Masimo Rad-5: troubleshooting and error codes
The ServiceForge knowledge base holds 29 documented symptoms for the Masimo Rad-5, with 39 ranked causes, 15 diagnostic checks with test points and expected values, and 7 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
No Sen or Sen Off displayed patient safety
Display shows No Sen or Sen Off message indicating sensor detection issue
Ranked causes
- 1. Defective sensor (common)
- 2. Faulty patient cable (occasional)
- 3. Oximeter board failure (occasional)
Verify cable is connected with sensor attached
- Test point
- sensor connector
- Expected
- cable connected with sensor attached
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No sensor displayed patient safety
No Sen or Sen Off message displayed on unit
Ranked causes
- 1. Sensor fault (common)
- 2. Patient cable fault (common)
- 3. Oximeter board fault (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No sensor message displayed
Display shows 'No Sen' or 'Sen Off' message indicating sensor not detected
Ranked causes
- 1. Defective sensor (common)
- 2. Defective patient cable (common)
- 3. Oximeter board failure (occasional)
Verify cable is connected with sensor attached
- Test point
- Sensor cable connection
- Expected
- Cable connected with sensor attached
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm condition but no audible alarm patient safety
Alarm condition detected but no audible sound
Ranked causes
- 1. Speaker fault (common)
- 2. System board fault (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display missing segments
LCD or LED display is missing or not showing segments properly
Ranked causes
- 1. Display failure (common)
- 2. System Board malfunction (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rad-5 will not turn off caution
Unit cannot be powered off using normal power control
Ranked causes
- 1. System Board malfunction (common)
- 2. Keypad malfunction (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not turn off
Rad-5 cannot be powered off
Ranked causes
- 1. System Board failure (common)
- 2. Keypad failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
5V Voltage Supply Out of Range
System error detected with audible alarm and error code displayed. 5V supply voltage outside acceptable range.
Ranked causes
- 1. 5V power supply voltage out of specification (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm mute cannot be disabled patient safety
Pressing Alarm button once does not disable alarm mute or no audible alarm is heard
Ranked causes
- 1. Alarm button function or mute disable logic malfunction (unknown)
Alarm mute disabled by pressing Alarm button
- Test point
- Alarm button and speaker
- Expected
- audible alarm heard after single Alarm button press
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm mute not functioning
Double-pressing Alarm Silence button does not mute alarm or yellow LED does not stay on steadily
Ranked causes
- 1. Alarm mute button logic or LED control malfunction (unknown)
Alarm mute activates on double button press
- Test point
- Alarm Silence button and yellow LED
- Expected
- momentary beep heard with yellow LED on and steady (not flashing)
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm not functioning patient safety
No audible alarm or SpO2 display is not flashing when alarm should trigger
Ranked causes
- 1. Alarm circuit or speaker malfunction (unknown)
Verify audible alarm occurs and SpO2 flashes when alarm threshold reached
- Test point
- Speaker output and display
- Expected
- audible alarm with flashing SpO2 display
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm Silence button not responding
Alarm does not silence when Alarm Silence button is pressed, or yellow LED does not flash
Ranked causes
- 1. Alarm Silence button or LED circuit malfunction (unknown)
Alarm silences and yellow LED flashes on single button press
- Test point
- Alarm Silence button and yellow LED
- Expected
- alarm silenced with yellow LED flashing
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm silence timeout not working patient safety
Audible alarm does not resume after 120 seconds when Alarm Silence is active
Ranked causes
- 1. Alarm timeout timer malfunction (unknown)
Alarm silence timeout at 120 seconds
- Test point
- Alarm output and yellow LED
- Expected
- audible alarm reactivates and yellow LED turns off after 120 seconds
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery Voltage Out of Range
System error detected with audible alarm and error code displayed. Battery voltage outside acceptable range.
Ranked causes
- 1. Battery voltage out of specification (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blue Voltage Supply Out of Range
System error detected with audible alarm and error code displayed. Blue wavelength LED supply voltage outside acceptable range.
Ranked causes
- 1. Blue wavelength supply voltage out of specification (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
MS-X Board Failure
System error detected, main software triggers audible alarm and displays error code on handheld display. MS-X board held in reset, no runtime functionality available until power cycled.
Ranked causes
- 1. MS-X Board malfunction (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No numeric display or Signal IQ bar pulsing
Display does not show numeric readings or Signal IQ bar is not pulsing after Masimo Set Tester connection
Ranked causes
- 1. Device not initialized or tester not properly connected (common)
Verify display shows numeric readings and Signal IQ bar pulsing
- Test point
- Display output
- Expected
- numeric readings visible and Signal IQ bar pulsing within 20 seconds
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No sensor detected or sensor off displayed
Display shows 'No Sen' or 'Sen Off' message indicating sensor detection failure
Ranked causes
- 1. Sensor failure or disconnection (common)
Verify cable is connected with sensor attached
- Test point
- Sensor cable connection
- Expected
- Cable connected securely with sensor attached
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No Sensor or Sensor Off displayed patient safety
Display shows "No Sen" or "Sen Off" message indicating sensor disconnection or failure
Ranked causes
- 1. Sensor fault or disconnection (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pulse rate measurement out of range
Pulse rate reading is not between 55 bpm and 65 bpm when Masimo Set Tester is connected
Ranked causes
- 1. Device malfunction or incorrect tester connection (unknown)
Verify pulse rate measurement within specification
- Test point
- Display numeric reading
- Expected
- between 55 bpm and 65 bpm, +10 bpm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pulse volume control not responding
Up and down arrows do not change pulse beep volume or volume cannot be silenced
Ranked causes
- 1. Volume control buttons or audio circuit malfunction (unknown)
Up arrow increases pulse beep volume
- Test point
- Up arrow button and speaker output
- Expected
- audible increase in pulse beep volume
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Red Voltage Supply Out of Range
System error detected with audible alarm and error code displayed. Red wavelength LED supply voltage outside acceptable range.
Ranked causes
- 1. Red wavelength supply voltage out of specification (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Speaker alarming loudly
Speaker producing very loud alarm, unit not functioning or displaying
Ranked causes
- 1. Watchdog error (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Speaker alarming loudly and unit not functioning
Speaker produces very loud alarm, unit does not function or display information
Ranked causes
- 1. Watchdog error (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Speaker alarming loudly and unit not functioning or displaying patient safety
Speaker emitting very loud alarm; unit does not function or display information
Ranked causes
- 1. Watchdog error (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Speaker alarming very loud and unit not functioning caution
Speaker outputs alarm at high volume; unit does not function or display information
Ranked causes
- 1. Watchdog error (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Speaker alarming very loud and unit not functioning or displaying
Speaker produces very loud alarm; unit does not function or display any information
Ranked causes
- 1. Watchdog error (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Speaker alarming very loudly and unit not functioning or displaying
Speaker produces loud alarm while device is non-responsive and display is blank
Ranked causes
- 1. Watchdog error (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 measurement out of range
SpO2 reading is not between 79% and 84% when measured with Masimo Set Tester
Ranked causes
- 1. Sensor or measurement circuit malfunction (unknown)
Verify SpO2 measurement within specification
- Test point
- Display SpO2 value
- Expected
- between 79% and 84%, +5%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
1 | MS-X Board Failure | System error detection during operation | error |
2 | MS-X Board Communication Failure | System error detection during operation | error |
3 | 5V Voltage Supply Out of Range | System error detection during operation | error |
4 | Red Voltage Supply Out of Range | System error detection during operation | error |
5 | Blue Voltage Supply Out of Range | System error detection during operation | error |
6 | Battery Drop Voltage Out of Range | System error detection during operation | warning |
7 | Battery Voltage Out of Range | System error detection during operation | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI