Hospira LifeCare PCA3: troubleshooting and error codes

The ServiceForge knowledge base holds 41 documented symptoms for the Hospira LifeCare PCA3, with 70 ranked causes, 18 diagnostic checks with test points and expected values, and 102 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Self test or module test failure caution

Self test or service mode tests fail indicating component defect

Ranked causes

  1. 1. MCU PWA defective (common)
  2. 2. LCD assembly defective (occasional)
  3. 3. Numeric display PWA defective (occasional)
  4. 4. Icon display PWA defective (occasional)
  5. 5. Keypad defective (occasional)
  6. 6. Door lock assembly defective (occasional)
  7. 7. Mechanism assembly defective (occasional)
  8. 8. Patient pendant assembly defective (rare)
First check

Turn pump off and on, allow self test to complete

Expected
Self test passes without errors

12 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Empty vial sensor test failure caution

Empty vial detection sensor fails self test diagnostic

Ranked causes

  1. 1. Vial sensor switch misaligned (common)
  2. 2. Defective mechanism assembly (common)
  3. 3. Defective MCU PWA (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Icon display indicator test failure caution

Icon display PWA fails self test diagnostic

Ranked causes

  1. 1. Defective icon display PWA (common)
  2. 2. Icon display PWA connection fault (common)
  3. 3. Defective MCU PWA (occasional)
First check

Verify icon display connection

Test point
Icon display connector to MCU PWA
Expected
Connection fully seated and secure

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

LCD backlight test failure caution

LCD backlight circuit or lamp fails self test

Ranked causes

  1. 1. Defective LCD assembly backlight (common)
  2. 2. LCD assembly connection fault (common)
  3. 3. Defective MCU PWA (occasional)
First check

Verify LCD assembly connection

Test point
LCD connector to MCU PWA
Expected
Connection fully seated and secure

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

LCD display test failure caution

LCD assembly or display electronics fails self test

Ranked causes

  1. 1. Defective LCD assembly (common)
  2. 2. LCD assembly connection fault (common)
  3. 3. Defective MCU PWA (occasional)
First check

Verify LCD assembly connection

Test point
LCD connector to MCU PWA
Expected
Connection fully seated and secure

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

LED numeric display test failure caution

LED numeric display PWA fails self test

Ranked causes

  1. 1. Defective numeric display PWA (common)
  2. 2. Numeric display PWA connection fault (common)
  3. 3. Defective MCU PWA (occasional)
First check

Verify numeric display PWA connection

Test point
Display connector to MCU PWA
Expected
Connection fully seated and secure

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Occlusion switch error patient safety

Defective occlusion sensor fails to detect line blockage

Ranked causes

  1. 1. Occlusion alarm switch maladjustment (common)
  2. 2. Mechanism assembly failure (occasional)
  3. 3. MCU PWA malfunction (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Occlusion test failure patient safety

Occlusion detection system fails diagnostic test

Ranked causes

  1. 1. Infusion set not properly primed or installed (common)
  2. 2. Occlusion alarm switch misaligned (common)
  3. 3. Defective mechanism assembly (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Screw rotation error caution

Motor steps and screw rotations not in proportion, possible lubrication or mechanism issue

Ranked causes

  1. 1. Lead screw lubrication inadequate (common)
  2. 2. Mechanism assembly failure (occasional)
  3. 3. MCU PWA malfunction (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Syringe mechanism test failure caution

Syringe drive mechanism fails self test diagnostic

Ranked causes

  1. 1. Vial sensor switch misaligned (common)
  2. 2. Defective mechanism assembly (common)
  3. 3. Defective MCU PWA (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Delivery accuracy test failure patient safety

Pump delivery volume or rate does not meet accuracy specifications

Ranked causes

  1. 1. PCA set not properly primed (common)
  2. 2. Mechanism assembly defective (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Door lock test failure caution

Door lock assembly fails self test diagnostic

Ranked causes

  1. 1. Defective door lock assembly (common)
  2. 2. Defective MCU PWA (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keypad test failure caution

Pump keypad or keypad interface fails self test

Ranked causes

  1. 1. Defective keypad (common)
  2. 2. Defective MCU PWA (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Self test fails during verification patient safety

Device fails self-test verification when powered on, with alarm message displayed and system not proceeding to SELF TEST COMPLETE screen.

Ranked causes

  1. 1. Software or memory integrity failure (unknown)
  2. 2. Electronic function malfunction (unknown)
First check

Verify AC power connection

Test point
AC power indicator on front panel
Expected
AC power indicator illuminated

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm circuit test failure caution

Alarm speaker or alarm circuit fails self test

Ranked causes

  1. 1. Defective MCU PWA (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery fault caution

Battery malfunction: open circuit, over-discharge, or short circuit

Ranked causes

  1. 1. Battery defective (open, over-discharged, or shorted) (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Injector sensor switch caution

Injector sensor switch requires positioning and varnish sealing to ensure proper detection

Ranked causes

  1. 1. Injector sensor switch misalignment (common)
First check

Verify switch bracket position parallel to surface

Test point
Switch bracket
Expected
Parallel to surface, 1/32 inch from front enclosure

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Lead screw lubrication caution

Lead screw requires cleaning and lubrication with thin coating of grease

Ranked causes

  1. 1. Lead screw dry or contaminated with old grease and residue (common)
First check

Remove old grease and residue from lead screw

Test point
Lead screw threads
Expected
All old grease and residue completely removed
Tools
Isopropyl alcohol, cotton swabs or small brush

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Occlusion alarm switch and pressure patient safety

Occlusion alarm switch must trigger at 15±5 psi to ensure proper line blockage detection

Ranked causes

  1. 1. Occlusion alarm switch pressure out of specification (common)
First check

Attach DPM to patient line

Test point
End of patient line via three-way stopcock
Expected
DPM connected and zeroed

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Vial sensor switch caution

Vial sensor switch requires adjustment to detect vial presence correctly

Ranked causes

  1. 1. Vial sensor switch misalignment (common)
First check

Move switch up until it clicks off

Test point
Vial sensor switch
Expected
Audible click-off

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Charger fault caution

Battery charger is faulty and cannot charge battery

Ranked causes

  1. 1. Charger failure (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

CPU test failure caution

Central processing unit fails self test diagnostic

Ranked causes

  1. 1. Defective CPU (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Door improperly secured

Door secured on wrong screen or no operator input for two minutes

Ranked causes

  1. 1. Door secured on wrong screen or timeout (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Door left open

Door is left open or unlocked for more than two minutes

Ranked causes

  1. 1. Door not secured (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Electrical safety test failure high voltage

AC power cord or electrical safety components defective

Ranked causes

  1. 1. Defective AC power cord (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Infusion paused

Infusion stopped by operator pressing START/PAUSE button

Ranked causes

  1. 1. Operator pressed START/PAUSE button (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Injector not detected patient safety

Vial is detected but injector is not properly seated in cradle assembly

Ranked causes

  1. 1. Improper injector seating (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low battery patient safety

Operating on battery power with less than 30 minutes of battery life remaining

Ranked causes

  1. 1. Battery capacity depleted (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low battery power during operation patient safety

Battery indicator flashing on front panel while operating on battery power, indicating insufficient battery capacity for reliable device operation.

Ranked causes

  1. 1. Battery requires recharge (common)
First check

Verify AC power connection and indicator status

Test point
AC power indicator on front panel
Expected
AC power indicator illuminated indicating charging

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

No battery installed in device patient safety

Device prompts BATTERY NOT INSTALLED message during startup or service mode, indicating battery compartment is empty.

Ranked causes

  1. 1. Battery missing from compartment (common)
First check

Verify battery compartment access

Test point
Battery access cover on device
Expected
Battery access cover present and accessible

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Occlusion detected patient safety

Tubing or delivery line is occluded, blocking fluid flow

Ranked causes

  1. 1. Occlusion in delivery path (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Patient pendant assembly test failure patient safety

Patient pendant assembly fails self test diagnostic

Ranked causes

  1. 1. Defective patient pendant assembly (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

RAM memory test failure caution

Pump RAM memory fails self test diagnostic

Ranked causes

  1. 1. Defective MCU RAM (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Real-time clock test failure caution

Real-time clock RTC circuit fails self test diagnostic

Ranked causes

  1. 1. Defective RTC (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

ROM memory test failure caution

Pump ROM/EPROM memory fails self test diagnostic

Ranked causes

  1. 1. Defective EPROM (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Rotation mechanism test failure caution

Syringe rotation drive mechanism fails self test

Ranked causes

  1. 1. Defective mechanism assembly (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Self test failure

PVT self test does not complete successfully

Ranked causes

  1. 1. Infusion pump self test error (common)
First check

Power cycle the pump and run self test

Test point
Pump power switch
Expected
Self test completes without errors

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Service mode MCU test failure caution

Microcontroller unit PWA fails service mode testing

Ranked causes

  1. 1. Defective microcontroller PWA (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Syringe empty patient safety

Empty vial detected, no medication available for infusion

Ranked causes

  1. 1. Empty vial installed (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Syringe not properly loaded patient safety

Syringe (vial and injector) not properly seated in cradle assembly

Ranked causes

  1. 1. Improper syringe seating (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Vial not properly loaded patient safety

Injector detected but vial not properly seated in cradle assembly

Ranked causes

  1. 1. Improper vial seating (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
170Interrupt OverlapInterrupt failureerror
171Invalid InterruptInvalid or unused interrupterror
172Stack ErrorRAM errorcritical
173Task Dead ErrorPeriodic task or interrupt stopped runningcritical
174Kernel ErrorKernel failurecritical
175Watchdog ErrorMCU fails to strobe watchdog within timeouterror
176Cancel Timer ErrorTimer message cancelled unsuccessfullyerror
200Motor Overdelivery ConditionMotor running faster than expectedcritical
201Motor underdelivery conditionMotor running slower than expectedcritical
202Motor runawayMotor running without permissioncritical
203Excessive rateProgrammed infusion rate exceeds maximum limiterror
204Mode errorUnexpected operating mode detectedcritical
205Motor step time errorMotor step timing out of valid rangecritical
206Rate errorUnexpected infusion rate detectedcritical
207Delivery Volume ErrorVolume infused out of rangeerror
3105-Volt ErrorTwo voltage readings outside 4.7 V - 5.3 V rangecritical
311Lithium Battery ErrorLithium battery voltage below minimumerror
313Battery current errorBattery charge current exceeds safe limitcritical
480Configuration DataFailed CRC of pump configuration datacritical
481Infusion DataFailed CRC of infusion datacritical
482Program DataFailed CRC of program datacritical
483Flash Checksum ErrorFailed checksum at startup or during operationcritical
484Stored Protocol DataFailed CRC of stored protocolscritical
485Motor state errorInvalid rate, restart state, or bad events in motor softwarecritical
486History CorruptionHistory data corruptederror
487Motor Data ErrorFailed CRC of motor dataerror
488Drug Table Data ErrorFailed CRC of drug table dataerror
489User Interface Data ErrorFailed CRC of saved user interface dataerror
490Error Log Data ErrorFailed error log checksumerror
491Background Data ErrorInconsistent background dataerror
492Switch Data ErrorCorrupted switch dataerror
493Mirror data mismatchData in RAM does not match data in FLASHcritical
493Mirror data mismatched - Data in RAM does not match data in FLASHerror
494UI state errorUser interface state ID out of valid rangecritical
494UI state error - User interface state ID out of rangeerror
495Backlight data errorInvalid display data for backlightcritical
496Vial manager data checksum errorCRC failure of vial manager datacritical
497Power data error - Failed CRC of battery charging dataerror
497Power data checksum errorCRC failure of battery charging datacritical
498Display data errorInvalid LCD display data detectedcritical
499Alarm data errorInvalid alarm message data detectedcritical
550Power down errorImproper system shutdown detectedcritical
551Intertask message errorIllegal message between software taskscritical
552Invalid function parameterInvalid parameter data passed to functioncritical
553Invalid messageInvalid kernel message detectedcritical
554Invalid mode - Invalid user interface modeerror
554Invalid modeInvalid UI mode state detectedcritical
555Clock change errorInvalid date/time set detectedwarning
556Invalid UI state actionInvalid UI action code detectedcritical
557Invalid alarm - Illegal alarm message dataerror
557Invalid alarmIllegal alarm message datacritical
558Mode change errorIllegal mode change messagecritical
559Invalid UI stack - User interface stack is corrupted or fullerror
559Invalid UI stackUI stack corrupted or fullcritical
620RAM address error - Address lines stuck high, stuck low, or shorted togethercritical
620RAM address errorRAM address lines stuck or shortedcritical
621Chip ID errorRegister does not contain expected valuecritical
621Chip ID error - Register does not contain expected valuecritical
622Clock RAM error - Data or address communication errorserror
622Clock RAM errorData or address communication errors in clock RAMcritical
623External RAM errorFailed read/write operations on external RAMcritical
623External RAM error - Failed read/write operations on external RAMerror
624LCD backlight errorBacklight current out of rangecritical
625Battery icon LED errorBattery icon LED current out of rangecritical
626Audible alarm errorAlarm buzzer voltage out of valid rangecritical
627LCD readback errorDisplay data does not match RAM datacritical
627LCD readback error - Display data does not match RAM dataerror
628LED numeric display errorDigit segment current out of valid rangecritical
629Clock time errorRTC and MCU are out of synchronizationcritical
629Clock time error - RTC and MCU out of synchronizationerror
630Screw rotation error - Motor steps and screw rotations not in proportionerror
630Screw rotation errorMotor steps and lead screw rotations not proportionalcritical
631LCD timeout error - LCD does not acknowledge data within hardware specific-timeerror
631LCD timeout errorLCD does not acknowledge data within timeout periodcritical
633Print error - Failed UART loopback testerror
633Print errorFailed UART loopback test with printercritical
634ADC voltage reference errorAnalog-to-digital converter voltage out of rangecritical
635Stuck keyKeypad button held for more than 2 minuteswarning
635Stuck key error - Key stuck for more than two minuteswarning
636Motor winding excessive currentMotor winding current exceeds 200 mA limitcritical
637Motor winding open circuitMotor winding current below 75 mA minimumcritical
638Motor voltage errorMotor supply voltage out of acceptable rangecritical
639Contrast voltage errorLCD contrast voltage out of rangecritical
640Internal RAM error - Failed read/write operations on internal RAMerror
640Internal RAM errorFailed read/write operations on internal RAMcritical
641Occlusion switch errorOcclusion sensor malfunction detectedcritical
642Empty syringe switch errorEmpty syringe sensor malfunction detectedcritical
643Injector switch errorInjector sensor malfunction detectedcritical
644Watchdog timer errorWatchdog timer failurecritical
645Barcode reader errorThree consecutive barcode read attempts failedcritical
645Barcode reader error - Three consecutive attempts to read barcode failederror
Battery FaultBattery internal fault detected (open circuit, over-discharged, or shorted)Battery monitoring detects OB (open), OV (over-discharged), or SH (short)warning
Check InjectorInjector not detected when vial is presentVial detected but injector missingwarning
Check SettingsDoor secured on wrong screen or timeout after no key press for 2 minutesDoor locked on screen other than Setup/Press Start, or inactivity timeoutwarning
Check SyringeSyringe not properly positioned in cradleSyringe detector sees improper loadingwarning
Check VialVial not properly positioned or injector mismatchInjector detected but vial not properly loadedwarning
Door OpenDoor has been left open or unlocked for over 2 minutesDoor open/unlocked timeoutwarning
Empty SyringeSyringe medication reservoir is emptyEmpty syringe sensor activatedwarning
Infuser PausedInfusion delivery has been paused by operatorSTART/PAUSE button pressed during infusioninfo
Low BatteryBattery power remaining is critically lowBattery charge falls below 30 minutes of operating timewarning
OcclusionDelivery line or mechanism is blockedOcclusion detected by sensorwarning
Pendant FaultPatient pendant connection or operation faultOpen circuit or short detected in pendant, or incompatible/stuck pendantwarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.