Impact Instrumentation Uni-Vent 754: troubleshooting and error codes
The ServiceForge knowledge base holds 347 documented symptoms for the Impact Instrumentation Uni-Vent 754, with 362 ranked causes, 730 diagnostic checks with test points and expected values, and 80 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Audible alarm comes on continuously patient safety
Continuous audible alarm activation during ventilator operation or failure conditions
Ranked causes
- 1. Failure Code 5: Excessive airway pressure above 40 cmH2O for 4 seconds (common)
- 2. Failure Code 2: No gas supply (COMP alarm) (occasional)
- 3. Failure Code 1: Self check failure during startup (rare)
- 4. Failure Code 3: Excessive airway pressure during startup (rare)
- 5. Failure Code 4: Memory check failure (rare)
Verify airway pressure level
- Test point
- Airway pressure transducer
- Expected
- Should be below 40 cmH2O
- Tools
- Pressure gauge or syringe manometer
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm mute does not restore after 30 seconds patient safety
After pressing Mute/Cancel, alarm remains silent or LED remains solid beyond 30-second period
Ranked causes
- 1. Alarm condition not resolved before mute timeout (common)
- 2. Mute/Cancel button not functioning (rare)
Verify alarm condition exists and measure time from Mute/Cancel button press
- Test point
- Alarm LED and audible alarm output
- Expected
- LED goes from blinking to solid; alarm sound stops for 30 seconds; both resume after 30 seconds
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Charge LED does not illuminate
LED fails to light after external power supply connected and battery disconnected
Ranked causes
- 1. External power supply not connected or faulty (common)
- 2. Dummy load not in place (occasional)
Verify external power supply connected
- Test point
- External power connector
- Expected
- Power supply plugged in and active
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Comp alarm occurs with external air switchover patient safety
Compressor alarm triggered during switching between external air and compressor operation, occurring simultaneously with possible Vt and Disconnect alarms.
Ranked causes
- 1. Compressor fails to start during switchover from external air (occasional)
- 2. External gas supply pressure below 35 PSI +/- 2 PSI (occasional)
Verify external gas supply pressure
- Test point
- External air inlet pressure gauge
- Expected
- >=35 PSI +/- 2 PSI, +/- 2 PSI
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
External Power Fail alarm does not clear when Mute/Cancel pressed patient safety
After pressing Mute/Cancel during External Power Fail alarm, alarm LED remains on or audible alarm continues
Ranked causes
- 1. External power not restored (common)
- 2. Mute/Cancel button malfunction (occasional)
Verify external power connection status
- Test point
- Main power input connector
- Expected
- External power connected and supplying voltage
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Failure Code 2 alarm verification patient safety
Verification test for COMP (No Gas) alarm functionality during gas supply loss
Ranked causes
- 1. Loss of external air supply (common)
- 2. Loss of oxygen supply (common)
Establish COMP alarm condition
- Test point
- Gas supply inlet connections
- Expected
- COMP alarm triggered when supplies are disconnected
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Manual Breath button does not deliver breath patient safety
Pressing Manual Breath button does not produce audible beep or generate control breath at specified settings
Ranked causes
- 1. Manual Breath button pressed at wrong time in breath cycle (common)
- 2. Manual Breath button malfunction or circuit failure (occasional)
Press Manual Breath button during expiratory pause
- Test point
- Manual Breath push-button control
- Expected
- Audible beep heard and control breath generated
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Manual Breath generates no control breath patient safety
Pressing Manual Breath button during expiratory pause produces beep but no breath delivery from unit
Ranked causes
- 1. Hose circuit not properly connected to test lung (common)
- 2. Incorrect mode or rate settings (occasional)
Verify hose circuit connected to test lung
- Test point
- Hose circuit inlet
- Expected
- Secure connection with no leaks visible
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System Failure LED and audible alarm do not activate caution
LED does not come on solid and continuous audible tone not heard when both fuses are removed
Ranked causes
- 1. Bottom fuse not removed or not fully seated (common)
- 2. Top fuse not removed (occasional)
Verify top fuse removed
- Test point
- Top fuse location
- Expected
- Fuse absent
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System Failure LED does not come on or LCD screen does not blank caution
LED fails to illuminate solid or screen does not go blank after both fuses are removed and alarm muted.
Ranked causes
- 1. Audible alarm not giving continuous tone (occasional)
- 2. Unit not shutting down after fuse removal (occasional)
Observe System Failure LED state
- Test point
- System Failure LED indicator
- Expected
- LED comes on solid
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System Failure LED does not come on solid caution
LED fails to illuminate when both top and bottom fuses are removed with unit in AC mode.
Ranked causes
- 1. Top fuse not removed before bottom fuse removal (common)
- 2. Bottom fuse not removed after top fuse removal (common)
Verify unit is in A/C mode with external power and battery connected
- Test point
- Unit mode indicator
- Expected
- A/C mode displayed
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Total Flow Backup Alarm does not occur patient safety
During partial compression inlet occlusion test with tidal volume less than half set volume, alarm should trigger after 4-5 breaths but does not
Ranked causes
- 1. Compressor inlet partially occluded incorrectly or insufficient occlusion (common)
- 2. Alarm detection circuit malfunction (rare)
Set ventilator parameters to test values
- Test point
- Control panel settings
- Expected
- Rate=15 BPM, I time=1.0 sec, Vt=900 ml, FIO2=21%, External Air=OFF
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air/Oxygen Mixer control inoperative patient safety
Air/Oxygen Mixer setting cannot be changed or does not affect delivery mixture
Ranked causes
- 1. Air/Oxygen Mixer control or LCD display failure (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm cannot be muted or alarm LED/audible does not turn off
Mute/Cancel button should silence LED and audible for Vt Settings alarm while message remains
Ranked causes
- 1. Mute/Cancel button malfunction (rare)
Press Mute/Cancel button
- Test point
- Mute/Cancel push-button
- Expected
- Button responds to press
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm deactivation after disconnect cleared
When disconnect condition is fixed, alarm should fully deactivate including LED and audible alert
Ranked causes
- 1. Disconnect condition resolved (common)
Restore disconnect condition
- Test point
- Circuit connection
- Expected
- Hose/circuit reconnected
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED and audible should turn off after Mute/Cancel
LED and audible alert silence when Mute/Cancel button is pressed while alarm is active
Ranked causes
- 1. Normal alarm mute function operation (common)
Activate alarm and press Mute/Cancel
- Test point
- Mute/Cancel push-button
- Expected
- LED and audible turn off; AMC message remains
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED behavior during mute/cancel
LED should transition from blinking to solid on for 30 seconds, then return to blinking after mute is released or timeout occurs
Ranked causes
- 1. Mute/Cancel button pressed while alarm active (common)
Verify alarm LED transitions to solid on when Mute/Cancel pressed
- Test point
- Alarm LED indicator
- Expected
- LED changes from blinking to solid on
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED blinking and solid response
Alarm LED should blink when a Disconnect alarm is generated, then turn solid after Mute/Cancel button is pressed.
Ranked causes
- 1. Disconnect alarm condition detected (common)
Generate Disconnect alarm condition
- Test point
- Patient circuit connection
- Expected
- Alarm LED blinking
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED does not blink on Disconnect alarm
LED fails to respond when Disconnect alarm is generated
Ranked causes
- 1. Alarm not generated or patient circuit properly connected (common)
Verify patient circuit disconnected from test lung
- Test point
- Patient circuit connection
- Expected
- Circuit completely disconnected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED does not come on solid when muted
LED fails to remain solid after Mute/Cancel button is pressed during disconnect alarm.
Ranked causes
- 1. Mute/Cancel button not functioning (occasional)
Press Mute/Cancel push-button during active alarm
- Test point
- Mute/Cancel button
- Expected
- Alarm LED transitions to solid state
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED does not remain solid after Mute/Cancel pressed
Alarm LED fails to change from blinking to solid state after Mute/Cancel button is pressed during Disconnect alarm
Ranked causes
- 1. Mute/Cancel push-button not functioning (occasional)
Press Mute/Cancel push-button during active Disconnect alarm
- Test point
- Mute/Cancel push-button
- Expected
- Alarm LED transitions from blinking to solid
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED does not respond to disconnect condition patient safety
Alarm LED fails to blink when disconnect alarm is generated or does not come on solid when Mute/Cancel is pressed
Ranked causes
- 1. Patient circuit disconnected from test lung (common)
Remove patient circuit from test lung
- Test point
- Patient circuit connection
- Expected
- Circuit disconnected
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED does not transition between states during mute
Alarm LED should transition from blinking to solid on when Mute/Cancel is pressed, remain solid for 30 seconds, then return to blinking
Ranked causes
- 1. Alarm LED control logic malfunction (unknown)
Observe Alarm LED during mute activation
- Test point
- Alarm LED
- Expected
- Transitions from blinking to solid on
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED does not transition to solid state when alarm muted
Alarm LED does not change from blinking to solid on state when Mute/Cancel button is pressed
Ranked causes
- 1. Alarm LED control circuit malfunction (unknown)
Verify alarm LED state during mute cycle
- Test point
- Alarm LED indicator
- Expected
- LED should transition from blinking to solid on for 30 seconds, then resume blinking
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED does not turn off after Cancel button during External Power Fail
LED remains lit or blinking after pressing Cancel button during External Power Fail alarm
Ranked causes
- 1. LED control circuit not responding to Cancel command (unknown)
Verify alarm LED turns off after Cancel button press
- Test point
- Alarm LED indicator
- Expected
- LED turns completely off
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED does not turn off after disconnect condition fixed patient safety
Alarm LED remains on after patient circuit is reconnected and disconnect condition is resolved
Ranked causes
- 1. Disconnect condition not fully resolved (common)
Reconnect patient circuit to test lung
- Test point
- Patient circuit connection
- Expected
- Positive airway pressure rise is sensed on next breath
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED does not turn off after disconnect is corrected
LED remains on after patient circuit is reconnected to test lung.
Ranked causes
- 1. Patient circuit not properly connected or breath not fired (common)
Reconnect patient circuit to test lung
- Test point
- Patient circuit connection
- Expected
- Circuit fully connected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED does not turn off when External Power Fail alarm canceled
Alarm LED remains on or does not turn off when Mute/Cancel button is pressed during External Power Fail alarm
Ranked causes
- 1. LED control circuit does not respond to cancel command (unknown)
Verify alarm LED turns off when cancel pressed
- Test point
- Alarm LED indicator
- Expected
- LED should turn off
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm Message Center continues displaying alarm after mute
Alarm Message Center still shows Disconnect alarm when Mute/Cancel button is pressed during muting operation
Ranked causes
- 1. Alarm mute function behavior as designed (common)
Confirm Alarm Message Center display persists during mute
- Test point
- Alarm Message Center display
- Expected
- Should still display the Disconnect alarm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm Message Center display during mute patient safety
AMC should continue displaying the active alarm condition even while audible alarm is muted
Ranked causes
- 1. Mute function active while alarm condition persists (common)
Verify AMC displays alarm message while mute active
- Test point
- Alarm Message Center display
- Expected
- Disconnect alarm message displayed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm Message Center does not clear after cancel
Alarm Message Center displays alarm message even after cancel function is activated during External Power Fail condition
Ranked causes
- 1. Display control circuit does not clear alarm display on cancel (unknown)
Verify Alarm Message Center clears on cancel
- Test point
- Alarm Message Center display
- Expected
- Should show no alarm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm Message Center does not display Disconnect alarm during Mute patient safety
AMC display is blank or shows different alarm while Disconnect condition is active and muted
Ranked causes
- 1. AMC display circuit not updating correctly (unknown)
Verify Alarm Message Center displays Disconnect alarm while muted
- Test point
- AMC display
- Expected
- Disconnect alarm message visible during 30 second mute period
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm Message Center does not display during mute
When alarm is muted using Mute/Cancel button, the Alarm Message Center should continue displaying the alarm condition
Ranked causes
- 1. Alarm Message Center display not updating during mute condition (unknown)
Observe Alarm Message Center display during mute period
- Test point
- Alarm Message Center display
- Expected
- Disconnect alarm message displayed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm mute/cancel not responding
Mute/Cancel button should silence audible alarm for 30 seconds and toggle alarm LED from blinking to solid, or fully cancel alarm and turn off alarm LED
Ranked causes
- 1. Mute/Cancel button not functional or electrical issue with alarm circuit (common)
Generate Disconnect alarm and press Mute/Cancel button
- Test point
- Mute/Cancel push-button control
- Expected
- Audible alarm mutes for 30 seconds; LED changes from blinking to solid
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarms cancel on Vt adjustment patient safety
Alarms clear when tidal volume setting decreased to 500 ml
Ranked causes
- 1. Tidal volume setting corrected to within acceptable range (common)
Verify alarms cleared
- Test point
- Ventilator display and audio
- Expected
- All alarms cancel
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarms cancel when oxygen concentration set to 21% patient safety
O2 Low/Fail and FIO2 alarms clear when FIO2 set to room air level
Ranked causes
- 1. Low oxygen pressure alarm requires higher FIO2 setting (common)
Verify alarms cancel
- Test point
- Alarm display and audible
- Expected
- All alarms cleared
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarms cancel when tidal volume decreased to 500 ml patient safety
All alarms clear when tidal volume setting is decreased to 500 ml from elevated levels
Ranked causes
- 1. Tidal volume setting 500 ml below alarm thresholds (common)
Decrease tidal volume to 500 ml and verify alarms cancel
- Test point
- Tidal volume setting
- Expected
- All alarms cancel ml
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarms cancel when tidal volume reduced patient safety
All tidal volume-related alarms cancel when Vt setting is decreased to 500 ml
Ranked causes
- 1. Tidal volume within normal operating range (common)
Decrease tidal volume to 500 ml
- Test point
- Vt control
- Expected
- 500 ml
Alarms cancel when tidal volume returns to normal patient safety
All alarms clear when Vt setting is reduced to 500 ml
Ranked causes
- 1. Tidal volume within safe limits (common)
Alarm cancellation at low Vt
- Test point
- alarm status
- Expected
- All alarms cancel ml
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarms cancel when tidal volume setting is decreased patient safety
All alarms clear when tidal volume is decreased to 500 ml from elevated setting
Ranked causes
- 1. Tidal volume setting returned to acceptable range (common)
Decrease tidal volume setting
- Test point
- Vt control
- Expected
- 500 ml
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarms cancel when Vt decreased to 500 ml patient safety
All tidal volume alarms clear when setting returned to safe range
Ranked causes
- 1. Tidal volume setting outside safe operating range (common)
Verify tidal volume alarms clear
- Test point
- Alarm display
- Expected
- All alarms cancelled
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarms cancel when Vt setting reduced patient safety
Alarms are canceled when tidal volume setting is decreased to 500 ml, which is within acceptable range
Ranked causes
- 1. Tidal volume setting decreased to 500 ml (within acceptable range) (common)
Verify alarms cancel at Vt setting of 500 ml
- Test point
- Tidal volume setting control
- Expected
- Alarms should cancel
Alarms do not cancel when oxygen concentration set to 21% patient safety
Alarms should clear when FIO2 is lowered to 21%
Ranked causes
- 1. Oxygen concentration at 21% with low tank pressure (common)
Set oxygen concentration
- Test point
- FIO2 control
- Expected
- 21%
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarms do not cancel when oxygen tank pressure returned to normal patient safety
Alarms should clear when oxygen tank pressure is restored to 45 PSI +/- 2 PSI
Ranked causes
- 1. Oxygen tank pressure restored to normal operating level (common)
Restore oxygen tank pressure
- Test point
- Oxygen tank pressure
- Expected
- 45 PSI +/- 2 PSI, +/- 2 PSI
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarms do not cancel when tidal volume reduced to safe level patient safety
When tidal volume decreased to 500 ml, all alarms should clear
Ranked causes
- 1. Alarm cancellation logic malfunction for tidal volume (unknown)
Decrease tidal volume to safe level
- Test point
- Vt control
- Expected
- 500 ml
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarms do not cancel when Vt decreased to 500 ml patient safety
All alarms should clear when tidal volume is reduced to 500 ml
Ranked causes
- 1. Tidal volume within normal range at 500 ml (common)
Decrease tidal volume setting
- Test point
- Vt control
- Expected
- 500 ml
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarms do not recur when oxygen concentration returned to 60% patient safety
O2 Low/Fail and FIO2 alarms should re-trigger when FIO2 is increased to 60% with tank pressure at 35 PSI
Ranked causes
- 1. Oxygen concentration increased to 60% with low tank pressure (common)
Increase oxygen concentration to 60%
- Test point
- FIO2 control
- Expected
- 60%
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AMC message does not remain displayed during test procedures caution
AMC indicator message fails to stay on display after Mute/Cancel button is pressed
Ranked causes
- 1. Display circuit or AMC indicator logic malfunction (unknown)
Observe AMC message persistence after alarm cancellation
- Test point
- LCD display
- Expected
- AMC message remains visible after Mute/Cancel button press
Audible alarm does not give continuous tone when both fuses removed
Audible alarm fails to produce continuous tone when both fuses are removed from the unit
Ranked causes
- 1. Both fuses not removed (common)
Verify both fuses removed
- Test point
- Top and bottom fuse locations
- Expected
- Both fuses out
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm does not mute when Mute/Cancel button pressed
Pressing Mute/Cancel button does not silence the audible alarm during a Disconnect alarm condition
Ranked causes
- 1. Mute/Cancel button not functioning (unknown)
Verify Mute/Cancel button silences audible alarm
- Test point
- Mute/Cancel push-button control
- Expected
- Audible alarm should mute for 30 seconds
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm does not sound at low voltage
Audible alarm should sound continuously when external power voltage drops below shutdown threshold (9.5V–10.5V)
Ranked causes
- 1. External power voltage drops below 9.5V–10.5V shutdown threshold (common)
Lower voltage to shutdown range and verify audible alarm
- Test point
- External power jack, audio output
- Expected
- Audible alarm sounds continuously at 9.50V–10.50V, –10.50V
- Tools
- Variable power source, DMM
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm does not sound continuously on shutdown
When system voltage drops below threshold and unit shuts down, audible alarm fails to come on continuously as expected.
Ranked causes
- 1. Low voltage causes system shutdown without alarm activation (unknown)
Verify system shutdown voltage range
- Test point
- External power jack input
- Expected
- 8.5V - 9.3V, 8.5–9.3V
- Tools
- Variable power source, DMM
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm does not sound continuously with both fuses removed caution
Continuous tone audible alarm fails to sound when both top and bottom fuses are removed
Ranked causes
- 1. Both fuses not removed (common)
Confirm both top and bottom fuses are removed
- Test point
- Fuse holders (top and bottom)
- Expected
- Both fuses absent
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm does not stop after mute
User presses Mute/Cancel button but alarm continues to sound; alarm LED behavior and message center state vary depending on alarm type
Ranked causes
- 1. External Power Fail alarm not properly canceled (common)
Verify alarm type and mute button response
- Test point
- Mute/Cancel push-button control
- Expected
- Audible alarm stops and alarm LED turns off
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm mute function
Audible alarm should mute for 30 seconds when Mute/Cancel button pressed, then resume after timeout
Ranked causes
- 1. Mute/Cancel button activated during active alarm (common)
Verify audible alarm mutes when Mute/Cancel pressed
- Test point
- Speaker/audible output
- Expected
- Alarm sound ceases
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm mutes temporarily instead of stopping
After pressing Mute/Cancel, alarm silences for 30 seconds then resumes; alarm LED transitions from blinking to solid for 30 seconds then blinks again
Ranked causes
- 1. Disconnect alarm pending; mute function working as designed (common)
Generate Disconnect alarm and press Mute/Cancel
- Test point
- Mute/Cancel push-button control
- Expected
- Audible alarm mutes for 30 seconds; alarm LED transitions from blinking to solid for 30 seconds
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm not producing expected pulse pattern
Audible alarm does not produce the expected 4 to 5 pulse repeating alarm pattern when expected to sound.
Ranked causes
- 1. Audible alarm circuit or speaker defective (unknown)
Backlight does not respond to ambient light
Backlight threshold adjustment needed; backlight should turn off in darkness and on when sensor is covered
Ranked causes
- 1. Backlight threshold requires calibration (common)
Enter User Program Mode
- Test point
- LCD menu
- Expected
- Menu appears at top of LCD
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Backlight does not respond to sensor
Backlight does not turn off when sensor is uncovered or turn on when covered, indicating backlight sensor or threshold calibration issue
Ranked causes
- 1. Backlight threshold not properly calibrated (common)
Enter User Program Mode
- Test point
- LCD menu
- Expected
- Menu appears along top of LCD
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Backlight sensor calibration
Backlight threshold must be set so the display backlight responds to ambient light levels appropriately.
Ranked causes
- 1. Backlight threshold not calibrated to user environment (unknown)
Enter User Program Mode
- Test point
- Unit controls
- Expected
- Menu appears along top of LCD
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Backup ventilation flow rate below specification
- Backup ventilation inspiratory time out of specification
- Backup ventilation rate out of specification
- Backup ventilator flow below specification
- Backup ventilator flow insufficient
- Backup ventilator flow rate out of tolerance
- Backup ventilator gas flow insufficient
- Backup ventilator I time incorrect
- Backup ventilator I time out of specification
- Backup Ventilator Inspiratory Time out of specification
- Backup ventilator inspiratory time out of tolerance
- Backup Ventilator Manual Trigger not responding
- Backup Ventilator PIP Relief not functioning correctly
- Backup ventilator PIP relief out of tolerance
- Backup ventilator rate incorrect
- Backup ventilator rate out of specification
- Backup ventilator rate out of tolerance
- Battery charging voltage out of specification
- Battery Fail alarm does not trigger
- Battery Fail condition not triggering or delayed
- Battery low alarm threshold out of tolerance
- Battery low shutdown threshold out of tolerance
- Battery voltage out of acceptable range
- Battery voltage out of range
- Battery voltage out of specification
- Battery voltage out of tolerance
- Calibration abort alarm triggered
- Calibration: Airway pressure baseline
- Calibration: LCD CAL mode verification
- Calibration: Manual breath control
- Calibration: Mode control and LCD verification
- Calibration: Mode control and system response
- Calibration: Mode verification
- Calibration: Real Time Clock Reset
- Calibration: Unit calibration
- Cancel alarm function does not stop audible alarm
- Cancel button does not clear External Power Fail alarm
- Cancel button does not stop External Power Fail alarm
- Cancel External Power Fail alarm
- Cancel function does not clear External Power Fail alarm
- Charge LED does not light
- Charge LED indicator setup and operation
- Charger voltage measurement
- Charger voltage out of range at battery connector
- Charger voltage out of specification
- Charger voltage out of tolerance
- Charger voltage output
- Charger voltage output incorrect
- Charger voltage output out of specification
- Comp alarm does not occur
- Comp alarm does not occur during compressor failure test
- Comp alarm does not occur during compressor switching test
- Comp alarm does not occur with external air and compressor switching
- Comp alarm generation during external air to compressor switching
- Comp alarm occurrence during external air switching
- Comp alarm occurs
- Comp alarm occurs during compressor operation
- Compressor alarm occurs unexpectedly
- Compressor fails to attempt start after external air button press
- Compressor fails to start
- CPAP apnea trigger time out of tolerance
- CPAP backup rate out of tolerance
- CPAP flow rate out of tolerance
- CPAP inspiratory time out of tolerance
- CPAP PIP relief pressure out of tolerance
- Disconnect alarm does not deactivate after condition is fixed
- Disconnect alarm on patient circuit removal
- Excessive airway pressure during startup
- Excessive negative airway pressure not detected
- Exhaust valve opens and stays open
- External Power Fail alarm cancel function
- External Power Fail alarm cannot be canceled
- External Power Fail alarm does not cancel
- External Power Fail alarm does not cancel after Mute/Cancel pressed
- External Power Fail alarm does not cancel properly
- External Power Fail alarm does not clear when external power restored
- External Power Fail alarm does not occur
- External Power Fail alarm does not occur when top fuse removed
- External Power Fail alarm occurs when top fuse removed
- External power fail shutdown threshold out of tolerance
- External Power Low alarm does not occur at expected voltage
- External Power Low alarm does not trigger
- External power low alarm does not trigger at correct voltage
- External Power Low alarm does not trigger at expected voltage
- External Power Low alarm does not trigger at threshold voltage
- External Power Low alarm not displayed
- External power low alarm not triggered at correct voltage
- External power low alarm occurs
- External Power Low alarm occurs at incorrect voltage
- External power low alarm threshold
- External power low alarm threshold out of tolerance
- Failure Code 1 occurred
- Failure Code 2 (No Gas, Comp) alarm not generated
- Failure Code 2 displayed
- Failure Code 2 displayed (No Gas, COMP alarm)
- Failure Code 2 displays (No Gas, COMP alarm)
- Failure Code 2 triggered
- Failure Code 2: No Gas (COMP alarm)
- Failure Code 2: No gas alarm (COMP)
- Failure Code 3
- Failure Code 3 (Excessive Airway Pressure)
- Failure Code 3 alarm
- Failure Code 3 displayed
- Failure Code 3 occurred
- Failure Code 3 occurs (Excessive Airway Pressure at startup)
- Failure Code 4
- Failure Code 4 (Memory Check Failure)
- Failure Code 4 alarm
- Failure Code 4 displayed
- Failure Code 4 occurred
- Failure Code 4 occurs (Memory check fails)
- Failure Code 5 (Airway Pressure Threshold Exceeded)
- Failure Code 5 (Exhaust Valve Failure) triggered
- Failure Code 5 alarm (Exhaust Valve Failure)
- Failure Code 5 alarm appears
- Failure Code 5 alarm on high airway pressure
- Failure Code 5 alarm with safety shutdown
- Failure Code 5 displayed
- Failure Code 5 occurs (Airway pressure exceeds 40 cmH2O for 4 seconds)
- Failure Code 5 triggered (Exhaust Valve Failure)
- Failure Code 5 triggered with safety shutdown
- Failure Code 5 with safety shutdown
- Failure Code 5: Excessive airway pressure shutdown
- Failure Code 6 (Excessive Negative Pressure) triggered
- Failure Code 6 alarm appears
- Failure Code 6 alarm on excessive negative pressure
- Failure Code 6 displayed
- Failure Code 6 triggered (Excessive Negative Pressure)
- FIO2 alarm does not occur with low oxygen pressure
- FIO2 alarm during low oxygen tank pressure
- FIO2 alarm occurs
- FIO2 alarm occurs at low oxygen pressure
- FIO2 alarm occurs at low oxygen pressure with high concentration setting
- FIO2 alarm occurs at low oxygen tank pressure
- FIO2 alarm occurs during low oxygen supply
- FIO2 alarm triggers at low oxygen pressure
- FiO2 reading out of tolerance
- Inspiratory time out of tolerance (range verification)
- Inspiratory/exhalation indicator malfunction
- LCD indicator does not read OFF after external air button press
- LCD screen does not blank when both fuses removed
- Low Battery Alarm
- Low battery alarm condition
- Low Battery Alarm does not occur at expected voltage
- Low Battery Alarm does not trigger
- Low battery alarm not activating at correct voltage
- Low Battery Alarm occurs at incorrect voltage threshold
- Low battery alarm threshold incorrect
- Low Battery Alarm triggered
- Low Battery Alarm triggers at incorrect voltage
- Manual Breath button activates during inhalation or exhalation
- Manual Breath button does not function
- Manual Breath button does not generate breath
- Manual Breath button does not generate controlled breath
- Manual breath button during expiratory pause
- Manual Breath button generates unwanted breath during inhalation or exhalation
- Manual Breath button held during inhalation does not function until baseline pressure
- Manual Breath button held during inhalation produces no effect until current breath completes
- Manual breath button hold during inhalation
- Manual breath button inoperative
- Manual breath button press during inhalation or exhalation
- Manual Breath button produces no audible beep
- Manual Breath button produces no effect during inhalation or exhalation
- Manual Breath control does not generate controlled breath
- Manual Breath control does not produce audible beep
- Manual Breath control generates breath at wrong time in cycle
- Manual Breath during held inhalation produces no beep after breath ends
- Manual Breath function does not generate audible beep
- Manual Breath function does not respond
- Manual Breath generates audible beep
- Manual Breath generates unwanted breath during inhalation or exhalation
- Manual breath not triggered
- Manual Breath push-button control verification
- Manual Trigger does not deliver gas flow
- Manual trigger does not function
- Manual trigger does not stop gas flow at pressure limit
- Manual trigger not functioning in backup ventilator
- Mode control switch to A/C
- Mode control switch to CPAP
- Mode control switch to OFF
- Mode control switch to SIMV
- Multiple alarms display priority with newest first
- Mute button does not silence audible alarm for 30 seconds
- Mute/Cancel button does not mute alarm
- Mute/Cancel button does not silence alarm
- Mute/Cancel button does not silence alarm LED and audible alarm
- Mute/Cancel button mutes alarm for 30 seconds
- Mute/Cancel does not silence alarm LED and audible tone
- Mute/Cancel function does not silence alarm
- Mute/Cancel function does not stop audible alarm
- No Gas (Comp) alarm - Failure Code 2
- No Gas alarm (Failure Code 2)
- O2 Low/Fail alarm does not occur at low oxygen pressure
- O2 Low/Fail alarm does not occur at low tank pressure
- O2 Low/Fail alarm during low oxygen tank pressure
- O2 Low/Fail alarm occurs
- O2 Low/Fail alarm occurs at low oxygen pressure
- O2 Low/Fail alarm occurs at low oxygen tank pressure
- O2 Low/Fail alarm occurs at low tank pressure
- O2 Low/Fail alarm triggers at low tank pressure
- Oxygen tank pressure below threshold
- Peak airway pressure display malfunction
- PEEP value out of tolerance
- PIC line 1 does not display CHK EXT FUSE after Mute pressed
- PIP relief does not occur at correct pressure
- PIP relief not occurring at correct pressure
- Positive inspiratory pressure relief setting out of specification
- Rate out of tolerance (A/C and SIMV modes)
- Rate out of tolerance (range verification)
- Rate, Inspiration Time, and Tidal Volume controls non-functional in CPAP mode
- Real time clock reset procedure
- RTC battery voltage low
- Safety shutdown does not occur at excessive airway pressure
- Safety shutdown not activating at pressure threshold
- Safety shutdown triggered by excessive airway pressure
- Safety shutdown with sustained high airway pressure
- Software version verification failure
- Spontaneous flow rate out of tolerance (SIMV/CPAP)
- System does not shut down at low external power voltage
- System does not shut down at low external voltage
- System fail LED comes on
- System Failure LED and alarm when both fuses removed
- System Failure LED does not activate when both fuses removed
- System failure LED does not come on continuously
- System failure LED does not illuminate
- System Failure LED does not illuminate when bottom fuse removed
- System Failure LED illuminated solid
- System Failure LED indicator not working
- System shutdown does not occur at expected voltage
- System shutdown during battery low voltage test
- System shutdown during low external power
- System shutdown occurs at incorrect external power voltage
- System shutdown occurs at low external voltage
- System shuts down at low battery voltage
- System shuts down at low external power voltage
- Tidal volume alarm at 1150 ml setting
- Tidal volume alarm cancellation at reduced setting
- Tidal volume alarm occurs when set above 1050 ml
- Tidal Volume control inoperative
- Tidal volume not maintained at low oxygen pressure
- Tidal volume not maintained during low oxygen pressure
- Tidal volume out of tolerance
- Tidal volume setting exceeds alarm threshold
- Total Flow Backup alarm does not occur during outlet occlusion test
- Total Flow Backup alarm does not occur with compressor inlet occlusion
- Total Flow Backup alarm does not occur with restricted flow
- Total Flow Backup alarm during low compressor inlet pressure
- Total Flow Backup alarm generation
- Total Flow Backup alarm occurs
- Unit does not give Failure Code 2 alarm
- Unit does not shut down after gas supply loss
- Unit does not shut down at +100 cmH2O airway pressure
- Unit does not shut down at excessive negative airway pressure
- Unit does not shut down at excessive negative pressure
- Unit does not shut down at expected voltage
- Unit does not shut down at external power low
- Unit does not shut down at low battery voltage
- Unit does not shut down when both fuses removed
- Unit shuts down after airway pressure decreases to -10 cmH2O
- Unit shuts down after airway pressure reaches +100 cmH2O
- Unit shuts down with Failure Code 2 alarm
- Unit shuts down with Failure Code 5 alarm
- Unit shuts down with Failure Code 6 alarm
- Vt alarm does not occur or LED does not reactivate at elevated tidal volume
- Vt alarm does not occur when Vt increased to 1150 ml
- Vt alarm occurs
- Vt alarm occurs above 1150 ml
- Vt alarm occurs at 1150 ml setting
- Vt alarm occurs at 1150 ml with LED and audible active
- Vt alarm occurs at high tidal volume setting
- Vt alarm occurs at very high tidal volume
- Vt alarm occurs when set above 1150 ml
- Vt alarm occurs when tidal volume exceeds 1150 ml
- Vt alarm occurs when tidal volume exceeds critical threshold
- Vt alarm occurs when tidal volume exceeds high threshold
- Vt alarm occurs when tidal volume exceeds limit
- Vt alarm occurs when tidal volume setting is increased further
- Vt Settings alarm does not occur at 1050 ml
- Vt Settings alarm does not occur when tidal volume exceeds threshold
- Vt Settings alarm occurs
- Vt Settings alarm occurs above 1050 ml
- Vt Settings alarm occurs at 1050 ml
- Vt Settings alarm occurs at 1050 ml setting
- Vt Settings alarm occurs at high tidal volume
- Vt Settings alarm occurs when tidal volume exceeds 1050 ml
- Vt Settings alarm occurs when tidal volume exceeds limit
- Vt Settings alarm should occur
287 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
1 | Transducer Cal Fail | Transducer calibration failure detected during startup or operation | error |
1 (Transducer Cal Fail) | Transducer calibration failed during power-up self-test | Unit power-up or software calibration attempted | error |
2 | No Gas, Comp | No gas supply detected to compressor or gas supply line disconnected | critical |
2 (No Gas, Comp) | No gas supply to compressor or compressor failure | Gas supply lost during operation | critical |
3 | Total Flow Backup / Excessive Airway Pressure | Flow cannot be delivered or airway pressure exceeds safe limits | critical |
3 (Total Flow Backup) / (Excessive Airway Pressure) | Ver1.83X: Total Flow Backup triggered; Ver1.85X: Excessive airway pressure detected | Flow exceeds limits or pressure exceeds safety threshold | error |
4 | Excessive Airway Pressure / Memory Check | Airway pressure exceeds 40 cmH2O or memory integrity failure | error |
4 (Excessive Airway Pressure) / (Memory Check) | Ver1.83X: Excessive airway pressure; Ver1.85X: Memory (EEPROM) check failed | Pressure exceeds safety limit or EEPROM integrity check failed | error |
5 | Memory check failure or exhalation valve failure | Internal memory test fails or exhalation valve malfunction detected | error |
5 (Memory Check) / (Exhalation Valve Failure) | Ver1.83X: Memory check failed; Ver1.85X: Exhalation valve does not open properly | EEPROM error or exhalation valve malfunction | error |
6 | Exhaust Valve Fail / Excessive Negative Pressure | Exhaust valve failure or excessive negative pressure in circuit | error |
6 (Exhaust Valve Fail) / (Excessive Negative Pressure) | Ver1.83X: Exhaust valve failure; Ver1.85X: Negative pressure exceeds safe level | Exhaust valve malfunction or patient breathing generates excessive negative pressure | error |
7 | Excessive Negative Pressure / Run-Time Cal Fail | Negative pressure exceeds limits or runtime calibration fails | error |
7 (Excessive Negative Pressure) / (Run-Time Cal Fail) | Ver1.83X: Excessive negative pressure; Ver1.85X: Run-time calibration failed | Negative pressure spike or runtime calibration procedure failure | error |
Audible alarm 4-5 pulse | Repeating 4 to 5 pulse alarm indicating system status | Audible alarm gives a repeating 4 to 5 pulse alarm | warning |
Audible alarm 4-5 pulse | System failure alarm | System failure condition detected | critical |
BATTERY LOW/FAIL | Battery voltage is critically low or battery has failed | Battery voltage drops below acceptable threshold or battery is disconnected | warning |
BATTERY LOW/FAIL | Low battery voltage detected; unit operating on battery below safe voltage threshold. | Battery voltage falls to 11.1V - 11.7V range. | warning |
BATTERY LOW/FAIL | Low battery condition detected | Battery voltage drops below 11.1V - 11.7V range | warning |
Calibration Abort Alarm | Calibration was interrupted or not completed successfully | Mode control switched away from CAL mode before calibration completes | warning |
CHK EXT FUSE | Check external power supply fuse | Displayed on PIC line 1 after External Power Fail alarm and Mute/Cancel press | warning |
Code 1 | Transducer calibration failure (firmware version 1.83X) or Transducer Cal Fail | Pressure transducer returns invalid or out-of-range values during sensor diagnostics; cannot be calibrated to expected values | error |
Code 2 | No gas or compressor failure (applicable to both firmware versions 1.83X and 1.85X) | Unit detects no gas supply pressure or compressor output failure during system initialization or operation | critical |
Code 3 | Total flow backup failure or Excessive airway pressure (varies by version: backup in 1.83X, pressure in 1.85X) | Version 1.83X: Total flow measurement circuit detects excessive or unmeasurable flow during backup mode. Version 1.85X: Airway pressure exceeds safety threshold (typically 50+ cmH2O) outside normal range. | error |
Code 4 | Excessive airway pressure (firmware version 1.83X) or Memory check failure (firmware version 1.85X) | Version 1.83X: Airway pressure exceeds safe limit (typically 50+ cmH2O) outside normal breath delivery window. Version 1.85X: EEPROM memory checksum or data integrity error detected during boot or operation. | error |
Code 5 | Memory check failure (firmware version 1.83X) or Exhalation valve failure (firmware version 1.85X) | Version 1.83X: EEPROM checksum or RAM integrity error detected. Version 1.85X: Exhalation valve does not open or close properly; stuck closed or mechanical binding detected. | error |
Code 6 | Exhaust valve failure (firmware version 1.83X) or Excessive negative pressure (firmware version 1.85X) | Version 1.83X: Exhaust valve does not vent during safety shutdown or pressure relief cycles; may not open when commanded. Version 1.85X: Inspiratory effort (negative airway pressure) exceeds safety limit during spontaneous or demand breathing (typically < -25 cmH2O). | warning |
Code 7 | Excessive negative pressure (firmware version 1.83X) or Run-time calibration failure (firmware version 1.85X) | Version 1.83X: Inspiratory effort (negative airway pressure) exceeds safe limit during spontaneous or demand breathing (<-25 cmH2O typical). Version 1.85X: On-device automatic calibration routine fails to converge or produce valid calibration values. | warning |
Comp | Compressor alarm | Compressor fails to start or external air pressure drops below threshold during operation | warning |
Comp alarm | Compressor alarm indicating compressor failure or inability to start | Switching from external air to compressor operation; external gas pressure below 35 PSI +/- 2 PSI | error |
Comp alarm | Compressor alarm | Compressor fails to start or cannot maintain operation when External Air is toggled off | warning |
Comp alarm | Compressor failure alarm indicating compressor cannot supply sufficient gas pressure | Compressor fails to provide adequate pressure during operation; external air valves take over gas supply during Comp alarm condition | warning |
Comp Alarm (Failure Code 2) | Compressor failure with No Gas condition | External gas pressure drops below 35 PSI +/- 2 PSI during compressor operation | error |
Disconnect | Patient circuit has become disconnected from the ventilator | Patient circuit is removed from test lung and next breath is fired with no positive airway pressure rise sensed | critical |
Disconnect | Disconnection detection alarm | Patient circuit becomes disconnected from ventilator outlet | warning |
Disconnect alarm | Alarm indicating loss of patient circuit connection or gas delivery path interruption | During external air and compressor switching test or when circuit is disconnected | warning |
Disconnect alarm | Circuit or patient disconnection detected | Hose circuit disconnected from test lung or patient | warning |
Disconnect alarm | Disconnect alarm | Loss of gas delivery or patient disconnect detected | warning |
Disconnect alarm | Patient circuit or breathing hose disconnected from patient or ventilator | Loss of airway pressure or circuit continuity | error |
External Power Fail | External power supply has failed or been interrupted | Top fuse is removed while in AC mode with battery connected | warning |
External Power Fail | Loss or failure of external power supply | Top fuse is removed or external power supply is disconnected while in A/C mode with both external power and battery connected | warning |
External Power Fail alarm | External electrical power disconnected or lost | External AC power supply unplugged or power loss | critical |
EXTERNAL POWER LOW | External power supply voltage has dropped below safe operating threshold | External power voltage drops below 10.5V - 11.1V range | warning |
Failure Code 1 | Self Check failure during startup | Self-check fails when unit powers on | critical |
Failure Code 2 | No Gas, COMP | Self Check fails during startup or gas supply is disconnected | error |
Failure Code 2 | No Gas, Comp alarm - external gas supply failure during compressor operation | External gas pressure drops below 35 PSI when compressor is active | critical |
Failure Code 2 (No Gas, Comp) | No gas available to compressor; backup to external air activated | External gas supply pressure falls below 35 PSI +/- 2 PSI during compressor operation attempt | error |
Failure Code 3 | Excessive airway pressure | Airway pressure exceeds safe limits during startup | critical |
Failure Code 4 | Memory self-check failure | Memory integrity check fails during startup | critical |
Failure Code 5 | Safety shutdown triggered due to excessive airway pressure exceeding 40 cmH2O for 4 seconds | Airway pressure sustained above 40 cmH2O for 4 seconds or more | critical |
Failure Code 5 | Excessive airway pressure above 40 cmH2O with safety shutdown | Airway pressure rises above 40 cmH2O and remains there for 4 seconds or more | critical |
Failure Code 5 | Ventilator Failure Code 5 | Pressure exceeds 40 cmH2O for 4 seconds | critical |
Failure Code 5 | Exhaust Valve Failure | Airway pressure reaches +100 cmH2O or above | error |
Failure Code 5 | Safety shutdown activated | Airway pressure exceeds 40 cmH2O for 4 seconds | critical |
Failure Code 6 | Excessive Negative Pressure | Airway pressure falls to -10 cmH2O or below | error |
Failure Code 7 | Run-Time Transducer Calibration Failure | Transducer calibration fails during operation | error |
FIO2 | Oxygen concentration setting cannot be maintained due to low oxygen supply pressure | FIO2 setting at 60% or higher with oxygen tank pressure below 35 PSI +/- 2 PSI | critical |
FIO2 | Fraction of inspired oxygen alarm; oxygen delivery out of specification | Oxygen tank pressure drops below 35 PSI +/- 2 PSI when high FIO2 set | critical |
FIO2 | Oxygen concentration delivery failure | Oxygen tank pressure insufficient to achieve set FIO2; pressure below 45 PSI at high concentration settings | critical |
FIO2 | FIO2 (inspired oxygen fraction) sensor detects insufficient oxygen | Oxygen tank pressure below threshold while FIO2 is set above 21% | warning |
O2 Low/Fail | Oxygen tank pressure too low or oxygen supply failed | Oxygen tank pressure drops below 45 PSI threshold at 60% FIO2 | critical |
System failure | System has shut down due to insufficient external power voltage | External power voltage drops below 9.50V - 10.50V range | critical |
System Failure | Both external power and battery power lost or fuses removed | Top and bottom fuses both removed | critical |
System Failure | Critical system failure detected | Bottom fuse is removed while top fuse is already removed | critical |
System Failure LED | System failure condition detected | Internal system fault or critical component failure | critical |
Total Flow Backup | Flow backup alarm - inadequate gas delivery detected | Delivered tidal volume falls below 50% of set tidal volume for 4-5 consecutive breaths | error |
Total Flow Backup | Total flow backup alarm | Actual delivered tidal volume drops below 50% of set tidal volume for 4+ consecutive breaths | warning |
Total Flow Backup | Alarm indicating insufficient flow delivery due to compressor inlet obstruction | Actual tidal volume less than half of set tidal volume for 4-5 consecutive breaths | warning |
Total Flow Backup Alarm | Insufficient gas flow backup when primary flow path is compromised | Actual delivered tidal volume drops below 50% of set tidal volume for 4-5 breaths | warning |
Total Flow Backup alarm | Actual delivered tidal volume insufficient | Actual tidal volume drops below half of set tidal volume after 4-5 breaths | warning |
Total Flow Backup alarm | Tidal volume insufficient due to compressor inlet restriction | Actual delivered tidal volume falls below half of set tidal volume for 4-5 consecutive breaths | warning |
Total Flow Backup Alarm | Alarm indicating total flow output is below expected tidal volume threshold | Compressor inlet occlusion or flow restriction reduces actual tidal volume to less than half of set tidal volume for 4-5 breaths | warning |
Vt | Actual tidal volume delivered differs significantly from setting | Delivered tidal volume diverges from set value during ventilation | error |
Vt | Tidal volume delivery is outside acceptable parameters | Tidal volume setting increased beyond alarm threshold (1150 ml or greater) | error |
Vt | Measured or set tidal volume exceeds alarm limit | Tidal volume setting increased to 1150 ml | error |
Vt | Measured tidal volume out of specification or range exceeded | Tidal volume setting set to 1150 ml or higher | error |
Vt | Tidal volume critically high | Vt set to 1150 ml or higher | error |
Vt alarm | Tidal Volume alarm | Actual delivered tidal volume deviates from set tidal volume | warning |
Vt alarm | Tidal volume alarm indicating delivered tidal volume does not meet set value | During external air and compressor switching test or when delivery is insufficient | warning |
Vt Settings | Tidal volume setting is outside configured parameters | Tidal volume setting changed to 1050 ml or greater | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI