Hologic Fluent: troubleshooting and error codes

The ServiceForge knowledge base holds 91 documented symptoms for the Hologic Fluent, with 143 ranked causes, 154 diagnostic checks with test points and expected values, and 38 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Cutting problems with MyoSure device patient safety

MyoSure tissue removal device fails to cut or resect tissue effectively

Ranked causes

  1. 1. MyoSure device not fully inserted into system (common)
  2. 2. Cutting window not aligned with tissue (occasional)
  3. 3. Outflow tubing blocked or clamped (common)
  4. 4. MyoSure device handle bent or damaged (rare)
First check

Verify MyoSure device insertion depth

Test point
MyoSure connector at system
Expected
Fully inserted and flush

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Insufficient uterine distension patient safety

Uterine cavity not adequately expanded during procedure

Ranked causes

  1. 1. Cervix over-dilated (occasional)
  2. 2. Fluid bag clamp closed (common)
  3. 3. Inadequate pressure setting (occasional)
  4. 4. Inflow tube kinked or obstructed (common)
First check

Check fluid bag clamp

Test point
Fluid bag clamp
Expected
Clamp open

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Poor tissue resection

MyoSure device not removing tissue effectively during procedure

Ranked causes

  1. 1. MyoSure device not fully inserted into system (common)
  2. 2. MyoSure window not aimed at tissue (occasional)
  3. 3. Outflow tube kinked or obstructed (common)
  4. 4. MyoSure handle bent (rare)
First check

Verify MyoSure device insertion depth

Test point
MyoSure device connection to system
Expected
Device fully inserted, aligned with system front

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Poor uterine distension patient safety

Inadequate intrauterine fluid distension; uterus not sufficiently expanded for visualization and instrumentation.

Ranked causes

  1. 1. Cervical canal excessively dilated (common)
  2. 2. Fluid bag clamp closed or partially closed (common)
  3. 3. Pressure setting inadequate (common)
  4. 4. Inflow tubing obstructed or kinked (occasional)
First check

Assess cervical dilation

Test point
Cervical os
Expected
Moderate cervical dilation without excessive opening

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Poor visibility during procedure patient safety

Inadequate visualization of surgical field; vision obscured or compromised.

Ranked causes

  1. 1. Fluid bag clamp closed or partially closed (common)
  2. 2. Inadequate intrauterine pressure for fluid distension (occasional)
  3. 3. MyoSure ablation device blocking fluid circulation (occasional)
  4. 4. Outflow stopcock not fully open (occasional)
First check

Verify fluid bag clamp position

Test point
Fluid bag clamp
Expected
Clamp fully open

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Poor visualization

Inadequate visual field during hysteroscopic procedure

Ranked causes

  1. 1. Fluid bag clamp closed (common)
  2. 2. Insufficient intrauterine pressure (occasional)
  3. 3. MyoSure device not advanced to fundus or blade not irrigated (occasional)
  4. 4. Outflow stopcock not fully open (common)
First check

Check fluid bag clamp position

Test point
Fluid bag clamp
Expected
Clamp fully open

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Poor visualization or image quality loss patient safety

Reduced or obscured hysteroscopic field of view during procedure

Ranked causes

  1. 1. Fluid bag clamp closed or partially closed (common)
  2. 2. Intrauterine pressure needs adjustment (occasional)
  3. 3. MyoSure blade obstructing view during tissue removal (occasional)
  4. 4. Outflow stopcock closed or partially closed (common)
First check

Check fluid bag clamp position

Test point
Fluid supply bag clamp
Expected
Fully open

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Problems with MyoSure cutting performance patient safety

MyoSure ablation device not cutting tissue effectively; reduced ablation capability during procedure.

Ranked causes

  1. 1. MyoSure device not fully inserted into system (common)
  2. 2. Cutting window not directed against tissue (common)
  3. 3. Output tubing obstructed or kinked (occasional)
  4. 4. MyoSure handle twisted or misaligned (occasional)
First check

Verify MyoSure device insertion depth

Test point
MyoSure cartridge in system housing
Expected
Device fully inserted into system

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Visibility problems

Poor visualization or obscured view during procedure

Ranked causes

  1. 1. Fluid bag clamp closed (common)
  2. 2. Intrauterine pressure requires adjustment (occasional)
  3. 3. MyoSure device blade area blocked (occasional)
  4. 4. Outflow valve not fully open (occasional)
First check

Check fluid bag clamp position

Test point
Fluid supply bag clamp
Expected
Open

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loss of negative pressure

System cannot maintain vacuum during operation

Ranked causes

  1. 1. Hysteroscope outflow tube not connected to hysteroscope (common)
  2. 2. Drape tubing not connected properly (common)
  3. 3. Outflow tube not connected to outflow port (common)
First check

Verify hysteroscope outflow tube connection

Test point
Hysteroscope outflow port
Expected
Tube securely connected

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loss of suction

Reduced or absent suction flow during procedure

Ranked causes

  1. 1. Outflow tubing disconnected from hysteroscope (common)
  2. 2. Drape tubing disconnected (common)
  3. 3. Outflow port disconnected (common)
First check

Verify hysteroscope outflow connection

Test point
Hysteroscope outflow tubing
Expected
Connected

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loss of suction during procedure patient safety

Hysteroscope or tissue trap suction not functioning; loss of aspiration detected during operation.

Ranked causes

  1. 1. Hysteroscope outlet tubing not connected to hysteroscope (common)
  2. 2. Tissue trap tubing not connected to tissue trap (common)
  3. 3. Output tubing not connected to outflow port (common)
First check

Verify hysteroscope outlet tubing connection

Test point
Hysteroscope outlet port
Expected
Tubing firmly connected to hysteroscope

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Priming error with scope blocked

System detects pressure buildup during priming procedure

Ranked causes

  1. 1. Hysteroscope outlet blocked or obstructed (common)
  2. 2. Hysteroscope in uterus during priming (common)
  3. 3. Inflow tubing clamp or stopcock closed (common)
First check

Verify hysteroscope outlet clear

Test point
Hysteroscope distal opening
Expected
Open, no blockage

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Tubing pressure high in inflow line high voltage

Elevated pressure detected in the inflow tubing during operation

Ranked causes

  1. 1. Hysteroscope tip blocked or dirty (common)
  2. 2. Inflow or outflow stopcock partially closed or blocked (common)
  3. 3. Tubing kinked or clamped near scope (common)
First check

Inspect hysteroscope tip for blockage

Test point
Hysteroscope tip opening
Expected
Clear, no tissue or debris

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flow limited by hysteroscope size - warning 010A000C

Flow volume limited by hysteroscope diameter; system cannot deliver requested flow rate

Ranked causes

  1. 1. Hysteroscope diameter too small for requested flow (common)
  2. 2. Inflow stopcock partially closed or blocked from priming (occasional)
First check

Verify inflow stopcock fully open

Test point
Inflow stopcock on hysteroscope
Expected
Stopcock completely open

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

High pressure condition - warning 010A0004 high voltage

Momentary high-pressure condition sustained for more than 2 seconds but less than 5 seconds detected

Ranked causes

  1. 1. Inflow tube between In-FloPak and hysteroscope inlet obstructed or closed (common)
  2. 2. Hysteroscope tip positioned against fundus (common)
First check

Verify inflow tube is open

Test point
Inflow line from In-FloPak to hysteroscope
Expected
Line open and fluid flowing

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

High pressure detected in inflow line patient safety

System detects high pressure in inflow tubing. Displays error with instruction to open inflow line or fluid bag clamp.

Ranked causes

  1. 1. Hysteroscope blocked or obstructed (common)
  2. 2. Inflow tubing kinked or clamped (common)
First check

Verify hysteroscope patency

Test point
Hysteroscope tip
Expected
Clear, no blockage or occlusion

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

High pressure in inlet tubing high voltage

High pressure detected in inlet tubing

Ranked causes

  1. 1. Scope blocked or clogged (common)
  2. 2. Inlet or outlet valve closed or tubing kinked (common)
First check

Inspect scope for blockage

Test point
Scope tip and channels
Expected
Clean and patent

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

High pressure warning - momentary caution

Momentary high pressure condition lasting 2-5 seconds

Ranked causes

  1. 1. Inflow line blocked or restricted (common)
  2. 2. Scope tip touching uterine fundus (common)
First check

Verify inflow line patency

Test point
In-FloPak to scope inlet connection
Expected
Open and unobstructed

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

High tube pressure detected - error 020A0005 high voltage

Excessive pressure measured in inflow tubing; blockage or obstruction detected

Ranked causes

  1. 1. Hysteroscope tip blocked or clogged (common)
  2. 2. Inflow or outflow stopcock closed or tube kinked (common)
First check

Inspect hysteroscope tip for blockage

Test point
Hysteroscope tip opening
Expected
Tip clear and open

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

In-FloPak not detected during setup

System fails to recognize In-FloPak insertion; error code 0208000A displayed on touchscreen monitor.

Ranked causes

  1. 1. In-FloPak not fully inserted or seating improperly (common)
  2. 2. In-FloPak cartridge not fully inserted or loose during procedure (occasional)
First check

Verify In-FloPak insertion during setup

Test point
In-FloPak cartridge slot
Expected
In-FloPak fully seated and detected by system

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inadequate fluid supply to pump

System uses more pump fluid than transducer fluid. Supply bag empty, clamped, or not inserted.

Ranked causes

  1. 1. Supply bag empty or clamped (common)
  2. 2. Supply bag not inserted into system (occasional)
First check

Verify supply bag presence and fluid level

Test point
Supply bag holder
Expected
Bag inserted and contains fluid

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

MyoSure device not detected - error 01070001

MyoSure unit not properly connected or malfunctioning; system cannot communicate with device

Ranked causes

  1. 1. MyoSure device not properly connected (common)
  2. 2. MyoSure device defective (rare)
First check

Verify MyoSure device connection

Test point
MyoSure device connector to system
Expected
Device firmly seated and engaged

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

MyoSure motor torque too high - warning 01070002

Excessive torque detected in MyoSure motor; operator instructed to reduce pressure and restart

Ranked causes

  1. 1. Excessive bending pressure applied to MyoSure handle (common)
  2. 2. MyoSure device defective (rare)
First check

Reduce bending pressure and reactivate footpedal

Test point
MyoSure device operation
Expected
Motor torque returns to normal range

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Out-FloPak not detected during setup or procedure

System fails to recognize Out-FloPak insertion; error code 0209000A displayed on touchscreen monitor.

Ranked causes

  1. 1. Waste bag tube connector not oriented vertically (common)
  2. 2. Out-FloPak not fully inserted or loose during procedure (occasional)
First check

Verify waste bag tube connector orientation

Test point
Waste bag tube connector
Expected
Connector oriented vertically

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pressure limited by hysteroscope size - warning 010A000B

Set pressure exceeds maximum allowed for selected hysteroscope diameter; system limiting pressure

Ranked causes

  1. 1. Hysteroscope diameter too small for requested pressure (common)
  2. 2. Inflow stopcock partially closed or blocked from priming (occasional)
First check

Verify inflow stopcock fully open

Test point
Inflow stopcock on hysteroscope
Expected
Stopcock completely open

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Startup error - hysteroscope obstruction patient safety

System detects elevated pressure during startup. Startup screen instructs to ensure hysteroscope tip is not blocked. Error code 020A0007.

Ranked causes

  1. 1. Hysteroscope tip blocked or obstructed at startup (common)
  2. 2. Hysteroscope tubing pinched or clamped (occasional)
First check

Verify hysteroscope position before startup

Test point
Hysteroscope location
Expected
Outside uterus, tip oriented to air, not inserted

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Supply bag empty or clamp closed

No fluid being drawn from supply bag or insufficient flow

Ranked causes

  1. 1. Supply bag empty or bag missing (common)
  2. 2. Supply bag tubing clamped or blocked (common)
First check

Check supply bag presence and fluid level

Test point
Supply bag on hooks
Expected
Bag present with fluid visible

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Waste bag not detected during setup

System fails to recognize the waste collection bag during initial setup

Ranked causes

  1. 1. Waste bag tubing connector oriented incorrectly (common)
  2. 2. Waste bag missing or not hung correctly (occasional)
First check

Check waste bag tubing orientation

Test point
Waste bag tubing connection
Expected
Oriented vertically downward

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Load Cell

Annual verification of supply weight and collection weight load cells using 500 gram test weight to ensure accurate fluid tracking

Ranked causes

  1. 1. Load cell calibration required annually (common)
First check

Verify supply weight after zeroing load cells

Test point
Supply Weight display
Expected
0 ± 25 g, ±25 g

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Load cell verification

Annual inspection procedure to verify supply and waste bag load cell accuracy using 500 g weight

Ranked causes

  1. 1. Load cells require calibration verification (unknown)
First check

Zero load cells

Test point
System display
Expected
Supply Weight and Collection Weight both 0 ± 25 g

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pressure caution

Annual verification of pressure sensor calibration using 3000 mL fluid bag to confirm accurate pressure measurement in intra-uterine pressure system

Ranked causes

  1. 1. Pressure sensor calibration required annually (common)
First check

Verify tubing is completely filled with no air bubbles

Test point
Hysteroscope inflow channel
Expected
Fluid flowing with no visible air bubbles

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pressure verification

Annual inspection procedure to verify pressure chamber, pressure sensor and measurement accuracy

Ranked causes

  1. 1. Pressure sensor requires calibration verification (unknown)
First check

Verify pressure reading

Test point
System display (pressure gauge)
Expected
20 ± 5 inH2O, ± 5

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check fluid bag hooks - error 02010026

Fluid bag hooks potentially damaged or excessively pressurized; detection issue noted

Ranked causes

  1. 1. Hooks accidentally touched or supply bag subjected to excessive force (occasional)
First check

Inspect fluid bag hooks for interference

Test point
Supply bag hooks
Expected
Hooks clear, only bag present

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check supply bag hooks caution

Supply bag hooks may be damaged or supply bag forced excessively

Ranked causes

  1. 1. Supply bag hooks manipulated or bag forced too hard (occasional)
First check

Inspect supply hooks for damage

Test point
Supply bag hooks
Expected
Intact, not bent or damaged

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check waste bag hooks caution

Waste bag hooks may be damaged or waste bag forced excessively

Ranked causes

  1. 1. Waste bag hooks manipulated or bag forced too hard (occasional)
First check

Inspect waste hooks for damage

Test point
Waste bag hooks
Expected
Intact, not bent or damaged

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check waste bag hooks - error 02010027

Waste bag hooks potentially damaged or excessively pressurized; detection issue noted

Ranked causes

  1. 1. Hooks accidentally touched or waste bag subjected to excessive force (occasional)
First check

Inspect waste bag hooks for interference

Test point
Waste bag hooks
Expected
Hooks clear, only bag present

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Collection bag overweight caution

More than 15 kg weight hanging on collection bag hook; system displays error

Ranked causes

  1. 1. Excessive weight on collection bag hook (common)
First check

Verify weight on collection bag hook

Test point
Collection bag hook
Expected
Less than 15 kg total

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Deficit error - sudden jump detected patient safety

System detected instantaneous deficit jump of 500 ml or more

Ranked causes

  1. 1. Sudden 500 ml or greater deficit jump (rare)
First check

Record final automatic deficit value

Test point
Deficit display before clearing
Expected
Note value marked as XXX

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Deficit error with sudden fluid loss patient safety

System detects unexpected jump in fluid deficit value of 500 ml or more

Ranked causes

  1. 1. Sudden fluid measurement error or leak (occasional)
First check

Record deficit value at error

Test point
System display
Expected
Specific numerical value shown (e.g., XXX ml)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Deficit limit approaching patient safety

Fluid deficit is within 75% of established limit

Ranked causes

  1. 1. Deficit accumulation near limit (common)
First check

Review current deficit against limit

Test point
Deficit monitoring display
Expected
Below 75% of established limit

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Deficit limit reached or exceeded patient safety

Cumulative fluid deficit has reached or exceeded established limit

Ranked causes

  1. 1. Deficit accumulation exceeds limit (common)
First check

Verify deficit against limit

Test point
Deficit monitoring display
Expected
Below established limit

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flow limited by scope size

Flow rate is limited by scope diameter being used

Ranked causes

  1. 1. Scope diameter too small for requested flow (common)
First check

Verify inlet flow valve open during priming

Test point
Inlet flow valve position
Expected
Fully open

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fluent system unexpectedly shuts down

System powers off or terminates procedure without operator action

Ranked causes

  1. 1. Temporary system fault or power interruption (occasional)
First check

Wait and restart

Test point
System power
Expected
System boots normally and displays menu

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fluid deficit near threshold - warning 01010037

Cumulative fluid deficit has reached 75% of configured limit; warning issued to operator

Ranked causes

  1. 1. Fluid deficit accumulation near limit (common)
First check

Review fluid deficit accumulation

Test point
System fluid deficit display
Expected
Below 75% of threshold

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fluid deficit threshold reached or exceeded - error 02010028 patient safety

Cumulative fluid deficit has met or exceeded configured threshold; procedure must stop

Ranked causes

  1. 1. Fluid deficit limit reached during procedure (common)
First check

Review fluid deficit and clinical status

Test point
System fluid deficit display
Expected
Deficit below threshold or procedure completed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fluid loss limit reached or exceeded patient safety

System indicates fluid loss has reached or exceeded the preset loss limit. Error code 02010028.

Ranked causes

  1. 1. Maximum fluid loss during procedure (common)
First check

Monitor cumulative fluid loss

Test point
Loss display screen
Expected
Loss within preset limit

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fluid loss measurement error detected

System detects sudden fluid loss jump of 500 mL or more. Displays error indicating loss value (XXX).

Ranked causes

  1. 1. Sudden fluid loss spike (occasional)
First check

Monitor fluid loss values during procedure

Test point
Loss display on screen
Expected
Gradual increase without sudden jumps

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

In-FloPak dislodged during procedure

System detects In-FloPak no longer completely inserted

Ranked causes

  1. 1. In-FloPak partially withdrawn or dislodged (common)
First check

Open inlet regulating valve

Test point
Inlet valve on hysteroscope
Expected
Fully open

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

In-FloPak not detected - error 0208000A

In-FloPak has become partially dislodged from system or connection lost during procedure

Ranked causes

  1. 1. In-FloPak not fully seated in system (common)
First check

Open inflow stopcock

Test point
Hysteroscope inflow stopcock
Expected
Stopcock fully open

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

In-FloPak not detected during assembly

System does not recognize In-FloPak insertion; assembly procedure cannot proceed

Ranked causes

  1. 1. In-FloPak not fully inserted or misaligned (common)
First check

Verify In-FloPak is fully inserted and aligned

Test point
In-FloPak connection to system
Expected
In-FloPak flush with system front face

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

In-FloPak not detected or not fully inserted

System detects In-FloPak is not completely inserted. Error code 0208000A.

Ranked causes

  1. 1. In-FloPak not fully seated in connector (common)
First check

Verify In-FloPak insertion

Test point
In-FloPak connector
Expected
Flush with system front, fully seated

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

MyoSure device not found

Device not detected or not fully connected

Ranked causes

  1. 1. Device not fully connected or defective (common)
First check

Verify MyoSure device connection

Test point
Device connector
Expected
Fully and securely connected

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

MyoSure excessive torque detected patient safety

Motor error - excessive force or resistance on MyoSure device

Ranked causes

  1. 1. Excessive bending force on device (common)
First check

Verify adequate clearance and positioning

Test point
MyoSure device in uterus
Expected
Proper positioning with minimal resistance

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Out-FloPak dislodged during procedure

System detects Out-FloPak no longer completely inserted

Ranked causes

  1. 1. Out-FloPak partially withdrawn or dislodged (common)
First check

Open inlet regulating valve

Test point
Inlet valve on hysteroscope
Expected
Fully open

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Out-FloPak not detected - error 0209000A

Out-FloPak has become partially dislodged from system or connection lost during procedure

Ranked causes

  1. 1. Out-FloPak not fully seated in system (common)
First check

Open inflow stopcock

Test point
Hysteroscope inflow stopcock
Expected
Stopcock fully open

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Out-FloPak not detected or not fully inserted

System detects Out-FloPak is not completely inserted. Error code 0209000A.

Ranked causes

  1. 1. Out-FloPak not fully seated in connector (common)
First check

Verify Out-FloPak insertion

Test point
Out-FloPak connector
Expected
Flush with system front, fully seated

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Out-FloPak not fully inserted

System reports Out-FloPak no longer completely inserted; error displayed on interface

Ranked causes

  1. 1. Out-FloPak not seated flush with system front (common)
First check

Verify Out-FloPak is fully seated and flush with system front

Test point
Out-FloPak front face
Expected
Flush with system front panel

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pressure limited by scope size

Adjusted pressure exceeds system capability for scope diameter being used

Ranked causes

  1. 1. Scope diameter too small for requested pressure (common)
First check

Verify inlet valve open during priming

Test point
Inlet flow valve position
Expected
Fully open

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Priming error - scope outlet blocked

System detected pressure buildup during priming procedure

Ranked causes

  1. 1. Scope outlet blocked during priming (common)
First check

Position scope open to air at patient level

Test point
Scope position during priming
Expected
Open to air, not in uterus

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Priming error, hysteroscope tip blocked - error 020A0007
  • Priming suggested
  • Repriming suggested - warning 010A0008
  • Runtime or system error - error 04XXXXXX
  • Supply bag exceeds weight limit
  • Supply bag hook compromised
  • Supply bag low - warning 01010033
  • Supply bag missing
  • Supply bag missing - error 02010023
  • Supply bag overweight
  • Supply bag running low
  • Supply bag too heavy - error 04010020
  • Supply empty, kinked or not punctured
  • Supply empty, kinked, or not connected - error 020A0006
  • System unexpected power loss
  • System unexpected shutdown
  • System unexpectedly powers off
  • Unexpected deficit spike detected - error 02010029
  • Waste bag almost full - warning 01010034
  • Waste bag exceeds weight limit
  • Waste bag full
  • Waste bag full - error 02010025
  • Waste bag full or at capacity
  • Waste bag hook compromised
  • Waste bag lost or misaligned
  • Waste bag missing - error 02010024
  • Waste bag missing or improperly hung
  • Waste bag nearly full
  • Waste bag not detected
  • Waste bag overweight
  • Waste bag too heavy - error 04010021

31 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
01010033Supply bag lowSupply bag contains less than 500 ml of fluidwarning
01010034Waste bag almost fullWaste collection bag is within 500 ml of full capacitywarning
01010037Approaching deficit limitFluid deficit is within 75% of the set limitwarning
01010037Fluid deficit near thresholdCumulative fluid deficit reaches 75% of configured limitwarning
01070001MyoSure device not foundSystem cannot detect MyoSure tissue removal devicewarning
01070002Excessive torque detected, reduce pressure on deviceMyoSure motor error from excessive rotational force on device shaftwarning
01070002MyoSure excessive torque detectedMotor error from excessive force on deviceerror
010A0004Warning - High PressureHigh pressure condition lasting more than 2 seconds but less than 5 secondswarning
010A0008Reprime suggestedUnexpected discrepancy between measured and expected uterine pressurewarning
010A000BPressure limited by scope size being usedSet pressure exceeds maximum allowed for the hysteroscope diameter selectedwarning
010A000CFlow limited by scope size being usedFlow rate is limited by hysteroscope diameter usedwarning
010A000CScope size restricts flow rateSelected scope size limits flow speed during operationwarning
02010023Supply bag missingSupply fluid bag not present or emptyerror
02010024No waste bagWaste bag not properly hung or missingerror
02010024Waste bag missing or improperly positionedWaste bag not correctly suspended on both hooks or absenterror
02010025Waste bag fullWaste bag is completely full and cannot accept more fluiderror
02010026Check supply bag hooksHooks handled or supply bag swung excessivelywarning
02010026Check fluid bag hooksSupply bag hooks detect unexpected pressure or movementerror
02010027Check waste bag hooksWaste bag hooks detect unexpected pressure or movementerror
02010028Deficit limit reached or exceededCumulative fluid deficit equals or exceeds set limiterror
02010029Deficit error with sudden jump in valueSystem detects sudden 500 ml or greater jump in measured deficiterror
02010029Deficit error. Record last value, zero deficit, proceed, use manual count.System detected instantaneous deficit jump of 500 mL or moreerror
02010029Waste value error; at least 500 mL change detectedSystem detects ≥500 mL change in waste valueerror
02010029Fluid loss measurement error - sudden loss jump detectedSystem detects 500mL or more instantaneous loss jumpwarning
02010029Unexpected fluid deficit spikeSystem detects 500 mL or greater sudden change in fluid balanceerror
0208000AIn-FloPak not detected - open stopcock, push in, clear errorSystem detects In-FloPak (blue inlet cartridge) no longer fully insertederror
0208000AIn-FloPak not detected. Open inflow stopcock, push blue In-FloPak back in, clear error.System detected In-FloPak no longer fully insertederror
0209000AOut-FloPak not detected - open stopcock, push in, clear errorSystem detects Out-FloPak (yellow outlet cartridge) no longer fully insertederror
0209000AOut-FloPak not detected. Open inflow stopcock, push yellow Out-FloPak back in, clear error.System detected Out-FloPak no longer fully insertederror
020A0005Inlet tubing pressure highHigh pressure detected in inlet tubingwarning
020A0005Tubing pressure high - open inflow or unclampHigh pressure detected in inflow tubingerror
020A0006Supply empty, clamped or not spikedPump-measured fluid quantity exceeds load cell-measured fluid quantityerror
020A0006Inadequate fluid supply from pumpMore pump fluid used than transducer fluid availableerror
020A0007Startup pressure error - hysteroscope obstructionElevated pressure detected during system startuperror
020A0007Priming error, ensure scope end not blockedSystem detected pressure buildup during primingerror
04010020Supply bag overweightMore than 15 kg of weight on supply bag hookscritical
04010021Waste bag overweightMore than 15 kg of weight on waste bag hookscritical
04XXXXXXVarious runtime or system errorsVarious system fault conditions during operationcritical

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI