FR FieldReady Academy
X-ray imaging consoles · REGIUS family

Konica Minolta CS-1: troubleshooting and error codes

The ServiceForge knowledge base holds 9 documented symptoms for the Konica Minolta CS-1, with 12 ranked causes, 72 diagnostic checks with test points and expected values, and 217 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Calibration: Sensitivity for regular cassette radiation

Correct sensitivity differences between REGIUS 110HQ readers or plate types by exposing a solid density image and processing through Service Tool.

Ranked causes

  1. 1. New plate type with different sensitivity group introduced (common)
  2. 2. X-ray generator device changed (occasional)
  3. 3. Calibrating sensitivity difference on each REGIUS 110HQ (occasional)
  4. 4. Plate type requires sensitivity settings different from regular cassette (common)
First check

Expose solid density image on cassette

Test point
Cassette (largest size at site)
Expected
Solid density image created

15 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AEC adjustment for C-Series Mammo Cassette Plate radiation

Automatic Exposure Control (AEC) must be adjusted when switching from RP-6M110 to CP-1M200 cassette due to lower transmission factor. Incorrect settings cause excessive patient X-ray dose or misdiagnosis.

Ranked causes

  1. 1. CP-1M200 cassette has 60% lower transmission factor than RP-6M110 (common)
First check

Measure average mammographic dose using 156 Phantom at reference exposure condition

Test point
156 Phantom standard position
Expected
2mGy or less for contact mammogram
Tools
Dosimeter

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Mammo system setup patient safety

Set up mammo system with high sharpness plate and high density film by configuring CS-3, printer, REGIUS unit, exposure device and operation checks.

Ranked causes

  1. 1. Mammo system requires dedicated calibration for plate version and printer settings (common)
First check

Verify CS-3 normal setting configured per Installation/Service manual

Test point
CS-3 configuration
Expected
Normal mammo system settings applied

4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Processing parameters setup for C-Series Mammo Cassette Plate

Image processing parameters must be configured according to C-series mammo plate specifications to achieve proper sharpness and image quality for mammary gland structure and calcified image visualization.

Ranked causes

  1. 1. C-series mammo plates require specific processing parameter settings for optimal image quality (common)
First check

Set G-value parameter

Test point
Processing parameter G-value
Expected
3.83

5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: REGIUS 170/190 plate version and uniformity correction

Verify that the plate version used for cassette calibration matches the actual examination plate, and check uniformity correction table for compatibility. If plate versions differ, recalibration is required.

Ranked causes

  1. 1. Plate version mismatch between calibration and examination (common)
First check

Access uniformity correction table

Test point
Service Tool screen Reader function
Expected
Service Tool screen (Reader) opens

7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: REGIUS 170/190 sensitivity correction

Verify that the plate version used for cassette sensitivity calibration matches the actual examination plate. If plate versions differ, recalibration is required.

Ranked causes

  1. 1. Plate version mismatch in sensitivity correction table (common)
First check

Access sensitivity calibration interface

Test point
Sensitivity button under Calibration
Expected
Reader Device screen displayed

6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Sensitivity calibration of regular cassette radiation

Compensate sensitivity differences between REGIUS 110 units or when X-ray generator condition changes or new plate version introduced

Ranked causes

  1. 1. X-ray generator condition altered or new plate version introduced (common)
First check

Align cassette long side with vertical direction of non-grid X-ray tube

Test point
X-ray tube and cassette orientation
Expected
Long side vertical

9 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Sensitivity for mammo cassette radiation

Correct sensitivity differences for mammo cassettes (RP-6M, CP-1M) using tungsten X-ray tube with dose adjustment coefficients for time elapsed.

Ranked causes

  1. 1. New mammo plate type introduced requiring sensitivity adjustment (common)
First check

Prepare X-ray exposure for mammo cassette

Test point
X-ray tube (must be tungsten filament)
Expected
Tungsten tube selected (not mammo exposure device)

17 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Unevenness High and Standard Resolutions radiation

Regular calibration procedure to maintain image quality; requires calibration at both High (87.5µm) and Standard (175µm) resolutions

Ranked causes

  1. 1. Routine maintenance: both resolutions require calibration for normal case (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
04001HP1 Sensor NG(Transport Unit)error
04002HP1 Sensor NG(Transport Unit)error
04011HP2 Sensor NG(Transport Unit)error
04012HP2 Sensor NG(Transport Unit)error
04013HP2 Sensor NG(Transport Unit)error
04014HP2 Sensor NG(Transport Unit)error
04015HP2 Sensor NG(Transport Unit)error
04020Alignment Sensor NG(Start)error
04021Alignment Sensor NG(Feeder)error
04022Alignment Sensor NG(Adhesion)error
04023Alignment Sensor NG(Shelter)error
04024Alignment Sensor NG(Discharge)error
04025Alignment Sensor NGerror
04026Alignment Left-side Sensor NG(Initialize)error
04027Alignment Right-side Sensor NG(Initialize)error
04030Elevator Detective Sensor NG(Start)error
04031Elevator Detective Sensor NG(Feeder)error
04032Elevator Detective Sensor NG(Adhesion)error
04033Elevator Detective Sensor NG(Shelter)error
04034Elevator Detective Sensor NG(Before Elevating)error
04035Lock HP Sensor NGerror
04036Elevator HP Sensor NGerror
04037Elevator Detective Sensor NG(After Elevating)error
04041Feeder's Shutter Opening Sensor NGerror
04042Feeder's Shutter Closing Sensor NGerror
04043Feeder's Shutter Opening Sensor NGerror
04044Feeder's Shutter Closing Sensor NGerror
04045Cassette Back Plate Sensor NGerror
04046Backboard's Warp Detective Sensor NGerror
04047Backboard's Adhesion Detective Sensor NGerror
04048Ejection Shutter Close Sensor NGerror
04061Slow Scan HP Sensor NGerror
04062Backboard's Adhesion Detective Sensor NGerror
04064V-Sync Sensor NGerror
04065Backboard's Adhesion Detective Sensor NGerror
04067Backboard's Adhesion Detective Sensor NGerror
04068Backboard's Adhesion Detective Sensor NGerror
04073Cassette Back Plate Sensor NG(Initialize)error
04074Backboard's Warp Detective Sensor NG(Initialize)error
04075Backboard's Adhesion Detective Sensor NG(Initialize)error
0407624V Power NG(Usually)error
0407724V Power NG(Door Close)error
04101Slow Scan Motor NGerror
04102Transport Unit Motor NGerror
04103Elevator Motor NGerror
04104Alignment Motor NGerror
04105Feeder's Shutter Open/Close Timeouterror
04106Lock Timeouterror
04107Load Motor NGerror
04150Lock Initial Processingerror
04151Feeder's Shutter Initial Processingerror
04152Elevator Initial Processingerror
04153Alignment Initial Processingerror
04154Transport Unit Initial Processingerror
04155Slow Scan Initial Processingerror
04156Transfer Action 1 Starting NGerror
04157Transfer Action 2 Starting NGerror
04158Feeder's Discharge Action Starting NGerror
04159Discharge Action Starting NGerror
04160Feeder's Shutter Opening NGerror
04161Feeder's Shutter Closing NGerror
04162Lock Starting NGerror
04163Cassette Falling Detective Time NGerror
04164No Notice of Permission on Erase-lamp ONerror
04165No Notice of Permission on Erase-lamp OFFerror
04167Load Motor Initial Processingerror
04200GetStatus Receiving Timeout(Initialize)error
04201GetStatus Receiving Timeout(Ready)error
04202GetStatus Receiving Timeout(Detect Barcode)error
04203GetStatus Receiving Timeout(High-voltage ON)error
04204GetStatus Receiving Timeout(Laser ON)error
04205GetStatus Receiving Timeout(Cancel Barcode)error
04206GetStatus Receiving Timeout(Processing)error
04207GetStatus Receiving Timeout(Pause)error
04208GetStatus Receiving Timeout(Standby)error
04209GetStatus Receiving Timeout(Close Front Door)error
04210GetStatus Receiving Timeout(Error)error
04211GetStatus Receiving Timeout(Standby)error
04212GetStatus Receiving Timeout(Transporting of Erase Mode)error
04213GetStatus Receiving Timeout(Power off after Ready)error
04216GoBackward Receiving Timeouterror
04217Laser ON without Sending High-voltage ON(1)error
04218Laser ON without Sending High-voltage ON(2)error
04219Invalid Commanderror
04221StartSubScan Receiving Timeouterror
04900Cassette Exists(Elevator)error
04902Cassette Exists(Open Door of Discharge)error
04903Cassette Exists(Backboard Adhesion)error
04904Turn Off Erase-lamps(2)error
04905SCB Communication Errorcritical
04906Warped Cassetteerror
04907Cassette Opening Errorerror
04908Falling Cassetteerror
04909Casette Gap Detection(Back plate magnetize)[04909]error
04910Casette Gap Detection(Casette peeling)[04910]error
04914Opening Extraction (Initialize)error
04917Feeder's Shutter Opening NG(Unlock)error
04918FAN Disconnection NG 1error
04920FAN Disconnection NG 3error
04921Erase-lamp NGerror
12010Serial communication Error (Mech Resend Overflow)error
12012Serial communication Error (Mech Overrun)error
12013Serial communication Error (Mech Timeout)error
12020Msg Error (Mech illegal Msg)error
12210Serial communication Error (Net Resend Overflow)error
12212Serial communication Error (Net Overrun)error
12213Serial communication Error (Net Timeout)error
14000Signal Proc Error (Gain Overflow)error
14001Signal Proc Error (Gain Underflow)error
14002Signal Proc Error (Offset Overflow)error
14003Signal Proc Error (Offset Underflow)error
14010Signal Proc Error (QRV Overflow)error
14011Signal Proc Error (QRV Underflow)error
23000Program Error (Main Control)critical
23200Program Error (Data File Size)critical
23900Program Error (Firmware Update)critical
24000Network Error (JM Connection)error
25000Network Error (Non-Timeout)error
25100Network Error (Timeout:in General Communication)error
25200Network Error (Timeout:in Image Transmission)error
26000Operation Error (Invalid Job)error
26010Operation Error (Invalid Job Parameter)error
26100Operation Error (Modeless Job)error
26200Operation Error (Invalid Console)error
26300Operation Error (Reader Busy)error
26900Operation Error (PowerOFF under Job Execution)critical
6000Justifier HP sensor NG (Move to HP)error
6001Justifier Motor Control NG (Move to HP)error
6002Justifier HP sensor NG (Cassette Fixed position)error
6003Justifier Motor Control NG (Cassette Fixed position)error
6004Justifier Motor Control NG (Cassette Support Position)error
6005Justifier Motor Control NG (Cassette Fixed position -2)error
6006Justifier HP sensor NG (Move to HP -2)error
6007Justifier Motor Control NG (Move to HP -2)error
6008Justifier HP sensor NG (Move to HP -3)error
6009Justifier Motor Control NG (Move to HP -3)error
6010Justifier HP sensor NG (Cassette Support Position)error
6011Justifier Motor Control NG (Cassette Support Position -2)error
6050Reciever HP sensor NG (Move to HP)error
6051Elevator Motor Control NG (Move to HP)error
6052Reciever HP sensor NG (Position 1)error
6053Elevator Motor Control NG (Position 1)error
6054Elevator Motor Control NG (Position 2)error
6055Elevator Motor Control NG (Slight descent at Position 2)error
6056Elevator Motor Control NG (Position 1 -2)error
6057Reciever HP sensor NG (Move to HP -2)error
6058Elevator Motor Control NG (Move to HP -2)error
6059Reciever HP sensor NG (Move to HP -3)error
6060Elevator Motor Control NG (Move to HP -3)error
6061Reciever HP sensor NG (Move to Position 2)error
6062Elevator Motor Control NG (Position 2 -2)error
6100Transporter Sensor NG (Reading Position)error
6101Transporter Motor Control NG (Reading Position)error
6102Transporter Sensor NG (Move to HP)error
6103Transporter HP sensor NG (Move to HP)error
6104Transporter Motor Control NG (Move to HP)error
6105Transporter HP sensor NG (Pressing Position)error
6106Transporter Motor Control NG (Pressing Position)error
6107Transporter Sensor NG (Reading Position 2)error
6108Transporter Motor Control NG (Reading Position 2)error
6109Transporter Sensor NG (Pressing Position 2)error
6110Transporter Motor Control NG (Pressing Position 2)error
6111Transporter Motor Control NG (HP Passing Position)error
6112Transporter HP sensor NG (Move to HP -2)error
6113Transporter Motor Control NG (Move to HP -2)error
6150Encoder Initializing NGerror
6151Subscan Motor Control NG (Mechanical HP)error
6152Subscan Motor Control NG (Reading Completion Position)error
6153Subscan Motor Control NG (Mechanical HP -2)error
6200Open Insertion Shutter NGerror
6201Close Insertion Shutter NGerror
6250Motor Lock Release NG (Initialize)error
6251Motor Lock Release NG (Lock)error
6252Motor Lock Release NG (Release)error
6300Justifier Sensor NG (Fixed Action)error
6301Justifier Sensor NG (Fixed)error
6302Casette Exists(Sensor Justifier)error
6303Casette Exists (Sensor Justifier)error
6350Reciever Sensor NG (Feeder)error
6351Reciever Sensor NG (Send)error
6352Reciever Sensor NG (Discharge)error
6353Reciever Sensor NG (Removed)error
6354Casette Exists (Reciever)error
6355Casette Exists (Reciever)error
6400Back Plate detection Sensor NG (Before moving to Reading Position)error
6401Back Plate detection Sensor NG (Reading Position is Stopping)error
6402Back Plate detection Sensor NG (After moving to Pressing Position)error
6450Peel Sensor Plate (Reading or Erasing)error
6451Peel Sensor NG (Non-operation)error
6452Peel Sensor NG (Start)error
6453Peel Sensor NG (Initialize)error
6500Cassette Insertion Sensor NG (Initialize)error
6501Cassette Insertion Sensor NG (Load)error
6502Cassette Insertion Sensor NG (Discharge)error
6503Back Plate Casette detectionerror
6504Cassette Insertion Sensor NG (Discharge -2)error
6550Polygon Rotation NGerror
6551Not recieving Reading Completion signalerror
65522 Erase lamps turn off (Start)error
65532 Erase lamps turn off (Erase)error
65541 Erase lamp turns off (Start)error
6600Cooler fan's digital power supply disconnectederror
6601Eraser cooling fan blowouterror
6602Digital power supply cooling fan blowouterror
6603MDU cooling fan blowouterror
6650FPGA Error (Event Driver Access failed)critical
6651FPGA Error (Fail to access MDU)critical
6652FPGA Error (Fail to access LMC)critical
6653FPGA Error (I/O Control failed)critical
6654FPGA Error (Fail to open Image Driver)critical
6655FPGA Error (Fail to access PMU)critical
6750Reception desk board opening and shutting position sensor NGerror
6751Reception desk board HP location sensor NGerror
6765Reception desk board opening and shutting position sensor NGerror
6766Reception desk board HP location sensor NGerror
6780Reception desk board opening and shutting position sensor NGerror
6781Reception desk board HP location sensor NGerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.