Hospira Plum A+ with MedNet: troubleshooting and error codes

The ServiceForge knowledge base holds 28 documented symptoms for the Hospira Plum A+ with MedNet, with 41 ranked causes, 39 diagnostic checks with test points and expected values, and 107 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Alarm loudness test fails

Alarm does not produce expected sound level during test

Ranked causes

  1. 1. Defective CPU (occasional)
  2. 2. Defective peripheral PWA (occasional)
  3. 3. Defective piezo alarm assembly (common)
First check

Run Alarm Loudness Test

Test point
Alarm sound output
Expected
Alarm produces required sound level

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm loudness test failure caution

Alarm does not meet volume requirements during loudness test

Ranked causes

  1. 1. Defective CPU (occasional)
  2. 2. Defective peripheral PWA (occasional)
  3. 3. Defective piezo alarm assembly (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unrestricted flow test failure patient safety

Pump does not pass unrestricted flow performance test; unable to deliver fluid without restriction

Ranked causes

  1. 1. Cassette not properly seated (common)
  2. 2. Defective cassette (occasional)
  3. 3. Defective or dirty valve pins (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cassette alarm test failure patient safety

Cassette alarm test does not complete properly during performance verification test

Ranked causes

  1. 1. Cassette not properly seated (common)
  2. 2. Defective cassette (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Delivery accuracy not within specification patient safety

Total fluid delivered differs from programmed VTBI by more than ±1 mL

Ranked causes

  1. 1. Pump mechanism wear or cassette chamber fatigue (occasional)
  2. 2. Pump mechanism mechanical failure (rare)
First check

Program Line A at 200 mL/hr, VTBI 10 mL and start pumping

Test point
Line A
Expected
PUMPING message and LED flashing

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Electrical safety test failure high voltage

Leakage current or chassis ground resistance exceeds specification

Ranked causes

  1. 1. Damaged or degraded power cord insulation (occasional)
  2. 2. Poor chassis ground connection (occasional)
First check

Measure enclosure leakage current (normal condition, ground intact)

Test point
Safety analyzer
Expected
Not to exceed 100 µA
Tools
Safety analyzer per specifications in Section 5.3.1

4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ethernet connection not available

Green LED on CE module is not illuminated; Ethernet connectivity is not established

Ranked causes

  1. 1. Ethernet cable not properly inserted into RJ-45 connector (common)
  2. 2. CE module failure or connection issue (occasional)
First check

Verify shielded Ethernet cable is plugged into RJ-45 connector

Test point
RJ-45 connector on CE module
Expected
Cable fully and securely inserted

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Self test fails patient safety

Device self test does not complete successfully

Ranked causes

  1. 1. Cassette not properly installed (common)
  2. 2. Defective CPU PWA (rare)
First check

Verify cassette installation

Test point
Cassette seating
Expected
Cassette fully seated in device

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Self test failure patient safety

Self-test does not complete successfully on startup; cassette or CPU issue detected

Ranked causes

  1. 1. Cassette not properly installed (common)
  2. 2. Defective CPU PWA (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unrestricted flow event patient safety

Infusion flows without control; regulator closer not properly seated or administration set clamp open

Ranked causes

  1. 1. Regulator closer not completely seated (common)
  2. 2. Administration set clamp not closed (common)
First check

Inspect regulator closer seating

Test point
Regulator closer position
Expected
Regulator closer completely seated

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display test fails

Device display does not function during test

Ranked causes

  1. 1. Defective display assembly (common)
First check

Run Display Test

Test point
Display output
Expected
Display functions and shows output

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display test failure caution

Display assembly does not respond to or pass functional test during PVT

Ranked causes

  1. 1. Defective display assembly (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Distal air-in-line not detected patient safety

Test cassette with distal air in line does not trigger detection alarm

Ranked causes

  1. 1. Distal air-in-line detection sensor malfunction (unknown)
First check

Insert distal test cassette and start pumping

Test point
Line A
Expected
PUMPING message displayed, LED above Line A flashes

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Distal occlusion not detected at set pressure patient safety

Occlusion alarm does not trigger at specified distal pressure limit

Ranked causes

  1. 1. Distal pressure sensor malfunction or miscalibration (unknown)
First check

Set distal pressure limit to 6 psi

Test point
Options/Pressure/Post Infusion Rate menu
Expected
Distal Pressure Limit set to 6 psi

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ethernet connectivity not established

Green LED on CE module is not illuminated when Ethernet cable is connected

Ranked causes

  1. 1. Shielded Ethernet cable not properly plugged into RJ-45 connector or cable malfunction (unknown)
First check

Verify shielded Ethernet cable is connected to RJ-45 connector

Test point
CE module RJ-45 connector
Expected
Shielded Ethernet cable firmly inserted

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ethernet connector cable tab binding caution

Ethernet cable connector with tabs cannot be easily removed from RJ-45 connector

Ranked causes

  1. 1. Cable connector designed with anti-tangling tabs (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keypad lockout switch malfunction

Keypad does not lock when switch moved to ON position, or invalid key presses do not generate alert

Ranked causes

  1. 1. Keypad lockout switch not in ON position (common)
First check

Verify keypad lockout switch position

Test point
Back of infuser keypad lockout switch
Expected
UP (ON) position

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keypad test fails

Keypad does not respond during functional test

Ranked causes

  1. 1. Defective keypad (common)
First check

Run Keypad Verification/Functional Test

Test point
Keypad response
Expected
Keypad responds to input

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keypad verification or functional test failure caution

Keypad does not respond properly during verification or functional test

Ranked causes

  1. 1. Defective keypad (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Nurse call function not triggering

Nurse call jack does not generate short circuit when infusion completes

Ranked causes

  1. 1. Nurse call jack contact failure or wiring disconnection (unknown)
First check

Connect DMM to nurse call jack for resistance measurement

Test point
Nurse call jack on back of infuser
Expected
DMM set to measure resistance

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Proximal air-in-line not detected patient safety

Test cassette with proximal air in line does not trigger detection alarm

Ranked causes

  1. 1. Proximal air-in-line detection sensor malfunction (unknown)
First check

Insert proximal test cassette and start pumping

Test point
Line A
Expected
PUMPING message displayed, LED above Line A flashes

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Proximal occlusion not detected patient safety

Line A tubing clamped proximal to cassette but PROXOCCL A/AIR alarm does not sound

Ranked causes

  1. 1. Proximal occlusion detection sensor malfunction (unknown)
First check

Verify pumping starts with Line A at 400 mL/hr, VTBI 50 mL

Test point
Line A LED and status bar
Expected
LED above Line A flashes, PUMPING message displayed

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Stop delivery, then turn off patient safety

Operator attempted to turn off infuser while a delivery is in progress.

Ranked causes

  1. 1. Operator error - incorrect shutdown sequence (common)
First check

Verify all infusion lines stopped

Test point
Infusion pump channels
Expected
All channels in stop state; no flow occurring

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Warning: Charger Service caution

Battery charging circuit hardware problem detected; charging circuitry not behaving as expected.

Ranked causes

  1. 1. Battery charging circuit malfunction (common)
First check

Verify AC adapter connection

Test point
Power connector
Expected
Secure connection at charging port

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Warning: Low Battery caution

Infuser battery is discharged with approximately 30 minutes of battery life remaining; warning message displayed on LCD screen.

Ranked causes

  1. 1. Battery depleted during operation (common)
First check

Verify AC power availability

Test point
Power outlet
Expected
AC power available and functional

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Warning: Replace Battery caution

Battery voltage has dropped below depleted threshold while charge level is above low charge threshold, indicating battery service is needed.

Ranked causes

  1. 1. Battery has reached end of service life (common)
First check

Verify battery replacement procedure

Test point
Battery compartment
Expected
Old battery removed; new battery installed per section 7.2.4

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Wireless connection not appearing on main delivery screen

Wireless Connection Available icon does not appear on the main delivery screen despite device being in wireless network environment

Ranked causes

  1. 1. Device not receiving wireless signal or network access point not recognized (unknown)
First check

Verify device is receiving wireless signal and network access point is recognized

Test point
Main delivery screen
Expected
Wireless Connection Available icon displayed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Wireless connection not available

Wireless Connection Available icon does not appear on main delivery screen; device is not receiving wireless signal

Ranked causes

  1. 1. Wireless network access point not recognized or signal not received (common)
First check

Verify Wireless Connection Available icon appears on main delivery screen

Test point
Main delivery screen display
Expected
Wireless Connection Available icon is displayed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
Charger Service WarningBattery charging hardware circuit malfunction detected; charger not functioning properlyCharging circuitry hardware detects abnormal behavior or failure conditionwarning
E160Line B VTBI completeProgrammed VTBI completed on line Berror
E161Line A VTBI completeProgrammed VTBI completed on line Aerror
E180Distal occlusion detected during non-deliveryPeak distal occlusion, non-deliveryerror
E181Distal occlusion detected during non-deliveryNegative distal occlusion, non-deliveryerror
E182Proximal occlusion B detected during non-deliveryNegative proximal occlusion B, non-deliveryerror
E183Proximal occlusion B detected during non-deliveryPeak proximal occlusion B, non-deliveryerror
E184Proximal occlusion A detected during non-deliveryNegative proximal occlusion A, non-deliveryerror
E185Proximal occlusion A detected during non-deliveryPeak proximal occlusion A, non-deliveryerror
E186Distal occlusion detected during deliveryPeak distal occlusion, deliveryerror
E187Distal occlusion detected during deliveryNegative distal occlusion, deliveryerror
E188Proximal occlusion B detected during deliveryNegative proximal occlusion B, deliveryerror
E189Proximal occlusion B detected during deliveryPeak proximal occlusion B, deliveryerror
E190Proximal occlusion A detected during deliveryNegative proximal occlusion A, deliveryerror
E191Proximal occlusion A detected during deliveryPeak proximal occlusion A, deliveryerror
E230Proximal air-in-line total detected500 microL of air has entered the cassetteerror
E231Proximal air-in-line on line B detected500 microL of air has entered the cassette on line Berror
E232Proximal air-in-line on line A detected500 microL of air has entered the cassette on line Aerror
E233Distal air cumulative detected500 microL of air detected in the last 5.3 mL of fluid deliverederror
E234Distal air-in-line detected (older firmware version)Air detected in distal line before 1 mL fluid delivered on older Plum A+ software versionswarning
E250Door opened while pumpingDoor opened while pumpingerror
E251Valve/cassette test failureValve/cassette fails the leak testerror
E252Depleted batteryBattery terminal voltage is less than 5.45 Verror
E253Lockout violation - hard lockoutThe use of [STOP] key or attempt to open door while lockout switch is lockederror
E254Lockout enabled - keypad lockedAny action not resulting in stopping delivery while lockout switch is lockederror
E300ADC failureAnalog to digital converter failureerror
E301Audio alarm failurePiezo is off but sensed on or Piezo is on but sensed offerror
E302Backlight failureBacklight (CCFT tube) is not at the expected rangeerror
E320Battery charge/discharge current out of rangeCurrent limiting circuitry in power supply board has failed or other hardware causing excessive current drawcritical
E321Failure of charging current to drop below thresholdCharging current does not drop below end of charge threshold within 8 hours or below 1A within 6 hourscritical
E322Battery current calibration value out of rangeBattery charging current measurement is out of acceptable rangecritical
E323Battery trickle charge current out of rangeBattery trickle charge current is out of acceptable rangecritical
E324Supply overvoltagePower supply voltage exceeds safe operating rangecritical
E325Battery overvoltageBattery voltage exceeds acceptable rangecritical
E326Battery disconnected or voltage too lowConnectors not seated on battery terminals or battery voltage critically low while infuser is oncritical
E340Critical instruction failurePower-up CPU register test failedcritical
E341Critical data memory failureCritical data memory failure detectedcritical
E342Display failureDisplay assembly malfunction detectedcritical
E343Distal air sensor failure 1With cassette removed, distal air sensor self test detects liquiderror
E344Distal air sensor failure 2With cassette inserted, distal air sensor self test detects sensor out of rangeerror
E345Distal pressure sensor failure 1Distal pressure sensor failed while infuser is offerror
E346Distal pressure sensor failure 2Distal pressure sensor failed while infuser is onerror
E347Hardware watchdog failureHardware watchdog timer failure detectedcritical
E378I/O valve phase lossGeneric I/O valve failure detectederror
E379L/S valve phase lossGeneric L/S valve failure detectederror
E380Plunger motor phase lossGeneric plunger motor failure detectederror
E430Proximal air sensor failure 1Proximal air sensor ongoing test detects liquid with cassette removederror
E431Proximal air sensor failure 2Proximal air sensor self test detects liquid with cassette removederror
E432Proximal pressure sensor 1Proximal pressure sensor failed while infuser is offerror
E433Proximal pressure sensor 2Proximal pressure sensor failed while infuser is onerror
E434RAM failureRAM memory failure detectedcritical
E435RTC failureReal-time clock failure detectedcritical
E436ROM failureROM checksum failure detectedcritical
E437Software failureGeneric software failure detectedcritical
E438Stack out-of-range failureStack pointer exceeds valid memory rangecritical
E439Stuck keyA key is sensed as pressed for over two minuteserror
E440Power hold stuckPower hold signal stuck in active statecritical
E443LCD failureLCD bias is out of acceptable rangeerror
E444CPU timebase inaccurateCPU timer 2 and RTC measured times disagreeerror
E445RTC memory failureReal-time clock memory is corruptederror
E446CPU timer failureCPU timer 1 and timer 2 measured times disagreeerror
E447Battery ADC reading failure16 consecutive battery ADC readings are all zero or max valueerror
E448SEEP write failureSEEP data write operation failedcritical
E449SEEP calibration data corruptedCalibration data block in SEEP is corruptedcritical
E450MMIO port read/write failureI/O port read or write operation failedcritical
E451Inaccurate deliveryOver-delivery or under-delivery detected during operationwarning
E452Software failureMiscellaneous software failures not categorized elsewherecritical
E453Two SEEP CRC errorsNVRAM data block CRC validation failed twicecritical
E454NVRAM over capacitySoftware attempting to write into non-existent NVRAM spacecritical
E455Invalid device configurationIncorrect flash memory detected on peripheral PWAcritical
E456Invalid drug libraryDrug library installation started but not completed successfullyerror
E457Drug library corruptedCRC failure detected on drug library dataerror
Low Battery WarningBattery charge critically low with approximately 30 minutes remaining before power lossBattery voltage drops to threshold indicating insufficient charge for extended operationwarning
N100Unrecognizable cassetteIncorrect cassette type insertedwarning
N101No action - operator idle during delivery parameters entryNo operator action and no delivery for two minutes during delivery parameters entrywarning
N102Infuser idle 2 minutesInfuser in reset or idle for over two minuteswarning
N103NV RAM lost therapy dataTherapy data is losterror
N104Nurse callback BDelivery line B has changed (if alarm is enabled)info
N105Nurse callback ADelivery line A has changed (if alarm is enabled)info
N160Line B VTBI completeProgrammed VTBI completed on line Binfo
N161Line A VTBI completeProgrammed VTBI completed on line Ainfo
N180Distal occlusion detected during non-deliveryPeak distal occlusion, non-deliverywarning
N181Distal occlusion detected during non-deliveryNegative distal occlusion, non-deliverywarning
N182Proximal occlusion B detected during non-deliveryNegative proximal occlusion B, non-deliverywarning
N183Proximal occlusion B detected during non-deliveryPeak proximal occlusion B, non-deliverywarning
N184Proximal occlusion A detected during non-deliveryNegative proximal occlusion A, non-deliverywarning
N185Proximal occlusion A detected during non-deliveryPeak proximal occlusion A, non-deliverywarning
N186Distal occlusion detected during deliveryPeak distal occlusion, deliverywarning
N187Distal occlusion detected during deliveryNegative distal occlusion, deliverywarning
N188Proximal occlusion B detected during deliveryNegative proximal occlusion B, deliverywarning
N189Proximal occlusion B detected during deliveryPeak proximal occlusion B, deliverywarning
N190Proximal occlusion A detected during deliveryNegative proximal occlusion A, deliverywarning
N191Proximal occlusion A detected during deliveryPeak proximal occlusion A, deliverywarning
N230Proximal air-in-line total detected500 microL of air has entered the cassettewarning
N231Proximal air-in-line on line B detected500 microL of air has entered the cassette on line Bwarning
N232Proximal air-in-line detected with backpriming requiredAir detected in proximal line before 1 mL fluid delivered during proximal air-in-line testwarning
N233Distal air cumulative detected500 microL of air detected in the last 5.3 mL of fluid deliveredwarning
N234Distal air-in-line detectedAir detected in distal line before 1 mL fluid delivered during distal air-in-line testwarning
N250Door opened while pumpingDoor opened while pumpingwarning
N251Valve/cassette test failureValve/cassette fails the leak testwarning
N252Depleted batteryBattery terminal voltage is less than 5.45 Vwarning
N253Lockout violation - hard lockoutThe use of [STOP] key or attempt to open door while lockout switch is lockedwarning
N254Lockout enabled - keypad lockedAny action not resulting in stopping delivery while lockout switch is lockedwarning
N255Lockout violation - soft lockoutThe use of [STOP] key or attempt to open door while lockout switch is lockedwarning
N256Lockout enabled - soft lockoutAny action not resulting in stopping delivery while lockout switch is lockedwarning
Replace Battery WarningBattery service or replacement required; voltage below depletion thresholdBattery voltage drops below minimum serviceable voltage while charge level above low thresholdwarning
Stop Delivery FirstShutdown blocked; infuser cannot power down during active deliveryPower off command received while infusion pump is actively delivering fluid to patientcritical

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.