Leica CM1850: troubleshooting and error codes
The ServiceForge knowledge base holds 36 documented symptoms for the Leica CM1850, with 87 ranked causes, 16 diagnostic checks with test points and expected values, and 8 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Ridged sections
Cut sections exhibit ridge marks or ridged patterns.
Ranked causes
- 1. Specimen insufficiently frozen onto specimen disc (common)
- 2. Specimen disc not clamped tightly (common)
- 3. Specimen holder ball joint not clamped (common)
- 4. Knife/blade not clamped tightly enough (common)
- 5. Specimen detached from disc due to excessive thickness (occasional)
- 6. Very hard or inhomogeneous specimen (occasional)
- 7. Blunt knife/blade (occasional)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sections not properly flattened
Cut sections fail to flatten properly or exhibit wrinkles and waves.
Ranked causes
- 1. Anti-roll plate poorly positioned (common)
- 2. Anti-roll plate poorly aligned with knife edge (common)
- 3. Incorrect clearance angle (common)
- 4. Knife/blade blunt or damaged (common)
- 5. Knife/blade and/or anti-roll plate dirty (occasional)
- 6. Edge of anti-roll plate damaged (occasional)
- 7. Blunt knife/blade (occasional)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Chatter during sectioning
Vibration or chattering sound during the cutting motion.
Ranked causes
- 1. Knife profile inappropriate for specimen (common)
- 2. Incorrect clearance angle (common)
- 3. Ice buildup in knife back (occasional)
- 4. Handwheel speed not uniform (occasional)
- 5. Knife/blade not clamped tightly enough (occasional)
- 6. Specimen holder not clamped tightly (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Thick-thin sections
Sections vary in thickness, alternating between thick and thin cuts.
Ranked causes
- 1. Cryocompound applied to cold disc, specimen detached after freezing (common)
- 2. Blunt cutting edge (common)
- 3. Incorrect clearance angle (occasional)
- 4. Microtome not properly dried before reinstallation (occasional)
- 5. Dried specimen (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No or insufficient refrigeration caution
Cooling output is reduced or absent; chamber does not reach set temperature
Ranked causes
- 1. Compressor defective (common)
- 2. Leak in cooling system (common)
- 3. Inappropriate site conditions (occasional)
- 4. Ventilation slits of liquefier dirty (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sections splinter
Cut tissue sections break or splinter into fragments during sectioning.
Ranked causes
- 1. Specimen too cold (common)
- 2. Static electricity or air currents affecting specimen (occasional)
- 3. Specimen not cold enough (occasional)
- 4. Large area specimen (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sections tear
Tissue sections tear or shred during cutting rather than producing clean sections
Ranked causes
- 1. Blunt part, dirt, dust, frost or rust on knife/blade (common)
- 2. Leading edge of anti-roll plate damaged (occasional)
- 3. Hard particles in tissue (rare)
- 4. Knife back dirty (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tissue sticks or crumbles on anti-roll plate
Tissue material adheres to or breaks apart on the anti-roll plate surface.
Ranked causes
- 1. Temperature incorrect for tissue being cut (common)
- 2. Knife profile inappropriate for specimen (common)
- 3. Anti-roll plate too warm or incorrectly positioned (occasional)
- 4. Static electricity affecting tissue (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor startup current requirement not met high voltage
Nominal voltage drops below specified values during compressor start-up, instrument may not operate properly or sustain damage due to insufficient power supply capacity
Ranked causes
- 1. Electric circuit at installation site does not meet start-up current requirements (common)
- 2. Other consumers connected to same circuit (occasional)
- 3. Inadequate or unstable mains power supply (occasional)
Verify electric circuit capacity with electrical engineer
- Test point
- Installation site power supply
- Expected
- Circuit rated for minimum 45-50 A start-up current and stable nominal voltage per Technical data
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cryostat inoperational caution
Unit does not cool or operate; no refrigeration function
Ranked causes
- 1. Mains plug not properly connected (common)
- 2. Defective fuses (common)
- 3. Temperature control switch activated (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fat on corner or edge of anti-roll plate
Visible fat deposits on anti-roll plate surfaces during sectioning
Ranked causes
- 1. Rusty knife/blade (common)
- 2. Static electricity or air currents (occasional)
- 3. Anti-roll plate too warm (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Frost on chamber walls and microtome
Visible frost accumulation on interior surfaces of the cryochamber and microtome.
Ranked causes
- 1. Cryostat exposed to air currents from open windows and doors, or air conditioning (common)
- 2. Sliding window left open too long, exposed to air currents (common)
- 3. Frost built up from breathing into cryochamber (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ice formation on bottom of cryochamber
Accumulation of ice at the base of the cryochamber.
Ranked causes
- 1. Condensation water drain obstructed (common)
- 2. Drain of quick freeze shelf defrosting water obstructed (common)
- 3. Cryostat not aligned level (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor start-up current exceeds supply capacity high voltage
Instrument may not start or operate smoothly if electrical supply cannot deliver required current during compressor initialization.
Ranked causes
- 1. Electrical circuit undersized or inadequate for compressor startup current (common)
- 2. Nominal voltage drops below specified minimum during startup (common)
Verify electrical circuit capacity and specifications
- Test point
- Installation site main circuit breaker and wiring
- Expected
- Circuit rated for 45–50 A startup current and capable of maintaining nominal voltage per instrument nameplate
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inconsistent or insufficient specimen feed
Specimen advance is erratic, slow or stops intermittently during sectioning cycle
Ranked causes
- 1. Microtome not entirely dry; ice in micrometer feed system (common)
- 2. Defective microtome (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Instrument damage due to inadequate power supply high voltage
Continuous or unstable power supply causes equipment failure
Ranked causes
- 1. Inadequate or interrupted power supply to instrument (common)
- 2. Other consumers sharing the electrical circuit (occasional)
Verify electrical circuit is dedicated and protected
- Test point
- installation site circuit breaker and wiring
- Expected
- Separate dedicated circuit with appropriate protection, no shared loads
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
LED (2) does not illuminate at front limit
LED (2) on advance button fails to illuminate when specimen reaches front limit
Ranked causes
- 1. LED (2) burned out or disconnected (unknown)
- 2. Front limit switch not triggering (unknown)
Observe LED (2) behavior during specimen advance to front limit
- Test point
- LED (2) on advance button and front limit position
- Expected
- LED (2) illuminates when specimen reaches front limit
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scraping noise during sectioning and specimen return movement
Audible scraping or grinding sound during cutting and return movement of the microtome.
Ranked causes
- 1. Knife/blade damaged (common)
- 2. Edge of anti-roll plate damaged (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sections curl on anti-roll plate
Cut tissue sections curl or wave on the anti-roll plate instead of lying flat.
Ranked causes
- 1. Anti-roll plate does not protrude far enough beyond cutting edge (common)
- 2. Anti-roll plate protrudes too far and scrapes specimen (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sections smear
Cut tissue sections are smeared or dragged during sectioning.
Ranked causes
- 1. Specimen not cold enough (common)
- 2. Knife/blade and/or anti-roll plate not yet cold enough, warming sections (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Specimen does not return to rear limit
Specimen fails to move rearward when return button is pressed; LED (1) should flash during rapid return or slow return buttons pressed
Ranked causes
- 1. Rapid or slow return button not functioning (unknown)
- 2. Return movement blocked or stalled (unknown)
Test rapid return button for LED (1) response
- Test point
- Return button and LED (1) on control panel 2
- Expected
- LED (1) flashes while specimen head is in motion; illuminates steady when rear limit reached
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Anti-roll plate damage after adjustment
Anti-roll plate becomes damaged during positioning or alignment procedures.
Ranked causes
- 1. Plate raised too high when aligning, struck by cutting edge (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Condensation on anti-roll plate and knife during cleaning
Moisture and condensation form on anti-roll plate and knife surfaces during cleaning operations.
Ranked causes
- 1. Brush, forceps and/or cloth too warm (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flattened sections curl up when anti-roll plate is picked up
Tissue sections begin curling immediately upon removal of anti-roll plate
Ranked causes
- 1. Temperature too low for tissue cut (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fuse replacement caution
Fuses need to be replaced with correct specification to avoid equipment damage.
Ranked causes
- 1. Blown or faulty fuse (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Instrument damaged after transport caution
Severe damage to compressor and internal components following shipment and immediate power-on without waiting period.
Ranked causes
- 1. Compressor oil displaced during transport, not allowed to settle before startup (common)
Verify waiting period observed after transport
- Test point
- Clock or timestamp
- Expected
- At least 4 hours elapsed since transport
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Instrument does not power on
No response when automatic mains fuse switch is activated to position 1
Ranked causes
- 1. Instrument powered on too soon after transport (common)
Verify wait time after transport
- Test point
- Transport completion time
- Expected
- At least 4 hours elapsed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lamp replacement high voltage
Broken lamp requires replacement by technical service due to high injury risk.
Ranked causes
- 1. Lamp broken or non-functional (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Manual defrost of quick freeze shelf required caution
Excessive frost formation on quick freeze shelf, particularly after spray disinfection, necessitates manual defrost cycle.
Ranked causes
- 1. Moisture accumulation on quick freeze shelf from spray disinfection (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Microtome malfunction or damage from humidity caution
Humidity inside condenses and freezes in the cold cryostat, causing malfunctions or damage.
Ranked causes
- 1. Microtome not entirely dry before reinstallation (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nominal voltage drops below specification at startup high voltage
Compressor start-up current draws 45–50 A; voltage drop occurs at installation site
Ranked causes
- 1. Inadequate electrical circuit capacity at installation site (common)
Verify mains voltage against Technical data specifications during compressor start-up
- Test point
- mains power outlet at wall
- Expected
- Nominal voltage per Technical data, minimum threshold not exceeded
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scraping noise at microtome caution
Audible scraping or grinding sound during microtome operation
Ranked causes
- 1. Friction between slot cover and microtome housing (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Severe instrument damage from inadequate power supply caution
Failure to ensure proper electrical circuit and constant adequate power supply causes irreversible damage to the instrument.
Ranked causes
- 1. Inadequate or unstable power supply connected to instrument (common)
Verify circuit is protected and dedicated exclusively to instrument
- Test point
- Installation site main circuit
- Expected
- Circuit has separate protection and no other loads connected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Severe instrument damage from power supply failure patient safety
Instrument sustains severe damage if nominal voltage drops during compressor start-up or if power supply is inadequate or unstable
Ranked causes
- 1. Non-compliance with power supply requirements (common)
Confirm electrical engineer inspection completed
- Test point
- Installation documentation
- Expected
- Written verification that circuit meets 45-50 A requirement
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Specimen disc cannot be removed patient safety
Specimen disc stuck to freezing shelf or specimen head and cannot be separated
Ranked causes
- 1. Moisture on underside froze disc to freezing shelf or specimen head (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Specimen does not advance toward knife patient safety
Specimen fails to move forward when coarse feed advance button is pressed; movement should occur as long as button is held
Ranked causes
- 1. Coarse feed advance button not functioning (unknown)
Verify specimen can advance toward knife with coarse feed button
- Test point
- Advance button on control panel 2
- Expected
- Specimen moves toward knife as long as button is pressed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
E0:20 | Calibrating error; possibly defective controller board | Calibration failure during operation | error |
E0:21 | Clock battery on the controller board empty | Battery voltage drops below threshold | error |
E0:22 | Microtome wet | Moisture detected in microtome chamber | warning |
E0:23 | Cryochamber temperature out of range of indication | Temperature sensor reads outside -35 °C to +55 °C range | error |
E0:24 | Short circuit at temperature sensor of chamber cooling system | Sensor circuit short detected | error |
E0:25 | Breaking of temperature sensor of chamber cooling system | Sensor circuit open detected | error |
E0:26 | Short circuit at temperature sensor of evaporator | Evaporator sensor circuit short detected | error |
E0:27 | Breaking of temperature sensor of evaporator | Evaporator sensor circuit open detected | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community