Hologic LORAD Premier Encore: troubleshooting and error codes
The ServiceForge knowledge base holds 19 documented symptoms for the Hologic LORAD Premier Encore, with 28 ranked causes, 82 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Totally blank video monitors
Both monitors display no image, text, or brightness after power on and initialization
Ranked causes
- 1. Power supply failure or blown fuses (common)
- 2. Loose integrated circuit in socket (occasional)
- 3. Video PCB failure (occasional)
- 4. Monitor brightness/contrast set too low (common)
Observe if system beeps when powered ON
- Test point
- System audio output
- Expected
- System beeps upon power ON
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No video signal from camera to console
Monitors receive no video input from camera; camera not outputting video or no power to camera
Ranked causes
- 1. Video cable between camera and J14 damaged or disconnected (common)
- 2. Camera lacks 12V dc power (common)
- 3. Image Intensifier failure (occasional)
Measure resistance across BNC connector J14 to camera with power off
- Test point
- BNC connector at J14 on Video PCB to camera return
- Expected
- About 6K ohms, approximately
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Positional laser crosshair not displayed radiation
One or both laser crosshairs missing when laser switch pressed on head assembly
Ranked causes
- 1. Faulty positional laser (side or rear) (occasional)
- 2. Laser Power Board defective (occasional)
- 3. Laser Switch defective (rare)
Measure voltage across positional laser connector
- Test point
- J3 or J4 pins on Laser Power Board
- Expected
- greater than +4 V, >
- Tools
- Digital voltmeter
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
DC power supply 5V output out of specification
5V supply voltage measured at power supply output or TP1 outside acceptable range
Ranked causes
- 1. 5V fuse blown or open (occasional)
- 2. Switching power supply regulator failure (occasional)
Measure +5V output voltage
- Test point
- TP1 or power supply output terminals
- Expected
- 5.15 V +0.00, -0.05 V, +0.00, -0.05
- Tools
- Digital voltmeter
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No beam on indicator or timer when footswitch depressed
Footswitch pressed but timer does not count and Beam On indicator does not light
Ranked causes
- 1. Footswitch or footswitch cabling failure (common)
- 2. Footswitch or cabling defect confirmed (common)
Press and hold footswitch and observe timer and Beam On indicator
- Test point
- Timer display, Beam On indicator light
- Expected
- Timer counts up, Beam On indicator lights
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Automatic Dose Settings radiation
Configuration of automatic dose rate (ADR) parameters for imaging optimization
Ranked causes
- 1. Incorrect ADR parameter values requiring adjustment for proper dose delivery (occasional)
Verify Top White setting
- Test point
- ADR settings menu
- Expected
- 105
6 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Beam Collimation radiation
Adjustment of collimator assembly to ensure round, centered X-ray beam at 6" and 4" aperture settings after HVPS or X-Ray Tube Assembly replacement
Ranked causes
- 1. Collimator misalignment after component replacement (common)
Verify system powers on and displays default 40 kV, 0.020 mA
- Test point
- Left monitor
- Expected
- 40 kV and 0.020 mA displayed
6 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: C-Arm Calibration radiation
Calibration of X-ray kilovolt and milliampere output, and feedback voltage measurement to establish proper dose parameters for imaging
Ranked causes
- 1. Miscalibrated kV and mA output or feedback voltage after system setup or component replacement (common)
Verify system startup and default power display
- Test point
- Left and right monitor screens
- Expected
- Left monitor displays 40 kV and 0.020 mA; right monitor prompts for password
4 more checks with pass and fail actions, 15 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Camera focus radiation
Image focus requires adjustment to achieve optimal resolution on monitor
Ranked causes
- 1. Camera lens out of focus (common)
Position C-Arm with Image Intensifier on top
- Test point
- C-Arm
- Expected
- Image Intensifier on top, locked into position
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Camera focus adjustment radiation
Fine adjustment of camera lens focus for optimal image resolution
Ranked causes
- 1. Focus ring needs fine adjustment (common)
Adjust focus ring while observing image on monitor
- Test point
- Focus ring (silver anodized) on camera lens assembly
- Expected
- Image sharpness improves with ring rotation lp/mm, 6.5-7.5 lp/mm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Camera Gain Adjustment radiation
Calibration of image intensifier iris and camera gain parameters for optimized image quality at 4-inch and 6-inch aperture settings
Ranked causes
- 1. Incorrect camera iris position or gain values requiring adjustment after collimation or C-arm calibration (common)
Verify 6-inch camera gain setting
- Test point
- Calibration menu
- Expected
- 165
4 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: kV and mA output verification radiation
Verify and calibrate kV and mA output settings to ensure accurate beam control in manual and auto modes.
Ranked causes
- 1. Manual verification required for preventative maintenance (common)
Manual mode: foot pedal at 0.050 mA
- Test point
- TP12 (Gnd) to TP2 (kV Out)
- Expected
- 2.38 ±5% VDC, ±5%
- Tools
- Digital voltmeter
12 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Positional laser alignment radiation
Side and rear laser crosshairs require alignment to target lines on Image Intensifier cover
Ranked causes
- 1. Side and rear positional lasers misaligned (common)
Copy Laser Target Template from Appendix B
- Test point
- Documentation
- Expected
- Template copied and ready
19 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
High voltage arcing during X-ray operation high voltage
Sharp muffled snap sound heard during X-ray operation; may occur once or repeat; indicates possible electrical arc
Ranked causes
- 1. Tube housing coolant leak allowing air into tube housing (common)
Listen for sharp muffled snap sound during X-ray operation
- Test point
- System audio during footswitch depression
- Expected
- No arcing sounds
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Kilovoltage control calibration radiation
Verify kV output control voltage accuracy at various kilovoltage settings
Ranked causes
- 1. Control circuitry miscalibration or drift (common)
Measure kV Out control voltage at 40 kV setting
- Test point
- TP1 pin 2 (kV Out) to ground (pin 10, 11 or 12)
- Expected
- 1.39 ±5% VDC, ±5%
- Tools
- Digital voltmeter
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low kV reading during X-ray operation high voltage
kV reading is lower than expected during mid-range power settings with footswitch depressed
Ranked causes
- 1. High voltage power supply failure (common)
Set system to 55 kV and 0.057 mA via F10 key; observe kV reading with footswitch depressed
- Test point
- Display kV readout
- Expected
- kV reading approximately 55
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low mA reading during X-ray operation radiation
mA reading is lower than expected during mid-range power settings with footswitch depressed
Ranked causes
- 1. X-ray tube failure or malfunction (common)
Set system to 55 kV and 0.057 mA via F10 key; observe mA reading with footswitch depressed
- Test point
- Display mA readout
- Expected
- mA reading approximately 0.057
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Milliamperage control calibration radiation
Verify mA output control voltage accuracy at various milliamperage settings
Ranked causes
- 1. Control circuitry miscalibration or drift (common)
Measure µA Out control voltage at 0.020 mA setting
- Test point
- TP1 pin 4 (µA Out) to ground (pin 12)
- Expected
- 0.95 ±5% VDC, ±5%
- Tools
- Digital voltmeter
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System does not initialize completely
System startup stalls with 'System Startup...Please Wait' message on left screen, communication difficulty between PC Controller and X-ray or Video PCB
Ranked causes
- 1. PC Controller wiring connections not properly seated (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.