ranked causes
- 1. Pads connected to patient lines prevent suction on fill tube (common)
- 2. Vacuum leak in patient line manifold valve (occasional)
- 3. Bypass or Fill Valve fault (occasional)
- 4. Low voltage circuitry ground fault (rare)
- 5. Electronics fault (rare)
Verify no pads are connected to patient lines
- Test point
- Patient line connectors
- Expected
- All pads physically disconnected from outlet manifold
- Tools
- Visual inspection
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient temperature not decreasing during cooling patient safety
Despite prolonged cooling procedure, patient temperature remains stable or increases; trend indicator unchanged or rising
Ranked causes
- 1. Water temperature not at or near minimum setting (common)
- 2. System in Manual Mode without adequate clinician monitoring (occasional)
- 3. Inadequate water flow rate during treatment (occasional)
- 4. Patient shivering interfering with cooling (non-surgical patients) (occasional)
- 5. Concurrent use of other cooling devices or water blankets (occasional)
Check water temperature on display screen
- Test point
- Display screen
- Expected
- At or near minimum water temperature setting
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient temperature not displayed on screen patient safety
Temperature reading missing or not shown on display despite probe connection
Ranked causes
- 1. Loose connection between probe and cable (common)
- 2. Probe connected to Patient Temperature 2 instead of Patient Temperature 1 (common)
- 3. Loose cable connection to control module back (common)
- 4. Wet temperature cable causing intermittent connection (occasional)
- 5. Temperature out of measurement range (occasional)
Verify probe to cable connection is secure
- Test point
- Temperature probe connector
- Expected
- Fully seated, secure connection
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System fails to purge completely
Pads do not empty fully during purge cycle; water remains in pads or lines.
Ranked causes
- 1. Vent line is kinked (common)
- 2. Water inadvertently entered purge line or valve (occasional)
- 3. Tank is overfilled (common)
- 4. Purge valve inoperative (occasional)
- 5. Electronics fault (rare)
Observe pads and listen during purge initiation
- Test point
- Pads and system audible output
- Expected
- Click sound heard and pads empty well
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Extended period of cold water delivery to patient patient safety
System has delivered water below 10°C/50°F for 8 of the previous 10 hours
Ranked causes
- 1. Prolonged therapeutic hypothermia at low temperature setting (common)
- 2. Insufficient water flow through pads (occasional)
- 3. Inadequate pad coverage or placement (occasional)
- 4. Environmental interference with cooling (occasional)
Assess patient's skin condition
- Expected
- No cold-related injuries or excessive shivering
4 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Patient temperature reading changes significantly within seconds patient safety
Large fluctuations in displayed patient temperature over short time periods, unstable readings
Ranked causes
- 1. Loose or wet connection between patient temperature probe and cable (common)
- 2. Foley probe reading affected by retrograde urine flow past sensor (occasional)
- 3. Probe displaced or in incorrect location (occasional)
- 4. Temperature probe defective or damaged (occasional)
Check connection between patient temperature probe and cable
- Test point
- Probe connection point
- Expected
- Secure, dry connection
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unequal flow through bilateral pads patient safety
Flow through one pad differs significantly from the other pad, causing efficiency loss.
Ranked causes
- 1. Restriction or kink in pad supply line (common)
- 2. Pads not fully adhered to patient (common)
- 3. Pads drying out causing loss of adhesion (occasional)
- 4. Incorrect pad size selection (occasional)
Inspect pad supply lines for kinks
- Test point
- Entire length of tubing between manifold and pads
- Expected
- No compression; open path
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Water cools but unable to cool effectively with patient
Chiller runs but cooling capacity inadequate during patient cooling.
Ranked causes
- 1. Chiller condenser clogged with lint (common)
- 2. Chiller Power Relay or cable fault (occasional)
- 3. Water reservoir level insufficient to run chiller (occasional)
- 4. Electronics or chiller fault (rare)
Visually inspect condenser for lint blockage
- Test point
- Chiller condenser area behind plastic grille
- Expected
- Condenser fins clear of debris
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Water does not cool to target temperature in Manual Mode
Cooling system unable to bring water temperature down to set point despite chiller operation.
Ranked causes
- 1. Chiller fault (occasional)
- 2. Pump cable or connection fault (common)
- 3. Mixing Pump internal filter clogged (occasional)
- 4. Electronics fault (rare)
Verify water temperature T4 with cooling command nonzero
- Test point
- Temperature sensor T4 on display
- Expected
- T4 stabilizes below 6°C with cooling command active
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Water does not heat to target temperature in Manual Mode high voltage
System unable to warm water to set temperature when heater is active and flow adequate.
Ranked causes
- 1. Flow rate less than 800 ml/min (common)
- 2. Heater fuse open due to heater circuit fault (common)
- 3. Power Module fault (occasional)
- 4. Heater thermal breaker tripped (rare)
Measure flow rate in Manual Mode at 42°C target
- Test point
- Diagnostic display flow indicator or collection vessel
- Expected
- 800 ml/min or greater
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low water flow, low pressure, or high PWM with pump audible
Pump can be heard but flow diagnostic shows no flow or minimal flow despite water appearing to flow.
Ranked causes
- 1. Flowmeter fault (common)
- 2. Pump connection fault (common)
- 3. Pump electronics failure (occasional)
Verify flowmeter operation in diagnostic mode
- Test point
- User Diagnostic Mode display
- Expected
- Flow indication present
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient Temperature 1 below control range patient safety
Patient temperature reading below 33°C when system enters Automatic Mode
Ranked causes
- 1. Patient temperature too low for automatic control (common)
- 2. Temperature probe or cable defective or displaced (occasional)
- 3. Control Module calibration error (occasional)
Use CTU to verify Control Module accuracy
- Expected
- Module reading matches actual temperature
3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Patient Temperature 1 erratic or discontinuous
Patient temperature reading shows significant discontinuity; reading changes erratically for more than 8 minutes
Ranked causes
- 1. Patient temperature probe or cable defective (occasional)
- 2. Control Module calibration error (occasional)
- 3. Intermittent connection at rear connector (occasional)
Use CTU to verify Control Module accuracy
- Expected
- Reading matches actual temperature
3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Patient temperature not increasing during warming patient safety
Despite prolonged warming procedure, patient temperature remains stable or decreases; trend indicator unchanged or falling
Ranked causes
- 1. Water temperature not at or near maximum setting (common)
- 2. Inadequate water flow rate during treatment (occasional)
- 3. Concurrent use of other warming devices or water blankets (occasional)
Check water temperature on display screen
- Test point
- Display screen
- Expected
- At or near maximum water temperature setting
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature probe 2 malfunction
Probe 2 reading outside display limits or not responding correctly
Ranked causes
- 1. Temperature probe defective (occasional)
- 2. Temperature probe cable defective or wet (occasional)
- 3. Isolation Card defective (rare)
Use CTU to verify Control Module accuracy
- Expected
- Reading matches actual temperature
3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Water flow rate suddenly decreases caution
Flow to pads drops significantly during operation, may indicate air in lines or damaged pads
Ranked causes
- 1. Fluid delivery or pad lines compressed, kinked, or not patent (common)
- 2. Loose connections between pad lines and fluid delivery line (common)
- 3. Energy Transfer Pad or patient pad line damaged (occasional)
Inspect fluid delivery lines and pad lines
- Test point
- All lines
- Expected
- No kinks, compressions, or blockages
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Water jets from Main Tank Vent in Fill Mode caution
High-pressure water spray exits tank vent during fill operation.
Ranked causes
- 1. Low conductivity of fluid due to missing Cleaning Solution (common)
- 2. Level sensor or cable fault (occasional)
- 3. Electronics fault (rare)
Verify Cleaning Solution was added to fill water
- Test point
- Fill fluid composition and conductivity
- Expected
- Cleaning Solution mixed into distilled water before filling
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Water temperature outside safe range caution
System water temperature above 42.5°C or below 3.5°C; unable to maintain setpoint
Ranked causes
- 1. Cooling system failure or transient condition (common)
- 2. Temperature sensor J5 connector loose or damaged (occasional)
- 3. Solid state relay or temperature cables defective (occasional)
Operate in User Diagnostic Mode and Manual Mode with CTU attached
- Test point
- T1, T2, T3, T4 temperatures
- Expected
- T4 above 3°C for at least 10 minutes after stability, > 3C
- Tools
- CTU
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Control Module will not power up high voltage
System does not respond to power switch; no lights, fans, or audible indicators.
Ranked causes
- 1. Power Card fault (common)
- 2. Service diagnostics failure (rare)
Verify AC power and circuit breaker status
- Test point
- Power cord and facility outlet
- Expected
- AC power available at outlet; no breaker trips
- Tools
- Outlet tester or DMM
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
High Temperature Alert condition patient safety
Patient temperature rises to 39.5°C (103.1°F) with water temperature higher and increasing, triggering alarm that reverts to Stop Mode
Ranked causes
- 1. Patient temperature approaching or above critical high threshold of 39.5°C (103.1°F) (common)
- 2. High Temperature Alert threshold requires adjustment for specific patient or procedure (occasional)
Check current patient temperature
- Test point
- Display screen Patient Temperature reading
- Expected
- Patient temperature below 39.5°C (103.1°F)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low Temperature Alert condition patient safety
Patient temperature drops to 31°C (87.8°F) with water temperature lower and decreasing, triggering alarm that reverts to Stop Mode
Ranked causes
- 1. Patient temperature approaching or below critical low threshold of 31°C (87.8°F) (common)
- 2. Low Temperature Alert threshold requires adjustment for specific patient or procedure (occasional)
Check current patient temperature
- Test point
- Display screen Patient Temperature reading
- Expected
- Patient temperature above 31°C (87.8°F)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No patient temperature displayed with probe installed patient safety
Temperature display absent or shows no reading despite probe being physically connected.
Ranked causes
- 1. Open circuit in probe or cable (common)
- 2. Electronics fault (rare)
Connect probe/cable to another Control Module to test functionality
- Test point
- Probe and cable assembly
- Expected
- Temperature displays correctly on test unit
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor water temperature control with good flow
Water temperature regulation inadequate despite normal flow rate through pads.
Ranked causes
- 1. Main Tank overfilled (common)
- 2. Heater or chiller circuit fault (common)
Check reservoir level in diagnostic mode
- Test point
- User Diagnostic Mode level display (Lx)
- Expected
- Level less than 5
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Prolonged cold exposure to patient patient safety
System has issued 11 extended cold water delivery alerts; patient experiencing prolonged hypothermia
Ranked causes
- 1. Therapeutic hypothermia at minimum temperature setting for extended period (common)
- 2. Patient intolerance to prolonged cold exposure (occasional)
Assess patient's skin condition and core temperature
- Expected
- Skin intact; core temperature stable at target
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Reservoir overfills caution
Tank continues to fill beyond normal full level and overflows.
Ranked causes
- 1. Level sensor or cable fault (common)
- 2. Electronics fault after sensor replacement (rare)
Inspect level sensor harness at I/O Card
- Test point
- Level sensor connector and cable
- Expected
- Secure connection with no damage or corrosion
- Tools
- Visual inspection
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC power lost caution
Notifies user if AC power was previously lost while power switch was in ON position. Current Mode continues. Audio alarm half-second tone every 2 minutes. Alarm/Alert indicator flashing.
Ranked causes
- 1. AC power was interrupted (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration and operational check required patient safety
System requires periodic calibration using Calibration and Test Unit (CTU) to verify Patient Temperature inputs accuracy, Temperature Output circuit, and internal water temperature sensors before operational use.
Ranked causes
- 1. Routine maintenance calibration (common)
Obtain CTU and verify factory calibration data
- Test point
- CTU unit label
- Expected
- Factory calibration data present on CTU
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Diagnostic mode access and monitoring patient safety
Enter diagnostic mode to view system parameters, logs, and test pump/valve operation without therapeutic function
Ranked causes
- 1. Need diagnostic access to system parameters (common)
Power on unit while holding mode keys
- Test point
- Control Module power button and remote Automatic and Purge Mode keys
- Expected
- LCD displays diagnostic home screen after power-up sequence
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Drained water appears green caution
Water drained from system has greenish discoloration indicating chemical contamination.
Ranked causes
- 1. Highly alkaline tap water caused copper chelation from chiller evaporator (rare)
Drain all water and inspect residue color
- Test point
- Drained water appearance
- Expected
- Clear or pale color when refilled with distilled water
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Extended period of cold water delivery patient safety
Circulating water temperature below 10°C/50°F for 8 of previous 10 hours. Reoccurs after 4 hours if condition continues. Current Mode continues. Audio alarm half-second tone every 2 minutes. Alarm/Alert indicator flashing.
Ranked causes
- 1. Extended cold water exposure (occasional)
Measure water flow rate
- Test point
- Water circulating through pads
- Expected
- At least 2.3 liters per minute L/min
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invalid system log entry caution
In Diagnostic Mode, one or more entries in internal system event log are invalid. Used by designated clinical engineering or Medivance service. Current Mode continues. Audio alarm half-second tone every 2 minutes. Alarm/Alert indicator flashing.
Ranked causes
- 1. Corrupt system event log (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low bypass flow caution
Insufficient internal flow during priming or bypass mode. Current Mode continues. Audio alarm half-second tone every 2 minutes. Alarm/Alert indicator flashing.
Ranked causes
- 1. Low bypass flow (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low fluid pressure in delivery lines caution
Fluid circuit delivering inadequate pressure to pads
Ranked causes
- 1. Delivery line open to air or leaking (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low water flow or air leaving pads
Air is present in pads or water lines affecting delivery.
Ranked causes
- 1. Air leak in pads or lines (common)
Visually inspect pads and tubing for air bubbles or moisture
- Test point
- Pad surface and all line connections
- Expected
- No air bubbles; water flows steadily
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low water flow to pads caution
Water circulation to pads decreased significantly
Ranked causes
- 1. Flow decreased by >50% or <0.3 liters per minute (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low water flow with pump inaudible
Pump cannot be heard; pressure sensor shows negative pressure exceeding -8 PSI for more than one minute.
Ranked causes
- 1. Pressure sensor or cable fault (common)
Measure pressure in diagnostic mode
- Test point
- System pressure output
- Expected
- Not more negative than -8 PSI for >1 min
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Manual mode disabled caution
Manual key pressed while Manual Mode is disabled. System changes to Stop Mode. Audio alarm half-second tone every 10 seconds. Alarm/Alert indicator flashing.
Ranked causes
- 1. Manual Mode is disabled (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Non-recoverable system error (codes 60-99) caution
System error that cannot be managed without speaking to Medivance. System changes to Stop Mode. Audio alarm half-second tone every 10 seconds. Alarm/Alert indicator flashing.
Ranked causes
- 1. Internal system failure (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient Temperature 1 above user-defined limit patient safety
Patient Temperature 1 reading equals or exceeds user-set threshold. Audio alarm half-second tone every 2 minutes, Alarm/Alert indicator flashing.
Ranked causes
- 1. Temperature setpoint too low or patient overheating (common)
Verify patient temperature from independent source
- Test point
- Patient monitor or thermometer
- Expected
- Confirm actual patient temperature
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient Temperature 1 below user-defined limit patient safety
Patient Temperature 1 reading equals or is less than user-set threshold. Audio alarm half-second tone every 2 minutes, Alarm/Alert indicator flashing.
Ranked causes
- 1. Temperature setpoint too high or patient cooling (common)
Verify patient temperature from independent source
- Test point
- Patient monitor or thermometer
- Expected
- Confirm actual patient temperature
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient temperature 1 erratic patient safety
Patient temperature discontinuity. Significant change in patient temperature reading for more than 8 minutes. Current Mode continues. Audio alarm half-second tone every 2 minutes. Alarm/Alert indicator flashing.
Ranked causes
- 1. Temperature probe misplaced or dislodged (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient Temperature 1 probe disconnected or out of range patient safety
Patient Temperature 1 reading outside lower limits of display range in Automatic Mode. System changes to Stop Mode. Audio alarm half-second tone every 10 seconds, Alarm/Alert indicator flashing.
Ranked causes
- 1. Temperature probe dislodged, disconnected, or cable wet/damaged (common)
Verify probe placement on patient
- Test point
- Patient temperature probe site
- Expected
- Probe secure and in proper contact with patient
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient Temperature 2 above user-defined limit patient safety
Patient Temperature 2 reading equal to or greater than user-set threshold. Current Mode continues. Audio alarm half-second tone every 2 minutes, Alarm/Alert indicator flashing.
Ranked causes
- 1. Temperature setpoint too low or patient overheating (common)
Verify patient temperature from independent source
- Test point
- Patient monitor or thermometer
- Expected
- Confirm actual patient temperature
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient Temperature 2 below user-defined limit patient safety
Patient Temperature 2 reading equal to or less than user-set threshold. Current Mode continues. Audio alarm half-second tone every 2 minutes, Alarm/Alert indicator flashing.
Ranked causes
- 1. Temperature setpoint too high or patient cooling (common)
Verify patient temperature from independent source
- Test point
- Patient monitor or thermometer
- Expected
- Confirm actual patient temperature
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient Temperature 2 high in Automatic Mode patient safety
Patient Temperature 2 reading above 39.5°C (103.1°F) and water temperature above 39.5°C (103.1°F), system unable to cool. System changes to Stop Mode. Audio alarm half-second tone every 10 seconds, Alarm/Alert indicator flashing.
Ranked causes
- 1. Inadequate cooling capacity, blocked vents, or cooling system failure (common)
Confirm high patient temperature from independent source
- Test point
- Patient monitor or external thermometer
- Expected
- Patient Temperature 2 reading >= 39.5°C confirmed
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient Temperature 2 high in Manual Mode patient safety
Patient Temperature 2 reading above 39.5°C (103.1°F) and water temperature above 39.5°C (103.1°F), system unable to cool in Manual Mode. Current Mode continues. Audio alarm half-second tone every 2 minutes, Alarm/Alert indicator flashing.
Ranked causes
- 1. Manual Water Target Temperature set too high, inadequate cooling, or blocked vents (common)
Inspect machine side vents for blockage
- Test point
- Side vents of Arctic Sun 2000
- Expected
- Vents open and clear
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient Temperature 2 low in Automatic Mode patient safety
Patient Temperature 2 reading below 31.0°C (87.8°F) and water temperature below 31.0°C (87.8°F), system unable to heat. System changes to Stop Mode. Audio alarm half-second tone every 10 seconds, Alarm/Alert indicator flashing.
Ranked causes
- 1. Inadequate heating capacity or heating system failure (common)
Confirm low patient temperature from independent source
- Test point
- Patient monitor or external thermometer
- Expected
- Patient Temperature 2 reading <= 31.0°C confirmed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient Temperature 2 low in Manual Mode patient safety
Patient Temperature 2 reading below 31.0°C (87.8°F) and water temperature below 31.0°C (87.8°F), system unable to heat in Manual Mode. Current Mode continues. Audio alarm half-second tone every 2 minutes, Alarm/Alert indicator flashing.
Ranked causes
- 1. Manual Water Target Temperature set too low (common)
Adjust Water Target Temperature upward
- Test point
- Water Target Temperature setting in Manual Mode
- Expected
- Setting increased from current value
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient Temperature 2 probe disconnected or out of range patient safety
Patient Temperature 2 reading outside lower limits of display range in Automatic Mode. System changes to Stop Mode. Audio alarm half-second tone every 10 seconds, Alarm/Alert indicator flashing.
Ranked causes
- 1. Temperature probe dislodged, disconnected, or cable wet/damaged (common)
Verify probe placement on patient
- Test point
- Patient temperature probe site
- Expected
- Probe secure and in proper contact
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient temperature above user-defined limit patient safety
Patient temperature reading exceeds user-configured high threshold
Ranked causes
- 1. Patient temperature exceeds configured limit (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient temperature and water temperature high patient safety
Both patient and water above 38C in manual mode, cooling ineffective
Ranked causes
- 1. Cooling system ineffective in manual mode (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient temperature and water temperature low patient safety
Both patient and water below 32C in manual mode, heating ineffective
Ranked causes
- 1. Heating system ineffective in manual mode (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient temperature below user-defined limit patient safety
Patient temperature reading below user-configured low threshold
Ranked causes
- 1. Patient temperature below configured limit (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient temperature calibration error patient safety
Internal temperature calibration not functioning
Ranked causes
- 1. Calibration circuit malfunction (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient Temperature calibration error in Automatic Mode patient safety
Internal temperature calibration from Patient Temperature 1 probe input not functioning in Automatic Mode. System changes to Stop Mode. Audio alarm half-second tone every 10 seconds, Alarm/Alert indicator flashing. Patient temperature displayed as three dashes.
Ranked causes
- 1. Probe or internal calibration circuit failure (rare)
Attempt restart in Manual Mode
- Test point
- System operation mode
- Expected
- System operates in Manual Mode with dashes for patient temperature
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient Temperature calibration error in Manual Mode patient safety
Internal temperature calibration from Patient Temperature 1 probe input not functioning in Manual Mode. Current Mode continues. Audio alarm half-second tone every 2 minutes, Alarm/Alert indicator flashing. Patient temperature displayed as three dashes.
Ranked causes
- 1. Probe or internal calibration circuit failure (rare)
Verify probe placement and connection
- Test point
- Temperature probe location and connector
- Expected
- Probe secure and properly positioned
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient temperature out invalid patient safety
Patient Temperature Out calibration data in non-volatile memory invalid. Current Mode continues. Audio alarm half-second tone every 2 minutes. Alarm/Alert indicator flashing.
Ranked causes
- 1. Invalid calibration data (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient temperature out system error patient safety
Calculated Patient Temperature Out differs 0.1°C from measured Patient Temperature 1 (primary patient temperature). Current Mode continues. Audio alarm half-second tone every 2 minutes. Alarm/Alert indicator flashing.
Ranked causes
- 1. Patient temperature calculation error (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Prolonged cold exposure
- Purge cycle incomplete
- Reservoir level below minimum after pad purge
- System continuous rebooting
- System cooling with low patient temperature
- System unable to maintain stable water temperature
- System warming with elevated patient temperature
- Temperature probe 1 malfunction
- Unable to obtain stable patient temperature for Automatic control
- User interface disconnect
- User interface disconnect from monitor
- User settings not saved
- Water jets from Main Tank Vent in Purge Mode
- Water reservoir level below minimum
- Water reservoir level low
- Water spills from Main Tank Vent in Manual or Automatic modes
- Water temperature above safety level
- Water temperature below safety level
- Water temperature decreases during warming in Automatic Mode before reaching target
- Water temperature high
- Water temperature high at 42.5°C threshold
- Water temperature increases during cooling in Automatic Mode before reaching target
- Water temperature low
23 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
01 | Low fluid pressure in lines | Patient line is open to air, leaking, or has significant air in line | warning |
02 | Low water flow to pads | Flow to pads decreased >50% from maximum during case or since last low flow alert, or flow <0.3 L/min (300 ml/min) | warning |
03 | Water reservoir level low | Reservoir fluid level almost empty, only enough for one procedure remaining | warning |
04 | Water reservoir level below minimum | Reservoir empty or below minimum level | critical |
05 | Water reservoir level below minimum, fill or purge | Reservoir level is empty or below minimum to run system | critical |
07 | Purge not complete, pads may contain extra water | Significant amount of water still being returned from pads at end of purge cycle | warning |
08 | System warming and Patient Temperature 1 high | Patient Temperature 1 reading above 39.5°C (103.1°F) and water temperature above 39.5°C (103.1°F) and system unable to cool in Automatic Mode | critical |
09 | Patient Temperature 1 above user-defined limit | Patient Temperature 1 reading equal to or greater than user-set threshold | warning |
10 | System cooling and Patient Temperature 1 low | Patient Temp 1 below 31.0°C (87.8°F) and water temp below 31.0°C (87.8°F) in Automatic Mode, unable to heat | error |
11 | Patient Temperature 1 below user-defined limit | Patient Temperature 1 reading equal to or less than user-set threshold | warning |
12 | Patient and water temperature high in Manual Mode | Patient Temp 1 above 39.5°C (103.1°F) and water temp above 39.5°C (103.1°F), unable to cool | warning |
13 | Patient and water temperature low in Manual Mode | Patient Temp 1 below 31.0°C (87.8°F) and water temp below 31.0°C (87.8°F), unable to heat | warning |
14 | Patient temperature 1 probe disconnected or out of range | Temperature 1 reading outside lower display limits in auto mode | critical |
15 | Unable to obtain stable patient temperature for Automatic control | Patient temperature discontinuity; significant change in reading for >10 minutes | error |
16 | Temperature probe 1 malfunction. Replace temperature probe or cable | Patient Temperature 1 reading outside high limit of display range in Automatic Mode | error |
17 | Patient Temperature calibration error. Restart system. (Automatic Mode) | Internal temperature calibration from Patient Temperature 1 probe not functioning in Automatic Mode | error |
18 | Patient Temperature calibration error. Restart system. (Manual Mode Alert) | Internal temperature calibration from Patient Temperature 1 probe not functioning in Manual Mode | warning |
19 | Patient Temperature calibration error. Restart system. (Automatic Mode Alarm 2) | Internal temperature calibration from Patient Temperature 1 probe not functioning in Automatic Mode | error |
20 | Patient Temperature calibration error. Restart system. (Manual Mode Alert 2) | Internal temperature calibration from Patient Temperature 1 probe not functioning in Manual Mode | warning |
21 | Patient Temperature 2 high in Automatic Mode | Patient Temp 2 above 39.5°C (103.1°F) and water temp above 39.5°C (103.1°F), unable to cool | error |
22 | Patient Temperature 2 above user-defined limit | Patient Temperature 2 reading equal to or greater than user-set threshold | warning |
23 | Patient Temperature 2 low in Automatic Mode | Patient Temp 2 below 31.0°C (87.8°F) and water temp below 31.0°C (87.8°F), unable to heat | error |
24 | Patient Temperature 2 below user-defined limit | Patient Temperature 2 reading equal to or greater than user-set threshold (likely typo in source) | warning |
25 | System is warming and Patient Temperature 2 high in Manual Mode | Patient Temp 2 above 39.5°C (103.1°F) and water temp above 39.5°C (103.1°F), unable to cool | warning |
26 | System is cooling and Patient Temperature 2 low in Manual Mode | Patient Temp 2 below 31.0°C (87.8°F) and water temp below 31.0°C (87.8°F), unable to heat | warning |
27 | Patient Temperature 2 probe disconnected or out of range | Patient Temperature 2 reading outside lower limits of display range in Automatic Mode | error |
28 | Temperature probe 2 malfunction. Replace temperature probe or cable | Patient Temperature 2 reading outside upper limits of display range in Automatic Mode | error |
29 | Patient Temperature 2 calibration error. Restart system. (Automatic Mode Alert) | Internal temperature calibration from Patient Temperature 2 probe not functioning in Automatic Mode | warning |
30 | Patient Temperature 2 calibration error. Restart system. (Manual Mode Alert) | Internal temperature calibration from Patient Temperature 2 probe not functioning in Manual Mode | warning |
31 | Patient Temperature 2 calibration error. Restart system. (Automatic Mode Alert 2) | Internal temperature calibration from Patient Temperature 2 probe not functioning in Automatic Mode | warning |
32 | Patient temperature 2 calibration error. Restart system. (Manual Mode Alert 2) | Internal temperature calibration from Patient Temperature 2 probe not functioning in Manual Mode | warning |
33 | Water temperature high. Allow system to cool. (44.0°C threshold) | Water temperature above 44.0°C (111.2°F) | error |
34 | Water temperature high | Water temperature above 42.5°C (108.5°F) | unknown |
35 | Water temperature low. Allow system to warm. (3.5°C threshold) | Water temperature below 3.5°C (38.3°F) | error |
36 | Water temperature above safety level. Allow system to cool. | Water temperature above 44°C (111.2°F) | critical |
37 | Water temperature above safety level | Water temperature exceeds 43°C (109.4°F) | critical |
38 | Water temperature below safety level | Water temperature drops below 3.0°C (37.4°F) | critical |
40 | System unable to maintain stable water temperature | In Manual Mode, system unable to control water within 1.0°C of setpoint after 25 minutes in current mode or at time of last temperature change | unknown |
41 | Low bypass flow. Heating not available. | Insufficient internal flow during priming or bypass mode | unknown |
43 | User settings not saved. System default settings restored. | Saved customer user profile settings are invalid | unknown |
44 | Invalid system log entry. Service information may be corrupt. | In Diagnostic Mode, one or more entries in internal system event log are invalid | unknown |
45 | AC power lost | AC power was previously lost while power switch was in ON position | warning |
46 | User interface disconnect | User interface not detected by control system | error |
46 | Non-recoverable system error. User interface not detected. | User interface not detected by control system | unknown |
47 | Non-recoverable system error. User interface not detected. | User interface not detected by monitor system | unknown |
47 | User interface disconnect from monitor | User interface not detected by monitor system | error |
48 | Patient temperature output invalid | Patient Temperature Output (Echo) calibration data in non-volatile memory is invalid | unknown |
49 | Patient temperature out system error | Calculated Patient Temperature Out differs 0.1°C from measured Patient Temperature 1 | error |
50 | Patient temperature 1 erratic | Significant change in patient temperature reading for more than 8 minutes | warning |
51 | Patient temperature 1 below control range | Patient temperature less than 31°C when Automatic mode entered | warning |
52 | Extended period of cold water delivery | Circulating water temperature below 10°C/50°F for 8 of previous 10 hours | error |
53 | Prolonged cold exposure | Arctic Sun has issued 11 alerts for extended periods of cold water delivery | critical |
54 | Manual mode disabled | Manual key pressed while Manual Mode is disabled | error |
60 | Non-recoverable system error. Control and Monitor start-up synchronization fault. | Control and Monitor start up synchronization fault | unknown |
60-99 | Non-recoverable system error | System error that cannot be managed without Medivance support | critical |
61 | Non-recoverable system error. Control Processor parameter memory fault. | Control Processor parameter memory fault | unknown |
62 | Non-recoverable system error. Monitor Processor parameter memory fault. | Monitor Processor parameter memory fault | unknown |
64 | Non-recoverable system error. Unable to enable pump power (Control). | Unable to enable pump power (Control) | unknown |
65 | Non-recoverable system error. Unable to enable pump power (Monitor). | Unable to enable pump power (Monitor) | unknown |
66 | Non-recoverable system error. Unable to disable pump power (Control). | Unable to disable pump power (Control) | unknown |
67 | Non-recoverable system error. Unable to disable pump power (Monitor). | Unable to disable pump power (Monitor) | unknown |
70 | Non-recoverable system error. Mains signal is missing or inconsistent. | Mains signal is missing or inconsistent | unknown |
71 | Non-recoverable system error. T2 Thermistor signal out of range - high resistance. | T2 Thermistor signal out of range - high resistance | unknown |
72 | Non-recoverable system error. T2 Thermistor signal out of range - low resistance. | T2 Thermistor signal out of range - low resistance | unknown |
73 | Non-recoverable system error. T1 Thermistor signal out of range - high resistance. | T1 Thermistor signal out of range - high resistance | unknown |
74 | Non-recoverable system error. T1 Thermistor signal out of range - low resistance. | T1 Thermistor signal out of range - low resistance | unknown |
75 | Non-recoverable system error. T3 Thermistor signal out of range - high resistance. | T3 Thermistor signal out of range - high resistance | unknown |
76 | Non-recoverable system error. T3 Thermistor signal out of range - low resistance. | T3 Thermistor signal out of range - low resistance | unknown |
77 | Non-recoverable system error. T4 Thermistor signal out of range - high resistance. | T4 Thermistor signal out of range - high resistance | unknown |
78 | Non-recoverable system error. T4 Thermistor signal out of range - low resistance. | T4 Thermistor signal out of range - low resistance | unknown |
79 | Non-recoverable system error | T1 and T2 temperature sensors differ by more than 1.0 degree C | critical |
80 | Non-recoverable system error | Control Processor failed to detect simulated water temperature fault | critical |
81 | Non-recoverable system error | Monitor Processor failed to detect simulated water temperature fault | critical |
82 | Non-recoverable system error | A/D converter fault | critical |
83 | Non-recoverable system error | Monitor Processor does not communicate | critical |
84 | Non-recoverable system error | Control Processor does not communicate | critical |
85 | Non-recoverable system error | Control Module model could not be verified | critical |
86 | Non-recoverable system error | Power supply voltage out of range | critical |
98.x | Non-recoverable system error | Processor fault | critical |
99 | Non-recoverable system error | Processor Fault | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.