The ServiceForge knowledge base holds 121 documented symptoms for the IRadimed 3860+, with 150 ranked causes, 221 diagnostic checks with test points and expected values, and 61 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Pump does not start infusion or flow rate incorrect patient safety

Pump fails to begin infusion after Start button pressed, or measured flow rate does not match set rate

Ranked causes

  1. 1. IV administration set not installed correctly (common)
  2. 2. Pump rate or VTBI not set correctly (common)
  3. 3. Inlet or patient line occlusion preventing flow (occasional)
  4. 4. Pump mechanism failure or motor malfunction (rare)
First check

Verify IV set installed and door closed properly

Test point
Pump door and IV set seating
Expected
Door closes and latches, IV set seated flush in pump chamber

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display not functioning or degraded image caution

LCD display fails to show information or displays garbled or dim image

Ranked causes

  1. 1. Backlight failure or 5V supply issue (common)
  2. 2. RF shield mesh displacement or damage (occasional)
  3. 3. LCD controller (1330 type) failure (rare)
First check

Verify 5V supply voltage to display connector

Test point
Display power connector pin 1 (positive) and pin 2 (ground)
Expected
5.0 V DC, +/-0.25 V
Tools
Digital multimeter

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flow rate outside acceptable tolerance patient safety

Measured infusion flow rate deviates more than 5% from set rate or specified tolerance

Ranked causes

  1. 1. IV solution viscosity or temperature different from test conditions (occasional)
  2. 2. Pump calibration drift or mechanical wear (occasional)
  3. 3. IV set or tubing restriction or diameter variation (occasional)
First check

Prepare IV set with water at room temperature

Test point
IV solution container and tubing
Expected
Water filled to 2/3 in drip chamber; tubing primed and air-free; water container hung 21 inches above pump

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pumping chamber door misalignment or improper fit caution

Door does not fit flush with case or has friction/resistance opening and closing

Ranked causes

  1. 1. Door hinge pin corrosion or wear (occasional)
  2. 2. Door latch axle screw loose (common)
  3. 3. Brass door axle corroded or obstructed (occasional)
First check

Inspect door flush fit and alignment

Test point
Pumping chamber door AS112 or AS113 top, bottom, and side edges
Expected
Door fits flush with case on all sides

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air in line detection failure patient safety

Pump does not detect air bubble or indicate Air In Line alarm within 50 seconds when running IV set without fluid

Ranked causes

  1. 1. Bubble detector (HE09/HE09A) dirty or blocked (common)
  2. 2. Bubble detector optical sensor malfunction (occasional)
First check

Test bubble detector air detection capability

Test point
Pump running at 10 mL/Hr with IV set containing no fluid
Expected
Air In Line alarm within 50 seconds

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm sound not functioning patient safety

Audible alarm does not sound during alarm conditions or alarm test

Ranked causes

  1. 1. Alarm sound disabled or muted in settings (occasional)
  2. 2. Speaker or audio circuit failure (rare)
First check

Momentarily press Off switch and listen for alarm audio

Test point
Pump speaker and audio circuit
Expected
Audible alarm sound heard; pump remains powered on

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery pack mechanical damage or puncture patient safety

1133 Battery Pack damaged by impact, dropping, overheating, compression, or puncture to internal cells, potentially causing overheating, smoking, or catastrophic battery failure.

Ranked causes

  1. 1. Impact damage, dropping, compression, or mechanical shock (occasional)
  2. 2. Puncture to internal battery cells from sharp objects (occasional)
First check

Inspect battery pack for cracks, dents, or visible physical damage

Test point
1133 Battery Pack case exterior
Expected
No cracks, dents, deformation, or damage visible

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bubble detection alarm not triggered or functioning patient safety

Pump fails to detect air bubbles in line or alarm does not activate when bubble passes through pump

Ranked causes

  1. 1. Bubble detector sensor malfunction or obstructed (rare)
  2. 2. Alarm display or LED circuit failure (rare)
First check

Set pump rate to 500 ml/hr and VTBI to 100 ml

Test point
Display settings
Expected
Display shows Rate: 500 ml and VTBI: 100 ml; Ready to Start flashing

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pump flow accuracy

Factory calibration of 3860+ and 3861 pump flow rate using Sartorius GD603 scale and encoder wheel alignment to achieve accurate delivery profile at various flow rates.

Ranked causes

  1. 1. Encoder wheel misalignment relative to pumping finger position (occasional)
  2. 2. Offset adjustment needed for step size correction (occasional)
First check

Verify raw curve taper using SartoConnect data

Test point
Scale measurement during calibration
Expected
Ideal curve per Figure A with appropriate taper profile
Tools
Sartorius GD603 scale, SartoConnect software, Microsoft Excel

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check Door alarm on flow detection failure patient safety

Alarm raised when flow stop switch does not respond within 0.75 seconds after door opens

Ranked causes

  1. 1. Free-flow detector switch (IVPAB016) not responding properly (occasional)
  2. 2. Door switch not opening properly (occasional)
First check

Measure flow stop switch opening time after door open

Test point
Flow Stop Switch IVPAB016 signal
Expected
Switch opens within 0.75 second of door switch detecting open
Tools
Timer or oscilloscope

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display not clear or unreadable

Boot display appears but text is garbled, dim, or pixels not functioning

Ranked causes

  1. 1. Display window or overlay scratched or cracked (occasional)
  2. 2. Internal display module failure or LCD connection loose (rare)
First check

Inspect display window covering and overlay for physical damage

Test point
Front display area
Expected
No scratches, cracks, or damage to window or overlay

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Door latch or clamp mechanism malfunction caution

Door handle/cam lock does not operate smoothly, does not latch securely, or does not release properly

Ranked causes

  1. 1. Latch cam worn or misaligned (occasional)
  2. 2. Door clamp latch paw worn or damaged (common)
First check

Verify door handle/cam lock ease of operation

Test point
Door latch handle/cam AS104
Expected
Smooth, easy operation without binding or resistance

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Door open alarm during pumping patient safety

Alarm triggered and pump shuts down when door is detected open during infusion operation

Ranked causes

  1. 1. Door not fully latched in closed position (common)
  2. 2. Door switch or optic sensor malfunction (occasional)
First check

Verify door closure and latch engagement

Test point
Door clamp mechanism
Expected
Hard rubber pin fully contacts switch or optic sensor when latched

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fluid delivery rate inaccurate patient safety

Pump delivers less than 100 mL in measured time period when set to 100 mL/Hr; flow rate outside specification

Ranked causes

  1. 1. Motor speed out of specification (common)
  2. 2. Pumping gasket worn or leaking (occasional)
First check

Measure fluid delivery at 100 mL/Hr over calibrated time period

Test point
Graduated cylinder or Infutest system; pump outlet
Expected
Delivered volume matches 100 mL/Hr setting within manufacturer tolerance
Tools
100 mL and/or 25 mL graduated cylinder, Infutest or calibrated fluid measurement system

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inlet occlusion alarm not triggered or functioning patient safety

Pump fails to detect inlet line obstruction or alarm does not activate when inlet tubing pinched

Ranked causes

  1. 1. Inlet occlusion detection sensor malfunction (rare)
  2. 2. Alarm display or LED circuit failure (rare)
First check

Start pump with water flowing

Test point
Display and LED indicators
Expected
Pulsing > symbol in upper display; Green flashing LED in door; water flowing

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

IV set not properly seated in Flo-Stop holder patient safety

IV set does not remain seated in Flo-Stop holder when door opened; 'Check Door' alarm occurs inappropriately

Ranked causes

  1. 1. Flo-Stop holder plastic damaged by drug degradation (occasional)
  2. 2. Flo-Stop holder recess clogged or obstructed (common)
First check

Inspect Flo-Stop holder for damage and obstruction

Test point
IVP1047 or AS120 Flo-Stop holder recess
Expected
No damage, obstructions, or fluid spillage; clean condition

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Patient occlusion alarm not triggered or functioning patient safety

Pump fails to detect outlet line obstruction or alarm does not activate when distal tubing blocked

Ranked causes

  1. 1. Patient occlusion detection sensor malfunction (rare)
  2. 2. Alarm display or LED circuit failure (rare)
First check

Clear any inlet occlusion, start pump with water flowing

Test point
Display and LED indicators
Expected
Pulsing > symbol in upper display; Green flashing LED in door; water flowing through distal tubing

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power indicator LED not illuminated

LED below power on/off switches does not light when External DC Power/Charger connected or battery installed

Ranked causes

  1. 1. External DC Power/Charger not connected or connectors loose (common)
  2. 2. Battery pack not installed (common)
First check

Verify External DC Power/Charger connected and line power active

Test point
DC input connector on pump
Expected
Connectors screwed in tightly, line power connected

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Prevent battery pack damage patient safety

Precautions to avoid internal damage to 1133 Battery Pack that could disable protective circuitry and cause catastrophic failure

Ranked causes

  1. 1. Physical damage from impact, dropping, overheating, compression, or mechanical abuse (common)
  2. 2. Disassembly or tool damage to battery pack (occasional)
First check

Inspect battery pack for damage

Test point
1133 Battery Pack exterior surfaces
Expected
No cracks, dents, punctures, corrosion, or signs of impact or thermal damage

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pump does not initialize or start

Pump fails to power on or display initial boot sequence after pressing power button

Ranked causes

  1. 1. Battery not charged sufficiently (common)
  2. 2. Power button not pressed or held correctly (occasional)
First check

Verify battery has been charged for minimum 6 hours

Test point
Battery pack 1133
Expected
Battery fully charged

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pumping chamber door and latch mechanical inspection

Verify pumping chamber door (AS112AOP or AS113AOP) fits flush with case, latches securely, and operates freely without friction.

Ranked causes

  1. 1. Door misalignment, binding, or latch failure (occasional)
  2. 2. Worn or damaged upper door clamp latch paw (IVP1038) (occasional)
First check

Pumping chamber door flush fit inspection

Test point
Door edges at top, bottom, and sides
Expected
Door fits flush with case; no gaps

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pumping mechanism integrity test

Test pumping chamber door springs (HM10) and seal integrity by occluding IV set and verifying proper pinch-off and occlusion pressure maintenance.

Ranked causes

  1. 1. Worn or damaged pumping gasket/seal (IVP1070) (common)
  2. 2. Weak or broken extension springs (HM10) (occasional)
First check

Pumping mechanism occlusion pressure test

Test point
IV set tube occlusion by pump
Expected
Pump reaches 10 PSI maximum occlusion pressure without backflow or free flow leakage
Tools
iRadimed Series 1000 IV set, pressure gauge

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pumping mechanism weak or occlusion failure patient safety

IV set not pinched off or pump cannot build occlusion pressure when patient tubing clamp is closed

Ranked causes

  1. 1. Extension spring (HM10) worn or weak (common)
  2. 2. Pumping fingers worn or misaligned (occasional)
First check

Test pumping mechanism integrity and spring force

Test point
IV set loaded in pump during Pump mode operation
Expected
Pump achieves 10 PSI occlusion pressure with patient tube clamp closed

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Swollen battery stuck in pump or remote patient safety

1133 Battery Pack expands excessively and becomes jammed or difficult to remove from 3860+ pump or 3865 remote display chassis

Ranked causes

  1. 1. Excessive battery expansion causing jamming (occasional)
  2. 2. Battery plastic case beginning to crack during removal attempts (rare)
First check

Assess degree of battery swelling and jamming

Test point
Battery fit in chassis
Expected
Battery moves slightly when pried with gentle screwdriver pressure

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air bubble detected in IV line patient safety

Unit measures air bubble of 100µL or larger in infusion line and triggers alarm when acoustic impedance increases due to air-liquid interface

Ranked causes

  1. 1. Air bolus in IV tubing (common)
First check

Verify acoustic sensor detects air bubble presence

Test point
Acoustic impedance detection element
Expected
Voltage signal reduced when air present compared to fluid-filled state

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery cell voltage imbalance

Battery cells not within 0.2 V of each other or any cell outside 3.8 to 4.25 V range

Ranked causes

  1. 1. Battery cell imbalance or failure (occasional)
First check

Verify all battery cells within range and balanced

Test point
Service mode F2 Battery Status display; all 4 cell voltages
Expected
3.8 to 4.25 V per cell, all within 0.2 V of each other, 3.8-4.25, delta 0.2

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery discharge current abnormal

Battery discharge current significantly different from approximately -240 mA when charger disconnected

Ranked causes

  1. 1. Battery or discharge circuit fault (rare)
First check

Verify battery discharge current with charger disconnected

Test point
Service mode F2 Battery Status display; Current field
Expected
Approximately -240 mA, nominal -240

Battery pack protective circuit failure or damage patient safety

1133 Battery Pack integral safety circuit damaged or failed, unable to prevent catastrophic battery failure.

Ranked causes

  1. 1. Physical damage to battery pack compromising protective circuitry (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery pack swollen or expanded patient safety

1133 Battery Pack case expands or swells due to internal gas release from aging lithium-polymer cells, potentially causing case failure, cell failure, or safety circuit failure.

Ranked causes

  1. 1. Aged lithium-polymer cells releasing internal gas (common)
First check

Inspect battery pack for visible swelling or expansion of case

Test point
1133 Battery Pack exterior
Expected
No visible swelling or case expansion

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery pack swollen or expanded patient safety

1133 Battery Pack case visibly expands or swells due to internal gas release from aging lithium-polymer cells

Ranked causes

  1. 1. Battery pack aging and internal gas release (common)
First check

Inspect battery case for visible swelling or expansion

Test point
1133 Battery Pack exterior
Expected
Smooth, flat battery case with no bulging

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery shows swelling patient safety

Battery pack exhibits visible swelling or deformation indicating internal damage

Ranked causes

  1. 1. Internal battery degradation or short circuit (rare)
First check

Visual inspection for battery swelling

Test point
Battery pack physical appearance
Expected
No visible swelling or deformation

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery state of charge low

Battery state of charge below 50% before testing

Ranked causes

  1. 1. Battery not pre-charged before test (common)
First check

Verify battery state of charge

Test point
Service mode F2 Battery Status display
Expected
50 to 100 %, 50-100

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery terminal voltage out of range

Battery terminal voltage outside 15.0 to 16.9 V range during service mode test

Ranked causes

  1. 1. Battery faulty or not fully charged (occasional)
First check

Verify battery terminal voltage

Test point
Service mode F2 Battery Status display with charger cable connected
Expected
15 to 16.9 V, 15-16.9

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bubble detection not functioning

Bubble detection indicator (upper case B or lower case b) not displaying correctly on screen

Ranked causes

  1. 1. Bubble detection sensor malfunction or calibration issue (unknown)
First check

Verify bubble detection indicator with no IV Set installed

Test point
Screen bottom center, bubble indicator
Expected
Upper case B displayed

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Bubble detection not working correctly patient safety

Bubble detection indicator at bottom center of screen does not show uppercase B with no IV set or lowercase b with fluid-filled set installed

Ranked causes

  1. 1. Bubble detection sensor malfunction (unknown)
First check

Verify bubble detection with no IV set installed

Test point
Bottom center of screen indicator
Expected
uppercase B

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Bubble detector functional test patient safety

Verify bubble detector (HE09A) ability to detect air in line within specified time.

Ranked causes

  1. 1. Bubble detector failure or malfunction (occasional)
First check

Air in line detection time

Test point
Bubble detector response with empty IV set running at 10 mL/Hr
Expected
Air In Line alarm within 50 seconds
Tools
IV set, pump operation

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Backup battery replacement

Replace CR2032 coin cell memory/clock backup battery to ensure proper timekeeping and history record retention.

Ranked causes

  1. 1. CR2032 backup battery depleted or aged (common)
First check

Backup battery replacement interval

Test point
Battery age and condition record
Expected
Replace every two years or if 'Lost History' message appears

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Battery MRI environment

Verification of backup battery functionality during service mode

Ranked causes

  1. 1. Battery out of specification (unknown)
First check

Verify battery charge and functionality

Test point
Battery
Expected
Battery meets specification

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Battery status verification patient safety

Verify battery state of charge, terminal voltage, individual cell voltages, current draw, and cycle count meet specifications.

Ranked causes

  1. 1. Battery voltage or charge state out of specification (occasional)
First check

Battery state of charge with charger connected

Test point
F2 key display: State of Charge (%)
Expected
50 to 100%

4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Bubble detection MRI environment

Verification of bubble detection functionality during service mode

Ranked causes

  1. 1. Bubble detector out of specification (unknown)
First check

Verify bubble detection sensor response

Test point
Bubble detection sensor
Expected
Sensor detects bubbles correctly

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Flow preventer switch MRI environment

Verification of flow preventer switch operation during normal operating mode

Ranked causes

  1. 1. Flow preventer switch malfunction (unknown)
First check

Verify flow preventer switch operation

Test point
Flow preventer switch
Expected
Switch functions correctly

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Flow stop pull test MRI environment

Verification of flow stop mechanism under pull load during normal operating mode

Ranked causes

  1. 1. Flow stop mechanism failure (unknown)
First check

Apply pull force and verify flow stops

Test point
Flow stop mechanism
Expected
Flow stops under pull test load

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Flow stop pull test patient safety

Verify black pinch clamp full retraction and flow preventer housing retention

Ranked causes

  1. 1. IVP1068 Door Clamp Clip worn (occasional)
First check

Load IVP1056 IV set and close door

Test point
Door and flow stop carrier
Expected
Door closes fully

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Infusion pump flow rate patient safety

Verify pump delivers correct flow rates at multiple programmed settings with volume and rate tolerance bands

Ranked causes

  1. 1. Pump flow rate out of specification (unknown)
First check

Measure delivered volume at 1400 ml/hr

Test point
Pump output
Expected
44.27 – 48.93 ml or 1330 – 1470 ml/hr, 44.27 – 48.93 ml

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Infusion rate accuracy at 125 ml/hr patient safety

Verification of pump accuracy at low-medium flow rate setting

Ranked causes

  1. 1. Flow rate out of tolerance at 125 ml/hr (unknown)
First check

Deliver fluid at 125 ml/hr and measure output

Test point
Pump output collection point
Expected
10.4 ml, 9.88 – 10.92 ml or 118.75 – 131.25 ml/hr

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Infusion rate accuracy at 1400 ml/hr patient safety

Verification of pump accuracy at high flow rate setting

Ranked causes

  1. 1. Flow rate out of tolerance at 1400 ml/hr (unknown)
First check

Deliver fluid at 1400 ml/hr and measure output

Test point
Pump output collection point
Expected
46.6 ml, 44.27 – 48.93 ml or 1330 – 1470 ml/hr

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Infusion rate accuracy at 25 ml/hr patient safety

Verification of pump accuracy at low flow rate setting

Ranked causes

  1. 1. Flow rate out of tolerance at 25 ml/hr (unknown)
First check

Deliver fluid at 25 ml/hr and measure output

Test point
Pump output collection point
Expected
4.2 ml, 3.99 – 4.41 ml or 23.75 – 26.25 ml/hr

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Infusion rate accuracy at 500 ml/hr patient safety

Verification of pump accuracy at medium flow rate setting

Ranked causes

  1. 1. Flow rate out of tolerance at 500 ml/hr (unknown)
First check

Deliver fluid at 500 ml/hr and measure output

Test point
Pump output collection point
Expected
16.6 ml, 15.77 – 17.43 ml or 475 – 525 ml/hr

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Infusion rate accuracy at 999 ml/hr patient safety

Verification of pump accuracy at medium-high flow rate setting

Ranked causes

  1. 1. Flow rate out of tolerance at 999 ml/hr (unknown)
First check

Deliver fluid at 999 ml/hr and measure output

Test point
Pump output collection point
Expected
33.3 ml, 31.63 – 34.97 ml or 949.05 – 1048.95 ml/hr

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Keypad MRI environment

Verification of keypad functionality during normal operating mode

Ranked causes

  1. 1. Keypad malfunction (unknown)
First check

Verify all keys respond correctly

Test point
Keypad
Expected
All keys functional

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Motor speed MRI environment

Verification of motor speed performance during service mode

Ranked causes

  1. 1. Motor speed out of specification (unknown)
First check

Verify motor speed within specification

Test point
Motor operation
Expected
Motor operates at correct speed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Motor speed verification

Verify pump motor speed is within specification during manual pump run in service mode.

Ranked causes

  1. 1. Motor speed outside specification (occasional)
First check

Motor speed measurement

Test point
Pump motor during MANUAL PUMP RUN
Expected
180 +/- 30 RPM, +/- 30 RPM

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Occlusion detection patient safety

Verification of line occlusion detection capability during normal operating mode

Ranked causes

  1. 1. Occlusion detection out of specification (unknown)
First check

Verify system detects line occlusion

Test point
Occlusion detection sensor
Expected
System detects occlusion within specification

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Occlusion limit verification

Verification that pump detects patient line occlusion and triggers alarm within 40 seconds when pressure reaches 10 psi.

Ranked causes

  1. 1. Occlusion limit setting incorrect or detection timing delayed (occasional)
First check

Verify occlusion limit setting

Test point
MENU F3 soft key display
Expected
10 psi
Tools
Pump display

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pressure sensor MRI environment

Verification of pressure sensor accuracy during service mode

Ranked causes

  1. 1. Pressure sensor out of specification (unknown)
First check

Verify pressure sensor reading

Test point
Pressure sensor
Expected
Pressure reading within specification

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pressure sensor tare and span

Verify and adjust pressure sensor P1 and P2 readings to ensure accurate occlusion pressure detection using tare offset and span calibration procedures in service mode.

Ranked causes

  1. 1. Pressure sensor calibration drift (common)
First check

Tare offset zero verification

Test point
P1 and P2 pressure displays
Expected
0 +/- 0.5 PSI with hand bulb open
Tools
Pressure gauge with hand bulb, circular IV test set

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pump accuracy verification at multiple flow rates

Verification of pump accuracy using Infutest 2000 device at 100 mL/hr, 10 mL/hr, 1 mL/hr, 1400 mL/hr, and 999 mL/hr rates to ensure flow output meets ±2-3% tolerance.

Ranked causes

  1. 1. Flow compensation adjustment needed at specific flow rates (occasional)
First check

Prepare IV set and Infutest 2000 connection

Test point
Patient end of IV set to Infutest 2000 channel A or B
Expected
IV set primed, all air bubbles removed, stop cock in test position
Tools
IV Set 1056, Infutest 2000, water container, syringe

5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pump flow rate 125 ml/hr

Verify infusion pump delivers correct volume at 125 ml/hr using Infusion Pump Analyzer or graduated cylinder with distilled water

Ranked causes

  1. 1. Pump flow rate out of specification (unknown)
First check

Measure output volume or flow rate at 125 ml/hr

Test point
IVP1056 infusion set output into graduated cylinder or Infusion Pump Analyzer
Expected
9.88 – 10.92 ml or 118.75 – 131.25 ml/hr, ±5%
Tools
Graduated cylinder or Infusion Pump Analyzer

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pump flow rate 1400 ml/hr

Verify infusion pump delivers correct volume at 1400 ml/hr using Infusion Pump Analyzer or graduated cylinder with distilled water

Ranked causes

  1. 1. Pump flow rate out of specification (unknown)
First check

Measure output volume or flow rate at 1400 ml/hr

Test point
IVP1056 infusion set output into graduated cylinder or Infusion Pump Analyzer
Expected
44.27 – 48.93 ml or 1330 – 1470 ml/hr, ±5%
Tools
Graduated cylinder or Infusion Pump Analyzer

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pump flow rate 25 ml/hr

Verify infusion pump delivers correct volume at 25 ml/hr using Infusion Pump Analyzer or graduated cylinder with distilled water

Ranked causes

  1. 1. Pump flow rate out of specification (unknown)
First check

Measure output volume or flow rate at 25 ml/hr

Test point
IVP1056 infusion set output into graduated cylinder or Infusion Pump Analyzer
Expected
3.99 – 4.41 ml or 23.75 – 26.25 ml/hr, ±5%
Tools
Graduated cylinder or Infusion Pump Analyzer

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Calibration: Pump flow rate 500 ml/hr
  • Calibration: Pump flow rate 999 ml/hr
  • Calibration: Pump flow rate linearization
  • Calibration: Sound, KVO rate, and occlusion limit
  • Calibration: SpO2 verification
  • Calibration: Time and date
  • Channel A door does not open properly
  • Channel B flow rate out of specification
  • Check Door B alarm verification
  • Coin cell battery low voltage
  • Coin cell battery voltage low
  • Determine current software version
  • Display readability inspection
  • Door handle/cam lock operation test
  • External case and chassis inspection
  • Flo-Stop holder inspection and assembly test
  • Flow preventer alarm not triggering when pinch clamp activated
  • Flow preventer does not trigger alarm on door open
  • Flow rate accuracy test at 125 ml/hr
  • Flow rate accuracy test at 1400 ml/hr
  • Flow rate accuracy test at 500 ml/hr
  • Flow rate accuracy test at 999 ml/hr
  • Flow stop does not fully retract after door open/close cycle
  • Flow stop does not fully retract on door close
  • Flow Stop Pull Test failure
  • Flow Stop Pull Test verification
  • Fluid delivery accuracy test
  • Inlet occlusion alarm does not trigger
  • Inlet Occlusion detection
  • Inlet occlusion not detected when tubing clamped above pump
  • IV set seating and flow control test
  • Keypad buttons not producing audible tones
  • Keypad control functional test
  • Keypad display not showing entered values
  • Keypad does not respond to input
  • Motor hinge pin misaligned
  • Motor speed out of specification
  • Motor speed outside specification
  • Motor speed verification
  • No probe alarm
  • No Probe alarm clears properly
  • No Probe alarm does not sound or display
  • Pole clamp mechanism freedom test
  • Pressure sensor reading incorrect
  • Pressure sensor readings incorrect
  • Pressure transducer offset error
  • Pressure transducer span error
  • Pressure transducer zero drift after adjustment
  • Preventive maintenance overdue
  • Probe off alarm
  • Probe Off alarm does not sound or display
  • PT occlusion alarm does not trigger
  • PT Occlusion detection
  • PT occlusion not detected when tubing clamped below pump
  • Pump does not start or infuse
  • Pump flow accuracy 1400 ml/hr test
  • Pump flow rate outside acceptable range
  • Pump motor speed out of specification
  • Service mode entry verification
  • Software update
  • SpO2 display not functioning

61 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
1Master/Slave communications failure on Channel AProcessor communication loss between master and slave controllers on Channel A pumpcritical
1000Shaft revolution time too short on Channel AShaft completed revolution in less than minimum required timecritical
1001Shaft revolution did not occur on Channel AExpected shaft revolution during programmed operation did not occurcritical
1002Shaft revolution timeout error on Channel AShaft revolution exceeded maximum allowed time on Channel Acritical
1003Shaft rotating when it should not on Channel AEncoder detects motor shaft rotation when motor command is offcritical
101Master/Slave communications failure on Channel BProcessor communication loss between master and slave controllers on Channel B pumpcritical
103Motor control processor watchdog timeout on Channel BMotor control processor failed to service watchdog timer on Channel Bcritical
1100Shaft revolution time too short on Channel BShaft completed revolution on Channel B in less than minimum required timecritical
1101Shaft revolution did not occur on Channel BExpected shaft revolution on Channel B during operation did not occurcritical
1102Shaft revolution timeout on Channel BShaft revolution on Channel B exceeded maximum allowed timecritical
1103Shaft rotating when it should not on Channel BEncoder detects motor shaft rotation on Channel B when motor command is offcritical
129Encoder index not found while seeking starting mark on Channel ASlave processor did not detect encoder index pulse during initial position seekcritical
130Motor rotating too slowly on Channel AMeasured motor speed during operation is below minimum thresholdcritical
132Starting index mark found in wrong position on Channel AEncoder starting index located at position misaligned with minor index markscritical
134Minor index mark not detected on Channel AProcessor expected to detect minor index mark during operation but did notcritical
135Minor index mark not detected while motor stopped on Channel AProcessor expected to detect minor index mark after motor stop but did notcritical
136Encoder index not detected during normal operation on Channel AEncoder index pulse missed during continuous pump operationcritical
2Watchdog timeout errorSystem watchdog timer expired indicating processor hang or freezecritical
229Encoder index not found while seeking starting mark on Channel BSlave processor did not detect encoder index pulse during initial position seek on Channel Bcritical
23Pump Malfunction - Multiple causes depending on alarm numberVarious processor, motor, or communications errors in Channels A or Bcritical
23Pump MalfunctionVarious motor control, processor, or encoder detection failurescritical
230Motor rotating too slowly on Channel BMeasured motor speed on Channel B below minimum threshold during operationcritical
232Starting index mark in wrong position on Channel BEncoder starting index located misaligned with minor index marks on Channel Bcritical
234Minor index mark not detected on Channel BProcessor expected minor index mark during operation on Channel B but did not detectcritical
235Minor index mark not detected while motor stopped on Channel BProcessor expected minor index mark detection after motor stop on Channel B but did notcritical
236Encoder index not detected during normal operation on Channel BEncoder index pulse missed during continuous operation on Channel Bcritical
23 (Alarm 1)Pump Malfunction - Master/Slave communications failure (Channel A)Master and slave processor communication breakdown on channel Acritical
23 (Alarm 1000)Pump Malfunction - Shaft revolution occurred too quickly (Channel A)Shaft completes revolution in less than required minimum time on channel Acritical
23 (Alarm 1001)Pump Malfunction - Shaft revolution did not occur (Channel A)Shaft did not complete expected revolution on channel Acritical
23 (Alarm 1002)Pump Malfunction - Shaft revolution timeout error (Channel A)Shaft revolution took longer than timeout threshold on channel Acritical
23 (Alarm 1003)Pump Malfunction - Shaft should not be turning but it is (Channel A)Shaft rotation detected when motor should be stopped on channel Acritical
23 (Alarm 101)Pump Malfunction - Master/Slave communications failure (Channel B)Master and slave processor communication breakdown on channel Bcritical
23 (Alarm 103)Pump Malfunction - Motor control processor watchdog timeout (Channel B)Motor controller watchdog timer expired on channel Bcritical
23 (Alarm 1100)Pump Malfunction - Shaft revolution occurred too quickly (Channel B)Shaft completes revolution in less than required minimum time on channel Bcritical
23 (Alarm 1101)Pump Malfunction - Shaft revolution did not occur (Channel B)Shaft did not complete expected revolution on channel Bcritical
23 (Alarm 1102)Pump Malfunction - Shaft revolution timeout error (Channel B)Shaft revolution took longer than timeout threshold on channel Bcritical
23 (Alarm 1103)Pump Malfunction - Shaft should not be turning but it is (Channel B)Shaft rotation detected when motor should be stopped on channel Bcritical
23 (Alarm 129)Pump Malfunction - Slave processor did not see encoder index while seeking (Channel A)Encoder index pulse not detected during starting index mark search on channel Acritical
23 (Alarm 130)Pump Malfunction - Motor is rotating too slow (Channel A)Motor speed falls below minimum threshold on channel Acritical
23 (Alarm 132)Pump Malfunction - Starting index mark in wrong position (Channel A)Starting index mark detected at incorrect position relative to minor marks on channel Acritical
23 (Alarm 134)Pump Malfunction - Minor index mark not detected (Channel A)Expected minor index mark not seen during motor operation on channel Acritical
23 (Alarm 135)Pump Malfunction - Minor index mark not seen while motor stopped (Channel A)Minor index mark expected but not detected when motor is stationary on channel Acritical
23 (Alarm 136)Pump Malfunction - Encoder index not detected during normal operation (Channel A)Encoder index pulse missing during normal pump operation on channel Acritical
23 (Alarm 2)Pump Malfunction - Watchdog timeoutMaster processor watchdog timer expiredcritical
23 (Alarm 229)Pump Malfunction - Slave processor did not see encoder index while seeking (Channel B)Encoder index pulse not detected during starting index mark search on channel Bcritical
23 (Alarm 230)Pump Malfunction - Motor is rotating too slow (Channel B)Motor speed falls below minimum threshold on channel Bcritical
23 (Alarm 232)Pump Malfunction - Starting index mark in wrong position (Channel B)Starting index mark detected at incorrect position relative to minor marks on channel Bcritical
23 (Alarm 234)Pump Malfunction - Minor index mark not detected (Channel B)Expected minor index mark not seen during motor operation on channel Bcritical
23 (Alarm 235)Pump Malfunction - Minor index mark not seen while motor stopped (Channel B)Minor index mark expected but not detected when motor is stationary on channel Bcritical
23 (Alarm 236)Pump Malfunction - Encoder index not detected during normal operation (Channel B)Encoder index pulse missing during normal pump operation on channel Bcritical
23 (Alarm 3)Pump Malfunction - Motor control processor watchdog timeout (Channel A)Motor controller watchdog timer expired on channel Acritical
23 (Alarm 4)Pump Malfunction - Nonmaskable interruptNonmaskable interrupt triggered on master processorcritical
23 (Alarm 5)Pump Malfunction - ROM checksum errorFirmware ROM checksum validation failure on startup or runtimecritical
3Motor control processor watchdog timeout on Channel AMotor control processor failed to service watchdog timer within timeout periodcritical
4Nonmaskable interrupt errorNonmaskable interrupt generated by processorcritical
5ROM checksum errorCalculated ROM checksum does not match stored value during bootcritical
Check Door BDoor opened while pump runningDoor opened during active infusionwarning
Inlet OcclusionBlockage detected in inlet (supply) lineTubing above pump is pinched or kinkederror
Inlet OcclusionInput line obstruction detectedTubing clamped, kinked or blocked above pump inletwarning
PT OcclusionOutput line obstruction detectedTubing clamped, kinked or blocked below pump outletwarning
PT OcclusionBlockage detected in outlet (patient) lineTubing below pump is pinched or kinkederror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI