Hill-Rom VersaCare: troubleshooting and error codes

The ServiceForge knowledge base holds 224 documented symptoms for the Hill-Rom VersaCare, with 277 ranked causes, 415 diagnostic checks with test points and expected values, and 158 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Alerts inactive when bed plugged in patient safety

SafeView alert lights do not illuminate after bed power-on

Ranked causes

  1. 1. Alerts deactivated in system settings (common)
  2. 2. Bed Exit System not activated (occasional)
  3. 3. Alert cables loose or disconnected (occasional)
  4. 4. Alert cable or assembly damaged (occasional)
  5. 5. UCB board failure (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Enable switch not functioning caution

Handle enable switch does not engage transport system or shows incorrect voltage readings

Ranked causes

  1. 1. Enable switch not plugged into Junction board (occasional)
  2. 2. Enable switch not generating correct voltage signal (occasional)
  3. 3. Defective enable switch or cable (occasional)
  4. 4. Low or absent battery voltage (occasional)
  5. 5. Problem on Junction board or P6 connector (occasional)
First check

Verify enable switch connection to Junction board

Expected
Both switches plugged into P2 and P5

5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SafeView alerts do not come on when plugged in patient safety

SafeView alert lights remain dark after bed is powered on

Ranked causes

  1. 1. Alerts deactivated (common)
  2. 2. Bed exit system not activated (occasional)
  3. 3. Both alert cables loose or disconnected (occasional)
  4. 4. Both alert cables damaged (rare)
  5. 5. Siderail interface PC board failure (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air Control Board or Blower Power Board failure high voltage

Air module not functioning; blower inoperative or incorrect voltage output; weight sensing inaccurate

Ranked causes

  1. 1. Defective Air Control Board (ACB) (unknown)
  2. 2. Defective Blower Power Board (unknown)
  3. 3. Blower power board voltage out of specification (occasional)
  4. 4. Patient weight offset not calibrated (occasional)
First check

Verify blower is operating

Test point
Green LED on blower power board (D)
Expected
LED is on when power present

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Enable switch not responding

Bed does not move when handle switch is depressed or pulled. Operator reports switch inoperative.

Ranked causes

  1. 1. Enable switch cable disconnected or not seated at P2 or P5 (common)
  2. 2. Enable switch or cable defective (always open or always closed) (occasional)
  3. 3. No voltage at P6.1 due to low or missing battery voltage (occasional)
  4. 4. Junction board or P6 connector failure (rare)
First check

Verify enable switches plugged into Junction board

Test point
P2 and P5
Expected
Both switches fully seated

7 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SafeView single alert flashing or off color patient safety

Only one SafeView alert light on or flashing yellow and green

Ranked causes

  1. 1. Alert cable loose or disconnected (common)
  2. 2. Alert cable damaged (occasional)
  3. 3. Alert light assembly PC board failure (occasional)
  4. 4. Siderail interface PC board failure (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Throttle output signal out of range

Handle force signal does not respond correctly to push/pull actions. Bed may not accelerate or decelerate as expected.

Ranked causes

  1. 1. Throttle zero offset requires adjustment (common)
  2. 2. Low supply voltage at P6.8 (occasional)
  3. 3. Low excitation voltage at P3.5 or P7.5 (red wire) (occasional)
  4. 4. Strain element in handle damaged (rare)
First check

Measure supply voltage at P6.8

Test point
P6.8
Expected
more than 22 V
Tools
DMM

8 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Throttle signal incorrect or out of range caution

Handle throttle signal does not produce expected voltage levels; bed speed control not working properly

Ranked causes

  1. 1. Throttle signal out of neutral range (occasional)
  2. 2. Low excitation voltage to throttle circuit (occasional)
  3. 3. Damaged handle strain element (occasional)
  4. 4. Cable continuity or Junction board fault (occasional)
First check

Measure throttle output at P6.10 with no handle force

Test point
P6.10
Expected
2.4-2.6 V, +/- 0.1 V
Tools
DC Volt Meter

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bed frame does not operate patient safety

Bed sections (head, knee, foot, or hilow) do not move when controls are pressed, or move unexpectedly

Ranked causes

  1. 1. Control lockout is engaged (common)
  2. 2. Motor or drive mechanism failure (occasional)
  3. 3. Electronic control system fault (occasional)
First check

Verify control lockout status

Test point
Lockout indicator LEDs
Expected
Indicator should be off (unlocked)

8 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Motor controller not receiving 4700 ohm resistor signal

Motor does not operate. PACM board may not detect Junction board connection or resistor continuity is lost.

Ranked causes

  1. 1. 4700 ohm resistor on Junction board open or Junction board failure (occasional)
  2. 2. 14-pin cable between Junction and PACM board defective or disconnected (occasional)
  3. 3. Connector P6 on Junction board pins damaged or corroded (rare)
First check

Unplug 14-pin cable from PACM board

Test point
PACM board 14-pin connector
Expected
Cable safely unplugged

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Motor controller not receiving resistance signal caution

Motor controller will not operate; connection to junction board 4700 ohm resistor is missing or open

Ranked causes

  1. 1. Open or defective PACM-to-Junction cable (occasional)
  2. 2. Faulty P6 connector on Junction board (occasional)
  3. 3. Defective Junction board resistor (occasional)
First check

Measure resistance across PACM cable pins 12-13

Test point
14-pin cable pins 12, 13
Expected
4700 ohms +/- 5%, +/- 5%
Tools
Ohm Meter

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Only one alert on or flashing yellow and green patient safety

Single SafeView alert light malfunctioning while other operates normally

Ranked causes

  1. 1. Alert cable loose, disconnected, or damaged (common)
  2. 2. Alert light assembly PCB damaged (occasional)
  3. 3. UCB board malfunction (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

SafeView both alerts flash yellow and green patient safety

Both SafeView alert lights flashing yellow and green simultaneously

Ranked causes

  1. 1. SideCom PC board malfunction (occasional)
  2. 2. Siderail interface PC board malfunction (occasional)
  3. 3. Control pod PC board malfunction (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Scale system malfunction after P.C. board work

System does not function properly after scale P.C. board removal and replacement.

Ranked causes

  1. 1. Improper connector seating or connector alignment (common)
  2. 2. Scale calibration not performed or incomplete (common)
  3. 3. Component damage from electrostatic discharge during handling (occasional)
First check

Verify all connectors fully seated

Test point
Connectors P1, P2, P3, P4 on P.C. board H
Expected
All connectors seated and latched

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery voltage out of range high voltage

Battery or charging system not providing correct voltage; controller unable to operate

Ranked causes

  1. 1. Charging system failure (common)
  2. 2. Battery discharged (common)
First check

Measure battery voltage at P6.8

Test point
P6.8
Expected
> 32 V when plugged into AC, > 22 V on battery, > 32V AC, > 22V battery

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Boost Position System does not operate patient safety

Bed does not move to middle height or assumes incorrect position when Boost control is activated

Ranked causes

  1. 1. Control lockout engaged (common)
  2. 2. Boost motor or control circuit failure (occasional)
First check

Verify lockouts are disengaged

Test point
Control lockout indicators
Expected
All lockout indicators are off

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Caster or wheel locks do not function patient safety

Wheels do not lock when brake is set, or casters do not lock in expected position

Ranked causes

  1. 1. Brake mechanism failure or wear (common)
  2. 2. Steer caster locking mechanism failure (non-IntelliDrive beds) (occasional)
First check

Test wheel brake engagement

Test point
Brake pedal, attempt to move bed
Expected
The four wheels are locked and prevent any movement

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Controller LED not responding high voltage

LED DS 1 is off; controller either not powered or not receiving KSI signal from wheel deployment

Ranked causes

  1. 1. Controller not powered (common)
  2. 2. KSI signal not received (occasional)
First check

Observe LED DS 1 state on PACM board after pressing Enable switch

Test point
DS 1 LED
Expected
LED should be either on steady or off

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

CPR cable removal and replacement caution

Complete removal and reinstallation procedure for CPR cable assembly including routing and cable tie attachment

Ranked causes

  1. 1. CPR cable requires removal or replacement (occasional)
  2. 2. CPR cable requires installation after removal (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

CPR handle does not function patient safety

CPR handle does not lower head section quickly or movement is not cushioned, or head section does not continue into Trendelenburg

Ranked causes

  1. 1. CPR actuation mechanism failure or wear (common)
  2. 2. Trendelenburg function not working after CPR activation (occasional)
First check

Test CPR handle head section descent

Test point
CPR handle, head section movement
Expected
The head section descends quickly up to the mid-travel then the movement is cushioned until the low position is reached without problem or abnormal noise

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Head FSR error or blower current error

Service Required LED flashing with pattern 1 long, 5 short flashes. A Model Bed: Head FSR1 and FSR2 error. H Model or Newer Bed with AMT: Blower current error.

Ranked causes

  1. 1. Head FSR1 or FSR2 sensor failure (A Model) (unknown)
  2. 2. Blower current error (H Model or Newer with AMT) (unknown)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Incorrect throttle voltage

Handle throttle input voltage outside expected range; bed does not respond to forward/back force correctly

Ranked causes

  1. 1. Incorrect throttle voltage at controller input (common)
  2. 2. Damaged Junction board or cable (occasional)
First check

Measure throttle input voltage at controller

Test point
controller throttle input
Expected
2.5 V neutral, 4.0 V pushed forward, 1.0 V pulled back, +/-0.5 V

4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Motor controller resistor fault

4700 ohm termination resistor on Junction board not present or damaged; controller will not operate

Ranked causes

  1. 1. Resistor disconnected or damaged (rare)
  2. 2. Cable continuity broken (occasional)
First check

Measure resistance between PACM board pins 12 and 13

Test point
PACM connector pins 12, 13
Expected
4700 ohms +/- 5%, +/- 5%

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

SafeView® Alert lights flashing yellow during configured siderail movements

Lights flash yellow when configured siderails are lowered or bed is raised, indicating proper alert system operation

Ranked causes

  1. 1. Normal operation during bed and siderail movement (common)
  2. 2. Head angle alarm active and bed lowered below alarm threshold (common)
First check

Verify siderail configuration is set correctly

Test point
Control pod buttons
Expected
Lights flash green for three seconds after five-second button press

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SafeView® Alert lights off when expected

Lights remain off when Bed Exit Alarm System deactivated or when specific deactivation procedure performed

Ranked causes

  1. 1. Bed Exit Alarm System not active or lights have been deactivated (common)
  2. 2. Bed on battery power or disconnected from AC power (common)
First check

Verify Bed Exit Alarm System is not active before deactivation procedure

Test point
System status indicator or control pod
Expected
Bed Exit Alarm System inactive

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Scale and bed exit alarm system does not function patient safety

Zero control, weight detection, Enable control, bed exit modes, or alarm sound fails to operate correctly

Ranked causes

  1. 1. Scale system malfunction (unknown)
  2. 2. Bed exit alarm system malfunction (unknown)
First check

Zero control display with 50 lb weight on bed center

Test point
Display
Expected
Shows 00.0

6 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Transport system not responding to handle input caution

IntelliDrive transport system does not move when handles are engaged; unable to move bed forward, backward, or steer

Ranked causes

  1. 1. Controller not powered or not receiving KSI signal (common)
  2. 2. Controller error condition - LED flashing (occasional)
First check

Monitor DS 1 LED indicator on PACM board

Expected
LED is on and steady when functioning correctly

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Trendelenburg function does not operate patient safety

Sleep surface does not tilt to Trendelenburg or reverse Trendelenburg position, or tilts abnormally

Ranked causes

  1. 1. Trendelenburg motor or actuator failure (common)
  2. 2. Mechanical linkage misalignment or damage (occasional)
First check

Test Trendelenburg function at intermediate bed height

Test point
Trendelenburg control, bed at intermediate position
Expected
The sleep surface gradually tilts to maximum trendelenburg without any problem or abnormal noise

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

30° Head Angle Alarm does not disable

Pressing Enable control followed by 30° Head Angle Alarm control does not turn off the 30° Head Angle Alarm indicator

Ranked causes

  1. 1. 30° Head Angle Alarm control or disable circuit failure (unknown)
First check

Press Enable control while alarm is active and indicator is on

Test point
Enable control
Expected
Enable mode activated

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

30° Head Angle Alarm does not sound when head angle below 30°

Lowering head section below 30° does not trigger alarm sounding and indicator flashing for 30° Head Angle Alarm option

Ranked causes

  1. 1. Head angle sensor or alarm circuit malfunction (unknown)
First check

Lower head section below 30°

Test point
Head Down control and head position
Expected
Head section angle is below 30°

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

30° Head Angle Alarm does not stop when head angle above 30°

Raising head section above 30° does not stop alarm and does not leave indicator on solid for 30° Head Angle Alarm option

Ranked causes

  1. 1. Head angle sensor or alarm circuit malfunction (unknown)
First check

Raise head section above 30°

Test point
Head Up control and head position
Expected
Head section angle is above 30°

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

30° Head Angle Alarm indicator does not activate patient safety

Enable and 30° Head Angle Alarm controls pressed but indicator does not illuminate.

Ranked causes

  1. 1. 30° Head Angle Alarm option not functional or wiring fault (unknown)
First check

Press Enable control then 30° Head Angle Alarm control

Test point
Alarm option indicator
Expected
Indicator comes on

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

30° Head Angle Alarm indicator does not illuminate

Pressing Enable control followed by 30° Head Angle Alarm control does not illuminate the indicator light for the 30° Head Angle Alarm option

Ranked causes

  1. 1. 30° Head Angle Alarm circuit or indicator lamp failure (unknown)
First check

Press Enable control

Test point
Enable control
Expected
Enable mode is activated

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

AC Power Fuses replacement high voltage

AC power fuses located in power supply module require replacement after disconnecting power

Ranked causes

  1. 1. AC power fuse blown or degraded (unknown)
First check

Set brakes

Test point
Brake controls
Expected
Brakes engaged

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

AC power present but no motor power high voltage

AC power available but DC motor power supply has failed

Ranked causes

  1. 1. DC power supply failure or connection issue (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Accumulator pressure error caution

Service Required LED flashing with pattern 2 long, 2 short flashes

Ranked causes

  1. 1. Accumulator pressure out of specification (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm does not sound when brake is not set patient safety

No steady alarm when bed is unplugged from power and brake is not set

Ranked causes

  1. 1. Alarm circuit or speaker failure (occasional)
First check

Test alarm with brake not set

Test point
Power source, brake pedal position
Expected
A steady alarm comes on when brake is not set and bed is plugged in

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm does not stop when brake is set

G model or newer bed continues sounding steady alarm after brake is set

Ranked causes

  1. 1. Brake safety circuit or sensor malfunction (unknown)
First check

Set brake

Test point
Brake control
Expected
Alarm stops immediately when brake is fully set

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm Silence feature does not function correctly patient safety

Alarm Silence indicator fails to activate or silence alarm, or system does not arm properly with weight on bed

Ranked causes

  1. 1. Alarm Silence feature system malfunction (unknown)
First check

Enable and Alarm Silence control activation

Test point
Alarm Silence indicator
Expected
Indicator comes on

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm Silence feature does not operate patient safety

Pressing Alarm Silence control during bed exit alarm does not stop sound or indicator.

Ranked causes

  1. 1. Alarm Silence control or circuit fault (unknown)
First check

Cause bed exit alarm condition with weight on bed

Test point
Alarm and indicators
Expected
Alarm sounds

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery backup does not operate patient safety

Bed functions do not work when unplugged from AC power and running on battery backup

Ranked causes

  1. 1. Battery failure or discharge (common)
First check

Test head and knee functions on battery power

Test point
Bed unplugged, head/knee controls operated
Expected
The bed works correctly on battery power for head and knee functions

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery backup does not operate bed functions patient safety

After unplugging bed and activating battery backup, functions do not work correctly

Ranked causes

  1. 1. Battery backup circuit failure (unknown)
First check

Unplug bed and activate battery backup; operate functions 10-12

Test point
Battery backup system
Expected
Head, knee, foot functions work correctly on battery power

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery charging error with AC power on caution

Battery charging indicator off despite AC mains present

Ranked causes

  1. 1. Battery charger malfunction or battery failure (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery replacement high voltage

Batteries must be replaced in pairs; procedure involves disconnecting power supply, removing cover, disconnecting cables, removing bracket and batteries, checking vent holes for blockage.

Ranked causes

  1. 1. Batteries depleted or faulty (unknown)
First check

Set brakes on bed

Test point
Brake control
Expected
Brakes engaged

11 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bed does not enter Trendelenburg when CPR handle held patient safety

Continuing to hold CPR handle after head descent does not transition bed into Trendelenburg position

Ranked causes

  1. 1. CPR control or trendelenburg sequencing circuit failure (unknown)
First check

Continue holding CPR handle after head descent completes

Test point
CPR handle held position
Expected
Bed transitions into Trendelenburg position

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bed does not lower to low position patient safety

Pressing Bed Down control does not lower bed to low position or bed stops during movement; Bed Not Down indicator does not turn off at maximum low position

Ranked causes

  1. 1. Motor, control, or mechanical failure (unknown)
First check

Press Bed Down control

Test point
Bed Down control
Expected
Bed lowers to low position without stopping

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bed does not raise to high position patient safety

Pressing Bed Up control does not raise bed to high position or bed stops during movement

Ranked causes

  1. 1. Motor, control, or mechanical failure (unknown)
First check

Press Bed Up control

Test point
Bed Up control
Expected
Bed rises to high position without stopping

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bed exit alarm does not activate or indicator malfunction patient safety

Enable control active but bed exit mode control does not arm system or alarm fails when patient leaves bed.

Ranked causes

  1. 1. Bed exit alarm circuit, sensor, or control fault (unknown)
First check

Place 50 lb weight on center of bed and press Enable control

Test point
Bed exit alarm Enable indicator
Expected
Indicator comes on

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bed flat function does not work correctly patient safety

Pressing Bed Flat control does not return head, knee, foot sections to horizontal; on G model or newer, sleep deck and bed do not return to flat and level position

Ranked causes

  1. 1. Motor, control, or mechanical failure (unknown)
First check

Press Bed Flat control

Test point
Bed Flat control
Expected
All sections gradually return to horizontal while control is pressed; G model beds return to flat and level

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bed not down indicator flashing patient safety

System indicates bed is not fully lowered when it should be

Ranked causes

  1. 1. Position sensor fault or hilow not fully down (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bed reverse trendelenburg function does not operate patient safety

Reverse Trendelenburg control pressed but sleep surface does not tilt to maximum reverse trendelenburg or produces abnormal noise

Ranked causes

  1. 1. Motor or drive mechanism failure (unknown)
First check

Operate the Reverse Trendelenburg control

Test point
Bed at intermediate position
Expected
Sleep surface gradually tilts to maximum reverse trendelenburg without problem or abnormal noise

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bed trendelenburg function does not operate patient safety

Trendelenburg control pressed but sleep surface does not tilt gradually to maximum trendelenburg angle or produces abnormal noise

Ranked causes

  1. 1. Motor or drive mechanism failure (unknown)
First check

Operate the Trendelenburg control at intermediate bed height

Test point
Bed raised to intermediate position
Expected
Sleep surface gradually tilts to maximum trendelenburg without problem or abnormal noise

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bed wheels do not lock when brake is set patient safety

Setting brake does not lock all four wheels; bed moves despite brake being engaged

Ranked causes

  1. 1. Brake mechanism failure or damaged caster (unknown)
First check

Set all bed brakes

Test point
Brake control
Expected
All four wheels lock and prevent bed movement

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bed will not drive

IntelliDrive powered mobility system fails to propel bed

Ranked causes

  1. 1. PACM board power loss or drive motor failure (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bed will not drive with wheel down

Mobility system fails to move with transport wheel in contact with floor

Ranked causes

  1. 1. Wheel deployment issue or motor brake engaged (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Boost position function does not work correctly patient safety

For beds with Boost Position System option, pressing and holding Boost control does not move bed to middle height with head and knee sections lowering and bed at ~7° Trendelenburg

Ranked causes

  1. 1. Boost circuit or control system failure; lockout engaged (unknown)
First check

Verify lockouts are disengaged

Test point
Head, knee, bed up/down lockout controls
Expected
All lockout indicators are off

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: AC power fuse replacement high voltage

Procedure to remove and replace AC power fuses in the power supply module

Ranked causes

  1. 1. AC power fuse requires replacement (unknown)
First check

Set brakes

Test point
Bed brake control
Expected
Brakes engaged

9 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Disable AMT mode

Configure Air Control Board to non-AMT operation after replacement or per specification

Ranked causes

  1. 1. AMT factory default enabled; must be disabled for this bed configuration (common)
First check

Enter AMT disable mode

Test point
Control pod: Turn R + Turn L + Max-Inflate + Normal
Expected
Single beep after ~5 seconds

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Drive Belt—IntelliDrive® Transport System caution

Procedure to remove and replace the drive belt on IntelliDrive® transport system. Includes disassembly of drive box, motor mounting, pulley plates, and belt removal/installation steps.

Ranked causes

  1. 1. Drive belt replacement procedure (unknown)
First check

Set the brakes

Test point
Bed brake mechanism
Expected
Brakes engaged and locked

11 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Drive Motor IntelliDrive Transport System caution

Removal and replacement procedure for drive motor assembly in VersaCare hospital bed IntelliDrive transport system.

Ranked causes

  1. 1. Drive motor replacement required (unknown)

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Calibration: Head Angle Sensor (G-J model beds)
  • Calibration: Head angle sensor for G model or newer
  • Calibration: Head Siderail P.C. Board
  • Calibration: Patient weight offset
  • Calibration: Scale
  • Calibration: Seat Section Bladder Assembly replacement
  • Calibration: Sensor Strip replacement
  • Calibration: Transformer and Battery Fuse
  • Calibration: Transport Handle IntelliDrive voltage adjustment
  • Calibration: Turn Assist Bladder Assembly replacement
  • Cannot view software version on optional pod
  • Caster removal foot end
  • Caster removal head end
  • Caster replacement
  • Chair position function does not work correctly
  • Control lockout indicator does not illuminate
  • Control Pod P.C. Board removal and replacement procedure
  • Control unlock indicator does not turn off
  • CPR cable adjustment
  • CPR function does not continue into Trendelenburg position
  • CPR handle function does not work correctly
  • CPR handle quick release not functioning properly
  • CRC error
  • Damaged handle cable or connector
  • Damaged handle strain element
  • Diagnostic or battery charging errors present
  • Drive Motor Removal Procedure
  • Error detected in dynamic air module
  • Foot section does not extend to full out position
  • Foot section does not retract to full in position
  • Head angle alarm does not sound when head exceeds 30°
  • Head Angle Display does not show 0° in level position
  • Head Angle Display does not show 62° at maximum head height
  • Head Angle Display shows incorrect angle
  • Head angle sensor error
  • Head section bladder assembly replacement
  • Head section does not deflate when squeezed
  • Head section does not inflate properly
  • Head section does not lower to low position
  • Head section does not re-inflate when pressure released
  • Head section does not rise smoothly after CPR
  • Head section does not rise to high position
  • Head section motor does not operate after CPR handle release
  • Head zone pressure lower than seat zone in articulated position
  • Head zone pressure not lower than seat zone
  • Head, knee, leg, or foot motor unexpected motion
  • Heartbeat error
  • Hilow motor position rate high
  • Hilow motor position rate low
  • Hilow motor unexpected motion
  • IntelliDrive battery check
  • IntelliDrive bed will not drive
  • IntelliDrive bed will not drive with wheel down
  • IntelliDrive controller check
  • IntelliDrive handle enable switch check
  • IntelliDrive handle gauge check
  • IntelliDrive junction board debugging
  • IntelliDrive motor check
  • IntelliDrive PACM board deployment check
  • IntelliDrive PACM board drive check
  • IntelliDrive PACM board power check
  • IntelliDrive PACM to junction board cable check
  • IntelliDrive steer switch check
  • IntelliDrive throttle check
  • IntelliDrive Transport System does not move
  • IntelliDrive visual inspection
  • IntelliDrive wheel will not stow
  • Knee section does not lower to low position
  • Knee section does not rise to high position
  • Left Turn Assist bladder does not inflate or alarm does not sound
  • Load Beam—J Models replacement
  • Logic P.C. Board All Lockout LED error
  • Logic P.C. Board Bootloader error
  • Logic P.C. Board CPR Switch failure
  • Logic P.C. Board Drive Current Too Low error
  • Logic P.C. Board EEPROM failure
  • Logic P.C. Board Foot Current Overload error
  • Logic P.C. Board Foot, Knee, Leg Unexpected Motion error
  • Logic P.C. Board Head Current Overload error
  • Logic P.C. Board Head Unexpected Motion error
  • Logic P.C. Board Head, Knee, Hilow Lockout LED error
  • Logic P.C. Board Heartbeat Failure error
  • Logic P.C. Board Hilow Foot Current Overload error
  • Logic P.C. Board Hilow Foot Position Rate High error
  • Logic P.C. Board Hilow Foot Position Rate Low error
  • Logic P.C. Board Hilow Foot Unexpected Motion error
  • Logic P.C. Board Hilow Head Current Overload error
  • Logic P.C. Board Hilow Head Position Rate High error
  • Logic P.C. Board Hilow Head Position Rate Low error
  • Logic P.C. Board Hilow Head Unexpected Motion error
  • Logic P.C. Board Knee Current Overload error
  • Logic P.C. Board Leg Current Overload error
  • Logic P.C. Board Logic Circuit Failure error
  • Logic P.C. Board Patient Pendant Failure error
  • Logic P.C. Board Supervisory Relay error
  • Mattress disconnected
  • Mattress three zones do not inflate to high pressure
  • Missing 24 VDC input
  • Motor controller removal and replacement procedure
  • Motor controller replacement needed
  • Motors do not operate
  • Night light cannot be disabled
  • No AC power
  • No DC power
  • No DC power available
  • Not reaching foot zone pressure
  • Not reaching head zone pressure
  • Not reaching left turn assist pressure
  • Not reaching right turn assist pressure
  • Not reaching seat zone pressure
  • Outlet cable routing verification
  • Patient control backlight LEDs cannot be disabled
  • Patient head down function does not work correctly
  • Patient head up function does not work correctly
  • Patient pendant malfunction
  • Power Cord replacement
  • Power supply board replacement
  • Power supply PCB error codes
  • PPM will not arm with scale display
  • PPM will not arm without scale display
  • Pressure Relief/Normal mode does not activate
  • Preventive Maintenance Checklist
  • Replace foot hilow motor
  • Replacement: Air Compressor
  • Replacement: Control Pod P.C. Board
  • Replacement: Foot Pedal Assembly
  • Replacement: Night Light
  • Replacement: Patient Control P.C. Board
  • Reverse Trendelenburg has problems or abnormal noise
  • Right Turn Assist bladder does not inflate or alarm does not sound
  • SafeView alerts flashing yellow and green
  • SafeView Alerts lights do not operate correctly
  • SafeView® Alert lights flash yellow and green
  • Scale display does not zero or shows incorrect weight
  • Scale display error Err0
  • Scale display error Err1
  • Scale display error Err2
  • Scale display error Err3
  • Scale error
  • Seat section bladder assembly replacement
  • Seat section bladder hose disconnection difficult
  • Service Required indicator is on after hardware removal
  • Service Required indicator solid
  • SideCom communication system error
  • SideCom communication system malfunction
  • Siderail detection switch error with SideCom
  • Siderail detection switch error without SideCom
  • Siderail detection switch malfunction
  • Siderail does not latch
  • Siderails do not lock in raised position
  • Steady alarm sounds when brake is not set
  • Steer caster does not lock in lengthways position
  • Transformer and Battery Fuse replacement
  • Transport handle cables require voltage adjustment
  • Transport handle replacement procedure
  • Transportation mode error code displayed
  • Treatment surface inflation not maximum
  • Treatment surface zones do not inflate to high pressure
  • Treatment surface zones not inflated in flat position
  • Trendelenburg at intermediate height has problems or abnormal noise
  • Trendelenburg function does not work
  • Turn Assist bladder does not inflate or alarm fails
  • Wheel will not stow
  • WIU does not connect to facility

164 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
¤ ¤¤Throttle Fault 1Controller not receiving correct throttle voltage from handleserror
¤¤¤ ¤¤¤¤¤HPDHigh Power Detect circuit faultcritical
¤¤¤ ¤¤¤¤Brake OFF FaultBrake circuit fails to releasecritical
¤¤¤ ¤PROC/Wiring FaultProcessor or internal wiring failurecritical
¤¤¤¤ ¤¤¤¤Power Section FaultPower output stage failureerror
¤¤¤¤ ¤¤HW FailsafeHardware failsafe triggered unexpectedlyerror
¤¤¤¤ ¤Current Sense FaultCurrent sensor or measurement circuit failureerror
¤¤ ¤¤¤¤Main ON FaultMain contactor ON circuit failureerror
¤¤ ¤¤¤Main Cont FLTSMain contactor fault detectederror
¤¤ ¤Main OFF FaultMain contactor OFF circuit failureerror
¤ ¤¤¤¤¤OvervoltageBattery voltage exceeds maximum operating rangewarning
¤ ¤¤¤¤Low Battery VoltageBattery voltage below acceptable operating levelwarning
¤ ¤¤¤SPD Limit Pot FaultSpeed limit potentiometer circuit failure in controller or junction boarderror
¤ ¤Thermal CutbackController detected overheating conditionerror
0NO ERRORSNormal operation stateinfo
1STUCK SWITCHCaregiver pedal stuck or not releasingerror
10ENCLOSURE TEMPERATUREEnclosure temperature out of rangeerror
11SMBUSSMBus communication failureerror
1-2Mattress DisconnectedMattress disconnected from bed control systemwarning
12LOW VCCSupply voltage below minimum thresholderror
1-3Head Angle Sensor ErrorHead angle sensor malfunctionwarning
13LOW BATTERY VOLTAGEBattery voltage below minimum levelwarning
14CHARGE TIMEBattery charging time exceeded limiterror
15D TO A COUNT LIMITDigital to analog conversion count exceedederror
1-5Head FSR1 and FSR2 ErrorHead force sensor 1 and 2 failurewarning
16CHARGE CURRENTCharge current out of specificationerror
18VOLT/TIME SLOPEVoltage change rate abnormalerror
19CURRENT/TIME SLOPECurrent change rate abnormalerror
20HIGH BATTERY VOLTAGEBattery voltage above maximum thresholderror
21D TO A OUTPUTDigital to analog converter output errorerror
22CURRENT MEASUREMENT DIFFERENCEMeasured current values divergeerror
2-2Accumulator Pressure ErrorPneumatic accumulator pressure out of rangeerror
2-3Error Detected In Dynamic Air ModuleAir Module control failureerror
23CURRENT CALCULATIONCurrent calculation algorithm errorerror
24EXTERNAL EEPROM BUSExternal EEPROM communication failureerror
25CHARGE TIME 2Secondary battery charging time limit exceedederror
2-6Missing 24 VDC Input24 VDC power supply not detectedcritical
3-1Not Reaching Head Zone PressureHead pressure zone unable to reach set pressureerror
3-2Not Reaching Seat Zone PressureSeat pressure zone unable to reach set pressureerror
3-3Not Reaching Foot Zone PressureFoot pressure zone unable to reach set pressureerror
3-4Not Reaching Right Turn Assist PressureRight turn assist zone unable to reach set pressureerror
3-5Not Reaching Left Turn Assist PressureLeft turn assist zone unable to reach set pressureerror
4-2Scale ErrorScale/weight measurement system malfunctionerror
4-3Scale ErrorScale/weight measurement system malfunctionerror
4-4Scale ErrorScale/weight measurement system malfunctionerror
ACB GeneralAir Control Board general fault or error conditionACB malfunction detectederror
ACB HeartbeatAir Control Board heartbeat signal missing or failedACB communication loss or malfunctioncritical
ADC ErrorAnalog-to-digital converter malfunctionBlue DS29=0, Green DS3=0, Red DS4=1, Amber DS1=1, DS5=0error
Air node heartbeat failureSCB lost heartbeat from pneumatic control nodeRed D47=0, Yellow D49=2, Green D46=2critical
All Lockout LED ErrorBed lockout controls failedAll articulation joints lockederror
Battery charging current over limitCharger output current exceeds maximum safe valueBlue D108=2, Green D45=2, Red D41=1, Amber D44=0, Service LED=1error
Battery charging current sensor faultCurrent sensor or ADC for battery charging brokenBlue D108=2, Green D45=2, Red D41=1, Amber D44=1, Service LED=1error
Battery voltage < 18 V and > 2 VBattery voltage critically low but not completely deadBlue D108=2, Green D45=0, Red D41=1, Amber D44=0, Service LED=1critical
Battery voltage < 2 VBattery voltage completely absent or severely disconnectedBlue D108=2, Green D45=0, Red D41=1, Amber D44=1, Service LED=1critical
Battery voltage too highBattery voltage exceeds safe charging rangeBlue D108=2, Green D45=0, Red D41=1, Amber D44=2, Service LED=1warning
Blower Current ErrorAir circulation motor current out of rangeBlue DS29=1, Green DS3=1, Red DS4=2, Amber DS1=1, DS5=0error
Bootloader errorBootloader failedSystem startup failurecritical
CAN Heartbeat FailureACB communication heartbeat lost on CAN busBlue DS29=1, Green DS3=0, Red DS4=1, Amber DS1=0, DS5=0critical
Circuit failureGeneral electrical circuit malfunction detectedBlue D108=0, Green D45=2, Red D41=1, Amber D44=2, Service LED=1error
Composer communication errorSCB cannot communicate with Composer facility network moduleRed D47=1, Yellow D49=2, Green D46=0error
CPR Switch FailureCPR mode switch malfunctionCPR switch circuit failureerror
CRCCRC Checksum ErrorCommunication or data integrity failurecritical
Deflate HeadCommand to deflate head section activeBlue DS29=1, Green DS3=0, Red DS4=1, Amber DS1=2, DS5=0info
Deflate HeelCommand to deflate heel section activeBlue DS29=1, Green DS3=1, Red DS4=1, Amber DS1=2, DS5=0info
Deflate Left Turn AssistLeft turn assist section deflation activeBlue DS29=0, Green DS3=1, Red DS4=2, Amber DS1=1, DS5=0info
Deflate Right Turn AssistRight turn assist section deflation activeBlue DS29=0, Green DS3=1, Red DS4=2, Amber DS1=0, DS5=0info
Deflate SeatCommand to deflate seat section activeBlue DS29=1, Green DS3=1, Red DS4=0, Amber DS1=2, DS5=0info
Disconnected MattressMattress unit not detected or disconnectedBlue DS29=0, Green DS3=0, Red DS4=1, Amber DS1=2, DS5=0error
Drive Current Too LowMotor drive current insufficientPower supply or drive circuit failurewarning
EEPROM access errorUCB cannot access EEPROM memoryBlue D108=0, Green D45=0, Red D41=2, Amber D44=2, Service LED=1error
EEPROM data corruptUCB detected EEPROM data corruption on startupBlue D108=0, Green D45=0, Red D41=0, Amber D44=1, Service LED=1critical
EEPROM ErrorMemory write/read or data corruption errorBlue DS29=0, Green DS3=0, Red DS4=1, Amber DS1=0, DS5=0error
EEPROM FailureEEPROM memory failureMemory read/write errorerror
EEPROM read errorSCB memory read operation failedRed D47=2, Yellow D49=2, Green D46=1error
EEPROM read error for bed articulation dataUCB cannot read articulation motor configuration from EEPROMBlue D108=0, Green D45=0, Red D41=2, Amber D44=1, Service LED=1error
EEPROM soft errorNon-critical EEPROM error detected and correctedRed D47=1, Yellow D49=1, Green D46=1warning
EEPROM Software DiagnosticsCritical EEPROM or software error detectedBlue DS29=2, Green DS3=2, Red DS4=2, Amber DS1=2, DS5=0critical
EEPROM write errorSCB memory write operation failedRed D47=2, Yellow D49=2, Green D46=0error
Err0Scale Error Err0Scale display shows Err0 codeerror
Err1Scale system error or calibration faultScale initialization or calibration failureerror
Err2Scale system error or measurement faultScale measurement failure or invalid readingerror
Err3Scale system error or communication faultScale communication loss or data errorerror
Err3Scale Error Err3Scale display shows Err3 codeerror
Err4Scale Error Err4Scale display shows Err4 codeerror
Err5Scale Error Err5Scale display shows Err5 codeerror
Flash: Brake OFF FaultBrake OFF control circuit faultBrake release control failureerror
Flash: Brake ON FaultBrake ON control circuit faultBrake engagement control failureerror
Flash: Current Sense FaultMotor current sensing circuit faultCurrent sensor failure or circuit openerror
Flash: EEPROM FaultEEPROM memory circuit faultEEPROM read/write failureerror
Flash: HPDHigh Power Discharge or hardware protection device triggeredExcessive current or temperature detectionerror
Flash: HW FailsafeHardware failsafe activatedHardware safety mechanism triggerederror
Flash: Low Battery VoltageController battery voltage below acceptable levelBattery voltage too low for controller operationwarning
Flash: Main Cont FLTSMain motor contactor faultMain motor contactor not engaging or disengagingerror
Flash: Main OFF FaultMain motor OFF control circuit faultMotor OFF control signal failureerror
Flash: Main ON FaultMain motor ON control circuit faultMotor ON control signal failureerror
Flash: OvervoltageController battery voltage exceeds safe operating limitBattery voltage too high for controller circuitswarning
Flash: Power Section FaultPower section circuit faultPower delivery circuit malfunctionerror
Flash: Precharge FaultPrecharge circuit faultCapacitor precharge circuit not functioningerror
Flash: PROC/Wiring FaultProcessor or wiring circuit faultProcessor internal fault or wiring integrity failureerror
Flash: SPD Limit Pot FaultSpeed limit potentiometer circuit faultSPD Limit Pot circuit failure or cable faulterror
Flash: Thermal CutbackIncubator temperature outside safe operating range or drive train bindingBed temperature outside 58-103°F or mechanical obstruction in drivewarning
Flash: Throttle Fault 1Incorrect throttle voltage or signal from handle controlsThrottle input voltage outside expected range or cable/junction board faulterror
Foot Current OverloadFoot motor current exceededMotor stall or mechanical jamwarning
Foot Knee Leg Unexpected MotionFoot, knee, or leg moving without commandUnintended motor actuationwarning
Foot overcurrentFoot articulation motor drawing excessive currentBlue D108=0, Green D45=1, Red D41=1, Amber D44=2, Service LED=0error
Head Angle ErrorHead articulation position sensor faultBlue DS29=1, Green DS3=0, Red DS4=2, Amber DS1=0, DS5=0error
Head Current OverloadHead motor current exceededMotor stall or mechanical jamwarning
Head Knee Hilow Lockout LED ErrorMultiple articulation lockout failureHead, knee, or hilow joint stuckwarning
Head overcurrentHead articulation motor drawing excessive currentBlue D108=0, Green D45=1, Red D41=1, Amber D44=1, Service LED=0error
Head Unexpected MotionHead section moving without commandUnintended motor actuationwarning
Heartbeat FailureNetwork heartbeat signal lostCommunication link failureerror
Heartbeat failure of the ACBUCB lost heartbeat signal from air control boardBlue D108=0, Green D45=2, Red D41=1, Amber D44=0, Service LED=1critical
Heartbeat failure of the SCBUCB lost heartbeat signal from SideCom boardBlue D108=2, Green D45=1, Red D41=2, Amber D44=1, Service LED=1critical
Hilo foot overcurrentFoot section motor drawing excessive currentBlue D108=0, Green D45=1, Red D41=0, Amber D44=2, Service LED=0error
Hilo head overcurrentHead section motor drawing excessive currentBlue D108=0, Green D45=1, Red D41=0, Amber D44=1, Service LED=0error
Hilow Foot Current OverloadHilow foot motor current exceededMotor stall or mechanical jamwarning
Hilow Foot Position Rate HighHilow foot moving too fastPosition feedback anomalywarning
Hilow Foot Position Rate LowHilow foot moving too slowlyPosition feedback anomalywarning
Hilow Foot Unexpected MotionFoot hilow section moving unexpectedlyUnintended motor actuationwarning
Hilow Head Current OverloadHilow head motor current exceededMotor stall or mechanical jamwarning
Hilow Head Position Rate HighHilow head moving too fastPosition feedback anomalywarning
Hilow Head Position Rate LowHilow head moving too slowlyPosition feedback anomalywarning
Hilow Head Unexpected MotionHead hilow section moving unexpectedlyUnintended motor actuationwarning
Inflate HeadCommand to inflate head section activeBlue DS29=0, Green DS3=1, Red DS4=0, Amber DS1=2, DS5=0info
Inflate HeelCommand to inflate heel section activeBlue DS29=1, Green DS3=0, Red DS4=0, Amber DS1=2, DS5=0info
Inflate Left Turn AssistLeft turn assist section inflation activeBlue DS29=0, Green DS3=0, Red DS4=2, Amber DS1=1, DS5=0info
Inflate Right Turn AssistRight turn assist section inflation activeBlue DS29=0, Green DS3=0, Red DS4=2, Amber DS1=0, DS5=0info
Inflate SeatCommand to inflate seat section activeBlue DS29=0, Green DS3=1, Red DS4=1, Amber DS1=2, DS5=0info
Knee Current OverloadKnee motor current exceededMotor stall or mechanical jamwarning
Knee overcurrentKnee section motor drawing excessive currentBlue D108=1, Green D45=1, Red D41=0, Amber D44=1, Service LED=0error
LCB SPI communication lostUCB cannot communicate with lower control board via SPIBlue D108=2, Green D45=2, Red D41=0, Amber D44=0, Service LED=1error
Leg Current OverloadLeg motor current exceededMotor stall or mechanical jamwarning
Leg overcurrentLeg section motor drawing excessive currentBlue D108=0, Green D45=1, Red D41=2, Amber D44=0, Service LED=0error
Logic Circuit FailureLogic PCB circuit malfunctionLogic board component failureerror
Motor H-Bridge FET short detectionMotor driver power transistor short circuit detectedBlue D108=1, Green D45=1, Red D41=2, Amber D44=1, Service LED=1error
Motor overvoltage detectionMotor supply voltage exceeds safe limit during operationBlue D108=1, Green D45=1, Red D41=2, Amber D44=2, Service LED=1error
No errorNormal condition with weight display in 100 lb incrementsBlue DS29=x, Green DS3=x, Red DS4=x, Amber DS1=x, DS5=2info
Patient Pendant FailurePatient pendant not communicatingWireless link failure or dead batterywarning
PPM will not arm no scale displayPatient Position Monitor cannot be armed and scale display is blankPPM activation attempted with scale communication losserror
PPM will not arm with scale displayPatient Position Monitor system cannot be armed despite scale operationalPPM activation attempted with scale display present but error conditionwarning
Scale display shows blank or distorted messageScale display shows no text or garbled/distorted messageDisplay initialization failure or data corruptionwarning
Scale HB DataScale sensor heartbeat data missingBlue DS29=1, Green DS3=1, Red DS4=2, Amber DS1=0, DS5=0warning
Scale Input ErrorWeight sensor input out of range or disconnectedBlue DS29=1, Green DS3=0, Red DS4=2, Amber DS1=1, DS5=0error
SCB GeneralScale Control Board general fault or error conditionSCB malfunction detectederror
SCB HeartbeatScale Control Board heartbeat signal missing or failedSCB communication loss or malfunctioncritical
Solenoid OvervoltageSolenoid valve drive voltage exceeds limitBlue DS29=0, Green DS3=1, Red DS4=1, Amber DS1=1, DS5=0error
Solenoid UndervoltageSolenoid valve drive voltage too lowBlue DS29=1, Green DS3=2, Red DS4=0, Amber DS1=0, DS5=0error
SolidService Required Indicator SolidService Required indicator continuously illuminatedcritical
Supervisory Overvoltage, OffSupply voltage exceeds safe limit in off stateBlue DS29=0, Green DS3=1, Red DS4=0, Amber DS1=0, DS5=0warning
Supervisory Overvoltage, OnSupply voltage exceeds safe limit during operationBlue DS29=0, Green DS3=1, Red DS4=0, Amber DS1=1, DS5=0warning
Supervisory Relay ErrorSupervisory relay malfunctionRelay circuit failureerror
Supervisory UndervoltageSupply voltage drops below safe minimumBlue DS29=0, Green DS3=1, Red DS4=1, Amber DS1=0, DS5=0error
System CPR switch failureCardiopulmonary resuscitation switch not respondingBlue D108=0, Green D45=0, Red D41=1, Amber D44=2, Service LED=1error
System pendant failurePatient pendant wireless module not respondingBlue D108=0, Green D45=2, Red D41=1, Amber D44=1, Service LED=1error
UCB HeartbeatUniversal Control Board heartbeat signal missing or failedUCB communication loss or malfunctioncritical
UCB heartbeat failureSCB lost communication with UCBRed D47=2, Yellow D49=1, Green D46=0critical
Valve Overcurrent, Head, Seat, HeelSolenoid current exceeds limit for mattress sectionsBlue DS29=1, Green DS3=1, Red DS4=0, Amber DS1=0, DS5=0error
Valve Overcurrent, Turn AssistSolenoid current exceeds limit for turn assist functionBlue DS29=1, Green DS3=1, Red DS4=1, Amber DS1=0, DS5=0error

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.