Lumenis VersaPulse Select II: troubleshooting and error codes

The ServiceForge knowledge base holds 45 documented symptoms for the Lumenis VersaPulse Select II, with 99 ranked causes, 113 diagnostic checks with test points and expected values, and 138 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

System overheating fault displayed caution

Overheating fault message appears on LCD screen during operation

Ranked causes

  1. 1. Coolant level low or improper (common)
  2. 2. Fan speed controller improperly adjusted (common)
  3. 3. Water particle filter dirty or discolored (common)
  4. 4. Air filter restricted or dirty (common)
  5. 5. Inadequate airflow around system (occasional)
  6. 6. Ambient temperature above 22.5°C (occasional)
  7. 7. Water temperature calibration incorrect (occasional)
  8. 8. Water hoses bent, kinked, or damaged (occasional)
  9. 9. Lamp calibration values too high (occasional)
  10. 10. Water pump head worn or flow restriction present (rare)
First check

Verify coolant level in upper water reservoir

Test point
Upper water reservoir
Expected
Between one-half and full

5 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System blowing blastshields patient safety

Frequent replacement of blastshield optic during operation; fiber endface explodes during use

Ranked causes

  1. 1. Blastshield optic contaminated or damaged (common)
  2. 2. Fiber optical alignment incorrect (common)
  3. 3. Fiber previously used, autoclaved, or worn beyond lifetime (common)
  4. 4. Fiber SMA end damaged or contaminated (common)
  5. 5. Blastshield glass fingerprint or prior contamination (occasional)
  6. 6. Laser optical alignment not centered or peaked (occasional)
  7. 7. Fiber focus lens contaminated or damaged (occasional)
First check

Visually inspect blastshield optic

Test point
Blastshield optic surface
Expected
Clean, no contamination or visible damage

5 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System does not start or shuts off during operation high voltage

Low voltage power supply failure or loss of coolant flow prevents system operation; may display OVERHEATING on LCD

Ranked causes

  1. 1. Blown F1 or F2 line fuse (common)
  2. 2. Blown F3 fuse (common)
  3. 3. Open T1 thermal switch (occasional)
  4. 4. Front door opened (interlock switch) (common)
  5. 5. Loss of coolant flow (SW3) (occasional)
  6. 6. Loss of facility power (common)
First check

Verify facility 220 VAC supply voltage present

Test point
T1 secondary
Expected
220 VAC

4 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Access interior components high voltage

Technician needs to access laser head, optics bench, cooling system, fuses, HVPS, and circuit breaker for service and maintenance

Ranked causes

  1. 1. Service access required to front door and internal PCBs (common)
  2. 2. Service access required to laser head and top cover (common)
  3. 3. Service access required to optics bench (occasional)
  4. 4. Service access required to right side cover (common)
  5. 5. Service access required to left side cover (common)
First check

Verify front door opens properly

Test point
Bottom left hole on front cover
Expected
Door swings open to right when tab released

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fiber endface explosion radiation

Fiber endface explodes during operation, causing system to blow blastshields; violent reaction ejects sparks and hot gases from fiber

Ranked causes

  1. 1. Fiber previously used or autoclaved, lifetime exceeded (common)
  2. 2. Fiber SMA end damaged or contaminated (common)
  3. 3. Blastshield contamination scattering laser energy (common)
  4. 4. Laser optical alignment off-center or not peaked (occasional)
  5. 5. Fiber focus lens contamination scattering laser energy (occasional)
First check

Verify fiber service history and condition

Test point
Fiber age and sterilization records
Expected
New fiber; not previously used or autoclaved

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System shutdown after turn-on high voltage

System de-energizes main contactor and shuts down after successfully turning on

Ranked causes

  1. 1. Interlock loop break (no circuit breaker trip) (common)
  2. 2. Thermal switch in isolation transformer open (occasional)
  3. 3. Waterflow switch S3 open or defective (occasional)
  4. 4. Fuse F3 open (occasional)
  5. 5. Circuit breaker tripping during shutdown (occasional)
First check

Consult Interlock schematic to identify break in 24 VAC contactor loop

Test point
Interlock schematic (Section 8)
Expected
Open point identified in thermal switch, waterflow switch, door interlock, K2, K1, or F3

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cavity alignment verification and optimization high voltage

Determine if Ho/Nd:YAG cavity optics (HR, OC mirrors) are properly aligned for maximum power output by visual inspection and burn pattern analysis on paper target.

Ranked causes

  1. 1. HR or OC mirror misalignment in X-axis (unknown)
  2. 2. HR or OC mirror misalignment in both X and Y axes (unknown)
  3. 3. Damaged HR, OC, or Relay mirrors (rare)
  4. 4. Damaged rod or failed flash lamp (rare)
First check

Visual inspection of optics for damage

Test point
HR, OC, and Relay mirrors
Expected
No visual sign of damage

6 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Service attenuator not inserted into beam path high voltage

Attenuator fails to move into position when SW1 is activated; 'ATTEN IN BEAM' message does not appear on LCD display

Ranked causes

  1. 1. SW1 switch on Shutter PCB not making contact or open (common)
  2. 2. 24 VDC supply failure to solenoid (common)
  3. 3. Solenoid coil malfunction or mechanical jam (occasional)
  4. 4. Slotted optical switch LPT1 blocked or misaligned (occasional)
First check

Verify SW1 switch contact on Shutter PCB

Test point
SW1 on Shutter PCB
Expected
Switch closes when activated, providing ground return path

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Coolant overtemperature detected caution

"OVERHEATING" message displayed on LCD color display when coolant temperature exceeds setpoint; system is disabled until temperature drops to reset point.

Ranked causes

  1. 1. Coolant temperature exceeds setpoint (Ho:YAG & Ho/Nd:YAG = 45.5°C) (common)
  2. 2. Heat exchanger fan not running or blocked air flow (occasional)
  3. 3. Coolant contamination or insufficient deionization (occasional)
First check

Monitor coolant temperature reading on control electronics

Test point
Coolant temperature sensor input to CPU PCB
Expected
Below 45.5°C (Ho:YAG & Ho/Nd:YAG models)

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Intermittent laser energy and power supply over temperature faults high voltage

Intermittent Laser Energy messages and Power Supply Over Temperature (Crowbar) faults occurring during operation

Ranked causes

  1. 1. Fiber optic cable improperly crimped causing fiber pull-out (common)
  2. 2. Fiber optic cable rough polished causing scatter and transmission losses (common)
  3. 3. High Voltage Power Supply controller PCB design issue requiring modifications (occasional)
First check

Inspect fiber optic cable connectors at HVPS and Controller PCB

Test point
both ends of fiber optic cable connectors
Expected
Connectors properly seated, fibers fully inserted without gaps

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System fault code reported patient safety

Safety processor detects fault and displays two-digit hexadecimal code on CPU PCB LED display and touch screen

Ranked causes

  1. 1. Informational fault condition (unknown)
  2. 2. Clearable fault condition (unknown)
  3. 3. Permanent fault condition (unknown)
First check

Understand symptom details before troubleshooting

Test point
Observation
Expected
Determine if fault appears during self-test or operation; occurrence pattern (firing only, certain energy/pulse settings); repeatability

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System turn-on failure high voltage

System fails to turn on and stay on when key switch is activated

Ranked causes

  1. 1. No power to unit or circuit breaker issue (common)
  2. 2. Main contactor not energizing (common)
  3. 3. Insufficient coolant flow (occasional)
First check

Verify mains power present at unit

Test point
Circuit breaker input
Expected
Power present

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Aiming beam adjustment high voltage

Aiming beam outputs inconsistent spot pattern or reduced power; appears doughnut-shaped rather than full bright red spot

Ranked causes

  1. 1. Aiming diode mount misaligned (common)
  2. 2. Folding mirror misaligned (common)
First check

Observe aiming beam output from fiber

Test point
Fiber output
Expected
Full spot of bright red light, no doughnut pattern

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Service Calibration (Nd:YAG laser gain adjustment) high voltage

Adjust reflectivity factors and optical transmission to match indicated power on display with external power meter reading for Nd:YAG laser output

Ranked causes

  1. 1. Reflectivity factors and transmission settings require calibration (common)
  2. 2. Nd:YAG software gain adjustment needed for dual wavelength systems (occasional)
First check

Verify all lasers enabled on service screen #1

Test point
Service screen #1
Expected
All lasers turned on

8 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Coolant system contamination or leakage caution

Visible leakage from fittings, hoses, or heat exchanger; coolant conductivity elevated; system performance degraded.

Ranked causes

  1. 1. Hose or fitting deterioration; seal degradation (occasional)
  2. 2. Contamination from external sources or corrosion deposits (occasional)
First check

Visually inspect coolant system for leaks

Test point
All hose connections, heat exchanger, pump inlet/outlet, drain hose
Expected
No visible moisture or coolant dripping; dry connections

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Integrator gain calibration required high voltage

Service screen power readings do not match external power meter measurements for ENERGY I or ENERGY III detectors

Ranked causes

  1. 1. ENERGY I integrator gain (R1) out of calibration (unknown)
  2. 2. ENERGY III integrator gain (R3) out of calibration (unknown)
First check

Fire laser and verify delivered power

Test point
External power meter
Expected
6.0 Watts
Tools
External power meter

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low coolant or coolant conductivity error (F6, F52) caution

Coolant conductivity out of range or coolant temperature exceeds limit

Ranked causes

  1. 1. Coolant conductivity out of specification (occasional)
  2. 2. Coolant overtemperature condition (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

OVERHEATING displayed on LCD caution

Coolant temperature exceeds set point threshold; system disabled until temperature falls to reset point

Ranked causes

  1. 1. Coolant temperature at or above set point (45.5°C for Ho:YAG or Ho/Nd:YAG) (common)
  2. 2. Ineffective heat exchanger or fan operation (occasional)
First check

Measure coolant temperature

Test point
Temperature sensor
Expected
Below 45.5°C (Ho:YAG/Ho/Nd:YAG)

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Service attenuator positioned but 'ATTEN IN BEAM' message not displayed

Attenuator physically enters beam path but LCD display does not show 'ATTEN IN BEAM' message

Ranked causes

  1. 1. Slotted optical switch LPT1 not signaling position correctly to Controller PCB (common)
  2. 2. Controller PCB digital input (SERATNO) not reading optical switch state (occasional)
First check

Measure SERATNO signal voltage at Shutter PCB with attenuator in beam path

Test point
SER ATEN NO output on Shutter PCB
Expected
Signal grounded (0 V) when attenuator is positioned in beam path
Tools
Digital multimeter

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unable to enter service mode

System does not enter service mode when attempting to activate it via Controller PCB SW2 switch

Ranked causes

  1. 1. SW2 switch left ON during power-up (no OFF to ON transition) (common)
  2. 2. Microprocessors halted due to system fault (occasional)
First check

Verify SW2 switch position and state during power-up

Test point
Controller PCB SW2 (left of main microprocessor)
Expected
Switch transitions from OFF to ON during system startup

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

ADC busy timeout (F1) patient safety

ADC (analog-to-digital converter) remains in busy state longer than expected

Ranked causes

  1. 1. ADC hardware malfunction or timing error (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Channel relay mirror peak adjustment high voltage

Procedure to adjust relay mirrors to maximize laser pulse amplitude on oscilloscope for each channel

Ranked causes

  1. 1. Relay mirror position requires adjustment to optimize beam coupling (unknown)
First check

Verify second wedge optic beam clearance with test aperture

Test point
Second wedge optic
Expected
Beam unclipped

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Energy monitor and auto laser calibration radiation

Adjusts energy monitor amplifier gains and optical transmission characteristics; establishes Volts/Joule conversion factor referencing delivered Holmium laser energy measured on calibrated external power meter.

Ranked causes

  1. 1. Energy monitor requires calibration adjustment (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Energy Monitor and Pick-Off Gain high voltage

Adjustment procedure for Energy Monitor Calibration and Pick-Off Gain on Controller PCB, Single & Dual Solenoid PCBs following hardware changes

Ranked causes

  1. 1. Hardware changes to Controller PCB, Single & Dual Solenoid PCBs (common)
First check

Set laser operating parameters

Test point
Laser control panel
Expected
40Hz repetition rate, one Holmium rod on, 50J lamp energy, 800V lamp voltage, READY state

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Fiber alignment high voltage

Positioning YAG beam down optical path into fiber center by adjusting relay mirrors and monitoring beam centering through aperture and detector box

Ranked causes

  1. 1. Relay mirror misalignment (common)
First check

Examine burn pattern at second wedge optic opening with test aperture

Test point
Second wedge optic opening
Expected
Burn is round and centered in aperture with no clipping

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Initial flash lamp drive calibration high voltage

Set start point flash lamp drive parameters for lamp energy normalization

Ranked causes

  1. 1. Flash lamp drive parameters require initial calibration (common)
First check

Verify NVRAM storage after CALIBRATE

Test point
Service screen #1 after DONE message
Expected
DONE message displayed; parameters persisted in NVRAM

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Laser pulse energy high voltage

Automatic calibration to determine cap charge voltages for lowest and highest selectable pulse energies for each of four laser rods.

Ranked causes

  1. 1. Calibration procedure required (common)
First check

Navigate to first service screen

Test point
LCD display
Expected
CALIBRATE button visible

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Optic replacement and realignment high voltage

Procedure after replacing a damaged optic to restore system alignment by adjusting the replaced optic rather than full procedure.

Ranked causes

  1. 1. Damaged optic requiring replacement (rare)
First check

Replace damaged optic using mounting and adjusting screws

Test point
Optic mount assembly
Expected
Optic properly seated in retainer with mounting screw secured

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pyro mirror alignment high voltage

Pyro detector mirrors require alignment to ensure accurate system calibration. Alignment may also be needed after solenoid board replacement.

Ranked causes

  1. 1. Misaligned pyro mirrors (common)
First check

Set up laser for Holmium: turn on one flashlamp, set lamp energy to 50 Joules, lamp volts to 800V, repetition rate to 40 Hz

Test point
Service screen
Expected
Lamp energy 50 J, lamp volts 800 V, repetition rate 40 Hz
Tools
Service screen

6 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Servo motor and mirror alignment high voltage

Procedure to remove, reinstall and align the servo motor/encoder assembly and servo mirror to ensure proper laser beam routing to all four channels

Ranked causes

  1. 1. Motor/encoder assembly requires replacement or realignment (unknown)
First check

Verify motor shaft rotation and index position detection

Test point
TP24 (U73-19)
Expected
Goes high at index position or home position

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: YAG channel alignment high voltage

YAG channel cavity HR positioning, output direction, relay mirror centering, and beam coaxiality with other YAG channels must be verified and adjusted to meet three criteria: HR positioned for center output to first relay mirror with maximum power OC position; first and second relay mirrors directing energy centered through wedge optic apertures and into fiber proximal end; each YAG beam coaxial with other channels at second wedge optic aperture and fiber port.

Ranked causes

  1. 1. YAG channel misalignment requiring adjustment (common)
First check

Verify cavity HR positioned to direct YAG output to center of first relay mirror

Test point
First relay mirror centerline
Expected
Output directed to center

6 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CPU test failed (F8) patient safety

Main processor CPU test fails during self-test

Ranked causes

  1. 1. Main processor CPU hardware failure (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display problems (screen lock up, jumbled/erratic operation)

Screen lock up or jumbled and erratic display operation on system display

Ranked causes

  1. 1. Outdated software revision (pre-revision E 0629-141-01) (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

EPROM socket compatibility issue caution

Controller PCB may have incompatible socket types (28 pin vs 32 pin) preventing proper software installation

Ranked causes

  1. 1. Controller PCB earlier than revision C with 28 pin sockets (common)
First check

Verify Controller PCB socket type in U79 and U87 positions

Test point
U79 and U87 socket pins
Expected
32 pin sockets for revision E software; 28 or 32 pin sockets acceptable for older software if properly positioned

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fan noise excessive

Fan running at high RPM producing loud noise during operation

Ranked causes

  1. 1. Fan speed controller not optimized for noise reduction (common)
First check

Verify fan controller idle speed

Test point
Visual inspection during operation
Expected
Fan idles slightly faster than before, speeds up more quickly when required

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

High voltage power supply error (F14) high voltage

HVPS (high voltage power supply) detects error condition

Ranked causes

  1. 1. HVPS overvoltage, overcurrent, or thermal condition (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Intermittent 20% energy messages

Intermittent 20% High and Low energy messages displayed on screen along with fault 36/56 and various display problems

Ranked causes

  1. 1. Outdated software revision (pre-revision E 0629-141-01) (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Intermittent 20% high and low energy messages

System displays intermittent messages indicating energy at 20% above or below expected levels

Ranked causes

  1. 1. Software revision earlier than E 0629-141-01 (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Laser overheating during continuous operation

System temperature rises excessively when operated continuously for more than a few minutes

Ranked causes

  1. 1. Fan speed controller not adequately improving cooling characteristics (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loss of coolant flow detected caution

System will not start or shuts down if coolant flow is insufficient; monitored by SW3 flow switch as part of 24 VAC loop to main contactor.

Ranked causes

  1. 1. Flow switch SW3 malfunction or inadequate coolant circulation (unknown)
First check

Verify coolant pump operation and flow

Test point
Laser head inlet/outlet, pump output
Expected
Visible coolant circulation through system

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Overheating message displays incorrectly caution

System displays overheating warning at higher than expected water temperature threshold

Ranked causes

  1. 1. Software earlier than revision E lacks temperature compensation (common)
First check

Monitor water temperature when overheating message appears

Test point
Water temperature sensor
Expected
Below 40½C

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Overheating message displays prematurely

Overheating message displays when water temperature reaches 40½C instead of expected higher threshold

Ranked causes

  1. 1. Water temperature exceeding 40½C setpoint (common)
First check

Monitor water temperature

Test point
water temperature sensor
Expected
below 38½C for reset, display triggers at 40½C

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Service mode entry fails

Main or safety microprocessor does not recognize service switch transition; system does not enter service mode.

Ranked causes

  1. 1. Service switch SW2 not toggled OFF to ON during power-up (common)
First check

Verify SW2 is to the left of main microprocessor and toggled during power-up

Test point
Controller PCB SW2
Expected
OFF to ON transition during power-up

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

UART break condition detected (F2) patient safety

Serial communication UART detects break condition on data line

Ranked causes

  1. 1. Serial communication line break or interference (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

YAG channel partial adjustment for minor misalignment high voltage

Quick alignment procedure when single channel is only slightly out of center at fiber port, using first relay mirror adjustment and test fiber burn confirmation.

Ranked causes

  1. 1. Single YAG channel slightly out of center at fiber port (occasional)
First check

Adjust first relay mirror using fiber detector signal

Test point
First relay mirror adjustment screws
Expected
Fiber detector signal optimized; beam centered into proximal end of fiber

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
F1ADC remains busy too longADC does not complete conversion within timeout perioderror
F100HVPS switches wrong. (UP = Ho, DN = Nd)HVPS laser selection switches are set to wrong laser module combinationerror
F101Fiber to HVPS badLaser delivery fiber connection or interface to HVPS is defective or disconnectedwarning
F102HVPS overheatingHVPS thermal temperature sensor indicates overtemperature conditionwarning
F103HVPS analog readback channel badHVPS analog feedback measurement circuit detects fault or signal errorwarning
F104Lamp voltage errorLamp (lamp-pumped laser) supply voltage outside acceptable rangeinfo
F105Pulsewidth too lowLaser pulse width measurement falls below minimum acceptable valuewarning
F109Model not recognizedSystem firmware cannot identify connected laser module or system varianterror
F11Error processor errorSafety/error processor detects internal fault during self-testerror
F110Safety DAC defectiveSafety-critical digital-to-analog converter self-test failserror
F111Firepulse activeFire pulse signal remains active when it should be inactive (informational)error
F112Pulsewidth too highLaser pulse width measurement exceeds maximum acceptable valuewarning
F113SP reports illegal pulse intervalSafety processor detects pulse interval outside valid rangeerror
F114SP reports bad max. rateSafety processor detects maximum pulse rate setting outside specificationerror
F115SP reports bad max. energySafety processor detects maximum energy setting outside specificationerror
F116SP reports illegal pulse rateSafety processor detects pulse rate outside valid operating rangeerror
F117SP reports illegal pulse rateSafety processor reports second occurrence of pulse rate outside specificationerror
F118MP finds SP not idleMain processor detects safety processor still processing previous commanderror
F119Failed energy low testEnergy calibration low-point test fails during self-testerror
F12Spurious firepulse interruptUnexpected fire pulse interrupt signal detectedwarning
F120Failed energy high testEnergy calibration high-point test fails during self-testerror
F121SP reports wrong error in safety testSafety processor reports incorrect error code during safety self-testerror
F122SP reports no error in safety testSafety processor fails to report error when safety test detects fault conditionerror
F123SP reports incorrect footswitch configurationSafety processor detects footswitch configuration mismatch at startuperror
F124SP not in test modeSafety processor not ready to enter test mode when commanded (informational)info
F125Calibration failedSystem energy or parameter calibration procedure fails to completeerror
F126HVPS stuck in softstartHVPS soft-start circuit cannot transition to full power outputerror
F127SP failed test(s)Safety processor internal self-test fails (multiple possible failures)error
F128SP calibration bad dataSafety processor calibration data fails integrity checkerror
F129SP CRC errorSafety processor program CRC checksum verification failserror
F13UART detected framing errorSerial data received with incorrect stop/start bit patternwarning
F130SP DRAM-check errorSafety processor dynamic RAM self-test failserror
F131Other SP error at startupSafety processor reports unspecified error during power-on startuperror
F132Marks end of error logLast entry in system error log reachedinfo
F133HVPS volts command channel badHVPS voltage command ADC input channel detects circuit failure or signal errorwarning
F134HVPS current command channel badHVPS current command ADC input channel detects circuit failure or signal errorwarning
F136SP reports incorrect FS statusSafety processor detects inconsistent footswitch statuserror
F137Crowbar not fired when it should beHVPS overvoltage detected but crowbar safety discharge circuit fails to activatecritical
F138Wrong lamp voltage (not displayed)Lamp supply voltage error detected internally (not shown on operator display)warning
F139Crowbar fired (not displayed)Crowbar safety discharge has fired; error not displayed to operator (informational)warning
F14High voltage power supply errorHVPS detects overvoltage, overcurrent, or shutdown conditionwarning
F140HVPS DC not setHigh voltage power supply initialization failureerror
F141No fire pulse detectedFire pulse detection failure during testingerror
F142Holmium laser not clearedHolmium laser residual energy at startuperror
F143Nd laser not calibratedNd laser output calibration missing or failederror
F144Holmium laser rod energy lowHolmium rod laser output below minimum energyunknown
F145Safety test failedCould not perform safety testcritical
F146All Nd lasers offAll Nd laser tubes are inactiveerror
F147Too many shots in self-testExcessive number of laser shots during self-testerror
F148Crowbar diagnostic no resultCrowbar circuit test produced no measurable resulterror
F149Temperature sensor disconnectedTemperature sensor not detected during self-testunknown
F15Main program deadMain processor program execution halted or infinite loop detectederror
F150Laser 1 cannot reach 90% of max energyLaser #1 output below 90 percent of rated maximumerror
F151Laser 2 cannot reach 90% of max energyLaser #2 output below 90 percent of rated maximumerror
F152Laser 3 cannot reach 90% of max energyLaser #3 output below 90 percent of rated maximumerror
F153Laser 4 cannot reach 90% of max energyLaser #4 output below 90 percent of rated maximumerror
F16-15V supply voltage wrong-15V power supply output outside acceptable rangewarning
F17Nonexistent error codeSafety processor reports undefined fault code numbererror
F18UART detected data overrun errorReceived serial data before previous byte was read (buffer overflow)warning
F19+5V supply voltage wrong+5V power supply output outside acceptable rangewarning
F2UART detected break conditionSerial communication line held low beyond break timeoutwarning
F20UART detected parity errorReceived serial data with incorrect parity bitwarning
F21+15V supply voltage wrong+15V power supply output outside acceptable rangewarning
F22ROM CRC check failedRead-only memory checksum verification fails during startuperror
F23Safety related data errorSafety processor detects corruption in critical safety dataerror
F24Shutter drive stuck offShutter solenoid or actuator fails to open when commandedwarning
F25Shutter drive stuck onShutter solenoid or actuator fails to close when commandederror
F26Servo errorServo motor feedback or control error detectederror
F27Software errorUnhandled exception or illegal instruction in main processor softwareerror
F28Safety shutter not closedShutter position sensor indicates shutter not closed when requirederror
F29Safety shutter not openShutter position sensor indicates shutter not open when required for beam patherror
F31st 68230 malfunction68230 parallel interface chip fails self-test or operation checkerror
F32DAC feedback check errorDigital-to-analog converter feedback check fails during self-testerror
F33HCTL servo controller brokenHCTL motion control IC fails internal test or communication checkwarning
F34DPRAM data incorrectDual-port RAM data verification check failserror
F35No fire setNo firing command set when firing expected during operationerror
F36Safety processor errorSafety processor reports internal error or fault condition (not elsewhere listed)error
F37Software detects energy highMeasured laser energy exceeds set point plus toleranceerror
F38Software detects energy lowMeasured laser energy below set point minus toleranceerror
F42nd 68230 malfunctionSecond 68230 parallel interface chip fails testerror
F40Pulse rate too highMeasured pulse rate exceeds maximum rate setting or hardware limiterror
F41Pulse rate too lowMeasured pulse rate below minimum rate setting or expected rateerror
F42Unexpected fire pulseFire pulse detected when system not in fire command modewarning
F43DPRAM handshake errorDual-port RAM synchronization handshake between processors failserror
F44MP - SP firing parameter mismatchMain processor and safety processor firing parameters disagreeerror
F45DPRAM failed testDPRAM self-test or read-write test fails during startuperror
F46No delivery system attachedDelivery system connector not seated or delivery system missingerror
F47Footswitch disconnect or brokenFootswitch signal is missing or does not respond to press during self-testerror
F48Footswitch down during self testFootswitch held down or stuck when system power-on self-test runserror
F49System inactiveSystem detects no activity for extended period (not used)warning
F50Remote interlock openRemote safety interlock circuit is openerror
F51No control panelControl panel not responding to communication from main processorerror
F52Coolant over temperatureCoolant temperature exceeds safe operating limitwarning
F53Performing self testSystem self-test sequence is running (informational only)info
F55System not calibratedEnergy or other critical system calibration has not been performederror
F56Failed safety test: energyEnergy safety test during startup detects out-of-range energy valueerror
F57Failed safety test: pulse intervalPulse interval safety test during startup detects incorrect timingerror
F58NVRAM checksum incorrectNon-volatile RAM checksum verification fails at startuperror
F59NVRAM system data incorrectSystem configuration data in NVRAM fails integrity checkerror
F6Coolant conductivity wrongCoolant conductivity measurement outside acceptable rangewarning
F60Energy monitor hardware errorEnergy monitoring circuit self-test failserror
F61Program flow errorMain processor detects unexpected program flow or jump to invalid addresserror
F62Energy 20% highMeasured laser energy 20% above set point (informational)info
F63Energy 20% lowMeasured laser energy 20% below set point (informational)info
F64Pulse rate 20% highMeasured pulse rate 20% above set point (informational)info
F65Pulse rate 20% lowMeasured pulse rate 20% below set point (informational)info
F66Front communication time outCommunication with front panel control or display times outwarning
F67Software too slowMain processor software execution time exceeds real-time deadlineerror
F68Blastshield missingBlastshield (protective enclosure) not installed or misaligned during self-testerror
F7Communication data errorData integrity check fails in inter-processor communicationerror
F70No fire test failedSystem cannot successfully block firing during no-fire safety test at startuperror
F71Autocalibration disabledAutocalibration feature has been disabled (informational)info
F72Service attenuator in beam pathService attenuator detected in laser beam path during operationinfo
F73All Holmium lasers are offAll Holmium laser tubes are detected in off state (informational)info
F74Max. energy reducedMaximum energy output has been automatically reduced (informational)info
F75Maximum pulse rate reducedMaximum pulse rate has been automatically limited (informational)info
F78Wrong laser firedFire pulse detected from laser other than one selected by operatorerror
F79Exposure time limit exceededCumulative exposure time during treatment session exceeds safe limiterror
F8CPU test failed main processor has failedMain processor CPU self-test failserror
F81Low energy attenuator position errorLow energy attenuator position sensor reports inconsistent positionerror
F82SP failed time exposure testSafety processor time/exposure measurement test fails during self-testerror
F83SP failed rod id testSafety processor laser rod identification test fails at startuperror
F84Pulse repetition mode errorPulse repetition mode setting inconsistent or invaliderror
F85SP failed repetition mode testSafety processor pulse repetition mode test fails during self-testerror
F86Brick 1 failureSolid-state laser brick 1 (HVPS module 1) detects fault conditionwarning
F87Brick 2 failureSolid-state laser brick 2 (HVPS module 2) detects fault conditionwarning
F88Brick 3 failureSolid-state laser brick 3 (HVPS module 3) detects fault conditionwarning
F89Brick 4 failureSolid-state laser brick 4 (HVPS module 4) detects fault conditionwarning
F90Dual footswitch badDual footswitch circuit or communication detects failureerror
F92Pulsewidth readback not the same as requestedLaser pulse width measurement does not match commanded settingwarning
F93Time out after firepulseSystem times out waiting for completion signal after fire pulse sentwarning
F94Crowbar firedCrowbar (emergency high-voltage discharge) safety circuit has fired (informational)info
F95Crowbar test failedCrowbar safety discharge circuit test fails during self-testerror
F99HVPS balance bad. (UP=Ho, DN=Nd)High voltage power supply output voltage imbalance between supply railswarning
Fault 10Unspecified system faultMultiple causes related to software or power supply issueserror
Fault 10General system fault or communication errorSoftware incompatibility or system communication failurewarning
Fault 36/56Laser energy or power supply over temperature faultInadequate HVPS communication or excessive operating temperatureerror
OverheatingWater temperature exceeds safe operating thresholdWater temperature reaches 40½Cwarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI