FR FieldReady Academy
C-arms · InSight family

Hologic InSight: troubleshooting and error codes

The ServiceForge knowledge base holds 26 documented symptoms for the Hologic InSight, with 28 ranked causes, 72 diagnostic checks with test points and expected values, and 401 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Unit does not power on high voltage

System cabinet shows no signs of power; check fuses in Power Input Module and DIN Rail

Ranked causes

  1. 1. Power Input Module fuse open (100-120V input) (common)
  2. 2. Power Input Module fuse open (220-240V input) (common)
  3. 3. DIN Rail fuse open (occasional)
First check

Inspect Power Input Module fuses for open condition

Test point
Power Input Module fuses (100-120V or 220-240V)
Expected
Intact filament; continuity across fuse
Tools
Visual inspection or multimeter

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Beam quality verification radiation

X-ray beam quality measurement using dosimeter to establish Half-Value Layer (HVL) compliance

Ranked causes

  1. 1. HVL below specification (unknown)
First check

Set system parameters for beam quality measurement

Test point
System Configuration - Physician Preferences tab
Expected
Simultaneous kV/mA setting Disabled

5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

C-arm brake not functioning properly caution

C-arm brake fails to stop or hold C-arm in position, or brake operation is sluggish

Ranked causes

  1. 1. C-arm brake wear or damage (occasional)
First check

Test C-arm brake engagement

Test point
C-arm brake assembly
Expected
Brake holds C-arm firmly when engaged

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

C-arm movement restricted when brake disengaged caution

C-arm rotation is not free or encounters resistance when brake is disengaged

Ranked causes

  1. 1. Friction or obstruction in C-arm rotation mechanism (occasional)
First check

Test C-arm movement with brake disengaged

Test point
C-arm rotation mechanism
Expected
Free and unrestricted movement in all directions

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

C-arm rotation mechanism has excessive end-play caution

C-arm rotates with loose or slack movement indicating excessive play in pivot pins or bearings

Ranked causes

  1. 1. Loose internal bolt in C-arm rotation pivot (occasional)
First check

Inspect C-arm rotation mechanism for end-play

Test point
C-arm rotation pivot
Expected
No excessive end-play; rotation smooth without slack

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Laser alignment

Side and rear laser beams must be aligned to crosshairs on template placed on image intensifier

Ranked causes

  1. 1. Laser misalignment requiring adjustment (unknown)
First check

Place laser alignment template on image intensifier

Test point
image intensifier
Expected
template vertical line points to C-arm, fixture does not move

6 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Monitor gamma settings patient safety

Monitor gamma settings must be verified and set to DICOM standard for proper display of medical imaging data.

Ranked causes

  1. 1. Gamma settings require DICOM verification (common)
First check

Access menu settings

Test point
Touchscreen base
Expected
Menu button pressed and menu displayed

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Touchscreen alignment

Touchscreen requires calibration alignment to ensure accurate touch response and coordinate mapping on the monitor display.

Ranked causes

  1. 1. Touchscreen alignment drift or initialization (common)
First check

ELO touchscreen utility accessible

Test point
Windows taskbar
Expected
ELO icon visible in taskbar or accessible via Control Panel

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: X-ray Controller radiation

X-ray controller must be calibrated using calibration software after XRC board service or replacement. Operator must wear lead apron during calibration as radiation is being produced.

Ranked causes

  1. 1. XRC board requires calibration (common)
First check

Exit InSight application and enter System Configuration

Test point
InSight application menu
Expected
System Configuration menu accessible

13 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Collimation adjustment required radiation

C-arm collimator misalignment causes uneven field coverage; circular image appears off-center or shaded unevenly when rotated 180 degrees

Ranked causes

  1. 1. Collimator assembly out of alignment (common)
First check

Disable Simultaneous kV/mA setting

Test point
System Configuration - Physician Preferences tab
Expected
Simultaneous kV/mA set to Disabled

9 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

External video monitor does not show entire screen

External monitor only displays Fluoroscan X-ray image, not full Windows desktop

Ranked causes

  1. 1. Video Controller card not set up properly for full display (common)
First check

Verify Video Playback Settings configuration

Test point
Video Playback Settings dialog, 'Use Pure Video DVD Max' checkbox
Expected
Use Pure Video DVD Max is de-selected (no checkmark)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flex-arm lateral movement restricted caution

Flex-arm does not move freely from right to left or left to right

Ranked causes

  1. 1. Loose or tight mounting hardware at flex-arm (common)
First check

Test flex-arm lateral movement

Test point
Flex-arm assembly
Expected
Free and unrestricted movement right to left and left to right

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flex-arm to C-arm shaft set-screw loose caution

Set-screw at flex-arm to C-arm shaft connection is not properly secured

Ranked causes

  1. 1. Set-screw vibration or incomplete tightening (occasional)
First check

Inspect set-screw security at flex-arm to C-arm shaft junction

Test point
Set-screw at the very end of the flex-arm
Expected
Set-screw is tight and secure

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flex-arm vertical lock not restricting movement caution

Flex-arm moves vertically when lock is engaged; lock does not prevent up and down motion

Ranked causes

  1. 1. Flex-arm lock mechanism malfunction (occasional)
First check

Test flex-arm lock function

Test point
Flex-arm lock mechanism
Expected
Vertical movement restricted when lock is engaged

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flex-arm vertical movement excessive force required caution

Vertical movement of flex-arm requires more than expected force or is difficult to operate when arm is extended

Ranked causes

  1. 1. Friction or mechanical resistance in flex-arm joints (occasional)
First check

Measure force required for vertical flex-arm movement

Test point
Flex-arm when fully extended from cabinet
Expected
Approximately 35 lbs. maximum force required

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Horizontal arm movement restricted caution

Horizontal arm does not move freely or encounters resistance during movement

Ranked causes

  1. 1. Loose or tight mounting hardware (common)
First check

Test horizontal arm movement freedom

Test point
Horizontal arm assembly
Expected
Free and unrestricted movement

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Image resolution verification - mesh test radiation

Measurement of system resolution using Resolution Test Tool Model 141 in 6-inch field

Ranked causes

  1. 1. Resolution below 40 Mesh minimum (unknown)
First check

Select field of view

Test point
InSight application controls
Expected
6-inch field selected

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Image resolution verification - Nuclear Associates test with ADR radiation

Measurement of system resolution using Nuclear Associates Test Tool Model 07-647 with ADR enabled

Ranked causes

  1. 1. Resolution below 30 Mesh minimum with ADR (unknown)
First check

Enable ADR and set noise suppression

Test point
InSight application controls
Expected
ADR enabled; Noise Suppression = High

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Image resolution verification - parallel line pairs radiation

Measurement of system resolution using parallel line pairs test tool in 6-inch field

Ranked causes

  1. 1. Resolution below 7.0 lp/mm minimum (unknown)
First check

Set system parameters for image quality measurement

Test point
System Configuration - Physician Preferences tab
Expected
Simultaneous kV/mA setting Disabled

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Laser does not turn off caution

Laser remains on after switch is pressed to turn off

Ranked causes

  1. 1. Laser power control circuit malfunction (unknown)
First check

Press laser switch to OFF

Test point
laser switch OFF position
Expected
laser turns off within 1 second

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Laser does not turn on

Laser switch does not activate laser beam

Ranked causes

  1. 1. Laser switch not functioning (unknown)
First check

Verify laser switch responds to manual activation

Test point
laser switch
Expected
laser turns on on first push, turns off on second push

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor brightness or contrast controls not functioning

Brightness and contrast adjustment controls on monitor are inoperative or do not change display

Ranked causes

  1. 1. Control potentiometer failure or loose connection (occasional)
First check

Test brightness control adjustment

Test point
Monitor brightness control
Expected
Display brightness changes smoothly across full range

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor X-ray On lights not functioning

X-ray On indicator lights on monitor do not illuminate or function improperly

Ranked causes

  1. 1. Burned out indicator light or electrical connection failure (occasional)
First check

Test X-ray On indicator lights

Test point
Monitor X-ray On light indicator
Expected
Lights illuminate when X-ray is on

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

No video output at external DVI port

External monitor connected to DVI port receives no signal from the system

Ranked causes

  1. 1. Video Controller card not set up properly (common)
First check

Verify Video Controller configuration in Matrox PowerDesk-HF

Test point
Multi-Display Setup dialog
Expected
2 displays. Clone mode (including features) is selected

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power cord insulation damaged high voltage

Power cord shows signs of damage such as cuts, splits, fraying, or other deterioration of insulation

Ranked causes

  1. 1. Mechanical damage to power cord (common)
First check

Visual inspection of power cord insulation

Test point
Full length of power cord from mains to system inlet
Expected
No cuts, splits, fraying, or other damage visible

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

X-ray image not circular during C-arm movement

X-ray image on monitor changes shape, distorts, or is not circular when C-arm is moved in any direction

Ranked causes

  1. 1. C-arm misalignment or mechanical binding (occasional)
First check

Observe X-ray image while moving C-arm through full range

Test point
Monitor image during C-arm movement
Expected
Image remains circular in all directions

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
02Expected next character to be end-of-message, but it wasn'tMessage parsing error during XRC communicationwarning
03The buffer for incoming characters is fullExcessive incoming message traffic or buffer not clearedwarning
04Bad value for a control code inside the receive message routineInvalid control code in incoming messagewarning
05Can't find a receive queue for an incoming message (they're all in use)All receive queues occupiedwarning
08Bad value for the first character in a sequence number (for an incoming message)Invalid sequence number first characterwarning
09Bad value for the second character in a sequence number or the whole number is badInvalid sequence number format in incoming messagewarning
0ANo transmit queue available for reply to Get Version Number commandGet Version Number command received with no available queuewarning
0BBad value for first character of command code (incoming message)Invalid command code first characterwarning
0CBad value for second character of command code (incoming message)Invalid command code second characterwarning
0DCommand code value is too big (incoming message)Command code exceeds maximum valid valuewarning
11Bad value for first character of check sum (incoming message)Invalid checksum first characterwarning
12Bad value for second character of check sum, or back check sumInvalid checksum format in incoming messagewarning
13Bad value for data character (incoming message)Invalid data character in messagewarning
14Bad value for string data character (incoming message)Invalid character in string data fieldwarning
20Could not find transmit queue for outgoing message (they're all in use)All transmit queues occupied when sending messagewarning
23Message has been transmitted, but transmit queue was not freedQueue not released after message transmissionwarning
24Message has been transmitted, but transmit queue was not freed (different place in code)Queue not released after message transmission (alternate code path)warning
26Current sequence number (for outgoing message) is still in useAttempting to reuse an active sequence numberwarning
27Current sequence number is still in use and all transmit queues are in useSequence number active and no available queueswarning
28Current sequence number is not available for outgoing message (different place in code)Sequence number allocation failure in alternate code pathwarning
29All the transmit queues are in useAll transmit queues occupiedwarning
2AAn unrecognized state has occurred in the function that finds transmit queuesInvalid state in queue allocation functionerror
30The event list is out of spaceEvent list buffer fullwarning
31The bus used to communicate with the I2C board is not freeI2C bus busy or lockedwarning
32The bus used to communicate with the I2C board is not free (different place in code)I2C bus busy or locked (alternate code path)warning
33The bus used to communicate with the I2C board is not free (different place in code)I2C bus busy or locked (another alternate code path)warning
38Error in I2C board communications codeI2C board communication failurewarning
39Error in I2C board communications codeI2C board communication failurewarning
3AError in I2C board communications codeI2C board communication failurewarning
3BError in I2C board communications codeI2C board communication failurewarning
3CError in I2C board communications codeI2C board communication failurewarning
3DError in I2C board communications codeI2C board communication failurewarning
3EError in I2C board communicationsI2C board communication failurewarning
40Initial KV value is too far from target (during calibration)X-ray tube kilovoltage initialization error during calibrationcritical
41Initial MA value is too far from target (during calibration)X-ray tube milliamperage initialization error during calibrationcritical
42Maximum possible input for KV does not reach target value (during calibration)KV cannot reach target at maximum input during calibrationcritical
43Maximum possible input for MA does not reach target value (during calibration)MA cannot reach target at maximum input during calibrationcritical
44Target KV value not obtained after maximum number of retries (during calibration)KV fails to stabilize after max retries during calibrationcritical
45Target MA value not obtained after maximum number of retries (during calibration)MA fails to stabilize after max retries during calibrationcritical
46Error in C-Arm I/O board communicationsC-Arm I/O board communication failurewarning
47Error in C-Arm I/O board communicationsC-Arm I/O board communication failurewarning
48Error in C-Arm I/O board communicationsC-Arm I/O board communication failurewarning
49Error in C-Arm I/O board communicationsC-Arm I/O board communication failurewarning
4AError in C-Arm I/O board communicationsC-Arm I/O board communication failurewarning
4BError in C-Arm I/O board communicationsC-Arm I/O board communication failurewarning
4CError in C-Arm I/O board communicationsC-Arm I/O board communication failurewarning
4DError in C-Arm I/O board communicationsC-Arm I/O board communication failurewarning
4EError in C-Arm I/O board communicationsC-Arm I/O board communication failurewarning
50Error transmitting characters to host PCHost PC communication transmission failurewarning
51Error transmitting characters to host PCHost PC communication transmission failurewarning
52Error transmitting characters to host PCHost PC communication transmission failurewarning
53Error transmitting characters to host PCHost PC communication transmission failurewarning
54Error transmitting characters to host PCHost PC communication transmission failurewarning
56Error communicating with the X-ray power supply hardwareX-ray power supply communication failurewarning
57Error communicating with the X-ray power supply hardwareX-ray power supply communication failurewarning
58Error communicating with the X-ray power supply hardwareX-ray power supply communication failurewarning
59Error communicating with the X-ray power supply hardwareX-ray power supply communication failurewarning
5 Minute X-Ray Limit ExceededContinuous X-ray time limit exceeded; X-rays turned offX-ray system left on continuously for 5 minuteswarning
80Error communicating with the EEPROMEEPROM communication failurewarning
81Error communicating with the EEPROMEEPROM communication failurewarning
82Error communicating with the EEPROMEEPROM communication failurewarning
83Error communicating with the X-ray power supply hardwareX-ray power supply communication failurewarning
84Error communicating with the X-ray power supply hardwareX-ray power supply communication failurewarning
85Error communicating with the X-ray power supply hardwareX-ray power supply communication failurewarning
86Error communicating with the X-ray power supply hardwareX-ray power supply communication failurewarning
87An invalid interrupt has been triggeredUnexpected interrupt event detectederror
89Error communicating with the EEPROMEEPROM communication failurewarning
8AError communicating with the EEPROMEEPROM communication failurewarning
8BError communicating with the EEPROMEEPROM communication failurewarning
8CError communicating with the EEPROMEEPROM communication failurewarning
8DError communicating with the EEPROMEEPROM communication failurewarning
8EError communicating with the EEPROMEEPROM communication failurewarning
8FError communicating with the EEPROMEEPROM communication failurewarning
90Error communicating with the EEPROMEEPROM communication failurewarning
91Error communicating with the EEPROMEEPROM communication failurewarning
92Error communicating with the EEPROMEEPROM communication failurewarning
93Error communicating with the EEPROMEEPROM communication failurewarning
94Error communicating with the EEPROMEEPROM communication failurewarning
95Error communicating with the EEPROMEEPROM communication failurewarning
96Error communicating with the EEPROMEEPROM communication failurewarning
99Error in I2C board communicationsI2C board communication failurewarning
9DError in I2C board communicationsI2C board communication failurewarning
9EError in I2C board communicationsI2C board communication failurewarning
9FError in I2C board communicationsI2C board communication failurewarning
A1Error in updating XRC statusXRC status register update failurewarning
A4An impossible condition has occuredUnexpected system state detectedcritical
A5An invalid interrupt has been triggeredUnexpected interrupt event detectederror
A6An invalid interrupt has been triggeredUnexpected interrupt event detectederror
A7An invalid interrupt has been triggeredUnexpected interrupt event detectederror
A8An invalid interrupt has been triggeredUnexpected interrupt event detectederror
A9An invalid interrupt has been triggeredUnexpected interrupt event detectederror
ABAn invalid interrupt has been triggeredUnexpected interrupt event detectederror
ACAn un-implemented function has been calledCode execution reached unimplemented functionerror
ACK reply unknown commandXRC received message but command code not recognizedImage acquisition operationwarning
Backup Not Performed 30 DaysSystem backup overdueMore than 30 days since last backup performedwarning
Calibration Check Failed (Non-Service User)System not calibrated; operator cannot proceedCalibration check failure when non-service user logs inerror
Calibration Check Failed (Service User)System needs calibrationCalibration check failure during system startup when logged in as service usererror
Camera communication port has not been specifiedNo serial port assigned for camera communicationMissing port configuration at startupcritical
Can not create a Worklist requestWorklist request creation failedError during DICOM worklist query request creationerror
Can not process Worklist QueryWorklist query processing failedError processing DICOM worklist query with invalid or missing dataerror
CheckCalibration file access errorCould not read calibration file at startupCalibration verification at system startuperror
CImageLoaderReview1::SetFilter() bad dataNULL image list received from session during image review filter setupReview image list operation when no images existwarning
CImageLoaderReview1::UpdateList() bad dataNULL image list for session during review list updateReview image list operation when no images existwarning
CLstListImages::Create() image window errorWindows operating system call failed during image window creationImage list image window creationerror
CLstListImages::CreateLayout() memory errorMemory allocation failed during image list layout creationList Control manipulation operation initiatederror
CLstListImages::CreateLoader() memory errorMemory allocation failed during image list loader creationList Control manipulation operation initiatederror
CLstListImages::Create() system errorWindows operating system call failed during main window creationImage list main window creationerror
CLstListImages::SetupScrollBar() system errorWindows operating system call failed during scroll bar parameter setupScroll bar initialization during list displayerror
CLstListImages::ShowPatientDetails() bad dataSelected image has no associated study recordAttempt to display patient details for image without study datawarning
CLstListImages::SignalDoubleClick() system errorWindows operating system call failed while posting double-click messageDouble-click event in image listerror
CLstListImages::SignalSelection() system errorWindows operating system call failed while posting selection messageImage selection event in list controlerror
CLstPatientDetails::SetStudy() bad dataError reading patient details from databasePatient details display or update operationwarning
CLstStudyList::CreateLayout() memory errorMemory allocation failed during study list layout creationList Control manipulation operation initiatederror
CLstStudyList::CreateLoader() memory errorMemory allocation failed during study list loader creationList Control manipulation operation initiatederror
CLstVistaDetails::SetStudy() bad dataError reading VISTA details for patientVISTA details retrieval operationwarning
Collimator Invalid PositionCollimator is in an invalid position; X-rays cannot be generatedDuring image acquisition or when starting X-rays with collimator not in valid positionerror
Collimator Invalid Position (Acquisition)Collimator in invalid position; X-ray generation blockedCollimator error prevents starting image acquisitionerror
Computation ErrorMathematical or computation operation failureError during data calculation or processingerror
Could not abort the image acquisition processFailed to terminate active acquisitionAcquisition abort command failurecritical
Could not create Sapera Acquisition objectImage acquisition driver object instantiation failedSapera acquisition module initialization failurecritical
Could not create Sapera Buffer objectImage buffer management object instantiation failedSapera buffer module initialization failurecritical
Could not create Sapera Flat Field objectFlat Field correction object instantiation failedSapera Flat Field module initialization failurecritical
Could not create Sapera Flat Field Transfer objectFlat Field data transfer object instantiation failedSapera Flat Field transfer module initialization failurecritical
Could not create Sapera Processor objectImage processing library failed to instantiate processorHost image processing initialization failurecritical
Could not create Sapera RtPro objectImage Board driver failed to instantiate imaging objectImage Board initialization failurecritical
Could not create Sapera Transfer objectData transfer object instantiation failedSapera transfer module initialization failurecritical
Could not create Sapera View objectImage display object instantiation failedSapera view module initialization failurecritical
Could not create the Image Board objectImage Board driver object instantiation failedImage Board object creation failurecritical
Could not create Video Recorder objectVideo recorder driver object instantiation failedVideo recorder module initialization failurecritical
Could not destroy Image Board buffersImage buffer cleanup operation failedImage Board buffer teardown failurecritical
Could not destroy Sapera Acquisition objectImage acquisition driver object cleanup failedSapera acquisition module shutdown failurecritical
Could not destroy Sapera Buffer objectImage buffer management object cleanup failedSapera buffer module shutdown failurecritical
Could not destroy Sapera Flat Field objectFlat Field correction object cleanup failedSapera Flat Field module shutdown failurecritical
Could not destroy Sapera Flat Field Transfer objectFlat Field data transfer object cleanup failedSapera Flat Field transfer module shutdown failurecritical
Could not destroy Sapera Processor objectImage processing library failed to clean up processorHost image processing shutdown failurecritical
Could not destroy Sapera RtPro objectImage Board driver failed to clean up imaging objectImage Board shutdown failurecritical
Could not destroy Sapera Transfer objectData transfer object cleanup failedSapera transfer module shutdown failurecritical
Could not destroy Sapera View objectImage display object cleanup failedSapera view module shutdown failurecritical
Could not get access to the image bufferFailed to obtain reference to image memory bufferBuffer access operation failurecritical
Could not load the image bitmapFailed to read image file from storageImage load operation failurecritical
Could not save the image bitmapFailed to write image data to storageImage save operation failurecritical
Could not start the image acquisition processImage Board failed to begin data captureAcquisition start command failurecritical
Could not start video recordingFailed to begin video file recording operationVideo recording start operation failurecritical
Could not stop the image acquisition process, abortingFailed to stop ongoing acquisition, attempting abortAcquisition stop command failurecritical
Create KeepAlive timer failedCould not restart keep-alive loopSystem startup or error recoverycritical
Create message timeout timer failedCould not create outgoing message timeout timerCreating message reply timererror
Create OVERLAPPED event handle failedError starting basic message transmit or receive loopSystem initializationcritical
Create reply timeout timer failedFailed to create timer for incoming message processingProcessing incoming messageerror
CStudyLayoutXxx::UpdateRow() bad dataFailed to update display row due to missing study dataPatient or worklist data display refreshwarning
Data file upgrade conversion failedData upgrade process creation errorError occurred during creation of data upgrade application processwarning
Delayed ReadFile failureLow-level read operation failed for incoming messageReceiving message from XRCerror
DICOM Printer ErrorDICOM printer status cannot be obtained or error occurredDICOM module unable to read DICOM printer statuserror
DICOM Printer Out of FilmDICOM printer has no film loadedDICOM printer out of film during print operationwarning
Drive Access ErrorCannot access requested drive or mediaError attempting to read/write to unknown path or removable mediaerror
E-22Bad data - format error in stringSystem cannot parse time-span from temp.log filewarning
E-22Memory allocation errorSystem fails to allocate memory for log processingerror
E-22File access error - file cannot be opened for readingSystem attempts to open log file for readingerror
EEProm Check Failed (Non-Service User)EEProm error; recovery requiredEEProm check failure with non-service user logged inerror
EEProm Check Failed (Service User)EEProm data error; system needs calibrationEEProm check failure during startup with service user logged inerror
Error acquiring Flat Field black imageBlack image acquisition failed during flat field calibrationError acquiring black image during flat field calibration when calling snap functionerror
Error calculating Flat Field gainsFlat field gain calculation failedError during flat field gain calculationerror
Error calculating Flat Field offsetsFlat field offset calculation failedError while calculating flat field gains and flat field offsetserror
Error changing the image displayFailed to modify image display parameters or configurationImage display modification operation failurecritical
Error cleaning up the temp directory for a new sessionFailed to remove temporary session filesTemporary directory cleanup operation failurecritical
Error communicating with a cameraCamera port communication failureError while opening the camera port during flat field or warp calibrationerror
Error communicating with X-ray controllerX-ray control communication failureError while enabling/disabling X-rays or setting power levelerror
Error copying warp bitmap fileWarp bitmap file copy failed in demo modeError copying warp bitmap file while in demo mode using previous acquired imageerror
Error creating a directory for cine loopFailed to create cine sequence storage directoryCine directory creation failurecritical
Error creating image directoryFailed to create image storage directoryImage directory creation failurecritical
Error creating Sapera objectsMultiple Sapera library object instantiation failedSapera framework initialization failurecritical
Error creating the image displayFailed to initialize image display window or surfaceImage display creation operation failurecritical
Error creating video fileFailed to create video output fileVideo file creation failurecritical
Error creating video streamFailed to initialize video stream objectVideo stream creation operation failurecritical
Error destroying Sapera objectsMultiple Sapera library object cleanup failedSapera framework shutdown failurecritical
Error destroying the Image Board objectImage Board driver object cleanup failedImage Board object destruction failurecritical
Error displaying bitmap for demo acquisitionFailed to display demonstration mode image dataDemo image display operation failurecritical
Error emptying bufferFailed to clear image buffer contentsBuffer clear operation failurecritical
Error getting buffer addressFailed to retrieve memory address of image bufferBuffer address lookup operation failurecritical
Error getting comm stateFailed to retrieve camera communication stateCamera state query operationcritical
Error getting raw image dataFailed to access unprocessed image buffer dataRaw image data access failurecritical
Error getting Xray power from XRCFailed to query X-ray power status from controllerX-ray power status query failurecritical
Error initializing Internal Image Board buffersInternal image board buffer initialization failedInternal buffer setup failurecritical
Error initializing Sapera Image Board buffersSapera image buffer initialization failedImage Board buffer setup failurecritical
Error initializing X-ray controllerXRC controller initialization failedError while initializing the XRC controllererror
Error loading bitmap file from the hard driveFailed to read bitmap file from storage deviceBitmap file read operation failurecritical
Error loading flat field calibration dataFailed to apply Flat Field calibration to systemFlat Field data application failurecritical
Error loading imageImage file load failed during flat field calibrationError while loading .bmp or .tif file selected by user during flat field calibrationerror
Error loading the image from fileFailed to retrieve image data from diskImage file read operation failurecritical
Error loading warp calibration dataFailed to apply warp calibration to systemWarp data application failurecritical
Error opening serial port for camera communicationsUnable to establish serial communication with imaging cameraCamera communications initialization failurecritical
Error processing camera commandCamera did not process commanded instruction correctlyCommand execution failurecritical
Error processing display flags for static imageStatic image display parameters could not be processedDisplay flags processing failure for static imagecritical
Error processing FPGA dark fluoroDark fluoroscopy filter failed in FPGAFPGA dark fluoro operationcritical
Error processing FPGA edge enhancementEdge enhancement filter failed in FPGAFPGA edge enhancement operationcritical
Error processing FPGA flipImage flip operation failed in FPGAFPGA flip operationcritical
Error processing FPGA motion detectionMotion detection function failed in FPGAFPGA motion detection operationcritical
Error processing FPGA noise suppressionNoise suppression filter failed in FPGAFPGA noise suppression operationcritical
Error processing FPGA rotationImage rotation function failed in FPGAFPGA image rotation operationcritical
Error processing live functionReal-time live image processing failedLive image processing operationcritical
Error processing live image function for Dark FluoroReal-time dark fluoroscopy filter failedLive dark fluoro operationcritical
Error processing live image function for Edge EnhancementReal-time edge enhancement filter failedLive edge enhancement operationcritical
Error processing live image function for FlipReal-time image flip operation failedLive flip operationcritical
Error processing live image function for Motion DetectionReal-time motion detection filter failedLive motion detection operationcritical
Error processing live image function for Noise ReductionReal-time noise reduction filter failedLive noise reduction operationcritical
Error processing live image function for Reverse VideoReal-time reverse video filter failedLive reverse video operationcritical
Error processing live image function for RotationReal-time image rotation operation failedLive rotation operationcritical
Error processing Sapera hardware overlayHardware-accelerated display overlay operation failedHardware overlay processing failurecritical
Error processing static image function for AOIStatic Area of Interest operation failed in host processorStatic AOI operation failurecritical
Error processing static image function for Dark FluoroStatic dark fluoroscopy filter failed in host processorStatic dark fluoro operation failurecritical
Error processing static image function for Edge EnhancementStatic image edge enhancement failed in host processorStatic edge enhancement operation failurecritical
Error processing static image function for FlipStatic image flip operation failed in host processorStatic image flip operation failurecritical
Error processing static image function for Reverse VideoStatic reverse video filter failed in host processorStatic reverse video operation failurecritical
Error processing static image function for RotationStatic image rotation operation failed in host processorStatic image rotation operation failurecritical
Error processing static image function for ZoomStatic image zoom operation failed in host processorStatic zoom operation failurecritical
Error reading backup from %1Backup archive read error during restoreError reading archive file while restoring to temporary foldererror
Error reading Flat Field calibration fileFailed to access or load Flat Field calibration dataFlat Field calibration file read operation failurecritical
Error reading from %1Disk read error during backup operationError occurs while reading free space on disk during backuperror
Error reading from camera serial portFailed to receive data from cameraSerial port read operation failurecritical
Error reading target mediaMedia read error during recoveryAny error reading media other than unformatted errorerror
Error reading warp calibration fileFailed to access or load warp calibration dataWarp calibration file read operation failurecritical
Error resetting static flagsFailed to reset static image display flagsStatic flags reset operation failurecritical
Error resetting the FPGA functionsFailed to reset FPGA image processing modulesFPGA reset operation failurecritical
Error response received from cameraCamera returned an error statusInvalid camera response receivedcritical
Error saving display flags into a fileFailed to store display parameters to fileDisplay flags file save operation failurecritical
Error saving Flat Field correction fileFlat field correction file save failedError saving flat field correction file (FF6_FILE or FF4_FILE)error
Error saving image to permanent directoryFailed to save image to long-term storage locationPermanent image save operation failurecritical
Error saving image to temp directoryFailed to save temporary image fileTemporary image save operation failurecritical
Error saving raw imageFailed to save unprocessed image dataRaw image save operation failurecritical
Error saving Warping bitmap fileWarp calibration bitmap save failedError while saving image bitmap for calibration during warp calibrationerror
Error saving Warping calibration fileWarp calibration file save failedError while calculating warp calibration and saving to calibration fileerror
Error setting a window to the Sapera View objectFailed to assign display window to view objectWindow assignment operation failurecritical
Error setting camera gainUnable to set or get camera gain parametersCamera gain adjustment during initialization or operationcritical
Error setting Image Board configuration parametersFailed to apply configuration to Image BoardConfiguration parameter application failurecritical
Error setting internal image board triggerFailed to configure image board acquisition triggerTrigger configuration operationcritical
Error setting internal image board trigger frequencyFailed to configure trigger frequency parameterTrigger frequency configuration operationcritical
Error setting video stream formatFailed to configure video stream parametersVideo stream format configuration failurecritical
Error starting cine loop acquisitionFailed to begin continuous cine sequence captureCine acquisition start failurecritical
Error starting image acquisitionFailed to begin standard image captureImage acquisition start failurecritical
Error starting image acquisitionImage acquisition snap/grab function failedError while calling snap and grab functions of image board during warp and flat field calibrationerror
Error stopping the acquisitionFailed to terminate image acquisition processAcquisition termination failurecritical
Error stopping video recordingFailed to terminate active video recording sessionVideo recording stop operation failurecritical
Error updating image display flagsFailed to modify image display parametersDisplay flags update operation failurecritical
Error updating image presentation state fileFailed to update image metadata state informationState file update operation failurecritical
Error writing backup to %1Backup write to media failedError while writing the archived file to backup mediaerror
Error writing to camera serial portFailed to transmit data to cameraSerial port write operation failurecritical
Error writing video frame to a fileFailed to write video frame data to storageVideo frame write operation failurecritical
Error zooming imageImage zoom operation failedImage zoom processing failurecritical
External Data ErrorInvalid data passed for processingBad data received for processing operationerror
Failed to connect to NetworkNetwork connection failedDICOM worklist query operation attempted without network accesserror
Failed to open association with BrokerDICOM association with broker failedError during DICOM worklist query when opening broker associationerror
Failed to start X-raysX-ray startup failed; calibration mode not setError turning on X-rays while performing calibrationerror
FakeCalibrate file access errorCould not read default calibration fileSystem startup or calibrationerror
File Access ErrorCannot open, read or write specified fileFile system access error during file operationerror
Flat Field buffer has invalid formatFlat Field data in memory has invalid structureFlat Field buffer data corruptioncritical
Flat Field calibration data has not been loadedSystem operating without Flat Field calibration activeFlat Field calibration data missing from memorycritical
Flat Field calibration file has invalid formatFlat Field calibration file format does not match expected structureFlat Field calibration file parsing failurecritical
Flat Field/Camera Gain calibration failedFlat field or camera gain calibration validation failedCalibration validation failed after pressing Start Flat Field Calibration buttonwarning
Hospital System Access FailedCannot access hospital system; network or PACS errorDICOM unable to open association with PACS or ping failserror
Image Board ErrorCommunication with image board failedError communicating with image board hardwareerror
Image Count WarningUser has reached memory barrier limit for images per sessionUser has taken majority of allowed images per sessioninfo
Image has invalid number of bits allocatedImage bit allocation parameter is invalid or corruptedImage header validation failurecritical
Image has invalid number of bits storedImage bit storage parameter is invalid or corruptedImage header validation failurecritical
Image has invalid number of columnsImage width parameter is invalid or corruptedImage dimension validation failurecritical
Image has invalid number of rowsImage height parameter is invalid or corruptedImage dimension validation failurecritical
Image Limit Per SessionMaximum images per session reached; X-rays disabledUser reaches 200 image limit per sessionwarning
Image Processor Start FailedImage acquisition could not be startedError from image processor module during acquisition starterror
Images Not Saved at Session EndOne or more images from session were not savedUser exits session or switches view without saving all imageswarning
Input buffer overflowIncoming messages from XRC came in too fast to processReceiving messages from XRCwarning
InSight software not foundInSight software not installed or not detectedSoftware verification failed during Image Calibration Dialog initiationerror
Invalid argument has been passed to the functionFunction received an invalid or incompatible parameterInvalid parameter in function callcritical
Invalid camera synchronization modeCamera synchronization setting is incorrect or unsupportedInvalid sync mode configurationcritical
Invalid checksumMessage from XRC has incorrect checksumReceiving message from XRCwarning
Invalid command codeMessage from XRC contains invalid command codeReceiving message from XRCwarning
Invalid Image Board configuration parameterConfiguration file contains invalid or unsupported parameterConfiguration file parsing failurecritical
Invalid image processing function specifiedImage processing request uses unsupported or invalid functionInvalid function parameter passed to processorcritical
Invalid User DataUser entered invalid value in data fieldUser enters illegal value in input field during operationwarning
KeepAlive failedX-ray source stopped responding to keep-alive messagesXRC communications monitoringcritical
Live image processing function called for static imageReal-time processing function invoked on non-live image dataWrong processing function used for image typecritical
Log File Export ErrorFailed to export log file to destination mediaError copying log file to export destinationerror
Lost X-Ray Source CommunicationCommunication with X-ray source lost; system cannot proceedCommunication error with X-ray source modulecritical
Maximum number of hits per query reachedQuery result limit exceeded; worklist entries ignoredDICOM worklist query returned more results than configured limitwarning
Maximum X-ray time exceededX-rays on continuously for too longContinuous X-ray operationwarning
Memory Access ErrorSystem error accessing memoryError during memory read/write operationerror
Memory Allocation ErrorSystem unable to create required object due to insufficient memoryMemory allocation failure during object creationerror
Message slot status errorIncoming message received before previous message processedReceiving messagewarning
Message timeout not foundFailed to reply to incoming message and message not foundIncoming message timeout exceededinfo
Message timer timeoutMessage timer timed out but message not foundXRC communication timeoutinfo
Minimum Patient Last Name and Patient ID requiredValidation failed; required patient info missingValidation of patient selection dialog when required fields emptywarning
Minor XRC ErrorMinor XRC error occurredMinor error detected in XRC modulewarning
Missing DICOM ConfigurationDICOM configuration data unavailable; hospital system access blockedDICOM module unable to read DICOM XML configuration fileserror
Modality value may not be emptyModality field validation failedValidation fails in patient worklist query dialogwarning
NAK reply to messageReceived a missed-message response from the XRCImage acquisition communication with XRC modulewarning
No Flat Field calibration file name specifiedFlat Field calibration file path not defined in configurationMissing flat field calibration file referencecritical
No matching patients foundPatient search returned no resultsDetailed query or worklist query returned no patientswarning
No message window specifiedMessage display window not configured in systemMessage window configuration missingcritical
No Storage for ImageCurrent image will not be saved due to storage limitStorage limit reached while attempting to save image during sessionwarning
No Studies matching the filter criteria were foundStudy search returned no resultsUser entered search criteria and selected update; no studies returnedwarning
No warp calibration file name specifiedWarp calibration file path not defined in configurationMissing warp calibration file referencecritical
No X-ray Off event has been receivedX-ray controller did not confirm X-ray shutdownX-ray Off confirmation timeoutcritical
No X-rays - reset timeoutSystem failed to recover from serious XRC errorXRC error recovery attemptcritical
No XRC Interface object specifiedX-ray controller interface not initializedXRC interface configuration missingcritical
Old Password IncorrectCurrent password does not match login passwordUser enters incorrect old password when changing password or fails 3 login attemptswarning
OnAttemptRecovery failedSystem failed to recover from serious XRC errorXRC error recovery attemptcritical
One or more Sapera objects has not been createdRequired Sapera library objects not instantiatedMultiple object creation failurecritical
Outgoing message reply timeoutFailed to reply to incoming message within timeoutProcessing incoming message responsewarning
Out of message slotsToo many pending messages to XRCAttempting to send message when queue fullwarning
Panel Button ErrorPanel control buttons (print, save, rotate, power adjust) not workingXRC error or hardware failure affecting panel controlserror
Password MismatchNew password and confirmation do not matchUser enters mismatched passwords in change password dialoginfo
Patient Selection /UpdateAccession Number validation errorAccession Number value contains backslash charactererror
Patient Selection /Update MessagePatient data modification confirmationUser updates existing patient datainfo
Patient Selection /Update MessageReview confirmation for updated patient dataUser selects review button after updating patient informationinfo
Printer ErrorPrinter settings, hardware, or status errorPrinter settings unreadable, not found, out of memory, or invalid bitmaperror
Printer Out of PaperPrinter has no paper loadedSystem attempts to print when paper tray emptywarning
Programming errorIllegal condition in code indicating software bugVarious, depends on exact error locationunknown
Programming error - illegal code conditionSoftware encountered an illegal code condition indicating a bugVarious list generator module operationscritical
Query failedDICOM worklist query operation failedError returned by DICOM worklist query when user selects query or query detail buttonerror
Query has been abortedUser or system aborted worklist queryDICOM worklist query operation was abortedwarning
Ram RW ErrorRAM read/write test failureError returned from XRC initializationcritical
ReadFile failedLow-level read operation failed for incoming messageReceiving message from XRCerror
Registry Access ErrorSystem unable to access Windows RegistryError reading or writing to Registry during operationerror
Requested association rejected by BrokerDICOM broker rejected the association requestBroker rejected association during DICOM worklist queryerror
Requested Procedure ID invalid charactersProcedure ID field validation failedValidation fails in patient worklist query dialog when invalid characters usedwarning
Resource File Access ErrorUnable to access .RES resource fileError reading or accessing resource fileerror
RestoreCalibration file access errorCould not read or write calibration fileRestoring calibration dataerror
Rom Check Sum ErrorROM data integrity check failedError returned from XRC initializationcritical
RtProDesign object has not been createdFPGA image processing object not instantiatedImage Board design object creation failurecritical
Serial Camera Communication ErrorError in serial communications with cameraCommunication failure with camera moduleerror
Service Log File Access ErrorFailed to add entry to Service Log due to file access errorCannot open specified log file during log entry writeerror
Session information is missingRequired session context data not available for operationSession data missing during processingcritical
Specified pixel depth is not supportedImage bit depth configuration is invalid or unsupportedUnsupported pixel depth value in configurationcritical
Stack Creep ErrorStack overflow detectedError returned from XRC initializationcritical
Starting OnAttemptRecoveryXRC module detected problem and attempting recoveryXRC error detection during operationwarning
Stop KeepAlive timer failedCould not stop keep-alive loopError handling or exit from acquisition modecritical
Stop message timeout timer failedFailed to remove timer for previous messageStopping message timeout timererror
Storage Limit ExceededNo storage available for new imagesImage storage limit reached during pre-login check or image savewarning
System ConfigurationPassword mismatch errorUser enters two non-matching passwordserror
System ConfigurationInvalid hour value errorUser enters invalid hour value in Worklist Configurationerror
System ConfigurationEmpty first name field errorUser leaves First Name field empty in Physician or User pageerror
System ConfigurationNetwork connection errorLocal ping failswarning
System ConfigurationDestination computer not found errorRemote ping returns unsuccessful resulterror
System ConfigurationUnsuccessful echo from destination errorC-echo function returns failureerror
System ConfigurationImage rotation value out of range errorImage rotation field value outside -360 to +360 rangeerror
System ConfigurationHost name or IP address required errorRemote ping or c-echo operation without host informationerror
System Configuration WarningBoth fluoroscopy modes disabled warningBoth continuous and snapshot fluoroscopy disabledwarning
System Failure detectedFluoroscan InSight instrument is not operableError detected during XRCinit call in pre-login checkcritical
System Not Connected to NetworkNetwork connectivity lostLocal ping from review screen failswarning
System Off 2 WeeksSystem has been off for extended period; warmup recommendedSystem has not been powered on for at least 2 weeksinfo
System recover completeRecovery operation successfulSuccessful system recovery from backupinfo
System Recovery SuccessfulSystem successfully recovered from XRC failureSystem recovery completed after XRC errorinfo
The acquisition process is not runningAcquisition operation requested but no active capture sessionAcquisition control command without active sessioncritical
The argument has invalid valueFunction called with invalid parameter valueInvalid argument value in function callcritical
The Image Board object has already been createdImage Board object instantiation attempted when already initializedDuplicate Image Board object creation attemptcritical
The Image Board object has not yet been createdImage Board object accessed before initializationPremature Image Board object accesscritical
There are no image buffers allocatedNo image memory buffers currently reserved for operationBuffer allocation not performedcritical
The video recorder has been already createdVideo recorder object instantiation attempted when already initializedDuplicate video recorder object creation attemptcritical
The video recorder has not yet been createdVideo recorder object accessed before initializationPremature video recorder object accesscritical
The view object has not been created yetImage display object not initialized before useView object accessed before creationcritical
Timeout has been reached waiting for X-ray Off eventX-ray controller did not confirm X-ray shutdown within timeout periodX-ray Off confirmation timeoutcritical
Timeout has been reached waiting for X-ray On eventX-ray controller did not confirm X-ray startup within timeout periodX-ray On confirmation timeoutcritical
Timeout waiting for camera responseCamera did not respond within expected timeframeCamera response timeout during communicationcritical
Trying to invoke real function in demo modeProduction function called when system in demonstration modeMode mismatch in function invocationcritical
Trying to start acquisition while previous one is still runnImage capture start requested while previous acquisition activeOverlapping acquisition start requestwarning
Trying to start cine acquisition while in image modeAttempted continuous cine capture when system in standard image modeMode mismatch during cine start requestcritical
Trying to start image acquisition while in cine modeAttempted standard image capture when system in continuous cine modeMode mismatch during acquisition start requestcritical
Unable to backup required file: %1Backup archive creation failedError while creating the backup archiveerror
Unable to complete system recoverSystem restore failed, reverted to original stateError occurred while extracting backup archiveerror
Unable to Restart SessionPrevious session cannot be recoveredError accessing images or raw data when restarting interrupted sessionerror
Unexpected ErrorUnexpected system error occurredUndefined or unexpected error during operationerror
Unrecoverable XRC ErrorX-rays disabled due to unrecoverable errorXRC error during acquisition that cannot be recoveredcritical
Unsaved Images WarningImages exist that have not been savedUser attempts to exit session without saving imageswarning
UtilitiesNo active DICOM print nodes configured errorUser attempts DICOM print without configured destinationerror
UtilitiesFailed to export image errorExport operation unable to write to destinationerror
UtilitiesInsufficient space on target media errorExport media does not have enough free spaceerror
UtilitiesInsufficient space with capacity details errorSpace validation fails during export with capacity infoerror
UtilitiesInvalid InSight option license key errorUser enters invalid license key during option installationerror
UtilitiesUnformatted media format promptExport operation detects unformatted DVD-RAM mediawarning
WaitForData timeoutXRC did not reply to command within timeout periodWaiting for XRC command responsewarning
Warning: A more recent backup was created on %1User attempting to restore older backupUser tries to recover backup dated before most recent backupwarning
Warp calibration data has not been loadedSystem operating without warp calibration activeWarp calibration data missing from memorycritical
Warp calibration failed to be completedWarp calibration validation failedCalibration validation failed after pressing Start Warp Calibration buttonwarning
Warp calibration file has invalid formatWarp calibration file format does not match expected structureWarp calibration file parsing failurecritical
Worklist Query FailedUnexpected response from hospital system worklist queryDICOM worklist query receives unexpected response from PACSerror
WriteFile failedLow-level message send operation failedSending message to XRCerror
Wrong message sizeMessage from XRC has incorrect formatReceiving message from XRCwarning
X-Ray Continuous Time LimitX-rays on for maximum continuous time allowedX-ray system left on continuously for maximum allowed timewarning
X-ray controller board not set to Calibration ModeXRC not in calibration mode for Image Calibration DialogUser initiated Image Calibration Dialog while XRC not in calibration modewarning
X-Ray Controller ErrorX-ray controller error; X-rays cannot be generatedError while enabling/disabling X-rays or setting power levelerror
X-Ray Enable ErrorX-ray enable circuit failureError returned from XRC initializationcritical
X-Ray General ErrorX-ray system general failureError returned from XRC initializationcritical
X-Ray kV ErrorX-ray tube voltage (kV) control failureError returned from XRC initializationcritical
X-Ray mA ErrorX-ray tube current (mA) control failureError returned from XRC initializationcritical
X-ray Off error received from XRCX-ray controller reported error when turning off X-rayX-ray shutdown command failurecritical
X-Ray Power FailureX-ray power supply failureError returned from XRC initializationcritical
X-Ray Recovery in ProgressX-rays off; system recovering from hardware errorSystem attempting recovery after X-ray disable due to hardware errorwarning
X-Ray Source Command FailedX-ray source command execution failedXRC module unable to execute command sent to X-ray sourceerror
X-Ray Source Communication ErrorError communicating with X-ray sourceCommunication failure with X-ray source during operationerror
X-Ray Source Hardware FailureX-ray source hardware failure; system cannot proceedHardware error in X-ray source detected after X-rays disabledcritical
XRC error codeXRC module reporting internal errorXRC internal operationwarning
XRC link lost - COM port openCannot open COM1 port for XRC communicationSystem startup or initializationcritical
XRC link lost - COM port propertiesCannot configure COM1 port for XRC communicationSystem startup or initializationcritical
XRC reset timeoutSystem failed to recover from serious XRC errorXRC error recovery attemptcritical

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.