Hospira Plum A+ with MedNet: troubleshooting and error codes
The ServiceForge knowledge base holds 28 documented symptoms for the Hospira Plum A+ with MedNet, with 41 ranked causes, 39 diagnostic checks with test points and expected values, and 107 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Alarm loudness test fails
Alarm does not produce expected sound level during test
Ranked causes
- 1. Defective CPU (occasional)
- 2. Defective peripheral PWA (occasional)
- 3. Defective piezo alarm assembly (common)
Run Alarm Loudness Test
- Test point
- Alarm sound output
- Expected
- Alarm produces required sound level
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm loudness test failure caution
Alarm does not meet volume requirements during loudness test
Ranked causes
- 1. Defective CPU (occasional)
- 2. Defective peripheral PWA (occasional)
- 3. Defective piezo alarm assembly (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unrestricted flow test failure patient safety
Pump does not pass unrestricted flow performance test; unable to deliver fluid without restriction
Ranked causes
- 1. Cassette not properly seated (common)
- 2. Defective cassette (occasional)
- 3. Defective or dirty valve pins (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cassette alarm test failure patient safety
Cassette alarm test does not complete properly during performance verification test
Ranked causes
- 1. Cassette not properly seated (common)
- 2. Defective cassette (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Delivery accuracy not within specification patient safety
Total fluid delivered differs from programmed VTBI by more than ±1 mL
Ranked causes
- 1. Pump mechanism wear or cassette chamber fatigue (occasional)
- 2. Pump mechanism mechanical failure (rare)
Program Line A at 200 mL/hr, VTBI 10 mL and start pumping
- Test point
- Line A
- Expected
- PUMPING message and LED flashing
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrical safety test failure high voltage
Leakage current or chassis ground resistance exceeds specification
Ranked causes
- 1. Damaged or degraded power cord insulation (occasional)
- 2. Poor chassis ground connection (occasional)
Measure enclosure leakage current (normal condition, ground intact)
- Test point
- Safety analyzer
- Expected
- Not to exceed 100 µA
- Tools
- Safety analyzer per specifications in Section 5.3.1
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ethernet connection not available
Green LED on CE module is not illuminated; Ethernet connectivity is not established
Ranked causes
- 1. Ethernet cable not properly inserted into RJ-45 connector (common)
- 2. CE module failure or connection issue (occasional)
Verify shielded Ethernet cable is plugged into RJ-45 connector
- Test point
- RJ-45 connector on CE module
- Expected
- Cable fully and securely inserted
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Self test fails patient safety
Device self test does not complete successfully
Ranked causes
- 1. Cassette not properly installed (common)
- 2. Defective CPU PWA (rare)
Verify cassette installation
- Test point
- Cassette seating
- Expected
- Cassette fully seated in device
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Self test failure patient safety
Self-test does not complete successfully on startup; cassette or CPU issue detected
Ranked causes
- 1. Cassette not properly installed (common)
- 2. Defective CPU PWA (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unrestricted flow event patient safety
Infusion flows without control; regulator closer not properly seated or administration set clamp open
Ranked causes
- 1. Regulator closer not completely seated (common)
- 2. Administration set clamp not closed (common)
Inspect regulator closer seating
- Test point
- Regulator closer position
- Expected
- Regulator closer completely seated
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display test fails
Device display does not function during test
Ranked causes
- 1. Defective display assembly (common)
Run Display Test
- Test point
- Display output
- Expected
- Display functions and shows output
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display test failure caution
Display assembly does not respond to or pass functional test during PVT
Ranked causes
- 1. Defective display assembly (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Distal air-in-line not detected patient safety
Test cassette with distal air in line does not trigger detection alarm
Ranked causes
- 1. Distal air-in-line detection sensor malfunction (unknown)
Insert distal test cassette and start pumping
- Test point
- Line A
- Expected
- PUMPING message displayed, LED above Line A flashes
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Distal occlusion not detected at set pressure patient safety
Occlusion alarm does not trigger at specified distal pressure limit
Ranked causes
- 1. Distal pressure sensor malfunction or miscalibration (unknown)
Set distal pressure limit to 6 psi
- Test point
- Options/Pressure/Post Infusion Rate menu
- Expected
- Distal Pressure Limit set to 6 psi
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ethernet connectivity not established
Green LED on CE module is not illuminated when Ethernet cable is connected
Ranked causes
- 1. Shielded Ethernet cable not properly plugged into RJ-45 connector or cable malfunction (unknown)
Verify shielded Ethernet cable is connected to RJ-45 connector
- Test point
- CE module RJ-45 connector
- Expected
- Shielded Ethernet cable firmly inserted
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ethernet connector cable tab binding caution
Ethernet cable connector with tabs cannot be easily removed from RJ-45 connector
Ranked causes
- 1. Cable connector designed with anti-tangling tabs (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keypad lockout switch malfunction
Keypad does not lock when switch moved to ON position, or invalid key presses do not generate alert
Ranked causes
- 1. Keypad lockout switch not in ON position (common)
Verify keypad lockout switch position
- Test point
- Back of infuser keypad lockout switch
- Expected
- UP (ON) position
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keypad test fails
Keypad does not respond during functional test
Ranked causes
- 1. Defective keypad (common)
Run Keypad Verification/Functional Test
- Test point
- Keypad response
- Expected
- Keypad responds to input
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keypad verification or functional test failure caution
Keypad does not respond properly during verification or functional test
Ranked causes
- 1. Defective keypad (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nurse call function not triggering
Nurse call jack does not generate short circuit when infusion completes
Ranked causes
- 1. Nurse call jack contact failure or wiring disconnection (unknown)
Connect DMM to nurse call jack for resistance measurement
- Test point
- Nurse call jack on back of infuser
- Expected
- DMM set to measure resistance
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Proximal air-in-line not detected patient safety
Test cassette with proximal air in line does not trigger detection alarm
Ranked causes
- 1. Proximal air-in-line detection sensor malfunction (unknown)
Insert proximal test cassette and start pumping
- Test point
- Line A
- Expected
- PUMPING message displayed, LED above Line A flashes
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Proximal occlusion not detected patient safety
Line A tubing clamped proximal to cassette but PROXOCCL A/AIR alarm does not sound
Ranked causes
- 1. Proximal occlusion detection sensor malfunction (unknown)
Verify pumping starts with Line A at 400 mL/hr, VTBI 50 mL
- Test point
- Line A LED and status bar
- Expected
- LED above Line A flashes, PUMPING message displayed
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Stop delivery, then turn off patient safety
Operator attempted to turn off infuser while a delivery is in progress.
Ranked causes
- 1. Operator error - incorrect shutdown sequence (common)
Verify all infusion lines stopped
- Test point
- Infusion pump channels
- Expected
- All channels in stop state; no flow occurring
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Warning: Charger Service caution
Battery charging circuit hardware problem detected; charging circuitry not behaving as expected.
Ranked causes
- 1. Battery charging circuit malfunction (common)
Verify AC adapter connection
- Test point
- Power connector
- Expected
- Secure connection at charging port
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Warning: Low Battery caution
Infuser battery is discharged with approximately 30 minutes of battery life remaining; warning message displayed on LCD screen.
Ranked causes
- 1. Battery depleted during operation (common)
Verify AC power availability
- Test point
- Power outlet
- Expected
- AC power available and functional
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Warning: Replace Battery caution
Battery voltage has dropped below depleted threshold while charge level is above low charge threshold, indicating battery service is needed.
Ranked causes
- 1. Battery has reached end of service life (common)
Verify battery replacement procedure
- Test point
- Battery compartment
- Expected
- Old battery removed; new battery installed per section 7.2.4
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Wireless connection not appearing on main delivery screen
Wireless Connection Available icon does not appear on the main delivery screen despite device being in wireless network environment
Ranked causes
- 1. Device not receiving wireless signal or network access point not recognized (unknown)
Verify device is receiving wireless signal and network access point is recognized
- Test point
- Main delivery screen
- Expected
- Wireless Connection Available icon displayed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Wireless connection not available
Wireless Connection Available icon does not appear on main delivery screen; device is not receiving wireless signal
Ranked causes
- 1. Wireless network access point not recognized or signal not received (common)
Verify Wireless Connection Available icon appears on main delivery screen
- Test point
- Main delivery screen display
- Expected
- Wireless Connection Available icon is displayed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Charger Service Warning | Battery charging hardware circuit malfunction detected; charger not functioning properly | Charging circuitry hardware detects abnormal behavior or failure condition | warning |
E160 | Line B VTBI complete | Programmed VTBI completed on line B | error |
E161 | Line A VTBI complete | Programmed VTBI completed on line A | error |
E180 | Distal occlusion detected during non-delivery | Peak distal occlusion, non-delivery | error |
E181 | Distal occlusion detected during non-delivery | Negative distal occlusion, non-delivery | error |
E182 | Proximal occlusion B detected during non-delivery | Negative proximal occlusion B, non-delivery | error |
E183 | Proximal occlusion B detected during non-delivery | Peak proximal occlusion B, non-delivery | error |
E184 | Proximal occlusion A detected during non-delivery | Negative proximal occlusion A, non-delivery | error |
E185 | Proximal occlusion A detected during non-delivery | Peak proximal occlusion A, non-delivery | error |
E186 | Distal occlusion detected during delivery | Peak distal occlusion, delivery | error |
E187 | Distal occlusion detected during delivery | Negative distal occlusion, delivery | error |
E188 | Proximal occlusion B detected during delivery | Negative proximal occlusion B, delivery | error |
E189 | Proximal occlusion B detected during delivery | Peak proximal occlusion B, delivery | error |
E190 | Proximal occlusion A detected during delivery | Negative proximal occlusion A, delivery | error |
E191 | Proximal occlusion A detected during delivery | Peak proximal occlusion A, delivery | error |
E230 | Proximal air-in-line total detected | 500 microL of air has entered the cassette | error |
E231 | Proximal air-in-line on line B detected | 500 microL of air has entered the cassette on line B | error |
E232 | Proximal air-in-line on line A detected | 500 microL of air has entered the cassette on line A | error |
E233 | Distal air cumulative detected | 500 microL of air detected in the last 5.3 mL of fluid delivered | error |
E234 | Distal air-in-line detected (older firmware version) | Air detected in distal line before 1 mL fluid delivered on older Plum A+ software versions | warning |
E250 | Door opened while pumping | Door opened while pumping | error |
E251 | Valve/cassette test failure | Valve/cassette fails the leak test | error |
E252 | Depleted battery | Battery terminal voltage is less than 5.45 V | error |
E253 | Lockout violation - hard lockout | The use of [STOP] key or attempt to open door while lockout switch is locked | error |
E254 | Lockout enabled - keypad locked | Any action not resulting in stopping delivery while lockout switch is locked | error |
E300 | ADC failure | Analog to digital converter failure | error |
E301 | Audio alarm failure | Piezo is off but sensed on or Piezo is on but sensed off | error |
E302 | Backlight failure | Backlight (CCFT tube) is not at the expected range | error |
E320 | Battery charge/discharge current out of range | Current limiting circuitry in power supply board has failed or other hardware causing excessive current draw | critical |
E321 | Failure of charging current to drop below threshold | Charging current does not drop below end of charge threshold within 8 hours or below 1A within 6 hours | critical |
E322 | Battery current calibration value out of range | Battery charging current measurement is out of acceptable range | critical |
E323 | Battery trickle charge current out of range | Battery trickle charge current is out of acceptable range | critical |
E324 | Supply overvoltage | Power supply voltage exceeds safe operating range | critical |
E325 | Battery overvoltage | Battery voltage exceeds acceptable range | critical |
E326 | Battery disconnected or voltage too low | Connectors not seated on battery terminals or battery voltage critically low while infuser is on | critical |
E340 | Critical instruction failure | Power-up CPU register test failed | critical |
E341 | Critical data memory failure | Critical data memory failure detected | critical |
E342 | Display failure | Display assembly malfunction detected | critical |
E343 | Distal air sensor failure 1 | With cassette removed, distal air sensor self test detects liquid | error |
E344 | Distal air sensor failure 2 | With cassette inserted, distal air sensor self test detects sensor out of range | error |
E345 | Distal pressure sensor failure 1 | Distal pressure sensor failed while infuser is off | error |
E346 | Distal pressure sensor failure 2 | Distal pressure sensor failed while infuser is on | error |
E347 | Hardware watchdog failure | Hardware watchdog timer failure detected | critical |
E378 | I/O valve phase loss | Generic I/O valve failure detected | error |
E379 | L/S valve phase loss | Generic L/S valve failure detected | error |
E380 | Plunger motor phase loss | Generic plunger motor failure detected | error |
E430 | Proximal air sensor failure 1 | Proximal air sensor ongoing test detects liquid with cassette removed | error |
E431 | Proximal air sensor failure 2 | Proximal air sensor self test detects liquid with cassette removed | error |
E432 | Proximal pressure sensor 1 | Proximal pressure sensor failed while infuser is off | error |
E433 | Proximal pressure sensor 2 | Proximal pressure sensor failed while infuser is on | error |
E434 | RAM failure | RAM memory failure detected | critical |
E435 | RTC failure | Real-time clock failure detected | critical |
E436 | ROM failure | ROM checksum failure detected | critical |
E437 | Software failure | Generic software failure detected | critical |
E438 | Stack out-of-range failure | Stack pointer exceeds valid memory range | critical |
E439 | Stuck key | A key is sensed as pressed for over two minutes | error |
E440 | Power hold stuck | Power hold signal stuck in active state | critical |
E443 | LCD failure | LCD bias is out of acceptable range | error |
E444 | CPU timebase inaccurate | CPU timer 2 and RTC measured times disagree | error |
E445 | RTC memory failure | Real-time clock memory is corrupted | error |
E446 | CPU timer failure | CPU timer 1 and timer 2 measured times disagree | error |
E447 | Battery ADC reading failure | 16 consecutive battery ADC readings are all zero or max value | error |
E448 | SEEP write failure | SEEP data write operation failed | critical |
E449 | SEEP calibration data corrupted | Calibration data block in SEEP is corrupted | critical |
E450 | MMIO port read/write failure | I/O port read or write operation failed | critical |
E451 | Inaccurate delivery | Over-delivery or under-delivery detected during operation | warning |
E452 | Software failure | Miscellaneous software failures not categorized elsewhere | critical |
E453 | Two SEEP CRC errors | NVRAM data block CRC validation failed twice | critical |
E454 | NVRAM over capacity | Software attempting to write into non-existent NVRAM space | critical |
E455 | Invalid device configuration | Incorrect flash memory detected on peripheral PWA | critical |
E456 | Invalid drug library | Drug library installation started but not completed successfully | error |
E457 | Drug library corrupted | CRC failure detected on drug library data | error |
Low Battery Warning | Battery charge critically low with approximately 30 minutes remaining before power loss | Battery voltage drops to threshold indicating insufficient charge for extended operation | warning |
N100 | Unrecognizable cassette | Incorrect cassette type inserted | warning |
N101 | No action - operator idle during delivery parameters entry | No operator action and no delivery for two minutes during delivery parameters entry | warning |
N102 | Infuser idle 2 minutes | Infuser in reset or idle for over two minutes | warning |
N103 | NV RAM lost therapy data | Therapy data is lost | error |
N104 | Nurse callback B | Delivery line B has changed (if alarm is enabled) | info |
N105 | Nurse callback A | Delivery line A has changed (if alarm is enabled) | info |
N160 | Line B VTBI complete | Programmed VTBI completed on line B | info |
N161 | Line A VTBI complete | Programmed VTBI completed on line A | info |
N180 | Distal occlusion detected during non-delivery | Peak distal occlusion, non-delivery | warning |
N181 | Distal occlusion detected during non-delivery | Negative distal occlusion, non-delivery | warning |
N182 | Proximal occlusion B detected during non-delivery | Negative proximal occlusion B, non-delivery | warning |
N183 | Proximal occlusion B detected during non-delivery | Peak proximal occlusion B, non-delivery | warning |
N184 | Proximal occlusion A detected during non-delivery | Negative proximal occlusion A, non-delivery | warning |
N185 | Proximal occlusion A detected during non-delivery | Peak proximal occlusion A, non-delivery | warning |
N186 | Distal occlusion detected during delivery | Peak distal occlusion, delivery | warning |
N187 | Distal occlusion detected during delivery | Negative distal occlusion, delivery | warning |
N188 | Proximal occlusion B detected during delivery | Negative proximal occlusion B, delivery | warning |
N189 | Proximal occlusion B detected during delivery | Peak proximal occlusion B, delivery | warning |
N190 | Proximal occlusion A detected during delivery | Negative proximal occlusion A, delivery | warning |
N191 | Proximal occlusion A detected during delivery | Peak proximal occlusion A, delivery | warning |
N230 | Proximal air-in-line total detected | 500 microL of air has entered the cassette | warning |
N231 | Proximal air-in-line on line B detected | 500 microL of air has entered the cassette on line B | warning |
N232 | Proximal air-in-line detected with backpriming required | Air detected in proximal line before 1 mL fluid delivered during proximal air-in-line test | warning |
N233 | Distal air cumulative detected | 500 microL of air detected in the last 5.3 mL of fluid delivered | warning |
N234 | Distal air-in-line detected | Air detected in distal line before 1 mL fluid delivered during distal air-in-line test | warning |
N250 | Door opened while pumping | Door opened while pumping | warning |
N251 | Valve/cassette test failure | Valve/cassette fails the leak test | warning |
N252 | Depleted battery | Battery terminal voltage is less than 5.45 V | warning |
N253 | Lockout violation - hard lockout | The use of [STOP] key or attempt to open door while lockout switch is locked | warning |
N254 | Lockout enabled - keypad locked | Any action not resulting in stopping delivery while lockout switch is locked | warning |
N255 | Lockout violation - soft lockout | The use of [STOP] key or attempt to open door while lockout switch is locked | warning |
N256 | Lockout enabled - soft lockout | Any action not resulting in stopping delivery while lockout switch is locked | warning |
Replace Battery Warning | Battery service or replacement required; voltage below depletion threshold | Battery voltage drops below minimum serviceable voltage while charge level above low threshold | warning |
Stop Delivery First | Shutdown blocked; infuser cannot power down during active delivery | Power off command received while infusion pump is actively delivering fluid to patient | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.