Hospira Plum A+3: troubleshooting and error codes
The ServiceForge knowledge base holds 12 documented symptoms for the Hospira Plum A+3, with 26 ranked causes, 12 diagnostic checks with test points and expected values, and 80 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Distal occlusion test failure patient safety
Infuser does not detect or respond to distal line occlusion
Ranked causes
- 1. Cassette not properly primed (common)
- 2. Dirty sensor pin (occasional)
- 3. Defective cassette (occasional)
- 4. Defective APP PWA (rare)
Distal occlusion test per Section 5.3.13
- Test point
- Distal sensor detection
- Expected
- Occlusion detected and alarm triggered
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Proximal occlusion test failure patient safety
Infuser does not detect or respond to proximal line occlusion
Ranked causes
- 1. Closed proximal clamp (common)
- 2. Cassette not properly primed (common)
- 3. Dirty sensor pin (occasional)
- 4. Defective APP PWA (rare)
Proximal occlusion test per Section 5.3.10
- Test point
- Proximal sensor detection
- Expected
- Occlusion detected and alarm triggered
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air-in-line test failure patient safety
Infuser does not detect air bubbles in proximal or distal line
Ranked causes
- 1. Defective special cassette (occasional)
- 2. Dirty sensors (occasional)
- 3. Defective APP PWA (rare)
Air-in-line tests per Section 5.3.11 and 5.3.12
- Test point
- Air detection sensors
- Expected
- Air bubbles detected
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm loudness test failure patient safety
Alarm does not sound or is too quiet
Ranked causes
- 1. Defective CPU PWA (rare)
- 2. Defective peripheral interface PWA (rare)
- 3. Defective piezo alarm assembly (occasional)
Alarm loudness test per Section 5.3.8
- Test point
- Piezo speaker output
- Expected
- Alarm sounds at specified loudness
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Delivery accuracy test failure patient safety
Infuser delivery volume or rate measurement outside specification
Ranked causes
- 1. Cassette not properly primed (common)
- 2. Damaged or faulty cassette (occasional)
- 3. Defective mechanism assembly (rare)
Run delivery accuracy test per Section 5.3.14
- Test point
- Fluid delivery measurement
- Expected
- Volume and rate within specification
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cassette alarm test failure patient safety
Cassette alarm functionality test does not pass
Ranked causes
- 1. Cassette not properly seated (common)
- 2. Defective cassette (occasional)
Run cassette alarm test per Section 5.3.4
- Test point
- Infuser display
- Expected
- Alarm test pass
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Self test failure patient safety
Infuser self test does not complete successfully on power-on
Ranked causes
- 1. Cassette not properly installed (common)
- 2. Defective CPU PWA (rare)
Perform self test from Section 5.3.3
- Test point
- Infuser display
- Expected
- Self test pass indication
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display test failure caution
Display does not show text or segments, or display is dark or unreadable
Ranked causes
- 1. Defective display assembly (occasional)
Display test per Section 5.3.6
- Test point
- Display screen
- Expected
- All pixels or segments illuminated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrical safety test failure high voltage
AC power cord insulation or grounding test fails
Ranked causes
- 1. Defective AC power cord (occasional)
Electrical safety test per Section 5.3.16
- Test point
- AC power cord
- Expected
- Insulation and grounding within specification
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keypad lockout switch test failure caution
Keypad lockout function does not operate
Ranked causes
- 1. Defective peripheral interface PWA (rare)
Keypad lockout test per Section 5.3.9
- Test point
- Lockout switch and keypad
- Expected
- Keys locked when switch engaged
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keypad test failure caution
Keypad keys do not register or functional test fails
Ranked causes
- 1. Defective keypad (occasional)
Keypad functional test per Section 5.3.7
- Test point
- All keypad keys
- Expected
- All keys register input
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unrestricted flow detected patient safety
Infuser delivers fluid without occlusion control; life-threatening condition
Ranked causes
- 1. Regulator closer not seated correctly (common)
Visual inspection of regulator closer position
- Test point
- Regulator closer seat in mechanism
- Expected
- Regulator closer completely seated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
E300 | Analog to digital converter failure | ADC conversion fails during operation | error |
E301 | Audio alarm piezo failure | Piezo off but sensed on, or piezo on but sensed off | error |
E302 | Backlight CCFT tube failure | CCFT tube not at expected range | error |
E320 | Battery charge/discharge current out of range | Current limiting circuitry fails or excessive current draw detected | error |
E321 | Battery charging current fails to drop below threshold | Charging current not below threshold in 8 hours or 1A in 6 hours | error |
E322 | Battery current calibration value out of range | Charge current measurement circuitry fails | error |
E323 | Battery trickle charge current out of range | Trickle charge current measurement fails | error |
E324 | Supply overvoltage condition | Power supply voltage exceeds acceptable range | critical |
E325 | Battery overvoltage condition | Battery voltage exceeds acceptable range | critical |
E326 | Battery disconnected while on or voltage too low | Battery removed during operation or voltage critically low | critical |
E340 | Critical CPU register self-test failure | Power-up CPU register test fails | critical |
E341 | Critical data memory failure | Critical data memory corruption or failure during self-test | critical |
E342 | Display assembly failure | Defective display detected | error |
E343 | Distal air sensor detects liquid without cassette | With cassette removed, self-test detects liquid | error |
E344 | Distal air sensor out of range with cassette | With cassette inserted, self-test detects out of range | error |
E345 | Distal pressure sensor failure while off | Distal pressure sensor failed during off-state test | error |
E346 | Distal pressure sensor failure while on | Distal pressure sensor failed during on-state test | error |
E347 | Hardware watchdog timer failure | Watchdog timer fails during self-test | critical |
E378 | I/O valve phase loss detected | Generic I/O valve failure or phase mismatch | error |
E379 | L/S valve phase loss detected | Generic L/S valve failure or phase mismatch | error |
E380 | Plunger motor phase loss detected | Generic plunger motor failure | error |
E430 | Proximal air sensor detects liquid without cassette | Ongoing test detects liquid with cassette removed | error |
E431 | Proximal air sensor self-test detects liquid | Self-test detects liquid with cassette removed | error |
E432 | Proximal pressure sensor failure while off | Sensor failed during off-state test | error |
E433 | Proximal pressure sensor failure while on | Sensor failed during on-state test | error |
E434 | RAM memory failure | RAM checksum or access failure during test | critical |
E435 | Real-time clock failure | RTC circuit fails during self-test | error |
E436 | ROM checksum failure | ROM data checksum verification fails | critical |
E437 | Generic software failure | Software error detected during operation | error |
E437-13 | Processor reset while infuser in ON state | Processor reset or restarted while device operating | critical |
E438 | Stack overflow or underflow condition | Stack pointer exceeds valid range | critical |
E439 | Keypad key stuck | Key sensed as pressed for over two minutes | error |
E440 | Power hold signal stuck | Power hold signal cannot turn device off | critical |
E443 | LCD display bias out of range | LCD bias voltage fails during self-test | error |
E444 | CPU timebase accuracy failure | CPU timer 2 and RTC measured times disagree | error |
E445 | RTC memory corruption | Real-time clock memory data corrupted | error |
E446 | CPU timer mismatch failure | CPU timer 1 and timer 2 measured times disagree | error |
E447 | Battery ADC reading stuck at extreme | 16 consecutive readings all zero or maximum value | error |
E448 | SEEP data write failure | SEEP nonvolatile memory write operation fails | error |
E449 | SEEP calibration data corrupted | Calibration data block CRC check fails | error |
E450 | Memory-mapped I/O port read/write failure | I/O port access fails during test | critical |
E451 | Over or under delivery detected | Infusion delivery rate significantly deviates from programmed | error |
E452 | Miscellaneous software failure | Generic software error detected | error |
E453 | Two SEEP CRC errors detected | NVRAM data block checksum verification fails twice | critical |
E454 | NVRAM storage capacity exceeded | Software attempts to write to non-existent NVRAM space | error |
E455 | Invalid flash memory configuration | Incorrect flash memory detected on peripheral PWA | error |
E456 | Drug library installation failed | Drug library install started but not completed | error |
E457 | Drug library data corrupted | Drug library CRC verification fails | error |
N100 | Unrecognizable cassette | Incorrect cassette is inserted | warning |
N101 | No action | No operator action and no delivery for two minutes during delivery parameters entry | warning |
N102 | Infuser idle 2 minutes | Infuser in reset or idle for over two minutes | info |
N103 | NV RAM lost thpy data | Infuser did not complete the previous non-volatile memory write successfully | error |
N104 | Nurse callback B | End of delivery step on line B other than VTBI complete while callback is enabled | warning |
N105 | Nurse callback A | End of delivery step on line A other than VTBI complete while callback is enabled | warning |
N160 or E160 | Line B VTBI complete | Programmed VTBI completed on line B | info |
N161 or E161 | Line A VTBI complete | Programmed VTBI completed on line A | info |
N180 or E180 | Distal Occl | Peak distal occlusion detected during non-delivery | warning |
N181 or E181 | Distal Occl | Negative distal occlusion detected during non-delivery | warning |
N182 or E182 | Prox. Occl B, Air | Negative proximal occlusion B detected during non-delivery | warning |
N183 or E183 | Prox. Occl B, Air | Peak proximal occlusion B detected during non-delivery | warning |
N184 or E184 | Prox. Occl A, Air | Negative proximal occlusion A detected during non-delivery | warning |
N185 or E185 | Prox. Occl A, Air | Peak proximal occlusion A detected during non-delivery | warning |
N186 or E186 | Distal Occl | Peak distal occlusion detected during delivery | error |
N187 or E187 | Distal Occl | Negative distal occlusion detected during delivery | error |
N188 or E188 | Prox. Occl B, Air | Negative proximal occlusion B detected during delivery | error |
N189 or E189 | Prox. Occl B, Air | Peak proximal occlusion B detected during delivery | error |
N190 or E190 | Prox. Occl A, Air | Negative proximal occlusion A detected during delivery | error |
N191 or E191 | Prox. Occl A, Air | Peak proximal occlusion A detected during delivery | error |
N230 or E230 | Prox. Air total | 500 µL of air has entered the cassette | warning |
N231 or E231 | Prox. Air on B, backprime | 500 µL of air has entered the cassette on line B | warning |
N232 or E232 | Prox. Air on A, Backprime | 500 µL of air has entered the cassette on line A | warning |
N233 or E233 | Distal air cumulative | 500 µL of air detected in the last 5.3 mL of fluid delivered | warning |
N234 or E234 | Distal air bolus | 100 µL bolus of air detected at distal sensor | warning |
N250 or E250 | Door opened while pumping | Door opened while pumping | error |
N251 or E251 | Valve/cass test fail | Valve/cassette fails leak test | error |
N252 or E252 | Depleted battery | Battery terminal voltage is less than 5.45 V | warning |
N253 or E253 | Lockout Violation | Use of the [STOP] key or door open attempt while lockout switch is locked | warning |
N254 or E254 | Lockout Enabled | Any action not resulting in stopping of delivery while lockout switch is locked | warning |
N255 | Lockout violation | Use of the [STOP] key or door open attempt while lockout switch is locked | warning |
N256 | Lockout enabled | Any action not resulting in stopping of delivery while lockout switch is locked | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.