Hospira LifeCare PCA: troubleshooting and error codes
The ServiceForge knowledge base holds 13 documented symptoms for the Hospira LifeCare PCA, with 13 ranked causes, 4 diagnostic checks with test points and expected values, and 90 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Battery fault caution
Battery malfunction: open circuit, over-discharge, or short circuit
Ranked causes
- 1. Battery defective (open, over-discharged, or shorted) (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Occlusion alarm switch test and adjustment patient safety
Test occlusion detection at specified pressure threshold; adjust pressure setpoint if needed
Ranked causes
- 1. Occlusion alarm switch not triggering at correct pressure (occasional)
Attach DPM to patient line via three-way stopcock
- Test point
- Stopcock valve
- Expected
- DPM connected and stopcock open to air
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Charger fault caution
Battery charger is faulty and cannot charge battery
Ranked causes
- 1. Charger failure (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Door improperly secured
Door secured on wrong screen or no operator input for two minutes
Ranked causes
- 1. Door secured on wrong screen or timeout (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Door left open
Door is left open or unlocked for more than two minutes
Ranked causes
- 1. Door not secured (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Infusion paused
Infusion stopped by operator pressing START/PAUSE button
Ranked causes
- 1. Operator pressed START/PAUSE button (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Injector not detected patient safety
Vial is detected but injector is not properly seated in cradle assembly
Ranked causes
- 1. Improper injector seating (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery patient safety
Operating on battery power with less than 30 minutes of battery life remaining
Ranked causes
- 1. Battery capacity depleted (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Occlusion detected patient safety
Tubing or delivery line is occluded, blocking fluid flow
Ranked causes
- 1. Occlusion in delivery path (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump stalls and cannot operate patient safety
Pump stalls, cannot be used, suspends communication with external devices, no alarm message displayed
Ranked causes
- 1. Null character sent by CE or MMU (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Syringe empty patient safety
Empty vial detected, no medication available for infusion
Ranked causes
- 1. Empty vial installed (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Syringe not properly loaded patient safety
Syringe (vial and injector) not properly seated in cradle assembly
Ranked causes
- 1. Improper syringe seating (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vial not properly loaded patient safety
Injector detected but vial not properly seated in cradle assembly
Ranked causes
- 1. Improper vial seating (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
170 | Interrupt Overlap | Interrupt failure | error |
171 | Invalid Interrupt | Invalid or unused interrupt | error |
172 | Stack Error | RAM error | critical |
173 | Task Dead Error | Periodic task or interrupt stopped running | critical |
174 | Kernel Error | Kernel failure | critical |
175 | Watchdog Error | MCU fails to strobe watchdog within timeout | error |
176 | Cancel Timer Error | Timer message cancelled unsuccessfully | error |
200 | Motor Overdelivery Condition | Motor running faster than expected | critical |
201 | Motor Underdelivery Condition | Motor running slower than expected | critical |
202 | Motor Runaway | Motor running without permission | critical |
203 | Excessive Rate | Rate/Mode Mismatches | error |
204 | Mode | Unexpected mode | error |
205 | Motor Step Time Error | Motor step time out of range | error |
206 | Rate | Unexpected rate | error |
207 | Delivery Volume Error | Volume infused out of range | error |
310 | 5-Volt Error | Two voltage readings outside 4.7 V - 5.3 V range | critical |
311 | Lithium Battery Error | Lithium battery voltage below minimum | error |
313 | Battery Current Error | Battery charge current exceeded limit | error |
480 | Configuration Data | Failed CRC of pump configuration data | critical |
481 | Infusion Data | Failed CRC of infusion data | critical |
482 | Program Data | Failed CRC of program data | critical |
483 | Flash Checksum Error | Failed checksum at startup or during operation | critical |
484 | Stored Protocol Data | Failed CRC of stored protocols | critical |
485 | Motor State | Invalid rate or restart or bad states/events in motor software | critical |
486 | History Corruption | History data corrupted | error |
487 | Motor Data Error | Failed CRC of motor data | error |
488 | Drug Table Data Error | Failed CRC of drug table data | error |
489 | User Interface Data Error | Failed CRC of saved user interface data | error |
490 | Error Log Data Error | Failed error log checksum | error |
491 | Background Data Error | Inconsistent background data | error |
492 | Switch Data Error | Corrupted switch data | error |
493 | Mirror Data Mismatched | Invalid date and time | error |
494 | UI State Error | User interface state ID out of range | error |
495 | Backlight Data Error | Invalid display data | error |
496 | Vial Manager Data Error | Failed CRC of vial manager data | error |
497 | Power Data Error | Failed CRC of battery charging data | error |
498 | Display Data Error | Invalid display data | error |
499 | Alarm Data Error | Invalid alarm data | error |
550 | Power Down Error | Improper shutdown | error |
551 | Intertask Message Error | Illegal intertask message | error |
552 | Invalid Function Parameter | Invalid parameter data | error |
553 | Invalid Message | Invalid kernel message | error |
554 | Invalid Mode | Invalid user interface mode | error |
555 | Clock Change Error | Invalid date and time during update | error |
556 | Invalid UI State Action | Invalid user interface action code | error |
557 | Invalid Alarm | Illegal alarm message data | error |
558 | Mode Change Error | Illegal mode change message | error |
559 | Invalid UI Stack | User interface stack corrupted or full | error |
560 | Drug Library Access Error | Invalid auto-programming order or invalid CCA index | error |
588 | Drug Table Data Error | Failed CRC of drug table data | error |
620 | RAM Address Error | Address lines stuck high, stuck low, or shorted | critical |
621 | Chip ID Error | Register does not contain expected value | critical |
622 | Clock RAM Error | Data or address communication errors | critical |
623 | External RAM Error | Failed read/write operations on external RAM | critical |
624 | LCD Backlight | Backlight current out of range | error |
625 | Battery Icon LED | Battery icon LED current out of range | error |
626 | Audible Alarm | On or off voltage out of range | error |
627 | LCD Readback Error | Display data does not match RAM data | error |
628 | LED Numeric Display Error | One or more digit segment current out of range | error |
629 | Clock Time Error | RTC and MCU out of synchronization | error |
630 | Screw Rotation Error | Motor steps and screw rotations not in proportion | critical |
631 | LCD Timeout Error | LCD does not acknowledge data within hardware specific time | error |
633 | Print Error | Failed UART loopback test | error |
634 | ADC Voltage Reference Error | Analog level out of range | error |
635 | Stuck Key | Key stuck for more than two minutes | error |
636 | Motor winding excessive current | Motor winding current above 200 mA | critical |
637 | Motor winding open circuit | Motor winding current below 75 mA | critical |
638 | Motor Voltage Error | Motor voltage out of range | error |
639 | Contrast Voltage Error | Contrast voltage out of range | error |
640 | Internal RAM read/write operation failure | Microcontroller RAM self-test | critical |
641 | Occlusion sensor malfunction or misalignment | Occlusion detection during priming or delivery | error |
642 | Empty syringe sensor malfunction | Syringe empty detection test | error |
643 | Injector sensor malfunction | Injector position detection | error |
644 | Watchdog timer circuit failure | Watchdog timer monitoring | critical |
645 | Barcode reader three consecutive read failures | Three failed barcode read attempts | error |
646 | Serial communication chip loopback test failure | UART loopback self-test | critical |
PVT-ALARM | Alarm circuit defective on MCU PWA | PVT alarm function test | error |
PVT-BACKLIGHT | LCD backlight circuit defective | PVT backlight test | warning |
PVT-DA | Delivery accuracy test failure | PVT Delivery Accuracy Test | error |
PVT-DOOR | Door lock assembly malfunction | PVT door lock test | warning |
PVT-EMPTY VIAL | Empty syringe detection mechanism failure | PVT empty syringe detection test | error |
PVT-INDICATOR | Icon display LEDs defective | PVT icon display test | warning |
PVT-KEYPAD | Keypad membrane or matrix defective | PVT keypad function test | error |
PVT-LCD | LCD assembly defective or disconnected | PVT LCD test | error |
PVT-LED | Numeric display LED array defective | PVT numeric display test | error |
PVT-OCCLUSION | Occlusion detection test failure | PVT Occlusion Test | error |
PVT-PATIENT PENDANT | Patient pendant assembly not responding | PVT patient pendant communication test | error |
PVT-ROTATION | Syringe rotation motor mechanism defective | PVT syringe rotation test | error |
PVT-SAFETY | Electrical safety test failure | PVT Electrical Safety Test | critical |
PVT-SYRINGE | Mechanism assembly position sensing failure | PVT syringe position test | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.