Liko Mobile lifts: troubleshooting and error codes
The ServiceForge knowledge base holds 64 documented symptoms for the Liko Mobile lifts, with 74 ranked causes, 81 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Emergency stop or electrical emergency-lowering device not functioning patient safety
Emergency stop out, but electrical emergency-lowering device does not function; lift arms up/down or base width in/out not responding
Ranked causes
- 1. Actuator or base motor cable position incorrect (occasional)
- 2. Output channel or motor fault (common)
- 3. Control box faulty (occasional)
Reposition actuator and base motor cables
- Test point
- Actuator and base motor connections
- Expected
- Function restored after cable repositioning
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No charging function
Battery does not charge when charger connected; no indication of charging activity
Ranked causes
- 1. Poor contact between control box and battery (common)
- 2. Battery contact plates defective or broken (occasional)
- 3. Charging function faulty in control box (occasional)
Verify battery and control box contact
- Test point
- Battery connection to control box
- Expected
- Green LED indicator light starts when battery reinstalled
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Operation panel functions not responding patient safety
Buttons on operation panel do not activate lift functions
Ranked causes
- 1. Emergency stop button activated (common)
- 2. Low battery capacity (common)
- 3. Faulty control box (occasional)
Verify emergency stop is not activated
- Test point
- Emergency stop button
- Expected
- Button released (not pressed in)
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
All hand control functions not working patient safety
All buttons on hand control are non-responsive
Ranked causes
- 1. Emergency stop not released (common)
- 2. Defective hand control (occasional)
Release emergency stop button
- Test point
- Emergency stop button
- Expected
- Button released and lifted
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery not charging caution
No charging function or battery not holding charge
Ranked causes
- 1. Poor contact between control box and battery (common)
- 2. Charger malfunction (occasional)
Verify contact between control box and battery
- Test point
- Battery and control box contacts
- Expected
- Green LED lights automatically after battery insertion
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Golvo mast locking screws loose patient safety
Upper locking screws in mast are not tight or lower holes have screws installed (not permitted)
Ranked causes
- 1. Upper mast locking screws loose (common)
- 2. Screws installed in lower mast holes (not permitted) (rare)
Verify upper mast locking screws are tight
- Test point
- Upper holes on mast where it connects to base
- Expected
- Both upper screws tight and secure
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Operation panel functions not working patient safety
Buttons on operation panel do not activate lift functions (CBL control unit models)
Ranked causes
- 1. Emergency stop not released (common)
- 2. Faulty control box (occasional)
Release emergency stop button
- Test point
- Emergency stop button
- Expected
- Button released and lifted
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Abnormal noise from actuator caution
Actuator produces unusual sounds during operation
Ranked causes
- 1. Loose components or wear (unknown)
Listen for abnormal noise during actuator operation
- Test point
- Actuator
- Expected
- No abnormal noise
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Actuator fasteners loose patient safety
Upper and lower actuator fasteners to the mast are not tight
Ranked causes
- 1. Fasteners not properly secured (common)
Verify upper and lower actuator fasteners are tight
- Test point
- Upper and lower actuator fasteners to mast
- Expected
- Fasteners tight
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Base widening mechanism not functioning patient safety
Base does not open and close smoothly or reaches incorrect maximum and minimum positions
Ranked causes
- 1. Loose screws or nuts on base assembly (common)
Inspect base widening mechanism
- Test point
- Base assembly
- Expected
- No visible damage or wear
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery takes longer than normal to charge or does not reach full charge
Charging cycle exceeds 5 hours or battery does not indicate full charge with green LED
Ranked causes
- 1. Battery aged or internal resistance increased (rare)
Observe charging indicator LEDs during charge cycle
- Test point
- Battery charging LEDs
- Expected
- Lamps blink alternately yellow/yellow/green during charging; blinking green indicates charging complete
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Brakes do not hold patient safety
When brakes are locked, wheels still turn or housing swivels when lift is pushed
Ranked causes
- 1. Brake mechanism failure (occasional)
Lock brakes and push lift horizontally
- Test point
- All wheels and swivel housing
- Expected
- Wheels do not turn and housing does not swivel
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cable cover incorrectly mounted caution
Cable cover on CBL20 control unit not mounted properly
Ranked causes
- 1. Cable cover loose or displaced (common)
Visually inspect cable cover mounting on CBL20
- Test point
- CBL20 control unit cable cover
- Expected
- Cable cover mounted securely and correctly
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cable not properly inserted caution
Cables loose or disconnected from sockets
Ranked causes
- 1. Loose cable connections (common)
Verify all cables are fully inserted
- Test point
- All cable connections
- Expected
- Cables firmly seated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: LikoLight mobile lift welding inspection caution
Visual inspection of welding between arm and fastening device for deformation or cracks on LikoLight lifts.
Ranked causes
- 1. Welding defect or fatigue crack from loading cycles (rare)
Visually inspect welding between arm and fastening device
- Test point
- LikoLight arm-fastening welded joint
- Expected
- No visible deformation or cracks in welding
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control unit replacement affects usage tracking data
When Control Unit CBL or Jumbo Care is replaced, historical lift usage data (total Ah or As cycles) reflects only the new control unit, not the complete lift history
Ranked causes
- 1. Control Unit replaced; memory cleared or transferred incomplete (occasional)
Review control unit replacement history in lift maintenance records
- Test point
- Service log and control unit serial number
- Expected
- Match serial number in documentation to verify original or replacement unit
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Deformities in sling bar attachment points patient safety
Attachment points show visible deformation or damage
Ranked causes
- 1. Stress, impact, or wear on attachment points (rare)
Visually inspect attachment points
- Test point
- Sling bar attachment points (A)
- Expected
- No visible deformities or damage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Emergency stop button not releasing patient safety
Red emergency stop button remains locked in pressed position and does not pop up
Ranked causes
- 1. Emergency stop button stuck or misaligned (unknown)
Turn emergency stop button in direction of arrows
- Test point
- Red emergency stop button
- Expected
- Button releases and pops up to open position
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive gap at M220/M230 fastening area patient safety
Gap at measured area exceeds maximum allowable distance on M220/M230 models
Ranked causes
- 1. Fastening screws loose or fastener wear (common)
Measure gap at specified area (M220/M230 only)
- Test point
- M220/M230 fastening area
- Expected
- Gap does not exceed 3 mm (1/8 inch), max 3 mm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flexlink arm binds or does not lower freely patient safety
Link arm or flexlink arm does not move freely; binding detected when raising, lowering, or pivoting sling bar.
Ranked causes
- 1. Flexlink screw too tight (common)
Manually raise sling bar and link arm with sling bar hanging lowest
- Test point
- Link arm
- Expected
- Freely raises and lowers without binding
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flexlink arm binds or does not move freely patient safety
Link arm or flexlink arm sticks or resists movement when manually raised/lowered or pivoted
Ranked causes
- 1. Flexlink screw too tight (common)
Manually raise sling bar and link arm
- Test point
- Link arm
- Expected
- Link arm lowers and raises freely
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot tray cracked or loose patient safety
Foot tray shows cracks or is not securely attached to metal frame (Sabina and RollOn models)
Ranked causes
- 1. Foot tray damaged or fasteners loose (occasional)
Inspect foot tray for cracks
- Test point
- Foot tray surface
- Expected
- No visible cracks
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Golvo middle beam distances A and B not equal patient safety
Distances A and B on Golvo middle beam are not equal (symmetry check)
Ranked causes
- 1. Middle beam misaligned or bent (rare)
Measure distances A and B on middle beam
- Test point
- Points A and B on Golvo middle beam
- Expected
- Distance A equals distance B
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Golvo stop screw lower not secured patient safety
K6S M8x16 screw on Golvo middle beam is not properly secured with nyloc nut
Ranked causes
- 1. K6S M8x16 screw loose or nyloc nut missing (occasional)
Verify lower stop screw is secured
- Test point
- K6S M8x16 screw and nyloc nut in middle beam
- Expected
- Stop screw tight with nyloc nut
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Golvo stop screw upper not secured patient safety
M6x8 Ezloc stop screw on Golvo middle beam is not properly secured with square nut (Golvo produced from Dec 2014 or with replaced middle beam)
Ranked causes
- 1. M6x8 Ezloc stop screw loose or nut missing (occasional)
Verify upper stop screw is secured
- Test point
- M6x8 Ezloc stop screw and square nut in middle beam
- Expected
- Stop screw tight with square nut
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hand control buttons do not respond with emergency stop pressed patient safety
Hand control buttons remain inactive when emergency stop is engaged
Ranked causes
- 1. Normal operation - emergency stop engaged (common)
Press emergency stop and attempt hand control button operation
- Test point
- Hand control buttons with emergency stop engaged
- Expected
- Hand control buttons have no effect on lift
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hand control buttons not operating lift patient safety
Control panel functions work but hand control has no function
Ranked causes
- 1. Defective or broken battery contact plates in hand control (common)
Visually inspect hand control battery contact plates
- Test point
- Hand control battery contacts
- Expected
- Contact plates intact and clean
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hand control has no function but control panel works patient safety
Control box function is operational but hand control does not respond to commands
Ranked causes
- 1. Hand control battery contact plates defective or broken (common)
Inspect hand control battery contact plates
- Test point
- Hand control battery contacts
- Expected
- Contact plates intact and properly positioned
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leg widening play excessive (Sabina II only) patient safety
Horizontal movement of legs from inner position exceeds allowable limit
Ranked causes
- 1. Leg support pins or hinges worn or loose (occasional)
With legs in inner position, push legs by hand in and out and measure total movement
- Test point
- Leg pivot assembly
- Expected
- 35 mm / 1.38 Inch total or less, <= 35 mm
- Tools
- Ruler or caliper
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lift arm binds or does not move freely patient safety
Arm does not move smoothly through full lifting cycle without load; binding or restricted motion detected.
Ranked causes
- 1. Mounting screw too tight (common)
Run lift arm through full cycle without load
- Test point
- Lift arm
- Expected
- Smooth motion without binding
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lift arm or base width function not working patient safety
Lift arms do not move up/down or base width does not adjust in/out
Ranked causes
- 1. Actuator or base motor cable disconnected or reversed (common)
Verify actuator and base motor cable positions
- Test point
- Actuator and base motor cable connectors
- Expected
- Cables in correct position for respective motors
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
LikoLight mobile lift has welding deformation or cracks patient safety
Visible deformation or cracks in welding between arm and fastening device on LikoLight model
Ranked causes
- 1. Fatigue, stress, or impact damage to welding (rare)
Visually inspect welding on LikoLight lift
- Test point
- Welding between arm and fastening device
- Expected
- No visible deformation or cracks in welding
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery capacity caution
Operating time noticeably shorter than normal; battery capacity below expected
Ranked causes
- 1. Battery aged or degraded (common)
Press battery capacity button to check LED indicators
- Test point
- Battery capacity LED indicator
- Expected
- Green LED indicates >50% capacity; yellow indicates 25-50% or 0-25%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery capacity or short operating time
Battery operates noticeably shorter than normal; reduced runtime between charges
Ranked causes
- 1. Battery capacity degraded or battery aged (common)
Check battery capacity indicator
- Test point
- Battery i button and LED indicators
- Expected
- Green light indicates more than 50% capacity; yellow light indicates 25-50% or 0-25%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
M220/M230 sling bar gap excessive caution
Gap in M220/M230 sling bar area exceeds maximum allowed tolerance.
Ranked causes
- 1. Wear or deformation of sling bar or fastening points (occasional)
Measure gap in sling bar area on M220/M230 lifts
- Test point
- Sling bar gap area
- Expected
- 3 mm (1/8 inch) or less, max 3 mm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mast holes deformed patient safety
Deformation of holes in mast detected, distance between sides unequal
Ranked causes
- 1. Mast holes deformed due to stress or impact (rare)
Measure distance from cover hat to top of ear on each side
- Test point
- Mast holes (X¹ and X²)
- Expected
- X¹=X² ±1 mm/0.04 inch, ±1 mm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Micro switch fault in LA31 caution
Micro switch malfunction in LA31 (Uno 200, Sabina 200 models)
Ranked causes
- 1. Defective micro switch in LA31 (unknown)
Inspect micro switch using SB001 service box
- Test point
- LA31 micro switch
- Expected
- Micro switch functions properly
- Tools
- Service box SB001 with adapter cable
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Outer tube stuck during operation patient safety
Outer tube gets stuck at some positions when arm is at highest position
Ranked causes
- 1. Outer tube obstruction or wear (unknown)
Pull outer tube to top and bottom positions
- Test point
- Outer tube
- Expected
- Outer tube moves smoothly without sticking
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overload indicator lamp lit patient safety
Overload warning lamp illuminates during lift operation (Viking, Golvo models)
Ranked causes
- 1. Electric end stop failure (common)
Run lift to full up and full down positions
- Test point
- Overload indicator lamp
- Expected
- Indicator lamp not lit
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Play between fork and wheel nut patient safety
Visible or detectable movement between castor fork and wheel fastener
Ranked causes
- 1. Wheel nut loose (common)
Manually move wheel side to side while holding fork
- Test point
- Wheel nut connection to castor fork
- Expected
- No play; wheel moves only with fork
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
RollOn cushion cover damaged caution
Cushion cover on RollOn support shows rips or tears
Ranked causes
- 1. Cushion cover worn or damaged (occasional)
Inspect cushion cover condition
- Test point
- Cushion cover surface
- Expected
- No rips or tears
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
RollOn support locking mechanism loose patient safety
Upper or lower locking knobs are not tight or back plate screws are loose (RollOn support)
Ranked causes
- 1. Locking knobs or plate screws not tightened (common)
Verify upper and lower locking knobs are tight
- Test point
- Locking knobs on RollOn support
- Expected
- Knobs secure with no play
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sabina latches not mounted patient safety
Latches (C) on Sabina sit-to-stand lift are missing or not installed.
Ranked causes
- 1. Latches missing or not installed (rare)
Visually inspect Sabina latches
- Test point
- Sabina lift latches (C)
- Expected
- Latches mounted and in place
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sabina lift hook screws and locking washers loose or missing patient safety
Lift hook fasteners are not properly tightened or are missing on Sabina sit-to-stand lift
Ranked causes
- 1. Fasteners loose due to vibration or inadequate initial assembly (occasional)
Inspect lift hook screws and locking washers
- Test point
- Lift hook fasteners (B)
- Expected
- All screws and locking washers tightened
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sabina lift hook screws loose patient safety
Lift hook screws or locking washers on Sabina lift are not tightened; fastening is loose.
Ranked causes
- 1. Lift hook fasteners loose (occasional)
Inspect and tighten lift hook screws and locking washers
- Test point
- Sabina lift hook fasteners (B)
- Expected
- All screws and locking washers tight
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sabina lift hooks are outdated design patient safety
Sabina sit-to-stand lift equipped with old design lift hooks instead of latest version.
Ranked causes
- 1. Lift hooks not upgraded to latest design (occasional)
Visually compare lift hooks against latest design reference
- Test point
- Sabina lift hooks (A)
- Expected
- Latest design hooks installed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sabina lift latches not mounted patient safety
Latches missing from Sabina sit-to-stand lift
Ranked causes
- 1. Latches missing or not installed (rare)
Verify latches are mounted on Sabina lift
- Test point
- Latches (C)
- Expected
- Latches mounted on Sabina sit-to-stand lift
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sabina sit-to-stand lift has old design lift hooks patient safety
Lift hooks are old design instead of latest version
Ranked causes
- 1. Lift hooks not updated to latest design (occasional)
Inspect lift hook design on Sabina sit-to-stand lift
- Test point
- Lift hooks (A)
- Expected
- Latest design lift hooks installed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Service Light illuminates on Control Unit patient safety
Service Light at the Control Unit illuminates, indicating that thorough inspection is required
Ranked causes
- 1. Scheduled maintenance interval reached or lift usage thresholds exceeded (common)
Verify Service Light status and reason for illumination
- Test point
- Control Unit display
- Expected
- Service Light illuminated or annual inspection due
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sling bar attachment points deformed caution
Deformities or damage detected at sling bar attachment points; may affect alignment or fastening.
Ranked causes
- 1. Impact, overload, or wear damage at attachment points (rare)
Visually inspect attachment points (A) for deformation
- Test point
- Sling bar attachment points
- Expected
- No visible deformation or damage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sling bar cover not attached caution
Sling bar cover is missing or not properly attached to sling bar on standard model
Ranked causes
- 1. Cover not installed or fasteners loose (occasional)
Inspect sling bar cover attachment
- Test point
- Sling bar cover fasteners
- Expected
- Cover securely attached to sling bar
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sling bar does not rotate freely patient safety
Sling bar rotates unevenly, wobbles, or spinning stops due to high friction or bearing issues
Ranked causes
- 1. Bearing friction or contamination (occasional)
Rotate sling bar and observe movement
- Test point
- Sling bar bearings
- Expected
- Sling bar rotates freely without wobble or uneven rotation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sling bar does not rotate freely on bearings patient safety
Sling bar rotates unevenly, wobbles, or stops due to high friction; bearing rotation not smooth.
Ranked causes
- 1. Bearing friction or contamination (unknown)
Rotate sling bar by hand and observe motion
- Test point
- Sling bar bearings
- Expected
- Smooth, even rotation without wobble, high friction, or stopping
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sling bar latches not mounted or not falling back patient safety
Latches are missing from sling bar or do not fall back against the body when released.
Ranked causes
- 1. Latches missing or damaged (rare)
Visually inspect both sling bar latches
- Test point
- Sling bar latches
- Expected
- Both latches mounted and falling back against sling bar body
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sling bar latches not mounted or not functioning patient safety
Latches are missing, not mounted, or do not fall back against body of sling bar
Ranked causes
- 1. Latches missing or loose (occasional)
Verify both latches are mounted and functional
- Test point
- Sling bar latches
- Expected
- Both latches mounted and fall back against sling bar body
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sling bar not correctly fastened patient safety
Sling bar fastening is loose or inadequate; unit may move or shift during operation.
Ranked causes
- 1. Fastening bolts or screws loose (occasional)
Check tightness of sling bar fastening bolts and screws
- Test point
- Sling bar fastening points
- Expected
- All fasteners tight and secure
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sling bar standard o-ring missing or mispositioned caution
O-ring is not present in center bolt groove or is not centered on SlingBar Standard model
Ranked causes
- 1. O-ring not installed or has shifted position (occasional)
Inspect o-ring on SlingBar Standard
- Test point
- Center bolt groove
- Expected
- O-ring present and centered in groove
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SlingBar Standard bearing rotation compromised patient safety
Unit does not rotate freely on bearings or latches are not functioning properly on standard sling bar
Ranked causes
- 1. Bearing contamination, wear, or loose latches (occasional)
Rotate unit and verify bearing function
- Test point
- Sling bar bearings and latches
- Expected
- Unit rotates freely; both latches fall back against body
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SlingBar Standard cover not attached caution
Sling bar cover is not attached or is missing from sling bar.
Ranked causes
- 1. Cover not attached, loose, or missing (rare)
Visually inspect sling bar cover attachment
- Test point
- SlingBar Standard cover
- Expected
- Cover firmly attached to sling bar
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SlingBar Standard does not rotate freely patient safety
SlingBar Standard unit does not rotate smoothly on bearings; uneven rotation, wobble, or high friction present.
Ranked causes
- 1. Bearing friction, contamination, or wear (unknown)
Rotate SlingBar Standard by hand and observe motion
- Test point
- SlingBar Standard bearings
- Expected
- Smooth, free rotation without wobble or high friction
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- SlingBar Standard latches not mounted
- SlingBar Standard o-ring missing or mispositioned
- Wheel height out of specification (Sabina II only)
- Wheels do not roll freely
4 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.