Hologic Fluent: troubleshooting and error codes
The ServiceForge knowledge base holds 91 documented symptoms for the Hologic Fluent, with 143 ranked causes, 154 diagnostic checks with test points and expected values, and 38 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Cutting problems with MyoSure device patient safety
MyoSure tissue removal device fails to cut or resect tissue effectively
Ranked causes
- 1. MyoSure device not fully inserted into system (common)
- 2. Cutting window not aligned with tissue (occasional)
- 3. Outflow tubing blocked or clamped (common)
- 4. MyoSure device handle bent or damaged (rare)
Verify MyoSure device insertion depth
- Test point
- MyoSure connector at system
- Expected
- Fully inserted and flush
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Insufficient uterine distension patient safety
Uterine cavity not adequately expanded during procedure
Ranked causes
- 1. Cervix over-dilated (occasional)
- 2. Fluid bag clamp closed (common)
- 3. Inadequate pressure setting (occasional)
- 4. Inflow tube kinked or obstructed (common)
Check fluid bag clamp
- Test point
- Fluid bag clamp
- Expected
- Clamp open
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor tissue resection
MyoSure device not removing tissue effectively during procedure
Ranked causes
- 1. MyoSure device not fully inserted into system (common)
- 2. MyoSure window not aimed at tissue (occasional)
- 3. Outflow tube kinked or obstructed (common)
- 4. MyoSure handle bent (rare)
Verify MyoSure device insertion depth
- Test point
- MyoSure device connection to system
- Expected
- Device fully inserted, aligned with system front
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor uterine distension patient safety
Inadequate intrauterine fluid distension; uterus not sufficiently expanded for visualization and instrumentation.
Ranked causes
- 1. Cervical canal excessively dilated (common)
- 2. Fluid bag clamp closed or partially closed (common)
- 3. Pressure setting inadequate (common)
- 4. Inflow tubing obstructed or kinked (occasional)
Assess cervical dilation
- Test point
- Cervical os
- Expected
- Moderate cervical dilation without excessive opening
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor visibility during procedure patient safety
Inadequate visualization of surgical field; vision obscured or compromised.
Ranked causes
- 1. Fluid bag clamp closed or partially closed (common)
- 2. Inadequate intrauterine pressure for fluid distension (occasional)
- 3. MyoSure ablation device blocking fluid circulation (occasional)
- 4. Outflow stopcock not fully open (occasional)
Verify fluid bag clamp position
- Test point
- Fluid bag clamp
- Expected
- Clamp fully open
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor visualization
Inadequate visual field during hysteroscopic procedure
Ranked causes
- 1. Fluid bag clamp closed (common)
- 2. Insufficient intrauterine pressure (occasional)
- 3. MyoSure device not advanced to fundus or blade not irrigated (occasional)
- 4. Outflow stopcock not fully open (common)
Check fluid bag clamp position
- Test point
- Fluid bag clamp
- Expected
- Clamp fully open
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor visualization or image quality loss patient safety
Reduced or obscured hysteroscopic field of view during procedure
Ranked causes
- 1. Fluid bag clamp closed or partially closed (common)
- 2. Intrauterine pressure needs adjustment (occasional)
- 3. MyoSure blade obstructing view during tissue removal (occasional)
- 4. Outflow stopcock closed or partially closed (common)
Check fluid bag clamp position
- Test point
- Fluid supply bag clamp
- Expected
- Fully open
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Problems with MyoSure cutting performance patient safety
MyoSure ablation device not cutting tissue effectively; reduced ablation capability during procedure.
Ranked causes
- 1. MyoSure device not fully inserted into system (common)
- 2. Cutting window not directed against tissue (common)
- 3. Output tubing obstructed or kinked (occasional)
- 4. MyoSure handle twisted or misaligned (occasional)
Verify MyoSure device insertion depth
- Test point
- MyoSure cartridge in system housing
- Expected
- Device fully inserted into system
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Visibility problems
Poor visualization or obscured view during procedure
Ranked causes
- 1. Fluid bag clamp closed (common)
- 2. Intrauterine pressure requires adjustment (occasional)
- 3. MyoSure device blade area blocked (occasional)
- 4. Outflow valve not fully open (occasional)
Check fluid bag clamp position
- Test point
- Fluid supply bag clamp
- Expected
- Open
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of negative pressure
System cannot maintain vacuum during operation
Ranked causes
- 1. Hysteroscope outflow tube not connected to hysteroscope (common)
- 2. Drape tubing not connected properly (common)
- 3. Outflow tube not connected to outflow port (common)
Verify hysteroscope outflow tube connection
- Test point
- Hysteroscope outflow port
- Expected
- Tube securely connected
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of suction
Reduced or absent suction flow during procedure
Ranked causes
- 1. Outflow tubing disconnected from hysteroscope (common)
- 2. Drape tubing disconnected (common)
- 3. Outflow port disconnected (common)
Verify hysteroscope outflow connection
- Test point
- Hysteroscope outflow tubing
- Expected
- Connected
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of suction during procedure patient safety
Hysteroscope or tissue trap suction not functioning; loss of aspiration detected during operation.
Ranked causes
- 1. Hysteroscope outlet tubing not connected to hysteroscope (common)
- 2. Tissue trap tubing not connected to tissue trap (common)
- 3. Output tubing not connected to outflow port (common)
Verify hysteroscope outlet tubing connection
- Test point
- Hysteroscope outlet port
- Expected
- Tubing firmly connected to hysteroscope
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Priming error with scope blocked
System detects pressure buildup during priming procedure
Ranked causes
- 1. Hysteroscope outlet blocked or obstructed (common)
- 2. Hysteroscope in uterus during priming (common)
- 3. Inflow tubing clamp or stopcock closed (common)
Verify hysteroscope outlet clear
- Test point
- Hysteroscope distal opening
- Expected
- Open, no blockage
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tubing pressure high in inflow line high voltage
Elevated pressure detected in the inflow tubing during operation
Ranked causes
- 1. Hysteroscope tip blocked or dirty (common)
- 2. Inflow or outflow stopcock partially closed or blocked (common)
- 3. Tubing kinked or clamped near scope (common)
Inspect hysteroscope tip for blockage
- Test point
- Hysteroscope tip opening
- Expected
- Clear, no tissue or debris
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow limited by hysteroscope size - warning 010A000C
Flow volume limited by hysteroscope diameter; system cannot deliver requested flow rate
Ranked causes
- 1. Hysteroscope diameter too small for requested flow (common)
- 2. Inflow stopcock partially closed or blocked from priming (occasional)
Verify inflow stopcock fully open
- Test point
- Inflow stopcock on hysteroscope
- Expected
- Stopcock completely open
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
High pressure condition - warning 010A0004 high voltage
Momentary high-pressure condition sustained for more than 2 seconds but less than 5 seconds detected
Ranked causes
- 1. Inflow tube between In-FloPak and hysteroscope inlet obstructed or closed (common)
- 2. Hysteroscope tip positioned against fundus (common)
Verify inflow tube is open
- Test point
- Inflow line from In-FloPak to hysteroscope
- Expected
- Line open and fluid flowing
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
High pressure detected in inflow line patient safety
System detects high pressure in inflow tubing. Displays error with instruction to open inflow line or fluid bag clamp.
Ranked causes
- 1. Hysteroscope blocked or obstructed (common)
- 2. Inflow tubing kinked or clamped (common)
Verify hysteroscope patency
- Test point
- Hysteroscope tip
- Expected
- Clear, no blockage or occlusion
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
High pressure in inlet tubing high voltage
High pressure detected in inlet tubing
Ranked causes
- 1. Scope blocked or clogged (common)
- 2. Inlet or outlet valve closed or tubing kinked (common)
Inspect scope for blockage
- Test point
- Scope tip and channels
- Expected
- Clean and patent
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
High pressure warning - momentary caution
Momentary high pressure condition lasting 2-5 seconds
Ranked causes
- 1. Inflow line blocked or restricted (common)
- 2. Scope tip touching uterine fundus (common)
Verify inflow line patency
- Test point
- In-FloPak to scope inlet connection
- Expected
- Open and unobstructed
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
High tube pressure detected - error 020A0005 high voltage
Excessive pressure measured in inflow tubing; blockage or obstruction detected
Ranked causes
- 1. Hysteroscope tip blocked or clogged (common)
- 2. Inflow or outflow stopcock closed or tube kinked (common)
Inspect hysteroscope tip for blockage
- Test point
- Hysteroscope tip opening
- Expected
- Tip clear and open
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
In-FloPak not detected during setup
System fails to recognize In-FloPak insertion; error code 0208000A displayed on touchscreen monitor.
Ranked causes
- 1. In-FloPak not fully inserted or seating improperly (common)
- 2. In-FloPak cartridge not fully inserted or loose during procedure (occasional)
Verify In-FloPak insertion during setup
- Test point
- In-FloPak cartridge slot
- Expected
- In-FloPak fully seated and detected by system
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inadequate fluid supply to pump
System uses more pump fluid than transducer fluid. Supply bag empty, clamped, or not inserted.
Ranked causes
- 1. Supply bag empty or clamped (common)
- 2. Supply bag not inserted into system (occasional)
Verify supply bag presence and fluid level
- Test point
- Supply bag holder
- Expected
- Bag inserted and contains fluid
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
MyoSure device not detected - error 01070001
MyoSure unit not properly connected or malfunctioning; system cannot communicate with device
Ranked causes
- 1. MyoSure device not properly connected (common)
- 2. MyoSure device defective (rare)
Verify MyoSure device connection
- Test point
- MyoSure device connector to system
- Expected
- Device firmly seated and engaged
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
MyoSure motor torque too high - warning 01070002
Excessive torque detected in MyoSure motor; operator instructed to reduce pressure and restart
Ranked causes
- 1. Excessive bending pressure applied to MyoSure handle (common)
- 2. MyoSure device defective (rare)
Reduce bending pressure and reactivate footpedal
- Test point
- MyoSure device operation
- Expected
- Motor torque returns to normal range
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Out-FloPak not detected during setup or procedure
System fails to recognize Out-FloPak insertion; error code 0209000A displayed on touchscreen monitor.
Ranked causes
- 1. Waste bag tube connector not oriented vertically (common)
- 2. Out-FloPak not fully inserted or loose during procedure (occasional)
Verify waste bag tube connector orientation
- Test point
- Waste bag tube connector
- Expected
- Connector oriented vertically
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure limited by hysteroscope size - warning 010A000B
Set pressure exceeds maximum allowed for selected hysteroscope diameter; system limiting pressure
Ranked causes
- 1. Hysteroscope diameter too small for requested pressure (common)
- 2. Inflow stopcock partially closed or blocked from priming (occasional)
Verify inflow stopcock fully open
- Test point
- Inflow stopcock on hysteroscope
- Expected
- Stopcock completely open
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Startup error - hysteroscope obstruction patient safety
System detects elevated pressure during startup. Startup screen instructs to ensure hysteroscope tip is not blocked. Error code 020A0007.
Ranked causes
- 1. Hysteroscope tip blocked or obstructed at startup (common)
- 2. Hysteroscope tubing pinched or clamped (occasional)
Verify hysteroscope position before startup
- Test point
- Hysteroscope location
- Expected
- Outside uterus, tip oriented to air, not inserted
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Supply bag empty or clamp closed
No fluid being drawn from supply bag or insufficient flow
Ranked causes
- 1. Supply bag empty or bag missing (common)
- 2. Supply bag tubing clamped or blocked (common)
Check supply bag presence and fluid level
- Test point
- Supply bag on hooks
- Expected
- Bag present with fluid visible
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Waste bag not detected during setup
System fails to recognize the waste collection bag during initial setup
Ranked causes
- 1. Waste bag tubing connector oriented incorrectly (common)
- 2. Waste bag missing or not hung correctly (occasional)
Check waste bag tubing orientation
- Test point
- Waste bag tubing connection
- Expected
- Oriented vertically downward
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Load Cell
Annual verification of supply weight and collection weight load cells using 500 gram test weight to ensure accurate fluid tracking
Ranked causes
- 1. Load cell calibration required annually (common)
Verify supply weight after zeroing load cells
- Test point
- Supply Weight display
- Expected
- 0 ± 25 g, ±25 g
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Load cell verification
Annual inspection procedure to verify supply and waste bag load cell accuracy using 500 g weight
Ranked causes
- 1. Load cells require calibration verification (unknown)
Zero load cells
- Test point
- System display
- Expected
- Supply Weight and Collection Weight both 0 ± 25 g
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pressure caution
Annual verification of pressure sensor calibration using 3000 mL fluid bag to confirm accurate pressure measurement in intra-uterine pressure system
Ranked causes
- 1. Pressure sensor calibration required annually (common)
Verify tubing is completely filled with no air bubbles
- Test point
- Hysteroscope inflow channel
- Expected
- Fluid flowing with no visible air bubbles
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pressure verification
Annual inspection procedure to verify pressure chamber, pressure sensor and measurement accuracy
Ranked causes
- 1. Pressure sensor requires calibration verification (unknown)
Verify pressure reading
- Test point
- System display (pressure gauge)
- Expected
- 20 ± 5 inH2O, ± 5
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check fluid bag hooks - error 02010026
Fluid bag hooks potentially damaged or excessively pressurized; detection issue noted
Ranked causes
- 1. Hooks accidentally touched or supply bag subjected to excessive force (occasional)
Inspect fluid bag hooks for interference
- Test point
- Supply bag hooks
- Expected
- Hooks clear, only bag present
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check supply bag hooks caution
Supply bag hooks may be damaged or supply bag forced excessively
Ranked causes
- 1. Supply bag hooks manipulated or bag forced too hard (occasional)
Inspect supply hooks for damage
- Test point
- Supply bag hooks
- Expected
- Intact, not bent or damaged
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check waste bag hooks caution
Waste bag hooks may be damaged or waste bag forced excessively
Ranked causes
- 1. Waste bag hooks manipulated or bag forced too hard (occasional)
Inspect waste hooks for damage
- Test point
- Waste bag hooks
- Expected
- Intact, not bent or damaged
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check waste bag hooks - error 02010027
Waste bag hooks potentially damaged or excessively pressurized; detection issue noted
Ranked causes
- 1. Hooks accidentally touched or waste bag subjected to excessive force (occasional)
Inspect waste bag hooks for interference
- Test point
- Waste bag hooks
- Expected
- Hooks clear, only bag present
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Collection bag overweight caution
More than 15 kg weight hanging on collection bag hook; system displays error
Ranked causes
- 1. Excessive weight on collection bag hook (common)
Verify weight on collection bag hook
- Test point
- Collection bag hook
- Expected
- Less than 15 kg total
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Deficit error - sudden jump detected patient safety
System detected instantaneous deficit jump of 500 ml or more
Ranked causes
- 1. Sudden 500 ml or greater deficit jump (rare)
Record final automatic deficit value
- Test point
- Deficit display before clearing
- Expected
- Note value marked as XXX
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Deficit error with sudden fluid loss patient safety
System detects unexpected jump in fluid deficit value of 500 ml or more
Ranked causes
- 1. Sudden fluid measurement error or leak (occasional)
Record deficit value at error
- Test point
- System display
- Expected
- Specific numerical value shown (e.g., XXX ml)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Deficit limit approaching patient safety
Fluid deficit is within 75% of established limit
Ranked causes
- 1. Deficit accumulation near limit (common)
Review current deficit against limit
- Test point
- Deficit monitoring display
- Expected
- Below 75% of established limit
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Deficit limit reached or exceeded patient safety
Cumulative fluid deficit has reached or exceeded established limit
Ranked causes
- 1. Deficit accumulation exceeds limit (common)
Verify deficit against limit
- Test point
- Deficit monitoring display
- Expected
- Below established limit
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow limited by scope size
Flow rate is limited by scope diameter being used
Ranked causes
- 1. Scope diameter too small for requested flow (common)
Verify inlet flow valve open during priming
- Test point
- Inlet flow valve position
- Expected
- Fully open
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fluent system unexpectedly shuts down
System powers off or terminates procedure without operator action
Ranked causes
- 1. Temporary system fault or power interruption (occasional)
Wait and restart
- Test point
- System power
- Expected
- System boots normally and displays menu
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fluid deficit near threshold - warning 01010037
Cumulative fluid deficit has reached 75% of configured limit; warning issued to operator
Ranked causes
- 1. Fluid deficit accumulation near limit (common)
Review fluid deficit accumulation
- Test point
- System fluid deficit display
- Expected
- Below 75% of threshold
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fluid deficit threshold reached or exceeded - error 02010028 patient safety
Cumulative fluid deficit has met or exceeded configured threshold; procedure must stop
Ranked causes
- 1. Fluid deficit limit reached during procedure (common)
Review fluid deficit and clinical status
- Test point
- System fluid deficit display
- Expected
- Deficit below threshold or procedure completed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fluid loss limit reached or exceeded patient safety
System indicates fluid loss has reached or exceeded the preset loss limit. Error code 02010028.
Ranked causes
- 1. Maximum fluid loss during procedure (common)
Monitor cumulative fluid loss
- Test point
- Loss display screen
- Expected
- Loss within preset limit
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fluid loss measurement error detected
System detects sudden fluid loss jump of 500 mL or more. Displays error indicating loss value (XXX).
Ranked causes
- 1. Sudden fluid loss spike (occasional)
Monitor fluid loss values during procedure
- Test point
- Loss display on screen
- Expected
- Gradual increase without sudden jumps
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
In-FloPak dislodged during procedure
System detects In-FloPak no longer completely inserted
Ranked causes
- 1. In-FloPak partially withdrawn or dislodged (common)
Open inlet regulating valve
- Test point
- Inlet valve on hysteroscope
- Expected
- Fully open
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
In-FloPak not detected - error 0208000A
In-FloPak has become partially dislodged from system or connection lost during procedure
Ranked causes
- 1. In-FloPak not fully seated in system (common)
Open inflow stopcock
- Test point
- Hysteroscope inflow stopcock
- Expected
- Stopcock fully open
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
In-FloPak not detected during assembly
System does not recognize In-FloPak insertion; assembly procedure cannot proceed
Ranked causes
- 1. In-FloPak not fully inserted or misaligned (common)
Verify In-FloPak is fully inserted and aligned
- Test point
- In-FloPak connection to system
- Expected
- In-FloPak flush with system front face
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
In-FloPak not detected or not fully inserted
System detects In-FloPak is not completely inserted. Error code 0208000A.
Ranked causes
- 1. In-FloPak not fully seated in connector (common)
Verify In-FloPak insertion
- Test point
- In-FloPak connector
- Expected
- Flush with system front, fully seated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
MyoSure device not found
Device not detected or not fully connected
Ranked causes
- 1. Device not fully connected or defective (common)
Verify MyoSure device connection
- Test point
- Device connector
- Expected
- Fully and securely connected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
MyoSure excessive torque detected patient safety
Motor error - excessive force or resistance on MyoSure device
Ranked causes
- 1. Excessive bending force on device (common)
Verify adequate clearance and positioning
- Test point
- MyoSure device in uterus
- Expected
- Proper positioning with minimal resistance
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Out-FloPak dislodged during procedure
System detects Out-FloPak no longer completely inserted
Ranked causes
- 1. Out-FloPak partially withdrawn or dislodged (common)
Open inlet regulating valve
- Test point
- Inlet valve on hysteroscope
- Expected
- Fully open
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Out-FloPak not detected - error 0209000A
Out-FloPak has become partially dislodged from system or connection lost during procedure
Ranked causes
- 1. Out-FloPak not fully seated in system (common)
Open inflow stopcock
- Test point
- Hysteroscope inflow stopcock
- Expected
- Stopcock fully open
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Out-FloPak not detected or not fully inserted
System detects Out-FloPak is not completely inserted. Error code 0209000A.
Ranked causes
- 1. Out-FloPak not fully seated in connector (common)
Verify Out-FloPak insertion
- Test point
- Out-FloPak connector
- Expected
- Flush with system front, fully seated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Out-FloPak not fully inserted
System reports Out-FloPak no longer completely inserted; error displayed on interface
Ranked causes
- 1. Out-FloPak not seated flush with system front (common)
Verify Out-FloPak is fully seated and flush with system front
- Test point
- Out-FloPak front face
- Expected
- Flush with system front panel
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure limited by scope size
Adjusted pressure exceeds system capability for scope diameter being used
Ranked causes
- 1. Scope diameter too small for requested pressure (common)
Verify inlet valve open during priming
- Test point
- Inlet flow valve position
- Expected
- Fully open
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Priming error - scope outlet blocked
System detected pressure buildup during priming procedure
Ranked causes
- 1. Scope outlet blocked during priming (common)
Position scope open to air at patient level
- Test point
- Scope position during priming
- Expected
- Open to air, not in uterus
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Priming error, hysteroscope tip blocked - error 020A0007
- Priming suggested
- Repriming suggested - warning 010A0008
- Runtime or system error - error 04XXXXXX
- Supply bag exceeds weight limit
- Supply bag hook compromised
- Supply bag low - warning 01010033
- Supply bag missing
- Supply bag missing - error 02010023
- Supply bag overweight
- Supply bag running low
- Supply bag too heavy - error 04010020
- Supply empty, kinked or not punctured
- Supply empty, kinked, or not connected - error 020A0006
- System unexpected power loss
- System unexpected shutdown
- System unexpectedly powers off
- Unexpected deficit spike detected - error 02010029
- Waste bag almost full - warning 01010034
- Waste bag exceeds weight limit
- Waste bag full
- Waste bag full - error 02010025
- Waste bag full or at capacity
- Waste bag hook compromised
- Waste bag lost or misaligned
- Waste bag missing - error 02010024
- Waste bag missing or improperly hung
- Waste bag nearly full
- Waste bag not detected
- Waste bag overweight
- Waste bag too heavy - error 04010021
31 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
01010033 | Supply bag low | Supply bag contains less than 500 ml of fluid | warning |
01010034 | Waste bag almost full | Waste collection bag is within 500 ml of full capacity | warning |
01010037 | Approaching deficit limit | Fluid deficit is within 75% of the set limit | warning |
01010037 | Fluid deficit near threshold | Cumulative fluid deficit reaches 75% of configured limit | warning |
01070001 | MyoSure device not found | System cannot detect MyoSure tissue removal device | warning |
01070002 | Excessive torque detected, reduce pressure on device | MyoSure motor error from excessive rotational force on device shaft | warning |
01070002 | MyoSure excessive torque detected | Motor error from excessive force on device | error |
010A0004 | Warning - High Pressure | High pressure condition lasting more than 2 seconds but less than 5 seconds | warning |
010A0008 | Reprime suggested | Unexpected discrepancy between measured and expected uterine pressure | warning |
010A000B | Pressure limited by scope size being used | Set pressure exceeds maximum allowed for the hysteroscope diameter selected | warning |
010A000C | Flow limited by scope size being used | Flow rate is limited by hysteroscope diameter used | warning |
010A000C | Scope size restricts flow rate | Selected scope size limits flow speed during operation | warning |
02010023 | Supply bag missing | Supply fluid bag not present or empty | error |
02010024 | No waste bag | Waste bag not properly hung or missing | error |
02010024 | Waste bag missing or improperly positioned | Waste bag not correctly suspended on both hooks or absent | error |
02010025 | Waste bag full | Waste bag is completely full and cannot accept more fluid | error |
02010026 | Check supply bag hooks | Hooks handled or supply bag swung excessively | warning |
02010026 | Check fluid bag hooks | Supply bag hooks detect unexpected pressure or movement | error |
02010027 | Check waste bag hooks | Waste bag hooks detect unexpected pressure or movement | error |
02010028 | Deficit limit reached or exceeded | Cumulative fluid deficit equals or exceeds set limit | error |
02010029 | Deficit error with sudden jump in value | System detects sudden 500 ml or greater jump in measured deficit | error |
02010029 | Deficit error. Record last value, zero deficit, proceed, use manual count. | System detected instantaneous deficit jump of 500 mL or more | error |
02010029 | Waste value error; at least 500 mL change detected | System detects ≥500 mL change in waste value | error |
02010029 | Fluid loss measurement error - sudden loss jump detected | System detects 500mL or more instantaneous loss jump | warning |
02010029 | Unexpected fluid deficit spike | System detects 500 mL or greater sudden change in fluid balance | error |
0208000A | In-FloPak not detected - open stopcock, push in, clear error | System detects In-FloPak (blue inlet cartridge) no longer fully inserted | error |
0208000A | In-FloPak not detected. Open inflow stopcock, push blue In-FloPak back in, clear error. | System detected In-FloPak no longer fully inserted | error |
0209000A | Out-FloPak not detected - open stopcock, push in, clear error | System detects Out-FloPak (yellow outlet cartridge) no longer fully inserted | error |
0209000A | Out-FloPak not detected. Open inflow stopcock, push yellow Out-FloPak back in, clear error. | System detected Out-FloPak no longer fully inserted | error |
020A0005 | Inlet tubing pressure high | High pressure detected in inlet tubing | warning |
020A0005 | Tubing pressure high - open inflow or unclamp | High pressure detected in inflow tubing | error |
020A0006 | Supply empty, clamped or not spiked | Pump-measured fluid quantity exceeds load cell-measured fluid quantity | error |
020A0006 | Inadequate fluid supply from pump | More pump fluid used than transducer fluid available | error |
020A0007 | Startup pressure error - hysteroscope obstruction | Elevated pressure detected during system startup | error |
020A0007 | Priming error, ensure scope end not blocked | System detected pressure buildup during priming | error |
04010020 | Supply bag overweight | More than 15 kg of weight on supply bag hooks | critical |
04010021 | Waste bag overweight | More than 15 kg of weight on waste bag hooks | critical |
04XXXXXX | Various runtime or system errors | Various system fault conditions during operation | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.