Impact Instrumentation Ventilator accessory: troubleshooting and error codes
The ServiceForge knowledge base holds 331 documented symptoms for the Impact Instrumentation Ventilator accessory, with 349 ranked causes, 702 diagnostic checks with test points and expected values, and 63 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Calibration: Mode control and CAL mode operation
Testing MODE control switch positions (OFF, A/C, SIMV, CPAP, CAL) and verifying unit response, LCD display, and calibration completion.
Ranked causes
- 1. Mode switch not transitioning properly or unit not responding to mode changes (unknown)
- 2. Calibration abort or incomplete calibration cycle (unknown)
- 3. Autocal Valve not functioning or pressure baseline incorrect (unknown)
Verify unit powers off
- Test point
- Unit status indicator
- Expected
- Unit turns off and is non-operational
7 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED does not blink during disconnect alarm
Alarm LED fails to blink when disconnect alarm is generated
Ranked causes
- 1. Disconnect alarm not generated (occasional)
- 2. Alarm LED circuit failure (occasional)
Generate disconnect alarm condition
- Test point
- Patient circuit connection
- Expected
- Disconnect alarm occurs on next breath
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED does not respond to Disconnect alarm
Alarm LED fails to blink when Disconnect alarm is generated
Ranked causes
- 1. Disconnect alarm not generated (common)
- 2. Alarm LED circuit malfunction (occasional)
Generate Disconnect alarm condition
- Test point
- Patient circuit connection
- Expected
- Alarm LED blinking
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED does not turn solid when mute button pressed
Alarm LED fails to change from blinking to solid state after Mute/Cancel button press during alarm condition
Ranked causes
- 1. Mute/Cancel button not responding (occasional)
- 2. Alarm LED control circuit malfunction (occasional)
Press Mute/Cancel button during alarm
- Test point
- Mute/Cancel push-button
- Expected
- Button depresses and registers
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Charge LED does not illuminate
Charge indicator LED fails to light when external power connected and battery disconnected
Ranked causes
- 1. External power supply not properly connected or insufficient power (common)
- 2. Charge LED requires warm-up time (occasional)
Verify external power supply connected
- Test point
- External power input
- Expected
- Power supply connected and active
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Comp alarm occurs patient safety
Compressor alarm activates when switching between external air and compressor operation, particularly when external air button is pressed after one external air breath.
Ranked causes
- 1. Compressor failure to start (common)
- 2. External gas pressure below minimum (35 PSI) (occasional)
Verify LCD indicator status after pressing External Air button
- Test point
- LCD display
- Expected
- OFF
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Comp alarm occurs during external air to compressor switching patient safety
Compressor alarm is artificially induced when switching between external air and compressor operation; external air valves supply gas during alarm
Ranked causes
- 1. Compressor fails to start or continues operation abnormally during source switching (unknown)
- 2. External gas supply below minimum threshold (Failure Code 2) (occasional)
Press External Air button after one external air breath
- Test point
- LCD indicator
- Expected
- OFF
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Comp alarm with external air and compressor switching
Comp alarm occurs during switching between external air and compressor operation, along with possible Vt and Disconnect alarms. External air valves supply gas source during Comp alarm.
Ranked causes
- 1. Compressor failure or malfunction (unknown)
- 2. External gas pressure below 35 PSI (common)
Press External Air push-button control after one external air breath
- Test point
- External Air push-button
- Expected
- LCD indicator reads OFF
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Disconnect alarm does not occur on next breath patient safety
Disconnect alarm not triggered when patient circuit is removed from test lung and next breath is fired
Ranked causes
- 1. Patient circuit not properly disconnected (common)
- 2. Pressure sensor not detecting lack of airway pressure rise (rare)
Remove patient circuit from test lung
- Test point
- patient circuit connection
- Expected
- Circuit disconnected
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Disconnect alarm does not occur on patient circuit removal patient safety
No alarm triggered when patient circuit is removed from test lung and breath is fired
Ranked causes
- 1. Pressure sensing circuit not detecting loss of airway pressure (common)
- 2. Alarm threshold not configured or microcontroller logic failure (occasional)
Remove patient circuit from test lung
- Test point
- Patient circuit connection
- Expected
- Circuit disconnected
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive airway pressure safety shutdown does not occur patient safety
Unit does not shut down when airway pressure exceeds 40 cmH2O for 4 seconds during test
Ranked causes
- 1. Airway pressure transducer not properly connected or sensing failure (unknown)
- 2. Safety shutdown threshold not triggered at 40 cmH2O (unknown)
Verify test setup with specified ventilator settings
- Test point
- Ventilator control panel
- Expected
- Mode=A/C, Rate=15 BPM, I time=1.0 sec, Vt=500 ml, FIO2=21%
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
External Power Fail alarm does not occur patient safety
No External Power Fail alarm is generated when top fuse is removed in A/C mode with external power and battery connected
Ranked causes
- 1. Top fuse not removed properly (occasional)
- 2. Alarm circuit not detecting power loss (rare)
Connect external power and battery
- Test point
- power connections
- Expected
- Both connected
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Failure Code 2 occurs (No Gas, COMP alarm) patient safety
Unit displays Failure Code 2 and COMP alarm after gas supplies are disconnected during test
Ranked causes
- 1. External air and oxygen supplies disconnected or absent (common)
- 2. Oxygen supply disconnected or absent (common)
Verify external air supply connected and flowing
- Test point
- External air inlet
- Expected
- Air supply connected
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Failure Code 5 occurs with safety shutdown patient safety
Unit displays Failure Code 5 and triggers 40 cmH2O safety shutdown after 4 seconds of excessive airway pressure
Ranked causes
- 1. Airway pressure exceeds 40 cmH2O for 4 seconds (common)
- 2. Transducer disconnection or obstruction during operation (occasional)
Set unit to Mode A/C, Rate 15 BPM, I time 1.0 sec, Vt 500 ml, FIO2 21%
- Test point
- Unit control panel
- Expected
- Mode=A/C, Rate=15 BPM, I time=1.0 sec, Vt=500 ml, FIO2=21%
6 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
LCD screen does not blank during system failure
LCD display remains active instead of blanking when both fuses are removed
Ranked causes
- 1. Fuses not fully removed or system failure condition not triggered (occasional)
- 2. LCD control circuit malfunction (occasional)
Verify both fuses removed
- Test point
- Top and bottom fuse locations
- Expected
- Both fuses completely removed
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
System Failure LED does not illuminate solid
System Failure LED fails to turn on solid when both top and bottom fuses are removed in A/C mode
Ranked causes
- 1. Top fuse not removed or external power fail alarm not cleared (occasional)
- 2. System Failure LED circuit malfunction (occasional)
Connect external power and battery in A/C mode
- Test point
- Power and mode settings
- Expected
- Unit in A/C mode with both power sources active
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not shut down during system failure patient safety
Unit remains operational instead of shutting down when both fuses are removed
Ranked causes
- 1. Fuses not both removed or system failure condition not triggered (occasional)
- 2. Shutdown control circuit malfunction (occasional)
Verify both fuses removed
- Test point
- Top and bottom fuse locations
- Expected
- Both fuses completely removed
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Alarm cancel function does not clear alarm
Pressing Mute/Cancel button does not stop audible alarm or clear alarm message display when external power is disconnected
Ranked causes
- 1. Cancel button not responding or alarm state not clearing (unknown)
Verify alarm fully cancels after button press
- Test point
- Mute/Cancel push-button during External Power Fail alarm
- Expected
- Audible alarm stops; LED turns off; AMC displays no alarm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm cancel function not working
Pressing Mute/Cancel button to cancel External Power Fail alarm does not stop audible alarm, turn off LED, or clear message
Ranked causes
- 1. Mute/Cancel push-button control malfunction or power supply fault (common)
Audible alarm stops when Mute/Cancel pressed during power fail
- Test point
- Mute/Cancel push-button, speaker output
- Expected
- Audible alarm stops
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm cancellation when tidal volume decreased to 500 ml patient safety
Test procedure to verify all alarms cancel when tidal volume is reduced to 500 ml
Ranked causes
- 1. Tidal volume setting decreased to 500 ml (common)
Decrease Vt setting to 500 ml
- Test point
- Vt control setting
- Expected
- 500 ml
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Alarm does not cancel when Mute/Cancel button pressed during External Power Fail
When Mute/Cancel button is pressed during an External Power Fail alarm, the audible alarm should stop completely and alarm LED should turn off, indicating the alarm has been cancelled rather than muted.
Ranked causes
- 1. Mute/Cancel button not canceling External Power Fail alarm (unknown)
Audible alarm stops after Mute/Cancel button press
- Test point
- Mute/Cancel push-button during External Power Fail
- Expected
- The audible alarm should stop
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm does not deactivate after clearing disconnect condition patient safety
Alarm remains active or reactivates after disconnect condition is fixed
Ranked causes
- 1. Disconnect sensor malfunction or circuit not detecting connection restoration (occasional)
Verify disconnect condition is fully restored
- Test point
- Patient circuit connection
- Expected
- Circuit securely connected with no gaps
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED and audible reactivate on new alarm condition patient safety
Previously muted alarm LED and audible reactivate when new alarm condition is detected
Ranked causes
- 1. New alarm condition triggered (common)
Verify alarm reactivation on new condition
- Test point
- Alarm status indicator
- Expected
- Alarm LED and audible turn on
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED does not blink on disconnect alarm patient safety
Alarm LED fails to blink when a disconnect alarm is generated
Ranked causes
- 1. Disconnect condition not properly triggered (common)
Generate Disconnect alarm condition
- Test point
- patient circuit
- Expected
- Patient circuit disconnected from test lung
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED does not blink on disconnect condition patient safety
Alarm LED fails to respond to disconnect alarm trigger
Ranked causes
- 1. Disconnect alarm not properly generated or patient circuit disconnected from test lung (common)
Remove patient circuit from test lung
- Test point
- Patient circuit connection
- Expected
- Circuit disconnected
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED does not change state during Mute
Alarm LED fails to transition from blinking to solid during 30-second mute period, or fails to resume blinking after timeout
Ranked causes
- 1. Alarm LED driver or control circuitry malfunction (unknown)
Observe Alarm LED state before Mute
- Test point
- Alarm LED indicator
- Expected
- LED blinking
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED does not come on solid after mute button press
Alarm LED fails to change to solid state when Mute/Cancel push-button is pressed during disconnect alarm
Ranked causes
- 1. Mute/Cancel button not functioning (occasional)
Press Mute/Cancel push-button
- Test point
- Mute/Cancel button
- Expected
- Button responds
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED does not go solid after Mute/Cancel button press
Alarm LED fails to change from blinking to solid state when Mute/Cancel push-button is pressed during active alarm.
Ranked causes
- 1. Mute/Cancel button malfunction or LED control circuit failure (unknown)
Verify Alarm LED response to Mute/Cancel button press
- Test point
- Alarm LED indicator
- Expected
- LED changes from blinking to solid state
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED does not respond correctly
Alarm LED does not transition from blinking to solid during mute or does not turn off during cancel
Ranked causes
- 1. Alarm LED circuit malfunction or firmware issue (unknown)
Monitor LED state during mute operation
- Test point
- Alarm LED indicator
- Expected
- LED transitions from blinking to solid for 30 seconds, then resumes blinking
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED does not transition during Mute period
Alarm LED does not change from blinking to solid on for 30 seconds or does not resume blinking afterwards
Ranked causes
- 1. LED state machine or control circuit malfunction (unknown)
Alarm LED state when Mute/Cancel pressed during Disconnect alarm
- Test point
- Alarm LED indicator
- Expected
- LED transitions from blinking to solid on seconds
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED does not transition from blinking to solid
LED behavior incorrect during mute/cancel operation; should go solid for 30 seconds then resume blinking
Ranked causes
- 1. Mute/Cancel button not functioning correctly (common)
Verify LED transitions to solid state after Mute/Cancel pressed
- Test point
- Alarm LED
- Expected
- Solid on for 30 seconds, then blinking
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED does not turn off after disconnect condition resolved
Alarm LED remains on after patient circuit is reconnected and disconnect alarm is fixed
Ranked causes
- 1. Pressure sensing not detecting reconnection (occasional)
Reconnect patient circuit and fire breath
- Test point
- Patient circuit connection and test lung
- Expected
- Positive airway pressure rise sensed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED does not turn off after Mute/Cancel pressed
When Mute/Cancel is pressed during External Power Fail alarm, LED remains on instead of turning off
Ranked causes
- 1. Alarm LED control circuit malfunction (unknown)
Alarm LED status after Mute/Cancel pressed during External Power Fail alarm
- Test point
- Alarm LED indicator
- Expected
- LED turns off
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm Message Center continues displaying alarm during Mute
Alarm Message Center displays alarm message during 30-second mute period when it should remain visible
Ranked causes
- 1. Mute function affects only audio, not display (expected behavior) (common)
Generate Disconnect alarm and press Mute/Cancel
- Test point
- Alarm Message Center display
- Expected
- Alarm message continues displaying during 30-second mute
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm Message Center displays alarm after Mute/Cancel pressed
AMC screen continues showing alarm after Mute/Cancel button pressed during External Power Fail condition
Ranked causes
- 1. AMC display logic malfunction (unknown)
Alarm Message Center display after Mute/Cancel pressed
- Test point
- Alarm Message Center screen
- Expected
- No alarm displayed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm message center does not display during mute patient safety
Alarm Message Center clears or stops showing disconnect alarm during mute period
Ranked causes
- 1. Display or message logic malfunction (unknown)
Verify alarm message center display during mute
- Test point
- Alarm Message Center display
- Expected
- Disconnect alarm still visible while muted
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm mute duration incorrect caution
Audible alarm does not mute for exactly 30 seconds or LED does not transition as expected
Ranked causes
- 1. Mute/Cancel button malfunction or timer defect (unknown)
Measure mute duration after Mute/Cancel button press
- Test point
- Audible alarm output
- Expected
- Mute for 30 seconds, then resume, ±1 second
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm mute function not working
Pressing Mute/Cancel button during active alarm does not silence audible alarm or change LED state
Ranked causes
- 1. Mute/Cancel push-button control malfunction (common)
Audible alarm mutes when Mute/Cancel pressed during active alarm
- Test point
- Mute/Cancel push-button, speaker output
- Expected
- Audible alarm silenced for 30 seconds
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm Mute/Cancel button does not silence alarm caution
Pressing Mute/Cancel button fails to silence alarm LED and audible tone during alarm condition
Ranked causes
- 1. Mute/Cancel button circuit or alarm driver malfunction (occasional)
Press Mute/Cancel button during alarm condition
- Test point
- Mute/Cancel push-button control
- Expected
- Button press registered
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarms cancel after oxygen concentration reduced patient safety
O2 Low/Fail and FIO2 alarms clear when concentration set to 21% despite low tank pressure
Ranked causes
- 1. Alarm inhibit when FIO2 concentration low (common)
Verify tidal volume maintained on external air/compressor
- Test point
- Volume delivery
- Expected
- Tidal volume sustained
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarms cancel after tidal volume decreased to 500 ml patient safety
Both Vt and Vt Settings alarms clear when tidal volume reduced to 500 ml
Ranked causes
- 1. Tidal volume returned within acceptable range (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarms cancel when oxygen concentration reduced to 21% patient safety
Both O2 Low/Fail and FIO2 alarms clear when oxygen concentration setting is reduced to room air level
Ranked causes
- 1. Alarm condition resolves at lower oxygen demand (common)
Set oxygen concentration to 21%
- Test point
- FIO2 control
- Expected
- 21%
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarms cancel when oxygen concentration set to 21% patient safety
O2 Low/Fail and FIO2 alarms clear when concentration reduced to room air level
Ranked causes
- 1. Oxygen concentration at non-emergency level (common)
Set oxygen concentration to room air level
- Test point
- Oxygen concentration control
- Expected
- 21%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarms cancel when tidal volume reduced patient safety
All active alarms clear when tidal volume decreased to 500 ml
Ranked causes
- 1. Tidal volume within acceptable range (common)
Decrease tidal volume setting
- Test point
- Vt control
- Expected
- 500 ml
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarms cancel when tidal volume setting decreased patient safety
All active alarms cancel when Vt setting is decreased to 500 ml
Ranked causes
- 1. Tidal volume setting returned to acceptable range (common)
Verify tidal volume setting
- Test point
- Vt control
- Expected
- 500 ml
Alarms cancel when Vt reduced to 500 ml patient safety
All active Vt and Settings alarms are cleared when tidal volume setting is decreased to 500 ml
Ranked causes
- 1. Tidal volume setting returns to acceptable range (common)
Decrease Vt setting to 500 ml
- Test point
- Vt control setting
- Expected
- All alarms cancel
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarms cancel when Vt setting normalized
When tidal volume setting decreased to 500 ml, all active alarms cancel
Ranked causes
- 1. Vt setting returned to acceptable range (common)
Decrease Vt setting
- Test point
- Vt control setting
- Expected
- 500 ml
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarms do not cancel when oxygen concentration returns to safe level patient safety
Alarms should cancel when oxygen concentration is set to 21% after low tank pressure condition
Ranked causes
- 1. Alarm reset logic malfunction (occasional)
Set oxygen concentration
- Test point
- O2 concentration control
- Expected
- 21%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
AMC does not display EXTERNAL POWER LOW alarm
AMC screen should display EXTERNAL POWER LOW alarm message when external power voltage falls to threshold
Ranked causes
- 1. Display communication or alarm message circuit failure (occasional)
Visually verify AMC alarm message display
- Test point
- AMC screen
- Expected
- EXTERNAL POWER LOW alarm message displayed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm comes on continuously patient safety
System generates continuous audible alarm during Failure Code 5 testing when airway pressure exceeds 40 cmH2O for 4 seconds
Ranked causes
- 1. Excessive airway pressure above 40 cmH2O sustained for 4 seconds (common)
Pressurize syringe to raise airway pressure above 40 cmH2O
- Test point
- Transducer fitting via syringe with hose
- Expected
- Airway pressure above 40 cmH2O sustained for 4 seconds
- Tools
- Syringe with hose, pressure transducer
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm does not mute for 30 seconds
Alarm sound continues or does not resume after mute period; should mute when Mute/Cancel pressed and resume after 30 seconds
Ranked causes
- 1. Mute/Cancel control circuit malfunction (common)
Verify audible alarm mutes when Mute/Cancel pressed
- Test point
- Alarm speaker
- Expected
- No audible alarm for 30 seconds
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm does not mute when Mute/Cancel button pressed
When Mute/Cancel push-button is pressed during an active alarm, the audible alarm should silence for 30 seconds before automatically resuming, but it may not mute or resume correctly.
Ranked causes
- 1. Mute/Cancel button malfunction or not responding (unknown)
Audible alarm mutes after Mute/Cancel button press
- Test point
- Mute/Cancel push-button
- Expected
- Audible alarm should mute for 30 seconds
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm does not sound at secondary shutdown caution
Audible alarm should sound continuously when unit shuts down at secondary voltage threshold (9.50V - 10.50V)
Ranked causes
- 1. Alarm circuit malfunction during secondary shutdown (occasional)
Audibly verify alarm activation
- Test point
- Audible alarm speaker
- Expected
- Continuous audible alarm at secondary shutdown
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm does not sound continuously at shutdown
System failure LED comes on but audible alarm fails to sound when unit shuts down due to low voltage
Ranked causes
- 1. Low battery voltage drops below shutdown threshold (common)
Verify system shuts down within specified voltage range
- Test point
- External power jack
- Expected
- 8.5V - 9.3V, 8.5-9.3V
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm does not sound continuously with both fuses removed
Continuous audible alarm tone not generated when both fuses are removed
Ranked causes
- 1. Both fuses not properly removed (occasional)
Confirm both fuses removed
- Test point
- fuse positions
- Expected
- Both top and bottom fuses removed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm does not sound during low battery shutdown
Continuous audible alarm does not activate when battery voltage drops below shutdown threshold
Ranked causes
- 1. Audible alarm circuit failure (rare)
Verify audible alarm sounds
- Test point
- Audible alarm output
- Expected
- Alarm on continuously
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm does not sound on dual fuse failure patient safety
Continuous tone from audible alarm not heard when both fuses are removed
Ranked causes
- 1. Top and bottom fuses not both removed or alarm circuit failure (common)
Ensure both fuses removed
- Test point
- Top and bottom fuse locations
- Expected
- Both fuses absent
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm does not sound on shutdown caution
Audible alarm should sound continuously when unit shuts down due to low voltage
Ranked causes
- 1. Alarm circuit malfunction or speaker failure (occasional)
Audibly verify alarm activates
- Test point
- Audible alarm speaker
- Expected
- Continuous audible alarm when unit shuts down
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm does not stop after Mute/Cancel pressed
Operator presses Mute/Cancel button but alarm continues sounding instead of stopping
Ranked causes
- 1. Mute/Cancel button not functioning (unknown)
Audible alarm status after Mute/Cancel pressed during External Power Fail alarm
- Test point
- Mute/Cancel push-button control
- Expected
- Audible alarm stops
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm repeating 4 to 5 pulse pattern caution
Ventilator emitting repeating 4 to 5 pulse audible alarm
Ranked causes
- 1. Unknown alarm condition (unknown)
More symptoms
- Backlight does not respond to ambient light
- Backlight does not respond to hand proximity sensor
- Backup Ventilator flow below minimum
- Backup ventilator flow insufficient
- Backup ventilator flow rate below specification
- Backup ventilator flow rate insufficient
- Backup ventilator flow verification
- Backup ventilator I time verification
- Backup ventilator inspiratory time incorrect
- Backup ventilator inspiratory time out of specification
- Backup Ventilator manual trigger does not function correctly
- Backup ventilator manual trigger not functioning
- Backup ventilator PIP relief pressure incorrect
- Backup Ventilator PIP relief pressure out of specification
- Backup ventilator rate incorrect
- Backup ventilator rate out of specification
- Backup ventilator rate verification
- Battery charger voltage output
- Battery fail alarm not functioning
- Battery fail condition not functioning
- Battery Fail condition test
- Battery shutdown voltage out of range
- Battery voltage below shutdown threshold
- Battery voltage measurement
- Battery voltage out of range
- Battery voltage out of specification
- Battery voltage too low
- Calibration abort when Mode switch turned away during CAL mode
- Calibration: Airway pressure baseline
- Calibration: Autocal valve check
- Calibration: Backlight Threshold
- Calibration: CAL mode verification
- Calibration: External Air OFF LCD indicator
- Calibration: Mode control and autocal valve
- Calibration: Real Time Clock Reset
- Calibration: Tidal volume alarm limits
- Calibration: Unit pressure baseline
- Cancel button does not stop External Power Fail alarm
- Cancel function does not clear External Power Fail alarm
- Canceling alarm clears all indicators
- Charge LED does not light
- Charger output voltage out of range
- Charger voltage measurement
- Charger voltage out of range
- Charger voltage out of specification
- Charger voltage output incorrect
- Comp alarm does not occur when compressor operation initiated
- Comp alarm during compressor operation
- Comp alarm occurs during compressor start
- Comp alarm occurs during External Air switching
- Comp alarm occurs with low external air pressure
- Compressor alarm (Comp) occurs
- Compressor fails to start
- Control breath generation at given settings
- Control breath not generated at set parameters
- Disconnect alarm does not cancel on circuit reconnection
- Disconnect alarm does not cancel when circuit reconnected
- Disconnect alarm does not deactivate after fix
- Disconnect alarm does not occur on patient circuit disconnection
- Disconnect alarm generated
- Excessive Negative Pressure alarm (Failure Code 6)
- Exhaust Valve Failure alarm (Failure Code 5)
- Exhaust valve opens and stays open
- Extended Non-Use alarm cannot be verified during final test
- External Power Fail alarm does not cancel
- External Power Fail alarm does not cancel after mute
- External Power Fail alarm does not cancel after Mute/Cancel button press
- External power fail alarm does not cancel properly
- External Power Fail alarm does not cancel when Mute pressed
- External power low alarm
- External Power Low alarm does not activate
- External Power Low alarm does not occur
- External Power Low alarm does not trigger
- External Power Low alarm does not trigger at correct voltage
- External power low alarm does not trigger at expected voltage
- External Power Low alarm does not trigger at threshold
- External power low alarm occurs
- External power low alarm threshold
- External power low alarm threshold incorrect
- External power shutdown
- External power shutdown threshold
- External power shutdown threshold not met
- External power shutdown voltage out of range
- Failure Code 2 - No Gas COMP alarm
- Failure Code 2 (No Gas, COMP) alarm does not occur
- Failure Code 2 (No Gas, Comp) alarm during low external gas pressure
- Failure Code 2 (No Gas, Comp) alarm generation
- Failure Code 2 displayed
- Failure Code 2 occurs during gas supply test
- Failure Code 3 (Excessive Airway Pressure at startup)
- Failure Code 3 (Excessive Airway Pressure) alarm
- Failure Code 3 displayed
- Failure Code 4 (Memory check failure)
- Failure Code 4 (Memory check failure) alarm
- Failure Code 4 displayed
- Failure Code 5 (Airway pressure exceeds 40 cmH2O for 4 seconds)
- Failure Code 5 (Excessive Airway Pressure) safety shutdown does not trigger
- Failure Code 5 (Exhaust Valve Failure) alarm
- Failure Code 5 (Safety Shutdown - High Airway Pressure) alarm
- Failure Code 5 alarm triggered
- Failure Code 5 displayed
- Failure Code 5 displayed with safety shutdown
- Failure Code 5 safety shutdown at high airway pressure
- Failure Code 5 triggered
- Failure Code 5 triggered with high airway pressure
- Failure Code 5 with safety shutdown
- Failure Code 5: Excessive airway pressure safety shutdown
- Failure Code 5: Exhaust Valve Failure
- Failure Code 6 (Excessive Negative Pressure) alarm
- Failure Code 6 alarm triggered
- Failure Code 6 displayed
- Failure Code 6 triggered
- Failure Code 6: Excessive Negative Pressure
- Failure Code 7 occurs during operation
- FIO2 alarm does not occur at low oxygen pressure
- FIO2 alarm does not occur at low oxygen tank pressure
- FIO2 alarm occurs
- FIO2 alarm occurs at low oxygen pressure
- FIO2 alarm occurs at low oxygen tank pressure
- FIO2 alarm occurs at low tank pressure
- Flow in Backup Ventilator test
- Gas flow does not stop at 40 cmH2O pressure
- Gas flow does not stop at pressure relief threshold
- I time in Backup Ventilator test
- LCD screen blanks
- LCD screen does not blank with both fuses removed
- Low battery alarm does not activate at correct threshold
- Low battery alarm does not trigger
- Low battery alarm does not trigger at correct voltage
- Low battery alarm not triggering
- Low Battery Alarm occurs
- Low Battery Alarm occurs at incorrect voltage
- Low battery alarm threshold out of specification
- Low Battery Alarm threshold test
- Low Battery Alarm triggers
- Manual breath beep not heard when button held during inhalation
- Manual Breath button does not generate breath
- Manual breath button does not produce audible beep
- Manual Breath button does not trigger breath
- Manual Breath button generates no audible beep
- Manual Breath button held during inhalation does not queue breath after cycle
- Manual Breath button holds breath when pressed during inhalation
- Manual Breath button triggers during inhalation or exhalation
- Manual Breath button unresponsive
- Manual Breath button unresponsive during inhalation or exhalation
- Manual Breath button unresponsive when held during inhalation
- Manual Breath control does not generate breath
- Manual breath delivery during setup
- Manual breath does not generate audible beep
- Manual Breath does not generate control breath
- Manual breath does not trigger
- Manual breath does not trigger during inspiration or expiration
- Manual Breath function inactive during inspiration or exhalation
- Manual Breath function not triggering
- Manual Breath function unresponsive when button held during inhalation
- Manual breath hold not functioning
- Manual breath not generated
- Manual breath not triggered
- Manual breath push-button control
- Manual breath triggered during inhalation or exhalation hangs or continues indefinitely
- Manual trigger does not control gas flow properly
- Manual trigger does not deliver gas flow
- Manual trigger function verification
- Manual trigger gas flow does not respond to pressure buildup
- Manual Trigger in Backup Ventilator test
- Mode switch to A/C
- Mode switch to CAL abort test
- Mode switch to CAL with successful calibration
- Mode switch to CPAP
- Mode switch to OFF
- Mode switch to SIMV
- Mode switch: A/C mode
- Mode switch: CAL mode abort
- Mode switch: CAL mode calibration
- Mode switch: CPAP mode
- Mode switch: OFF mode
- Mode switch: SIMV mode
- Mute button does not silence alarm for 30 seconds
- Mute/Cancel button does not silence alarm message
- Mute/Cancel function suppresses alarm audible and LED
- Muting alarm for 30 seconds
- No Gas alarm (Failure Code 2) does not trigger at low external gas pressure
- O2 Low/Fail alarm does not occur at low oxygen tank pressure
- O2 Low/Fail alarm does not occur at low tank pressure
- O2 Low/Fail alarm occurs
- O2 Low/Fail alarm occurs at low oxygen pressure
- O2 Low/Fail alarm occurs at low oxygen tank pressure
- O2 Low/Fail alarm occurs at low tank pressure
- O2 Low/Fail alarm triggers at low tank pressure
- PIC Line 1 does not display EXT PWR LOW message
- PIP relief does not occur at correct pressure
- PIP relief in Backup Ventilator test
- PIP relief not occurring at correct pressure
- PIP relief pressure verification
- Preventative Maintenance alarm cannot be verified during final test
- Rate in Backup Ventilator test
- Rate, Inspiration Time, Tidal Volume controls non-functional in CPAP mode
- System does not shut down at correct external power threshold
- System does not shut down at low external power
- System does not shut down at secondary threshold with external power low
- System fail LED comes on
- System Failure LED and alarm activate incorrectly
- System failure LED does not activate on shutdown
- System failure LED does not come on during shutdown
- System Failure LED does not come on solid with both fuses removed
- System failure LED does not illuminate at secondary shutdown
- System failure LED does not illuminate during low battery
- System Failure LED does not illuminate on dual fuse failure
- System failure LED does not illuminate on shutdown
- System failure LED should come on during shutdown
- System Failure LED solid on
- System shuts down during external power test
- Tidal volume not maintained during low oxygen pressure
- Total Flow Backup alarm after compressor inlet occlusion
- Total Flow Backup alarm does not occur after compressor occlusion
- Total Flow Backup alarm during low compressor inlet flow
- Total Flow Backup Alarm not occurring
- Total Flow Backup alarm not triggered
- Total Flow Backup alarm occurs
- Total Flow Backup alarm occurs during compressor inlet occlusion
- Total Flow Backup alarm occurs with compressor inlet occlusion
- Unit does not shut down after gas supply disconnected
- Unit does not shut down at -10 cmH2O airway pressure
- Unit does not shut down at +100 cmH2O airway pressure
- Unit does not shut down at expected low voltage
- Unit does not shut down at expected voltage
- Unit does not shut down at External Power Low shutdown threshold
- Unit does not shut down at low external power voltage
- Unit does not shut down at negative airway pressure
- Unit does not shut down at second low voltage threshold
- Unit does not shut down on dual fuse failure
- Unit does not shut down with both fuses removed
- Unit should shut down at excessive negative pressure
- Unit should shut down at high airway pressure
- Ventilator failure code 2 (No Gas, COMP alarm)
- Verify tidal volume maintained after oxygen concentration change to 21%
- Verify tidal volume maintained during oxygen pressure low condition
- Verify tidal volume maintained when oxygen concentration returned to 60%
- Vt alarm cascades with Vt Settings alarm on display
- Vt alarm clears at 500 ml setting
- Vt alarm displaces Settings alarm on display
- Vt alarm does not occur at high tidal volume
- Vt alarm does not occur when tidal volume further increased
- Vt alarm occurs
- Vt alarm occurs above 1150 ml
- Vt alarm occurs at 1150 ml
- Vt alarm occurs at high tidal volume setting
- Vt alarm occurs when setting increased
- Vt alarm occurs when tidal volume exceeds 1150 ml
- Vt alarm occurs when tidal volume exceeds critical limit
- Vt alarm occurs when tidal volume exceeds higher limit
- Vt alarm occurs when tidal volume exceeds secondary limit
- Vt alarm occurs when tidal volume increased further
- Vt alarm replaces Vt Settings alarm
- Vt alarm when tidal volume increased to 1150 ml
- Vt or Disconnect alarms occur unexpectedly during external air/compressor switching
- Vt Settings alarm does not occur
- Vt Settings alarm does not occur when tidal volume exceeds limits
- Vt Settings alarm does not occur when tidal volume increased
- Vt Settings alarm occurs
- Vt Settings alarm occurs above 1050 ml
- Vt Settings alarm occurs at 1050 ml
- Vt Settings alarm occurs at elevated tidal volume
- Vt Settings alarm occurs at high tidal volume
- Vt Settings alarm occurs at high tidal volume setting
- Vt Settings alarm occurs when tidal volume exceeds 1050 ml
- Vt Settings alarm occurs when tidal volume exceeds limit
- Vt Settings alarm occurs when tidal volume exceeds range
- Vt Settings alarm triggered
- Vt Settings alarm triggers on parameter change
- Vt Settings alarm when tidal volume set above 1000 ml
271 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
5 | Exhaust Valve Failure | Airway pressure reaches +100 cmH2O | error |
6 | Excessive Negative Pressure | Airway pressure drops to -10 cmH2O | error |
7 | Run-Time Transducer Calibration Failure | Transducer calibration fails during operation | error |
Audible alarm | Continuous audible alarm indicating system shutdown or critical condition | System voltage drops below shutdown threshold (8.5V battery or 9.5V external), or critical fault condition | critical |
Audible Alarm 4-5 pulse | Repeating audible alarm with 4 to 5 pulse pattern | System failure or alert condition | error |
BATTERY LOW/FAIL | Battery voltage too low or battery failure detected | Battery voltage drops below 11.1V or battery fails during low power condition | warning |
Calibration Abort Alarm | Calibration sequence was interrupted or aborted | Mode control switch is moved away from CAL mode before calibration completes | warning |
Calibration Abort Alarm | Mode switch turned away from CAL mode during calibration | User turns Mode control switch from CAL to another mode before calibration completes | warning |
CHK EXT FUSE | Check external power fuse | External Power Fail alarm with only top fuse removed and bottom fuse installed | warning |
Comp | Compressor alarm | External gas supply pressure drops below minimum threshold or compressor fails to start | warning |
Comp | Compressor failure or inability to provide gas at required pressure | Compressor startup when external air supply is insufficient or compressor malfunction | error |
Comp | Compressor alarm or failure code related to compressor operation | Compressor fails to start or operate during source switching; external gas supply drops below 35 PSI +/- 2 PSI | warning |
Comp alarm | Compressor operation alarm during mode transition or startup failure | Compressor attempts to start during external air to compressor switching | warning |
Comp alarm | Compressor failure or unavailable; gas source switched to external air | Compressor fails to start or external air supply pressure drops below 35 PSI +/- 2 PSI | warning |
Comp alarm | Compressor alarm - compressor failed to provide gas or pressure backup needed | External air supply depleted or compressor unable to maintain pressure during switch from external air to compressor operation | error |
Comp alarm | Compressor alarm condition | Compressor activation during external air operation or external gas pressure below 35 PSI ±2 PSI | warning |
Disconnect | Patient circuit disconnected or no positive airway pressure detected | Patient circuit removed from test lung or pressure rise not sensed after breath is fired | error |
Disconnect Alarm | Patient circuit disconnected from ventilator outlet | Breath initiated with no positive airway pressure rise sensed | critical |
Extended Non-Use | Alarm indicating prolonged non-operation of ventilator | Ventilator inactive for extended period | warning |
External Power Fail | External power supply has been disconnected or is not functioning | Top fuse removed or external power supply disconnected while in A/C mode | warning |
External Power Fail alarm | External AC power supply lost or disconnected | Disconnection of external power source | warning |
EXTERNAL POWER LOW | External power supply voltage has dropped below safe operating level | External power voltage drops below 10.5V - 11.1V range | warning |
Failure Code 1 | Self Check failure during startup | Self check fails during system startup | critical |
Failure Code 2 | No Gas, COMP | Gas supplies (external air and oxygen) are disconnected | error |
Failure Code 2 | No gas from compressor during alarm condition | External gas supply pressure drops below 35 PSI +/- 2 PSI while compressor is expected to provide backup | error |
Failure Code 2 (No Gas, Comp) | No gas supply available to compressor; compressor cannot operate | External gas pressure drops below 35 PSI +/- 2 PSI during compressor operation attempt | critical |
Failure Code 3 | Excessive airway pressure detected during startup | Airway pressure exceeds safe limits at system startup | critical |
Failure Code 4 | Memory check failure alarm | Memory check fails during startup Self Check | critical |
Failure Code 5 | Exhaust Valve Failure | Airway pressure exceeds +100 cmH2O for 2 seconds without unit shutdown | error |
Failure Code 5 | Excessive airway pressure detected - safety shutdown triggered | Airway pressure exceeds 40 cmH2O for 4 consecutive seconds during operation | critical |
Failure Code 5 | Safety shutdown triggered by excessive airway pressure | Airway pressure exceeds 40 cmH2O for 4 seconds or longer | critical |
Failure Code 6 | Excessive Negative Pressure | Airway pressure drops to -10 cmH2O or below for 1.5 seconds without unit shutdown | error |
Failure Code 7 | Run-Time Transducer Calibration Failure | Transducer calibration check fails during operation | error |
FIO2 | Insufficient oxygen flow to achieve requested FIO2 concentration | Oxygen tank pressure below setpoint when FIO2 setting is 60% or higher | warning |
FIO2 | Fractional inspired oxygen out of specification | Oxygen concentration setting and supply pressure mismatch or oxygen delivery failure | critical |
FIO2 | Oxygen concentration cannot be delivered at requested level | Oxygen supply pressure insufficient to meet target FIO2 setting | critical |
FIO2 alarm | Actual oxygen concentration cannot match the set FIO2 value | Oxygen tank pressure is insufficient to deliver the set oxygen concentration percentage | warning |
FIO2 alarm | Insufficient oxygen pressure for requested oxygen concentration | Oxygen tank pressure below 45 PSI during high oxygen concentration setting (60%) | error |
LOW (Battery Icon on PIC Line 2) | Battery voltage low indication on display | Battery voltage drops into low range during operation | warning |
O2 Low/Fail | Oxygen supply pressure is below acceptable threshold | Oxygen tank pressure drops below 35 PSI +/- 2 PSI | critical |
O2 Low/Fail | Oxygen supply pressure below threshold | Oxygen tank pressure drops to 35 PSI +/- 2 PSI or below | warning |
O2 Low/Fail alarm | Oxygen tank pressure is below minimum threshold | Oxygen tank pressure drops below 35 PSI while ventilator is operating | critical |
Preventative Maintenance | Alarm indicating scheduled maintenance is due | Maintenance interval elapsed or operational hours exceeded | info |
System Failure | Critical system failure; both external power and internal power sources have failed | Both top and bottom fuses removed simultaneously | critical |
System Failure | Critical system failure requiring unit shutdown | Both top and bottom fuses removed or both power sources failed | critical |
System Failure | Multiple system fuses have failed indicating critical system failure | Both top and bottom fuses removed or open circuit | critical |
System failure LED | System failure detected, LED illuminated continuously | System voltage drops below minimum safe level (8.5V battery or 9.5V external), or critical system fault detected | error |
Total Flow Backup | Total flow backup alarm during low tidal volume condition | Actual delivered tidal volume drops below 50% of set tidal volume for multiple breaths | warning |
Total Flow Backup | Actual delivered tidal volume falls significantly below set value | Compressor inlet obstruction or gas flow restriction causes delivered volume to drop below 50% of set value after 4-5 breaths | warning |
Total Flow Backup | Backup flow alarm due to insufficient delivered tidal volume | Actual tidal volume less than half of set tidal volume for 4-5 breaths | error |
Total Flow Backup Alarm | Actual tidal volume fell below threshold during backup flow operation | Actual tidal volume drops to less than half of the set tidal volume for 4-5 breaths | warning |
Total Flow Backup Alarm | Actual delivered tidal volume is significantly less than set value indicating breathing circuit obstruction or compressor failure | Actual tidal volume measured less than half of set tidal volume for 4-5 breaths | warning |
Total Flow Backup alarm | Total flow backup threshold exceeded | Actual delivered tidal volume falls below 50% of set tidal volume for 4-5 breaths | warning |
Total Flow Backup alarm | Compressor output flow insufficient; actual tidal volume below minimum threshold | Actual delivered tidal volume falls below half the set tidal volume for 4-5 breath cycles | error |
Vt | Measured tidal volume exceeds safe operating limit | Tidal volume set to 1150 ml or higher | error |
Vt | Delivered tidal volume is out of specification | Vt setting increased to 1150 ml; exceeds critical threshold | error |
Vt | Tidal volume setting significantly exceeds safe maximum limit | Tidal volume setting increased to 1150 ml or higher | error |
Vt alarm | Measured tidal volume has deviated significantly from set value | Tidal volume setting is increased above approximately 1100 ml or actual delivered volume exceeds limit | error |
Vt Alarm | Tidal volume out of acceptable range | Delivered tidal volume deviates from set value during normal operation | warning |
Vt alarm | Tidal volume alarm | Delivered tidal volume deviates from set value | warning |
Vt Settings | Tidal volume setting outside acceptable operating range | Tidal volume setting increased above approximately 1025 ml | warning |
Vt Settings | Tidal volume setting has been changed from baseline configuration | Vt changed from 800 ml to 1050 ml or other non-baseline value | warning |
Vt Settings alarm | Tidal volume setting is outside acceptable operating range | Tidal volume setting is increased above approximately 1000 ml | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI