Integra MPM-1: troubleshooting and error codes
The ServiceForge knowledge base holds 24 documented symptoms for the Integra MPM-1, with 60 ranked causes, 9 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
CPP reading displays as dashes patient safety
Cerebral perfusion pressure reading shows '---' or is unavailable
Ranked causes
- 1. Bedside monitor connection issue (common)
- 2. Defective adapter cable (occasional)
- 3. Defective PMIO cable (occasional)
- 4. Faultybedside monitor (occasional)
- 5. Faulty PMIO/Analogue Board connection (rare)
- 6. Analogue Board malfunction (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ICP/Systolic/Diastolic reads out of range or as dashes patient safety
Pressure readings display as '---' or outside operating range (-50 to 300 mmHg)
Ranked causes
- 1. Catheter connection issue (common)
- 2. Imperfect Pre-Amp cable connection (common)
- 3. Damaged catheter (occasional)
- 4. Damaged Pre-Amp cable (occasional)
- 5. Faulty Analogue/Pre-Amp flex connection (rare)
- 6. Faulty Digital or Analogue Board (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ICT temperature reading out of range or as dashes patient safety
Intracranial temperature reading shows '---' or is outside valid range (15-45°C)
Ranked causes
- 1. Catheter temperature connector issue (common)
- 2. Faulty catheter (common)
- 3. Wrong catheter in use (occasional)
- 4. Damaged Pre-Amp cable (occasional)
- 5. Faulty Temperature/Analogue Board connection (rare)
- 6. Analogue Board malfunction (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Catheter connection error patient safety
Display reads 'CHECK CATHETER CONNECTION'
Ranked causes
- 1. Defective external connection (common)
- 2. Damaged catheter (common)
- 3. Faulty Pre-Amp cable (common)
- 4. Faulty Analogue/Digital Board connection (occasional)
- 5. Faulty Digital or Analogue Board (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not operate on AC power, but system works on battery high voltage
MPM-1 fails to power on when connected to AC mains, but operates normally when on battery.
Ranked causes
- 1. AC power cord not connected (common)
- 2. Faulty AC power cord (common)
- 3. Blown fuse (common)
- 4. Power module malfunctioning (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Discrepancy in pressure reading between MPM-1 and host monitor patient safety
Pressure values displayed on host monitor differ from MPM-1 display; readings do not agree.
Ranked causes
- 1. Host monitor out of calibration (common)
- 2. Incorrect host monitor adapter cable (occasional)
- 3. Faulty PMIO cable (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Discrepancy in temperature reading between MPM-1 and host monitor patient safety
Temperature values displayed on host monitor differ from MPM-1 display; readings do not match.
Ranked causes
- 1. Host monitor out of calibration (common)
- 2. Incorrect host monitor adapter cable (occasional)
- 3. Faulty PMIO cable (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
MPM-1 displays numerical ICP but no CPP display patient safety
MPM-1 shows intracranial pressure (ICP) value but cerebral perfusion pressure (CPP) is missing from host monitor.
Ranked causes
- 1. Loose cable connection (common)
- 2. Faulty cable (occasional)
- 3. Faulty connection between PMIO flex cable and Analogue Board (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No Isolated Analogue Output
Isolated analogue output signal is absent or not functioning.
Ranked causes
- 1. Faulty connection between Analogue Board and Isolated Analogue Output cable (common)
- 2. Loose analogue output cable (common)
- 3. Faulty cable (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No pressure waveform or numerical display on host monitor but MPM-1 displays mean pressure patient safety
Host monitor screen does not show pressure waveform or numbers; MPM-1 pressure display is normal.
Ranked causes
- 1. Faulty connection between PMIO flex cable and Analogue Board (occasional)
- 2. Faulty PMIO or monitor adapter cable (common)
- 3. Host monitor malfunction (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No temperature waveform or numerical display on host monitor but MPM-1 displays mean temperature patient safety
Host monitor does not show temperature waveform or numbers; MPM-1 temperature is normal.
Ranked causes
- 1. Imperfect connection between PMIO flex cable and Analogue Board (occasional)
- 2. Faulty PMIO or monitor adapter cable (common)
- 3. Host monitor malfunction (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery does not charge
Battery fails to charge when unit is connected to AC power.
Ranked causes
- 1. Faulty battery (common)
- 2. Battery terminal wires reversed (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not operate on battery, but system operates on AC power
MPM-1 fails to power on when using battery, but operates normally when connected to AC mains.
Ranked causes
- 1. Faulty battery (common)
- 2. Battery terminal wires reversed (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm disabled when both ICP and CPP set to OFF patient safety
ALARM DISABLED INDICATOR remains ON; both alarms disabled as warning when parameters set to OFF
Ranked causes
- 1. Both CPP ALARM and ICP ALARM levels set to OFF (common)
Verify CPP ALARM level setting
- Test point
- CPP ALARM parameter
- Expected
- Not set to OFF
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm silenced for limited duration patient safety
Press SILENCE ALARM button; ALARM DISABLED INDICATOR flashes; parameter in alarm mode silences for 3 minutes and flashes on screen; silence expires after 3 minutes
Ranked causes
- 1. Alarm silencing feature active (common)
Verify ALARM DISABLED INDICATOR state
- Test point
- ALARM DISABLED INDICATOR
- Expected
- Flashing when alarm silenced
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarms silenced for 3 minutes patient safety
Parameter in alarm mode is silenced and flashes on screen after SILENCE ALARM button is pressed; ALARM DISABLED INDICATOR flashes
Ranked causes
- 1. Operator activated SILENCE ALARM button (common)
Verify ALARM DISABLED INDICATOR status
- Test point
- ALARM DISABLED INDICATOR
- Expected
- Flashing when alarm is silenced
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Both alarms disabled patient safety
ALARM DISABLED INDICATOR is ON and both ICP and CPP alarms are set to OFF; no alarms will sound
Ranked causes
- 1. Both CPP ALARM and ICP ALARM set to OFF (common)
Verify alarm settings
- Test point
- CPP ALARM and ICP ALARM level controls
- Expected
- Both set to OFF
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: External bedside monitor zero balance patient safety
Calibrate or balance external bedside monitor using CAL STEP button to advance through mmHg values and synchronize zero point
Ranked causes
- 1. External bedside monitor requires zero calibration (common)
Press CAL STEP button repeatedly
- Test point
- MPM-1 Display and CAL STEP button
- Expected
- 0 mmHg displayed on MPM-1 (cycles through 0, 20, 40, 100, 200)
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: ICP pressure zero adjustment patient safety
Before catheter insertion, ensure MPM-1 displays ICP pressure of 0 mmHg and temperature at reasonable value (room temperature)
Ranked causes
- 1. Transducer zero offset requires adjustment (common)
Verify ICP pressure display
- Test point
- MPM-1 Display
- Expected
- 0 mmHg
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dark display while unit operating
LCD screen appears dark despite MPM-1 powered on and functioning
Ranked causes
- 1. Faulty LCD or connections (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dominant color on LCD display
LCD screen displays with one color dominant or abnormal color representation
Ranked causes
- 1. Faulty LCD or connections (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
LCD displays white vertical lines
LCD screen shows white vertical lines across display
Ranked causes
- 1. Faulty LCD or connections (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No RS-232 communication
RS-232 serial communication between MPM-1 Digital Board and external device is not functioning.
Ranked causes
- 1. Faulty connection between Digital Board and RS-232 cable (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ON and OFF button does not work properly
Power button fails to respond or operates intermittently.
Ranked causes
- 1. Faulty membrane switch and/or power supply (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.