Hospira Plum A+3: troubleshooting and error codes

The ServiceForge knowledge base holds 12 documented symptoms for the Hospira Plum A+3, with 26 ranked causes, 12 diagnostic checks with test points and expected values, and 80 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Distal occlusion test failure patient safety

Infuser does not detect or respond to distal line occlusion

Ranked causes

  1. 1. Cassette not properly primed (common)
  2. 2. Dirty sensor pin (occasional)
  3. 3. Defective cassette (occasional)
  4. 4. Defective APP PWA (rare)
First check

Distal occlusion test per Section 5.3.13

Test point
Distal sensor detection
Expected
Occlusion detected and alarm triggered

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Proximal occlusion test failure patient safety

Infuser does not detect or respond to proximal line occlusion

Ranked causes

  1. 1. Closed proximal clamp (common)
  2. 2. Cassette not properly primed (common)
  3. 3. Dirty sensor pin (occasional)
  4. 4. Defective APP PWA (rare)
First check

Proximal occlusion test per Section 5.3.10

Test point
Proximal sensor detection
Expected
Occlusion detected and alarm triggered

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air-in-line test failure patient safety

Infuser does not detect air bubbles in proximal or distal line

Ranked causes

  1. 1. Defective special cassette (occasional)
  2. 2. Dirty sensors (occasional)
  3. 3. Defective APP PWA (rare)
First check

Air-in-line tests per Section 5.3.11 and 5.3.12

Test point
Air detection sensors
Expected
Air bubbles detected

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm loudness test failure patient safety

Alarm does not sound or is too quiet

Ranked causes

  1. 1. Defective CPU PWA (rare)
  2. 2. Defective peripheral interface PWA (rare)
  3. 3. Defective piezo alarm assembly (occasional)
First check

Alarm loudness test per Section 5.3.8

Test point
Piezo speaker output
Expected
Alarm sounds at specified loudness

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Delivery accuracy test failure patient safety

Infuser delivery volume or rate measurement outside specification

Ranked causes

  1. 1. Cassette not properly primed (common)
  2. 2. Damaged or faulty cassette (occasional)
  3. 3. Defective mechanism assembly (rare)
First check

Run delivery accuracy test per Section 5.3.14

Test point
Fluid delivery measurement
Expected
Volume and rate within specification

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cassette alarm test failure patient safety

Cassette alarm functionality test does not pass

Ranked causes

  1. 1. Cassette not properly seated (common)
  2. 2. Defective cassette (occasional)
First check

Run cassette alarm test per Section 5.3.4

Test point
Infuser display
Expected
Alarm test pass

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Self test failure patient safety

Infuser self test does not complete successfully on power-on

Ranked causes

  1. 1. Cassette not properly installed (common)
  2. 2. Defective CPU PWA (rare)
First check

Perform self test from Section 5.3.3

Test point
Infuser display
Expected
Self test pass indication

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display test failure caution

Display does not show text or segments, or display is dark or unreadable

Ranked causes

  1. 1. Defective display assembly (occasional)
First check

Display test per Section 5.3.6

Test point
Display screen
Expected
All pixels or segments illuminated

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Electrical safety test failure high voltage

AC power cord insulation or grounding test fails

Ranked causes

  1. 1. Defective AC power cord (occasional)
First check

Electrical safety test per Section 5.3.16

Test point
AC power cord
Expected
Insulation and grounding within specification

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keypad lockout switch test failure caution

Keypad lockout function does not operate

Ranked causes

  1. 1. Defective peripheral interface PWA (rare)
First check

Keypad lockout test per Section 5.3.9

Test point
Lockout switch and keypad
Expected
Keys locked when switch engaged

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keypad test failure caution

Keypad keys do not register or functional test fails

Ranked causes

  1. 1. Defective keypad (occasional)
First check

Keypad functional test per Section 5.3.7

Test point
All keypad keys
Expected
All keys register input

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unrestricted flow detected patient safety

Infuser delivers fluid without occlusion control; life-threatening condition

Ranked causes

  1. 1. Regulator closer not seated correctly (common)
First check

Visual inspection of regulator closer position

Test point
Regulator closer seat in mechanism
Expected
Regulator closer completely seated

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
E300Analog to digital converter failureADC conversion fails during operationerror
E301Audio alarm piezo failurePiezo off but sensed on, or piezo on but sensed offerror
E302Backlight CCFT tube failureCCFT tube not at expected rangeerror
E320Battery charge/discharge current out of rangeCurrent limiting circuitry fails or excessive current draw detectederror
E321Battery charging current fails to drop below thresholdCharging current not below threshold in 8 hours or 1A in 6 hourserror
E322Battery current calibration value out of rangeCharge current measurement circuitry failserror
E323Battery trickle charge current out of rangeTrickle charge current measurement failserror
E324Supply overvoltage conditionPower supply voltage exceeds acceptable rangecritical
E325Battery overvoltage conditionBattery voltage exceeds acceptable rangecritical
E326Battery disconnected while on or voltage too lowBattery removed during operation or voltage critically lowcritical
E340Critical CPU register self-test failurePower-up CPU register test failscritical
E341Critical data memory failureCritical data memory corruption or failure during self-testcritical
E342Display assembly failureDefective display detectederror
E343Distal air sensor detects liquid without cassetteWith cassette removed, self-test detects liquiderror
E344Distal air sensor out of range with cassetteWith cassette inserted, self-test detects out of rangeerror
E345Distal pressure sensor failure while offDistal pressure sensor failed during off-state testerror
E346Distal pressure sensor failure while onDistal pressure sensor failed during on-state testerror
E347Hardware watchdog timer failureWatchdog timer fails during self-testcritical
E378I/O valve phase loss detectedGeneric I/O valve failure or phase mismatcherror
E379L/S valve phase loss detectedGeneric L/S valve failure or phase mismatcherror
E380Plunger motor phase loss detectedGeneric plunger motor failureerror
E430Proximal air sensor detects liquid without cassetteOngoing test detects liquid with cassette removederror
E431Proximal air sensor self-test detects liquidSelf-test detects liquid with cassette removederror
E432Proximal pressure sensor failure while offSensor failed during off-state testerror
E433Proximal pressure sensor failure while onSensor failed during on-state testerror
E434RAM memory failureRAM checksum or access failure during testcritical
E435Real-time clock failureRTC circuit fails during self-testerror
E436ROM checksum failureROM data checksum verification failscritical
E437Generic software failureSoftware error detected during operationerror
E437-13Processor reset while infuser in ON stateProcessor reset or restarted while device operatingcritical
E438Stack overflow or underflow conditionStack pointer exceeds valid rangecritical
E439Keypad key stuckKey sensed as pressed for over two minuteserror
E440Power hold signal stuckPower hold signal cannot turn device offcritical
E443LCD display bias out of rangeLCD bias voltage fails during self-testerror
E444CPU timebase accuracy failureCPU timer 2 and RTC measured times disagreeerror
E445RTC memory corruptionReal-time clock memory data corruptederror
E446CPU timer mismatch failureCPU timer 1 and timer 2 measured times disagreeerror
E447Battery ADC reading stuck at extreme16 consecutive readings all zero or maximum valueerror
E448SEEP data write failureSEEP nonvolatile memory write operation failserror
E449SEEP calibration data corruptedCalibration data block CRC check failserror
E450Memory-mapped I/O port read/write failureI/O port access fails during testcritical
E451Over or under delivery detectedInfusion delivery rate significantly deviates from programmederror
E452Miscellaneous software failureGeneric software error detectederror
E453Two SEEP CRC errors detectedNVRAM data block checksum verification fails twicecritical
E454NVRAM storage capacity exceededSoftware attempts to write to non-existent NVRAM spaceerror
E455Invalid flash memory configurationIncorrect flash memory detected on peripheral PWAerror
E456Drug library installation failedDrug library install started but not completederror
E457Drug library data corruptedDrug library CRC verification failserror
N100Unrecognizable cassetteIncorrect cassette is insertedwarning
N101No actionNo operator action and no delivery for two minutes during delivery parameters entrywarning
N102Infuser idle 2 minutesInfuser in reset or idle for over two minutesinfo
N103NV RAM lost thpy dataInfuser did not complete the previous non-volatile memory write successfullyerror
N104Nurse callback BEnd of delivery step on line B other than VTBI complete while callback is enabledwarning
N105Nurse callback AEnd of delivery step on line A other than VTBI complete while callback is enabledwarning
N160 or E160Line B VTBI completeProgrammed VTBI completed on line Binfo
N161 or E161Line A VTBI completeProgrammed VTBI completed on line Ainfo
N180 or E180Distal OcclPeak distal occlusion detected during non-deliverywarning
N181 or E181Distal OcclNegative distal occlusion detected during non-deliverywarning
N182 or E182Prox. Occl B, AirNegative proximal occlusion B detected during non-deliverywarning
N183 or E183Prox. Occl B, AirPeak proximal occlusion B detected during non-deliverywarning
N184 or E184Prox. Occl A, AirNegative proximal occlusion A detected during non-deliverywarning
N185 or E185Prox. Occl A, AirPeak proximal occlusion A detected during non-deliverywarning
N186 or E186Distal OcclPeak distal occlusion detected during deliveryerror
N187 or E187Distal OcclNegative distal occlusion detected during deliveryerror
N188 or E188Prox. Occl B, AirNegative proximal occlusion B detected during deliveryerror
N189 or E189Prox. Occl B, AirPeak proximal occlusion B detected during deliveryerror
N190 or E190Prox. Occl A, AirNegative proximal occlusion A detected during deliveryerror
N191 or E191Prox. Occl A, AirPeak proximal occlusion A detected during deliveryerror
N230 or E230Prox. Air total500 µL of air has entered the cassettewarning
N231 or E231Prox. Air on B, backprime500 µL of air has entered the cassette on line Bwarning
N232 or E232Prox. Air on A, Backprime500 µL of air has entered the cassette on line Awarning
N233 or E233Distal air cumulative500 µL of air detected in the last 5.3 mL of fluid deliveredwarning
N234 or E234Distal air bolus100 µL bolus of air detected at distal sensorwarning
N250 or E250Door opened while pumpingDoor opened while pumpingerror
N251 or E251Valve/cass test failValve/cassette fails leak testerror
N252 or E252Depleted batteryBattery terminal voltage is less than 5.45 Vwarning
N253 or E253Lockout ViolationUse of the [STOP] key or door open attempt while lockout switch is lockedwarning
N254 or E254Lockout EnabledAny action not resulting in stopping of delivery while lockout switch is lockedwarning
N255Lockout violationUse of the [STOP] key or door open attempt while lockout switch is lockedwarning
N256Lockout enabledAny action not resulting in stopping of delivery while lockout switch is lockedwarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI