Impact Instrumentation Uni-Vent 754: troubleshooting and error codes

The ServiceForge knowledge base holds 347 documented symptoms for the Impact Instrumentation Uni-Vent 754, with 362 ranked causes, 730 diagnostic checks with test points and expected values, and 80 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Audible alarm comes on continuously patient safety

Continuous audible alarm activation during ventilator operation or failure conditions

Ranked causes

  1. 1. Failure Code 5: Excessive airway pressure above 40 cmH2O for 4 seconds (common)
  2. 2. Failure Code 2: No gas supply (COMP alarm) (occasional)
  3. 3. Failure Code 1: Self check failure during startup (rare)
  4. 4. Failure Code 3: Excessive airway pressure during startup (rare)
  5. 5. Failure Code 4: Memory check failure (rare)
First check

Verify airway pressure level

Test point
Airway pressure transducer
Expected
Should be below 40 cmH2O
Tools
Pressure gauge or syringe manometer

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm mute does not restore after 30 seconds patient safety

After pressing Mute/Cancel, alarm remains silent or LED remains solid beyond 30-second period

Ranked causes

  1. 1. Alarm condition not resolved before mute timeout (common)
  2. 2. Mute/Cancel button not functioning (rare)
First check

Verify alarm condition exists and measure time from Mute/Cancel button press

Test point
Alarm LED and audible alarm output
Expected
LED goes from blinking to solid; alarm sound stops for 30 seconds; both resume after 30 seconds

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Charge LED does not illuminate

LED fails to light after external power supply connected and battery disconnected

Ranked causes

  1. 1. External power supply not connected or faulty (common)
  2. 2. Dummy load not in place (occasional)
First check

Verify external power supply connected

Test point
External power connector
Expected
Power supply plugged in and active

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Comp alarm occurs with external air switchover patient safety

Compressor alarm triggered during switching between external air and compressor operation, occurring simultaneously with possible Vt and Disconnect alarms.

Ranked causes

  1. 1. Compressor fails to start during switchover from external air (occasional)
  2. 2. External gas supply pressure below 35 PSI +/- 2 PSI (occasional)
First check

Verify external gas supply pressure

Test point
External air inlet pressure gauge
Expected
>=35 PSI +/- 2 PSI, +/- 2 PSI

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

External Power Fail alarm does not clear when Mute/Cancel pressed patient safety

After pressing Mute/Cancel during External Power Fail alarm, alarm LED remains on or audible alarm continues

Ranked causes

  1. 1. External power not restored (common)
  2. 2. Mute/Cancel button malfunction (occasional)
First check

Verify external power connection status

Test point
Main power input connector
Expected
External power connected and supplying voltage

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Failure Code 2 alarm verification patient safety

Verification test for COMP (No Gas) alarm functionality during gas supply loss

Ranked causes

  1. 1. Loss of external air supply (common)
  2. 2. Loss of oxygen supply (common)
First check

Establish COMP alarm condition

Test point
Gas supply inlet connections
Expected
COMP alarm triggered when supplies are disconnected

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Manual Breath button does not deliver breath patient safety

Pressing Manual Breath button does not produce audible beep or generate control breath at specified settings

Ranked causes

  1. 1. Manual Breath button pressed at wrong time in breath cycle (common)
  2. 2. Manual Breath button malfunction or circuit failure (occasional)
First check

Press Manual Breath button during expiratory pause

Test point
Manual Breath push-button control
Expected
Audible beep heard and control breath generated

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Manual Breath generates no control breath patient safety

Pressing Manual Breath button during expiratory pause produces beep but no breath delivery from unit

Ranked causes

  1. 1. Hose circuit not properly connected to test lung (common)
  2. 2. Incorrect mode or rate settings (occasional)
First check

Verify hose circuit connected to test lung

Test point
Hose circuit inlet
Expected
Secure connection with no leaks visible

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System Failure LED and audible alarm do not activate caution

LED does not come on solid and continuous audible tone not heard when both fuses are removed

Ranked causes

  1. 1. Bottom fuse not removed or not fully seated (common)
  2. 2. Top fuse not removed (occasional)
First check

Verify top fuse removed

Test point
Top fuse location
Expected
Fuse absent

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System Failure LED does not come on or LCD screen does not blank caution

LED fails to illuminate solid or screen does not go blank after both fuses are removed and alarm muted.

Ranked causes

  1. 1. Audible alarm not giving continuous tone (occasional)
  2. 2. Unit not shutting down after fuse removal (occasional)
First check

Observe System Failure LED state

Test point
System Failure LED indicator
Expected
LED comes on solid

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System Failure LED does not come on solid caution

LED fails to illuminate when both top and bottom fuses are removed with unit in AC mode.

Ranked causes

  1. 1. Top fuse not removed before bottom fuse removal (common)
  2. 2. Bottom fuse not removed after top fuse removal (common)
First check

Verify unit is in A/C mode with external power and battery connected

Test point
Unit mode indicator
Expected
A/C mode displayed

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Total Flow Backup Alarm does not occur patient safety

During partial compression inlet occlusion test with tidal volume less than half set volume, alarm should trigger after 4-5 breaths but does not

Ranked causes

  1. 1. Compressor inlet partially occluded incorrectly or insufficient occlusion (common)
  2. 2. Alarm detection circuit malfunction (rare)
First check

Set ventilator parameters to test values

Test point
Control panel settings
Expected
Rate=15 BPM, I time=1.0 sec, Vt=900 ml, FIO2=21%, External Air=OFF

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air/Oxygen Mixer control inoperative patient safety

Air/Oxygen Mixer setting cannot be changed or does not affect delivery mixture

Ranked causes

  1. 1. Air/Oxygen Mixer control or LCD display failure (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm cannot be muted or alarm LED/audible does not turn off

Mute/Cancel button should silence LED and audible for Vt Settings alarm while message remains

Ranked causes

  1. 1. Mute/Cancel button malfunction (rare)
First check

Press Mute/Cancel button

Test point
Mute/Cancel push-button
Expected
Button responds to press

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm deactivation after disconnect cleared

When disconnect condition is fixed, alarm should fully deactivate including LED and audible alert

Ranked causes

  1. 1. Disconnect condition resolved (common)
First check

Restore disconnect condition

Test point
Circuit connection
Expected
Hose/circuit reconnected

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LED and audible should turn off after Mute/Cancel

LED and audible alert silence when Mute/Cancel button is pressed while alarm is active

Ranked causes

  1. 1. Normal alarm mute function operation (common)
First check

Activate alarm and press Mute/Cancel

Test point
Mute/Cancel push-button
Expected
LED and audible turn off; AMC message remains

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LED behavior during mute/cancel

LED should transition from blinking to solid on for 30 seconds, then return to blinking after mute is released or timeout occurs

Ranked causes

  1. 1. Mute/Cancel button pressed while alarm active (common)
First check

Verify alarm LED transitions to solid on when Mute/Cancel pressed

Test point
Alarm LED indicator
Expected
LED changes from blinking to solid on

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LED blinking and solid response

Alarm LED should blink when a Disconnect alarm is generated, then turn solid after Mute/Cancel button is pressed.

Ranked causes

  1. 1. Disconnect alarm condition detected (common)
First check

Generate Disconnect alarm condition

Test point
Patient circuit connection
Expected
Alarm LED blinking

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LED does not blink on Disconnect alarm

LED fails to respond when Disconnect alarm is generated

Ranked causes

  1. 1. Alarm not generated or patient circuit properly connected (common)
First check

Verify patient circuit disconnected from test lung

Test point
Patient circuit connection
Expected
Circuit completely disconnected

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LED does not come on solid when muted

LED fails to remain solid after Mute/Cancel button is pressed during disconnect alarm.

Ranked causes

  1. 1. Mute/Cancel button not functioning (occasional)
First check

Press Mute/Cancel push-button during active alarm

Test point
Mute/Cancel button
Expected
Alarm LED transitions to solid state

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LED does not remain solid after Mute/Cancel pressed

Alarm LED fails to change from blinking to solid state after Mute/Cancel button is pressed during Disconnect alarm

Ranked causes

  1. 1. Mute/Cancel push-button not functioning (occasional)
First check

Press Mute/Cancel push-button during active Disconnect alarm

Test point
Mute/Cancel push-button
Expected
Alarm LED transitions from blinking to solid

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LED does not respond to disconnect condition patient safety

Alarm LED fails to blink when disconnect alarm is generated or does not come on solid when Mute/Cancel is pressed

Ranked causes

  1. 1. Patient circuit disconnected from test lung (common)
First check

Remove patient circuit from test lung

Test point
Patient circuit connection
Expected
Circuit disconnected

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LED does not transition between states during mute

Alarm LED should transition from blinking to solid on when Mute/Cancel is pressed, remain solid for 30 seconds, then return to blinking

Ranked causes

  1. 1. Alarm LED control logic malfunction (unknown)
First check

Observe Alarm LED during mute activation

Test point
Alarm LED
Expected
Transitions from blinking to solid on

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LED does not transition to solid state when alarm muted

Alarm LED does not change from blinking to solid on state when Mute/Cancel button is pressed

Ranked causes

  1. 1. Alarm LED control circuit malfunction (unknown)
First check

Verify alarm LED state during mute cycle

Test point
Alarm LED indicator
Expected
LED should transition from blinking to solid on for 30 seconds, then resume blinking

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LED does not turn off after Cancel button during External Power Fail

LED remains lit or blinking after pressing Cancel button during External Power Fail alarm

Ranked causes

  1. 1. LED control circuit not responding to Cancel command (unknown)
First check

Verify alarm LED turns off after Cancel button press

Test point
Alarm LED indicator
Expected
LED turns completely off

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LED does not turn off after disconnect condition fixed patient safety

Alarm LED remains on after patient circuit is reconnected and disconnect condition is resolved

Ranked causes

  1. 1. Disconnect condition not fully resolved (common)
First check

Reconnect patient circuit to test lung

Test point
Patient circuit connection
Expected
Positive airway pressure rise is sensed on next breath

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LED does not turn off after disconnect is corrected

LED remains on after patient circuit is reconnected to test lung.

Ranked causes

  1. 1. Patient circuit not properly connected or breath not fired (common)
First check

Reconnect patient circuit to test lung

Test point
Patient circuit connection
Expected
Circuit fully connected

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LED does not turn off when External Power Fail alarm canceled

Alarm LED remains on or does not turn off when Mute/Cancel button is pressed during External Power Fail alarm

Ranked causes

  1. 1. LED control circuit does not respond to cancel command (unknown)
First check

Verify alarm LED turns off when cancel pressed

Test point
Alarm LED indicator
Expected
LED should turn off

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm Message Center continues displaying alarm after mute

Alarm Message Center still shows Disconnect alarm when Mute/Cancel button is pressed during muting operation

Ranked causes

  1. 1. Alarm mute function behavior as designed (common)
First check

Confirm Alarm Message Center display persists during mute

Test point
Alarm Message Center display
Expected
Should still display the Disconnect alarm

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm Message Center display during mute patient safety

AMC should continue displaying the active alarm condition even while audible alarm is muted

Ranked causes

  1. 1. Mute function active while alarm condition persists (common)
First check

Verify AMC displays alarm message while mute active

Test point
Alarm Message Center display
Expected
Disconnect alarm message displayed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm Message Center does not clear after cancel

Alarm Message Center displays alarm message even after cancel function is activated during External Power Fail condition

Ranked causes

  1. 1. Display control circuit does not clear alarm display on cancel (unknown)
First check

Verify Alarm Message Center clears on cancel

Test point
Alarm Message Center display
Expected
Should show no alarm

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm Message Center does not display Disconnect alarm during Mute patient safety

AMC display is blank or shows different alarm while Disconnect condition is active and muted

Ranked causes

  1. 1. AMC display circuit not updating correctly (unknown)
First check

Verify Alarm Message Center displays Disconnect alarm while muted

Test point
AMC display
Expected
Disconnect alarm message visible during 30 second mute period

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm Message Center does not display during mute

When alarm is muted using Mute/Cancel button, the Alarm Message Center should continue displaying the alarm condition

Ranked causes

  1. 1. Alarm Message Center display not updating during mute condition (unknown)
First check

Observe Alarm Message Center display during mute period

Test point
Alarm Message Center display
Expected
Disconnect alarm message displayed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm mute/cancel not responding

Mute/Cancel button should silence audible alarm for 30 seconds and toggle alarm LED from blinking to solid, or fully cancel alarm and turn off alarm LED

Ranked causes

  1. 1. Mute/Cancel button not functional or electrical issue with alarm circuit (common)
First check

Generate Disconnect alarm and press Mute/Cancel button

Test point
Mute/Cancel push-button control
Expected
Audible alarm mutes for 30 seconds; LED changes from blinking to solid

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarms cancel on Vt adjustment patient safety

Alarms clear when tidal volume setting decreased to 500 ml

Ranked causes

  1. 1. Tidal volume setting corrected to within acceptable range (common)
First check

Verify alarms cleared

Test point
Ventilator display and audio
Expected
All alarms cancel

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarms cancel when oxygen concentration set to 21% patient safety

O2 Low/Fail and FIO2 alarms clear when FIO2 set to room air level

Ranked causes

  1. 1. Low oxygen pressure alarm requires higher FIO2 setting (common)
First check

Verify alarms cancel

Test point
Alarm display and audible
Expected
All alarms cleared

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarms cancel when tidal volume decreased to 500 ml patient safety

All alarms clear when tidal volume setting is decreased to 500 ml from elevated levels

Ranked causes

  1. 1. Tidal volume setting 500 ml below alarm thresholds (common)
First check

Decrease tidal volume to 500 ml and verify alarms cancel

Test point
Tidal volume setting
Expected
All alarms cancel ml

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarms cancel when tidal volume reduced patient safety

All tidal volume-related alarms cancel when Vt setting is decreased to 500 ml

Ranked causes

  1. 1. Tidal volume within normal operating range (common)
First check

Decrease tidal volume to 500 ml

Test point
Vt control
Expected
500 ml

Alarms cancel when tidal volume returns to normal patient safety

All alarms clear when Vt setting is reduced to 500 ml

Ranked causes

  1. 1. Tidal volume within safe limits (common)
First check

Alarm cancellation at low Vt

Test point
alarm status
Expected
All alarms cancel ml

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarms cancel when tidal volume setting is decreased patient safety

All alarms clear when tidal volume is decreased to 500 ml from elevated setting

Ranked causes

  1. 1. Tidal volume setting returned to acceptable range (common)
First check

Decrease tidal volume setting

Test point
Vt control
Expected
500 ml

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarms cancel when Vt decreased to 500 ml patient safety

All tidal volume alarms clear when setting returned to safe range

Ranked causes

  1. 1. Tidal volume setting outside safe operating range (common)
First check

Verify tidal volume alarms clear

Test point
Alarm display
Expected
All alarms cancelled

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarms cancel when Vt setting reduced patient safety

Alarms are canceled when tidal volume setting is decreased to 500 ml, which is within acceptable range

Ranked causes

  1. 1. Tidal volume setting decreased to 500 ml (within acceptable range) (common)
First check

Verify alarms cancel at Vt setting of 500 ml

Test point
Tidal volume setting control
Expected
Alarms should cancel

Alarms do not cancel when oxygen concentration set to 21% patient safety

Alarms should clear when FIO2 is lowered to 21%

Ranked causes

  1. 1. Oxygen concentration at 21% with low tank pressure (common)
First check

Set oxygen concentration

Test point
FIO2 control
Expected
21%

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarms do not cancel when oxygen tank pressure returned to normal patient safety

Alarms should clear when oxygen tank pressure is restored to 45 PSI +/- 2 PSI

Ranked causes

  1. 1. Oxygen tank pressure restored to normal operating level (common)
First check

Restore oxygen tank pressure

Test point
Oxygen tank pressure
Expected
45 PSI +/- 2 PSI, +/- 2 PSI

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarms do not cancel when tidal volume reduced to safe level patient safety

When tidal volume decreased to 500 ml, all alarms should clear

Ranked causes

  1. 1. Alarm cancellation logic malfunction for tidal volume (unknown)
First check

Decrease tidal volume to safe level

Test point
Vt control
Expected
500 ml

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarms do not cancel when Vt decreased to 500 ml patient safety

All alarms should clear when tidal volume is reduced to 500 ml

Ranked causes

  1. 1. Tidal volume within normal range at 500 ml (common)
First check

Decrease tidal volume setting

Test point
Vt control
Expected
500 ml

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarms do not recur when oxygen concentration returned to 60% patient safety

O2 Low/Fail and FIO2 alarms should re-trigger when FIO2 is increased to 60% with tank pressure at 35 PSI

Ranked causes

  1. 1. Oxygen concentration increased to 60% with low tank pressure (common)
First check

Increase oxygen concentration to 60%

Test point
FIO2 control
Expected
60%

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

AMC message does not remain displayed during test procedures caution

AMC indicator message fails to stay on display after Mute/Cancel button is pressed

Ranked causes

  1. 1. Display circuit or AMC indicator logic malfunction (unknown)
First check

Observe AMC message persistence after alarm cancellation

Test point
LCD display
Expected
AMC message remains visible after Mute/Cancel button press

Audible alarm does not give continuous tone when both fuses removed

Audible alarm fails to produce continuous tone when both fuses are removed from the unit

Ranked causes

  1. 1. Both fuses not removed (common)
First check

Verify both fuses removed

Test point
Top and bottom fuse locations
Expected
Both fuses out

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarm does not mute when Mute/Cancel button pressed

Pressing Mute/Cancel button does not silence the audible alarm during a Disconnect alarm condition

Ranked causes

  1. 1. Mute/Cancel button not functioning (unknown)
First check

Verify Mute/Cancel button silences audible alarm

Test point
Mute/Cancel push-button control
Expected
Audible alarm should mute for 30 seconds

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarm does not sound at low voltage

Audible alarm should sound continuously when external power voltage drops below shutdown threshold (9.5V–10.5V)

Ranked causes

  1. 1. External power voltage drops below 9.5V–10.5V shutdown threshold (common)
First check

Lower voltage to shutdown range and verify audible alarm

Test point
External power jack, audio output
Expected
Audible alarm sounds continuously at 9.50V–10.50V, –10.50V
Tools
Variable power source, DMM

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarm does not sound continuously on shutdown

When system voltage drops below threshold and unit shuts down, audible alarm fails to come on continuously as expected.

Ranked causes

  1. 1. Low voltage causes system shutdown without alarm activation (unknown)
First check

Verify system shutdown voltage range

Test point
External power jack input
Expected
8.5V - 9.3V, 8.5–9.3V
Tools
Variable power source, DMM

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarm does not sound continuously with both fuses removed caution

Continuous tone audible alarm fails to sound when both top and bottom fuses are removed

Ranked causes

  1. 1. Both fuses not removed (common)
First check

Confirm both top and bottom fuses are removed

Test point
Fuse holders (top and bottom)
Expected
Both fuses absent

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarm does not stop after mute

User presses Mute/Cancel button but alarm continues to sound; alarm LED behavior and message center state vary depending on alarm type

Ranked causes

  1. 1. External Power Fail alarm not properly canceled (common)
First check

Verify alarm type and mute button response

Test point
Mute/Cancel push-button control
Expected
Audible alarm stops and alarm LED turns off

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarm mute function

Audible alarm should mute for 30 seconds when Mute/Cancel button pressed, then resume after timeout

Ranked causes

  1. 1. Mute/Cancel button activated during active alarm (common)
First check

Verify audible alarm mutes when Mute/Cancel pressed

Test point
Speaker/audible output
Expected
Alarm sound ceases

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarm mutes temporarily instead of stopping

After pressing Mute/Cancel, alarm silences for 30 seconds then resumes; alarm LED transitions from blinking to solid for 30 seconds then blinks again

Ranked causes

  1. 1. Disconnect alarm pending; mute function working as designed (common)
First check

Generate Disconnect alarm and press Mute/Cancel

Test point
Mute/Cancel push-button control
Expected
Audible alarm mutes for 30 seconds; alarm LED transitions from blinking to solid for 30 seconds

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarm not producing expected pulse pattern

Audible alarm does not produce the expected 4 to 5 pulse repeating alarm pattern when expected to sound.

Ranked causes

  1. 1. Audible alarm circuit or speaker defective (unknown)

Backlight does not respond to ambient light

Backlight threshold adjustment needed; backlight should turn off in darkness and on when sensor is covered

Ranked causes

  1. 1. Backlight threshold requires calibration (common)
First check

Enter User Program Mode

Test point
LCD menu
Expected
Menu appears at top of LCD

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Backlight does not respond to sensor

Backlight does not turn off when sensor is uncovered or turn on when covered, indicating backlight sensor or threshold calibration issue

Ranked causes

  1. 1. Backlight threshold not properly calibrated (common)
First check

Enter User Program Mode

Test point
LCD menu
Expected
Menu appears along top of LCD

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Backlight sensor calibration

Backlight threshold must be set so the display backlight responds to ambient light levels appropriately.

Ranked causes

  1. 1. Backlight threshold not calibrated to user environment (unknown)
First check

Enter User Program Mode

Test point
Unit controls
Expected
Menu appears along top of LCD

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Backup ventilation flow rate below specification
  • Backup ventilation inspiratory time out of specification
  • Backup ventilation rate out of specification
  • Backup ventilator flow below specification
  • Backup ventilator flow insufficient
  • Backup ventilator flow rate out of tolerance
  • Backup ventilator gas flow insufficient
  • Backup ventilator I time incorrect
  • Backup ventilator I time out of specification
  • Backup Ventilator Inspiratory Time out of specification
  • Backup ventilator inspiratory time out of tolerance
  • Backup Ventilator Manual Trigger not responding
  • Backup Ventilator PIP Relief not functioning correctly
  • Backup ventilator PIP relief out of tolerance
  • Backup ventilator rate incorrect
  • Backup ventilator rate out of specification
  • Backup ventilator rate out of tolerance
  • Battery charging voltage out of specification
  • Battery Fail alarm does not trigger
  • Battery Fail condition not triggering or delayed
  • Battery low alarm threshold out of tolerance
  • Battery low shutdown threshold out of tolerance
  • Battery voltage out of acceptable range
  • Battery voltage out of range
  • Battery voltage out of specification
  • Battery voltage out of tolerance
  • Calibration abort alarm triggered
  • Calibration: Airway pressure baseline
  • Calibration: LCD CAL mode verification
  • Calibration: Manual breath control
  • Calibration: Mode control and LCD verification
  • Calibration: Mode control and system response
  • Calibration: Mode verification
  • Calibration: Real Time Clock Reset
  • Calibration: Unit calibration
  • Cancel alarm function does not stop audible alarm
  • Cancel button does not clear External Power Fail alarm
  • Cancel button does not stop External Power Fail alarm
  • Cancel External Power Fail alarm
  • Cancel function does not clear External Power Fail alarm
  • Charge LED does not light
  • Charge LED indicator setup and operation
  • Charger voltage measurement
  • Charger voltage out of range at battery connector
  • Charger voltage out of specification
  • Charger voltage out of tolerance
  • Charger voltage output
  • Charger voltage output incorrect
  • Charger voltage output out of specification
  • Comp alarm does not occur
  • Comp alarm does not occur during compressor failure test
  • Comp alarm does not occur during compressor switching test
  • Comp alarm does not occur with external air and compressor switching
  • Comp alarm generation during external air to compressor switching
  • Comp alarm occurrence during external air switching
  • Comp alarm occurs
  • Comp alarm occurs during compressor operation
  • Compressor alarm occurs unexpectedly
  • Compressor fails to attempt start after external air button press
  • Compressor fails to start
  • CPAP apnea trigger time out of tolerance
  • CPAP backup rate out of tolerance
  • CPAP flow rate out of tolerance
  • CPAP inspiratory time out of tolerance
  • CPAP PIP relief pressure out of tolerance
  • Disconnect alarm does not deactivate after condition is fixed
  • Disconnect alarm on patient circuit removal
  • Excessive airway pressure during startup
  • Excessive negative airway pressure not detected
  • Exhaust valve opens and stays open
  • External Power Fail alarm cancel function
  • External Power Fail alarm cannot be canceled
  • External Power Fail alarm does not cancel
  • External Power Fail alarm does not cancel after Mute/Cancel pressed
  • External Power Fail alarm does not cancel properly
  • External Power Fail alarm does not clear when external power restored
  • External Power Fail alarm does not occur
  • External Power Fail alarm does not occur when top fuse removed
  • External Power Fail alarm occurs when top fuse removed
  • External power fail shutdown threshold out of tolerance
  • External Power Low alarm does not occur at expected voltage
  • External Power Low alarm does not trigger
  • External power low alarm does not trigger at correct voltage
  • External Power Low alarm does not trigger at expected voltage
  • External Power Low alarm does not trigger at threshold voltage
  • External Power Low alarm not displayed
  • External power low alarm not triggered at correct voltage
  • External power low alarm occurs
  • External Power Low alarm occurs at incorrect voltage
  • External power low alarm threshold
  • External power low alarm threshold out of tolerance
  • Failure Code 1 occurred
  • Failure Code 2 (No Gas, Comp) alarm not generated
  • Failure Code 2 displayed
  • Failure Code 2 displayed (No Gas, COMP alarm)
  • Failure Code 2 displays (No Gas, COMP alarm)
  • Failure Code 2 triggered
  • Failure Code 2: No Gas (COMP alarm)
  • Failure Code 2: No gas alarm (COMP)
  • Failure Code 3
  • Failure Code 3 (Excessive Airway Pressure)
  • Failure Code 3 alarm
  • Failure Code 3 displayed
  • Failure Code 3 occurred
  • Failure Code 3 occurs (Excessive Airway Pressure at startup)
  • Failure Code 4
  • Failure Code 4 (Memory Check Failure)
  • Failure Code 4 alarm
  • Failure Code 4 displayed
  • Failure Code 4 occurred
  • Failure Code 4 occurs (Memory check fails)
  • Failure Code 5 (Airway Pressure Threshold Exceeded)
  • Failure Code 5 (Exhaust Valve Failure) triggered
  • Failure Code 5 alarm (Exhaust Valve Failure)
  • Failure Code 5 alarm appears
  • Failure Code 5 alarm on high airway pressure
  • Failure Code 5 alarm with safety shutdown
  • Failure Code 5 displayed
  • Failure Code 5 occurs (Airway pressure exceeds 40 cmH2O for 4 seconds)
  • Failure Code 5 triggered (Exhaust Valve Failure)
  • Failure Code 5 triggered with safety shutdown
  • Failure Code 5 with safety shutdown
  • Failure Code 5: Excessive airway pressure shutdown
  • Failure Code 6 (Excessive Negative Pressure) triggered
  • Failure Code 6 alarm appears
  • Failure Code 6 alarm on excessive negative pressure
  • Failure Code 6 displayed
  • Failure Code 6 triggered (Excessive Negative Pressure)
  • FIO2 alarm does not occur with low oxygen pressure
  • FIO2 alarm during low oxygen tank pressure
  • FIO2 alarm occurs
  • FIO2 alarm occurs at low oxygen pressure
  • FIO2 alarm occurs at low oxygen pressure with high concentration setting
  • FIO2 alarm occurs at low oxygen tank pressure
  • FIO2 alarm occurs during low oxygen supply
  • FIO2 alarm triggers at low oxygen pressure
  • FiO2 reading out of tolerance
  • Inspiratory time out of tolerance (range verification)
  • Inspiratory/exhalation indicator malfunction
  • LCD indicator does not read OFF after external air button press
  • LCD screen does not blank when both fuses removed
  • Low Battery Alarm
  • Low battery alarm condition
  • Low Battery Alarm does not occur at expected voltage
  • Low Battery Alarm does not trigger
  • Low battery alarm not activating at correct voltage
  • Low Battery Alarm occurs at incorrect voltage threshold
  • Low battery alarm threshold incorrect
  • Low Battery Alarm triggered
  • Low Battery Alarm triggers at incorrect voltage
  • Manual Breath button activates during inhalation or exhalation
  • Manual Breath button does not function
  • Manual Breath button does not generate breath
  • Manual Breath button does not generate controlled breath
  • Manual breath button during expiratory pause
  • Manual Breath button generates unwanted breath during inhalation or exhalation
  • Manual Breath button held during inhalation does not function until baseline pressure
  • Manual Breath button held during inhalation produces no effect until current breath completes
  • Manual breath button hold during inhalation
  • Manual breath button inoperative
  • Manual breath button press during inhalation or exhalation
  • Manual Breath button produces no audible beep
  • Manual Breath button produces no effect during inhalation or exhalation
  • Manual Breath control does not generate controlled breath
  • Manual Breath control does not produce audible beep
  • Manual Breath control generates breath at wrong time in cycle
  • Manual Breath during held inhalation produces no beep after breath ends
  • Manual Breath function does not generate audible beep
  • Manual Breath function does not respond
  • Manual Breath generates audible beep
  • Manual Breath generates unwanted breath during inhalation or exhalation
  • Manual breath not triggered
  • Manual Breath push-button control verification
  • Manual Trigger does not deliver gas flow
  • Manual trigger does not function
  • Manual trigger does not stop gas flow at pressure limit
  • Manual trigger not functioning in backup ventilator
  • Mode control switch to A/C
  • Mode control switch to CPAP
  • Mode control switch to OFF
  • Mode control switch to SIMV
  • Multiple alarms display priority with newest first
  • Mute button does not silence audible alarm for 30 seconds
  • Mute/Cancel button does not mute alarm
  • Mute/Cancel button does not silence alarm
  • Mute/Cancel button does not silence alarm LED and audible alarm
  • Mute/Cancel button mutes alarm for 30 seconds
  • Mute/Cancel does not silence alarm LED and audible tone
  • Mute/Cancel function does not silence alarm
  • Mute/Cancel function does not stop audible alarm
  • No Gas (Comp) alarm - Failure Code 2
  • No Gas alarm (Failure Code 2)
  • O2 Low/Fail alarm does not occur at low oxygen pressure
  • O2 Low/Fail alarm does not occur at low tank pressure
  • O2 Low/Fail alarm during low oxygen tank pressure
  • O2 Low/Fail alarm occurs
  • O2 Low/Fail alarm occurs at low oxygen pressure
  • O2 Low/Fail alarm occurs at low oxygen tank pressure
  • O2 Low/Fail alarm occurs at low tank pressure
  • O2 Low/Fail alarm triggers at low tank pressure
  • Oxygen tank pressure below threshold
  • Peak airway pressure display malfunction
  • PEEP value out of tolerance
  • PIC line 1 does not display CHK EXT FUSE after Mute pressed
  • PIP relief does not occur at correct pressure
  • PIP relief not occurring at correct pressure
  • Positive inspiratory pressure relief setting out of specification
  • Rate out of tolerance (A/C and SIMV modes)
  • Rate out of tolerance (range verification)
  • Rate, Inspiration Time, and Tidal Volume controls non-functional in CPAP mode
  • Real time clock reset procedure
  • RTC battery voltage low
  • Safety shutdown does not occur at excessive airway pressure
  • Safety shutdown not activating at pressure threshold
  • Safety shutdown triggered by excessive airway pressure
  • Safety shutdown with sustained high airway pressure
  • Software version verification failure
  • Spontaneous flow rate out of tolerance (SIMV/CPAP)
  • System does not shut down at low external power voltage
  • System does not shut down at low external voltage
  • System fail LED comes on
  • System Failure LED and alarm when both fuses removed
  • System Failure LED does not activate when both fuses removed
  • System failure LED does not come on continuously
  • System failure LED does not illuminate
  • System Failure LED does not illuminate when bottom fuse removed
  • System Failure LED illuminated solid
  • System Failure LED indicator not working
  • System shutdown does not occur at expected voltage
  • System shutdown during battery low voltage test
  • System shutdown during low external power
  • System shutdown occurs at incorrect external power voltage
  • System shutdown occurs at low external voltage
  • System shuts down at low battery voltage
  • System shuts down at low external power voltage
  • Tidal volume alarm at 1150 ml setting
  • Tidal volume alarm cancellation at reduced setting
  • Tidal volume alarm occurs when set above 1050 ml
  • Tidal Volume control inoperative
  • Tidal volume not maintained at low oxygen pressure
  • Tidal volume not maintained during low oxygen pressure
  • Tidal volume out of tolerance
  • Tidal volume setting exceeds alarm threshold
  • Total Flow Backup alarm does not occur during outlet occlusion test
  • Total Flow Backup alarm does not occur with compressor inlet occlusion
  • Total Flow Backup alarm does not occur with restricted flow
  • Total Flow Backup alarm during low compressor inlet pressure
  • Total Flow Backup alarm generation
  • Total Flow Backup alarm occurs
  • Unit does not give Failure Code 2 alarm
  • Unit does not shut down after gas supply loss
  • Unit does not shut down at +100 cmH2O airway pressure
  • Unit does not shut down at excessive negative airway pressure
  • Unit does not shut down at excessive negative pressure
  • Unit does not shut down at expected voltage
  • Unit does not shut down at external power low
  • Unit does not shut down at low battery voltage
  • Unit does not shut down when both fuses removed
  • Unit shuts down after airway pressure decreases to -10 cmH2O
  • Unit shuts down after airway pressure reaches +100 cmH2O
  • Unit shuts down with Failure Code 2 alarm
  • Unit shuts down with Failure Code 5 alarm
  • Unit shuts down with Failure Code 6 alarm
  • Vt alarm does not occur or LED does not reactivate at elevated tidal volume
  • Vt alarm does not occur when Vt increased to 1150 ml
  • Vt alarm occurs
  • Vt alarm occurs above 1150 ml
  • Vt alarm occurs at 1150 ml setting
  • Vt alarm occurs at 1150 ml with LED and audible active
  • Vt alarm occurs at high tidal volume setting
  • Vt alarm occurs at very high tidal volume
  • Vt alarm occurs when set above 1150 ml
  • Vt alarm occurs when tidal volume exceeds 1150 ml
  • Vt alarm occurs when tidal volume exceeds critical threshold
  • Vt alarm occurs when tidal volume exceeds high threshold
  • Vt alarm occurs when tidal volume exceeds limit
  • Vt alarm occurs when tidal volume setting is increased further
  • Vt Settings alarm does not occur at 1050 ml
  • Vt Settings alarm does not occur when tidal volume exceeds threshold
  • Vt Settings alarm occurs
  • Vt Settings alarm occurs above 1050 ml
  • Vt Settings alarm occurs at 1050 ml
  • Vt Settings alarm occurs at 1050 ml setting
  • Vt Settings alarm occurs at high tidal volume
  • Vt Settings alarm occurs when tidal volume exceeds 1050 ml
  • Vt Settings alarm occurs when tidal volume exceeds limit
  • Vt Settings alarm should occur

287 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
1Transducer Cal FailTransducer calibration failure detected during startup or operationerror
1 (Transducer Cal Fail)Transducer calibration failed during power-up self-testUnit power-up or software calibration attemptederror
2No Gas, CompNo gas supply detected to compressor or gas supply line disconnectedcritical
2 (No Gas, Comp)No gas supply to compressor or compressor failureGas supply lost during operationcritical
3Total Flow Backup / Excessive Airway PressureFlow cannot be delivered or airway pressure exceeds safe limitscritical
3 (Total Flow Backup) / (Excessive Airway Pressure)Ver1.83X: Total Flow Backup triggered; Ver1.85X: Excessive airway pressure detectedFlow exceeds limits or pressure exceeds safety thresholderror
4Excessive Airway Pressure / Memory CheckAirway pressure exceeds 40 cmH2O or memory integrity failureerror
4 (Excessive Airway Pressure) / (Memory Check)Ver1.83X: Excessive airway pressure; Ver1.85X: Memory (EEPROM) check failedPressure exceeds safety limit or EEPROM integrity check failederror
5Memory check failure or exhalation valve failureInternal memory test fails or exhalation valve malfunction detectederror
5 (Memory Check) / (Exhalation Valve Failure)Ver1.83X: Memory check failed; Ver1.85X: Exhalation valve does not open properlyEEPROM error or exhalation valve malfunctionerror
6Exhaust Valve Fail / Excessive Negative PressureExhaust valve failure or excessive negative pressure in circuiterror
6 (Exhaust Valve Fail) / (Excessive Negative Pressure)Ver1.83X: Exhaust valve failure; Ver1.85X: Negative pressure exceeds safe levelExhaust valve malfunction or patient breathing generates excessive negative pressureerror
7Excessive Negative Pressure / Run-Time Cal FailNegative pressure exceeds limits or runtime calibration failserror
7 (Excessive Negative Pressure) / (Run-Time Cal Fail)Ver1.83X: Excessive negative pressure; Ver1.85X: Run-time calibration failedNegative pressure spike or runtime calibration procedure failureerror
Audible alarm 4-5 pulseRepeating 4 to 5 pulse alarm indicating system statusAudible alarm gives a repeating 4 to 5 pulse alarmwarning
Audible alarm 4-5 pulseSystem failure alarmSystem failure condition detectedcritical
BATTERY LOW/FAILBattery voltage is critically low or battery has failedBattery voltage drops below acceptable threshold or battery is disconnectedwarning
BATTERY LOW/FAILLow battery voltage detected; unit operating on battery below safe voltage threshold.Battery voltage falls to 11.1V - 11.7V range.warning
BATTERY LOW/FAILLow battery condition detectedBattery voltage drops below 11.1V - 11.7V rangewarning
Calibration Abort AlarmCalibration was interrupted or not completed successfullyMode control switched away from CAL mode before calibration completeswarning
CHK EXT FUSECheck external power supply fuseDisplayed on PIC line 1 after External Power Fail alarm and Mute/Cancel presswarning
Code 1Transducer calibration failure (firmware version 1.83X) or Transducer Cal FailPressure transducer returns invalid or out-of-range values during sensor diagnostics; cannot be calibrated to expected valueserror
Code 2No gas or compressor failure (applicable to both firmware versions 1.83X and 1.85X)Unit detects no gas supply pressure or compressor output failure during system initialization or operationcritical
Code 3Total flow backup failure or Excessive airway pressure (varies by version: backup in 1.83X, pressure in 1.85X)Version 1.83X: Total flow measurement circuit detects excessive or unmeasurable flow during backup mode. Version 1.85X: Airway pressure exceeds safety threshold (typically 50+ cmH2O) outside normal range.error
Code 4Excessive airway pressure (firmware version 1.83X) or Memory check failure (firmware version 1.85X)Version 1.83X: Airway pressure exceeds safe limit (typically 50+ cmH2O) outside normal breath delivery window. Version 1.85X: EEPROM memory checksum or data integrity error detected during boot or operation.error
Code 5Memory check failure (firmware version 1.83X) or Exhalation valve failure (firmware version 1.85X)Version 1.83X: EEPROM checksum or RAM integrity error detected. Version 1.85X: Exhalation valve does not open or close properly; stuck closed or mechanical binding detected.error
Code 6Exhaust valve failure (firmware version 1.83X) or Excessive negative pressure (firmware version 1.85X)Version 1.83X: Exhaust valve does not vent during safety shutdown or pressure relief cycles; may not open when commanded. Version 1.85X: Inspiratory effort (negative airway pressure) exceeds safety limit during spontaneous or demand breathing (typically < -25 cmH2O).warning
Code 7Excessive negative pressure (firmware version 1.83X) or Run-time calibration failure (firmware version 1.85X)Version 1.83X: Inspiratory effort (negative airway pressure) exceeds safe limit during spontaneous or demand breathing (<-25 cmH2O typical). Version 1.85X: On-device automatic calibration routine fails to converge or produce valid calibration values.warning
CompCompressor alarmCompressor fails to start or external air pressure drops below threshold during operationwarning
Comp alarmCompressor alarm indicating compressor failure or inability to startSwitching from external air to compressor operation; external gas pressure below 35 PSI +/- 2 PSIerror
Comp alarmCompressor alarmCompressor fails to start or cannot maintain operation when External Air is toggled offwarning
Comp alarmCompressor failure alarm indicating compressor cannot supply sufficient gas pressureCompressor fails to provide adequate pressure during operation; external air valves take over gas supply during Comp alarm conditionwarning
Comp Alarm (Failure Code 2)Compressor failure with No Gas conditionExternal gas pressure drops below 35 PSI +/- 2 PSI during compressor operationerror
DisconnectPatient circuit has become disconnected from the ventilatorPatient circuit is removed from test lung and next breath is fired with no positive airway pressure rise sensedcritical
DisconnectDisconnection detection alarmPatient circuit becomes disconnected from ventilator outletwarning
Disconnect alarmAlarm indicating loss of patient circuit connection or gas delivery path interruptionDuring external air and compressor switching test or when circuit is disconnectedwarning
Disconnect alarmCircuit or patient disconnection detectedHose circuit disconnected from test lung or patientwarning
Disconnect alarmDisconnect alarmLoss of gas delivery or patient disconnect detectedwarning
Disconnect alarmPatient circuit or breathing hose disconnected from patient or ventilatorLoss of airway pressure or circuit continuityerror
External Power FailExternal power supply has failed or been interruptedTop fuse is removed while in AC mode with battery connectedwarning
External Power FailLoss or failure of external power supplyTop fuse is removed or external power supply is disconnected while in A/C mode with both external power and battery connectedwarning
External Power Fail alarmExternal electrical power disconnected or lostExternal AC power supply unplugged or power losscritical
EXTERNAL POWER LOWExternal power supply voltage has dropped below safe operating thresholdExternal power voltage drops below 10.5V - 11.1V rangewarning
Failure Code 1Self Check failure during startupSelf-check fails when unit powers oncritical
Failure Code 2No Gas, COMPSelf Check fails during startup or gas supply is disconnectederror
Failure Code 2No Gas, Comp alarm - external gas supply failure during compressor operationExternal gas pressure drops below 35 PSI when compressor is activecritical
Failure Code 2 (No Gas, Comp)No gas available to compressor; backup to external air activatedExternal gas supply pressure falls below 35 PSI +/- 2 PSI during compressor operation attempterror
Failure Code 3Excessive airway pressureAirway pressure exceeds safe limits during startupcritical
Failure Code 4Memory self-check failureMemory integrity check fails during startupcritical
Failure Code 5Safety shutdown triggered due to excessive airway pressure exceeding 40 cmH2O for 4 secondsAirway pressure sustained above 40 cmH2O for 4 seconds or morecritical
Failure Code 5Excessive airway pressure above 40 cmH2O with safety shutdownAirway pressure rises above 40 cmH2O and remains there for 4 seconds or morecritical
Failure Code 5Ventilator Failure Code 5Pressure exceeds 40 cmH2O for 4 secondscritical
Failure Code 5Exhaust Valve FailureAirway pressure reaches +100 cmH2O or aboveerror
Failure Code 5Safety shutdown activatedAirway pressure exceeds 40 cmH2O for 4 secondscritical
Failure Code 6Excessive Negative PressureAirway pressure falls to -10 cmH2O or belowerror
Failure Code 7Run-Time Transducer Calibration FailureTransducer calibration fails during operationerror
FIO2Oxygen concentration setting cannot be maintained due to low oxygen supply pressureFIO2 setting at 60% or higher with oxygen tank pressure below 35 PSI +/- 2 PSIcritical
FIO2Fraction of inspired oxygen alarm; oxygen delivery out of specificationOxygen tank pressure drops below 35 PSI +/- 2 PSI when high FIO2 setcritical
FIO2Oxygen concentration delivery failureOxygen tank pressure insufficient to achieve set FIO2; pressure below 45 PSI at high concentration settingscritical
FIO2FIO2 (inspired oxygen fraction) sensor detects insufficient oxygenOxygen tank pressure below threshold while FIO2 is set above 21%warning
O2 Low/FailOxygen tank pressure too low or oxygen supply failedOxygen tank pressure drops below 45 PSI threshold at 60% FIO2critical
System failureSystem has shut down due to insufficient external power voltageExternal power voltage drops below 9.50V - 10.50V rangecritical
System FailureBoth external power and battery power lost or fuses removedTop and bottom fuses both removedcritical
System FailureCritical system failure detectedBottom fuse is removed while top fuse is already removedcritical
System Failure LEDSystem failure condition detectedInternal system fault or critical component failurecritical
Total Flow BackupFlow backup alarm - inadequate gas delivery detectedDelivered tidal volume falls below 50% of set tidal volume for 4-5 consecutive breathserror
Total Flow BackupTotal flow backup alarmActual delivered tidal volume drops below 50% of set tidal volume for 4+ consecutive breathswarning
Total Flow BackupAlarm indicating insufficient flow delivery due to compressor inlet obstructionActual tidal volume less than half of set tidal volume for 4-5 consecutive breathswarning
Total Flow Backup AlarmInsufficient gas flow backup when primary flow path is compromisedActual delivered tidal volume drops below 50% of set tidal volume for 4-5 breathswarning
Total Flow Backup alarmActual delivered tidal volume insufficientActual tidal volume drops below half of set tidal volume after 4-5 breathswarning
Total Flow Backup alarmTidal volume insufficient due to compressor inlet restrictionActual delivered tidal volume falls below half of set tidal volume for 4-5 consecutive breathswarning
Total Flow Backup AlarmAlarm indicating total flow output is below expected tidal volume thresholdCompressor inlet occlusion or flow restriction reduces actual tidal volume to less than half of set tidal volume for 4-5 breathswarning
VtActual tidal volume delivered differs significantly from settingDelivered tidal volume diverges from set value during ventilationerror
VtTidal volume delivery is outside acceptable parametersTidal volume setting increased beyond alarm threshold (1150 ml or greater)error
VtMeasured or set tidal volume exceeds alarm limitTidal volume setting increased to 1150 mlerror
VtMeasured tidal volume out of specification or range exceededTidal volume setting set to 1150 ml or highererror
VtTidal volume critically highVt set to 1150 ml or highererror
Vt alarmTidal Volume alarmActual delivered tidal volume deviates from set tidal volumewarning
Vt alarmTidal volume alarm indicating delivered tidal volume does not meet set valueDuring external air and compressor switching test or when delivery is insufficientwarning
Vt SettingsTidal volume setting is outside configured parametersTidal volume setting changed to 1050 ml or greaterwarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI