Kodak ACR-2000: troubleshooting and error codes
The ServiceForge knowledge base holds 65 documented symptoms for the Kodak ACR-2000, with 98 ranked causes, 166 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
No Galvanometer Movement caution
Galvanometer beam does not sweep across the scan plate during laser operation.
Ranked causes
- 1. Failed drive circuitry on Data Acquisition PCA (common)
- 2. DACQ adjusted to zero Galvo span (occasional)
- 3. Galvo drive signal cable damaged or disconnected (occasional)
- 4. Galvo Driver PCA failed (occasional)
- 5. Galvanometer mechanical failure (rare)
Disable PMT High Voltage and install laser interlock
- Test point
- AUX/IO connector on Data Acquisition PCA
- Expected
- Connector removed and Laser interlock key installed
8 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Galvo motor not operating high voltage
Laser galvo mirror motor does not move or scan across phosphor screen
Ranked causes
- 1. DACQ board malfunction (common)
- 2. Galvo driver board failure (common)
- 3. Galvo motor failure (common)
- 4. Drive cable J2 malfunction (occasional)
Observe laser beam movement during scan test
- Test point
- Light Collection Chamber width
- Expected
- Laser beam extends full width in both directions
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Laser scan linearity out of specification radiation
Parameter values CAL+, CAL−, Sum of Deltas, or Distance to first edge exceed specified tolerances when tested with linearity test pattern
Ranked causes
- 1. GALVO SPAN potentiometer R59 out of adjustment (common)
- 2. GALVO OFFSET potentiometer R58 out of adjustment (common)
- 3. GALVO DRIVER BOARD potentiometers out of adjustment (occasional)
- 4. Sum of Deltas parameter out of tolerance (occasional)
Ensure room has minimum light and 5-minute warmup complete
- Test point
- Equipment ambient
- Expected
- Room dark, READER at normal operating temperature
8 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
DI-2000 Self Test Failure
Autozero and/or AGC diagnostic tests fail during self-test execution.
Ranked causes
- 1. Excessive ambient light in room (common)
- 2. PMT offset adjustment out of specification (occasional)
- 3. Photomultiplier tubes (PMTs) defective (occasional)
Run DDT option 13/5 and verify Autozero voltage
- Test point
- Autozero voltage reading
- Expected
- within +1 volts (closer to 0 is better), +1 volts
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
LUT (Look-Up Table) transfer or configuration fails
LUT file does not transfer to DACQ BOARD or LUT settings incorrect; affects image contrast and tone mapping
Ranked causes
- 1. LUT file format or compatibility issue (occasional)
- 2. DACQ BOARD hardware issue (occasional)
- 3. Invalid calibration table (occasional)
Load LUT file to DACQ BOARD
- Test point
- DDT option 4; DACQ BOARD LUT memory
- Expected
- LUT file transferred successfully; no errors reported
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No Photomultiplier Tube Output high voltage
PMT does not produce output signal; PMT Output Voltage mean data value is abnormally high or near zero.
Ranked causes
- 1. Laser not emitting light (common)
- 2. PMT High Voltage Power Supply failed (common)
- 3. Bent pins on cable between PMT Preamp and Data Acquisition PCA (occasional)
Perform PMT Output Voltage Adjustment from Chapter 5
- Test point
- PMT Output Voltage mean data value
- Expected
- between 600 and 1000 normally digital units, 600–1000
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scan file image capture fails or produces poor quality radiation
Captured image is missing data, corrupted, or poor quality; image stored on DATA CONTROL BOARD memory
Ranked causes
- 1. IMAGE CONTROL BOARD or memory issue (occasional)
- 2. SCANNER DRIVER error (occasional)
- 3. Phosphor screen contamination or mechanical jam during scan (occasional)
Initiate scan file capture
- Test point
- SCANFILE.EXE or DDT option 1
- Expected
- Scan completes; image stored in memory as default file lsdt.img
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scan line noise or degraded signal quality detected
Scan line displays noise or weak signal; noise value high; affects image clarity
Ranked causes
- 1. High sensor noise or contamination (occasional)
- 2. Laser power degraded or misaligned (occasional)
- 3. Scan linearity drift requiring adjustment (occasional)
Display noise value to assess signal quality
- Test point
- DDT option 20; scan line signal
- Expected
- Noise level within acceptable baseline range
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Screen/phosphor plate motor control failure
Motor does not move screen forward/backward during scan; jammed or unresponsive
Ranked causes
- 1. Screen motor not receiving power or control signal (common)
- 2. Screen mechanically jammed or obstructed (occasional)
- 3. Screen motor parameters misaligned or out of spec (occasional)
Energize screen motor and verify movement
- Test point
- DDT option 12; screen motor and phosphor plate
- Expected
- Motor activates; screen moves in commanded direction smoothly
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System will not detect phosphor screens radiation
Screen moves into system several inches and stops without backing up; detector fails to register screen presence
Ranked causes
- 1. Failed laser module (common)
- 2. Insufficient laser light output (occasional)
- 3. Laser misalignment with screen detectors (rare)
Visually confirm laser is emitting light
- Test point
- Laser beam path inside scanner
- Expected
- Visible laser light emitting from laser module
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Dosimeter TL-4818 exposure measurement radiation
Procedure to verify and adjust X-ray exposure using TL-4818 dosimeter before adjustment or calibration
Ranked causes
- 1. Exposure measurement verification required (common)
- 2. LCD display does not indicate correct value (occasional)
Verify LCD display initial reading
- Test point
- LCD display on dosimeter
- Expected
- 0.00
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: PMT output voltage adjustment and reference blue LED adjustment radiation
Plate exposed at standard X-ray dosage must produce consistent count level during scanning. PMT high voltage adjusted to give ~600 counts; blue reference LED adjusted to maintain sensitivity via AGC circuit.
Ranked causes
- 1. PMT sensitivity requires calibration to standard dose exposure (common)
- 2. Blue reference LED requires adjustment for AGC function (common)
Verify X-ray exposure dose accuracy
- Test point
- High resolution dose meter reading
- Expected
- 8 mR ±0.05 mR, ±0.05 mR
- Tools
- High resolution X-ray dose meter
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: PMT output voltage balancing radiation
Two PMT tubes must have matched output amplitudes for consistent image acquisition. If one PMT cannot match the other, both must be replaced as a set.
Ranked causes
- 1. PMT tube gain mismatch requires balancing (occasional)
- 2. PMT tubes cannot be matched and require replacement (rare)
Verify X-ray exposure dose for balancing plate
- Test point
- High resolution dose meter
- Expected
- 4 mR ±0.05 mR, ±0.05 mR
- Tools
- High resolution X-ray dose meter
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image is white or blank
Acquired image appears completely white or blank
Ranked causes
- 1. PMT power supply malfunction (common)
- 2. Bent pins in cable between PMT preamplifier and DAC acquisition board (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Laser beam not centered in light collection chamber radiation
Beam does not project to center of light collection chamber within tolerance
Ranked causes
- 1. Galvo mirror misalignment (common)
- 2. Laser diode module misalignment (occasional)
Verify beam center position
- Test point
- Light collection chamber opening
- Expected
- Beam at center +/- 0.5 inches
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pixel clock or scan timing parameters incorrect
Scan timing off; pixels not captured at correct intervals; image distorted or incomplete
Ranked causes
- 1. Scan timing parameters drift or misconfiguration (occasional)
- 2. DACQ BOARD timing circuit fault (rare)
Display current pixel clock and scan parameters
- Test point
- DDT option 11; parameter menu
- Expected
- Current values displayed; note deviation from expected defaults
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PMT output voltage out of specification radiation
Voltage measured at PMT PREAMP BOARD test point is not -300V DC or higher; output signal is weak or absent.
Ranked causes
- 1. Bent pins on cable between PMT PREAMP and DACQ boards (common)
- 2. Missing or low 24V DC at PMT POWER SUPPLY (occasional)
Measure voltage at TP2 on PMT PREAMP BOARD
- Test point
- TP2 (PMT PREAMP BOARD)
- Expected
- -300 V DC or higher VDC
- Tools
- Digital Multimeter set to DC
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Start/End of film detection not triggering radiation
PHOTOSENSOR fails to detect laser light off PHOSPHOR SCREEN or ISFILM signal does not become positive when film is detected
Ranked causes
- 1. PHOTOSENSOR misalignment or obstruction (common)
- 2. DACQ BOARD signal detection failure (occasional)
Verify PHOTOSENSOR detects laser light off PHOSPHOR SCREEN
- Test point
- PHOTOSENSOR output
- Expected
- Signal present
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vertical lines in images
Random but repeatable vertical lines appear in successive radiographic images
Ranked causes
- 1. Lint blocking laser beam near focal point (common)
- 2. Laser beam misalignment (occasional)
Digitize test image and compare with baseline
- Test point
- System output
- Expected
- No vertical lines or fewer lines than before cleaning
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Autozero test does not complete radiation
Autozero diagnostic fails to finish or complete successfully
Ranked causes
- 1. Excessive ambient light in room (common)
Measure ambient light level
- Test point
- room environment
- Expected
- 2 EV or less
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blue REF/ISL film board replacement high voltage
Remove and install Blue REF/ISL film board with main power disconnected and dangerous voltage warning observed
Ranked causes
- 1. Blue REF/ISL film board requires replacement (unknown)
De-energize equipment
- Test point
- Main power switch
- Expected
- Equipment powered off
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: AGC adjustment radiation
Automatic Gain Control compensation for ambient light during Start of Scan; PMT output adjusted to 6-V maximum with laser beam at center of light collection chamber
Ranked causes
- 1. Ambient light variation requires AGC recalibration (occasional)
Verify PMT output voltage at light collection chamber center
- Test point
- PMT output
- Expected
- 6-V maximum
- Tools
- Diagnostic tool, voltmeter
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Calibration table generation radiation
Generate new calibration (CAL) table after PMT balancing or maintenance. Procedure varies by software version (NT driver <2.00 vs ≥2.00 or DOS/95/98 ≥3.60).
Ranked causes
- 1. CAL table must be regenerated after PMT adjustments (common)
Verify X-ray exposure for CAL plate
- Test point
- High resolution dose meter
- Expected
- 4 mR ±0.05 mR, ±0.05 mR
- Tools
- High resolution X-ray dose meter
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Laser beam alignment in light collection chamber radiation
Laser beam position and alignment in the optical path must be verified and adjusted to specification before operation.
Ranked causes
- 1. Laser beam misaligned after installation or service (unknown)
Laser beam visible on GALVO MIRROR after energizing equipment
- Test point
- GALVO MIRROR
- Expected
- Laser beam visible
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: PMT output voltage balance reference blue LED high voltage
Adjust blue LED reference voltage to mean value of 1071±10 using potentiometer R2 on REF BLUE LED/SCREEN PRESENT BOARD
Ranked causes
- 1. Calibration procedure for blue LED reference output (unknown)
Energize blue LED
- Test point
- REF BLUE LED/SCREEN PRESENT BOARD
- Expected
- Blue LED energized
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: PMT preamp offset high voltage
Set offset value for photomultiplier tube preamplifier to specification 0.0 ± 0.1 mV
Ranked causes
- 1. PMT preamp offset requires calibration (unknown)
Remove main cover
- Test point
- Main cover
- Expected
- Cover removed
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: REF BLUE LED/SCREEN PRESENT BOARD high voltage
Adjust potentiometer R2 to achieve Mean value of 1071±10 after energizing blue LED via command-line tool
Ranked causes
- 1. Potentiometer R2 requires calibration after board replacement or installation (common)
Open command prompt and navigate to DDT tool directory
- Test point
- Command line
- Expected
- c:\lsdt32\tools> prompt visible
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Scan Linearity Adjustment
Ensure laser beam reads single line of data in linear region; verify 14 test bars are evenly sampled and positioned within scan range. Adjust beam movement rate and position using potentiometers.
Ranked causes
- 1. Beam movement rate incorrect or bars mispositioned (common)
Run LNADJDT program and observe Cal+ and Cal- error percentages
- Test point
- Screen display from LNADJDT
- Expected
- CAL+ and CAL- errors less than 2% (nominal), less than 3% (preferred), or less than 5% (allowable), nominal <2%, preferred <3%, allowable <5%
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Scan Linearity setup and verification
Prepare unit environment and verify test pattern placement for scan linearity measurement
Ranked causes
- 1. Ambient light interference with linearity test (common)
Place 14 inch Linearity Adjust Pattern in input slot
- Test point
- Input slot
- Expected
- Vertical lines facing unit, widest red bar toward abort button
- Tools
- 14 inch Linearity Adjust Pattern P/N 0070-966
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CLUT calibration out of specification
CHKCLUT program reports pixel-to-optical-density conversion is incorrect; scanner produces wrong density values
Ranked causes
- 1. CLUT file incorrect or outdated (common)
Verify CLUT file correctness using CHKCLUT utility
- Test point
- Command line interface
- Expected
- CHKCLUT passes without out-of-specification errors
- Tools
- CHKCLUT utility, step wedge film, density file
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Debug screen display toggle
Enable or disable debug screen display mode to view internal program status information during operation
Ranked causes
- 1. Need internal program status information for troubleshooting (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Diagnostic command execution and system troubleshooting
Access to scanner microprocessor via serial port terminal connection to run diagnostic utilities and troubleshooting commands
Ranked causes
- 1. Diagnostic access requires terminal connection to serial port (common)
Verify serial port connection parameters
- Test point
- RS-232 serial line (RJ-11 connector next to SCSI connector)
- Expected
- Connection established at 9600 baud, 8 bits, No Parity, 1 stop bit (or 56K baud if configured)
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Diagnostic scan to obtain system information radiation
Run diagnostic scan to collect diagnostic data and system status information
Ranked causes
- 1. Need system diagnostic data for troubleshooting (common)
Launch DDT diagnostic program
- Test point
- DOS command prompt
- Expected
- DDT menu displays with options including Diagnostic Scan
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
DMA controller register dump with caution caution
Dump internal DMA (Direct Memory Access) controller registers; use with caution as command can cause data loss during active transfers
Ranked causes
- 1. Advanced troubleshooting of DMA controller for data transfer issues (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exit diagnostic program and return to command line
Quit LSDTISO.EXE diagnostic program to access command line prompt for system configuration and diagnostic utilities
Ranked causes
- 1. Need to exit program to access command line diagnostics or utilities (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Horizontal lines and banding in images
Dark horizontal bands or horizontal lines appear in radiographic images, particularly at top and bottom
Ranked causes
- 1. Excessive ambient light in room (common)
Measure ambient light level in room
- Test point
- Scanner room environment
- Expected
- 2EV maximum
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image density non-uniform or side lighter than expected
Image brightness uneven; one side of image noticeably lighter
Ranked causes
- 1. PMT malfunction (common)
Run DDT.EXE diagnostic on PMTs
- Test point
- PMT output value
- Expected
- Output value not approximately 4095
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image has light horizontal lines
Light horizontal lines appear in the image.
Ranked causes
- 1. Binds occur in the transport (common)
Clean the transport path
- Test point
- Transport mechanism
- Expected
- Transport path clear of obstructions
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image has vertical lines
Vertical lines visible in acquired image
Ranked causes
- 1. Beam blocked by particles in optics path (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect dosimeter reading due to missing lead shielding radiation
Dosimeter value may be false by 20% if lead sheet or lead apron not used under dosimeter during measurement
Ranked causes
- 1. Lead shielding not placed under dosimeter (common)
Verify lead shielding present under dosimeter
- Test point
- Floor area under dosimeter
- Expected
- Lead sheet or lead apron visible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Input guide and output tray cleaning
Daily cleaning of input guide and output tray with anti-static spray maintains proper mechanical operation and prevents dust accumulation.
Ranked causes
- 1. Dust and contamination accumulate in input guide and output tray (common)
Visually inspect input guide and output tray
- Test point
- Guide and tray surfaces
- Expected
- Clean, dust-free, no debris
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Laser beam out of focus radiation
Laser beam is not visibly in focus on the scanning surface
Ranked causes
- 1. Connector misalignment or loose setscrews (common)
Verify laser beam focus on paper surface
- Test point
- Post-it notes placed on slot
- Expected
- Laser beam is visibly in focus
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Laser beam position verification before operation radiation
Verify laser beam is visible and properly positioned in optics after energizing equipment before proceeding with adjustment.
Ranked causes
- 1. Laser beam not visible on GALVO MIRROR (occasional)
Interlock key installed
- Test point
- INTERLOCK KEY location
- Expected
- Key is installed
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Laser radiation exposure hazard radiation
Operator or technician may be exposed to laser beam if equipment covers are removed during operation
Ranked causes
- 1. Equipment operated without protective covers (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No laser light output radiation
Power is on and interlock key installed, but no laser beam projects onto galvo mirror
Ranked causes
- 1. Laser Module failure (unknown)
Verify laser light presence
- Test point
- Galvo mirror
- Expected
- Beam projects onto galvo mirror
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Non-uniformity of image density
One side of image is significantly darker or lighter than the other side; uneven grayscale across radiograph
Ranked causes
- 1. Failed Photo-Multiplier Tube (PMT) (common)
Disconnect first PMT cable and digitize test image
- Test point
- PMT #1 cable connection
- Expected
- Image quality unchanged if PMT #1 is functional; significant degradation if PMT #1 is failed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Photomultiplier Tube not detecting laser radiation
PMT (Photomultiplier Tube) module not responding to laser input; laser diode module not emitting beam.
Ranked causes
- 1. Laser Diode Module not functioning (common)
Run DDT.EXE diagnostic and verify LASER DIODE MODULE emits beam
- Test point
- Laser diode output
- Expected
- Visible laser beam
- Tools
- DDT.EXE software
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Plate detector not receiving laser beam radiation
Isfilm signal spikes not visible or beam not hitting three detectors
Ranked causes
- 1. Roof mirror height misalignment (common)
Measure Isfilm signal
- Test point
- DDT menu option 20
- Expected
- Three spikes approximately 1600 in amplitude
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PMT damage from light exposure caution
Photomultiplier tube (PMT) can be damaged if exposed to light during maintenance when main cover is removed and equipment is energized.
Ranked causes
- 1. Main cover removed with J14 connected and equipment energized without light protection (common)
Verify room light level is minimized
- Test point
- Room ambient
- Expected
- Minimum light conditions
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PMT light damage risk during maintenance patient safety
Photomultiplier tube (PMT) exposure to light when main cover is removed with J14 connected and equipment energized
Ranked causes
- 1. Main cover removal without light protection (common)
Verify photographic cloth is in place before removing main cover
- Test point
- Equipment main cover
- Expected
- PHOTOGRAPHIC CLOTH 1F1622 covering PMT area
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
PMT Preamp Offset Adjustment caution
Adjust PMT preamp offset voltage to 0mV ±0.1mV to ensure proper photomultiplier tube signal conditioning
Ranked causes
- 1. PMT Preamp offset voltage not at 0mV (occasional)
Measure voltage at TP9 on DACQ PCA with PMT HV disabled
- Test point
- TP9 DACQ PCA
- Expected
- 0mV, ±0.1mV
- Tools
- DVM set to mV range
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Replace DATA ACQUISITION (DACQ) BOARD high voltage
DACQ board removed and reinstalled with driver software and calibration files loaded
Ranked causes
- 1. DACQ board failure or replacement (unknown)
De-energize and disconnect main power
- Test point
- Main power switch and connector
- Expected
- Equipment powered off, main connector disconnected
7 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Replace lamps in light box high voltage
Lamps in the light box have failed or need replacement as part of maintenance
Ranked causes
- 1. Lamp end-of-life or failure (common)
De-energize equipment
- Test point
- main power
- Expected
- de-energized
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Replace reader unit high voltage
Reader unit removal and installation procedure for maintenance or replacement
Ranked causes
- 1. Reader unit requires replacement or service (unknown)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Replace SCSI board high voltage
SCSI board removal and installation procedure with file transfer requirements
Ranked causes
- 1. SCSI board failure or obsolescence requiring replacement (unknown)
De-energize the equipment and disconnect main power
- Test point
- Equipment power switch and main power connector
- Expected
- Equipment powered down, no mains voltage present
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Replace STATUS LED GREEN BOARD high voltage
Board removal and installation procedure for maintenance or fault correction.
Ranked causes
- 1. STATUS LED GREEN BOARD requires replacement (unknown)
De-energize the equipment
- Test point
- Main power
- Expected
- Equipment powered off
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Room lighting verification for PMT protection patient safety
Ensure adequate darkness in work area before performing PMT-sensitive maintenance
Ranked causes
- 1. Inadequate room darkness during PMT-exposed service (occasional)
Assess ambient room lighting conditions
- Test point
- Service work area
- Expected
- Minimum light level in room
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scanner MTF performance out of specification
CHKMTF program detects modulation transfer function does not meet specifications; image sharpness degraded
Ranked causes
- 1. Scanner optical or mechanical performance degraded (unknown)
Measure scanner MTF using CHKMTF utility
- Test point
- MTF test film scan
- Expected
- MTF values within specification limits
- Tools
- CHKMTF utility, MTF test film
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scanner X/Y linearity out of specification
CHKXYLIN program detects pixel or line spacing is non-linear; scan geometry is distorted
Ranked causes
- 1. X/Y line spacing measurement error or optical distortion (unknown)
Measure X and Y axis linearity using CHKXYLIN utility
- Test point
- Linearity test film scan
- Expected
- X and Y line spacing within specification; default threshold 1.25
- Tools
- CHKXYLIN utility, linearity test film
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SCSI FIFO error flag status check
Monitor internal FIFO (First In First Out) buffer error flag status to diagnose data overflow or buffer communication errors
Ranked causes
- 1. FIFO buffer overflow or communication error suspected (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- SCSI formatted data dump and display
- SCSI interface chip register inspection
- Storage phosphor screen cleaning and maintenance
- System does not detect phosphor screen
- Top or bottom of image has dark horizontal lines or bands
5 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI