Hospira LifeCare PCA3: troubleshooting and error codes
The ServiceForge knowledge base holds 41 documented symptoms for the Hospira LifeCare PCA3, with 70 ranked causes, 18 diagnostic checks with test points and expected values, and 102 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Self test or module test failure caution
Self test or service mode tests fail indicating component defect
Ranked causes
- 1. MCU PWA defective (common)
- 2. LCD assembly defective (occasional)
- 3. Numeric display PWA defective (occasional)
- 4. Icon display PWA defective (occasional)
- 5. Keypad defective (occasional)
- 6. Door lock assembly defective (occasional)
- 7. Mechanism assembly defective (occasional)
- 8. Patient pendant assembly defective (rare)
Turn pump off and on, allow self test to complete
- Expected
- Self test passes without errors
12 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Empty vial sensor test failure caution
Empty vial detection sensor fails self test diagnostic
Ranked causes
- 1. Vial sensor switch misaligned (common)
- 2. Defective mechanism assembly (common)
- 3. Defective MCU PWA (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Icon display indicator test failure caution
Icon display PWA fails self test diagnostic
Ranked causes
- 1. Defective icon display PWA (common)
- 2. Icon display PWA connection fault (common)
- 3. Defective MCU PWA (occasional)
Verify icon display connection
- Test point
- Icon display connector to MCU PWA
- Expected
- Connection fully seated and secure
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
LCD backlight test failure caution
LCD backlight circuit or lamp fails self test
Ranked causes
- 1. Defective LCD assembly backlight (common)
- 2. LCD assembly connection fault (common)
- 3. Defective MCU PWA (occasional)
Verify LCD assembly connection
- Test point
- LCD connector to MCU PWA
- Expected
- Connection fully seated and secure
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
LCD display test failure caution
LCD assembly or display electronics fails self test
Ranked causes
- 1. Defective LCD assembly (common)
- 2. LCD assembly connection fault (common)
- 3. Defective MCU PWA (occasional)
Verify LCD assembly connection
- Test point
- LCD connector to MCU PWA
- Expected
- Connection fully seated and secure
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
LED numeric display test failure caution
LED numeric display PWA fails self test
Ranked causes
- 1. Defective numeric display PWA (common)
- 2. Numeric display PWA connection fault (common)
- 3. Defective MCU PWA (occasional)
Verify numeric display PWA connection
- Test point
- Display connector to MCU PWA
- Expected
- Connection fully seated and secure
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Occlusion switch error patient safety
Defective occlusion sensor fails to detect line blockage
Ranked causes
- 1. Occlusion alarm switch maladjustment (common)
- 2. Mechanism assembly failure (occasional)
- 3. MCU PWA malfunction (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Occlusion test failure patient safety
Occlusion detection system fails diagnostic test
Ranked causes
- 1. Infusion set not properly primed or installed (common)
- 2. Occlusion alarm switch misaligned (common)
- 3. Defective mechanism assembly (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Screw rotation error caution
Motor steps and screw rotations not in proportion, possible lubrication or mechanism issue
Ranked causes
- 1. Lead screw lubrication inadequate (common)
- 2. Mechanism assembly failure (occasional)
- 3. MCU PWA malfunction (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Syringe mechanism test failure caution
Syringe drive mechanism fails self test diagnostic
Ranked causes
- 1. Vial sensor switch misaligned (common)
- 2. Defective mechanism assembly (common)
- 3. Defective MCU PWA (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Delivery accuracy test failure patient safety
Pump delivery volume or rate does not meet accuracy specifications
Ranked causes
- 1. PCA set not properly primed (common)
- 2. Mechanism assembly defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Door lock test failure caution
Door lock assembly fails self test diagnostic
Ranked causes
- 1. Defective door lock assembly (common)
- 2. Defective MCU PWA (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keypad test failure caution
Pump keypad or keypad interface fails self test
Ranked causes
- 1. Defective keypad (common)
- 2. Defective MCU PWA (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Self test fails during verification patient safety
Device fails self-test verification when powered on, with alarm message displayed and system not proceeding to SELF TEST COMPLETE screen.
Ranked causes
- 1. Software or memory integrity failure (unknown)
- 2. Electronic function malfunction (unknown)
Verify AC power connection
- Test point
- AC power indicator on front panel
- Expected
- AC power indicator illuminated
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm circuit test failure caution
Alarm speaker or alarm circuit fails self test
Ranked causes
- 1. Defective MCU PWA (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery fault caution
Battery malfunction: open circuit, over-discharge, or short circuit
Ranked causes
- 1. Battery defective (open, over-discharged, or shorted) (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Injector sensor switch caution
Injector sensor switch requires positioning and varnish sealing to ensure proper detection
Ranked causes
- 1. Injector sensor switch misalignment (common)
Verify switch bracket position parallel to surface
- Test point
- Switch bracket
- Expected
- Parallel to surface, 1/32 inch from front enclosure
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Lead screw lubrication caution
Lead screw requires cleaning and lubrication with thin coating of grease
Ranked causes
- 1. Lead screw dry or contaminated with old grease and residue (common)
Remove old grease and residue from lead screw
- Test point
- Lead screw threads
- Expected
- All old grease and residue completely removed
- Tools
- Isopropyl alcohol, cotton swabs or small brush
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Occlusion alarm switch and pressure patient safety
Occlusion alarm switch must trigger at 15±5 psi to ensure proper line blockage detection
Ranked causes
- 1. Occlusion alarm switch pressure out of specification (common)
Attach DPM to patient line
- Test point
- End of patient line via three-way stopcock
- Expected
- DPM connected and zeroed
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Vial sensor switch caution
Vial sensor switch requires adjustment to detect vial presence correctly
Ranked causes
- 1. Vial sensor switch misalignment (common)
Move switch up until it clicks off
- Test point
- Vial sensor switch
- Expected
- Audible click-off
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Charger fault caution
Battery charger is faulty and cannot charge battery
Ranked causes
- 1. Charger failure (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CPU test failure caution
Central processing unit fails self test diagnostic
Ranked causes
- 1. Defective CPU (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Door improperly secured
Door secured on wrong screen or no operator input for two minutes
Ranked causes
- 1. Door secured on wrong screen or timeout (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Door left open
Door is left open or unlocked for more than two minutes
Ranked causes
- 1. Door not secured (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrical safety test failure high voltage
AC power cord or electrical safety components defective
Ranked causes
- 1. Defective AC power cord (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Infusion paused
Infusion stopped by operator pressing START/PAUSE button
Ranked causes
- 1. Operator pressed START/PAUSE button (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Injector not detected patient safety
Vial is detected but injector is not properly seated in cradle assembly
Ranked causes
- 1. Improper injector seating (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery patient safety
Operating on battery power with less than 30 minutes of battery life remaining
Ranked causes
- 1. Battery capacity depleted (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery power during operation patient safety
Battery indicator flashing on front panel while operating on battery power, indicating insufficient battery capacity for reliable device operation.
Ranked causes
- 1. Battery requires recharge (common)
Verify AC power connection and indicator status
- Test point
- AC power indicator on front panel
- Expected
- AC power indicator illuminated indicating charging
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No battery installed in device patient safety
Device prompts BATTERY NOT INSTALLED message during startup or service mode, indicating battery compartment is empty.
Ranked causes
- 1. Battery missing from compartment (common)
Verify battery compartment access
- Test point
- Battery access cover on device
- Expected
- Battery access cover present and accessible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Occlusion detected patient safety
Tubing or delivery line is occluded, blocking fluid flow
Ranked causes
- 1. Occlusion in delivery path (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient pendant assembly test failure patient safety
Patient pendant assembly fails self test diagnostic
Ranked causes
- 1. Defective patient pendant assembly (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
RAM memory test failure caution
Pump RAM memory fails self test diagnostic
Ranked causes
- 1. Defective MCU RAM (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Real-time clock test failure caution
Real-time clock RTC circuit fails self test diagnostic
Ranked causes
- 1. Defective RTC (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ROM memory test failure caution
Pump ROM/EPROM memory fails self test diagnostic
Ranked causes
- 1. Defective EPROM (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rotation mechanism test failure caution
Syringe rotation drive mechanism fails self test
Ranked causes
- 1. Defective mechanism assembly (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Self test failure
PVT self test does not complete successfully
Ranked causes
- 1. Infusion pump self test error (common)
Power cycle the pump and run self test
- Test point
- Pump power switch
- Expected
- Self test completes without errors
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Service mode MCU test failure caution
Microcontroller unit PWA fails service mode testing
Ranked causes
- 1. Defective microcontroller PWA (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Syringe empty patient safety
Empty vial detected, no medication available for infusion
Ranked causes
- 1. Empty vial installed (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Syringe not properly loaded patient safety
Syringe (vial and injector) not properly seated in cradle assembly
Ranked causes
- 1. Improper syringe seating (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vial not properly loaded patient safety
Injector detected but vial not properly seated in cradle assembly
Ranked causes
- 1. Improper vial seating (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
170 | Interrupt Overlap | Interrupt failure | error |
171 | Invalid Interrupt | Invalid or unused interrupt | error |
172 | Stack Error | RAM error | critical |
173 | Task Dead Error | Periodic task or interrupt stopped running | critical |
174 | Kernel Error | Kernel failure | critical |
175 | Watchdog Error | MCU fails to strobe watchdog within timeout | error |
176 | Cancel Timer Error | Timer message cancelled unsuccessfully | error |
200 | Motor Overdelivery Condition | Motor running faster than expected | critical |
201 | Motor underdelivery condition | Motor running slower than expected | critical |
202 | Motor runaway | Motor running without permission | critical |
203 | Excessive rate | Programmed infusion rate exceeds maximum limit | error |
204 | Mode error | Unexpected operating mode detected | critical |
205 | Motor step time error | Motor step timing out of valid range | critical |
206 | Rate error | Unexpected infusion rate detected | critical |
207 | Delivery Volume Error | Volume infused out of range | error |
310 | 5-Volt Error | Two voltage readings outside 4.7 V - 5.3 V range | critical |
311 | Lithium Battery Error | Lithium battery voltage below minimum | error |
313 | Battery current error | Battery charge current exceeds safe limit | critical |
480 | Configuration Data | Failed CRC of pump configuration data | critical |
481 | Infusion Data | Failed CRC of infusion data | critical |
482 | Program Data | Failed CRC of program data | critical |
483 | Flash Checksum Error | Failed checksum at startup or during operation | critical |
484 | Stored Protocol Data | Failed CRC of stored protocols | critical |
485 | Motor state error | Invalid rate, restart state, or bad events in motor software | critical |
486 | History Corruption | History data corrupted | error |
487 | Motor Data Error | Failed CRC of motor data | error |
488 | Drug Table Data Error | Failed CRC of drug table data | error |
489 | User Interface Data Error | Failed CRC of saved user interface data | error |
490 | Error Log Data Error | Failed error log checksum | error |
491 | Background Data Error | Inconsistent background data | error |
492 | Switch Data Error | Corrupted switch data | error |
493 | Mirror data mismatch | Data in RAM does not match data in FLASH | critical |
493 | Mirror data mismatched - Data in RAM does not match data in FLASH | error | |
494 | UI state error | User interface state ID out of valid range | critical |
494 | UI state error - User interface state ID out of range | error | |
495 | Backlight data error | Invalid display data for backlight | critical |
496 | Vial manager data checksum error | CRC failure of vial manager data | critical |
497 | Power data error - Failed CRC of battery charging data | error | |
497 | Power data checksum error | CRC failure of battery charging data | critical |
498 | Display data error | Invalid LCD display data detected | critical |
499 | Alarm data error | Invalid alarm message data detected | critical |
550 | Power down error | Improper system shutdown detected | critical |
551 | Intertask message error | Illegal message between software tasks | critical |
552 | Invalid function parameter | Invalid parameter data passed to function | critical |
553 | Invalid message | Invalid kernel message detected | critical |
554 | Invalid mode - Invalid user interface mode | error | |
554 | Invalid mode | Invalid UI mode state detected | critical |
555 | Clock change error | Invalid date/time set detected | warning |
556 | Invalid UI state action | Invalid UI action code detected | critical |
557 | Invalid alarm - Illegal alarm message data | error | |
557 | Invalid alarm | Illegal alarm message data | critical |
558 | Mode change error | Illegal mode change message | critical |
559 | Invalid UI stack - User interface stack is corrupted or full | error | |
559 | Invalid UI stack | UI stack corrupted or full | critical |
620 | RAM address error - Address lines stuck high, stuck low, or shorted together | critical | |
620 | RAM address error | RAM address lines stuck or shorted | critical |
621 | Chip ID error | Register does not contain expected value | critical |
621 | Chip ID error - Register does not contain expected value | critical | |
622 | Clock RAM error - Data or address communication errors | error | |
622 | Clock RAM error | Data or address communication errors in clock RAM | critical |
623 | External RAM error | Failed read/write operations on external RAM | critical |
623 | External RAM error - Failed read/write operations on external RAM | error | |
624 | LCD backlight error | Backlight current out of range | critical |
625 | Battery icon LED error | Battery icon LED current out of range | critical |
626 | Audible alarm error | Alarm buzzer voltage out of valid range | critical |
627 | LCD readback error | Display data does not match RAM data | critical |
627 | LCD readback error - Display data does not match RAM data | error | |
628 | LED numeric display error | Digit segment current out of valid range | critical |
629 | Clock time error | RTC and MCU are out of synchronization | critical |
629 | Clock time error - RTC and MCU out of synchronization | error | |
630 | Screw rotation error - Motor steps and screw rotations not in proportion | error | |
630 | Screw rotation error | Motor steps and lead screw rotations not proportional | critical |
631 | LCD timeout error - LCD does not acknowledge data within hardware specific-time | error | |
631 | LCD timeout error | LCD does not acknowledge data within timeout period | critical |
633 | Print error - Failed UART loopback test | error | |
633 | Print error | Failed UART loopback test with printer | critical |
634 | ADC voltage reference error | Analog-to-digital converter voltage out of range | critical |
635 | Stuck key | Keypad button held for more than 2 minutes | warning |
635 | Stuck key error - Key stuck for more than two minutes | warning | |
636 | Motor winding excessive current | Motor winding current exceeds 200 mA limit | critical |
637 | Motor winding open circuit | Motor winding current below 75 mA minimum | critical |
638 | Motor voltage error | Motor supply voltage out of acceptable range | critical |
639 | Contrast voltage error | LCD contrast voltage out of range | critical |
640 | Internal RAM error - Failed read/write operations on internal RAM | error | |
640 | Internal RAM error | Failed read/write operations on internal RAM | critical |
641 | Occlusion switch error | Occlusion sensor malfunction detected | critical |
642 | Empty syringe switch error | Empty syringe sensor malfunction detected | critical |
643 | Injector switch error | Injector sensor malfunction detected | critical |
644 | Watchdog timer error | Watchdog timer failure | critical |
645 | Barcode reader error | Three consecutive barcode read attempts failed | critical |
645 | Barcode reader error - Three consecutive attempts to read barcode failed | error | |
Battery Fault | Battery internal fault detected (open circuit, over-discharged, or shorted) | Battery monitoring detects OB (open), OV (over-discharged), or SH (short) | warning |
Check Injector | Injector not detected when vial is present | Vial detected but injector missing | warning |
Check Settings | Door secured on wrong screen or timeout after no key press for 2 minutes | Door locked on screen other than Setup/Press Start, or inactivity timeout | warning |
Check Syringe | Syringe not properly positioned in cradle | Syringe detector sees improper loading | warning |
Check Vial | Vial not properly positioned or injector mismatch | Injector detected but vial not properly loaded | warning |
Door Open | Door has been left open or unlocked for over 2 minutes | Door open/unlocked timeout | warning |
Empty Syringe | Syringe medication reservoir is empty | Empty syringe sensor activated | warning |
Infuser Paused | Infusion delivery has been paused by operator | START/PAUSE button pressed during infusion | info |
Low Battery | Battery power remaining is critically low | Battery charge falls below 30 minutes of operating time | warning |
Occlusion | Delivery line or mechanism is blocked | Occlusion detected by sensor | warning |
Pendant Fault | Patient pendant connection or operation fault | Open circuit or short detected in pendant, or incompatible/stuck pendant | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.