Laborie Aquarius XT-CT-LT-CTS: troubleshooting and error codes
The ServiceForge knowledge base holds 47 documented symptoms for the Laborie Aquarius XT-CT-LT-CTS, with 74 ranked causes, 42 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
No response from devices
Devices do not respond to commands or inputs
Ranked causes
- 1. No power on electrical outlet (common)
- 2. Damaged power cord (occasional)
- 3. Power cord not connected properly (common)
- 4. Devices not connected in software (occasional)
- 5. Wrong connection type selected (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure not responding
Pressure signal absent or not updating
Ranked causes
- 1. Catheter position incorrect (common)
- 2. Catheters not securely connected to cables (common)
- 3. Zeroes not set properly (occasional)
- 4. Catheter is kinked (occasional)
- 5. Using water-based disposables (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer Error LED on steady or flashing
Printer error indicator illuminated continuously or blinking
Ranked causes
- 1. Printer paper is out (common)
- 2. Printer cover is open (common)
- 3. Printer is out of ink (common)
- 4. Printer paper jammed or mis-fed (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
EMG reading too high or low
EMG signal level is not within expected range
Ranked causes
- 1. EMG channel scale not optimized (common)
- 2. Electrodes wet or not sticking (common)
- 3. Detached electrode (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect flow or volume readings patient safety
UroCAP V reports inaccurate flow rate or volume measurement
Ranked causes
- 1. Beaker not properly seated on uroflowmeter dish (common)
- 2. Funnel touching beaker (common)
- 3. Incorrect beaker in use (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
UDS report not transferring to i-LIST Office Reporter
Generated UDS report cannot be sent to i-LIST reporting system
Ranked causes
- 1. i-LIST MsgQListener not running (common)
- 2. Missing Windows updates (occasional)
- 3. .NET Framework file corruption (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Urocap V device not waking or entering sleep mode
Device does not respond to ON button for wake-up or does not enter sleep when Reset button is held; device appears unresponsive
Ranked causes
- 1. Battery empty or critically low (common)
- 2. ON button not pressed or held correctly (occasional)
- 3. Reset button not held long enough for sleep (occasional)
Check battery voltage display on Urocap V screen
- Test point
- Battery display on message screen
- Expected
- Battery voltage 4.2 (full) to 3.6 (empty); status showing Battery Full, Charging, or Battery Powered V
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Urocap V LED indicators not responding or unclear
LED status lights (Bluetooth, Equipment, Power) display incorrect colors or states; technician cannot interpret device status from indicators
Ranked causes
- 1. LED indicators need interpretation reference (common)
- 2. Equipment status LED unclear (common)
- 3. Power LED status unclear (common)
Verify Bluetooth LED indicator color
- Test point
- BT LED on front of Urocap V
- Expected
- blue = connected; green = system at work but not connected; orange = system fault; light off = not connected
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Uroflowmeter signal shows vibration and spike patterns
UroCAP V signal displays unwanted oscillations or spikes
Ranked causes
- 1. Plastic beaker touching flow funnel (common)
- 2. Patient touched uroflowmeter with feet (occasional)
- 3. Floor is unsteady (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: EMG electrode check patient safety
EMG channel does not display activity during muscle movement and contraction tests
Ranked causes
- 1. EMG leads expired or require recalibration (common)
- 2. EMG system requires calibration (unknown)
Place two EMG electrodes 1 inch apart on forearm where muscle flexes most, third patch on elbow
- Test point
- EMG electrodes
- Expected
- Electrodes properly positioned
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Orange Battery LED displaying with no screen icon or keyboard response
USB power connected, orange Battery LED on, no screen icon, no keyboard response
Ranked causes
- 1. Battery is empty (common)
- 2. No external power connected and cannot wake-up ROAM using OK button or Bluetooth (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer LED does not light up
Printer power indicator LED remains off
Ranked causes
- 1. Printer power cable unplugged (common)
- 2. Printer is not turned on (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Uroflowmeter calibration failure: flow out of range
Calibration routine fails with flow measurement out of acceptable range
Ranked causes
- 1. Water poured too quickly into beaker (common)
- 2. More than 500mL poured into beaker (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Infusion transducer check with uroflowmeter
Infused volume reading does not match volume emptied into beaker on uroflowmeter
Ranked causes
- 1. Infusion transducer requires calibration (unknown)
Ensure uroflowmeter is checked and calibrated before starting infusion check
- Test point
- Uroflowmeter status
- Expected
- Uroflowmeter verified and calibrated
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pressure transducer check high voltage
Pressure channel reading does not match expected values at water surface and beaker bottom
Ranked causes
- 1. Pressure transducer requires calibration (unknown)
Switch T-DOC charger to OPEN position
- Test point
- T-DOC charger
- Expected
- Charger in OPEN position
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pressure transducers
Verify pressure transducer calibration and battery/connection status per maintenance schedule
Ranked causes
- 1. Calibration drift or suspected inaccuracy (occasional)
Inspect external surfaces of transducers and cables for contamination
- Test point
- External surfaces and cables
- Expected
- Clean, no fluids or foreign matter
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pump volume check
Pump fill volume reading does not match expected 150 cc when measured in beaker
Ranked causes
- 1. Pump requires calibration (unknown)
Prime pump tubing and set up with catheter into empty beaker
- Test point
- Pump outlet into beaker
- Expected
- Tubing primed and ready
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: UPP Puller
UPP puller requires calibration using on-screen instructions with ruler and pencil
Ranked causes
- 1. UPP puller calibration procedure (unknown)
Gather calibration tools
- Test point
- bench
- Expected
- ruler and pencil available
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Uroflowmeter flow rate check
Flow rate reading does not match expected 500 ml when water poured at 25-50 ml/second rate
Ranked causes
- 1. Uroflowmeter requires calibration (unknown)
Place empty beaker on Uroflow transducer and set zeroes
- Test point
- AQUARIUS software
- Expected
- Zero reference established
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Uroflowmeter volume check
Volume reading on AQUARIUS software does not match expected 500 ml when 500 cc water placed on transducer
Ranked causes
- 1. Uroflowmeter requires calibration (unknown)
Place empty beaker on Uroflow transducer and set zeroes
- Test point
- AQUARIUS software
- Expected
- Zero reference established
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cleaning: Urocap V uroflowmeter patient safety
Clean Urocap V to remove urine contamination using approved methods
Ranked causes
- 1. Urine contamination on Urocap V surfaces (common)
Verify Urocap V is not soaked or immersed
- Test point
- Urocap V body
- Expected
- Device remains dry internally; damp cloth used only externally
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Computer or printer cannot power on
Computer or printer fails to power on
Ranked causes
- 1. No power on outlet or damaged/disconnected power cord (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device displays partial charge when battery fully charged
Battery shows partial charge after left in charger for more than 5 hours
Ranked causes
- 1. System needs restart (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
EMG response too high and channel saturated
EMG signal exceeds channel capacity, no further signal capture
Ranked causes
- 1. Electrodes not picking up proper muscle group (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error Code 66a: .NET Framework installation failure
Software returns Error Code 66a indicating KB2160841 cannot be installed
Ranked causes
- 1. .NET Framework file needs repair (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error message after system restart: devices not found
Error appears on screen after system restarts indicating devices not detected
Ranked causes
- 1. Temporary connectivity issue during boot (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No spike response in EMG and baseline flat
EMG signal shows no activity, baseline remains flat
Ranked causes
- 1. Electrodes not picking up proper muscle group (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
PC does not power on after power outage
Computer fails to turn on following power interruption
Ranked causes
- 1. Power outage left system in hung state (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pediatric pump settings incorrect patient safety
Pump settings not configured for pediatric patient
Ranked causes
- 1. Pump switched to adult settings instead of pediatric (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump calibration incorrect
Pump calibration values out of specification
Ranked causes
- 1. Pump requires recalibration (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump not running
Pump fails to start or operate
Ranked causes
- 1. Pump not calibrated (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump pressure limit reached patient safety
Pump stops due to pressure exceeding limit
Ranked causes
- 1. High pressure in bladder circuit (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump rate set at 1ml/min not responding
Pump set to minimum rate fails to operate correctly
Ranked causes
- 1. Software state issue at minimum pump rate (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump volume reading incorrect
Volume channel displays value outside ±10 ml tolerance when pump fills to 150 cc
Ranked causes
- 1. Pump off calibration (common)
Measure actual volume output when pump fills to 150 cc
- Test point
- Volume channel
- Expected
- 150 cc (+/- 10 ml), +/- 10 ml
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Red light on bubble detector sensor lit
Bubble detector red indicator light illuminated without visible bubbles
Ranked causes
- 1. Pump tubing not inserted securely into sensor groove (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Roam™ DX low battery or disconnected
LED lights indicate broken connection and/or low battery status on wireless device
Ranked causes
- 1. Low battery or disconnected connection (common)
Check LED light status
- Test point
- LED lights on Roam™ DX device
- Expected
- Normal operation indicator (refer to Equipment Status and Light Signals section page 25)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Script buttons stopped working
Control panel script buttons (e.g., Uroflow Summary, Summary Review) non-functional
Ranked causes
- 1. Program file for formatting buttons not using AutoHotKey.ahk format (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touchscreen on tower frozen
Central Hub Computer touchscreen not responding to input
Ranked causes
- 1. System requires reset (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transducer change event marked on graph
Marker appears on signal graph indicating cable connection change
Ranked causes
- 1. Cable connected or disconnected from device (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to connect devices to Bluetooth at first setup
Initial Bluetooth pairing fails between device and system
Ranked causes
- 1. Connection broken during initial setup (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to connect to PC via touchscreen
Touchscreen cannot establish connection to personal computer
Ranked causes
- 1. Connection broken between touchscreen and PC (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to connect via Bluetooth
Bluetooth wireless connection fails to establish or maintain
Ranked causes
- 1. Connection broken due to distance or obstruction (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Uroflowmeter calibration failure: volume delta out of range
Calibration routine fails with volume change measurement out of range
Ranked causes
- 1. Less than 500mL of water poured into beaker (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Uroflowmeter flow reading incorrect
Volume channel displays value outside ±5 ml tolerance when 500 ml water poured at 25-50 ml/second rate
Ranked causes
- 1. Uroflowmeter off calibration (common)
Measure volume displayed when 500 ml water poured at controlled rate
- Test point
- Volume channel
- Expected
- 500ml (+/- 5 ml), +/- 5 ml
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Uroflowmeter volume reading incorrect
Volume channel displays value outside ±5 ml tolerance when 500 ml water placed on transducer
Ranked causes
- 1. Uroflowmeter off calibration (common)
Measure volume displayed on Uroflow transducer with 500 ml water
- Test point
- Volume channel
- Expected
- 500ml (+/- 5 ml), +/- 5 ml
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Volume limit exceeded patient safety
Fluid volume in pump circuit exceeds maximum
Ranked causes
- 1. VH2O channel not zeroed (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Water or urine leaked on electrodes patient safety
Electrodes contaminated with water or urine
Ranked causes
- 1. Liquid contamination of electrodes (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.