Hospira LifeCare 4100 PCA PLUS: troubleshooting and error codes
The ServiceForge knowledge base holds 56 documented symptoms for the Hospira LifeCare 4100 PCA PLUS, with 106 ranked causes, 41 diagnostic checks with test points and expected values, and 24 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Alarm test failure
Alarm does not sound or indicates test failure during PVT
Ranked causes
- 1. Loose/defective ribbon cable at power supply (common)
- 2. Defective alarm volume switch (occasional)
- 3. Defective piezoelectric alarm (occasional)
- 4. Defective power supply PWA (occasional)
- 5. Defective CPU/display PWA (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keypad test failure
Touchswitch or keypad does not respond during PVT test
Ranked causes
- 1. Loose/damaged ribbon cable at touchswitch (common)
- 2. Loose/damaged ribbon cable at power supply (common)
- 3. Defective touchswitch (occasional)
- 4. Defective CPU/display PWA (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low/empty syringe test failure
Syringe level sensors fail or do not detect during PVT test
Ranked causes
- 1. Loose connection (common)
- 2. Syringe low sensor switch out of adjustment or defective (common)
- 3. Syringe empty sensor switch out of adjustment or defective (common)
- 4. Defective power supply PWA (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Motor rotation test failure
Motor does not rotate or stalls during PVT test
Ranked causes
- 1. Lead screw dirty or needs lubrication (common)
- 2. Loose connection (common)
- 3. Defective power supply PWA (occasional)
- 4. Defective motor assembly (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Occlusion test failure
Occlusion detection system fails during PVT test
Ranked causes
- 1. Syringe not modified properly (common)
- 2. Set not installed or primed properly (common)
- 3. Occlusion pressure switch out of adjustment or defective (occasional)
- 4. Defective power supply PWA (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Syringe test failure
Syringe detection or positioning fails during PVT test
Ranked causes
- 1. Loose connection (common)
- 2. Injector sensor switch defective (common)
- 3. Vial sensor switch out of adjustment or defective (occasional)
- 4. Defective power supply PWA (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery charger test failure
Battery charging system fails during PVT test
Ranked causes
- 1. Loose connection (common)
- 2. Defective boost charger PWA (occasional)
- 3. Defective power supply PWA (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Delivery accuracy test failure
PVT delivery accuracy test fails
Ranked causes
- 1. PCA set not primed properly (common)
- 2. Defective power supply PWA (occasional)
- 3. Defective motor assembly (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
DOOR OPEN alarm patient safety
Security door left open for more than two minutes without button press
Ranked causes
- 1. Security door left open during operation (common)
- 2. Door closure not sensed by infuser (occasional)
- 3. Defective door lock switch (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrical safety test failure
Electrical safety test fails during PVT
Ranked causes
- 1. Insufficient earth connection (common)
- 2. Defective AC power cord plug (common)
- 3. Defective AC power cord (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Indicator test failure
Status indicator does not light or operate during PVT test
Ranked causes
- 1. Loose/damaged ribbon cable (common)
- 2. Defective CPU/display PWA (occasional)
- 3. Defective power supply PWA (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
LOW BATTERY alarm (flashing battery symbol) caution
Battery power level approaching critical depletion
Ranked causes
- 1. Battery power nearly depleted (common)
- 2. No battery operation or short operating time (occasional)
- 3. Battery pack is defective or exceeded useful life (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
OCCLUSION alarm patient safety
Back pressure detected in delivery line or syringe assembly
Ranked causes
- 1. Kinked, occluded tubing, closed slide clamp, or blocked venipuncture device (common)
- 2. Occlusion pressure setting out of calibration (occasional)
- 3. Defective occlusion pressure switch (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient control jack test failure
Patient control pendant jack does not operate during PVT test
Ranked causes
- 1. Loose connection (common)
- 2. Defective interconnect PWA cable (common)
- 3. Defective patient pendant jack (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Security door test failure
Door lock or security door does not operate during PVT test
Ranked causes
- 1. Loose connection (common)
- 2. Door lock switch out of adjustment or defective (common)
- 3. Defective power supply PWA (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Service test mode test failure
PVT test fails during service mode initialization
Ranked causes
- 1. Infuser not in cold start (common)
- 2. Defective touchswitch (occasional)
- 3. Defective CPU/display PWA (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CHECK INJECTOR message displayed patient safety
Injector improperly installed, missing, or sensor not detecting injector
Ranked causes
- 1. Injector improperly installed or missing (common)
- 2. Injector sensor misaligned or out of adjustment (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CHECK SYRINGE message displayed patient safety
Syringe assembly (vial and injector) improperly installed or missing
Ranked causes
- 1. Syringe assembly improperly installed or missing (common)
- 2. Defective or misadjusted injector sensor switch (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CHECK VIAL message displayed patient safety
Vial improperly installed, missing, or sensor not detecting vial
Ranked causes
- 1. Vial improperly installed or missing (common)
- 2. Vial sensor misaligned or out of adjustment (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
EMPTY SYRINGE alarm patient safety
Syringe empty or nearly empty (approximately 1 mL may remain)
Ranked causes
- 1. Syringe depleted (common)
- 2. Empty syringe switch out of calibration (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INFUSER IN RESET message displayed
Infuser in reset mode after alarm or user reset action
Ranked causes
- 1. Normal reset state after alarm cleared (common)
- 2. RESET/START pressed while security door locked (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
LOW SYRINGE alarm (PCA Plus only) patient safety
Approximately 5 mL of solution remaining in syringe
Ranked causes
- 1. Low syringe level reached during infusion (common)
- 2. Low syringe switch out of calibration (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No AC or battery operation caution
Infuser cannot operate on AC line power or battery
Ranked causes
- 1. Fuse blown (common)
- 2. Defective power supply PWA (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No audible alarm or audible alarm level control not working patient safety
Audio alarm does not sound or volume control does not function
Ranked causes
- 1. Defective audible level switch (common)
- 2. Defective power supply PWA (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Occlusion alarm does not trigger at expected pressure patient safety
Infuser fails to sound alarm and display OCCLUSION message when system pressure reaches threshold.
Ranked causes
- 1. Pressure sensor failure or miscalibration (common)
- 2. Alarm circuitry malfunction (occasional)
Monitor pressure gauge when stopcock is closed after fluid discharge
- Test point
- Digital pressure meter (DPM) reading
- Expected
- -008 and lower: 14.0 to 19.0 psig; -010 and higher: 12.5 to 17.5 psig when alarm triggers, -008: 14.0-19.0 psig; -010+: 12.5-17.5 psig
- Tools
- Digital pressure meter (0 to 50 psig), Bio-Tek DPM II
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PCA plus continuous test failure
PCA plus continuous infusion mode test fails during PVT
Ranked causes
- 1. Set not installed or primed properly (common)
- 2. Infuser not set up properly (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not deliver correct volume patient safety
Infuser fails to deliver expected fluid volume during loading dose administration.
Ranked causes
- 1. Pump mechanism wear or calibration drift (common)
- 2. Air in administration set or vial (occasional)
Measure delivered volume during two 10 mL loading doses
- Test point
- Graduate or marked test tube
- Expected
- 20±1 mL total volume, ±1 mL
- Tools
- Graduate or marked test tube readable to 0.2 mL increments or smaller
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
12/24 hour clock selection test failure
PCA Plus fails to verify clock format selection during PVT
Ranked causes
- 1. Defective CPU/display PWA (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Door lock switch caution
Adjustment of door lock switch to ensure switch actuation is heard when door is locked and unlocked. Recommended tool: needle nose pliers.
Ranked causes
- 1. Door lock switch not actuating properly when door is locked or unlocked (common)
Test switch actuation with door lock actuator arm pressed
- Test point
- Door lock actuator arm and switch
- Expected
- Switch actuation audible (click or signal)
- Tools
- Needle nose pliers
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Injector sensor switch adjustment caution
Adjust injector sensor switch position and gap for proper operation on PCA Plus Series Infusers with final assemblies 850-04250-002 through 850-04250-008.
Ranked causes
- 1. Injector sensor switch requires repositioning for proper detection (common)
Verify switch bracket position and gap from case
- Test point
- Injector sensor switch bracket
- Expected
- Parallel to surface and 1/32 inch from infuser case, 1/32 inch
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: LCD intensity caution
Adjustment of LCD display brightness using resistor R19 trim pot so screen is readable from approximately three feet distance.
Ranked causes
- 1. LCD brightness out of specification (common)
Verify LCD readability during adjustment
- Test point
- LCD display screen
- Expected
- Display readable from distance of approximately three feet
- Tools
- Trim pot tool, AC power
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Occlusion alarm switch and pressure high voltage
Occlusion alarm switch test and adjustment using modified syringe with water and pressure meter to verify alarm triggers within specified pressure range
Ranked causes
- 1. Occlusion alarm switch triggering outside specified pressure range (unknown)
Modify syringe plunger for test
- Test point
- Syringe plunger
- Expected
- Top two ribs removed cleanly, no rough material; light silicone oil coating applied to remaining rib
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Occlusion alarm switch and pressure (-008 and lower) caution
Adjustment of occlusion alarm switch contact and pressure threshold to trigger between 14–19 psig. Switch button must contact actuator without opening switch prematurely.
Ranked causes
- 1. Occlusion alarm switch misaligned or pressure threshold out of specification (common)
Examine switch button contact with switch actuator
- Test point
- Occlusion alarm switch
- Expected
- Switch button in contact with actuator, not opened
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Occlusion alarm switch and pressure (-010 and higher) caution
Adjustment of occlusion alarm switch contact and pressure threshold to trigger between 12.5–17.5 psig. Switch actuator must contact button without prematurely opening switch.
Ranked causes
- 1. Occlusion alarm switch misaligned or pressure threshold out of specification (common)
Examine switch actuator contact with switch button
- Test point
- Occlusion alarm switch
- Expected
- Switch actuator in contact with button, not tripping switch
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Real-time clock (PCA Plus II)
Setting real-time clock time and date via LCD menu after initial installation, time change, or complete battery discharge. No tools required.
Ranked causes
- 1. Real-time clock not set or reset required (common)
Navigate to CHANGE screen for time/date entry
- Test point
- LCD touchswitches and display menu
- Expected
- LCD displays CHANGE? screen with flashing time/date fields
- Tools
- Touchswitches [NO], [YES/ENTER], [REVIEW CHANGE]
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Real-time clock verification (PCA Plus II)
Optional verification via PVT if infuser may have been damaged during clock setting procedure.
Ranked causes
- 1. Infuser may have been damaged during real-time clock setting (rare)
Execute full PVT self-test
- Test point
- Full infuser system
- Expected
- PVT passes all checks per Section 5.2
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Syringe empty switch caution
Syringe empty switch adjustment using syringe limit gauge to trigger alarm when syringe reaches empty state
Ranked causes
- 1. Syringe empty switch not triggering at correct empty level (unknown)
Install gauge and move cradle to 0.3 inch position
- Test point
- Syringe limit gauge dial
- Expected
- Dial reads approximately 0.3 inch; LCD displays PURGE THE SYSTEM NOW?
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Syringe low switch caution
Syringe low switch adjustment using syringe limit gauge to trigger alarm at correct fill level
Ranked causes
- 1. Syringe low switch not triggering at correct fill level (unknown)
Install syringe limit gauge in cradle
- Test point
- Cradle with gauge installed
- Expected
- Dial reads approximately 0.05 inch
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Syringe plunger modification
Modify syringe plunger by removing top two ribs and applying silicone oil to prevent friction during occlusion pressure testing
Ranked causes
- 1. Unmodified syringe plunger causes friction and false pressure readings (unknown)
Remove syringe plunger
- Test point
- Syringe plunger
- Expected
- Plunger cleanly separated from syringe barrel
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Vial sensor switch caution
Vial sensor switch adjustment to detect vial presence with proper bracket positioning
Ranked causes
- 1. Vial sensor switch misaligned or not detecting vial (unknown)
Verify vial sensor switch detects vial
- Test point
- Vial sensor switch actuator arm
- Expected
- Switch clicks when rod inserted approximately 1/16 inch
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Vial sensor switch adjustment caution
Adjust vial sensor switch position for proper detection on PCA Plus Series Infusers with final assemblies 850-04250-002 through 850-04250-008.
Ranked causes
- 1. Vial sensor switch requires repositioning for proper detection (common)
Verify switch click off position
- Test point
- Vial sensor switch
- Expected
- Switch clicks off when moved up slowly
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CHECK 4-HR LIMIT message displayed patient safety
Four-hour dose limit parameter error detected
Ranked causes
- 1. Four-hour dose limit less than PCA dose (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CHECK SETTINGS message displayed patient safety
Security door closed and locked before setup was complete
Ranked causes
- 1. Security door closed prematurely during setup (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Injector sensor switch malfunction patient safety
Injector bolus function does not work or sensor does not respond
Ranked causes
- 1. Injector sensor switch failure or improper soldering (common)
Verify wire soldering to replacement switch terminals
- Test point
- Injector sensor switch terminals
- Expected
- Two black wires soldered to center terminal, brown wire to outside terminal
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
LCD test failure
LCD display does not respond during PVT test
Ranked causes
- 1. Defective CPU/display PWA (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
LED test failure
Indicator LED does not light during PVT test
Ranked causes
- 1. Defective CPU/display PWA (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
MALFUNCTION XX alarm (internal failure detected) caution
Infuser self-test detected internal system failure with error code
Ranked causes
- 1. Internal system failure detected by self-test (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Memory test failure
Internal RAM, External RAM, or ROM test fails
Ranked causes
- 1. Defective CPU/display PWA (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Motor assembly failure or improper coupling patient safety
Syringe does not advance, infusion stalled or motor does not run
Ranked causes
- 1. Motor assembly worn, damaged, or improperly coupled to lead screw (common)
Verify coupling alignment before motor installation
- Test point
- Lead screw and motor gear case shaft flats
- Expected
- Flats aligned; coupling pin engages both flats
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Occlusion pressure switch malfunction patient safety
Occlusion alarm does not trigger or triggers incorrectly during infusion
Ranked causes
- 1. Occlusion pressure switch failure or incorrect wiring (common)
Verify correct wire soldering to switch terminals
- Test point
- Occlusion pressure switch terminals
- Expected
- Wires soldered per model version: -008/Lower (orange front, black center, blue front) or -010/Higher (orange front, blue center, black front)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient pendant assembly test failure
Patient pendant assembly does not function during PVT test
Ranked causes
- 1. Defective patient pendant assembly (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient pendant tone selection test failure
PVT fails to verify patient pendant tone selection configuration
Ranked causes
- 1. Defective CPU/display PWA (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Real time clock test failure
PVT test fails on real time clock or elapsed days check
Ranked causes
- 1. Defective CPU/display PWA (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Select RX concentration test failure
Software version 3.1 or higher required for RX concentration selection test
Ranked causes
- 1. Incorrect software revision (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Slide assembly wear or mechanical failure patient safety
Syringe does not move smoothly, grinding noise, or motor stalls during infusion
Ranked causes
- 1. Lead screw worn, binding, or insufficient lubrication (common)
Verify lead screw lubrication
- Test point
- Lead screw surface
- Expected
- Even grease coating along entire length
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software version test failure
PVT test fails on software version check
Ranked causes
- 1. Defective CPU/display PWA (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
1A | MALF line went low could not return to normal | Line state failure during operation | critical |
1B | MALF line went low could not return to normal | Line state failure during operation | critical |
1C | MALF line did not go low soon enough after power-up | Power-up timing failure | critical |
1D | Malfunction line went high too soon | Line timing error | critical |
1E | Watchdog circuit failure during real-time operation | Watchdog timer circuit failure | critical |
2A | Power-up diagnostic failed | Startup self-test failure | critical |
2B | Nonvolatile RAM checksum test failed | RAM checksum error | critical |
3A | Motor failed to rotate | Motor does not turn when commanded | error |
3B | MOTOR BAD signal from motor driver circuitry | Motor driver circuit failure | error |
4A | LCD timeout error | LCD communication timeout | error |
4B | LCD data readback error | LCD data does not match readback | error |
4C | Nonvolatile RAM test failed | RAM self-test failure | error |
4D | EPROM checksum test failed | EPROM checksum error | error |
4F | LED segment defective | LED segment self-test failure | warning |
4G | Real-time clock stopped | RTC clock not running | error |
4H | Touchswitch stuck on front panel | Touchswitch sensed as continuously pressed | error |
4K | LED error | LED self-test failure | warning |
4L | Voltage error | error | |
4N | Error in timer T1 | error | |
4O | Error in timer T2 | error | |
4P | Program error | error | |
4Q | CPU stack error | error | |
4R | Analog-to-digital converter error | error | |
5A | Undetermined error | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.