Hospira LifeCare PCA: troubleshooting and error codes

The ServiceForge knowledge base holds 13 documented symptoms for the Hospira LifeCare PCA, with 13 ranked causes, 4 diagnostic checks with test points and expected values, and 90 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Battery fault caution

Battery malfunction: open circuit, over-discharge, or short circuit

Ranked causes

  1. 1. Battery defective (open, over-discharged, or shorted) (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Occlusion alarm switch test and adjustment patient safety

Test occlusion detection at specified pressure threshold; adjust pressure setpoint if needed

Ranked causes

  1. 1. Occlusion alarm switch not triggering at correct pressure (occasional)
First check

Attach DPM to patient line via three-way stopcock

Test point
Stopcock valve
Expected
DPM connected and stopcock open to air

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Charger fault caution

Battery charger is faulty and cannot charge battery

Ranked causes

  1. 1. Charger failure (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Door improperly secured

Door secured on wrong screen or no operator input for two minutes

Ranked causes

  1. 1. Door secured on wrong screen or timeout (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Door left open

Door is left open or unlocked for more than two minutes

Ranked causes

  1. 1. Door not secured (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Infusion paused

Infusion stopped by operator pressing START/PAUSE button

Ranked causes

  1. 1. Operator pressed START/PAUSE button (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Injector not detected patient safety

Vial is detected but injector is not properly seated in cradle assembly

Ranked causes

  1. 1. Improper injector seating (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low battery patient safety

Operating on battery power with less than 30 minutes of battery life remaining

Ranked causes

  1. 1. Battery capacity depleted (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Occlusion detected patient safety

Tubing or delivery line is occluded, blocking fluid flow

Ranked causes

  1. 1. Occlusion in delivery path (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pump stalls and cannot operate patient safety

Pump stalls, cannot be used, suspends communication with external devices, no alarm message displayed

Ranked causes

  1. 1. Null character sent by CE or MMU (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Syringe empty patient safety

Empty vial detected, no medication available for infusion

Ranked causes

  1. 1. Empty vial installed (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Syringe not properly loaded patient safety

Syringe (vial and injector) not properly seated in cradle assembly

Ranked causes

  1. 1. Improper syringe seating (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Vial not properly loaded patient safety

Injector detected but vial not properly seated in cradle assembly

Ranked causes

  1. 1. Improper vial seating (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
170Interrupt OverlapInterrupt failureerror
171Invalid InterruptInvalid or unused interrupterror
172Stack ErrorRAM errorcritical
173Task Dead ErrorPeriodic task or interrupt stopped runningcritical
174Kernel ErrorKernel failurecritical
175Watchdog ErrorMCU fails to strobe watchdog within timeouterror
176Cancel Timer ErrorTimer message cancelled unsuccessfullyerror
200Motor Overdelivery ConditionMotor running faster than expectedcritical
201Motor Underdelivery ConditionMotor running slower than expectedcritical
202Motor RunawayMotor running without permissioncritical
203Excessive RateRate/Mode Mismatcheserror
204ModeUnexpected modeerror
205Motor Step Time ErrorMotor step time out of rangeerror
206RateUnexpected rateerror
207Delivery Volume ErrorVolume infused out of rangeerror
3105-Volt ErrorTwo voltage readings outside 4.7 V - 5.3 V rangecritical
311Lithium Battery ErrorLithium battery voltage below minimumerror
313Battery Current ErrorBattery charge current exceeded limiterror
480Configuration DataFailed CRC of pump configuration datacritical
481Infusion DataFailed CRC of infusion datacritical
482Program DataFailed CRC of program datacritical
483Flash Checksum ErrorFailed checksum at startup or during operationcritical
484Stored Protocol DataFailed CRC of stored protocolscritical
485Motor StateInvalid rate or restart or bad states/events in motor softwarecritical
486History CorruptionHistory data corruptederror
487Motor Data ErrorFailed CRC of motor dataerror
488Drug Table Data ErrorFailed CRC of drug table dataerror
489User Interface Data ErrorFailed CRC of saved user interface dataerror
490Error Log Data ErrorFailed error log checksumerror
491Background Data ErrorInconsistent background dataerror
492Switch Data ErrorCorrupted switch dataerror
493Mirror Data MismatchedInvalid date and timeerror
494UI State ErrorUser interface state ID out of rangeerror
495Backlight Data ErrorInvalid display dataerror
496Vial Manager Data ErrorFailed CRC of vial manager dataerror
497Power Data ErrorFailed CRC of battery charging dataerror
498Display Data ErrorInvalid display dataerror
499Alarm Data ErrorInvalid alarm dataerror
550Power Down ErrorImproper shutdownerror
551Intertask Message ErrorIllegal intertask messageerror
552Invalid Function ParameterInvalid parameter dataerror
553Invalid MessageInvalid kernel messageerror
554Invalid ModeInvalid user interface modeerror
555Clock Change ErrorInvalid date and time during updateerror
556Invalid UI State ActionInvalid user interface action codeerror
557Invalid AlarmIllegal alarm message dataerror
558Mode Change ErrorIllegal mode change messageerror
559Invalid UI StackUser interface stack corrupted or fullerror
560Drug Library Access ErrorInvalid auto-programming order or invalid CCA indexerror
588Drug Table Data ErrorFailed CRC of drug table dataerror
620RAM Address ErrorAddress lines stuck high, stuck low, or shortedcritical
621Chip ID ErrorRegister does not contain expected valuecritical
622Clock RAM ErrorData or address communication errorscritical
623External RAM ErrorFailed read/write operations on external RAMcritical
624LCD BacklightBacklight current out of rangeerror
625Battery Icon LEDBattery icon LED current out of rangeerror
626Audible AlarmOn or off voltage out of rangeerror
627LCD Readback ErrorDisplay data does not match RAM dataerror
628LED Numeric Display ErrorOne or more digit segment current out of rangeerror
629Clock Time ErrorRTC and MCU out of synchronizationerror
630Screw Rotation ErrorMotor steps and screw rotations not in proportioncritical
631LCD Timeout ErrorLCD does not acknowledge data within hardware specific timeerror
633Print ErrorFailed UART loopback testerror
634ADC Voltage Reference ErrorAnalog level out of rangeerror
635Stuck KeyKey stuck for more than two minuteserror
636Motor winding excessive currentMotor winding current above 200 mAcritical
637Motor winding open circuitMotor winding current below 75 mAcritical
638Motor Voltage ErrorMotor voltage out of rangeerror
639Contrast Voltage ErrorContrast voltage out of rangeerror
640Internal RAM read/write operation failureMicrocontroller RAM self-testcritical
641Occlusion sensor malfunction or misalignmentOcclusion detection during priming or deliveryerror
642Empty syringe sensor malfunctionSyringe empty detection testerror
643Injector sensor malfunctionInjector position detectionerror
644Watchdog timer circuit failureWatchdog timer monitoringcritical
645Barcode reader three consecutive read failuresThree failed barcode read attemptserror
646Serial communication chip loopback test failureUART loopback self-testcritical
PVT-ALARMAlarm circuit defective on MCU PWAPVT alarm function testerror
PVT-BACKLIGHTLCD backlight circuit defectivePVT backlight testwarning
PVT-DADelivery accuracy test failurePVT Delivery Accuracy Testerror
PVT-DOORDoor lock assembly malfunctionPVT door lock testwarning
PVT-EMPTY VIALEmpty syringe detection mechanism failurePVT empty syringe detection testerror
PVT-INDICATORIcon display LEDs defectivePVT icon display testwarning
PVT-KEYPADKeypad membrane or matrix defectivePVT keypad function testerror
PVT-LCDLCD assembly defective or disconnectedPVT LCD testerror
PVT-LEDNumeric display LED array defectivePVT numeric display testerror
PVT-OCCLUSIONOcclusion detection test failurePVT Occlusion Testerror
PVT-PATIENT PENDANTPatient pendant assembly not respondingPVT patient pendant communication testerror
PVT-ROTATIONSyringe rotation motor mechanism defectivePVT syringe rotation testerror
PVT-SAFETYElectrical safety test failurePVT Electrical Safety Testcritical
PVT-SYRINGEMechanism assembly position sensing failurePVT syringe position testerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI