Konica Minolta S-SRM: troubleshooting and error codes
The ServiceForge knowledge base holds 2 documented symptoms for the Konica Minolta S-SRM, with 3 ranked causes, 17 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Hand SW signal voltage not measurable between X-ray Console and Hand SW high voltage
Unable to detect voltage change when Hand SW is pressed; multimeter shows no signal variation
Ranked causes
- 1. FG and grounding from Hand SW may be isolated (occasional)
- 2. Connection error or short circuit in check jig (common)
Set all DS1 and DS2 dip switches to ON position
- Test point
- DS1, DS2
- Expected
- All switches ON
8 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Signal voltage measurement between hand switch terminals high voltage
Procedure to measure voltage between hand switch terminals when stable voltage cannot be measured in procedure 5-2. Tests signal levels at TP1 and TP2 test points with SW1 and SW2 in various states (OFF, SW1 ON, both ON).
Ranked causes
- 1. Unable to measure stable voltage in procedure 5-2 (unknown)
Set jig configuration
- Test point
- DS1, DS2
- Expected
- Both ON
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI