Lumisys LUMISCAN LSDT: troubleshooting and error codes
The ServiceForge knowledge base holds 15 documented symptoms for the Lumisys LUMISCAN LSDT, with 19 ranked causes, 47 diagnostic checks with test points and expected values, and 6 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Calibration: DACQ offset adjustment
Adjust OFFSET potentiometer to calibrate minVal to specification using DDT display feedback
Ranked causes
- 1. A/D offset reading out of specification (common)
- 2. Resistor misadjustment prevents specification attainment (occasional)
Verify A/D display values via DDT
- Test point
- Computer monitor display from DDT option 10
- Expected
- minVal @ pixel 512 = 30, maxVal = 3675 (actual values may differ) decimal, ±2
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: PMT output and high voltage setup high voltage
Set PMT input high voltage and verify PMT high voltage is within safe operating limits
Ranked causes
- 1. PMT output voltage requires adjustment (common)
- 2. PMT high voltage exceeds safe limit (rare)
Verify beam stationary at horizontal center
- Test point
- Galvo slot, horizontal center
- Expected
- Within +0.50 inches of center
- Tools
- Visual inspection
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: PMT Preamp offset high voltage
Adjust PMT Preamp offset voltage to specification for proper photomultiplier tube signal conditioning
Ranked causes
- 1. PMT Preamp offset voltage out of specification (common)
- 2. PMT high voltage connector disconnection required (common)
Measure PMT Preamp offset voltage
- Test point
- TP1 on PMT Preamp PCA
- Expected
- Per specification table on page 5-9, ±0.1 mV
- Tools
- DVM on mV range
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Reference Amplifier adjustment caution
Adjust Reference Amplifier to compensate for laser output power variations
Ranked causes
- 1. Reference Amplifier voltage requires adjustment (common)
- 2. Laser warm-up period not met (occasional)
Measure REF test point voltage
- Test point
- DACQ PCA REF test point
- Expected
- Record minimum and maximum voltage levels
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Correction LUT generation
Generate custom correction lookup table to map indicated optical densities to actual densities after DACQ PCA replacement or adjustment
Ranked causes
- 1. Correction LUT missing or needs refresh (common)
Verify conditions for CLUT generation
- Test point
- System configuration and history
- Expected
- DACQ PCA replaced OR SLOPE/R103/R105 adjusted OR existing CLUT lost/deleted OR Density Verification failed
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Density accuracy verification
Verify measured optical densities match actual densities using step wedge film and batch verification file
Ranked causes
- 1. Correction Lookup Table needs verification or generation (common)
Verify Slope specification
- Test point
- results.dat file for LS50/75
- Expected
- Min 0.960 Max 1.040 to 1.040
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Folding mirror beam centering and parallelism radiation
Post-Galvo alignment of final beam position in film guide; ensures horizontal centering and parallel beam path
Ranked causes
- 1. Galvo scanner or folding mirror repositioned; beam not centered in film guide (common)
Verify horizontal beam centering in film guide
- Test point
- film guide opening in 15-inch collection cylinder
- Expected
- horizontally centered within ±0.5 inches
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Galvo alignment
Final alignment of Galvo system to position laser spot within slot parameters
Ranked causes
- 1. Galvo offset and span require adjustment (common)
Verify CAL+, CAL-, and Sum of Deltas values
- Test point
- DACQ board measurements
- Expected
- Meet specification
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Galvo PCA resistor reset radiation
Emergency reset of R47 and R45 potentiometers to predetermined starting position when severely out of adjustment
Ranked causes
- 1. Galvo PCA extremely out of adjustment; downward slope cannot be straightened (rare)
Set R47 resistance per model and frequency
- Test point
- Top of R42 to bottom pin of any test point
- Expected
- 115 Hz LS50/75: 2.86K Ω; LS85LF: 3.68K Ω; LS85: 3.68K Ω; 75 Hz LS50/75: 12 CCW turns from fully CW, per model
- Tools
- Ohmmeter
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: LS20 Correction LUT Generation
Generate custom correction lookup table (CLUT) mapping optical densities after DACQ PCA replacement, slope adjustments, or when existing CLUT is lost or density verification fails.
Ranked causes
- 1. DACQ PCA replaced or slope/offset adjustments made since last CLUT generation (common)
Verify Slope specification from MKCLUT results
- Test point
- Reported MKCLUT output Slope value
- Expected
- Min 0.985 Max 1.015, 0.985-1.015
- Tools
- Command line output
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: LS20 DACQ PCA Final Alignment
Verify and adjust LED output voltage and A/D offset for proper CCD signal acquisition after DACQ PCA, LED PCA, or CCD PCA replacement or failed density test.
Ranked causes
- 1. DACQ PCA or LED PCA or CCD PCA replaced, or failed density verification (common)
Verify LED output peak voltage range
- Test point
- TP9 of DACQ PCA (with trigger on TP8 SOS)
- Expected
- greater than 1.6V and less than 3.0V, 1.6V to 3.0V
- Tools
- Oscilloscope
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: LS20 Density Accuracy Verification
Compare measured optical densities to actual densities using step wedge film and reference file to verify digitizer calibration meets specifications.
Ranked causes
- 1. Data Acquisition PCA replaced or calibration verification required (common)
Verify Slope specification
- Test point
- results.dat file Slope value
- Expected
- Min 0.960 Max 1.040, 0.960-1.040
- Tools
- Text editor or command line
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: LS20 Setting the Dynamic Range
Adjust DACQ PCA and LED PCA controls to set proper signal levels and dynamic range when density verification test fails.
Ranked causes
- 1. Failed density accuracy verification test requires dynamic range adjustment (common)
Verify LED peak voltage
- Test point
- TP9 of DACQ PCA with trigger on TP8 SOS
- Expected
- 2.8V minimum, above 1.6V acceptable, ≥1.6V, target 2.8V
- Tools
- Oscilloscope
6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Optical alignment radiation
Laser beam path adjustment after component replacement in optical system; beam must be centered through all optical elements to PMT subsystem
Ranked causes
- 1. Optical component replaced or misaligned (common)
Verify laser beam position on corner mirror
- Test point
- corner mirror
- Expected
- beam centered
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Scan linearity adjustment radiation
Galvo motor sweep calibration to ensure linear pixel sampling across 14-bar test film; adjusts beam movement rate and position
Ranked causes
- 1. DACQ PCA or Galvo PCA replaced; beam linearity not verified (common)
Verify laser spot within center tolerance
- Test point
- Collection Chamber center
- Expected
- within +0.50 inch from center mark
11 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
10 beeps | DCB not present | LSDT driver loads with error | critical |
15 beeps | DCB memory error | LSDT driver loads with error | critical |
1 beep | Driver Previously Loaded | LSDT driver loads with error | warning |
25 beeps | LUT failure | LSDT driver loads with error | critical |
5 beeps | CLUT or CALCR file failure | LSDT driver loads with error | error |
Data Lost Error | Image data was not read by host computer in timely fashion during scan operation | Host computer fails to read buffered image data before circular buffer fills | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI