Invivo 3160: troubleshooting and error codes
The ServiceForge knowledge base holds 147 documented symptoms for the Invivo 3160, with 230 ranked causes, 347 diagnostic checks with test points and expected values, and 23 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Agents parameter failure or not working (with EtCO2)
CO2 measurement not functioning; no air flow from gas input port or occlusion detected
Ranked causes
- 1. Gas input tubing disconnected from water trap (common)
- 2. Water trap full or occluded (common)
- 3. Internal gas tubing disconnected or loose (occasional)
- 4. Agent Bench not operating properly (occasional)
- 5. WPU Electronics Assembly failure (occasional)
- 6. DCU assembly failure (occasional)
Check water trap for proper inlet air flow
- Test point
- GAS Input port on water trap
- Expected
- A mild air vacuum is detected at the GAS Input port
4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG 5 Volt digital pulse not present or incorrect amplitude high voltage
Digital scope does not display positive 5V pulse train on ECG signal channel during bench testing
Ranked causes
- 1. Patient simulator not outputting ECG signal at 60 BPM (common)
- 2. TF0240 test fixture not connected to PA558A WPU gating port (common)
- 3. Scope probe not connected to correct signal pin (occasional)
- 4. Scope settings incorrect (occasional)
- 5. WPU gating source not set to ECG (occasional)
Verify patient simulator ECG output at 60 BPM
- Test point
- medSim 300B display
- Expected
- 60 BPM flashing; ECG signal being output
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP parameter failure or not working patient safety
NIBP measurement not functioning; air not inflating or pressure readings not obtained
Ranked causes
- 1. NIBP cuff accessory disconnected or loose (common)
- 2. WPU Electronics Assembly failure (occasional)
- 3. NIBP Primary transducer calibration error (occasional)
- 4. NIBP Secondary transducer calibration error (occasional)
- 5. DCU assembly failure (occasional)
Press NIBP START/STOP key to initiate manual NIBP reading
- Test point
- DCU NIBP START/STOP key control panel and Inflate port
- Expected
- A moderate stream of air exits the Inflate port. A flashing red NOT INFLATING error message is displayed in the DCU's NIBP icon
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SPO2 20 millivolt analog signal not present or incorrect amplitude
Digital scope does not display 20 mV analog signal on SPO2 channel during bench testing
Ranked causes
- 1. Bapco pulse oximetry simulator not powered on or configured (common)
- 2. SPO2 sensor not connected or WSPO2 module not powered on (common)
- 3. Scope vertical scale not set to 20 mV/Div (occasional)
- 4. AS131 adapter not installed between scope and probe (occasional)
- 5. Scope probe connected to wrong signal pin (occasional)
Power on Bapco and configure for pulse oximetry
- Test point
- Bapco controls and display
- Expected
- Bapco powered; PULSE OXIMETRY selected; SPO2 > 90%; heart rate 60-120 BPM
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AS201 Power Adapter won't power up on facility line power high voltage
Power adapter does not illuminate DC OUTPUT indicator light when connected to facility AC power
Ranked causes
- 1. Facility power absent or out of tolerance (common)
- 2. AS201 DC OUTPUT indicator not green (internal fault) (common)
- 3. AS201 AC fuse(s) defective (occasional)
- 4. AS201 Power Adapter internal failure (occasional)
Verify facility AC power availability and within tolerance
- Test point
- AC wall outlet
- Expected
- 90–264 VAC @ 47–63 Hz continuous
- Tools
- DMM
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG parameter failure or not working patient safety
ECG transmitter module not communicating or transmitting data; status indicator displayed in red box with X
Ranked causes
- 1. WECG transmitter module battery voltage low (common)
- 2. DCU assembly failure (occasional)
- 3. WPU Electronics Assembly failure (occasional)
- 4. WECG transmitter module hardware failure (rare)
Check WECG transmitter module battery voltage
- Test point
- WECG transmitter module battery terminals
- Expected
- > +3.3 Vdc
- Tools
- DMM
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
IBP 200 millivolt analog signal not present or incorrect amplitude high voltage
Digital scope does not display 200 mV analog signal on IBP channel during bench testing
Ranked causes
- 1. IBP cable not connected from simulator BP1 output to P1 input of Interface Unit (common)
- 2. Simulator BP settings not configured correctly (common)
- 3. Scope vertical scale not appropriate for 200 mV signal (occasional)
- 4. Scope probe connected to wrong signal pin (occasional)
Verify P1 turned on and displayed on WPU display
- Test point
- WPU display
- Expected
- P1 indicator on and visible
6 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Respiration 1 Volt analog signal not present or incorrect amplitude
Digital scope does not display 1V analog signal on respiration channel during bench testing
Ranked causes
- 1. 3160 SPO2 and RESP parameters not enabled (common)
- 2. Respiration sensor bellows not connected to WSPO2 module (common)
- 3. Respiration signal not being generated at correct rate (common)
- 4. Scope probe connected to wrong signal pin (occasional)
Verify 3160 SPO2 and RESP parameters are enabled
- Test point
- 3160 system parameter settings
- Expected
- SPO2 parameter = ON; RESP parameter = ON
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SPO2 parameter failure or not working patient safety
SPO2 transmitter module not communicating or transmitting data; status indicator displayed in red box with X
Ranked causes
- 1. WSPO2 transmitter module battery voltage low (common)
- 2. DCU assembly failure (occasional)
- 3. WPU Electronics Assembly failure (occasional)
- 4. WSPO2 transmitter module hardware failure (rare)
Check WSPO2 transmitter module battery voltage
- Test point
- WSPO2 transmitter module battery terminals
- Expected
- > +3.3 Vdc
- Tools
- DMM
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG 1 millivolt analog signal not present or incorrect amplitude high voltage
Digital scope does not display 1 mV analog signal on ECG channel during bench testing
Ranked causes
- 1. Scope vertical scale not set to lowest possible scale (common)
- 2. AS131 adapter not installed between scope and probe (common)
- 3. Scope probe connected to wrong signal pin (occasional)
Verify scope vertical scale set to 2 mV/Div
- Test point
- Scope vertical scale control
- Expected
- Vertical scale 2 mV/Div (lowest possible)
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG negative 5 Volt digital pulse not present or incorrect amplitude high voltage
Digital scope does not display negative -5V pulse train on ECG signal channel during bench testing
Ranked causes
- 1. Scope probe connected to wrong signal pin (common)
- 2. Scope baseline position not adjusted for negative signal display (occasional)
- 3. Gating source not set to ECG on DCU display (occasional)
Verify scope probe connected to pin F on TF0240 with pin B as ground
- Test point
- TF0240 pin F (signal) and pin B (common)
- Expected
- Probe connected to pin F; ground connected to pin B
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
EtCO2 parameter failure or not working patient safety
Exhaled CO2 measurement not displayed or functional
Ranked causes
- 1. Water trap occluded or full (common)
- 2. Internal gas tubing disconnected or improperly seated (common)
- 3. WPU Electronics Assembly defective (occasional)
Disconnect patient tubing from GAS Input port
- Test point
- Water trap GAS Input port
- Expected
- Tubing disconnected
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ETCO2 reading out of expected range with calibration gas patient safety
Monitor ETCO2 reading does not fall within calculated tolerance range when 10% CO2 check gas is applied
Ranked causes
- 1. Barometric pressure not accounted for in expected value (common)
- 2. Check gas cylinder depleted or regulator malfunction (occasional)
- 3. Sensor calibration drift or sensor failure (occasional)
Apply 9091A check gas to sample line input with air vented delivery system
- Test point
- Sample line input
- Expected
- Monitor detects ETCO2 and displays reading
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
IBP parameter failure or not working patient safety
Invasive blood pressure measurement not displayed or functional
Ranked causes
- 1. WPU IBP output voltage out of specification (common)
- 2. IBP fiber optic cable defective or disconnected (occasional)
- 3. PCU assembly or AS226 IBP assembly defective (occasional)
Measure WPU IBP output voltage
- Test point
- PCU/DCU connector Pin 3 to Pin 4
- Expected
- +12 +0.5 Vdc, +/-0.5
- Tools
- DMM
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor internal cleaning and inspection required high voltage
Internal dust, debris, loose hardware, unseated cards, loose cables, or damaged EMI gaskets detected; affects cooling or electrical safety.
Ranked causes
- 1. Dust and debris accumulation inside monitor enclosure (common)
- 2. Loose or improperly seated hardware or components (occasional)
- 3. Damaged EMI gaskets affecting shielding (occasional)
Turn monitor OFF and disconnect AS201 DC Input power before opening
- Test point
- Power switch and AS201 connector
- Expected
- Monitor powered off and disconnected from DC power source
9 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 measurements not working patient safety
Oxygen sensor output not displayed or functional on monitor
Ranked causes
- 1. Waste gas exhaust not detected from O2 sensor block (common)
- 2. Internal connections to O2 sensor block faulty (common)
- 3. WPU Electronics Assembly defective (occasional)
Inspect for waste gas exhaust from O2 sensor block
- Test point
- WPU rear panel O2 sensor block
- Expected
- Air flow is detected (minimal)
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
RESP parameter failure or not working patient safety
Respiration measurement not displayed or functional
Ranked causes
- 1. WSPO2 transmitter module battery low or depleted (common)
- 2. DCU assembly defective (occasional)
- 3. WSPO2 transmitter module communication lost or defective (occasional)
Remove WSPO2 transmitter battery
- Test point
- WSPO2 transmitter module battery compartment
- Expected
- Battery removed safely
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Respiration rate not displayed on WPU patient safety
Respiration option enabled, bellows tubing connected to WSPO2 module and bellows positioned on chest/stomach, but WPU does not display consistent respiration rate
Ranked causes
- 1. Respiration bellows not properly connected or positioned (common)
- 2. Respiration option not enabled in menu (occasional)
- 3. WSPO2 module or respiration sensor malfunction (occasional)
Enable Respiration option in menu
- Test point
- SETUPS / PARAMETER SELECTION menu
- Expected
- RESPIRATION turned on
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SPO2 2 Volt analog signal not present or incorrect amplitude
Digital scope does not display 2V analog signal on SPO2 2V channel during bench testing
Ranked causes
- 1. AS131 adapter not removed from scope (common)
- 2. Scope vertical scale not set to 1 V/Div (occasional)
- 3. Scope probe connected to wrong signal pin (occasional)
Remove AS131 adapter and connect probe directly to scope channel 1
- Test point
- Oscilloscope channel 1 input
- Expected
- AS131 adapter removed; probe directly connected to channel 1
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Adult NIBP cuff inflation or deflation not working patient safety
Adult cuff fails to inflate to expected pressure or does not deflate after measurement
Ranked causes
- 1. Cuff not properly seated on arm or tubing obstruction (common)
- 2. Dead-man cuff detection failure (cuff not on limb) (occasional)
Measure adult cuff inflation pressure during normal measurement cycle
- Test point
- NIBP cuff pressure, NIBP display menu
- Expected
- Inflates to approximately 170 mmHg
- Tools
- NIBP system display
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
DCU not keeping accurate time caution
DCU system clock drifts or displays incorrect time and date
Ranked causes
- 1. DCU system clock out of calibration or battery backed time circuit failure (common)
- 2. DCU Display Assembly time circuit failure if verification fails (occasional)
Perform time set verification on DCU
- Test point
- DCU system clock and calendar
- Expected
- Time is recalibrated to the correct time and date
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
DCU sound system not working caution
DCU does not produce audible click tones or alarm sounds during operation
Ranked causes
- 1. CLICK TONE disabled in SOUND ADJUST SETUPS menu (common)
- 2. DCU Display Assembly or audio circuitry failure (occasional)
Verify CLICK TONE enabled in SOUND ADJUST SETUPS Menu
- Test point
- DCU SOUND ADJUST SETUPS Menu CLICK TONE setting
- Expected
- CLICK TONE is ON
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
DCU system crash alarm occurs patient safety
DCU generates a crash alarm and fails to boot properly to normal operation screen
Ranked causes
- 1. DCU firmware or software fault; cold start may recover (common)
- 2. DCU Display Assembly hardware failure if cold start fails (occasional)
Perform DCU cold start
- Test point
- DCU power and control sequence
- Expected
- Monitor boots up correctly and Normal Screen is displayed
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
DCU won't power up on AS201 +18 Vdc output power caution
DCU (Display/Control Unit) fails to power on when connected to AS201 +18 Vdc output
Ranked causes
- 1. DCU DC fuse defective or open (common)
- 2. DCU power board or assembly failure (occasional)
Test DCU DC fuse
- Test point
- DCU DC fuse
- Expected
- Fuse continuity good
- Tools
- DMM
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dual DCU communication verification
Verify WPU communication with multiple DCUs and parameter synchronization between DCU units
Ranked causes
- 1. Loss of dual communication between WPU and multiple DCUs (unknown)
- 2. Parameter display not synchronizing between DCUs (unknown)
Press NIBP START/STOP on first DCU
- Test point
- WPU NIBP pump
- Expected
- WPU NIBP pump starts
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG 1 Volt analog signal not present or incorrect amplitude high voltage
Digital scope does not display 1V analog signal on ECG channel during bench testing
Ranked causes
- 1. Scope probe connected to wrong signal pin (common)
- 2. Scope horizontal scale not set to 250 ms per division (occasional)
Verify scope probe connected to pin D with pin B as ground
- Test point
- TF0240 pin D (signal) and pin B (common)
- Expected
- Probe connected to pin D; ground connected to pin B
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG cable not connected to WECG module caution
ECG cable not properly connected to WECG module or leads not connected to patient simulator
Ranked causes
- 1. Loose or incorrect ECG cable connection (common)
- 2. Incorrect ECG lead connections to patient simulator (common)
Verify ECG cable connected to WECG module
- Test point
- WECG module connector
- Expected
- ECG cable fully seated in connector
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG heart rate not displaying correctly patient safety
ECG numerics do not display correct heart rate count or shows outside expected 60 ±1 BPM range
Ranked causes
- 1. Patient simulator output not set to 60 BPM (common)
- 2. ECG rate counter malfunction (occasional)
Verify patient simulator heart rate is 60 BPM
- Test point
- Patient simulator output
- Expected
- 60 BPM
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
ECG trace not displaying QRS waveform patient safety
ECG display does not show moving QRS pulses or shows jittering waveform
Ranked causes
- 1. ECG lead connection issue or cable malfunction (common)
- 2. ECG hardware filter or processor malfunction (occasional)
Set TRACE A LEAD to II in ECG menu
- Test point
- ECG menu configuration
- Expected
- TRACE A LEAD set to II
6 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
ETCO2 flow measurement out of specification patient safety
Flowmeter reading does not match expected flow rates at sample line or water trap ports during verification
Ranked causes
- 1. Sample line obstruction or water trap blockage (common)
- 2. Waste gas port blocked (occasional)
Verify ETCO2 WARMING UP message cleared and O2 cell installed
- Test point
- Monitor display
- Expected
- ETCO2 WARMING UP message no longer displayed, exhaust port open
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ETCO2 reading out of range patient safety
Check gas applied to CO2 sample line but displayed ETCO2 reading not between 67 and 85 mmHg at sea level after 2 minutes of flow
Ranked causes
- 1. Check gas not properly connected or flowing (common)
- 2. CO2 monitor or ETCO2 sensor malfunction (occasional)
Verify check gas is connected to BLUE CO2 sample line input
- Test point
- BLUE CO2 sample line input
- Expected
- 9091A check gas (10% CO2, BAL, N2) connected via air vented delivery system
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ETCO2 waveform absent or inconsistent with respiration patient safety
No ETCO2 waveform displayed or waveform does not correlate with operator breathing when nasal cannula is used
Ranked causes
- 1. Sample line disconnected or damaged (common)
- 2. Sample line blockage or water trap obstruction (occasional)
Connect nasal cannula to sample line and breathe normally
- Test point
- Monitor ETCO2 waveform display
- Expected
- ETCO2 waveform displayed, consistent with respiration rate 10-16 BPM
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ETCO2 waveform not displayed during nasal breathing patient safety
Nasal cannula connected to CO2 sample line and operator breathing normally but ETCO2 waveform not displayed or inconsistent with respiration rate
Ranked causes
- 1. Nasal cannula not properly connected or positioned (common)
- 2. ETCO2 monitor not functioning properly (occasional)
Connect nasal cannula and position on operator's nose
- Test point
- BLUE CO2 sample line connection; cannula position
- Expected
- Cannula connected and positioned for normal nasal breathing
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
F.O. TEMP parameter failure or not working patient safety
Fiber optic temperature measurement not displayed or functional
Ranked causes
- 1. DCU Display Assembly defective (occasional)
- 2. WPU Electronics Assembly with defective HE113 module (common)
Check Error Code Display status on DCU AB213 Power Board
- Test point
- DCU AB213 Power Board LEDs
- Expected
- Left decimal On (lit) solid green, right decimal flashing green, numeric segments Off
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gating signal not observed at MR console MRI environment
MR Tech cannot confirm gating pulse at MR console after ECG cable connection and ECG test signal selection
Ranked causes
- 1. Gating cable not connected or defective (common)
- 2. ECG test signal not selected correctly (occasional)
Verify gating cable connections at WPU and magnet interface
- Test point
- Gating cable WPU connector and magnet gating interface
- Expected
- Cable fully seated and secure at both ends
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
LEAD FAIL message displayed during ECG operation patient safety
LEAD FAIL error message appears on ECG display during testing or normal operation
Ranked causes
- 1. ECG lead connection loose or disconnected (common)
- 2. ECG cable malfunction (occasional)
Verify all ECG lead connections to patient simulator terminals
- Test point
- Patient simulator terminals
- Expected
- All leads firmly connected per color code: GREEN=RL, WHITE=RA, BLACK=LA, RED=LL
Neonatal NIBP cuff inflation or deflation not working patient safety
Neonatal cuff fails to inflate to expected pressure or does not deflate after measurement
Ranked causes
- 1. Cuff not properly seated on finger or tubing obstruction (common)
- 2. Dead-man cuff detection failure (cuff not on limb) (occasional)
Measure neonatal cuff inflation pressure during normal measurement cycle
- Test point
- NIBP cuff pressure, NIBP display menu
- Expected
- Inflates to approximately 130 mmHg
- Tools
- NIBP system display
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 sensor reading out of specification patient safety
O2 reading does not display 21%±1 at ambient air after initialization calibration
Ranked causes
- 1. O2 sensor degradation or failure (occasional)
- 2. Calibration procedure not completed correctly (occasional)
Access O2 initialization calibration menu
- Test point
- SETUPS > SERVICE (BIO-MED) > GAS CAL > O2 INIT CAL
- Expected
- Menu accessible, 2 MINUTE WARNING message appears
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient accessories cleaning and inspection patient safety
Reusable patient accessories show visible dirt, debris, cracks, or contamination; require cleaning and disinfection per protocol.
Ranked causes
- 1. Normal accumulation of dirt and debris on reusable accessories (common)
- 2. Visible damage to accessory (cracks, holes, tears, cuts) (occasional)
Remove reusable accessory from monitor
- Test point
- Accessory connection points on monitor
- Expected
- Accessory cleanly disconnected and removed
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Perkin Elmer agent bench flow measurement out of specification patient safety
Exhaust flow rate does not meet 230 ml/min ±40 ml/min specification or waste gas port blockage test fails
Ranked causes
- 1. Sample line obstruction or flowmeter miscalibration (common)
- 2. Waste gas port blockage (occasional)
Verify agents parameter is ON and water trap cleared
- Test point
- AGENTS parameter, CLEAR WATER TRAP, CLEAR SAMPLE LINE
- Expected
- ETCO2 WARMING UP message no longer displayed; exhaust port open
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
QRS detection failing in NEO patient mode at low amplitude patient safety
QRS count does not display or shows no flashing heart and tone when 0.5 mV signal input in NEO patient mode
Ranked causes
- 1. NEO mode sensitivity threshold not properly configured (occasional)
- 2. Signal level below NEO mode detection threshold (common)
Enter SETUPS menu and set PATIENT to NEO
- Test point
- SETUPS menu PATIENT setting
- Expected
- PATIENT set to NEO
5 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
QRS HR rate tone not producing sound patient safety
No QRS heart rate tone output or double tones detected
Ranked causes
- 1. Audio circuit malfunction or speaker failure (occasional)
- 2. ECG signal processing error producing double tones (rare)
Verify QRS HR rate tone is present and audible
- Test point
- Speaker output
- Expected
- Audible QRS rate tone
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Speaker crackling or distortion at maximum alarm volume caution
Speaker produces crackling noise or excessive distortion when ALARM VOLUME set to level 10
Ranked causes
- 1. Speaker damaged or audio amplifier distorting (rare)
- 2. Audio amplifier circuit fault (occasional)
Set ALARM VOLUME to level 10 and listen for distortion
- Test point
- SOUNDADJUST menu, ALARM VOLUME setting
- Expected
- Loud tone with no crackling or excessive distortion
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature reading not displayed patient safety
Temperature option enabled and temperature probe connected to WPU port but ambient temperature reading not displayed
Ranked causes
- 1. Temperature probe not connected or probe malfunction (common)
- 2. Temperature option not enabled in menu (occasional)
Enable Temperature option in menu
- Test point
- SETUPS / PARAMETER SELECTION menu
- Expected
- TEMPERATURE turned on
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Time and date not accurate or not retained
DCU time and date display is incorrect, loses time when powered off, or gains/loses time during operation
Ranked causes
- 1. Real-time clock battery depleted or failing (occasional)
- 2. RTC crystal or oscillator circuit fault (rare)
Enter SETUPS menu and verify time and date displayed
- Test point
- DCU SETUPS menu, SET TIME option
- Expected
- Current time and date accurately displayed
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
WECG module not communicating patient safety
WECG module with 9065 battery does not communicate with 3160 system
Ranked causes
- 1. WECG battery not inserted or depleted (common)
- 2. WECG module communication link failure (occasional)
Verify 9065 battery inserted in WECG module
- Test point
- WECG module battery compartment
- Expected
- 9065 battery properly seated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
WECG module not communicating with 3160 system patient safety
WECG module with 9065 battery inserted does not establish communication link to 3160 system
Ranked causes
- 1. 9065 battery not properly inserted or depleted (common)
- 2. WECG wireless communication module failure (occasional)
Verify 9065 battery properly seated in WECG module
- Test point
- WECG battery compartment
- Expected
- Battery fully inserted with positive terminal contacting spring
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
WECG module not receiving battery power or communicating with system
WECG module does not operate when 9065 battery inserted, no communication with 3160 system
Ranked causes
- 1. 9065 battery not inserted or depleted (common)
- 2. WECG module or communication circuit failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
WPU program update interrupted or halted
WPU loses power or USB device removed during reprogramming; unit may no longer function properly or communicate with DCU.
Ranked causes
- 1. Power loss during WPU reprogramming (occasional)
- 2. USB device removed during WPU reprogramming (occasional)
Verify WPU establishes DCU communication within 1 minute of power-on
- Test point
- DCU display
- Expected
- DCU shows communication with WPU established
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
WPU won't power up on AS201 +18 Vdc output power caution
WPU (Wireless Patient Unit) fails to power on when connected to AS201 +18 Vdc output
Ranked causes
- 1. WPU DC fuse defective or open (common)
- 2. WPU Electronics Assembly failure (occasional)
Test WPU DC fuse
- Test point
- WPU DC fuse
- Expected
- Fuse continuity good
- Tools
- DMM
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AB213 Power Board Error Code Display and Recovery high voltage
Fatal error code displayed on AB213 power board LED displays (U8 and U9) requiring diagnosis and corrective action
Ranked causes
- 1. Various power board malfunctions indicated by specific error codes in Table G-1 (unknown)
Power off monitor and disconnect power adapter
- Test point
- PCU and DCU power switches
- Expected
- Both switches in OFF position, AS201 adapter unplugged, batteries removed
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AGENTS or ETCO2 parameter not displayed after power up
AGENTS or ETCO2 parameter missing from DCU display following system startup
Ranked causes
- 1. AGENTS or ETCO2 not enabled in SETUPS menu (common)
Verify WARMING UP message displayed at top center of DCU display
- Test point
- DCU display top center
- Expected
- WARMING UP message shown
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm volume not producing audible tone caution
No audible tone or distorted sound output from alarm at various volume levels
Ranked causes
- 1. Speaker failure or audio circuit malfunction (unknown)
Press SETUP function key and navigate to SOUNDADJUST > ALARM VOLUME
- Test point
- DCU display menu
- Expected
- SOUND ADJUST menu opens
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Alarm volume not producing sound or has distortion
Speaker produces no tone at low level, crackling, or excessive distortion at high level
Ranked causes
- 1. Speaker failure or audio circuit defect (unknown)
Verify audible tone at ALARM VOLUME level 1
- Test point
- Speaker output
- Expected
- Audible tone produced
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ambient temperature not displayed
WPU does not show ambient temperature reading after temperature probe is connected
Ranked causes
- 1. Temperature probe not connected or not recognized (common)
Enable TEMPERATURE in menu
- Test point
- WPU menu
- Expected
- TEMPERATURE turned on
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Anesthetic agents verification not performed
Verification step not applicable if gas option not installed; requires checking software revision to determine which agent bench is installed
Ranked causes
- 1. No anesthetic agents option installed (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible tone not produced at alarm volume level 1
ALARM VOLUME setting at level 1 does not produce an audible tone from speaker
Ranked causes
- 1. Speaker not functioning or muted (unknown)
Set ALARM VOLUME and verify audible tone
- Test point
- SOUNDADJUST menu, ALARM VOLUME setting
- Expected
- Audible tone produced at level 1
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Anesthetic agents span adjustment patient safety
Agent (Desflurane, Sevoflurane, Isoflurane, Halothane, Enflurane) or CO2/N2O reading does not match expected value; requires manual menu adjustment for each gas calibrated.
Ranked causes
- 1. Agent bench requires span calibration after repair or installation (common)
Vent residual gases from water trap
- Test point
- GAS (EtCO2) port
- Expected
- Minimum 90 seconds with port open
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CO2 Occlusion Test
Verify pump motor responds to CO2 port occlusion by increasing speed and displaying CO2 OCCLUSION message, then decreases speed and shows READJUSTING CO2 ZERO when occlusion is cleared.
Ranked causes
- 1. CO2 port occlusion detection malfunction (unknown)
CO2 OCCLUSION message appears at top of display screen
- Test point
- Display screen
- Expected
- CO2 OCCLUSION
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: DCU verification
Verify proper operation of AS240 DCU Assembly by visual inspection of front panel indicators and controls.
Ranked causes
- 1. AS240 DCU Assembly not functioning properly (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Calibration: EtCO2 span adjustment
- Calibration: Fiber Optic Temperature Verification
- Calibration: Gas Monitor Test
- Calibration: Monitor Calibration Verification
- Calibration: Power Adapter verification
- Calibration: Pump Air Flow Test
- Calibration: Pump Speed Test
- Calibration: Span Calibration Test
- Click tone volume not producing audible tone
- CO2 occlusion not detected
- Communication verification between DCU, WPU, and remote modules
- DCU control panel not working
- DCU powers up but has no display
- DCU program update fails to complete
- DCU won't power up on +14.4 Vdc battery power
- ECG cable or leads not connected correctly to Patient Simulator
- ECG lead I verification failing
- ECG lead III verification failing
- ETCO2 hardware failure indicated
- ETCO2 HW FAIL message displayed
- ETCO2 not detected
- ETCO2 or AGENTS parameter not active during test
- ETCO2 parameter not displayed after power up
- ETCO2 reading out of tolerance
- ETCO2 reading outside expected range
- ETCO2 warming up message persists
- ETCO2 waveform not displayed
- ETCO2 zero calibration flow out of specification
- Keypad function not responding
- Keypad not functioning
- Keypad not responding or non-functional
- Low-flow ETCO2 module replacement
- Monitor external cleaning required
- NIBP calibration accuracy out of tolerance
- NIBP calibration error detection not working
- NIBP offset out of range
- NIBP over-pressure threshold not triggered (Adult)
- NIBP over-pressure threshold not triggered (Neonatal)
- NIBP pneumatic system leaks
- NIBP pump does not shut off on sense port timeout
- No ETCO2 waveform detected during respiration
- O2 calibration reading outside tolerance
- O2 sensor expired
- Patient mode not switching to ADULT mode correctly
- Patient simulator QRS output not configured correctly
- Power icon verification under different supply conditions
- Pressure transducer simulator reading out of tolerance
- Pulse volume not functioning
- Pulse volume not producing audible tone
- Pulse volume or click tone volume not correct
- QRS count not zeroing at low amplitude in ADULT mode
- QRS symbol (flashing heart) not displaying
- Readjusting CO2 zero
- Respiration rate not measured
- RF noise in MRI image during test
- RF noise test failure at bench
- Sample line flow rate out of tolerance
- Sample line occlusion not detected
- Speaker distortion or crackling at high volume
- STATUS code E-1: Blocked channel does not detect heat
- STATUS code E-10: EEPROM version different than 3
- STATUS code E-11: EEPROM version not supported
- STATUS code E-12: EEPROM chip test failed
- STATUS code E-13: Agent bench not present
- STATUS code E-2: Sensor voltage out of range
- STATUS code E-3: Invalid thermal gains
- STATUS code E-4: EEPROM failure
- STATUS code E-5: Cannot distinguish between DES and SEV
- STATUS code E-6: EEPROM agent type not supported
- STATUS code E-7: EEPROM checksum test failed
- STATUS code E-8: EEPROM root solution failed
- STATUS code E-9: EEPROM interference matrix inversion failed
- Time and date not displayed or incorrect
- Time and date not displaying correctly
- Time and date not set or incorrect
- Time and date reset after power cycle
- Time drift or clock loss over time
- Waste gas port flow restriction
- Water trap left port flow out of tolerance
- Water trap right port flow out of tolerance
- WECG battery not communicating with system
- WPU Electronics Assembly setup before adjustment
- WPU NAV mode entry required before USB update
- WPU program elements not correctly loaded after update
- WPU won't communicate with DCU
- WPU won't power up on +14.4 Vdc battery power
- Zero port flow out of tolerance
87 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
01 | +3.3 Vdc System Power Start-Up Error | System power voltage error detected during startup initialization | critical |
02 | +3.3 Vdc System Power Run Error | System power voltage error detected during operation | critical |
03 | +5 Vdc System Power Start-Up Error | 5V rail voltage error detected during startup | critical |
04 | +5 Vdc System Power Run Error | 5V rail voltage error detected during operation | critical |
05 | +12 Vdc System Power Start-Up Error | 12V rail voltage error detected during startup | critical |
06 | +12 Vdc System Power Run Error | 12V rail voltage error detected during operation | critical |
07 | Recorder Power Start-Up Error | Recorder power supply error during startup | critical |
08 | Recorder Power Run Error | Recorder power supply error during operation | critical |
09 | Power Board 5VL to Processor 3.3VL conversion error | Voltage conversion circuit error detected | critical |
11 | Motherboard Keep Alive Circuit Run Error | Keep alive circuit failed during operation | critical |
12 | Motherboard Keep Alive Circuit Start Error | Keep alive circuit failed during startup | critical |
13 | Motherboard Communications Error | Communication failure between motherboard components | error |
14 | Power Board Memory Error | Memory test failure detected on power board | critical |
15 | Power Board Clock Oscillator Error | Clock reference signal failure detected | critical |
16 | 9064 Battery Low Voltage Error | Backup battery voltage below minimum threshold | warning |
17 | System Overcurrent Error | System current draw exceeded safe limits | critical |
18 | Power Adapter Over/Under-Voltage Error | AS201 Power Adapter input or output voltage out of range | critical |
19 | Power Board Switched Power Error | Switched power supply circuit error | critical |
31 | 9064 Battery No Start Bit Error | Battery communication protocol error at startup | error |
32 | 9064 Battery No Acknowledge Bit Error | Battery communication protocol error during operation | error |
34 | 9064 Battery No Stop Bit Error | Battery communication protocol error during shutdown | error |
ETCO2: LOW FLOW | ETCO2 sample line or water trap is disconnected or not attached | Sample line or water trap separation during operation | warning |
FAIL | RF Noise Test did not pass acceptable limits | One or more frequency spectrums measured RF noise levels exceeding acceptable thresholds during automated bench testing | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI