Kodak DryView 8700: troubleshooting and error codes
The ServiceForge knowledge base holds 93 documented symptoms for the Kodak DryView 8700, with 195 ranked causes, 150 diagnostic checks with test points and expected values, and 188 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
EC85: Image memory is full
Insufficient image memory to store or process image; normal condition but may indicate operational issues
Ranked causes
- 1. Film pickup difficulties causing repeated pickup attempts (occasional)
- 2. Out of film not noticed by operator until memory full (common)
- 3. System pushed to operational limits (occasional)
- 4. Memory management setting not optimized (occasional)
- 5. User priority not set appropriately (rare)
- 6. Siemens interface delay parameters causing slow response (rare)
- 7. Host MemFull message setting incorrect (occasional)
- 8. Film pickup problems from jams or other issues (occasional)
Verify available memory capacity
- Test point
- MPC Image/Show or System Status screen
- Expected
- Available memory space greater than largest anticipated image size
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
EC68: Timeout waiting for pixels from EIB or IB (digital)
Image acquisition fails; error logged to system; subsequent acquires may succeed
Ranked causes
- 1. No power to DEIB/DIB (occasional)
- 2. DEIB loses power and parameter set after host power-down (occasional)
- 3. Host system not ready with digital data in correct mode (occasional)
- 4. Siemens camera board tri-stating STROBE signal (rare)
- 5. DEIB/DIB parameters incorrect or not downloaded (occasional)
- 6. Host reply timing slower than host wait timeout specification (occasional)
- 7. 8700/8500 IMAGER cannot process data faster than 4 Mpixels/second (occasional)
Verify DEIB/DIB power LED is lit
- Test point
- DEIB/DIB front panel POWER LED
- Expected
- LED lit
2 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
EIB not detected by IPB
Error code EC39 logged; input module sends command to DEIB or VEIB on fiber A but does not receive response on fiber B; imager cannot initialize EIB
Ranked causes
- 1. EIB is locked up (common)
- 2. No power to the EIB (common)
- 3. Wrong type EIB selected in MPC (occasional)
- 4. Fiber optic cable is bad (occasional)
- 5. DEIB transfer clock speed mismatch (occasional)
- 6. EIB Parameter Write Error (occasional)
- 7. Siemens camera board tri-stating strobe signal (rare)
Verify EIB power indicator lights
- Test point
- EIB power lights
- Expected
- Power lights on
2 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
EC68: Timeout waiting for pixels from EIB or IB (video)
Image acquisition timeout with video source; IMS did not receive full number of pixels within timeout period
Ranked causes
- 1. Video cable disconnected or bad (common)
- 2. VEIB parameter settings incorrect; digitization during sync pulse (occasional)
- 3. Video signal unstable due to intermittent connection (occasional)
- 4. Insufficient video amplitude in VEIB for consistent digitization (occasional)
- 5. VEIB number of passes parameter incorrect for 8700/8500 IMAGER (occasional)
- 6. PLL not fully seated; plastic standoffs too long (rare)
Inspect and test video cable
- Test point
- Video cable from VEIB/VIB to source and to IMAGER
- Expected
- Cable intact, properly connected, no damage
1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Jam in processor caution
Media jammed in processing drum. Processor entrance sensor triggered but densitometer sensor did not. Message EC542: 'Media jam at processor. Lift top cover and remove all media in area 4.'
Ranked causes
- 1. Film adhering too aggressively to drum at drum/stripper interface (common)
- 2. Stripper gap larger than 7 mils at corners (common)
- 3. Intermittent processor entry sensor toggle causing film stall in cooling section (occasional)
- 4. Hump or debris in cooling section felt pad causing jam (occasional)
- 5. Processor motor shorted MIB drive board or wiring harness rubbing frame (rare)
- 6. Developer housing not closing properly, idler rollers not fully contacting drum (occasional)
Measure stripper gap at processor drum/stripper interface
- Test point
- Stripper edge at four corners of drum
- Expected
- 0.007 inches ±0.001 inches, ±0.001 inches
1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SCB unable to detect EOL signal from optics module caution
SCB not receiving expected EOL (end-of-line) signal from optics for each line during imaging; error L126 OPTICS_NO_EOL logged
Ranked causes
- 1. IPB parity error in image data halts optics transmission (occasional)
- 2. Loose or bad connection between IPB and Optics or Optics and SCB (common)
- 3. Control cable bulkhead connection loose (occasional)
- 4. Bad fiber optic cable between 8800 and imager (rare)
- 5. Bad optics module (occasional)
Set clear border lines and power cycle imager
- Test point
- MPC/SCB/Config/Clear Border lines
- Expected
- Number of lines matches between SCB and IPB; imaging resumes normally
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System does not communicate between IPB and SCB boards
IPB cannot communicate with SCB, keypads, or MPC; EC910 declared on local panel; communication timeout after 45 seconds
Ranked causes
- 1. IPB board locked up or unresponsive (common)
- 2. IPB/SCB communication parameters misconfigured (occasional)
- 3. Peripheral device (MPC, keypad, or SCB) interfering with IPB (occasional)
- 4. IPB power supply voltage out of specification (occasional)
- 5. IPB Xilinx EPROM defective (rare)
Press reset button on IPB
- Test point
- IPB reset button
- Expected
- EC910 clears and communication restored
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Film jam in exposure area
Local panel displays 'Media jam in platen. Open left door and remove jam in area 3' (P164). Host displays 'K211: Media is jammed in laser imager X'. Error log shows 'L164 JAM_EXPOSURE'. Media jammed in exposure area while entering or leaving platen. Platen entry sensor did not turn off or platen bottom/exit sensor did not come on.
Ranked causes
- 1. Platen entry sensor loose or rounded mounting legs (common)
- 2. Shipping screw left in optics (occasional)
- 3. Optics misaligned or pushed towards rear (occasional)
- 4. Platen kick solenoid nylon bushings too tight (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Supply cartridge could not be closed caution
Cartridge failed to close during rollback process. Operator message warns of exposed sheets if door opened.
Ranked causes
- 1. Film jam in cartridge (common)
- 2. Tongue depressor arm bent (occasional)
- 3. Transport interlock SW407 faulty (occasional)
- 4. MIB driver U8C1 blown; motor leads shorted (rare)
Inspect cartridge for visible film jam
- Test point
- Cartridge interior
- Expected
- No film obstruction during closure
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Supply cartridge could not be opened caution
Cartridge failed to open during rollback process. Local panel message states film jam possible. Tongue depressor sensor failed to go high.
Ranked causes
- 1. Film jam in cartridge (common)
- 2. Tongue depressor arm bent (occasional)
- 3. Transport interlock SW407 faulty (occasional)
- 4. MIB driver U8C1 blown; motor leads shorted (rare)
Inspect cartridge for visible film jam
- Test point
- Cartridge interior
- Expected
- No film obstruction
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Processor drum brushes high voltage
Brushes must maintain full contact with slip rings; insufficient height or misalignment causes arcing and power loss to RPB.
Ranked causes
- 1. Brush height below minimum specification (common)
- 2. Brush misalignment with slip rings (common)
- 3. Arcing damage to brushes or slip rings (occasional)
Verify brush alignment side-to-side and front-to-back with respect to slip rings
- Test point
- Brushes and slip rings
- Expected
- Alignment as shown in Figure 3-17; no sign of arcing (burns)
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot connect with the MPC
System unable to establish communication with MPC (Multi-Purpose Controller)
Ranked causes
- 1. MPC and IMAGER/8800 baud rates do not match (common)
- 2. IPB MPC port is locked up (occasional)
- 3. IPB to MPC communication parameters are invalid (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dpatch density out of control limits (midtone fast)
Dpatch test patch density outside control limits; inconsistent recorded densities in log
Ranked causes
- 1. Densitometer mounting bracket bent, causing timing error (common)
- 2. Dmax setting incorrect (8800 only) (occasional)
- 3. Dpatch position out of specification on film (occasional)
Review Dpatch log from MPC/Log/Printer Log/Dpatch
- Test point
- Dpatch log in MPC
- Expected
- Consistent density readings
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dpatch density out of control limits (midtone slow)
Dpatch test patch density outside control limits; inconsistent readings in Dpatch log
Ranked causes
- 1. Densitometer mounting bracket bent, causing timing error (common)
- 2. Dmax setting incorrect (8800 only) (occasional)
- 3. Dpatch position out of specification on film (occasional)
Retrieve Dpatch log from MPC/Log/Printer Log/Dpatch
- Test point
- Dpatch log in MPC
- Expected
- Consistent density readings across Dpatch entries
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
EC67: Too many pixels received from EIB
Acquisition fails because more pixels received than expected from EIB/IB; FIFO overflow
Ranked causes
- 1. Parameter settings incorrect; VEIB digitizing during sync pulse (common)
- 2. Video signal not stable; intermittent connection (occasional)
- 3. Insufficient video amplitude from double gain parameter or VEIB gain pot (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Film density too high after calibration
Film speed above 3.4 from calibrate sheet reading; automatic image quality control off
Ranked causes
- 1. Film cartridge defective or poor quality (common)
- 2. Processor temperature incorrect (occasional)
- 3. Beam Power Monitor board contaminated or obstructed (occasional)
Try different film cartridge
- Test point
- Film cartridge
- Expected
- Film speed 3.4 or below
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Film density too low after calibration
Darkest density step 26 below 3.1 for blue film or 3.0 for clear film; automatic image quality control may be off
Ranked causes
- 1. Film cartridge defective or poor quality (common)
- 2. Processor temperature incorrect (too low or too high) (occasional)
- 3. Densitometer reading invalid or miscalibrated (occasional)
Try different film cartridge or lot
- Test point
- Film cartridge
- Expected
- 3.1 or 3.0 density achievable
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Film jammed on multifeed detector
Local panel displays 'Media jam at supply cartridge. Open left door and remove jam in area 1' (P161). Host displays 'K211: Media is jammed in laser imager X'. Error log shows 'L161 JAM_SUPPLY'. Film is jammed on the multifeed detector.
Ranked causes
- 1. Films stick together (common)
- 2. Multifeed board out of adjustment (occasional)
- 3. Debris in multifeed rollers (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Film pickup problems
Local panel displays 'Imager has difficulty picking up media. Replace cartridge' (P166). Host displays 'K211: Media is jammed in laser imager X'. Error log shows 'L166 PICKUP_TRY', 'L166 DOUBLE_PICKUP' or 'L166 NO_PICKUP'. One or more pickup attempts have failed.
Ranked causes
- 1. Films stick together or mispositioned (common)
- 2. Weak solenoid (early machines) (rare)
- 3. Multifeed board misadjusted (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Film speed near lower calibration limit
Film speed between 2.6 and 2.7; may indicate film variability or other factors
Ranked causes
- 1. Film variability or marginal cartridge quality (common)
- 2. Processor temperature out of specification (occasional)
- 3. Beam Power Monitor board contaminated (occasional)
Try different film cartridge or lot
- Test point
- Film cartridge
- Expected
- Film speed above 2.7
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Film speed near upper calibration limit
Film speed between 3.3 and 3.4; may indicate film variability or other factors
Ranked causes
- 1. Film variability or marginal cartridge quality (common)
- 2. Processor temperature out of specification (occasional)
- 3. Beam Power Monitor board contaminated (occasional)
Try different film cartridge or lot
- Test point
- Film cartridge
- Expected
- Film speed below 3.3
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Interlock switch fails or returns abnormal state patient safety
Safety interlock switch does not respond correctly to SCB commands; watchdog timer shutdowns or intermittent safety malfunctions occur.
Ranked causes
- 1. Interlock switch receives SCB low signal and returns high (short condition) (common)
- 2. Interlock switch mechanically stuck or damaged (occasional)
- 3. SCB CPU locked up; watchdog timer activates (rare)
Monitor SCB command to interlock switch
- Test point
- Interlock switch signal line at SCB
- Expected
- SCB sends High signal when door is closed; Low signal when checking interlock status
- Tools
- Logic probe or oscilloscope
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
IPB unable to load parameters to EIB
Error code EC59 logged; IPB cannot write parameter set to EIB/IB, or EOT (End of Transmission) not received from host during acquisition
Ranked causes
- 1. EIB locked up or not responding (common)
- 2. MPC not configured for correct EIB type (occasional)
- 3. Siemens camera board tri-stating strobe signal (rare)
Power cycle EIB and IMAGER to clear locked state
- Test point
- Power switch on both EIB and IMAGER
- Expected
- Both units power off and on successfully; EIB LED indicators show normal startup sequence
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No End of File timeout during film imaging
Error code EC48 logged; IPB times out waiting for EOF signal from optics and SCB after issuing print command; typically accompanied by film jam, parity error, or optics error
Ranked causes
- 1. Clear Border Lines not initialized in MPC (common)
- 2. Film jam in imaging chain (common)
- 3. Optics module malfunction (occasional)
Verify Clear Border Lines initialization on every film type including density tests
- Test point
- MPC/SCB/Config menu Clear Border Lines field
- Expected
- Clear Border Lines enabled and initialized
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Optics module PLL error patient safety
Local panel displays 'Printer malfunction' (P151). Host displays 'K202: Laser imager X is not operational'. Error log shows 'L151 OPTICS_OUT_OF_LOCK'. XPLL line from optics module does not pull low before print, indicating X galvo is not in sync.
Ranked causes
- 1. Optics module failure (occasional)
- 2. Loose control cable at bulkhead (occasional)
- 3. SCB circuit malfunction (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Platen door interlock open high voltage
Exposure cover is open, preventing laser operation. Console displays EC173. LED DS6K3 shows amber when power is present.
Ranked causes
- 1. Platen door not fully closed or latch not fully clockwise (common)
- 2. Platen door interlock switch faulty (occasional)
- 3. Left door laser interlock open or faulty (occasional)
Verify platen door position and latch engagement
- Test point
- Platen door latch
- Expected
- Fully clockwise
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rollback motor failed
Error code EC227; rollback motor not operating during cartridge load process
Ranked causes
- 1. Harness to rollback motor incorrectly mounted, rubbing on frame (common)
- 2. Rollback home sensor failure (occasional)
- 3. Rollback rack sensor failure (occasional)
Close cartridge and open left side door to inspect rollback bar area
- Test point
- Rollback bar mechanism
- Expected
- No jams; rollback bar can rotate to home position (arrow up)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Supply cartridge empty
Local panel displays 'Cartridge is empty' (P134). Host displays 'K218. Supply cartridge is empty on laser imager 1'. Supply door solenoid is energized to open door. SCB has counted 125 sheets since cartridge was loaded.
Ranked causes
- 1. Cartridge depleted (common)
- 2. Film jam in cartridge (occasional)
- 3. Film Out or Pickup Bottom Sensor malfunction (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bad densitometer data
Dmin of film greater than 0.3 or calibration sheet steps not monotonically increasing in density. May be caused by expired/fogged film, out-of-spec storage temperatures, or high developer temperature.
Ranked causes
- 1. Fogged film (occasional)
- 2. Film stored at out-of-spec temperatures, past expiration, or developer temp out of spec (occasional)
Try another film cartridge
- Test point
- Film cartridge
- Expected
- EC624 error cleared with different film
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Power supply voltages high voltage
Verify and adjust DC output voltages from PS901 (+5V, +17V, -17V, +24V) and PS902 (-5.2V) supplies to ensure they remain within specification.
Ranked causes
- 1. PS901 output voltages out of specification (unknown)
- 2. PS902 output voltage out of specification (unknown)
Measure +5V output from PS901
- Test point
- TP5H1 to TP7G1
- Expected
- +5.1 ± 0.1 VDC, ±0.1 V
- Tools
- Digital Voltmeter
5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cartridge open sensor failed
Error code EC225; cartridge open (rollback rack) sensor not changing states when cartridge opened or closed
Ranked causes
- 1. Harness to rollback motor incorrectly mounted, rubbing on frame (common)
- 2. MIB board chip U8C1 failure (common)
Power cycle the machine
- Test point
- System startup
- Expected
- Error clears and cartridge operates normally
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Communication failure between 8800 Multi-Input Manager and 8700/8500 imager
SCB declares EC910 when 8800 does not send command within 45 seconds; communication failure between systems
Ranked causes
- 1. Output module configuration incorrect for 8700/8500 system (common)
- 2. Communication link broken or improperly terminated (occasional)
Set Output Module config to 8700/8500 printer
- Test point
- MPC Output module configuration menu
- Expected
- Config set to '8700/8500 printer'
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code: Arcing at processor drum brushes high voltage
Arcing detected at brush-to-slip-ring interface, causing intermittent or lost power to RPB.
Ranked causes
- 1. Brush misalignment or poor contact (common)
- 2. Worn brush surfaces or slip rings (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Film jam before processor
Local panel displays 'Media jam in transport. Open left door and remove jam in area 2b' (P165). Host displays 'K211: Media is jammed in laser imager X'. Error log shows 'L165 Jam to Processor'. Media jammed after exposure. Platen exit sensor came on but mid transport or processor entry sensor did not come on.
Ranked causes
- 1. Mid transport or processor entry sensor loose (common)
- 2. Drive motor set screw loose (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flag-type sensor signal incorrect or open
Flag-type sensor output does not match expected logic states; sensor may be open, disconnected, or providing inverted signal.
Ranked causes
- 1. Flag sensor disconnected or open circuit (occasional)
- 2. Flag sensor actuator or flag stuck or damaged (occasional)
Measure flag sensor output voltage when physically blocked (deactuated)
- Test point
- SCB sensor cable for flag sensor
- Expected
- High state: 5.0 volts V
- Tools
- Voltmeter
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invalid laser dynamic range high voltage
If Laser Dynamic Range outside 200 to 4000 range, EC604 declared. Machine must be power cycled to clear error.
Ranked causes
- 1. Optics module is bad (common)
- 2. Loose control cable connection at firewall (occasional)
Check Beam Power Monitor aperture for dust or debris
- Test point
- BPM aperture chute
- Expected
- Clean, no dust or debris
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invalid power monitor offset high voltage
Beam power monitor reading taken with no laser beam hitting BPM board. Reading should be 13±3. If BPM offset outside 1 through 25 reading, EC601 declared.
Ranked causes
- 1. Bad BPM board (occasional)
- 2. Bad peak detector circuit on SCB (occasional)
Check BPM offset value at power up
- Test point
- BPM reading
- Expected
- 13±3 at power up, or 1 through 25 range for valid operation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
MFB actuator adjustment high voltage
Fine-tune actuator position to equalize ADC channel readings
Ranked causes
- 1. Channel 10 reading higher than Channel 9 (occasional)
- 2. Channel 9 reading higher than Channel 10 (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No communication from processor high voltage
Error code EC506; SCB does not receive communication from drum Rotating Processor Board (RPB) after 10 attempts
Ranked causes
- 1. 60 Hz phasing error in wiring to RPB (occasional)
- 2. Slip ring/brush misalignment or arcing (occasional)
Lift and lower top cover to clear error
- Test point
- Top cover
- Expected
- Error clears
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Optics module attenuator error caution
At power up, attenuator cannot move to home position after multiple attempts; local panel reports printer malfunction; error L121 logged
Ranked causes
- 1. Poor cable connection at optics, SCB or MIB (common)
- 2. Bad optics module (occasional)
Test attenuator home position movement via SCB command
- Test point
- SCB command window
- Expected
- Attenuator moves to home position; MPC sensor display shows home position box checked
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Optics module temperature error caution
Optics module fails to reach operating temperature within 15 minutes; local panel displays 'Printer performing Self Test'; error L123 OPTICS_TEMP_FAIL logged
Ranked causes
- 1. Optics module temperature below 46–56°C within 15 minutes (common)
- 2. Poor optics control cable connection (occasional)
Monitor YTempe signal at SCB
- Test point
- SCB resistor R7N3 top leg
- Expected
- Signal high with optics cable disconnected; pulled low by optics module when up to temperature
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pickup middle sensor failed caution
Pickup middle sensor (SW504) failed to operate during pickup cycle run at powerup diagnostics. System declares error and imager not operational.
Ranked causes
- 1. Pickup middle sensor SW504 intermittent or faulty (occasional)
- 2. Pickup assembly vibration causing intermittent sensor operation (occasional)
Test pickup middle sensor operation
- Test point
- SW504 sensor state during pickup cycle test
- Expected
- Sensor triggers high during pickup cycle run
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pickup top sensor failed caution
Pickup top sensor (SW503) failed to operate during pickup cycle run at powerup diagnostics. System declares error and imager not operational.
Ranked causes
- 1. Pickup top sensor SW503 intermittent or faulty (occasional)
- 2. Pickup assembly vibration causing intermittent sensor operation (occasional)
Test pickup top sensor operation
- Test point
- SW503 sensor state during pickup cycle test
- Expected
- Sensor triggers high during pickup cycle run
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Processor drum could not be moved caution
Error code EC501; SCB did not detect drum rotation in last 4 seconds; processor motor powered off
Ranked causes
- 1. Clear plastic lens cap on DS4 photointerrupter emitter/collector missing or loose (common)
- 2. Reflective plate on RPB dirty or deformed (occasional)
Open and close top cover to clear error
- Test point
- Top cover
- Expected
- Error clears; processor resumes operation
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Processor failed to warm up in allotted time
Drum temperature remains below 122.25°C after calculated warm-up period. EC509 generated; top cover must be lifted to clear.
Ranked causes
- 1. Heater fuse blown due to age (common)
- 2. Heat sink for transistor not mounted correctly on Rotating Processor Board (occasional)
Verify heater fuse continuity and condition
- Test point
- Heater zone fuse terminals
- Expected
- Continuous, no visible burn or corrosion
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Service override switch out or fail patient safety
Service override switch is pulled out (electrically closed) or pulsing erratically. SCB interprets state and may trigger P202 or P203 error. P203 causes safety shutdown of motors, solenoids, and local panel.
Ranked causes
- 1. Service override switch in pulled out (service mode) position (common)
- 2. Service override switch pulsing erratically or faulty wiring (occasional)
Verify service override switch position
- Test point
- Service override switch physical position
- Expected
- Switch is pushed in (normal non-service mode)
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tongue depressor sensor failed
Error code EC228; tongue depressor sensor did not change states during cartridge lid opening, causing left door solenoid to energize
Ranked causes
- 1. Tongue depressor arm bent (common)
- 2. Transport interlock SW407 faulty (occasional)
Test tongue depressor sensor with MPC
- Test point
- MPC diagnostic sensors screen
- Expected
- Sensor responds properly during cartridge opening
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
U-type sensor signal incorrect or open
U-type sensor output does not match expected logic states; sensor blocked/unblocked state does not produce expected voltage at SCB.
Ranked causes
- 1. U sensor disconnected or open circuit (occasional)
- 2. U sensor target or flag obstructed or misaligned (occasional)
Measure U sensor output voltage when target is blocked
- Test point
- SCB sensor cable for U-type sensor
- Expected
- Low state: typically 53 mV, typically 53 mV
- Tools
- Voltmeter
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Acquires fail with EC59
Acquisition fails with error code EC59, typically on new setup or when DEIB/DIB parameters are incorrect
Ranked causes
- 1. DEIB/DIB parameters incorrect (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Acquires fail with EC68 or EC59
Acquisition fails with error EC68 or EC59 due to host timeout waiting for IMAGER reply to header data
Ranked causes
- 1. IMAGER reply to host header takes too long; host timing out (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Attenuator test failed high voltage
Optics module attenuator could not rotate to position which gives proper output power. EC605 declared.
Ranked causes
- 1. Optics module is bad (common)
Check Beam Power Monitor aperture for dust or debris
- Test point
- BPM aperture chute
- Expected
- Clean, no dust or debris
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bad film contrast
Contrast of calibrate sheet unacceptable. Densities of steps in relation to each other are out of spec. EC625 declared.
Ranked causes
- 1. Fogged film (common)
Check densitometer adjustment
- Test point
- Densitometer
- Expected
- Densitometer properly adjusted
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Densitometer
A/D Out (Gain) and (Offset) values out of specification; suspected AIQC problems
Ranked causes
- 1. Densitometer gain and offset potentiometers require adjustment (unknown)
Connect MPC to MPC port and launch MPC for Windows; navigate to Operations > Diagnostics > PROC > Densi
- Test point
- MPC Densitometer Test window
- Expected
- A/D Out (Gain) = 1640±100, A/D Out (Offset) = 20±2, ±100 (Gain); ±2 (Offset)
- Tools
- Maintenance Personal Computer (MPC), MPC for Windows software
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Density Patch Offset high voltage
Top edge of density patch is not located 7.0 ± 1.0 mm from top edge of film
Ranked causes
- 1. Offset potentiometer out of adjustment (common)
Measure distance from top edge of test print to top edge of density patch
- Test point
- Density patch on test print
- Expected
- 7.0 ± 1.0 mm, ±1.0 mm
- Tools
- Ruler or caliper
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: MFB Actuator ADC values high voltage
ADC readings from Multifeed Board actuator sensor require adjustment to fall within specification after MFB or hardware replacement.
Ranked causes
- 1. MFB or associated hardware replaced, requiring recalibration (common)
Measure ADC value from channel 9
- Test point
- MPC command 'show adc=9'
- Expected
- Average of first 8 numbers; must be within 200 of adc=10 and within range 1530–1850
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: MFB Actuator R6 adjustment high voltage
Fine adjustment of R6 resistor when ADC range cannot be brought within specification by mechanical actuator positioning alone.
Ranked causes
- 1. ADC range cannot be brought within 1530–1850 by actuator position adjustment (occasional)
ADC range out of specification after actuator adjustment
- Test point
- MPC command 'show adc=9' and 'show adc=10' averages
- Expected
- Values outside range 1530–1850
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Multifeed Board (MFB) Actuator high voltage
Adjust MFB sensor actuator position to balance ADC channel readings after MFB or hardware replacement
Ranked causes
- 1. MFB actuator sensor misaligned after replacement (common)
Measure ADC channel 9 average reading
- Test point
- MPC software Command window, show adc=9
- Expected
- Average of first 8 numbers in range 1000 to 3000
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pickup Arm Height
Adjust pickup arm height after multifeed board actuator adjustment or system control board replacement
Ranked causes
- 1. Multifeed board actuator adjustment performed or system control board replaced (common)
Connect MPC to MPC port on rear of imager
- Test point
- MPC port
- Expected
- Connection established
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Processor Temperature caution
Three temperature zones (left, center, right) in processor must be verified to be within specification range. Procedure performed after processor assembly, drum, RPB, PCB or SCB replacement, or when drum temperatures suspected of causing image quality problems.
Ranked causes
- 1. Processor temperature out of specification (unknown)
Raise top cover of 8700/8500
- Test point
- Top cover
- Expected
- Fully raised
5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Digital acquires work with 969 but fail with IMAGER
Digital acquisition succeeds with 969 IMAGER but fails with 8700/8500 IMAGER due to processor speed differences
Ranked causes
- 1. 8700/8500 IMAGER has lower digital transfer rate limit (4 Mpixels/sec max vs 969 at 6 Mpixels/sec) (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- EC64: IPB detected parity error during acquisition
- EC65: Data Overflow or Underflow
- EC75: Local panel failed
- Film jam in transport to exposure
- Film present sensor failed
- Gain/Black level numbers do not change or work correctly
- HIB lookup table memory failed
- HIB parity error
- HIB register read/write failed
- Invalid densitometer data
- Invalid optical density range
- Invalid power monitor range
- Jam at exit
- Left side door could not be opened
- Left side door interlock failed
- Left side door open
- Motor not receiving correct drive signal
- Motor not responding to enable command
- No supply cartridge detected
- Optics module not present
- Pickup motor failed
- Processor heater zones 1, 2, or 3 failed
- Processor over temp
- Processor returning bad data
- RETRIEVE image procedure takes longer than normal
- Rollback home sensor failed
- SCB receiving data LED status check
- Sensor signal abnormal or inconsistent
- Supply door could not be opened
- Supply door interlock failed
- Top cover interlock failed
- Top cover open
- Video Gain/Black Level adjustment takes longer than normal
33 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
EC116 | Pickup arm could not be moved to proper position | Pickup cup arm upper mechanism fails to reach correct position | error |
EC121 | Optics module attenuator error | At power up, attenuator cannot be stepped to home position after multiple attempts | error |
EC122 | Optics module not present | OMCON signal not detected low by SCB during power up | error |
EC123 | Optics module temperature error | Optics module fails to reach 46–56°C within 15 minutes after power up | warning |
EC126 | SCB unable to detect EOL signal from optics module | SCB does not receive expected LINESYNC (EOL) signal from optics for each film line during imaging | error |
EC130 | Supply door open | Supply door sensor indicates door is not fully closed | warning |
EC132 | No supply cartridge | Cartridge Present Sensor is not actuated | warning |
EC133 | Less than 20 sheets remain in supply cartridge | Media supply low | info |
EC134 | Supply cartridge empty | SCB has counted 125 sheets since cartridge loaded or film out sensor deactuated | error |
EC139 | Bad cartridge ID | SCB unable to read RF tag information in cartridge | error |
EC145 | Unsupported media type | Cartridge contains incompatible media type | error |
EC151 | Optics module PLL error | XPLL line does not pull low at power-up or before print | error |
EC154 | NVRAM error | SCB NVRAM error detected | error |
EC161 | Film jammed on multifeed detector | Film detected jammed on multifeed detector | error |
EC163 | Jam to exposure | Film jammed after doublefeed sensor and before platen entry; platen entry sensor did not come on | error |
EC164 | Jam exposure | Film jammed in exposure area; platen entry sensor did not turn off or platen sensors did not come on | error |
EC165 | Jam before processor | Film jammed after exposure; platen exit sensor came on but mid transport or processor entry sensor did not | error |
EC166 | Film pickup problems | One or more pickup attempts failed; may be transparent or display message after max retries | warning |
EC173 | Platen door open | Platen door laser interlock or left door laser interlock is open | error |
EC176 | Supply cartridge could not be opened | Tongue depressor sensor failed to go high during cartridge lid rollback process | error |
EC177 | Supply cartridge could not be closed | Tongue depressor sensor or motor control failed during cartridge lid closure | error |
EC178 | Top cover open | Top cover of 8700/8500 is open, opening interlock switch SW802 | error |
EC179 | Left side door open | Left side door is open or not properly latched | error |
EC202 | Override switch is open | Service override switch is in the pulled out (electrically closed) position | warning |
EC203 | Printer malfunction - service override switch pulsing | Service override switch is pulsed high and low alternately every 1 second; SCB detects low when expecting high | critical |
EC203 | Interlock failed (primary) | SCB sends low command and receives high feedback from primary interlock switch after 1 second | critical |
EC204 | Interlock failed (secondary) | SCB sends low command and receives high feedback from secondary interlock switch after 1 second | critical |
EC205 | Interlock failed (tertiary) | SCB sends low command and receives high feedback from tertiary interlock switch after 1 second | critical |
EC206 | Interlock failed (service override) | Service override interlock switch malfunctions or returns unexpected state | critical |
EC208 | Left side door could not be opened | Left side door failed to open during normal operation | error |
EC221 | Pickup top sensor failed | Pickup top sensor (SW503) failed to operate during pickup cycle run at powerup diagnostics | critical |
EC222 | Pickup middle sensor failed | Pickup middle sensor (SW504) failed to operate during pickup cycle run at powerup diagnostics | critical |
EC223 | Film present sensor failed | Powerup diagnostic cycle; film out sensor SW502 fails to operate during pickup | error |
EC224 | Pickup motor failed | Powerup diagnostic cycle; pickup motor does not operate | error |
EC225 | Cartridge open sensor failed | Cartridge opened or closed; sensor does not change states | error |
EC226 | Rollback home sensor failed | Cartridge load sequence; rollback bar home sensor not responding | error |
EC227 | Rollback motor failed | Cartridge load sequence; rollback motor does not operate | error |
EC228 | Tongue depressor sensor failed | Cartridge lid opening; sensor does not change states properly | error |
EC229 | Supply door could not be opened | Supply door open command execution | error |
EC301 | HIB parity error | System initialization or communication with HIB module | error |
EC302 | HIB register read/write failed | HIB register access attempt | error |
EC303 | HIB lookup table memory failed | HIB memory access during initialization or operation | error |
EC39 | EIB (or IB) not detected by IPB | Input module sends command to DEIB or VEIB on fiber A; no response received on fiber B after Enable LED lights | error |
EC48 | No End of File (EOF) | IPB times out waiting for EOF signal from optics and SCB after print command issued; no full SOL sequence received within 1 minute 45 seconds | error |
EC501 | Processor drum could not be moved | Powerup diagnostic; SCB does not detect drum rotation in last 4 seconds | error |
EC506 | No communication from processor | SCB fails to receive communication from RPB after 10 attempts | error |
EC507 | Processor returning bad data | Invalid response from Rotating Processor Board: parity error, corrupted data, or drum temperature outside 50–300°F | warning |
EC509 | Processor failed to warm up in allotted time | Drum temperature does not reach 122.25°C within calculated warm-up period | error |
EC511 | Processor warming | Drum temperature below 122.2°C at power-on or after cool-down | info |
EC522 | Bad densitometer | Densitometer optical sensor failure or signal loss | error |
EC542 | Jam in processor | Processor entrance sensor triggered but densitometer sensor not triggered; film stuck in processing drum or cooling section | error |
EC543 | Jam in densitometer | Densitometer sensor triggered but exit sensor not triggered; film stuck at densitometer location | error |
EC544 | Jam at exit | Exit sensor triggered but never cleared; film stuck at exit assembly | error |
EC549 | Processor filter not present | Filter installation sensor not triggered at power-on or after filter removal | warning |
EC550 | Clean processor and replace filter | Filter count in SCB NVRAM exceeds 10,000 prints | info |
EC551 | Processor heater zone 1 failed | Temperature of heater zone 1 not within 70°F of zones 2 and 3 | error |
EC552 | Processor heater zone 2 failed | Temperature of heater zone 2 not within 70°F of zones 1 and 3 | error |
EC553 | Processor heater zone 3 failed | Temperature of heater zone 3 not within 70°F of zones 1 and 2 | error |
EC554 | Processor over temp | Temperature from any of three heater zones reported above 290°F | error |
EC561 | Invalid densitometer data | Densitometer gain not 1400-2000 or offset not greater than 0 during auto calibration | warning |
EC59 | IPB unable to load parameters to EIB (or IB) | IPB sends parameter set to EIB/IB but cannot confirm write completion; or EOT (End of Transmission) not received from host during acquisition (ACQERR EIB variant) | error |
EC601 | Invalid power monitor offset | BPM offset reading outside 1 through 25 range at power up | error |
EC602 | Invalid power monitor range | Maximum laser power reading outside 50 to 70 mw with attenuator in wide open position | error |
EC603 | Invalid optical density range | Attenuator Dynamic range outside 80-120 | error |
EC604 | Invalid laser dynamic range | Laser Dynamic Range outside 200 to 4000 range | error |
EC605 | Attenuator test failed | Attenuator cannot rotate to position giving proper output power | error |
EC620 | No film model found | Film model for film in cartridge not found in SCB NVRAM and default film model not found | warning |
EC622 | Corrupt film tables | Transfer Function Tables on SCB fail power up checksum test | error |
EC623 | Bad transfer function | Transfer Function Tables on SCB contain invalid data | error |
EC624 | Bad densitometer data | Dmin greater than 0.3 or calibration sheet steps not monotonically increasing in density | warning |
EC625 | Bad film contrast | Contrast of calibration sheet unacceptable, step densities out of spec | warning |
EC631 | Film slow | Darkest density of step 26 not above 3.1 for blue film or 3.0 for clear film | warning |
EC632 | Film fast | Film speed from calibrate sheet above 3.4 | warning |
EC633 | System near calibration limits (film almost fast) | Film speed between 3.3 and 3.4 | info |
EC634 | System near calibration limits (film almost slow) | Film speed between 2.6 and 2.7 | info |
EC635 | Dpatch slow (midtone slow) | Dpatch density outside control limits | warning |
EC636 | Dpatch fast (midtone fast) | Dpatch density outside control limits | warning |
EC64 | IPB parity error during acquisition | Parity error detected in data from EIB | error |
EC65 | Data Overflow or Underflow | VEIB internal pixel clock configured but external clock required | error |
EC66 | Image memory address boundary crossed | Image memory address exceeds available memory | error |
EC67 | Too many pixels received from EIB (FIFO overflow) | More pixels received than expected during acquire | error |
EC910 | Printer malfunction; laser imager not communicating | IPB does not receive command from SCB within 45 seconds; IPB/SCB communication failure; or 8800 Multi-Input Manager loses contact with 8700/8500 imager | critical |
EC913 | HIB parity error | Communication error detected on HIB interface | error |
K203 / L75 | Local panel failed | During panel initialization or operation | error |
K59 | FAI,SET | During image acquisition | error |
K65 | FAI,SET - Error reading EIB parameters from NVRAM | During module initialization | error |
K65 / L39 | ALM,39 - EIB (or IB) not detected | Module initialization or communication | error |
K65 / L59 | ALM,65 - EIB returned incorrect EOT | After image acquisition from wrong input port | error |
K65 / L59 | ALM,39 - IPB unable to load parameters to EIB | Module parameter initialization | error |
K68 / L68 | ALM,68 - Timeout waiting for pixels from EIB | During image acquisition | error |
K72 | FAI,SET - IMS system memory error during acquisition | During image data transfer to main system | error |
K72 / L59 | ALM,61 - Error reading EIB parameters from NVRAM | During module initialization | error |
K72 / L64 | ALM,64 - IPB parity error during acquisition | During image acquisition | error |
K72 / L66 | ALM,66 - Image memory boundary crossed | During image acquisition | error |
K72 / L67 | ALM,67 - Too many pixels received from EIB | During image acquisition | error |
K85 / L85 | FAI,MOV / FAI,BSY - Image memory full | When attempting to store image | warning |
L111 | Cartridge feed roller open failed | During cartridge feed mechanism operation | error |
L200 | Transmit parity error | During data transmission | error |
L201 | Module synchronization error | During module synchronization | error |
L202 | Transmit memory out of bounds | During memory write operations | error |
L210 | Acquire line buffer overflow | During image line acquisition | error |
L212 | Handshake acquisition error | During module handshake during acquisition | error |
L214 | Line length error | During line length validation | error |
L216 | Hardware real time error | During real-time hardware operation | error |
L217 | Software loading real time error | During software load or execution | error |
L48 | End of film timeout error | Film processing | error |
L635 | Dpatch slow | When calibration patch response slower than expected | warning |
L636 | Dpatch fast | When calibration patch response faster than expected | warning |
L65 | Data overflow or underflow | During image data transfer | error |
P116 / K202 / L116 | ALM,60 - Pickup arm could not be moved to proper position | During pickup arm positioning | error |
P121 / K202 / L121 | ALM,60 - Optics module attenuator error | During attenuator operation | error |
P122 / K202 / L122 | ALM,60 - SCB cannot communicate with optics module | During optics module communication | error |
P123 / K202 / L123 | ALM,60 - Optics module temperature error | During thermal monitoring | error |
P126 / K202 / L126 | ALM,60 - SCB unable to detect EOL signal from optics module | After laser print operation | error |
P130 / K215 | ALM,14 - Supply door open | When door open sensor is triggered | warning |
P132 / K215 | ALM,12 - No supply cartridge | When supply cartridge absent or not detected | warning |
P133 / K204 | FLO - Less than 20 sheets remain in supply cartridge | When cartridge reaches low film threshold | info |
P134 / K218 | ALM,10 / ALM,16 - Supply cartridge empty | When cartridge contains no usable film remaining | error |
P139 / K209 / L139 | ALM,31 - Bad cartridge ID | When cartridge ID cannot be read or is invalid | error |
P145 / K209 / L145 | ALM,31 - Unsupported media type | When cartridge media type not supported by system | error |
P146 / K209 / L145 | ALM,31 - Wrong media size | When cartridge film size does not match configuration | error |
P151 / K202 / L151 | ALM,60 - Optics module PLL error | During laser frequency synchronization | error |
P154 / K202 / L154 | ALM,60 - NVRAM error | During NVRAM access or self-test | error |
P161 / K211 / L161 | ALM,21 - Film jammed on multifeed detector | When multifeed sensor detects jam | error |
P163 / K211 / L163 | ALM,22 - Film jammed after multifeed and before platen entrance | When film path sensor detects jam in this section | error |
P164 / K211 / L164 | ALM,23 - Film jammed on platen entrance or exit sensor | When platen area jam sensor triggered | error |
P165 / K211 / L165 | ALM,26 - Film jammed on mid transport sensor | When mid-transport jam sensor triggered | error |
P166 | Multiple pickup attempts | When pickup mechanism fails multiple times consecutively | warning |
P171 | Supply door failed to open | When door open command not executed | error |
P173 / K215 | ALM,16 - Platen door open | When platen door open sensor activated | warning |
P176 / K210 / L176 | ALM,30 - Supply cartridge could not be opened | When cartridge cover open mechanism fails | error |
P177 / K210 / L177 | ALM,31 - Supply cartridge could not be closed | When cartridge cover close mechanism fails | error |
P178 / K215 / L178 | ALM,16 - Top cover open | When top cover open sensor triggered | warning |
P179 / K215 | ALM,16 - Left side door open | When left side door open sensor activated | warning |
P202 | Service override switch out | When service override switch not in correct position | warning |
P203 / K202 / L203 | ALM,60 - Service override interlock failed | When service override safety interlock fails | error |
P204 / K202 / L204 | ALM,60 - Supply door interlock failed | When supply door safety interlock fails | error |
P205 / K202 / L205 | ALM,60 - Top cover interlock failed | When top cover safety interlock fails | error |
P206 / K202 / L206 | ALM,60 - Left side door interlock failed | When left side door safety interlock fails | error |
P208 / L208 | Left side door could not be opened | When left door open command not executed | error |
P221 / K202 / L221 | ALM,60 - Pickup top sensor failed | When pickup top position sensor not responding | error |
P222 / K202 / L222 | ALM,60 - Pickup middle sensor failed | When pickup middle position sensor not responding | error |
P223 / K202 / L223 | ALM,60 - Film present sensor failed | When film presence sensor not responding | error |
P224 / K202 / L224 | ALM,60 - Pickup motor failed | When pickup motor not responding to commands | error |
P225 / K202 / L225 | ALM,60 - Cartridge open (rack) sensor failed | When cartridge open/closed sensor not responding | error |
P226 / K202 / L226 | ALM,60 - Rollback home sensor failed | When rollback home position sensor not responding | error |
P227 / K202 / L227 | ALM,60 - Rollback motor failed | When rollback motor not responding | error |
P228 / K210 / L176-L177 | ALM,60 - Tongue depressor sensor failed | When tongue depressor sensor not responding | error |
P229 / K202 / L229 | ALM,60 - Supply door could not be opened | When supply door open mechanism fails | error |
P230 / K202 / L230 | ALM,60 - Cartridge pickup calibration failed | During pickup mechanism calibration | error |
P301 / K202 / L301 | ALM,60 - HIB version not compatible with current software | During HIB module initialization | error |
P302 / K202 / L302 | ALM,60 - Read or write to HIB register failed | During HIB register access | error |
P303 / K202 / L303 | ALM,60 - HIB lookup table memory test failed | During HIB memory self-test | error |
P304 | HIB test print requested | When HIB test print command received | info |
P501 / K202 / L501 | ALM,17 - Processor drum could not be moved | When processor drum positioning fails | error |
P506 / K202 / L506 | ALM,17 - Processor communication error | When processor communication link fails | error |
P507 / K202 / L507 | ALM,17 - Processor returning bad data | When processor sends invalid response data | error |
P509 / K202 / L509 | ALM,17 - Processor failed to warm up in allotted time | During processor power-up thermal stabilization | error |
P510 / L510 | Calibration sheet printing | During calibration sheet print cycle | info |
P511 / K222 | ALM,17 - Processor warming | During processor thermal startup | info |
P512 | Calibration sheet requested | When calibration sheet print is initiated | info |
P522 / L522 | Bad densitometer | When densitometer returns invalid readings or fails test | error |
P542 / K227 / L542 | ALM,17 - Jam in processor | When film jam detected in processor module | error |
P543 / K211 / L543 | ALM,17 - Jam in densitometer | When film jam detected in densitometer area | error |
P544 / K211 / L544 | ALM,17 - Jam in exit area | When jam detected near output path | error |
P549 / L549 | Processor filter not present | When processor filter is missing or not detected | warning |
P550 / L550 | Clean processor and replace filter | Scheduled maintenance reminder | info |
P551 / K202 / L551 | ALM,17 - Processor heater 1 failed | When processor heater 1 not responding | error |
P552 / K202 / L552 | ALM,17 - Processor heater 2 failed | When processor heater 2 not responding | error |
P553 / K202 / L553 | ALM,17 - Processor heater 3 failed | When processor heater 3 not responding | error |
P554 / K202 / L554 | ALM,17 - Processor temperature too high | When processor temperature exceeds safe operating range | error |
P561 / L561 | Invalid densitometer readings | When densitometer produces out-of-range values | warning |
P601 / K202 / L601 | ALM,60 - Invalid power monitor offset | During power monitoring calibration | error |
P602 / K202 / L602 | ALM,60 - Invalid power monitor range | During power monitoring validation | error |
P603 / K202 / L603 | ALM,60 - Invalid optical density range | During optical density validation | error |
P604 / K202 / L604 | ALM,60 - Invalid laser dynamic range | During laser calibration validation | error |
P605 / K202 / L605 | ALM,60 - Attenuator test failed | During attenuator function test | error |
P620 / L620 | ALM,57 - No film model found | When film type cannot be identified | error |
P622 / K202 / L622 | ALM,57 - Corrupt film tables | When film characteristic tables fail integrity check | error |
P623 / L623 | ALM,57 - Bad transfer function | When film transfer function validation fails | error |
P624 / L624 | AIQC OFF - Bad densitometer data | When densitometer provides invalid measurement data | warning |
P625 / L625 | AIQC OFF - Bad film contrast | When film contrast outside acceptable range | warning |
P631 / L631 | AIQC OFF - Film slow | When film response slower than expected | warning |
P632 / L632 | AIQC OFF - Film fast | When film response faster than expected | warning |
P633 / K230 / L633 | System near calibration limits (film almost fast) | When film response approaching upper calibration limit | warning |
P634 / K230 / L634 | System near calibration limits (film almost slow) | When film response approaching lower calibration limit | warning |
P910 / K202-K1 | ALM,60 - Imager detects no communication from IMS | When imager module cannot reach main IMS system | error |
P913 / K202 / L913 | ALM,60 - HIB parity error | When data parity error detected in HIB transmission | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.