Masimo Rad-97: troubleshooting and error codes
The ServiceForge knowledge base holds 37 documented symptoms for the Masimo Rad-97, with 52 ranked causes, 53 diagnostic checks with test points and expected values, and 1 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Gas readings are questionable patient safety
Capnography gas measurement values appear unreliable or inaccurate
Ranked causes
- 1. Incorrect sampling line used (common)
- 2. Sampling line attached incorrectly (common)
- 3. Sampling line leaking (occasional)
- 4. Internal error (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
LEGI indicator dark (off)
LEGI indicator light is not illuminated
Ranked causes
- 1. Rad-97 with Capnography requires a power cycle (common)
- 2. Sampling line not fully plugged in (common)
- 3. Internal error (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pre-repair general procedures caution
General preparation steps for Rad-97 pulse oximeter disassembly and repair including damage inspection, marking documentation, and software version recording.
Ranked causes
- 1. Need to prevent equipment damage and ensure proper repair documentation (common)
- 2. Ensure software compatibility after board replacement (common)
- 3. Verify warranty coverage (occasional)
Verify precautions before disassembly
- Test point
- AC power cord and unit condition
- Expected
- AC power cord removed; unit cleaned per procedures; ESD precautions in place
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery does not charge
Rad-97 connected to AC power but battery does not accept or maintain charge
Ranked causes
- 1. AC power cable disconnected (common)
- 2. Internal charging circuit failure (rare)
Verify AC power cable is fully inserted in AC outlet and Rad-97
- Test point
- AC power cable connectors at wall outlet and Rad-97 AC inlet
- Expected
- Cable inserted firmly with no play
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP Module Pressure Accuracy
NIBP module pressure readings compared to calibrated manometer standard across 0-280 mmHg range; must be within ±3 mmHg tolerance per international standards.
Ranked causes
- 1. NIBP module requires zero point calibration (occasional)
- 2. NIBP module requires span point calibration (occasional)
Connect calibration apparatus and set up manometer
- Test point
- Manometer, 500 mL bottle or cuff, hand bulb connected to NIBP module
- Expected
- All connections secure via T adapters and tubing
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Capnography values not displayed on host monitor
Capnography measurements do not appear on connected host device display
Ranked causes
- 1. Rad-97 with Capnography requires a power cycle (common)
- 2. Host device incorrectly configured or incompatible (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
LEGI indicator steady red caution
LEGI indicator is illuminated with steady red color
Ranked causes
- 1. Rad-97 with Capnography requires a power cycle (common)
- 2. Internal error (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Main screen is dim or blank
Rad-97 powers on but display is not visible; screen shows minimal or no content
Ranked causes
- 1. Brightness setting too low (common)
- 2. Internal display failure (rare)
Locate brightness setting in menu and increase value
- Test point
- Brightness control in system settings menu
- Expected
- Display becomes brighter and readable
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nurse Call connection not triggering or showing incorrect resistance patient safety
Nurse Call interconnection cable shows open circuit or incorrect resistance values when alarm triggered or device in idle state
Ranked causes
- 1. Nurse Call cable misconfiguration or connection fault (occasional)
- 2. Defective Nurse Call port or internal signal routing failure (rare)
Set Nurse Call polarity to Normal
- Test point
- Device Output settings Nurse Call Polarity
- Expected
- Polarity set to Normal
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software and battery fuel gauge mismatch after battery replacement
Battery replacement requires coordinated software and fuel gauge firmware updates; incorrect sequencing may cause improper battery performance.
Ranked causes
- 1. Device software update performed after battery replacement instead of before (occasional)
- 2. Battery fuel gauge firmware below minimum version V1.0.0.0 (occasional)
Perform device software upgrade
- Test point
- Software version in Main Menu > About screen
- Expected
- Version matches yellow 4600 mAh battery kit version (minimum recommended)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on
Rad-97 does not respond to power button press; no display, no LED indicators
Ranked causes
- 1. Power button not pressed long enough (common)
- 2. Battery depleted (common)
Press and hold Power Button
- Test point
- Power Button on unit
- Expected
- System powers on after 2 seconds
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm limit parameter not updating on display patient safety
Changed High SpO2 Alarm parameter does not appear on display screen or shows incorrect value
Ranked causes
- 1. Display software or parameter memory issue (rare)
Connect sensor and obtain SpO2 reading
- Test point
- Sensor on finger; Display SpO2 value
- Expected
- Valid SpO2 value displayed on screen
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery charge indicator does not illuminate
Battery charge indicator on home button of Rad-97 front panel does not light up after connecting power cord to power entry module on rear of device.
Ranked causes
- 1. Battery not properly connected (common)
Verify battery charge indicator illuminates on home button
- Test point
- Home button on front panel
- Expected
- Battery charge indicator illuminates
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery icon shows red X on Main Screen
Battery icon displayed on Main Screen shows a red X through it, indicating battery connection issue.
Ranked causes
- 1. Battery not properly connected (common)
Verify battery icon on Main Screen does not show red X
- Test point
- Main Screen battery icon
- Expected
- Battery icon displays without red X, indicates charging
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery not holding sufficient charge
Run time is below specification
Ranked causes
- 1. Battery aged or degraded below Full Charge Capacity (used) limit (common)
Read Full Charge Capacity from Battery Diagnostics screen
- Test point
- Battery Diagnostics menu
- Expected
- At or above Full Charge Capacity (used) limit
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery not identified after replacement
Battery charge indicator on home button does not illuminate, or battery icon shows red X on main screen
Ranked causes
- 1. Battery cable not properly connected (common)
Verify battery charge indicator on home button illuminates
- Test point
- Home button on front panel
- Expected
- Battery charge indicator illuminates
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery replacement requires software and firmware updates
Device software and battery fuel gauge firmware must be updated when battery is replaced to ensure proper operation
Ranked causes
- 1. Battery replacement with yellow 4600 mAh battery (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery replacement requires specific firmware versions for 5400 mAh battery
Device software and battery fuel gauge firmware minimum versions must be met when replacing with blue 5400 mAh battery
Ranked causes
- 1. Blue 5400 mAh battery replacement (occasional)
Verify device software version meets minimum V1.6.4.0 for 5400 mAh battery
- Test point
- Main Screen > Main Menu > About screen > Processor field
- Expected
- V1.6.4.0 or higher
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery replacement software requirements unknown
Uncertainty about which software and firmware versions are required when replacing Rad-97 battery
Ranked causes
- 1. Unclear minimum firmware versions for battery replacement (unknown)
Confirm battery replacement kit contents
- Test point
- Battery replacement kit
- Expected
- Both device software upgrade USB drive and battery fuel gauge firmware USB drive included
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery replacement software update needed
After battery replacement, device software and battery fuel gauge firmware must be updated in correct sequence
Ranked causes
- 1. Battery replaced requiring firmware synchronization (unknown)
Verify minimum software version for yellow 4600 mAh battery
- Test point
- Main Menu, About screen, Processor field
- Expected
- Device software and battery fuel gauge firmware V1.0.0.0 or higher
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP Air Leak Test
Pneumatic system air leakage measured over test period; must not exceed 6 mmHg/minute per international standards.
Ranked causes
- 1. Pneumatic system leakage exceeds tolerance (rare)
Connect calibration apparatus and power on manometer
- Test point
- Manometer, 500 mL bottle, hand bulb connected to NIBP module
- Expected
- Manometer powered on, set to mmHg, calibrated within 12 months
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP Overpressure Safety Test (Adults/Pediatrics) patient safety
Overpressure relief valve function verified; pressure relief valves must open within specified pressure ranges (300±10 mmHg for adults/pediatrics, 150±5 mmHg for neonates).
Ranked causes
- 1. Overpressure relief valves open beyond tolerance (rare)
Connect apparatus; verify appropriate cuff size for patient population
- Test point
- Manometer, 500 mL bottle or cuff, hand bulb; cuff type
- Expected
- Adult or Pediatric cuff confirmed; do not use Neonate cuff for adults/pediatrics
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Charging indicator shows red
Rad-97 charging indicator LED illuminates red instead of expected green/amber; indicates charge fault condition
Ranked causes
- 1. Internal charging circuit fault (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Continuous speaker tone sounds caution
Continuous audible alarm tone emits from unit; cannot silence by normal alarm dismiss
Ranked causes
- 1. Internal audio controller malfunction (rare)
Press and hold Power Button for 8 seconds to force shutdown
- Test point
- Power Button on unit
- Expected
- Alarm silences and unit powers off
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrical safety test required after disassembly high voltage
Mandatory electrical safety testing required following any disassembly of Rad-97 unit.
Ranked causes
- 1. Verify electrical safety after service (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ESD protection procedures caution
Precautions required to minimize electrostatic discharge damage during service including grounding and anti-static equipment.
Ranked causes
- 1. Prevent ESD damage to sensitive electronic components (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Home button does not work
Home button on control panel does not respond when pressed; no navigation occurs
Ranked causes
- 1. Home button switch failure (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
LEGI indicator blinking red caution
LEGI indicator is blinking with red color
Ranked causes
- 1. Sampling line obstructed (occlusion) (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low or degraded battery run time
Battery runtime significantly reduced from normal operation
Ranked causes
- 1. Battery memory effect from incomplete charge/discharge cycles (occasional)
Perform complete discharge and full recharge cycle
- Test point
- Battery charging system
- Expected
- Battery runtime restored to normal levels after full cycle
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nurse Call does not communicate patient safety
Nurse Call feature does not establish connection to hospital communication system; no signal transmitted
Ranked causes
- 1. Nurse Call connector loose or disconnected (common)
Visually inspect Nurse Call connector at Rad-97 and wall outlet
- Test point
- Nurse Call connector at both ends
- Expected
- Connector fully seated and locked
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software upgrade fails or device corrupted
Device power is interrupted during software upgrade process, potentially causing device software corruption and malfunction
Ranked causes
- 1. Power interrupted during software upgrade (rare)
Verify software version after upgrade
- Test point
- Main Screen > Main Menu > About screen > Processor field
- Expected
- Software version matches installed version
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software upgrade fails or device corrupts
Device power interrupted during software upgrade process, risking firmware corruption and device malfunction
Ranked causes
- 1. Power loss or device power-off during upgrade (rare)
Verify Rad-97 is powered and plugged in
- Test point
- Power status
- Expected
- Device powered on or ready to power on
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software upgrade interruption caution
Device powers off unexpectedly during software update process, risking firmware corruption.
Ranked causes
- 1. Power loss or manual shutdown during upgrade (occasional)
Verify Rad-97 has power before starting upgrade
- Test point
- Power connection
- Expected
- Device plugged in and functional
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software upgrade needed caution
Device requires software update via USB drive for proper operation or after battery replacement
Ranked causes
- 1. Outdated or incompatible software version (unknown)
Verify current software version
- Test point
- Main Screen, Main Menu, About screen, Processor field
- Expected
- Software version matches intended upgrade version
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Speaker does not produce sound
Speaker fails to respond to volume adjustments; no audio output during Speaker Test
Ranked causes
- 1. Speaker defective or disconnected (unknown)
Perform Speaker Test with volume adjustments
- Test point
- Sounds settings menu
- Expected
- Speaker responds and sounds in relationship to volume adjustment
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Speaker not producing audio during speaker test
Speaker fails to respond or produce sound when alarm volume or pulse tone volume adjusted
Ranked causes
- 1. Speaker defect or internal connection failure (unknown)
Access Sounds settings with Rad-97 powered on
- Test point
- Sounds settings menu
- Expected
- Settings menu accessible
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch screen is not responsive
Touch functionality does not respond to finger input; no commands execute from screen touches
Ranked causes
- 1. Internal touch controller failure (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Red X battery icon | Battery not properly connected or identified | Power-on after battery replacement when cable is unseated | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI