Impact Instrumentation Ventilator accessory: troubleshooting and error codes

The ServiceForge knowledge base holds 331 documented symptoms for the Impact Instrumentation Ventilator accessory, with 349 ranked causes, 702 diagnostic checks with test points and expected values, and 63 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Calibration: Mode control and CAL mode operation

Testing MODE control switch positions (OFF, A/C, SIMV, CPAP, CAL) and verifying unit response, LCD display, and calibration completion.

Ranked causes

  1. 1. Mode switch not transitioning properly or unit not responding to mode changes (unknown)
  2. 2. Calibration abort or incomplete calibration cycle (unknown)
  3. 3. Autocal Valve not functioning or pressure baseline incorrect (unknown)
First check

Verify unit powers off

Test point
Unit status indicator
Expected
Unit turns off and is non-operational

7 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LED does not blink during disconnect alarm

Alarm LED fails to blink when disconnect alarm is generated

Ranked causes

  1. 1. Disconnect alarm not generated (occasional)
  2. 2. Alarm LED circuit failure (occasional)
First check

Generate disconnect alarm condition

Test point
Patient circuit connection
Expected
Disconnect alarm occurs on next breath

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LED does not respond to Disconnect alarm

Alarm LED fails to blink when Disconnect alarm is generated

Ranked causes

  1. 1. Disconnect alarm not generated (common)
  2. 2. Alarm LED circuit malfunction (occasional)
First check

Generate Disconnect alarm condition

Test point
Patient circuit connection
Expected
Alarm LED blinking

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LED does not turn solid when mute button pressed

Alarm LED fails to change from blinking to solid state after Mute/Cancel button press during alarm condition

Ranked causes

  1. 1. Mute/Cancel button not responding (occasional)
  2. 2. Alarm LED control circuit malfunction (occasional)
First check

Press Mute/Cancel button during alarm

Test point
Mute/Cancel push-button
Expected
Button depresses and registers

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Charge LED does not illuminate

Charge indicator LED fails to light when external power connected and battery disconnected

Ranked causes

  1. 1. External power supply not properly connected or insufficient power (common)
  2. 2. Charge LED requires warm-up time (occasional)
First check

Verify external power supply connected

Test point
External power input
Expected
Power supply connected and active

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Comp alarm occurs patient safety

Compressor alarm activates when switching between external air and compressor operation, particularly when external air button is pressed after one external air breath.

Ranked causes

  1. 1. Compressor failure to start (common)
  2. 2. External gas pressure below minimum (35 PSI) (occasional)
First check

Verify LCD indicator status after pressing External Air button

Test point
LCD display
Expected
OFF

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Comp alarm occurs during external air to compressor switching patient safety

Compressor alarm is artificially induced when switching between external air and compressor operation; external air valves supply gas during alarm

Ranked causes

  1. 1. Compressor fails to start or continues operation abnormally during source switching (unknown)
  2. 2. External gas supply below minimum threshold (Failure Code 2) (occasional)
First check

Press External Air button after one external air breath

Test point
LCD indicator
Expected
OFF

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Comp alarm with external air and compressor switching

Comp alarm occurs during switching between external air and compressor operation, along with possible Vt and Disconnect alarms. External air valves supply gas source during Comp alarm.

Ranked causes

  1. 1. Compressor failure or malfunction (unknown)
  2. 2. External gas pressure below 35 PSI (common)
First check

Press External Air push-button control after one external air breath

Test point
External Air push-button
Expected
LCD indicator reads OFF

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Disconnect alarm does not occur on next breath patient safety

Disconnect alarm not triggered when patient circuit is removed from test lung and next breath is fired

Ranked causes

  1. 1. Patient circuit not properly disconnected (common)
  2. 2. Pressure sensor not detecting lack of airway pressure rise (rare)
First check

Remove patient circuit from test lung

Test point
patient circuit connection
Expected
Circuit disconnected

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Disconnect alarm does not occur on patient circuit removal patient safety

No alarm triggered when patient circuit is removed from test lung and breath is fired

Ranked causes

  1. 1. Pressure sensing circuit not detecting loss of airway pressure (common)
  2. 2. Alarm threshold not configured or microcontroller logic failure (occasional)
First check

Remove patient circuit from test lung

Test point
Patient circuit connection
Expected
Circuit disconnected

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Excessive airway pressure safety shutdown does not occur patient safety

Unit does not shut down when airway pressure exceeds 40 cmH2O for 4 seconds during test

Ranked causes

  1. 1. Airway pressure transducer not properly connected or sensing failure (unknown)
  2. 2. Safety shutdown threshold not triggered at 40 cmH2O (unknown)
First check

Verify test setup with specified ventilator settings

Test point
Ventilator control panel
Expected
Mode=A/C, Rate=15 BPM, I time=1.0 sec, Vt=500 ml, FIO2=21%

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

External Power Fail alarm does not occur patient safety

No External Power Fail alarm is generated when top fuse is removed in A/C mode with external power and battery connected

Ranked causes

  1. 1. Top fuse not removed properly (occasional)
  2. 2. Alarm circuit not detecting power loss (rare)
First check

Connect external power and battery

Test point
power connections
Expected
Both connected

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Failure Code 2 occurs (No Gas, COMP alarm) patient safety

Unit displays Failure Code 2 and COMP alarm after gas supplies are disconnected during test

Ranked causes

  1. 1. External air and oxygen supplies disconnected or absent (common)
  2. 2. Oxygen supply disconnected or absent (common)
First check

Verify external air supply connected and flowing

Test point
External air inlet
Expected
Air supply connected

4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Failure Code 5 occurs with safety shutdown patient safety

Unit displays Failure Code 5 and triggers 40 cmH2O safety shutdown after 4 seconds of excessive airway pressure

Ranked causes

  1. 1. Airway pressure exceeds 40 cmH2O for 4 seconds (common)
  2. 2. Transducer disconnection or obstruction during operation (occasional)
First check

Set unit to Mode A/C, Rate 15 BPM, I time 1.0 sec, Vt 500 ml, FIO2 21%

Test point
Unit control panel
Expected
Mode=A/C, Rate=15 BPM, I time=1.0 sec, Vt=500 ml, FIO2=21%

6 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

LCD screen does not blank during system failure

LCD display remains active instead of blanking when both fuses are removed

Ranked causes

  1. 1. Fuses not fully removed or system failure condition not triggered (occasional)
  2. 2. LCD control circuit malfunction (occasional)
First check

Verify both fuses removed

Test point
Top and bottom fuse locations
Expected
Both fuses completely removed

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

System Failure LED does not illuminate solid

System Failure LED fails to turn on solid when both top and bottom fuses are removed in A/C mode

Ranked causes

  1. 1. Top fuse not removed or external power fail alarm not cleared (occasional)
  2. 2. System Failure LED circuit malfunction (occasional)
First check

Connect external power and battery in A/C mode

Test point
Power and mode settings
Expected
Unit in A/C mode with both power sources active

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not shut down during system failure patient safety

Unit remains operational instead of shutting down when both fuses are removed

Ranked causes

  1. 1. Fuses not both removed or system failure condition not triggered (occasional)
  2. 2. Shutdown control circuit malfunction (occasional)
First check

Verify both fuses removed

Test point
Top and bottom fuse locations
Expected
Both fuses completely removed

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Alarm cancel function does not clear alarm

Pressing Mute/Cancel button does not stop audible alarm or clear alarm message display when external power is disconnected

Ranked causes

  1. 1. Cancel button not responding or alarm state not clearing (unknown)
First check

Verify alarm fully cancels after button press

Test point
Mute/Cancel push-button during External Power Fail alarm
Expected
Audible alarm stops; LED turns off; AMC displays no alarm

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm cancel function not working

Pressing Mute/Cancel button to cancel External Power Fail alarm does not stop audible alarm, turn off LED, or clear message

Ranked causes

  1. 1. Mute/Cancel push-button control malfunction or power supply fault (common)
First check

Audible alarm stops when Mute/Cancel pressed during power fail

Test point
Mute/Cancel push-button, speaker output
Expected
Audible alarm stops

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm cancellation when tidal volume decreased to 500 ml patient safety

Test procedure to verify all alarms cancel when tidal volume is reduced to 500 ml

Ranked causes

  1. 1. Tidal volume setting decreased to 500 ml (common)
First check

Decrease Vt setting to 500 ml

Test point
Vt control setting
Expected
500 ml

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Alarm does not cancel when Mute/Cancel button pressed during External Power Fail

When Mute/Cancel button is pressed during an External Power Fail alarm, the audible alarm should stop completely and alarm LED should turn off, indicating the alarm has been cancelled rather than muted.

Ranked causes

  1. 1. Mute/Cancel button not canceling External Power Fail alarm (unknown)
First check

Audible alarm stops after Mute/Cancel button press

Test point
Mute/Cancel push-button during External Power Fail
Expected
The audible alarm should stop

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm does not deactivate after clearing disconnect condition patient safety

Alarm remains active or reactivates after disconnect condition is fixed

Ranked causes

  1. 1. Disconnect sensor malfunction or circuit not detecting connection restoration (occasional)
First check

Verify disconnect condition is fully restored

Test point
Patient circuit connection
Expected
Circuit securely connected with no gaps

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LED and audible reactivate on new alarm condition patient safety

Previously muted alarm LED and audible reactivate when new alarm condition is detected

Ranked causes

  1. 1. New alarm condition triggered (common)
First check

Verify alarm reactivation on new condition

Test point
Alarm status indicator
Expected
Alarm LED and audible turn on

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LED does not blink on disconnect alarm patient safety

Alarm LED fails to blink when a disconnect alarm is generated

Ranked causes

  1. 1. Disconnect condition not properly triggered (common)
First check

Generate Disconnect alarm condition

Test point
patient circuit
Expected
Patient circuit disconnected from test lung

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LED does not blink on disconnect condition patient safety

Alarm LED fails to respond to disconnect alarm trigger

Ranked causes

  1. 1. Disconnect alarm not properly generated or patient circuit disconnected from test lung (common)
First check

Remove patient circuit from test lung

Test point
Patient circuit connection
Expected
Circuit disconnected

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LED does not change state during Mute

Alarm LED fails to transition from blinking to solid during 30-second mute period, or fails to resume blinking after timeout

Ranked causes

  1. 1. Alarm LED driver or control circuitry malfunction (unknown)
First check

Observe Alarm LED state before Mute

Test point
Alarm LED indicator
Expected
LED blinking

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LED does not come on solid after mute button press

Alarm LED fails to change to solid state when Mute/Cancel push-button is pressed during disconnect alarm

Ranked causes

  1. 1. Mute/Cancel button not functioning (occasional)
First check

Press Mute/Cancel push-button

Test point
Mute/Cancel button
Expected
Button responds

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LED does not go solid after Mute/Cancel button press

Alarm LED fails to change from blinking to solid state when Mute/Cancel push-button is pressed during active alarm.

Ranked causes

  1. 1. Mute/Cancel button malfunction or LED control circuit failure (unknown)
First check

Verify Alarm LED response to Mute/Cancel button press

Test point
Alarm LED indicator
Expected
LED changes from blinking to solid state

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LED does not respond correctly

Alarm LED does not transition from blinking to solid during mute or does not turn off during cancel

Ranked causes

  1. 1. Alarm LED circuit malfunction or firmware issue (unknown)
First check

Monitor LED state during mute operation

Test point
Alarm LED indicator
Expected
LED transitions from blinking to solid for 30 seconds, then resumes blinking

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LED does not transition during Mute period

Alarm LED does not change from blinking to solid on for 30 seconds or does not resume blinking afterwards

Ranked causes

  1. 1. LED state machine or control circuit malfunction (unknown)
First check

Alarm LED state when Mute/Cancel pressed during Disconnect alarm

Test point
Alarm LED indicator
Expected
LED transitions from blinking to solid on seconds

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LED does not transition from blinking to solid

LED behavior incorrect during mute/cancel operation; should go solid for 30 seconds then resume blinking

Ranked causes

  1. 1. Mute/Cancel button not functioning correctly (common)
First check

Verify LED transitions to solid state after Mute/Cancel pressed

Test point
Alarm LED
Expected
Solid on for 30 seconds, then blinking

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LED does not turn off after disconnect condition resolved

Alarm LED remains on after patient circuit is reconnected and disconnect alarm is fixed

Ranked causes

  1. 1. Pressure sensing not detecting reconnection (occasional)
First check

Reconnect patient circuit and fire breath

Test point
Patient circuit connection and test lung
Expected
Positive airway pressure rise sensed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LED does not turn off after Mute/Cancel pressed

When Mute/Cancel is pressed during External Power Fail alarm, LED remains on instead of turning off

Ranked causes

  1. 1. Alarm LED control circuit malfunction (unknown)
First check

Alarm LED status after Mute/Cancel pressed during External Power Fail alarm

Test point
Alarm LED indicator
Expected
LED turns off

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm Message Center continues displaying alarm during Mute

Alarm Message Center displays alarm message during 30-second mute period when it should remain visible

Ranked causes

  1. 1. Mute function affects only audio, not display (expected behavior) (common)
First check

Generate Disconnect alarm and press Mute/Cancel

Test point
Alarm Message Center display
Expected
Alarm message continues displaying during 30-second mute

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm Message Center displays alarm after Mute/Cancel pressed

AMC screen continues showing alarm after Mute/Cancel button pressed during External Power Fail condition

Ranked causes

  1. 1. AMC display logic malfunction (unknown)
First check

Alarm Message Center display after Mute/Cancel pressed

Test point
Alarm Message Center screen
Expected
No alarm displayed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm message center does not display during mute patient safety

Alarm Message Center clears or stops showing disconnect alarm during mute period

Ranked causes

  1. 1. Display or message logic malfunction (unknown)
First check

Verify alarm message center display during mute

Test point
Alarm Message Center display
Expected
Disconnect alarm still visible while muted

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm mute duration incorrect caution

Audible alarm does not mute for exactly 30 seconds or LED does not transition as expected

Ranked causes

  1. 1. Mute/Cancel button malfunction or timer defect (unknown)
First check

Measure mute duration after Mute/Cancel button press

Test point
Audible alarm output
Expected
Mute for 30 seconds, then resume, ±1 second

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm mute function not working

Pressing Mute/Cancel button during active alarm does not silence audible alarm or change LED state

Ranked causes

  1. 1. Mute/Cancel push-button control malfunction (common)
First check

Audible alarm mutes when Mute/Cancel pressed during active alarm

Test point
Mute/Cancel push-button, speaker output
Expected
Audible alarm silenced for 30 seconds

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm Mute/Cancel button does not silence alarm caution

Pressing Mute/Cancel button fails to silence alarm LED and audible tone during alarm condition

Ranked causes

  1. 1. Mute/Cancel button circuit or alarm driver malfunction (occasional)
First check

Press Mute/Cancel button during alarm condition

Test point
Mute/Cancel push-button control
Expected
Button press registered

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarms cancel after oxygen concentration reduced patient safety

O2 Low/Fail and FIO2 alarms clear when concentration set to 21% despite low tank pressure

Ranked causes

  1. 1. Alarm inhibit when FIO2 concentration low (common)
First check

Verify tidal volume maintained on external air/compressor

Test point
Volume delivery
Expected
Tidal volume sustained

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarms cancel after tidal volume decreased to 500 ml patient safety

Both Vt and Vt Settings alarms clear when tidal volume reduced to 500 ml

Ranked causes

  1. 1. Tidal volume returned within acceptable range (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarms cancel when oxygen concentration reduced to 21% patient safety

Both O2 Low/Fail and FIO2 alarms clear when oxygen concentration setting is reduced to room air level

Ranked causes

  1. 1. Alarm condition resolves at lower oxygen demand (common)
First check

Set oxygen concentration to 21%

Test point
FIO2 control
Expected
21%

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarms cancel when oxygen concentration set to 21% patient safety

O2 Low/Fail and FIO2 alarms clear when concentration reduced to room air level

Ranked causes

  1. 1. Oxygen concentration at non-emergency level (common)
First check

Set oxygen concentration to room air level

Test point
Oxygen concentration control
Expected
21%

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarms cancel when tidal volume reduced patient safety

All active alarms clear when tidal volume decreased to 500 ml

Ranked causes

  1. 1. Tidal volume within acceptable range (common)
First check

Decrease tidal volume setting

Test point
Vt control
Expected
500 ml

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarms cancel when tidal volume setting decreased patient safety

All active alarms cancel when Vt setting is decreased to 500 ml

Ranked causes

  1. 1. Tidal volume setting returned to acceptable range (common)
First check

Verify tidal volume setting

Test point
Vt control
Expected
500 ml

Alarms cancel when Vt reduced to 500 ml patient safety

All active Vt and Settings alarms are cleared when tidal volume setting is decreased to 500 ml

Ranked causes

  1. 1. Tidal volume setting returns to acceptable range (common)
First check

Decrease Vt setting to 500 ml

Test point
Vt control setting
Expected
All alarms cancel

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarms cancel when Vt setting normalized

When tidal volume setting decreased to 500 ml, all active alarms cancel

Ranked causes

  1. 1. Vt setting returned to acceptable range (common)
First check

Decrease Vt setting

Test point
Vt control setting
Expected
500 ml

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarms do not cancel when oxygen concentration returns to safe level patient safety

Alarms should cancel when oxygen concentration is set to 21% after low tank pressure condition

Ranked causes

  1. 1. Alarm reset logic malfunction (occasional)
First check

Set oxygen concentration

Test point
O2 concentration control
Expected
21%

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

AMC does not display EXTERNAL POWER LOW alarm

AMC screen should display EXTERNAL POWER LOW alarm message when external power voltage falls to threshold

Ranked causes

  1. 1. Display communication or alarm message circuit failure (occasional)
First check

Visually verify AMC alarm message display

Test point
AMC screen
Expected
EXTERNAL POWER LOW alarm message displayed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarm comes on continuously patient safety

System generates continuous audible alarm during Failure Code 5 testing when airway pressure exceeds 40 cmH2O for 4 seconds

Ranked causes

  1. 1. Excessive airway pressure above 40 cmH2O sustained for 4 seconds (common)
First check

Pressurize syringe to raise airway pressure above 40 cmH2O

Test point
Transducer fitting via syringe with hose
Expected
Airway pressure above 40 cmH2O sustained for 4 seconds
Tools
Syringe with hose, pressure transducer

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarm does not mute for 30 seconds

Alarm sound continues or does not resume after mute period; should mute when Mute/Cancel pressed and resume after 30 seconds

Ranked causes

  1. 1. Mute/Cancel control circuit malfunction (common)
First check

Verify audible alarm mutes when Mute/Cancel pressed

Test point
Alarm speaker
Expected
No audible alarm for 30 seconds

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarm does not mute when Mute/Cancel button pressed

When Mute/Cancel push-button is pressed during an active alarm, the audible alarm should silence for 30 seconds before automatically resuming, but it may not mute or resume correctly.

Ranked causes

  1. 1. Mute/Cancel button malfunction or not responding (unknown)
First check

Audible alarm mutes after Mute/Cancel button press

Test point
Mute/Cancel push-button
Expected
Audible alarm should mute for 30 seconds

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarm does not sound at secondary shutdown caution

Audible alarm should sound continuously when unit shuts down at secondary voltage threshold (9.50V - 10.50V)

Ranked causes

  1. 1. Alarm circuit malfunction during secondary shutdown (occasional)
First check

Audibly verify alarm activation

Test point
Audible alarm speaker
Expected
Continuous audible alarm at secondary shutdown

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarm does not sound continuously at shutdown

System failure LED comes on but audible alarm fails to sound when unit shuts down due to low voltage

Ranked causes

  1. 1. Low battery voltage drops below shutdown threshold (common)
First check

Verify system shuts down within specified voltage range

Test point
External power jack
Expected
8.5V - 9.3V, 8.5-9.3V

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarm does not sound continuously with both fuses removed

Continuous audible alarm tone not generated when both fuses are removed

Ranked causes

  1. 1. Both fuses not properly removed (occasional)
First check

Confirm both fuses removed

Test point
fuse positions
Expected
Both top and bottom fuses removed

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarm does not sound during low battery shutdown

Continuous audible alarm does not activate when battery voltage drops below shutdown threshold

Ranked causes

  1. 1. Audible alarm circuit failure (rare)
First check

Verify audible alarm sounds

Test point
Audible alarm output
Expected
Alarm on continuously

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarm does not sound on dual fuse failure patient safety

Continuous tone from audible alarm not heard when both fuses are removed

Ranked causes

  1. 1. Top and bottom fuses not both removed or alarm circuit failure (common)
First check

Ensure both fuses removed

Test point
Top and bottom fuse locations
Expected
Both fuses absent

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarm does not sound on shutdown caution

Audible alarm should sound continuously when unit shuts down due to low voltage

Ranked causes

  1. 1. Alarm circuit malfunction or speaker failure (occasional)
First check

Audibly verify alarm activates

Test point
Audible alarm speaker
Expected
Continuous audible alarm when unit shuts down

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarm does not stop after Mute/Cancel pressed

Operator presses Mute/Cancel button but alarm continues sounding instead of stopping

Ranked causes

  1. 1. Mute/Cancel button not functioning (unknown)
First check

Audible alarm status after Mute/Cancel pressed during External Power Fail alarm

Test point
Mute/Cancel push-button control
Expected
Audible alarm stops

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarm repeating 4 to 5 pulse pattern caution

Ventilator emitting repeating 4 to 5 pulse audible alarm

Ranked causes

  1. 1. Unknown alarm condition (unknown)

More symptoms

  • Backlight does not respond to ambient light
  • Backlight does not respond to hand proximity sensor
  • Backup Ventilator flow below minimum
  • Backup ventilator flow insufficient
  • Backup ventilator flow rate below specification
  • Backup ventilator flow rate insufficient
  • Backup ventilator flow verification
  • Backup ventilator I time verification
  • Backup ventilator inspiratory time incorrect
  • Backup ventilator inspiratory time out of specification
  • Backup Ventilator manual trigger does not function correctly
  • Backup ventilator manual trigger not functioning
  • Backup ventilator PIP relief pressure incorrect
  • Backup Ventilator PIP relief pressure out of specification
  • Backup ventilator rate incorrect
  • Backup ventilator rate out of specification
  • Backup ventilator rate verification
  • Battery charger voltage output
  • Battery fail alarm not functioning
  • Battery fail condition not functioning
  • Battery Fail condition test
  • Battery shutdown voltage out of range
  • Battery voltage below shutdown threshold
  • Battery voltage measurement
  • Battery voltage out of range
  • Battery voltage out of specification
  • Battery voltage too low
  • Calibration abort when Mode switch turned away during CAL mode
  • Calibration: Airway pressure baseline
  • Calibration: Autocal valve check
  • Calibration: Backlight Threshold
  • Calibration: CAL mode verification
  • Calibration: External Air OFF LCD indicator
  • Calibration: Mode control and autocal valve
  • Calibration: Real Time Clock Reset
  • Calibration: Tidal volume alarm limits
  • Calibration: Unit pressure baseline
  • Cancel button does not stop External Power Fail alarm
  • Cancel function does not clear External Power Fail alarm
  • Canceling alarm clears all indicators
  • Charge LED does not light
  • Charger output voltage out of range
  • Charger voltage measurement
  • Charger voltage out of range
  • Charger voltage out of specification
  • Charger voltage output incorrect
  • Comp alarm does not occur when compressor operation initiated
  • Comp alarm during compressor operation
  • Comp alarm occurs during compressor start
  • Comp alarm occurs during External Air switching
  • Comp alarm occurs with low external air pressure
  • Compressor alarm (Comp) occurs
  • Compressor fails to start
  • Control breath generation at given settings
  • Control breath not generated at set parameters
  • Disconnect alarm does not cancel on circuit reconnection
  • Disconnect alarm does not cancel when circuit reconnected
  • Disconnect alarm does not deactivate after fix
  • Disconnect alarm does not occur on patient circuit disconnection
  • Disconnect alarm generated
  • Excessive Negative Pressure alarm (Failure Code 6)
  • Exhaust Valve Failure alarm (Failure Code 5)
  • Exhaust valve opens and stays open
  • Extended Non-Use alarm cannot be verified during final test
  • External Power Fail alarm does not cancel
  • External Power Fail alarm does not cancel after mute
  • External Power Fail alarm does not cancel after Mute/Cancel button press
  • External power fail alarm does not cancel properly
  • External Power Fail alarm does not cancel when Mute pressed
  • External power low alarm
  • External Power Low alarm does not activate
  • External Power Low alarm does not occur
  • External Power Low alarm does not trigger
  • External Power Low alarm does not trigger at correct voltage
  • External power low alarm does not trigger at expected voltage
  • External Power Low alarm does not trigger at threshold
  • External power low alarm occurs
  • External power low alarm threshold
  • External power low alarm threshold incorrect
  • External power shutdown
  • External power shutdown threshold
  • External power shutdown threshold not met
  • External power shutdown voltage out of range
  • Failure Code 2 - No Gas COMP alarm
  • Failure Code 2 (No Gas, COMP) alarm does not occur
  • Failure Code 2 (No Gas, Comp) alarm during low external gas pressure
  • Failure Code 2 (No Gas, Comp) alarm generation
  • Failure Code 2 displayed
  • Failure Code 2 occurs during gas supply test
  • Failure Code 3 (Excessive Airway Pressure at startup)
  • Failure Code 3 (Excessive Airway Pressure) alarm
  • Failure Code 3 displayed
  • Failure Code 4 (Memory check failure)
  • Failure Code 4 (Memory check failure) alarm
  • Failure Code 4 displayed
  • Failure Code 5 (Airway pressure exceeds 40 cmH2O for 4 seconds)
  • Failure Code 5 (Excessive Airway Pressure) safety shutdown does not trigger
  • Failure Code 5 (Exhaust Valve Failure) alarm
  • Failure Code 5 (Safety Shutdown - High Airway Pressure) alarm
  • Failure Code 5 alarm triggered
  • Failure Code 5 displayed
  • Failure Code 5 displayed with safety shutdown
  • Failure Code 5 safety shutdown at high airway pressure
  • Failure Code 5 triggered
  • Failure Code 5 triggered with high airway pressure
  • Failure Code 5 with safety shutdown
  • Failure Code 5: Excessive airway pressure safety shutdown
  • Failure Code 5: Exhaust Valve Failure
  • Failure Code 6 (Excessive Negative Pressure) alarm
  • Failure Code 6 alarm triggered
  • Failure Code 6 displayed
  • Failure Code 6 triggered
  • Failure Code 6: Excessive Negative Pressure
  • Failure Code 7 occurs during operation
  • FIO2 alarm does not occur at low oxygen pressure
  • FIO2 alarm does not occur at low oxygen tank pressure
  • FIO2 alarm occurs
  • FIO2 alarm occurs at low oxygen pressure
  • FIO2 alarm occurs at low oxygen tank pressure
  • FIO2 alarm occurs at low tank pressure
  • Flow in Backup Ventilator test
  • Gas flow does not stop at 40 cmH2O pressure
  • Gas flow does not stop at pressure relief threshold
  • I time in Backup Ventilator test
  • LCD screen blanks
  • LCD screen does not blank with both fuses removed
  • Low battery alarm does not activate at correct threshold
  • Low battery alarm does not trigger
  • Low battery alarm does not trigger at correct voltage
  • Low battery alarm not triggering
  • Low Battery Alarm occurs
  • Low Battery Alarm occurs at incorrect voltage
  • Low battery alarm threshold out of specification
  • Low Battery Alarm threshold test
  • Low Battery Alarm triggers
  • Manual breath beep not heard when button held during inhalation
  • Manual Breath button does not generate breath
  • Manual breath button does not produce audible beep
  • Manual Breath button does not trigger breath
  • Manual Breath button generates no audible beep
  • Manual Breath button held during inhalation does not queue breath after cycle
  • Manual Breath button holds breath when pressed during inhalation
  • Manual Breath button triggers during inhalation or exhalation
  • Manual Breath button unresponsive
  • Manual Breath button unresponsive during inhalation or exhalation
  • Manual Breath button unresponsive when held during inhalation
  • Manual Breath control does not generate breath
  • Manual breath delivery during setup
  • Manual breath does not generate audible beep
  • Manual Breath does not generate control breath
  • Manual breath does not trigger
  • Manual breath does not trigger during inspiration or expiration
  • Manual Breath function inactive during inspiration or exhalation
  • Manual Breath function not triggering
  • Manual Breath function unresponsive when button held during inhalation
  • Manual breath hold not functioning
  • Manual breath not generated
  • Manual breath not triggered
  • Manual breath push-button control
  • Manual breath triggered during inhalation or exhalation hangs or continues indefinitely
  • Manual trigger does not control gas flow properly
  • Manual trigger does not deliver gas flow
  • Manual trigger function verification
  • Manual trigger gas flow does not respond to pressure buildup
  • Manual Trigger in Backup Ventilator test
  • Mode switch to A/C
  • Mode switch to CAL abort test
  • Mode switch to CAL with successful calibration
  • Mode switch to CPAP
  • Mode switch to OFF
  • Mode switch to SIMV
  • Mode switch: A/C mode
  • Mode switch: CAL mode abort
  • Mode switch: CAL mode calibration
  • Mode switch: CPAP mode
  • Mode switch: OFF mode
  • Mode switch: SIMV mode
  • Mute button does not silence alarm for 30 seconds
  • Mute/Cancel button does not silence alarm message
  • Mute/Cancel function suppresses alarm audible and LED
  • Muting alarm for 30 seconds
  • No Gas alarm (Failure Code 2) does not trigger at low external gas pressure
  • O2 Low/Fail alarm does not occur at low oxygen tank pressure
  • O2 Low/Fail alarm does not occur at low tank pressure
  • O2 Low/Fail alarm occurs
  • O2 Low/Fail alarm occurs at low oxygen pressure
  • O2 Low/Fail alarm occurs at low oxygen tank pressure
  • O2 Low/Fail alarm occurs at low tank pressure
  • O2 Low/Fail alarm triggers at low tank pressure
  • PIC Line 1 does not display EXT PWR LOW message
  • PIP relief does not occur at correct pressure
  • PIP relief in Backup Ventilator test
  • PIP relief not occurring at correct pressure
  • PIP relief pressure verification
  • Preventative Maintenance alarm cannot be verified during final test
  • Rate in Backup Ventilator test
  • Rate, Inspiration Time, Tidal Volume controls non-functional in CPAP mode
  • System does not shut down at correct external power threshold
  • System does not shut down at low external power
  • System does not shut down at secondary threshold with external power low
  • System fail LED comes on
  • System Failure LED and alarm activate incorrectly
  • System failure LED does not activate on shutdown
  • System failure LED does not come on during shutdown
  • System Failure LED does not come on solid with both fuses removed
  • System failure LED does not illuminate at secondary shutdown
  • System failure LED does not illuminate during low battery
  • System Failure LED does not illuminate on dual fuse failure
  • System failure LED does not illuminate on shutdown
  • System failure LED should come on during shutdown
  • System Failure LED solid on
  • System shuts down during external power test
  • Tidal volume not maintained during low oxygen pressure
  • Total Flow Backup alarm after compressor inlet occlusion
  • Total Flow Backup alarm does not occur after compressor occlusion
  • Total Flow Backup alarm during low compressor inlet flow
  • Total Flow Backup Alarm not occurring
  • Total Flow Backup alarm not triggered
  • Total Flow Backup alarm occurs
  • Total Flow Backup alarm occurs during compressor inlet occlusion
  • Total Flow Backup alarm occurs with compressor inlet occlusion
  • Unit does not shut down after gas supply disconnected
  • Unit does not shut down at -10 cmH2O airway pressure
  • Unit does not shut down at +100 cmH2O airway pressure
  • Unit does not shut down at expected low voltage
  • Unit does not shut down at expected voltage
  • Unit does not shut down at External Power Low shutdown threshold
  • Unit does not shut down at low external power voltage
  • Unit does not shut down at negative airway pressure
  • Unit does not shut down at second low voltage threshold
  • Unit does not shut down on dual fuse failure
  • Unit does not shut down with both fuses removed
  • Unit should shut down at excessive negative pressure
  • Unit should shut down at high airway pressure
  • Ventilator failure code 2 (No Gas, COMP alarm)
  • Verify tidal volume maintained after oxygen concentration change to 21%
  • Verify tidal volume maintained during oxygen pressure low condition
  • Verify tidal volume maintained when oxygen concentration returned to 60%
  • Vt alarm cascades with Vt Settings alarm on display
  • Vt alarm clears at 500 ml setting
  • Vt alarm displaces Settings alarm on display
  • Vt alarm does not occur at high tidal volume
  • Vt alarm does not occur when tidal volume further increased
  • Vt alarm occurs
  • Vt alarm occurs above 1150 ml
  • Vt alarm occurs at 1150 ml
  • Vt alarm occurs at high tidal volume setting
  • Vt alarm occurs when setting increased
  • Vt alarm occurs when tidal volume exceeds 1150 ml
  • Vt alarm occurs when tidal volume exceeds critical limit
  • Vt alarm occurs when tidal volume exceeds higher limit
  • Vt alarm occurs when tidal volume exceeds secondary limit
  • Vt alarm occurs when tidal volume increased further
  • Vt alarm replaces Vt Settings alarm
  • Vt alarm when tidal volume increased to 1150 ml
  • Vt or Disconnect alarms occur unexpectedly during external air/compressor switching
  • Vt Settings alarm does not occur
  • Vt Settings alarm does not occur when tidal volume exceeds limits
  • Vt Settings alarm does not occur when tidal volume increased
  • Vt Settings alarm occurs
  • Vt Settings alarm occurs above 1050 ml
  • Vt Settings alarm occurs at 1050 ml
  • Vt Settings alarm occurs at elevated tidal volume
  • Vt Settings alarm occurs at high tidal volume
  • Vt Settings alarm occurs at high tidal volume setting
  • Vt Settings alarm occurs when tidal volume exceeds 1050 ml
  • Vt Settings alarm occurs when tidal volume exceeds limit
  • Vt Settings alarm occurs when tidal volume exceeds range
  • Vt Settings alarm triggered
  • Vt Settings alarm triggers on parameter change
  • Vt Settings alarm when tidal volume set above 1000 ml

271 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
5Exhaust Valve FailureAirway pressure reaches +100 cmH2Oerror
6Excessive Negative PressureAirway pressure drops to -10 cmH2Oerror
7Run-Time Transducer Calibration FailureTransducer calibration fails during operationerror
Audible alarmContinuous audible alarm indicating system shutdown or critical conditionSystem voltage drops below shutdown threshold (8.5V battery or 9.5V external), or critical fault conditioncritical
Audible Alarm 4-5 pulseRepeating audible alarm with 4 to 5 pulse patternSystem failure or alert conditionerror
BATTERY LOW/FAILBattery voltage too low or battery failure detectedBattery voltage drops below 11.1V or battery fails during low power conditionwarning
Calibration Abort AlarmCalibration sequence was interrupted or abortedMode control switch is moved away from CAL mode before calibration completeswarning
Calibration Abort AlarmMode switch turned away from CAL mode during calibrationUser turns Mode control switch from CAL to another mode before calibration completeswarning
CHK EXT FUSECheck external power fuseExternal Power Fail alarm with only top fuse removed and bottom fuse installedwarning
CompCompressor alarmExternal gas supply pressure drops below minimum threshold or compressor fails to startwarning
CompCompressor failure or inability to provide gas at required pressureCompressor startup when external air supply is insufficient or compressor malfunctionerror
CompCompressor alarm or failure code related to compressor operationCompressor fails to start or operate during source switching; external gas supply drops below 35 PSI +/- 2 PSIwarning
Comp alarmCompressor operation alarm during mode transition or startup failureCompressor attempts to start during external air to compressor switchingwarning
Comp alarmCompressor failure or unavailable; gas source switched to external airCompressor fails to start or external air supply pressure drops below 35 PSI +/- 2 PSIwarning
Comp alarmCompressor alarm - compressor failed to provide gas or pressure backup neededExternal air supply depleted or compressor unable to maintain pressure during switch from external air to compressor operationerror
Comp alarmCompressor alarm conditionCompressor activation during external air operation or external gas pressure below 35 PSI ±2 PSIwarning
DisconnectPatient circuit disconnected or no positive airway pressure detectedPatient circuit removed from test lung or pressure rise not sensed after breath is firederror
Disconnect AlarmPatient circuit disconnected from ventilator outletBreath initiated with no positive airway pressure rise sensedcritical
Extended Non-UseAlarm indicating prolonged non-operation of ventilatorVentilator inactive for extended periodwarning
External Power FailExternal power supply has been disconnected or is not functioningTop fuse removed or external power supply disconnected while in A/C modewarning
External Power Fail alarmExternal AC power supply lost or disconnectedDisconnection of external power sourcewarning
EXTERNAL POWER LOWExternal power supply voltage has dropped below safe operating levelExternal power voltage drops below 10.5V - 11.1V rangewarning
Failure Code 1Self Check failure during startupSelf check fails during system startupcritical
Failure Code 2No Gas, COMPGas supplies (external air and oxygen) are disconnectederror
Failure Code 2No gas from compressor during alarm conditionExternal gas supply pressure drops below 35 PSI +/- 2 PSI while compressor is expected to provide backuperror
Failure Code 2 (No Gas, Comp)No gas supply available to compressor; compressor cannot operateExternal gas pressure drops below 35 PSI +/- 2 PSI during compressor operation attemptcritical
Failure Code 3Excessive airway pressure detected during startupAirway pressure exceeds safe limits at system startupcritical
Failure Code 4Memory check failure alarmMemory check fails during startup Self Checkcritical
Failure Code 5Exhaust Valve FailureAirway pressure exceeds +100 cmH2O for 2 seconds without unit shutdownerror
Failure Code 5Excessive airway pressure detected - safety shutdown triggeredAirway pressure exceeds 40 cmH2O for 4 consecutive seconds during operationcritical
Failure Code 5Safety shutdown triggered by excessive airway pressureAirway pressure exceeds 40 cmH2O for 4 seconds or longercritical
Failure Code 6Excessive Negative PressureAirway pressure drops to -10 cmH2O or below for 1.5 seconds without unit shutdownerror
Failure Code 7Run-Time Transducer Calibration FailureTransducer calibration check fails during operationerror
FIO2Insufficient oxygen flow to achieve requested FIO2 concentrationOxygen tank pressure below setpoint when FIO2 setting is 60% or higherwarning
FIO2Fractional inspired oxygen out of specificationOxygen concentration setting and supply pressure mismatch or oxygen delivery failurecritical
FIO2Oxygen concentration cannot be delivered at requested levelOxygen supply pressure insufficient to meet target FIO2 settingcritical
FIO2 alarmActual oxygen concentration cannot match the set FIO2 valueOxygen tank pressure is insufficient to deliver the set oxygen concentration percentagewarning
FIO2 alarmInsufficient oxygen pressure for requested oxygen concentrationOxygen tank pressure below 45 PSI during high oxygen concentration setting (60%)error
LOW (Battery Icon on PIC Line 2)Battery voltage low indication on displayBattery voltage drops into low range during operationwarning
O2 Low/FailOxygen supply pressure is below acceptable thresholdOxygen tank pressure drops below 35 PSI +/- 2 PSIcritical
O2 Low/FailOxygen supply pressure below thresholdOxygen tank pressure drops to 35 PSI +/- 2 PSI or belowwarning
O2 Low/Fail alarmOxygen tank pressure is below minimum thresholdOxygen tank pressure drops below 35 PSI while ventilator is operatingcritical
Preventative MaintenanceAlarm indicating scheduled maintenance is dueMaintenance interval elapsed or operational hours exceededinfo
System FailureCritical system failure; both external power and internal power sources have failedBoth top and bottom fuses removed simultaneouslycritical
System FailureCritical system failure requiring unit shutdownBoth top and bottom fuses removed or both power sources failedcritical
System FailureMultiple system fuses have failed indicating critical system failureBoth top and bottom fuses removed or open circuitcritical
System failure LEDSystem failure detected, LED illuminated continuouslySystem voltage drops below minimum safe level (8.5V battery or 9.5V external), or critical system fault detectederror
Total Flow BackupTotal flow backup alarm during low tidal volume conditionActual delivered tidal volume drops below 50% of set tidal volume for multiple breathswarning
Total Flow BackupActual delivered tidal volume falls significantly below set valueCompressor inlet obstruction or gas flow restriction causes delivered volume to drop below 50% of set value after 4-5 breathswarning
Total Flow BackupBackup flow alarm due to insufficient delivered tidal volumeActual tidal volume less than half of set tidal volume for 4-5 breathserror
Total Flow Backup AlarmActual tidal volume fell below threshold during backup flow operationActual tidal volume drops to less than half of the set tidal volume for 4-5 breathswarning
Total Flow Backup AlarmActual delivered tidal volume is significantly less than set value indicating breathing circuit obstruction or compressor failureActual tidal volume measured less than half of set tidal volume for 4-5 breathswarning
Total Flow Backup alarmTotal flow backup threshold exceededActual delivered tidal volume falls below 50% of set tidal volume for 4-5 breathswarning
Total Flow Backup alarmCompressor output flow insufficient; actual tidal volume below minimum thresholdActual delivered tidal volume falls below half the set tidal volume for 4-5 breath cycleserror
VtMeasured tidal volume exceeds safe operating limitTidal volume set to 1150 ml or highererror
VtDelivered tidal volume is out of specificationVt setting increased to 1150 ml; exceeds critical thresholderror
VtTidal volume setting significantly exceeds safe maximum limitTidal volume setting increased to 1150 ml or highererror
Vt alarmMeasured tidal volume has deviated significantly from set valueTidal volume setting is increased above approximately 1100 ml or actual delivered volume exceeds limiterror
Vt AlarmTidal volume out of acceptable rangeDelivered tidal volume deviates from set value during normal operationwarning
Vt alarmTidal volume alarmDelivered tidal volume deviates from set valuewarning
Vt SettingsTidal volume setting outside acceptable operating rangeTidal volume setting increased above approximately 1025 mlwarning
Vt SettingsTidal volume setting has been changed from baseline configurationVt changed from 800 ml to 1050 ml or other non-baseline valuewarning
Vt Settings alarmTidal volume setting is outside acceptable operating rangeTidal volume setting is increased above approximately 1000 mlwarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.