IRadimed 3880: troubleshooting and error codes

The ServiceForge knowledge base holds 585 documented symptoms for the IRadimed 3880, with 674 ranked causes, 1015 diagnostic checks with test points and expected values, and 6 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

ECG Pod interface or RF link malfunction

ePOD connection LED not illuminated, battery indicator not responding, RF link not established, or ECG waveform not received from simulator

Ranked causes

  1. 1. ePOD not connected or RF link failure (common)
  2. 2. ePOD battery depleted or indicator faulty (occasional)
  3. 3. Simulator not connected or waveform generation failure (occasional)
  4. 4. RA/R lead disconnected or damaged (common)
First check

Verify CONNECTED LED illuminated on ePOD

Test point
ePOD CONNECTED LED
Expected
LED illuminated (solid or blinking)

6 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP sensor reading out of specification

NIBP analog voltage readings or pressure accuracy do not meet specification at test points or during pressure application

Ranked causes

  1. 1. NIBP-DC analog voltage out of range (occasional)
  2. 2. NIBP-AC analog voltage out of range (occasional)
  3. 3. Pressure measurement accuracy at 150 mmHg exceeds tolerance (occasional)
  4. 4. Pressure measurement accuracy at 250 mmHg exceeds tolerance (occasional)
First check

Measure NIBP-DC analog voltage at test point AN00

Test point
AN00
Expected
0.060V +/- 0.010V, +/- 0.010V
Tools
DMM

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CONNECTED LED not illuminated on oPOD

The CONNECTED LED on the SpO2 Pod (oPOD) does not light up, indicating no communication between pod and main unit.

Ranked causes

  1. 1. oPOD not properly connected or seated (common)
  2. 2. RF link communication failure (occasional)
  3. 3. oPOD battery depleted (occasional)
First check

Inspect oPOD seating in dock

Test point
oPOD connector
Expected
Fully seated with no gaps

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

ECG waveform not displayed on monitor

ECG waveform does not appear on monitor screen when ePOD is connected to Fluke ProSim 8 simulator

Ranked causes

  1. 1. ePOD channel not set to match monitor channel (common)
  2. 2. ECG lead connection incorrect or loose (common)
  3. 3. Fluke ProSim 8 simulator not powered or not set to ECG mode (occasional)
First check

Verify ePOD and monitor channels match

Test point
Channel selection on monitor and ePOD
Expected
Same channel selected on both devices

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

POD Charger LED Verification

Status indicator LEDs on rear of monitor and POD charging indicators during operation

Ranked causes

  1. 1. Normal operation with no PODs inserted (common)
  2. 2. Normal operation with POD inserted and charging (common)
  3. 3. Normal operation with Right POD inserted (common)
First check

Observe rear panel LEDs with no PODs inserted

Test point
Rear panel Yellow/Orange and Green LEDs
Expected
Alternately flashing

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 Probe Off message displayed

The unit displays 'SpO2 Probe Off' message, indicating no active SpO2 probe connection or detection.

Ranked causes

  1. 1. SpO2 probe not connected to oPOD (common)
  2. 2. SpO2 probe connector loose or damaged (occasional)
  3. 3. oPOD not detecting probe (rare)
First check

Verify SpO2 probe is connected to oPOD

Test point
oPOD SpO2 probe connector
Expected
Probe connector firmly attached

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

3881 ePOD CONNECTED LED not illuminated

CONNECTED LED on ePOD does not light when powered on and channel is set

Ranked causes

  1. 1. ePOD channel not matching Monitor channel (common)
  2. 2. ePOD not powered on (common)
First check

Power on ePOD

Test point
Power button on 3881 ePOD
Expected
Button pressed

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

3881 ePOD ECG waveform not displaying

No ECG waveform visible on Monitor screen after ePOD and simulator connection

Ranked causes

  1. 1. Incorrect ECG Lead Set connection or wiring (common)
  2. 2. Fluke ProSim 8 simulator not powered or not set to ECG mode (common)
First check

Connect ePOD to Fluke ProSim 8 simulator with ECG lead set

Test point
ePOD and simulator connectors
Expected
1811 or 1811E ECG Lead Set connected with correct color code

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

3881 ePOD not connecting to monitor

CONNECTED LED not illuminated on ePOD or no radio link established after power on

Ranked causes

  1. 1. ePOD channel not set to match monitor channel (common)
  2. 2. Poor radio link or distance exceeds range (occasional)
First check

Verify CONNECTED LED illumination on ePOD

Test point
LED on 3881 ePOD
Expected
CONNECTED LED illuminated

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

3882 oPOD CONNECTED LED not illuminated

CONNECTED LED on oPOD does not light when powered on and channel is set to Monitor channel

Ranked causes

  1. 1. oPOD channel not matching Monitor channel (common)
  2. 2. oPOD not powered on (common)
First check

Power on oPOD

Test point
Power button on 3882 oPOD
Expected
Button pressed

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

3886 multigas unit wireless communication or sensor failure caution

3886 wireless LED or CONN AGENTS LED not lit or sample line does not show port status

Ranked causes

  1. 1. Wireless module failure or disconnection (occasional)
  2. 2. Sample line port disconnected or sensor fault (common)
First check

Verify 3886 wireless LED and CONN AGENTS LED on 1887 module

Test point
LED indicators on 1887 module
Expected
Both wireless LED and CONN AGENTS LED lit

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

3886 wireless communication LED not lit

Wireless LED on 3886 unit or CONN AGENTS LED on 1887 Module not illuminated; indicates connection failure

Ranked causes

  1. 1. 3886 unit powered off or wireless module disabled (common)
  2. 2. Wireless connection error or 1887 Module failure (occasional)
First check

Observe 3886 wireless LED and 1887 Module CONN AGENTS LED status

Test point
LED indicators on 3886 unit and 1887 Module
Expected
Both wireless LED and CONN AGENTS LED illuminated (green or solid)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

3886 wireless or CONN AGENTS LED not illuminated patient safety

3886 wireless LED or CONN AGENTS LED on 1887 module does not light during operation

Ranked causes

  1. 1. 3886 unit not powered or wireless communication disabled (common)
  2. 2. LED or wireless module hardware failure (occasional)
First check

Verify 3886 wireless LED illuminated

Test point
3886 unit LED indicator
Expected
Wireless LED lit (solid or blinking)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm volume not adjustable or missing patient safety

Alarm volume does not change when adjusted or alarm does not sound

Ranked causes

  1. 1. Audio amplifier circuit failure (rare)
  2. 2. Volume control potentiometer failure (occasional)
First check

Adjust alarm volume on Monitor across full range

Test point
Monitor volume control and audio output
Expected
Alarm volume changes at 100%, 75%, 50%, 25%, and 0%

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Base, Monitor and Tablet communication loss

No green signal-strength bars; devices not communicating; communication restored when channel mismatch corrected

Ranked causes

  1. 1. Base Station channel mismatch (common)
  2. 2. Wireless module failure or RF interference (occasional)
First check

Verify green signal-strength bars on Tablet, Base and Monitor

Test point
Display signal indicators
Expected
Green signal-strength bars visible on all three devices

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Base, monitor and tablet communication lost

Wireless communication between tablet, base station and monitor is not present or unstable

Ranked causes

  1. 1. Wireless interference or channel conflict (occasional)
  2. 2. Base station, monitor or tablet hardware failure (rare)
First check

Observe signal-strength indicators on devices

Test point
Monitor/Tablet/Base display
Expected
Green signal-strength bars visible

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery indicator display malfunction

A/C plug and battery indicator icons do not appear on display, or battery indicator does not show expected symbol

Ranked causes

  1. 1. Display or graphics rendering fault (occasional)
  2. 2. Firmware or software issue (occasional)
First check

Verify A/C plug and battery indicator icons display

Test point
Display screen
Expected
A/C plug icon and battery indicator icon visible

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery indicator does not show correct state

ON/CHARGE LED color incorrect or battery status icons not displayed on screen

Ranked causes

  1. 1. Battery 1133 not installed or depleted (common)
  2. 2. Battery charging circuit failure (rare)
First check

Verify ON/CHARGE LED color with 1133 battery installed

Test point
ON/CHARGE LED
Expected
Orange

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery indicator shows incorrect color

ON/CHARGE LED does not display expected color (Orange with battery inserted, Green with battery removed)

Ranked causes

  1. 1. Battery not properly inserted or defective (common)
  2. 2. Indicator LED or control circuit malfunction (occasional)
First check

Verify ON/CHARGE LED color with battery installed

Test point
ON/CHARGE LED
Expected
Orange

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Clock time not displayed

Top row of display does not show current time

Ranked causes

  1. 1. Internal coin cell battery depleted (common)
  2. 2. Incorrect time and date settings (occasional)
First check

Verify time displayed on top row

Test point
Display top row
Expected
Current time shown

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

CO2 activity not detected on Multigas unit

3886 Multigas unit does not show CO2 activity when sample line is used

Ranked causes

  1. 1. Sample line disconnected or blocked (common)
  2. 2. Multigas unit not communicating with base (occasional)
First check

Verify 3886 wireless LED and CONN AGENTS LED are lit

Test point
3886 Module wireless LED and 1887 CONN AGENTS LED
Expected
Both LEDs lit

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CO2 activity not shown during breathing test patient safety

Monitor does not display CO2 activity when operator breathes into sample line

Ranked causes

  1. 1. Sample line blockage or kinked tubing (common)
  2. 2. CO2 sensor not functioning or not calibrated (occasional)
First check

Breathe into sample line and observe display

Test point
Monitor CO2 waveform and value display
Expected
CO2 activity visible (waveform or numeric value changes during breathing)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

CO2 reading inaccurate

Nasal cannula breathing test shows CO2 level outside tolerance; expected 4.0% +/- 0.75%

Ranked causes

  1. 1. Calibration gas source not correct or expired (common)
  2. 2. Sensor drift or contamination (occasional)
First check

Breathe into Nasal Cannula and measure CO2 level

Test point
Nasal Cannula port
Expected
4.0% +/- 0.75%, +/- 0.75%

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

CO2 suction flow rate low or out of specification patient safety

CO2 suction flow meter shows flow rate outside specification of 0.095 LPM +/- 0.015 LPM

Ranked causes

  1. 1. Pump worn or suction filter clogged (common)
  2. 2. Tubing leak or valve malfunction (occasional)
First check

Measure CO2 suction flow rate

Test point
Flow meter at suction outlet
Expected
0.095 LPM, +/- 0.015 LPM
Tools
Flow Meter

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

CO2 zeroing initialization failure

3886 Multigas Unit startup shows CO2 Zeroing message that does not clear after 20 seconds; expected automatic clearing

Ranked causes

  1. 1. CO2 sensor not reaching zero baseline or calibration timeout (occasional)
  2. 2. 3886 Multigas Unit communication or sensor failure (rare)
First check

Power on 3886 Multigas Unit and observe CO2 Zeroing message

Test point
3886 display and status LEDs
Expected
"CO2 Zeroing" message appears then clears after 20 seconds

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Communication between tablet, base, and monitor not established patient safety

Green signal-strength bars indicating communication are not visible on display

Ranked causes

  1. 1. Wireless communication module failure or configuration error (unknown)
  2. 2. Base station and tablet/monitor on different channels (common)
First check

Verify wireless signal presence

Test point
Monitor display signal-strength indicator
Expected
Green signal-strength bars visible

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display touch screen not responding

Touch screen does not respond to user input or menus do not appear when display areas are touched

Ranked causes

  1. 1. Display window damaged or scratched (occasional)
  2. 2. Touch screen calibration needed (rare)
First check

Inspect display window for physical damage

Test point
Display window surface
Expected
No visible damage or scratches

5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display touch screen unresponsive

Touch screen does not respond to user input or display menus not appearing

Ranked causes

  1. 1. Display window contamination or damage (common)
  2. 2. Touch controller circuit failure (rare)
First check

Inspect display window area for contamination or damage

Test point
Display surface
Expected
Clean, undamaged

5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display touch screen unresponsive or non-functional

Display window damaged or touch functionality not responding correctly in various regions

Ranked causes

  1. 1. Display window physically damaged or contaminated (occasional)
  2. 2. Touch sensor or display controller fault (occasional)
First check

Inspect display window area for damage

Test point
Display window
Expected
Clear, undamaged window; no visible cracks or contamination

5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display touchscreen not responding

Touchscreen does not respond to touches or menu items not displayed when screen areas touched

Ranked causes

  1. 1. Display damage or contamination preventing touch response (occasional)
  2. 2. Touchscreen firmware or calibration issue (rare)
First check

Inspect display window for physical damage

Test point
Display window area
Expected
No visible damage or scratches

5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display Touchscreen Verification

Display window integrity and touchscreen control responsiveness

Ranked causes

  1. 1. Visual inspection required (common)
  2. 2. Touchscreen responsiveness verification (common)
First check

Visually inspect display window

Test point
Display window area
Expected
No damage or scratches visible

5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

ECG Pod interface not connecting MRI environment

ePOD CONNECTED LED not illuminating or RF link not establishing

Ranked causes

  1. 1. ePOD battery depleted (common)
  2. 2. ePOD RF transceiver failure (rare)
First check

Verify ePOD battery indicator

Test point
ePOD battery LED
Expected
LED illuminated

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

ECG waveform missing or incorrect MRI environment

No ECG signal displayed on monitor from simulator or device

Ranked causes

  1. 1. ECG lead disconnected or damaged (common)
  2. 2. ePOD ECG channel circuit failure (rare)
First check

Verify ECG waveform from Simulator

Test point
Monitor display
Expected
ECG waveform displayed

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

ECG waveform not displayed for specific leads

No waveform detected when selecting leads III, aVR, aVL, aVF, V, or I on Monitor display

Ranked causes

  1. 1. ePOD not tuned to same channel as Monitor (common)
  2. 2. Wireless link not established (occasional)
First check

Select each lead (III, aVR, aVL, aVF, V, I) from upper left display

Test point
Monitor display lead selection buttons
Expected
Waveform detected for each lead

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

ECG waveform not displayed from simulator

After reconnecting RA/R lead, verify ECG waveform appears on display without alarms

Ranked causes

  1. 1. RA/R lead not properly connected (common)
  2. 2. ECG pod not connected to monitor (occasional)
First check

Visually inspect RA/R lead connection at ECG pod

Test point
RA/R lead connector
Expected
Lead fully seated, no gaps

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

ePOD does not connect to monitor patient safety

3881 ePOD CONNECTED LED not illuminated or radio link not established; ECG signal not received by monitor.

Ranked causes

  1. 1. ePOD Channel setting does not match Monitor Channel setting (common)
  2. 2. Radio link not established or RF signal weak (occasional)
First check

Verify Monitor and ePOD Channel settings match

Test point
Monitor upper left Channel selection; ePOD Channel setting
Expected
Both set to same Channel

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

ePOD not connecting to Monitor

CONNECTED LED not illuminated on 3881 ePOD or 'e' battery indicator not shown on display

Ranked causes

  1. 1. ePOD channel not set to same channel as Monitor (common)
  2. 2. Radio link not established or out of range (occasional)
First check

Verify ePOD Power button is ON

Test point
3881 ePOD power button
Expected
Power button in ON (I) position

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gas calibration out of specification caution

Verification gas readings do not match expected values: CO2 5.0%, N2O 42.0%, DES 2.0%, O2 51.0%

Ranked causes

  1. 1. Verification gas bottle depleted or incorrect mix (common)
  2. 2. Sensor calibration error or multigas unit malfunction (occasional)
First check

In Agents Maintenance menu start Cal Zero

Test point
Agents Maintenance menu, Cal Zero status
Expected
Port light flashing indicating zeroing in progress

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gas readings do not hold steady in closed-loop test patient safety

Gas readings drop more than 0.5% over 2-minute closed-loop hold test indicating leak or drift

Ranked causes

  1. 1. Leak in sample line or exhaust port connection (common)
  2. 2. Gas sensor drift or calibration error (occasional)
First check

Create closed loop and monitor gas reading stability

Test point
Monitor gas values over 2-minute interval
Expected
Readings hold steady with < 0.5% total drop, < 0.5% drift in 2 min

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Heart rate tone not present at all volume levels patient safety

HR tone does not sound or sounds at only some volume settings

Ranked causes

  1. 1. Audio amplifier or speaker failure (rare)
  2. 2. HR tone volume control malfunction (occasional)
First check

Verify HR tone at all volume levels with SpO2 source

Test point
Audio output from SpO2 module
Expected
HR tone present at 100%, 75%, 50%, 25%, and 0% settings

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

HR tone or alarm volume controls not functional patient safety

Heart rate tone or alarm volume does not respond to level changes or fails to play at correct volume

Ranked causes

  1. 1. Audio amplifier or speaker failure (occasional)
  2. 2. Keypad volume adjustment buttons stuck or broken (occasional)
First check

Set HR Source to SpO2 on monitor

Test point
HR source selector
Expected
HR Source set to SpO2

4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

IBP calibration out of range patient safety

Invasive blood pressure readings do not match transducer zero or pressure source values

Ranked causes

  1. 1. Pressure transducer not calibrated or drift (common)
  2. 2. IBP analog electronics drift or failure (occasional)
First check

IBP1 zero calibration with pressure source

Test point
Pressure transducer output at 0 mmHg
Expected
0/0 (0) +/- 2 mmHg, +/- 2 mmHg

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

IBP channel disconnect messages present patient safety

IBP1 Disconnect or IBP2 Disconnect message appears on monitor display

Ranked causes

  1. 1. Pressure transducer not connected or disconnected (common)
  2. 2. Transducer malfunction or poor connection (occasional)
First check

Verify IBP1 and IBP2 disconnect messages appear

Test point
Monitor display
Expected
IBP1 Disconnect and IBP2 Disconnect messages

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

IBP disconnect message not clearing patient safety

Invasive blood pressure interface shows disconnect alert even with transducer connected

Ranked causes

  1. 1. Pressure transducer not connected or faulty (common)
  2. 2. IBP interface RF or sensor failure (occasional)
First check

IBP1 disconnect detection with transducer absent

Test point
Monitor display
Expected
IBP1 Disconnect message displayed

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

IBP simulator test failed patient safety

IBP1 or IBP2 waveforms do not display or reading does not match simulator arterial pressure 120/80 mmHg within tolerance

Ranked causes

  1. 1. Simulator channels not zeroed or calibrated (common)
  2. 2. Simulator or channel connection malfunction (occasional)
First check

Zero both simulator and monitor IBP channels

Test point
Simulator output and monitor input
Expected
Zero pressure established

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

IBP1 calibration zero error patient safety

IBP1 zero calibration does not read 0 mmHg or reads outside tolerance after calibration

Ranked causes

  1. 1. Transducer zero offset out of calibration (common)
  2. 2. Pressure transducer defective or contaminated (occasional)
First check

Calibrate IBP1 zero

Test point
IBP1 transducer output at 0 mmHg
Expected
0/0 (0) reading mmHg, +/- 2 mmHg

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

IBP1 pressure reading out of accuracy patient safety

IBP1 pressure reading does not match simulator or source setting within tolerance (+/- 12 mmHg at 300 mmHg)

Ranked causes

  1. 1. Transducer calibration error or drift (common)
  2. 2. Pressure source inaccurate or transducer defective (occasional)
First check

Set pressure source to 300 mmHg and verify IBP1 reading

Test point
IBP1 transducer output
Expected
300 mmHg reading, +/- 12 mmHg

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

IBP2 calibration zero error patient safety

IBP2 zero calibration does not read 0 mmHg or reads outside tolerance after calibration

Ranked causes

  1. 1. Transducer zero offset out of calibration (common)
  2. 2. Pressure transducer defective or contaminated (occasional)
First check

Calibrate IBP2 zero

Test point
IBP2 transducer output at 0 mmHg
Expected
0/0 (0) reading mmHg, +/- 2 mmHg

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

IBP2 pressure reading out of accuracy patient safety

IBP2 pressure reading does not match simulator or source setting within tolerance (+/- 12 mmHg at 300 mmHg)

Ranked causes

  1. 1. Transducer calibration error or drift (common)
  2. 2. Pressure source inaccurate or transducer defective (occasional)
First check

Set pressure source to 300 mmHg and verify IBP2 reading

Test point
IBP2 transducer output
Expected
300 mmHg reading, +/- 12 mmHg

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Invasive pressure reading not displaying or disconnect message persists patient safety

IBP1 or IBP2 channels show disconnect message and do not display pressure readings after connecting pressure transducer and source.

Ranked causes

  1. 1. Pressure transducer not connected or not recognized (common)
  2. 2. Faulty IBP pod or interface (occasional)
First check

Verify IBP1 and IBP2 disconnect messages appear

Test point
Monitor display
Expected
IBP1 Disconnect and IBP2 Disconnect messages

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP adult cuff inflation too slow patient safety

NIBP adult cuff does not inflate to >165 mmHg within 60 second time limit

Ranked causes

  1. 1. Pump flow insufficient or valve leak (common)
  2. 2. Cuff obstruction or tubing kinked (occasional)
First check

Time NIBP adult cuff inflation to >165 mmHg

Test point
Cuff pressure vs. elapsed time
Expected
Inflates to >165 mmHg in <60 seconds, <60 seconds

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP adult over-pressure protection failure patient safety

Monitor does not display "NIBP Over Press" message when cuff pressure exceeds 300 mmHg during adult test

Ranked causes

  1. 1. Pressure sensor malfunction or threshold out of calibration (common)
  2. 2. Display or alarm circuit malfunction (occasional)
First check

Create NIBP pressure above 300 mmHg and monitor for alarm message

Test point
Monitor display and pressure gauge
Expected
"NIBP Over Press" message appears mmHg

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP adult pressure reading out of range or accuracy patient safety

NIBP adult channel reads outside tolerance when simulator set to specified pressure and heart rate values

Ranked causes

  1. 1. Simulator settings incorrect or channel calibration drifted (common)
  2. 2. Simulator malfunction or channel signal processing error (occasional)
First check

Set simulator to adult test values and verify monitor reading at 150/100 mmHg

Test point
Monitor display and simulator settings
Expected
150/100/120 reading mmHg, +/- 10 mmHg
Tools
NIBP simulator

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP cuff inflation time out of specification patient safety

Neonatal or adult cuff inflates slower or faster than required

Ranked causes

  1. 1. NIBP pump or valve malfunction (occasional)
  2. 2. Adult NIBP pump or valve malfunction (occasional)
First check

Neonatal cuff inflation rate

Test point
NIBP cuff pressure
Expected
reaches >100 mmHg in <15 seconds, <15 seconds

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP cuff leak detected patient safety

NIBP cuff loses pressure during inflation cycle; leak test fails or reports excessive leakage

Ranked causes

  1. 1. Cuff or tubing leak (common)
  2. 2. Valve or pump seal deteriorated (occasional)
First check

Perform NIBP leak test with 200 mL fixed volume chamber

Test point
Pressure chamber volume and time
Expected
No gross leaks; Fail ≤ 2% of run cycles

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP flow rate low patient safety

NIBP system does not achieve required flow rate; flow meter shows insufficient flow during test

Ranked causes

  1. 1. Pump worn or filter clogged (common)
  2. 2. Tubing obstruction or valve malfunction (occasional)
First check

Measure NIBP system flow rate

Test point
Flow meter at pump outlet
Expected
About 1 LPM

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP neonatal cuff inflation too slow patient safety

NIBP neonatal cuff does not inflate to >100 mmHg within 15 second time limit

Ranked causes

  1. 1. Pump flow insufficient or valve leak (common)
  2. 2. Cuff obstruction or tubing kinked (occasional)
First check

Time NIBP neonatal cuff inflation to >100 mmHg

Test point
Cuff pressure vs. elapsed time
Expected
Inflates to >100 mmHg in <15 seconds, <15 seconds

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP neonatal over-pressure protection failure patient safety

Monitor does not display "NIBP Over Press" message when cuff pressure exceeds 150 mmHg during neonatal test

Ranked causes

  1. 1. Pressure sensor malfunction or threshold out of calibration (common)
  2. 2. Display or alarm circuit malfunction (occasional)
First check

Create NIBP pressure above 150 mmHg and monitor for alarm message

Test point
Monitor display and pressure gauge
Expected
"NIBP Over Press" message appears mmHg

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP neonatal pressure reading out of range or accuracy patient safety

NIBP neonatal channel reads outside tolerance when simulator set to specified pressure and heart rate values

Ranked causes

  1. 1. Simulator settings incorrect or channel calibration drifted (common)
  2. 2. Simulator malfunction or channel signal processing error (occasional)
First check

Set simulator to neonatal test values and verify monitor reading at 100/65 mmHg

Test point
Monitor display and simulator settings
Expected
100/65/240 reading mmHg, +/- 10 mmHg
Tools
NIBP simulator

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP Over Press message not displayed patient safety

Neonatal or adult cuff pressure exceeds safe limits without alerting operator

Ranked causes

  1. 1. NIBP over-pressure detection failure (unknown)
  2. 2. Adult NIBP over-pressure detection failure (unknown)
First check

Neonatal over-pressure alert detection

Test point
NIBP cuff
Expected
"NIBP Over Press" message displayed mmHg, above threshold

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • NIBP pressure reading out of specification at 150 mmHg
  • NIBP pressure sensor voltage out of specification
  • NIBP pump not starting
  • NIBP readings inaccurate or out of range
  • NIBP-DC pressure reading out of specification at 150 mmHg
  • ON/CHARGE LED not lighting correctly
  • POD charger status not indicating
  • POD charging indicator malfunction
  • POD charging indicators malfunction
  • Printer does not produce output
  • Respiration rate reading inaccurate
  • Sample line gas reading not displayed
  • Sample line gas readings drift over time
  • SpO2 Pod interface not responding
  • SpO2 Pod interface or probe detection malfunction
  • SpO2 readings out of accuracy specification
  • SpO2 waveform and HR tone verification
  • System pressure loss detected
  • System shows pressure loss during leak test
  • Temperature measurement inaccurate
  • Wireless communication loss between tablet, base and monitor
  • 1849 CO2/Agents Sample Line connectivity verification
  • 3881 ePOD not establishing connection
  • 3881 ePOD radio link not established
  • 3882 oPOD not establishing connection
  • 3882 oPOD SpO2 Probe Off message displayed
  • 3886 CO2 zeroing message does not clear
  • 3886 CO2 zeroing message not appearing or not clearing
  • 3886 does not detect breathing into sample line
  • 3886 gas calibration zero not starting
  • 3886 gas readings out of specification with calibration gas
  • 3886 multigas CO2 zeroing not completing
  • 3886 multigas unit CO2 zero message not clearing
  • 3886 Multigas Unit CO2 zeroing not completing
  • 3886 Multigas Unit initial verification
  • 3886 multigas unit wireless communication not established
  • 3886 multigas unit wireless connection lost
  • 3886 Multigas unit wireless connection not established
  • 3886 multigas wireless LED or module LED not lit
  • 3886 sample line port does not illuminate green or shows error message
  • 3886 sample line port does not respond to connection
  • 3886 sample line readings drift during closed loop
  • 3886 wireless communication not established
  • 3886 wireless LEDs or CONN AGENTS LED not illuminating
  • 3886 wireless or module connection LED not illuminated
  • Adjust Date and Time
  • Adult cuff inflation time
  • Adult cuff inflation time exceeds specification
  • Adult cuff inflation time out of specification
  • Adult cuff inflation time verification
  • Adult NIBP cuff inflation speed outside specification
  • Adult NIBP cuff inflation time exceeds specification
  • Adult NIBP cuff inflation time out of specification
  • Adult NIBP over-pressure alarm
  • Adult NIBP over-pressure alarm does not trigger
  • Adult NIBP over-pressure alarm not triggering
  • Adult NIBP over-pressure detection
  • Adult NIBP over-pressure not detected
  • Adult NIBP over-pressure safety alarm not activating
  • Adult NIBP overpressure alarm not triggering
  • Adult NIBP pressure and heart rate readings inaccurate
  • Adult NIBP range and accuracy at 100/65
  • Adult NIBP range and accuracy at 150/100
  • Adult NIBP range and accuracy at 60/30
  • Adult NIBP range and accuracy verification
  • Adult NIBP reading out of range or accuracy specification
  • Adult NIBP readings out of specification
  • Adult NIBP readings out of tolerance across pressure range
  • Adult NIBP readings outside tolerance range
  • ALARM OFF / STANDBY key not functional
  • ALARM OFF / STANDBY key not functioning
  • ALARM OFF or STANDBY key not functional
  • Alarm off/standby key not functional
  • Alarm off/standby key not responding
  • Alarm volume adjustment on Monitor
  • Alarm volume control not functional
  • Alarm volume control not responsive
  • Alarm volume does not adjust on monitor
  • Alarm volume does not adjust on tablet
  • Alarm volume not adjustable
  • Alarm volume not adjustable on monitor or tablet
  • Alarm volume not adjustable or non-functional
  • Alarm volume not adjustable to all levels
  • Alarm volume not functioning across levels
  • Anesthetic gas readings out of specification during calibration
  • Audio alarm cannot be disabled
  • AUDIO ALARM OFF / SILENCE key not functional
  • AUDIO ALARM OFF / SILENCE key not functioning
  • Audio alarm off or silence key not functional
  • Audio alarm silence function non-functional
  • Audio alarm silence key not functional
  • Audio alarm silence key not responding
  • Battery indicator and RF link verification (IBP)
  • Battery indicator and RF link verification (SpO2)
  • Battery indicator LED does not light orange during charging
  • Battery indicator not displayed on monitor screen
  • Battery indicator not displaying correctly
  • Battery indicator not showing in display
  • Battery indicator not showing or RF link unreliable beyond 10 meters
  • Battery indicator not visible or RF link not established
  • Battery indicator or RF link not functioning on invasive pressure pod
  • Battery indicator or RF link not functioning on SpO2 pod
  • Battery indicator shows incorrect status
  • Battery indicator verification
  • Battery indicator verification failure
  • Breathing into Sample Line CO2 detection test
  • Calibration: 3880 MRI Patient Monitor System
  • Calibration: 3886 Multigas Unit gas readings verification
  • Calibration: NIBP-DC offset voltage
  • Calibration: NIBP-DC pressure sensor offset
  • Calibration: Pressure Sensor NIBP-DC Offset
  • Calibration: Pressure Sensor Verification & Calibration
  • Closed loop sample gas circulation and stability test
  • Closed loop test shows excessive gas drift
  • CO2 activity not detected during breathing
  • CO2 activity not displayed on monitor
  • CO2 activity not displayed when breathing into sample line
  • CO2 reading incorrect
  • CO2 reading out of calibration
  • CO2 reading out of specification
  • CO2 suction flow out of specification
  • CO2 suction flow rate not within specification
  • CO2 suction flow rate out of specification
  • CO2 suction flow rate out of tolerance
  • CO2 suction flow rate outside specification
  • CO2 suction flow rate verification
  • CO2 suction flow verification
  • CO2 zeroing message does not clear
  • CO2 zeroing message does not clear after 20 seconds
  • CO2 zeroing process does not complete
  • Comm Loss message and alarm on Tablet
  • Communication loss between Base Station and Tablet
  • Communication loss between devices
  • Communication loss between Tablet and Base Station
  • Communication loss between tablet, base and monitor
  • Communication loss between Tablet, Base Station, and Monitor
  • Communication loss message does not appear when base station channel changed
  • Communication loss message not triggered on channel change
  • Communication loss not detected after channel change
  • Communication not restored after channel alignment
  • Communication not restored after channel realignment
  • Communication not restored after channel synchronization
  • CONNECTED LED on ipPOD not lit
  • Date and time not set correctly
  • DES (Desflurane AA1) reading out of specification
  • Device communication not established
  • Diastolic Low Limit adjustment not working
  • Display physical inspection
  • Display touch function not responding
  • Display touch function verification
  • Display touch response not working
  • Display touchscreen responsiveness verification
  • Display window damage inspection
  • Display window damaged or scratched
  • ECG heart rate alarm settings not adjustable
  • ECG heart rate limits not adjustable
  • ECG HR limit adjustment not functional
  • ECG HR Limit adjustment not functioning
  • ECG lead disconnection not detected
  • ECG Lead Fail alarm not triggered
  • ECG Lead Fail alarm not triggered on lead removal
  • ECG Lead Fail message displayed
  • ECG lead fail message displayed without lead disconnection
  • ECG Lead Fail message does not appear on lead disconnect
  • ECG Lead Fail message not displayed on disconnection
  • ECG Lead Fail message present but RA/R lead is connected
  • ECG lead fault not detected
  • ECG lead not detected
  • ECG lead recovery not functioning
  • ECG lead restoration not detected
  • ECG lead selection not responding
  • ECG POD CONNECTED LED not illuminated
  • ECG Pod RF link not working
  • ECG RF link not established
  • ECG waveform and HR not restored after lead reconnection
  • ECG waveform not detected on specific leads
  • ECG waveform not displayed
  • ECG waveform not displaying or lead failure not detected
  • ePOD CONNECTED LED not illuminated or RF link not established
  • Gas calibration readings out of specification
  • Gas calibration values out of specification
  • Gas calibration zero not starting
  • Gas calibration zero procedure fails
  • Gas calibration zero start not responding
  • Gas readings do not hold stable in closed loop test
  • Gas readings drift during closed loop test
  • Gas readings drift during verification
  • Gas readings drift in closed-loop sample test
  • Gas readings out of specification
  • Gas readings out of specification during calibration
  • Gas readings out of specification during verification
  • Gas sample line port does not activate
  • Gas sample line port not responding to connection
  • Gas verification calibration readings out of specification
  • Heart rate tone levels not adjustable or non-functional
  • Heart rate tone not functioning across volume levels
  • Heart rate tone not functioning at all levels
  • Heart rate tone not functioning at various levels
  • Heart rate tone not producing at various volumes
  • Heart rate tone not working at all levels
  • Heart rate tone volume levels incorrect or missing
  • HR Limit bar adjustment not functioning
  • HR Limit bar-line not adjustable
  • HR limit controls not functional
  • HR Limit settings not adjustable
  • HR tone audible volume not adjustable
  • HR tone independent control between Tablet and Monitor
  • HR tone not audible at all volume levels
  • HR tone not audible on tablet speaker
  • HR tone or alarm volume not functioning
  • HR tone volume control not responsive
  • IBP arterial pressure simulation
  • IBP battery indicator not showing or RF link insufficient
  • IBP battery indicator not showing or RF link unreliable beyond 10 meters
  • IBP channels zero calibration and waveforms using simulator
  • IBP disconnect message does not clear when transducer connected
  • IBP disconnect message not appearing or not clearing
  • IBP disconnect message verification
  • IBP disconnect messages not appearing or not clearing
  • IBP disconnect messages not displayed
  • IBP disconnect messages persist after transducer connection
  • IBP high pressure reading out of tolerance
  • IBP module CONNECTED LED not illuminated
  • IBP module LED verification
  • IBP module not connected or detected
  • IBP pod CONNECTED LED not illuminating
  • IBP pod not showing connected status
  • IBP RF link fails to establish or drops within 10 meters
  • IBP simulator arterial pressure reading out of tolerance
  • IBP simulator arterial reading out of tolerance
  • IBP simulator arterial waveform out of specification
  • IBP simulator reading does not match expected arterial pressure
  • IBP simulator test: waveforms not displaying or inaccurate
  • IBP simulator waveform not displayed
  • IBP simulator waveform not displaying
  • IBP simulator waveform not displaying or out of range
  • IBP transducer connection verification
  • IBP waveform verification using simulator
  • IBP waveforms not displayed on both channels
  • IBP waveforms not displaying on simulator connection
  • IBP waveforms not displaying via simulator
  • IBP zero calibration out of tolerance
  • IBP1 accuracy at 300 mmHg
  • IBP1 accuracy at 300 mmHg using transducer
  • IBP1 and IBP2 disconnect message verification
  • IBP1 and IBP2 waveforms not displayed with simulator
  • IBP1 and IBP2 waveforms not displaying
  • IBP1 calibration zero out of specification
  • IBP1 calibration zero reading out of tolerance
  • IBP1 disconnect message does not appear or clear
  • IBP1 disconnect message not clearing
  • IBP1 high pressure reading out of specification
  • IBP1 high pressure reading out of tolerance
  • IBP1 pressure accuracy out of specification
  • IBP1 pressure accuracy out of tolerance at 300 mmHg
  • IBP1 pressure reading at 300 mmHg out of tolerance
  • IBP1 pressure reading out of tolerance at 300 mmHg
  • IBP1 reading inaccurate at 300 mmHg
  • IBP1 shows disconnect message
  • IBP1 zero calibration and pressure verification
  • IBP1 zero calibration out of specification
  • IBP1 zero calibration out of tolerance
  • IBP1 zero calibration reading out of tolerance
  • IBP1 zero calibration using transducer
  • IBP2 accuracy at 300 mmHg
  • IBP2 accuracy at 300 mmHg using transducer
  • IBP2 calibration verification
  • IBP2 calibration zero out of specification
  • IBP2 calibration zero reading out of tolerance
  • IBP2 high pressure reading out of specification
  • IBP2 high pressure reading out of tolerance
  • IBP2 pressure accuracy out of specification
  • IBP2 pressure accuracy out of tolerance at 300 mmHg
  • IBP2 pressure accuracy verification
  • IBP2 pressure reading at 300 mmHg out of tolerance
  • IBP2 pressure reading out of tolerance at 300 mmHg
  • IBP2 reading inaccurate at 300 mmHg
  • IBP2 zero calibration and pressure verification
  • IBP2 zero calibration out of specification
  • IBP2 zero calibration out of tolerance
  • IBP2 zero calibration reading out of tolerance
  • IBP2 zero calibration using transducer
  • Incorrect waveform detected on specific ECG leads
  • Invasive arterial pressure reading outside tolerance
  • Invasive Pressure module CONNECTED LED not illuminated
  • Invasive Pressure Pod (ipPOD) not connecting
  • Invasive pressure pod CONNECTED LED does not illuminate
  • Invasive Pressure Pod not connected or inoperable
  • Invasive pressure pod not connected or LED not illuminated
  • Invasive pressure pod not connecting or LED not illuminating
  • Invasive pressure pod RF link not operational beyond 10 meters
  • Invasive pressure reading outside tolerance at high pressure setting
  • Invasive pressure reading outside tolerance at zero calibration
  • Invasive pressure waveforms not displaying on monitor
  • ipPOD battery indicator not visible or RF link beyond 10 meters
  • ipPOD battery indicator or RF link out of range
  • ipPOD CONNECTED LED not illuminating
  • Monitor alarm volume levels not adjustable or inaudible
  • Monitor alarm volume not adjustable
  • Multigas calibration drift or reading inaccuracy
  • Multigas calibration out of specification
  • Multigas calibration verification out of specification
  • Multigas closed loop readings drift beyond specification
  • Multigas closed-loop calibration readings drift
  • Multigas readings drift during closed loop test
  • Multigas sample line not detected
  • Multigas unit calibration gas readings out of specification
  • Multigas unit CO2 activity not displayed
  • Multigas unit CO2 zeroing failure
  • Multigas unit sample line connection failure
  • Multigas unit wireless communication failure
  • Multigas unit wireless connectivity failure
  • Multiple ECG leads not detected
  • N2O reading out of specification
  • Nasal Cannula CO2 and respiration not detected
  • Nasal cannula CO2 and respiration not reading
  • Neonatal cuff inflation time
  • Neonatal cuff inflation time exceeds specification
  • Neonatal cuff inflation time out of specification
  • Neonatal cuff inflation time verification
  • Neonatal NIBP cuff inflation speed outside specification
  • Neonatal NIBP cuff inflation time exceeds specification
  • Neonatal NIBP cuff inflation time out of specification
  • Neonatal NIBP over-pressure alarm
  • Neonatal NIBP over-pressure alarm does not trigger
  • Neonatal NIBP over-pressure alarm not triggering
  • Neonatal NIBP over-pressure detection
  • Neonatal NIBP over-pressure not detected
  • Neonatal NIBP over-pressure safety alarm not activating
  • Neonatal NIBP overpressure alarm not triggering
  • Neonatal NIBP pressure and heart rate readings inaccurate
  • Neonatal NIBP range and accuracy at 100/65
  • Neonatal NIBP range and accuracy at 60/30
  • Neonatal NIBP range and accuracy at 80/50
  • Neonatal NIBP range and accuracy verification
  • Neonatal NIBP reading out of range or accuracy specification
  • Neonatal NIBP readings out of specification
  • Neonatal NIBP readings out of tolerance across pressure range
  • Neonatal NIBP readings outside tolerance range
  • NIBP AC offset out of specification
  • NIBP adult cuff inflation time
  • NIBP adult cuff inflation time exceeds limit
  • NIBP adult cuff inflation time out of specification
  • NIBP adult cuff inflation time too long
  • NIBP adult over-pressure alarm does not trigger
  • NIBP adult over-pressure alarm not functioning
  • NIBP adult over-pressure alarm not triggering
  • NIBP adult over-pressure protection does not trigger
  • NIBP adult pressure and heart rate readings inaccurate
  • NIBP adult pressure and heart rate readings out of tolerance
  • NIBP adult pressure reading out of tolerance at 150/100
  • NIBP adult readings do not meet range and accuracy specifications
  • NIBP cuff does not inflate or grossly leaks
  • NIBP cuff leaking during cycle test
  • NIBP cuff leaking or unable to hold pressure
  • NIBP DC offset calibration
  • NIBP dead man function not working
  • NIBP dead man reading not zero
  • NIBP dead man test
  • NIBP dead man test does not read zero
  • NIBP dead man test failed
  • NIBP dead man test fails
  • NIBP dead man test fails to show zero reading
  • NIBP dead man test failure
  • NIBP dead man test not showing zero reading
  • NIBP dead man test reading not zero
  • NIBP dead man test shows non-zero reading
  • NIBP flow rate below specification
  • NIBP flow rate incorrect
  • NIBP flow rate not within specification
  • NIBP flow rate out of specification
  • NIBP flow rate outside specification
  • NIBP flow rate verification
  • NIBP gross leaks detected or leak test failures exceed tolerance
  • NIBP integrity test fails or shows excessive leaks
  • NIBP key does not start pump
  • NIBP leak detection failure
  • NIBP leak test
  • NIBP module leaks during cycle test
  • NIBP neonatal cuff inflation time
  • NIBP neonatal cuff inflation time exceeds limit
  • NIBP neonatal cuff inflation time out of specification
  • NIBP neonatal cuff inflation time too long
  • NIBP neonatal over-pressure alarm does not trigger
  • NIBP neonatal over-pressure alarm not functioning
  • NIBP neonatal over-pressure alarm not triggering
  • NIBP neonatal over-pressure protection does not trigger
  • NIBP neonatal pressure and heart rate readings inaccurate
  • NIBP neonatal pressure and heart rate readings out of tolerance
  • NIBP neonatal pressure reading out of tolerance at 100/65
  • NIBP neonatal readings do not meet range and accuracy specifications
  • NIBP over-pressure alarm triggered
  • NIBP over-pressure detected
  • NIBP pressure reading out of specification
  • NIBP pressure reading out of specification at 250 mmHg
  • NIBP pressure sensor out of calibration
  • NIBP pump does not activate
  • NIBP pump does not start
  • NIBP pump does not start from NIBP key
  • NIBP pump not operating
  • NIBP pump not responding to key press
  • NIBP pump operation
  • NIBP system has excessive leaks
  • NIBP system has gross leaks or excessive failure rate
  • NIBP system leaking during cuff cycle test
  • NIBP system shows gross leaks or excessive test failures
  • NIBP-AC measurement out of specification
  • NIBP-DC voltage out of specification
  • No CO2 activity detected during breathing into sample line
  • No CO2 activity detected from breathing sample
  • No communication between Tablet, Base, and Monitor
  • No time displayed on top row
  • O2 reading out of specification
  • oPOD CONNECTED LED not illuminated or SpO2 probe off message displayed
  • oPOD not connecting to Monitor
  • Out of paper message not displaying
  • Out of paper message not triggered
  • Paper out detected in printer
  • POD charger LED status verification
  • POD charger LEDs flashing abnormally
  • POD charger LEDs not functioning as expected
  • POD charging indicators not functioning
  • POD charging LED not lighting continuously
  • Pressure Sensor Verification and Calibration
  • Printer does not produce output after paper installation
  • Printer not producing output
  • Printer not responding
  • Printer not responding to record command
  • Printer out of paper
  • Printer out of paper alarm
  • Printer out of paper alert not displayed
  • Printer out of paper condition not detected
  • Printer Out of Paper detection
  • Printer out of paper detection not functioning
  • Printer out of paper message does not display
  • Printer out of paper message not generated
  • Printer out of paper not detected
  • Printer out of paper or paper jam
  • Printer out-of-paper message not displayed
  • Printer paper detection
  • Printer paper installation and operation
  • Printer paper installation and printout verification
  • Printer paper out of service
  • Printer printout not produced
  • Printer reports out of paper when paper installed
  • Radio link not established on ePOD
  • Respiration rate out of specification
  • Respiration rate reading incorrect
  • Respiration rate reading out of specification
  • RF link fails to establish or drops within 10 meters
  • Sample line connection not detected or port light off
  • Sample line port does not light green
  • Sample line port does not light green when connected
  • Sample line port does not light or lights wrong color
  • Sample line port not responding or displaying wrong color
  • Selected ECG lead not detected
  • Simulator arterial waveform reading not within specification
  • Specific ECG leads not detected
  • SpO2 accuracy at 100% saturation
  • SpO2 accuracy at 70% saturation
  • SpO2 accuracy at 85% saturation
  • SpO2 alarm threshold and Monitor alarm volume verification
  • SpO2 battery indicator not showing
  • SpO2 battery indicator not visible
  • SpO2 module not searching for signal
  • SpO2 module not searching or displaying waveform
  • SpO2 monitor does not display waveform after attachment to simulator
  • SpO2 monitor does not display waveform after simulator attachment
  • SpO2 monitor does not search or display waveform
  • SpO2 monitor not displaying waveform or searching indicator
  • SpO2 monitoring not operational
  • SpO2 POD CONNECTED LED not illuminated
  • SpO2 Pod connection or probe status not displayed
  • SpO2 POD not connecting to monitor
  • SpO2 readings do not match simulator at 100% and 240 bpm
  • SpO2 readings do not match simulator at 100% SpO2
  • SpO2 readings do not match simulator at 70% and 30 bpm
  • SpO2 readings do not match simulator at 70% SpO2
  • SpO2 readings do not match simulator at 85% and 150 bpm
  • SpO2 readings do not match simulator at 85% SpO2
  • SpO2 readings do not match simulator settings
  • SpO2 readings do not match simulator settings at 100% / 240bpm
  • SpO2 readings do not match simulator settings at 70% / 30bpm
  • SpO2 readings do not match simulator settings at 85% / 150bpm
  • SpO2 readings do not match simulator values
  • SpO2 readings out of tolerance
  • SpO2 readings outside tolerance range
  • SpO2 RF link does not reach 10 meters
  • SpO2 sensor not searching or displaying waveform
  • SpO2 sensor not searching or waveform missing
  • STANDBY and Alarm OFF mode verification
  • System date and time incorrect
  • System pressure loss during verification
  • Tablet alarm volume levels not adjustable or inaudible
  • Tablet alarm volume not adjustable
  • Tablet Alarm Volume verification and audio alarm silencing test
  • Tablet heart rate tone not functioning
  • Tablet, Base and Monitor communication loss
  • Tablet, Base and Monitor communication lost
  • Tablet, base station, and monitor communication not established
  • Tablet, Base, and Monitor communication not established
  • Temperature accuracy out of specification
  • Temperature option not turning on
  • Temperature probe does not report failure when disconnected
  • Temperature probe does not show fail indication
  • Temperature probe fail alert with no sensor connected
  • Temperature probe fail indication not triggered
  • Temperature probe fail not detected
  • Temperature probe failure not detected
  • Temperature reading inaccurate
  • Temperature reading out of specification
  • Temperature sensor fail message does not appear
  • Temperature sensor not responding
  • Temperature sensor reading inaccurate
  • Time and Date incorrect or not updating
  • Time and Date not current
  • Time and date not set correctly
  • Time and Date settings incorrect
  • Time not displayed on monitor
  • Time not displayed on Tablet
  • Trend History log key not functional
  • Trend history log key not responding
  • TREND log display verification
  • Waveform not detected on selected leads
  • Wireless communication lost between devices

525 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
Comm LossCommunication lost between Tablet and Base StationBase Station channel setting differs from Tablet channelwarning
ECG Lead FailECG lead disconnected or not properly connected to simulatorRA/R lead removed or disconnected from Fluke ProSim 8 Simulator during ECG verificationwarning
NIBP Over PressNon-invasive blood pressure measurement exceeded safe pressure limitCuff pressure exceeds safe threshold during neonatal blood pressure measurementwarning
Out of PaperPrinter paper roll is empty or depletedRECORD key pressed when no paper is loaded in Base Station printerwarning
SpO2 Probe OffSpO2 probe is not connected or not making proper contact with 3882 oPODSpO2 probe disconnected or oPOD powered on without probe connectedwarning
STANDBY, All Alarms OFFMonitor entered standby mode with all alarms disabledALARM OFF / STANDBY key pressedinfo

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI