The ServiceForge knowledge base holds 40 documented symptoms for the Hospira Plum A+, with 75 ranked causes, 66 diagnostic checks with test points and expected values, and 145 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Occlusion not detected patient safety

Proximal or distal occlusion test fails, system does not detect tubing blockage

Ranked causes

  1. 1. Cassette not properly primed (common)
  2. 2. Defective cassette (common)
  3. 3. Dirty sensor pin (occasional)
  4. 4. Defective APP PWA (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Proximal occlusion test failure patient safety

Proximal line occlusion detection does not function

Ranked causes

  1. 1. Closed proximal clamp (common)
  2. 2. Cassette priming issue (common)
  3. 3. Defective cassette or sensor (occasional)
  4. 4. Defective APP PWA (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air in line not detected patient safety

Proximal or distal air-in-line test fails, system does not detect air bubbles in tubing

Ranked causes

  1. 1. Defective special cassette (common)
  2. 2. Dirty sensors (common)
  3. 3. Defective APP PWA (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air-in-line detection failure patient safety

Proximal or distal air detection does not function

Ranked causes

  1. 1. Defective special cassette (common)
  2. 2. Dirty sensors (common)
  3. 3. Defective APP PWA (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm loudness test fails

Alarm does not produce expected sound level during test

Ranked causes

  1. 1. Defective CPU (occasional)
  2. 2. Defective peripheral PWA (occasional)
  3. 3. Defective piezo alarm assembly (common)
First check

Run Alarm Loudness Test

Test point
Alarm sound output
Expected
Alarm produces required sound level

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm loudness test failure caution

Alarm audible output does not meet specification

Ranked causes

  1. 1. Defective CPU (common)
  2. 2. Defective peripheral PWA (occasional)
  3. 3. Defective piezo alarm (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm not sounding patient safety

Alarm loudness test fails or alarm inaudible during operation

Ranked causes

  1. 1. Defective CPU PWA (common)
  2. 2. Defective peripheral PWA (occasional)
  3. 3. Defective piezo alarm assembly (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Delivery accuracy error patient safety

Delivery accuracy test fails, infusion rate or volume measurement inaccurate

Ranked causes

  1. 1. Cassette not properly primed (common)
  2. 2. Damaged or faulty cassette (common)
  3. 3. Defective mechanism assembly (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Delivery accuracy test failure patient safety

Infusion delivery rate does not meet specification

Ranked causes

  1. 1. Cassette priming issue (common)
  2. 2. Damaged or faulty cassette (common)
  3. 3. Defective mechanism (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Distal occlusion test failure patient safety

Distal line occlusion detection does not function

Ranked causes

  1. 1. Cassette priming issue (common)
  2. 2. Defective cassette or sensor (occasional)
  3. 3. Defective APP PWA (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air-in-line detected patient safety

Air bubbles detected in delivery cassette by proximal or distal air sensors; alarm N230-N234 generated

Ranked causes

  1. 1. Proximal air accumulation (> 500 µL) (common)
  2. 2. Distal air bolus or cumulative air (≥ 100 µL) (occasional)
First check

Verify air volume in proximal chamber

Test point
Cassette proximal reservoir
Expected
Less than 500 µL air visible

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Nurse call function test

Verify nurse call circuit operates correctly. Set primary rate to 400 mL/hr with 1 mL dose limit, observe short circuit reading on DMM after DOSE END and KVO appear.

Ranked causes

  1. 1. Nurse call test procedure as part of service verification (common)
  2. 2. Nurse call function not used (bypass test) (occasional)
First check

Primary rate and dose limit configuration

Test point
Display
Expected
Rate 400 mL/hr, dose limit 1 mL

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cassette alarm test failure caution

Cassette alarm function does not operate during test

Ranked causes

  1. 1. Cassette seating issue (common)
  2. 2. Defective cassette (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cassette test fails or CASSETTE TEST FAIL message displayed

LCD displays CASSETTE TEST FAIL flashing message and alarm sounds after inserting empty or primed cassette

Ranked causes

  1. 1. Empty cassette inserted for alarm test (common)
  2. 2. Cassette hardware or sensor malfunction (rare)
First check

Verify infuser is powered on before cassette insertion

Test point
LCD display and power indicator
Expected
Infuser on with DELIVERY screen or CCA selection screen displayed

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Door open while pumping or cassette insertion failure patient safety

Door opened during delivery or cassette not properly inserted; alarm N250 or N251 generated

Ranked causes

  1. 1. Door physically open or cassette loose (common)
  2. 2. Cassette or valve leak test failure (occasional)
First check

Verify door is closed

Test point
Door sensor/latch
Expected
Door fully closed and latched

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ethernet connection not functioning

Green LED on CE module is not illuminated when Ethernet connection should be active.

Ranked causes

  1. 1. Ethernet cable not properly connected to RJ-45 connector (common)
  2. 2. Green LED on CE module not illuminated (unknown)
First check

Verify shielded Ethernet cable is connected to RJ-45 connector

Test point
RJ-45 Ethernet connector on CE module
Expected
Shielded Ethernet cable firmly inserted

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Free flow test failure patient safety

Free flow detection does not function properly

Ranked causes

  1. 1. Cassette seating or defect (common)
  2. 2. Dirty or defective valve pins (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Minimal or no fluid flow during free flow test patient safety

Distal tubing shows no flow or excessive flow instead of minimal drops when cassette door open or closed

Ranked causes

  1. 1. Normal operation during free flow test (common)
  2. 2. Excessive free flow indicating improper cassette seating or valve malfunction (occasional)
First check

Prepare test setup with primed cassette and collection container

Test point
Cassette, tubing, collection container positioning
Expected
Primed cassette installed, distal tubing end minimum 36 inches below cassette in collection container

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Occlusion detected during delivery or non-delivery patient safety

Proximal or distal occlusion alarm generated (N180-N191); delivery line blocked or kinked

Ranked causes

  1. 1. Distal occlusion (delivery line blocked) (common)
  2. 2. Proximal occlusion (cassette inlet blocked) (common)
First check

Inspect distal line for occlusion

Test point
IV line from cassette to patient
Expected
Line patent; no kinks, clamps or visible blockage

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power on failure or unexpected shutdown patient safety

Infuser does not power on or powers off unexpectedly; low battery warnings or depleted battery message displayed

Ranked causes

  1. 1. Battery discharged or depleted (common)
  2. 2. Battery charging circuit failure (occasional)
First check

Verify battery terminal voltage

Test point
Battery terminals
Expected
Voltage greater than 5.45 V, > specified

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Self test fails patient safety

Device self test does not complete successfully

Ranked causes

  1. 1. Cassette not properly installed (common)
  2. 2. Defective CPU PWA (rare)
First check

Verify cassette installation

Test point
Cassette seating
Expected
Cassette fully seated in device

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Self test failure

Device power-on self test fails or cannot complete

Ranked causes

  1. 1. Cassette not properly installed or seated (common)
  2. 2. Defective CPU PWA (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unrestricted flow event patient safety

Infusion flows without control; regulator closer not properly seated or administration set clamp open

Ranked causes

  1. 1. Regulator closer not completely seated (common)
  2. 2. Administration set clamp not closed (common)
First check

Inspect regulator closer seating

Test point
Regulator closer position
Expected
Regulator closer completely seated

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Delivery accuracy verification patient safety

Verify infusion pump delivers correct volume at programmed rate. Test dual-line system with 200 mL/hr rate and 10 mL dose on each line, confirm final delivery is 20 mL ± 1 mL total and KVO/alarm sound at completion.

Ranked causes

  1. 1. Calibration procedure as part of maintenance and service tests (common)
First check

Cassette door inserted and NEW PATIENT confirmed

Test point
Display screen
Expected
NEW PATIENT? message appears, then NEW PATIENT confirmed with YES

8 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Electrical safety verification high voltage

Verify AC mains leakage current and ground resistance meet safety specifications. Test enclosure leakage (normal and open ground), earth leakage, and chassis ground resistance using a safety analyzer.

Ranked causes

  1. 1. Electrical safety test procedure required for service verification (common)
First check

Power cord and analyzer connection

Test point
AC mains inlet and ground test point
Expected
Power cord connected to safety analyzer, ground lead connected to rear ground screw

6 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display backlight or contrast not adjustable or illegible

LCD backlight does not illuminate, cannot be adjusted, or display contrast cannot be changed; screen not legible from 18 inches at eye level

Ranked causes

  1. 1. Backlight or display hardware malfunction (rare)
First check

Verify LCD backlight is on and display legible

Test point
LCD display at eye level
Expected
Backlight illuminated, display clearly legible from approximately 18 inches distance

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display test fails

Device display does not function during test

Ranked causes

  1. 1. Defective display assembly (common)
First check

Run Display Test

Test point
Display output
Expected
Display functions and shows output

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fluid shield tab bent or damaged patient safety

Visual inspection reveals fluid shield tab is bent or damaged, which may unseat regulator closer and cause unrestricted flow

Ranked causes

  1. 1. Fluid shield tab bent or damaged during removal or installation (occasional)
First check

Visually inspect fluid shield tab

Test point
Fluid shield tab visible from open corner of mechanism assembly
Expected
Tab is straight and not bent or damaged

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gasket tape installation on mechanism chassis caution

Installation of gasket tape to cover opening on mechanism chassis to protect Switch PWA from fluid ingress

Ranked causes

  1. 1. Missing or inadequate gasket tape coverage on mechanism chassis opening (occasional)
First check

Verify gasket tape presence and alignment

Test point
Mechanism chassis opening
Expected
Gasket tape installed and covering opening, secure and not protruding beyond chassis

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keypad functional test failure caution

Keypad does not respond correctly during testing

Ranked causes

  1. 1. Defective keypad (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keypad not responding

Keypad verification or functional test fails, keys not responding to input

Ranked causes

  1. 1. Defective keypad (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keypad or softkey function not responding or producing incorrect results patient safety

Keypad entries not registered, softkeys not functioning, START/STOP/programmed rates not executing, LED indicators not responding, or incorrect rate/VTBI values displayed

Ranked causes

  1. 1. Keypad or firmware malfunction preventing input or operation (occasional)
First check

Select line A and verify PROGRAM screen displays

Test point
LCD display, softkey A
Expected
PROGRAM screen displayed after pressing A softkey from DELIVERY screen

10 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keypad test fails

Keypad does not respond during functional test

Ranked causes

  1. 1. Defective keypad (common)
First check

Run Keypad Verification/Functional Test

Test point
Keypad response
Expected
Keypad responds to input

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Lockout switch test failure caution

Cassette lockout mechanism does not function

Ranked causes

  1. 1. Defective peripheral PWA (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Self test fails or alarm condition during startup patient safety

LCD does not display SELF TEST screen or alarm condition recurs when power cycled

Ranked causes

  1. 1. Hardware or firmware malfunction preventing self test completion (unknown)
First check

Verify AC power connection and CHARGE indicator

Test point
CHARGE/LINE indicator LED
Expected
CHARGE illuminates and alarm tone sounds

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not deliver fluid or displays error during priming patient safety

Pump fails to deliver fluid or shows malfunction when executing the delivery accuracy test sequence, including cassette insertion and programming.

Ranked causes

  1. 1. Pump mechanism malfunction or cassette issue (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unrestricted flow (regulator closer disengaged) patient safety

No alarm generated; door opening with disengaged regulator closer may cause uncontrolled fluid delivery

Ranked causes

  1. 1. Regulator closer mechanism disengaged or not fully engaged (rare)
First check

Remove mechanism and inspect regulator closer

Test point
Regulator closer assembly
Expected
Regulator closer fully engaged and moving freely

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unrestricted flow risk patient safety

Regulator closer not properly seated, creates risk of uncontrolled fluid flow

Ranked causes

  1. 1. Regulator closer not properly seated (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

VTBI complete (programmed volume infused) patient safety

Alarm N160 or N161 generated; infusion of programmed volume complete on line A or B

Ranked causes

  1. 1. VTBI (volume to be infused) reached (common)
First check

Verify programmed volume delivered

Test point
Volume counter on display
Expected
Volume infused equals VTBI setting

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Wireless connection not available

Wireless Connection Available icon does not appear on the main delivery screen when device should be receiving a wireless signal.

Ranked causes

  1. 1. Wireless network access point not recognized or signal not being received (unknown)
First check

Verify Wireless Connection Available icon appears on main delivery screen

Test point
Main delivery screen
Expected
Wireless Connection Available icon is displayed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
DEPLETED BATTERY (Info)Battery completely dischargedDevice turned on without being plugged in while battery is depletedinfo
E300ADC failure - analog to digital converter malfunctionADC converter failure detectedcritical
E300ADC failureAnalog to digital converter failureerror
E301Audio alarm failurePiezo is off but sensed on or piezo is on but sensed offerror
E302Backlight failureBacklight CCFT tube not at expected rangeerror
E320Battery charge/discharge current out of rangeCurrent limiting circuitry malfunction or excessive current drawerror
E321Charging current failure to drop within thresholdCharging current does not drop below end-of-charge threshold within 8 hours or below 1A within 6 hourswarning
E321Battery not chargingBattery charging timed out or complete battery discharge has occurredcritical
E322Battery current calibration value out of rangeCharge current measurement circuitry failureerror
E323Battery trickle charge current out of rangeTrickle charge current outside specificationerror
E324Supply overvoltage - charging circuit faultCharging circuit produces excessive voltagecritical
E325Battery overvoltageBattery voltage exceeds specificationerror
E326Battery disconnected while infuser on or voltage too lowBattery connection loss detected during operation or voltage below minimumcritical
E327Brownout conditionBrownout condition detectederror
E340Critical instruction failure - CPU register test failedPower-up CPU register test failurecritical
E341Critical data memory failureCritical data memory checksum or access failurecritical
E342Display failure - defective displayDisplay malfunction detectederror
E343Distal air sensor failure 1 - detects liquid with cassette removedSelf-test detects liquid with cassette outerror
E344Distal air sensor failure 2 - sensor out of rangeSelf-test with cassette inserted detects out-of-rangeerror
E345Distal pressure sensor failure 1 - failed while offSensor failure detected during standbyerror
E346Distal pressure sensor failure 2 - failed while onSensor failure detected during operationerror
E347Hardware watchdog failureHardware watchdog timer malfunctioncritical
E378I/O valve phase lossI/O valve motor phase detection failureerror
E379L/S valve phase lossL/S valve motor phase detection failureerror
E380Plunger motor phase loss - generic motor failureMotor output loss detectederror
E430Proximal air sensor failure 1 - detects liquid cassette removedOngoing test detects liquid with cassette outerror
E431Proximal air sensor failure 2 - self test detects liquidSelf-test detects liquid with cassette removederror
E432Proximal pressure sensor failure 1 - failed while offSensor failure detected during standbyerror
E433Proximal pressure sensor failure 2 - failed while onSensor failure detected during operationerror
E434RAM failureRAM memory test failureerror
E435RTC failureReal-time clock malfunctionerror
E435RTC failure - real-time clock malfunctionReal-time clock circuit failureerror
E436ROM failureROM checksum verification failureerror
E437Software failure - generic software malfunctionGeneric software fault detectederror
E438Stack out-of-range failureSoftware stack pointer overflow/underflowcritical
E439Stuck key - key pressed for >2 minutesKeypad stuck in pressed state for over 2 minuteserror
E439Stuck keyKey sensed as pressed for over two minuteswarning
E440Power hold stuckPower hold signal remains assertedcritical
E443LCD failure - LCD bias out of rangeDisplay bias voltage out of specificationerror
E444CPU timebase inaccurateCPU timer 2 and RTC measured times disagreewarning
E445RTC memory failureReal-time clock memory corruption detectedwarning
E446CPU timer failureCPU timer 1 and timer 2 measured times disagreeerror
E447Battery ADC reading failure - consecutive zero/max readings16 consecutive ADC readings all zero or maximumerror
E448SEEP write failureSEEP data write operation failederror
E449SEEP calibration data corruptedCalibration data block checksum failureerror
E450MMIO port read/write failureMemory-mapped I/O port read/write errorerror
E451Inaccurate deliveryOver/under delivery detected during operationerror
E452Software failure - miscellaneous software failuresVarious software fault conditionserror
E453Two SEEP CRC errors - NVRAM data corruptedMultiple non-volatile RAM checksum failurescritical
E454NVRAM over capacity - write to non-existent spaceSoftware attempts to write beyond NVRAM limitserror
E455Invalid device configurationIncorrect flash memory on peripheral PWAerror
E456Invalid drug libraryDrug library install was started but not completed successfullywarning
E457Drug library corruptedCRC failure on drug libraryerror
LOW BATTERY (Warning)Battery has reduced capacityReduced battery capacity detected; first or second occurrencewarning
N100Unrecognizable cassetteIncorrect cassette type insertedwarning
N100 (URC)Unrecognizable cassetteIncorrect cassette type insertederror
N101No Action AlarmNo operator action and no delivery for 2 minutes during parameter entrywarning
N101 (NAA)No action - infuser idleNo operator action and no delivery for two minutes during parameter entrywarning
N102Infuser Idle 2 MinutesInfuser in reset or idle for over 2 minutes without start commandwarning
N102 (RL)Infuser idle over 2 minutesInfuser in reset or idle state for over 2 minuteswarning
N103NV RAM lost therapy dataTherapy data is losterror
N103 (SEEP CRC)NV RAM lost therapy dataTherapy data lost due to incomplete NVRAM writeerror
N104Nurse callback BDelivery line B has changed (if alarm enabled)warning
N104 (NC2)Nurse callback BDelivery line B changed while callback enabledwarning
N105Nurse callback ADelivery line A has changed (if alarm enabled)warning
N105 (NC1)Nurse callback ADelivery line A changed while callback enabledwarning
N160Line B VTBI completeProgrammed volume to be infused completed on line Binfo
N160/E160Line B VTBI completeProgrammed VTBI completed on line Binfo
N160/E160 (VTB2)Line B VTBI completeProgrammed VTBI completed on line Binfo
N161Line A VTBI completeProgrammed volume to be infused completed on line Ainfo
N161/E161Line A VTBI completeProgrammed VTBI completed on line Ainfo
N161/E161 (VTB1)Line A VTBI completeProgrammed VTBI completed on line Ainfo
N180Distal OcclusionPeak distal occlusion detected during non-delivery phasewarning
N180/E180Distal occlusion, peakDistal occlusion detected during non-deliveryerror
N180/E180 (OD1)Distal occlusion - peak, non-deliveryPeak distal occlusion detected during non-deliveryerror
N181/E181Distal occlusion, negativeDistal occlusion detected during non-deliveryerror
N181/E181 (OD1)Distal occlusion - negative, non-deliveryNegative distal occlusion detected during non-deliveryerror
N182/E182Proximal occlusion B, negativeProximal occlusion detected on line B during non-deliveryerror
N182/E182 (OP2)Proximal occlusion B, air - negative, non-deliveryNegative proximal occlusion B detected during non-deliveryerror
N183Prox Occl B At StartupPeak proximal occlusion B detected during non-delivery startupwarning
N183/E183Proximal occlusion B, peakProximal occlusion detected on line B during non-deliveryerror
N183/E183 (OP2)Proximal occlusion B, air - peak, non-deliveryPeak proximal occlusion B detected during non-deliveryerror
N184Prox Occl A / AirNegative proximal occlusion A detected during non-deliverywarning
N184/E184Proximal occlusion A, negativeProximal occlusion detected on line A during non-deliveryerror
N184/E184 (OP1)Proximal occlusion A, air - negative, non-deliveryNegative proximal occlusion A detected during non-deliveryerror
N185Prox Occl A At StartupPeak proximal occlusion A detected at startup non-deliverywarning
N185/E185Proximal occlusion A, peakProximal occlusion detected on line A during non-deliveryerror
N185/E185 (OP1)Proximal occlusion A, air - peak, non-deliveryPeak proximal occlusion A detected during non-deliveryerror
N186Distal OcclusionPeak distal occlusion detected during deliveryerror
N186/E186Distal occlusion during delivery, peakDistal occlusion detected during deliveryerror
N186/E186 (OD1)Distal occlusion - peak, deliveryPeak distal occlusion detected during deliveryerror
N187Neg Distal OcclusionNegative distal occlusion during deliveryerror
N187/E187Distal occlusion during delivery, negativeDistal occlusion detected during deliveryerror
N187/E187 (OD1)Distal occlusion - negative, deliveryNegative distal occlusion detected during deliveryerror
N188Prox Occl B / AirNegative proximal occlusion B during deliveryerror
N188/E188Proximal occlusion B during delivery, negativeProximal occlusion detected during delivery on line Berror
N188/E188 (OP2)Proximal occlusion B - negative, deliveryNegative proximal occlusion B detected during deliveryerror
N189Prox Occl B / AirPeak proximal occlusion B during deliveryerror
N189/E189Proximal occlusion B during delivery, peakProximal occlusion detected during delivery on line Berror
N189/E189 (OP2)Proximal occlusion B - peak, deliveryPeak proximal occlusion B detected during deliveryerror
N190Prox Occl A / AirNegative proximal occlusion A during deliveryerror
N190/E190Proximal occlusion A during delivery, negativeProximal occlusion detected during delivery on line Aerror
N190/E190 (OP1)Proximal occlusion A - negative, deliveryNegative proximal occlusion A detected during deliveryerror
N191Prox Occl A / AirPeak proximal occlusion A during deliveryerror
N191/E191Proximal occlusion A during delivery, peakProximal occlusion detected during delivery on line Aerror
N191/E191 (OP1)Proximal occlusion A - peak, deliveryPeak proximal occlusion A detected during deliveryerror
N230Prox Air, Backprime500 µL air accumulated in proximal line (combined both lines)warning
N230/E230Proximal air total500 µL of air has entered the cassetteerror
N230/E230 (APT)Proximal air-in-line total500 microliter of air entered cassetteerror
N231Prox Air B, Backprime500 µL air accumulated in proximal line Bwarning
N231/E231Proximal air on line B500 µL of air has entered the cassette on line Berror
N231/E231 (APB)Proximal air on B, backprime500 microliter of air entered cassette on line Berror
N232Prox Air A, Backprime500 µL air accumulated in proximal line Awarning
N232/E232Proximal air on line A500 µL of air has entered the cassette on line Aerror
N232/E232 (APA)Proximal air on A, backprime500 microliter of air entered cassette on line Aerror
N233Distal Air500 µL cumulative air detected in last 5.3 mL fluid deliveredwarning
N233/E233Distal air cumulative500 µL of air detected in the last 5.3 mL of fluid deliverederror
N233/E233 (ADC)Distal air cumulative500 microliter air detected in last 5.3 mL deliverederror
N234Distal Air100 µL air bolus detected at distal sensorwarning
N234/E234Distal air bolus100 µL bolus of air detected at distal sensorerror
N234/E234 (ADB)Distal air bolus100 microliter bolus of air detected at distal sensorerror
N250Door open while pumpingDoor opened during active deliveryerror
N250/E250Door opened while pumpingDoor opened while pumpingcritical
N250/E250 (DCO1)Door opened while pumpingInfuser door opened during active pumpingerror
N251Cassette Test FailureValve/cassette fails leak testerror
N251/E251Valve/cassette test failureValve/cassette fails the leak testerror
N251/E251 (CS1)Valve/cassette test failureValve/cassette fails leak testerror
N252Depleted BatteryLow battery terminal voltage during operationcritical
N252/E252Depleted batteryBattery terminal voltage is less than 5.45 Verror
N252/E252 (BDP)Depleted batteryBattery terminal voltage less than 5.45 Verror
N253Hard lockout ViolationAttempt to use [STOP] key or open door while hard lockout engagedwarning
N253/E253Lockout violation, hardUse of [STOP] key or attempt to open door while lockout switch is lockedwarning
N253/E253 (LOV)Lockout violation - hardSTOP key pressed or door open attempt while lockout activeunknown
N254Hard Lockout EnabledKeypad action taken while hard lockout switch lockedwarning
N254/E254Lockout enabled, keypad lockedAny action not resulting in stopping of delivery while lockout switch is lockedwarning
N254/E254 (FPL)Lockout enabled - keypad lockedAction not stopping delivery while lockout enabledunknown
N255Soft lockout ViolationAttempt to use [STOP] key or open door while soft lockout enabledwarning
N255 (SLV)Soft lockout violationSTOP key pressed or door open attempt while soft lockout activewarning
N256Soft lockout EnabledKeypad action taken while soft lockout switch lockedwarning
N256 (SLE)Soft lockout enabledAction not stopping delivery while soft lockout enabledwarning
No AlarmUnrestricted flowRegulator closer disengaged and door opensunknown
REPLACE BATTERY (Warning)Battery must be replaced immediatelyThird and subsequent occurrences of reduced battery capacitywarning
Warning: Charger ServiceCharger hardware problemBattery charging circuit malfunction detectedunknown
Warning: Low BatteryLow battery warningBattery discharged to approximately 30 minutes of operation remainingwarning
Warning: Replace BatteryBattery service neededBattery voltage less than depleted threshold and charge above low thresholdwarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI