IRadimed 3860+: troubleshooting and error codes
The ServiceForge knowledge base holds 121 documented symptoms for the IRadimed 3860+, with 150 ranked causes, 221 diagnostic checks with test points and expected values, and 61 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Pump does not start infusion or flow rate incorrect patient safety
Pump fails to begin infusion after Start button pressed, or measured flow rate does not match set rate
Ranked causes
- 1. IV administration set not installed correctly (common)
- 2. Pump rate or VTBI not set correctly (common)
- 3. Inlet or patient line occlusion preventing flow (occasional)
- 4. Pump mechanism failure or motor malfunction (rare)
Verify IV set installed and door closed properly
- Test point
- Pump door and IV set seating
- Expected
- Door closes and latches, IV set seated flush in pump chamber
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display not functioning or degraded image caution
LCD display fails to show information or displays garbled or dim image
Ranked causes
- 1. Backlight failure or 5V supply issue (common)
- 2. RF shield mesh displacement or damage (occasional)
- 3. LCD controller (1330 type) failure (rare)
Verify 5V supply voltage to display connector
- Test point
- Display power connector pin 1 (positive) and pin 2 (ground)
- Expected
- 5.0 V DC, +/-0.25 V
- Tools
- Digital multimeter
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow rate outside acceptable tolerance patient safety
Measured infusion flow rate deviates more than 5% from set rate or specified tolerance
Ranked causes
- 1. IV solution viscosity or temperature different from test conditions (occasional)
- 2. Pump calibration drift or mechanical wear (occasional)
- 3. IV set or tubing restriction or diameter variation (occasional)
Prepare IV set with water at room temperature
- Test point
- IV solution container and tubing
- Expected
- Water filled to 2/3 in drip chamber; tubing primed and air-free; water container hung 21 inches above pump
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pumping chamber door misalignment or improper fit caution
Door does not fit flush with case or has friction/resistance opening and closing
Ranked causes
- 1. Door hinge pin corrosion or wear (occasional)
- 2. Door latch axle screw loose (common)
- 3. Brass door axle corroded or obstructed (occasional)
Inspect door flush fit and alignment
- Test point
- Pumping chamber door AS112 or AS113 top, bottom, and side edges
- Expected
- Door fits flush with case on all sides
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air in line detection failure patient safety
Pump does not detect air bubble or indicate Air In Line alarm within 50 seconds when running IV set without fluid
Ranked causes
- 1. Bubble detector (HE09/HE09A) dirty or blocked (common)
- 2. Bubble detector optical sensor malfunction (occasional)
Test bubble detector air detection capability
- Test point
- Pump running at 10 mL/Hr with IV set containing no fluid
- Expected
- Air In Line alarm within 50 seconds
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm sound not functioning patient safety
Audible alarm does not sound during alarm conditions or alarm test
Ranked causes
- 1. Alarm sound disabled or muted in settings (occasional)
- 2. Speaker or audio circuit failure (rare)
Momentarily press Off switch and listen for alarm audio
- Test point
- Pump speaker and audio circuit
- Expected
- Audible alarm sound heard; pump remains powered on
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery pack mechanical damage or puncture patient safety
1133 Battery Pack damaged by impact, dropping, overheating, compression, or puncture to internal cells, potentially causing overheating, smoking, or catastrophic battery failure.
Ranked causes
- 1. Impact damage, dropping, compression, or mechanical shock (occasional)
- 2. Puncture to internal battery cells from sharp objects (occasional)
Inspect battery pack for cracks, dents, or visible physical damage
- Test point
- 1133 Battery Pack case exterior
- Expected
- No cracks, dents, deformation, or damage visible
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bubble detection alarm not triggered or functioning patient safety
Pump fails to detect air bubbles in line or alarm does not activate when bubble passes through pump
Ranked causes
- 1. Bubble detector sensor malfunction or obstructed (rare)
- 2. Alarm display or LED circuit failure (rare)
Set pump rate to 500 ml/hr and VTBI to 100 ml
- Test point
- Display settings
- Expected
- Display shows Rate: 500 ml and VTBI: 100 ml; Ready to Start flashing
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pump flow accuracy
Factory calibration of 3860+ and 3861 pump flow rate using Sartorius GD603 scale and encoder wheel alignment to achieve accurate delivery profile at various flow rates.
Ranked causes
- 1. Encoder wheel misalignment relative to pumping finger position (occasional)
- 2. Offset adjustment needed for step size correction (occasional)
Verify raw curve taper using SartoConnect data
- Test point
- Scale measurement during calibration
- Expected
- Ideal curve per Figure A with appropriate taper profile
- Tools
- Sartorius GD603 scale, SartoConnect software, Microsoft Excel
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Door alarm on flow detection failure patient safety
Alarm raised when flow stop switch does not respond within 0.75 seconds after door opens
Ranked causes
- 1. Free-flow detector switch (IVPAB016) not responding properly (occasional)
- 2. Door switch not opening properly (occasional)
Measure flow stop switch opening time after door open
- Test point
- Flow Stop Switch IVPAB016 signal
- Expected
- Switch opens within 0.75 second of door switch detecting open
- Tools
- Timer or oscilloscope
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display not clear or unreadable
Boot display appears but text is garbled, dim, or pixels not functioning
Ranked causes
- 1. Display window or overlay scratched or cracked (occasional)
- 2. Internal display module failure or LCD connection loose (rare)
Inspect display window covering and overlay for physical damage
- Test point
- Front display area
- Expected
- No scratches, cracks, or damage to window or overlay
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Door latch or clamp mechanism malfunction caution
Door handle/cam lock does not operate smoothly, does not latch securely, or does not release properly
Ranked causes
- 1. Latch cam worn or misaligned (occasional)
- 2. Door clamp latch paw worn or damaged (common)
Verify door handle/cam lock ease of operation
- Test point
- Door latch handle/cam AS104
- Expected
- Smooth, easy operation without binding or resistance
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Door open alarm during pumping patient safety
Alarm triggered and pump shuts down when door is detected open during infusion operation
Ranked causes
- 1. Door not fully latched in closed position (common)
- 2. Door switch or optic sensor malfunction (occasional)
Verify door closure and latch engagement
- Test point
- Door clamp mechanism
- Expected
- Hard rubber pin fully contacts switch or optic sensor when latched
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fluid delivery rate inaccurate patient safety
Pump delivers less than 100 mL in measured time period when set to 100 mL/Hr; flow rate outside specification
Ranked causes
- 1. Motor speed out of specification (common)
- 2. Pumping gasket worn or leaking (occasional)
Measure fluid delivery at 100 mL/Hr over calibrated time period
- Test point
- Graduated cylinder or Infutest system; pump outlet
- Expected
- Delivered volume matches 100 mL/Hr setting within manufacturer tolerance
- Tools
- 100 mL and/or 25 mL graduated cylinder, Infutest or calibrated fluid measurement system
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inlet occlusion alarm not triggered or functioning patient safety
Pump fails to detect inlet line obstruction or alarm does not activate when inlet tubing pinched
Ranked causes
- 1. Inlet occlusion detection sensor malfunction (rare)
- 2. Alarm display or LED circuit failure (rare)
Start pump with water flowing
- Test point
- Display and LED indicators
- Expected
- Pulsing > symbol in upper display; Green flashing LED in door; water flowing
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
IV set not properly seated in Flo-Stop holder patient safety
IV set does not remain seated in Flo-Stop holder when door opened; 'Check Door' alarm occurs inappropriately
Ranked causes
- 1. Flo-Stop holder plastic damaged by drug degradation (occasional)
- 2. Flo-Stop holder recess clogged or obstructed (common)
Inspect Flo-Stop holder for damage and obstruction
- Test point
- IVP1047 or AS120 Flo-Stop holder recess
- Expected
- No damage, obstructions, or fluid spillage; clean condition
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient occlusion alarm not triggered or functioning patient safety
Pump fails to detect outlet line obstruction or alarm does not activate when distal tubing blocked
Ranked causes
- 1. Patient occlusion detection sensor malfunction (rare)
- 2. Alarm display or LED circuit failure (rare)
Clear any inlet occlusion, start pump with water flowing
- Test point
- Display and LED indicators
- Expected
- Pulsing > symbol in upper display; Green flashing LED in door; water flowing through distal tubing
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power indicator LED not illuminated
LED below power on/off switches does not light when External DC Power/Charger connected or battery installed
Ranked causes
- 1. External DC Power/Charger not connected or connectors loose (common)
- 2. Battery pack not installed (common)
Verify External DC Power/Charger connected and line power active
- Test point
- DC input connector on pump
- Expected
- Connectors screwed in tightly, line power connected
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Prevent battery pack damage patient safety
Precautions to avoid internal damage to 1133 Battery Pack that could disable protective circuitry and cause catastrophic failure
Ranked causes
- 1. Physical damage from impact, dropping, overheating, compression, or mechanical abuse (common)
- 2. Disassembly or tool damage to battery pack (occasional)
Inspect battery pack for damage
- Test point
- 1133 Battery Pack exterior surfaces
- Expected
- No cracks, dents, punctures, corrosion, or signs of impact or thermal damage
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump does not initialize or start
Pump fails to power on or display initial boot sequence after pressing power button
Ranked causes
- 1. Battery not charged sufficiently (common)
- 2. Power button not pressed or held correctly (occasional)
Verify battery has been charged for minimum 6 hours
- Test point
- Battery pack 1133
- Expected
- Battery fully charged
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pumping chamber door and latch mechanical inspection
Verify pumping chamber door (AS112AOP or AS113AOP) fits flush with case, latches securely, and operates freely without friction.
Ranked causes
- 1. Door misalignment, binding, or latch failure (occasional)
- 2. Worn or damaged upper door clamp latch paw (IVP1038) (occasional)
Pumping chamber door flush fit inspection
- Test point
- Door edges at top, bottom, and sides
- Expected
- Door fits flush with case; no gaps
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pumping mechanism integrity test
Test pumping chamber door springs (HM10) and seal integrity by occluding IV set and verifying proper pinch-off and occlusion pressure maintenance.
Ranked causes
- 1. Worn or damaged pumping gasket/seal (IVP1070) (common)
- 2. Weak or broken extension springs (HM10) (occasional)
Pumping mechanism occlusion pressure test
- Test point
- IV set tube occlusion by pump
- Expected
- Pump reaches 10 PSI maximum occlusion pressure without backflow or free flow leakage
- Tools
- iRadimed Series 1000 IV set, pressure gauge
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pumping mechanism weak or occlusion failure patient safety
IV set not pinched off or pump cannot build occlusion pressure when patient tubing clamp is closed
Ranked causes
- 1. Extension spring (HM10) worn or weak (common)
- 2. Pumping fingers worn or misaligned (occasional)
Test pumping mechanism integrity and spring force
- Test point
- IV set loaded in pump during Pump mode operation
- Expected
- Pump achieves 10 PSI occlusion pressure with patient tube clamp closed
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Swollen battery stuck in pump or remote patient safety
1133 Battery Pack expands excessively and becomes jammed or difficult to remove from 3860+ pump or 3865 remote display chassis
Ranked causes
- 1. Excessive battery expansion causing jamming (occasional)
- 2. Battery plastic case beginning to crack during removal attempts (rare)
Assess degree of battery swelling and jamming
- Test point
- Battery fit in chassis
- Expected
- Battery moves slightly when pried with gentle screwdriver pressure
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air bubble detected in IV line patient safety
Unit measures air bubble of 100µL or larger in infusion line and triggers alarm when acoustic impedance increases due to air-liquid interface
Ranked causes
- 1. Air bolus in IV tubing (common)
Verify acoustic sensor detects air bubble presence
- Test point
- Acoustic impedance detection element
- Expected
- Voltage signal reduced when air present compared to fluid-filled state
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery cell voltage imbalance
Battery cells not within 0.2 V of each other or any cell outside 3.8 to 4.25 V range
Ranked causes
- 1. Battery cell imbalance or failure (occasional)
Verify all battery cells within range and balanced
- Test point
- Service mode F2 Battery Status display; all 4 cell voltages
- Expected
- 3.8 to 4.25 V per cell, all within 0.2 V of each other, 3.8-4.25, delta 0.2
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery discharge current abnormal
Battery discharge current significantly different from approximately -240 mA when charger disconnected
Ranked causes
- 1. Battery or discharge circuit fault (rare)
Verify battery discharge current with charger disconnected
- Test point
- Service mode F2 Battery Status display; Current field
- Expected
- Approximately -240 mA, nominal -240
Battery pack protective circuit failure or damage patient safety
1133 Battery Pack integral safety circuit damaged or failed, unable to prevent catastrophic battery failure.
Ranked causes
- 1. Physical damage to battery pack compromising protective circuitry (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery pack swollen or expanded patient safety
1133 Battery Pack case expands or swells due to internal gas release from aging lithium-polymer cells, potentially causing case failure, cell failure, or safety circuit failure.
Ranked causes
- 1. Aged lithium-polymer cells releasing internal gas (common)
Inspect battery pack for visible swelling or expansion of case
- Test point
- 1133 Battery Pack exterior
- Expected
- No visible swelling or case expansion
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery pack swollen or expanded patient safety
1133 Battery Pack case visibly expands or swells due to internal gas release from aging lithium-polymer cells
Ranked causes
- 1. Battery pack aging and internal gas release (common)
Inspect battery case for visible swelling or expansion
- Test point
- 1133 Battery Pack exterior
- Expected
- Smooth, flat battery case with no bulging
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery shows swelling patient safety
Battery pack exhibits visible swelling or deformation indicating internal damage
Ranked causes
- 1. Internal battery degradation or short circuit (rare)
Visual inspection for battery swelling
- Test point
- Battery pack physical appearance
- Expected
- No visible swelling or deformation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery state of charge low
Battery state of charge below 50% before testing
Ranked causes
- 1. Battery not pre-charged before test (common)
Verify battery state of charge
- Test point
- Service mode F2 Battery Status display
- Expected
- 50 to 100 %, 50-100
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery terminal voltage out of range
Battery terminal voltage outside 15.0 to 16.9 V range during service mode test
Ranked causes
- 1. Battery faulty or not fully charged (occasional)
Verify battery terminal voltage
- Test point
- Service mode F2 Battery Status display with charger cable connected
- Expected
- 15 to 16.9 V, 15-16.9
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bubble detection not functioning
Bubble detection indicator (upper case B or lower case b) not displaying correctly on screen
Ranked causes
- 1. Bubble detection sensor malfunction or calibration issue (unknown)
Verify bubble detection indicator with no IV Set installed
- Test point
- Screen bottom center, bubble indicator
- Expected
- Upper case B displayed
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Bubble detection not working correctly patient safety
Bubble detection indicator at bottom center of screen does not show uppercase B with no IV set or lowercase b with fluid-filled set installed
Ranked causes
- 1. Bubble detection sensor malfunction (unknown)
Verify bubble detection with no IV set installed
- Test point
- Bottom center of screen indicator
- Expected
- uppercase B
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Bubble detector functional test patient safety
Verify bubble detector (HE09A) ability to detect air in line within specified time.
Ranked causes
- 1. Bubble detector failure or malfunction (occasional)
Air in line detection time
- Test point
- Bubble detector response with empty IV set running at 10 mL/Hr
- Expected
- Air In Line alarm within 50 seconds
- Tools
- IV set, pump operation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Backup battery replacement
Replace CR2032 coin cell memory/clock backup battery to ensure proper timekeeping and history record retention.
Ranked causes
- 1. CR2032 backup battery depleted or aged (common)
Backup battery replacement interval
- Test point
- Battery age and condition record
- Expected
- Replace every two years or if 'Lost History' message appears
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Battery MRI environment
Verification of backup battery functionality during service mode
Ranked causes
- 1. Battery out of specification (unknown)
Verify battery charge and functionality
- Test point
- Battery
- Expected
- Battery meets specification
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Battery status verification patient safety
Verify battery state of charge, terminal voltage, individual cell voltages, current draw, and cycle count meet specifications.
Ranked causes
- 1. Battery voltage or charge state out of specification (occasional)
Battery state of charge with charger connected
- Test point
- F2 key display: State of Charge (%)
- Expected
- 50 to 100%
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Bubble detection MRI environment
Verification of bubble detection functionality during service mode
Ranked causes
- 1. Bubble detector out of specification (unknown)
Verify bubble detection sensor response
- Test point
- Bubble detection sensor
- Expected
- Sensor detects bubbles correctly
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Flow preventer switch MRI environment
Verification of flow preventer switch operation during normal operating mode
Ranked causes
- 1. Flow preventer switch malfunction (unknown)
Verify flow preventer switch operation
- Test point
- Flow preventer switch
- Expected
- Switch functions correctly
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Flow stop pull test MRI environment
Verification of flow stop mechanism under pull load during normal operating mode
Ranked causes
- 1. Flow stop mechanism failure (unknown)
Apply pull force and verify flow stops
- Test point
- Flow stop mechanism
- Expected
- Flow stops under pull test load
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Flow stop pull test patient safety
Verify black pinch clamp full retraction and flow preventer housing retention
Ranked causes
- 1. IVP1068 Door Clamp Clip worn (occasional)
Load IVP1056 IV set and close door
- Test point
- Door and flow stop carrier
- Expected
- Door closes fully
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Infusion pump flow rate patient safety
Verify pump delivers correct flow rates at multiple programmed settings with volume and rate tolerance bands
Ranked causes
- 1. Pump flow rate out of specification (unknown)
Measure delivered volume at 1400 ml/hr
- Test point
- Pump output
- Expected
- 44.27 – 48.93 ml or 1330 – 1470 ml/hr, 44.27 – 48.93 ml
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Infusion rate accuracy at 125 ml/hr patient safety
Verification of pump accuracy at low-medium flow rate setting
Ranked causes
- 1. Flow rate out of tolerance at 125 ml/hr (unknown)
Deliver fluid at 125 ml/hr and measure output
- Test point
- Pump output collection point
- Expected
- 10.4 ml, 9.88 – 10.92 ml or 118.75 – 131.25 ml/hr
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Infusion rate accuracy at 1400 ml/hr patient safety
Verification of pump accuracy at high flow rate setting
Ranked causes
- 1. Flow rate out of tolerance at 1400 ml/hr (unknown)
Deliver fluid at 1400 ml/hr and measure output
- Test point
- Pump output collection point
- Expected
- 46.6 ml, 44.27 – 48.93 ml or 1330 – 1470 ml/hr
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Infusion rate accuracy at 25 ml/hr patient safety
Verification of pump accuracy at low flow rate setting
Ranked causes
- 1. Flow rate out of tolerance at 25 ml/hr (unknown)
Deliver fluid at 25 ml/hr and measure output
- Test point
- Pump output collection point
- Expected
- 4.2 ml, 3.99 – 4.41 ml or 23.75 – 26.25 ml/hr
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Infusion rate accuracy at 500 ml/hr patient safety
Verification of pump accuracy at medium flow rate setting
Ranked causes
- 1. Flow rate out of tolerance at 500 ml/hr (unknown)
Deliver fluid at 500 ml/hr and measure output
- Test point
- Pump output collection point
- Expected
- 16.6 ml, 15.77 – 17.43 ml or 475 – 525 ml/hr
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Infusion rate accuracy at 999 ml/hr patient safety
Verification of pump accuracy at medium-high flow rate setting
Ranked causes
- 1. Flow rate out of tolerance at 999 ml/hr (unknown)
Deliver fluid at 999 ml/hr and measure output
- Test point
- Pump output collection point
- Expected
- 33.3 ml, 31.63 – 34.97 ml or 949.05 – 1048.95 ml/hr
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Keypad MRI environment
Verification of keypad functionality during normal operating mode
Ranked causes
- 1. Keypad malfunction (unknown)
Verify all keys respond correctly
- Test point
- Keypad
- Expected
- All keys functional
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Motor speed MRI environment
Verification of motor speed performance during service mode
Ranked causes
- 1. Motor speed out of specification (unknown)
Verify motor speed within specification
- Test point
- Motor operation
- Expected
- Motor operates at correct speed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Motor speed verification
Verify pump motor speed is within specification during manual pump run in service mode.
Ranked causes
- 1. Motor speed outside specification (occasional)
Motor speed measurement
- Test point
- Pump motor during MANUAL PUMP RUN
- Expected
- 180 +/- 30 RPM, +/- 30 RPM
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Occlusion detection patient safety
Verification of line occlusion detection capability during normal operating mode
Ranked causes
- 1. Occlusion detection out of specification (unknown)
Verify system detects line occlusion
- Test point
- Occlusion detection sensor
- Expected
- System detects occlusion within specification
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Occlusion limit verification
Verification that pump detects patient line occlusion and triggers alarm within 40 seconds when pressure reaches 10 psi.
Ranked causes
- 1. Occlusion limit setting incorrect or detection timing delayed (occasional)
Verify occlusion limit setting
- Test point
- MENU F3 soft key display
- Expected
- 10 psi
- Tools
- Pump display
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pressure sensor MRI environment
Verification of pressure sensor accuracy during service mode
Ranked causes
- 1. Pressure sensor out of specification (unknown)
Verify pressure sensor reading
- Test point
- Pressure sensor
- Expected
- Pressure reading within specification
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pressure sensor tare and span
Verify and adjust pressure sensor P1 and P2 readings to ensure accurate occlusion pressure detection using tare offset and span calibration procedures in service mode.
Ranked causes
- 1. Pressure sensor calibration drift (common)
Tare offset zero verification
- Test point
- P1 and P2 pressure displays
- Expected
- 0 +/- 0.5 PSI with hand bulb open
- Tools
- Pressure gauge with hand bulb, circular IV test set
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pump accuracy verification at multiple flow rates
Verification of pump accuracy using Infutest 2000 device at 100 mL/hr, 10 mL/hr, 1 mL/hr, 1400 mL/hr, and 999 mL/hr rates to ensure flow output meets ±2-3% tolerance.
Ranked causes
- 1. Flow compensation adjustment needed at specific flow rates (occasional)
Prepare IV set and Infutest 2000 connection
- Test point
- Patient end of IV set to Infutest 2000 channel A or B
- Expected
- IV set primed, all air bubbles removed, stop cock in test position
- Tools
- IV Set 1056, Infutest 2000, water container, syringe
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pump flow rate 125 ml/hr
Verify infusion pump delivers correct volume at 125 ml/hr using Infusion Pump Analyzer or graduated cylinder with distilled water
Ranked causes
- 1. Pump flow rate out of specification (unknown)
Measure output volume or flow rate at 125 ml/hr
- Test point
- IVP1056 infusion set output into graduated cylinder or Infusion Pump Analyzer
- Expected
- 9.88 – 10.92 ml or 118.75 – 131.25 ml/hr, ±5%
- Tools
- Graduated cylinder or Infusion Pump Analyzer
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pump flow rate 1400 ml/hr
Verify infusion pump delivers correct volume at 1400 ml/hr using Infusion Pump Analyzer or graduated cylinder with distilled water
Ranked causes
- 1. Pump flow rate out of specification (unknown)
Measure output volume or flow rate at 1400 ml/hr
- Test point
- IVP1056 infusion set output into graduated cylinder or Infusion Pump Analyzer
- Expected
- 44.27 – 48.93 ml or 1330 – 1470 ml/hr, ±5%
- Tools
- Graduated cylinder or Infusion Pump Analyzer
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pump flow rate 25 ml/hr
Verify infusion pump delivers correct volume at 25 ml/hr using Infusion Pump Analyzer or graduated cylinder with distilled water
Ranked causes
- 1. Pump flow rate out of specification (unknown)
Measure output volume or flow rate at 25 ml/hr
- Test point
- IVP1056 infusion set output into graduated cylinder or Infusion Pump Analyzer
- Expected
- 3.99 – 4.41 ml or 23.75 – 26.25 ml/hr, ±5%
- Tools
- Graduated cylinder or Infusion Pump Analyzer
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Calibration: Pump flow rate 500 ml/hr
- Calibration: Pump flow rate 999 ml/hr
- Calibration: Pump flow rate linearization
- Calibration: Sound, KVO rate, and occlusion limit
- Calibration: SpO2 verification
- Calibration: Time and date
- Channel A door does not open properly
- Channel B flow rate out of specification
- Check Door B alarm verification
- Coin cell battery low voltage
- Coin cell battery voltage low
- Determine current software version
- Display readability inspection
- Door handle/cam lock operation test
- External case and chassis inspection
- Flo-Stop holder inspection and assembly test
- Flow preventer alarm not triggering when pinch clamp activated
- Flow preventer does not trigger alarm on door open
- Flow rate accuracy test at 125 ml/hr
- Flow rate accuracy test at 1400 ml/hr
- Flow rate accuracy test at 500 ml/hr
- Flow rate accuracy test at 999 ml/hr
- Flow stop does not fully retract after door open/close cycle
- Flow stop does not fully retract on door close
- Flow Stop Pull Test failure
- Flow Stop Pull Test verification
- Fluid delivery accuracy test
- Inlet occlusion alarm does not trigger
- Inlet Occlusion detection
- Inlet occlusion not detected when tubing clamped above pump
- IV set seating and flow control test
- Keypad buttons not producing audible tones
- Keypad control functional test
- Keypad display not showing entered values
- Keypad does not respond to input
- Motor hinge pin misaligned
- Motor speed out of specification
- Motor speed outside specification
- Motor speed verification
- No probe alarm
- No Probe alarm clears properly
- No Probe alarm does not sound or display
- Pole clamp mechanism freedom test
- Pressure sensor reading incorrect
- Pressure sensor readings incorrect
- Pressure transducer offset error
- Pressure transducer span error
- Pressure transducer zero drift after adjustment
- Preventive maintenance overdue
- Probe off alarm
- Probe Off alarm does not sound or display
- PT occlusion alarm does not trigger
- PT Occlusion detection
- PT occlusion not detected when tubing clamped below pump
- Pump does not start or infuse
- Pump flow accuracy 1400 ml/hr test
- Pump flow rate outside acceptable range
- Pump motor speed out of specification
- Service mode entry verification
- Software update
- SpO2 display not functioning
61 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
1 | Master/Slave communications failure on Channel A | Processor communication loss between master and slave controllers on Channel A pump | critical |
1000 | Shaft revolution time too short on Channel A | Shaft completed revolution in less than minimum required time | critical |
1001 | Shaft revolution did not occur on Channel A | Expected shaft revolution during programmed operation did not occur | critical |
1002 | Shaft revolution timeout error on Channel A | Shaft revolution exceeded maximum allowed time on Channel A | critical |
1003 | Shaft rotating when it should not on Channel A | Encoder detects motor shaft rotation when motor command is off | critical |
101 | Master/Slave communications failure on Channel B | Processor communication loss between master and slave controllers on Channel B pump | critical |
103 | Motor control processor watchdog timeout on Channel B | Motor control processor failed to service watchdog timer on Channel B | critical |
1100 | Shaft revolution time too short on Channel B | Shaft completed revolution on Channel B in less than minimum required time | critical |
1101 | Shaft revolution did not occur on Channel B | Expected shaft revolution on Channel B during operation did not occur | critical |
1102 | Shaft revolution timeout on Channel B | Shaft revolution on Channel B exceeded maximum allowed time | critical |
1103 | Shaft rotating when it should not on Channel B | Encoder detects motor shaft rotation on Channel B when motor command is off | critical |
129 | Encoder index not found while seeking starting mark on Channel A | Slave processor did not detect encoder index pulse during initial position seek | critical |
130 | Motor rotating too slowly on Channel A | Measured motor speed during operation is below minimum threshold | critical |
132 | Starting index mark found in wrong position on Channel A | Encoder starting index located at position misaligned with minor index marks | critical |
134 | Minor index mark not detected on Channel A | Processor expected to detect minor index mark during operation but did not | critical |
135 | Minor index mark not detected while motor stopped on Channel A | Processor expected to detect minor index mark after motor stop but did not | critical |
136 | Encoder index not detected during normal operation on Channel A | Encoder index pulse missed during continuous pump operation | critical |
2 | Watchdog timeout error | System watchdog timer expired indicating processor hang or freeze | critical |
229 | Encoder index not found while seeking starting mark on Channel B | Slave processor did not detect encoder index pulse during initial position seek on Channel B | critical |
23 | Pump Malfunction - Multiple causes depending on alarm number | Various processor, motor, or communications errors in Channels A or B | critical |
23 | Pump Malfunction | Various motor control, processor, or encoder detection failures | critical |
230 | Motor rotating too slowly on Channel B | Measured motor speed on Channel B below minimum threshold during operation | critical |
232 | Starting index mark in wrong position on Channel B | Encoder starting index located misaligned with minor index marks on Channel B | critical |
234 | Minor index mark not detected on Channel B | Processor expected minor index mark during operation on Channel B but did not detect | critical |
235 | Minor index mark not detected while motor stopped on Channel B | Processor expected minor index mark detection after motor stop on Channel B but did not | critical |
236 | Encoder index not detected during normal operation on Channel B | Encoder index pulse missed during continuous operation on Channel B | critical |
23 (Alarm 1) | Pump Malfunction - Master/Slave communications failure (Channel A) | Master and slave processor communication breakdown on channel A | critical |
23 (Alarm 1000) | Pump Malfunction - Shaft revolution occurred too quickly (Channel A) | Shaft completes revolution in less than required minimum time on channel A | critical |
23 (Alarm 1001) | Pump Malfunction - Shaft revolution did not occur (Channel A) | Shaft did not complete expected revolution on channel A | critical |
23 (Alarm 1002) | Pump Malfunction - Shaft revolution timeout error (Channel A) | Shaft revolution took longer than timeout threshold on channel A | critical |
23 (Alarm 1003) | Pump Malfunction - Shaft should not be turning but it is (Channel A) | Shaft rotation detected when motor should be stopped on channel A | critical |
23 (Alarm 101) | Pump Malfunction - Master/Slave communications failure (Channel B) | Master and slave processor communication breakdown on channel B | critical |
23 (Alarm 103) | Pump Malfunction - Motor control processor watchdog timeout (Channel B) | Motor controller watchdog timer expired on channel B | critical |
23 (Alarm 1100) | Pump Malfunction - Shaft revolution occurred too quickly (Channel B) | Shaft completes revolution in less than required minimum time on channel B | critical |
23 (Alarm 1101) | Pump Malfunction - Shaft revolution did not occur (Channel B) | Shaft did not complete expected revolution on channel B | critical |
23 (Alarm 1102) | Pump Malfunction - Shaft revolution timeout error (Channel B) | Shaft revolution took longer than timeout threshold on channel B | critical |
23 (Alarm 1103) | Pump Malfunction - Shaft should not be turning but it is (Channel B) | Shaft rotation detected when motor should be stopped on channel B | critical |
23 (Alarm 129) | Pump Malfunction - Slave processor did not see encoder index while seeking (Channel A) | Encoder index pulse not detected during starting index mark search on channel A | critical |
23 (Alarm 130) | Pump Malfunction - Motor is rotating too slow (Channel A) | Motor speed falls below minimum threshold on channel A | critical |
23 (Alarm 132) | Pump Malfunction - Starting index mark in wrong position (Channel A) | Starting index mark detected at incorrect position relative to minor marks on channel A | critical |
23 (Alarm 134) | Pump Malfunction - Minor index mark not detected (Channel A) | Expected minor index mark not seen during motor operation on channel A | critical |
23 (Alarm 135) | Pump Malfunction - Minor index mark not seen while motor stopped (Channel A) | Minor index mark expected but not detected when motor is stationary on channel A | critical |
23 (Alarm 136) | Pump Malfunction - Encoder index not detected during normal operation (Channel A) | Encoder index pulse missing during normal pump operation on channel A | critical |
23 (Alarm 2) | Pump Malfunction - Watchdog timeout | Master processor watchdog timer expired | critical |
23 (Alarm 229) | Pump Malfunction - Slave processor did not see encoder index while seeking (Channel B) | Encoder index pulse not detected during starting index mark search on channel B | critical |
23 (Alarm 230) | Pump Malfunction - Motor is rotating too slow (Channel B) | Motor speed falls below minimum threshold on channel B | critical |
23 (Alarm 232) | Pump Malfunction - Starting index mark in wrong position (Channel B) | Starting index mark detected at incorrect position relative to minor marks on channel B | critical |
23 (Alarm 234) | Pump Malfunction - Minor index mark not detected (Channel B) | Expected minor index mark not seen during motor operation on channel B | critical |
23 (Alarm 235) | Pump Malfunction - Minor index mark not seen while motor stopped (Channel B) | Minor index mark expected but not detected when motor is stationary on channel B | critical |
23 (Alarm 236) | Pump Malfunction - Encoder index not detected during normal operation (Channel B) | Encoder index pulse missing during normal pump operation on channel B | critical |
23 (Alarm 3) | Pump Malfunction - Motor control processor watchdog timeout (Channel A) | Motor controller watchdog timer expired on channel A | critical |
23 (Alarm 4) | Pump Malfunction - Nonmaskable interrupt | Nonmaskable interrupt triggered on master processor | critical |
23 (Alarm 5) | Pump Malfunction - ROM checksum error | Firmware ROM checksum validation failure on startup or runtime | critical |
3 | Motor control processor watchdog timeout on Channel A | Motor control processor failed to service watchdog timer within timeout period | critical |
4 | Nonmaskable interrupt error | Nonmaskable interrupt generated by processor | critical |
5 | ROM checksum error | Calculated ROM checksum does not match stored value during boot | critical |
Check Door B | Door opened while pump running | Door opened during active infusion | warning |
Inlet Occlusion | Blockage detected in inlet (supply) line | Tubing above pump is pinched or kinked | error |
Inlet Occlusion | Input line obstruction detected | Tubing clamped, kinked or blocked above pump inlet | warning |
PT Occlusion | Output line obstruction detected | Tubing clamped, kinked or blocked below pump outlet | warning |
PT Occlusion | Blockage detected in outlet (patient) line | Tubing below pump is pinched or kinked | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI