Hill-Rom VersaCare: troubleshooting and error codes
The ServiceForge knowledge base holds 224 documented symptoms for the Hill-Rom VersaCare, with 277 ranked causes, 415 diagnostic checks with test points and expected values, and 158 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Alerts inactive when bed plugged in patient safety
SafeView alert lights do not illuminate after bed power-on
Ranked causes
- 1. Alerts deactivated in system settings (common)
- 2. Bed Exit System not activated (occasional)
- 3. Alert cables loose or disconnected (occasional)
- 4. Alert cable or assembly damaged (occasional)
- 5. UCB board failure (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Enable switch not functioning caution
Handle enable switch does not engage transport system or shows incorrect voltage readings
Ranked causes
- 1. Enable switch not plugged into Junction board (occasional)
- 2. Enable switch not generating correct voltage signal (occasional)
- 3. Defective enable switch or cable (occasional)
- 4. Low or absent battery voltage (occasional)
- 5. Problem on Junction board or P6 connector (occasional)
Verify enable switch connection to Junction board
- Expected
- Both switches plugged into P2 and P5
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SafeView alerts do not come on when plugged in patient safety
SafeView alert lights remain dark after bed is powered on
Ranked causes
- 1. Alerts deactivated (common)
- 2. Bed exit system not activated (occasional)
- 3. Both alert cables loose or disconnected (occasional)
- 4. Both alert cables damaged (rare)
- 5. Siderail interface PC board failure (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air Control Board or Blower Power Board failure high voltage
Air module not functioning; blower inoperative or incorrect voltage output; weight sensing inaccurate
Ranked causes
- 1. Defective Air Control Board (ACB) (unknown)
- 2. Defective Blower Power Board (unknown)
- 3. Blower power board voltage out of specification (occasional)
- 4. Patient weight offset not calibrated (occasional)
Verify blower is operating
- Test point
- Green LED on blower power board (D)
- Expected
- LED is on when power present
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Enable switch not responding
Bed does not move when handle switch is depressed or pulled. Operator reports switch inoperative.
Ranked causes
- 1. Enable switch cable disconnected or not seated at P2 or P5 (common)
- 2. Enable switch or cable defective (always open or always closed) (occasional)
- 3. No voltage at P6.1 due to low or missing battery voltage (occasional)
- 4. Junction board or P6 connector failure (rare)
Verify enable switches plugged into Junction board
- Test point
- P2 and P5
- Expected
- Both switches fully seated
7 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SafeView single alert flashing or off color patient safety
Only one SafeView alert light on or flashing yellow and green
Ranked causes
- 1. Alert cable loose or disconnected (common)
- 2. Alert cable damaged (occasional)
- 3. Alert light assembly PC board failure (occasional)
- 4. Siderail interface PC board failure (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Throttle output signal out of range
Handle force signal does not respond correctly to push/pull actions. Bed may not accelerate or decelerate as expected.
Ranked causes
- 1. Throttle zero offset requires adjustment (common)
- 2. Low supply voltage at P6.8 (occasional)
- 3. Low excitation voltage at P3.5 or P7.5 (red wire) (occasional)
- 4. Strain element in handle damaged (rare)
Measure supply voltage at P6.8
- Test point
- P6.8
- Expected
- more than 22 V
- Tools
- DMM
8 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Throttle signal incorrect or out of range caution
Handle throttle signal does not produce expected voltage levels; bed speed control not working properly
Ranked causes
- 1. Throttle signal out of neutral range (occasional)
- 2. Low excitation voltage to throttle circuit (occasional)
- 3. Damaged handle strain element (occasional)
- 4. Cable continuity or Junction board fault (occasional)
Measure throttle output at P6.10 with no handle force
- Test point
- P6.10
- Expected
- 2.4-2.6 V, +/- 0.1 V
- Tools
- DC Volt Meter
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed frame does not operate patient safety
Bed sections (head, knee, foot, or hilow) do not move when controls are pressed, or move unexpectedly
Ranked causes
- 1. Control lockout is engaged (common)
- 2. Motor or drive mechanism failure (occasional)
- 3. Electronic control system fault (occasional)
Verify control lockout status
- Test point
- Lockout indicator LEDs
- Expected
- Indicator should be off (unlocked)
8 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Motor controller not receiving 4700 ohm resistor signal
Motor does not operate. PACM board may not detect Junction board connection or resistor continuity is lost.
Ranked causes
- 1. 4700 ohm resistor on Junction board open or Junction board failure (occasional)
- 2. 14-pin cable between Junction and PACM board defective or disconnected (occasional)
- 3. Connector P6 on Junction board pins damaged or corroded (rare)
Unplug 14-pin cable from PACM board
- Test point
- PACM board 14-pin connector
- Expected
- Cable safely unplugged
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Motor controller not receiving resistance signal caution
Motor controller will not operate; connection to junction board 4700 ohm resistor is missing or open
Ranked causes
- 1. Open or defective PACM-to-Junction cable (occasional)
- 2. Faulty P6 connector on Junction board (occasional)
- 3. Defective Junction board resistor (occasional)
Measure resistance across PACM cable pins 12-13
- Test point
- 14-pin cable pins 12, 13
- Expected
- 4700 ohms +/- 5%, +/- 5%
- Tools
- Ohm Meter
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Only one alert on or flashing yellow and green patient safety
Single SafeView alert light malfunctioning while other operates normally
Ranked causes
- 1. Alert cable loose, disconnected, or damaged (common)
- 2. Alert light assembly PCB damaged (occasional)
- 3. UCB board malfunction (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SafeView both alerts flash yellow and green patient safety
Both SafeView alert lights flashing yellow and green simultaneously
Ranked causes
- 1. SideCom PC board malfunction (occasional)
- 2. Siderail interface PC board malfunction (occasional)
- 3. Control pod PC board malfunction (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scale system malfunction after P.C. board work
System does not function properly after scale P.C. board removal and replacement.
Ranked causes
- 1. Improper connector seating or connector alignment (common)
- 2. Scale calibration not performed or incomplete (common)
- 3. Component damage from electrostatic discharge during handling (occasional)
Verify all connectors fully seated
- Test point
- Connectors P1, P2, P3, P4 on P.C. board H
- Expected
- All connectors seated and latched
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery voltage out of range high voltage
Battery or charging system not providing correct voltage; controller unable to operate
Ranked causes
- 1. Charging system failure (common)
- 2. Battery discharged (common)
Measure battery voltage at P6.8
- Test point
- P6.8
- Expected
- > 32 V when plugged into AC, > 22 V on battery, > 32V AC, > 22V battery
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Boost Position System does not operate patient safety
Bed does not move to middle height or assumes incorrect position when Boost control is activated
Ranked causes
- 1. Control lockout engaged (common)
- 2. Boost motor or control circuit failure (occasional)
Verify lockouts are disengaged
- Test point
- Control lockout indicators
- Expected
- All lockout indicators are off
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Caster or wheel locks do not function patient safety
Wheels do not lock when brake is set, or casters do not lock in expected position
Ranked causes
- 1. Brake mechanism failure or wear (common)
- 2. Steer caster locking mechanism failure (non-IntelliDrive beds) (occasional)
Test wheel brake engagement
- Test point
- Brake pedal, attempt to move bed
- Expected
- The four wheels are locked and prevent any movement
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Controller LED not responding high voltage
LED DS 1 is off; controller either not powered or not receiving KSI signal from wheel deployment
Ranked causes
- 1. Controller not powered (common)
- 2. KSI signal not received (occasional)
Observe LED DS 1 state on PACM board after pressing Enable switch
- Test point
- DS 1 LED
- Expected
- LED should be either on steady or off
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CPR cable removal and replacement caution
Complete removal and reinstallation procedure for CPR cable assembly including routing and cable tie attachment
Ranked causes
- 1. CPR cable requires removal or replacement (occasional)
- 2. CPR cable requires installation after removal (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CPR handle does not function patient safety
CPR handle does not lower head section quickly or movement is not cushioned, or head section does not continue into Trendelenburg
Ranked causes
- 1. CPR actuation mechanism failure or wear (common)
- 2. Trendelenburg function not working after CPR activation (occasional)
Test CPR handle head section descent
- Test point
- CPR handle, head section movement
- Expected
- The head section descends quickly up to the mid-travel then the movement is cushioned until the low position is reached without problem or abnormal noise
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head FSR error or blower current error
Service Required LED flashing with pattern 1 long, 5 short flashes. A Model Bed: Head FSR1 and FSR2 error. H Model or Newer Bed with AMT: Blower current error.
Ranked causes
- 1. Head FSR1 or FSR2 sensor failure (A Model) (unknown)
- 2. Blower current error (H Model or Newer with AMT) (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect throttle voltage
Handle throttle input voltage outside expected range; bed does not respond to forward/back force correctly
Ranked causes
- 1. Incorrect throttle voltage at controller input (common)
- 2. Damaged Junction board or cable (occasional)
Measure throttle input voltage at controller
- Test point
- controller throttle input
- Expected
- 2.5 V neutral, 4.0 V pushed forward, 1.0 V pulled back, +/-0.5 V
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Motor controller resistor fault
4700 ohm termination resistor on Junction board not present or damaged; controller will not operate
Ranked causes
- 1. Resistor disconnected or damaged (rare)
- 2. Cable continuity broken (occasional)
Measure resistance between PACM board pins 12 and 13
- Test point
- PACM connector pins 12, 13
- Expected
- 4700 ohms +/- 5%, +/- 5%
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SafeView® Alert lights flashing yellow during configured siderail movements
Lights flash yellow when configured siderails are lowered or bed is raised, indicating proper alert system operation
Ranked causes
- 1. Normal operation during bed and siderail movement (common)
- 2. Head angle alarm active and bed lowered below alarm threshold (common)
Verify siderail configuration is set correctly
- Test point
- Control pod buttons
- Expected
- Lights flash green for three seconds after five-second button press
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SafeView® Alert lights off when expected
Lights remain off when Bed Exit Alarm System deactivated or when specific deactivation procedure performed
Ranked causes
- 1. Bed Exit Alarm System not active or lights have been deactivated (common)
- 2. Bed on battery power or disconnected from AC power (common)
Verify Bed Exit Alarm System is not active before deactivation procedure
- Test point
- System status indicator or control pod
- Expected
- Bed Exit Alarm System inactive
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scale and bed exit alarm system does not function patient safety
Zero control, weight detection, Enable control, bed exit modes, or alarm sound fails to operate correctly
Ranked causes
- 1. Scale system malfunction (unknown)
- 2. Bed exit alarm system malfunction (unknown)
Zero control display with 50 lb weight on bed center
- Test point
- Display
- Expected
- Shows 00.0
6 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transport system not responding to handle input caution
IntelliDrive transport system does not move when handles are engaged; unable to move bed forward, backward, or steer
Ranked causes
- 1. Controller not powered or not receiving KSI signal (common)
- 2. Controller error condition - LED flashing (occasional)
Monitor DS 1 LED indicator on PACM board
- Expected
- LED is on and steady when functioning correctly
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Trendelenburg function does not operate patient safety
Sleep surface does not tilt to Trendelenburg or reverse Trendelenburg position, or tilts abnormally
Ranked causes
- 1. Trendelenburg motor or actuator failure (common)
- 2. Mechanical linkage misalignment or damage (occasional)
Test Trendelenburg function at intermediate bed height
- Test point
- Trendelenburg control, bed at intermediate position
- Expected
- The sleep surface gradually tilts to maximum trendelenburg without any problem or abnormal noise
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
30° Head Angle Alarm does not disable
Pressing Enable control followed by 30° Head Angle Alarm control does not turn off the 30° Head Angle Alarm indicator
Ranked causes
- 1. 30° Head Angle Alarm control or disable circuit failure (unknown)
Press Enable control while alarm is active and indicator is on
- Test point
- Enable control
- Expected
- Enable mode activated
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
30° Head Angle Alarm does not sound when head angle below 30°
Lowering head section below 30° does not trigger alarm sounding and indicator flashing for 30° Head Angle Alarm option
Ranked causes
- 1. Head angle sensor or alarm circuit malfunction (unknown)
Lower head section below 30°
- Test point
- Head Down control and head position
- Expected
- Head section angle is below 30°
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
30° Head Angle Alarm does not stop when head angle above 30°
Raising head section above 30° does not stop alarm and does not leave indicator on solid for 30° Head Angle Alarm option
Ranked causes
- 1. Head angle sensor or alarm circuit malfunction (unknown)
Raise head section above 30°
- Test point
- Head Up control and head position
- Expected
- Head section angle is above 30°
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
30° Head Angle Alarm indicator does not activate patient safety
Enable and 30° Head Angle Alarm controls pressed but indicator does not illuminate.
Ranked causes
- 1. 30° Head Angle Alarm option not functional or wiring fault (unknown)
Press Enable control then 30° Head Angle Alarm control
- Test point
- Alarm option indicator
- Expected
- Indicator comes on
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
30° Head Angle Alarm indicator does not illuminate
Pressing Enable control followed by 30° Head Angle Alarm control does not illuminate the indicator light for the 30° Head Angle Alarm option
Ranked causes
- 1. 30° Head Angle Alarm circuit or indicator lamp failure (unknown)
Press Enable control
- Test point
- Enable control
- Expected
- Enable mode is activated
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC Power Fuses replacement high voltage
AC power fuses located in power supply module require replacement after disconnecting power
Ranked causes
- 1. AC power fuse blown or degraded (unknown)
Set brakes
- Test point
- Brake controls
- Expected
- Brakes engaged
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC power present but no motor power high voltage
AC power available but DC motor power supply has failed
Ranked causes
- 1. DC power supply failure or connection issue (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Accumulator pressure error caution
Service Required LED flashing with pattern 2 long, 2 short flashes
Ranked causes
- 1. Accumulator pressure out of specification (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm does not sound when brake is not set patient safety
No steady alarm when bed is unplugged from power and brake is not set
Ranked causes
- 1. Alarm circuit or speaker failure (occasional)
Test alarm with brake not set
- Test point
- Power source, brake pedal position
- Expected
- A steady alarm comes on when brake is not set and bed is plugged in
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm does not stop when brake is set
G model or newer bed continues sounding steady alarm after brake is set
Ranked causes
- 1. Brake safety circuit or sensor malfunction (unknown)
Set brake
- Test point
- Brake control
- Expected
- Alarm stops immediately when brake is fully set
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm Silence feature does not function correctly patient safety
Alarm Silence indicator fails to activate or silence alarm, or system does not arm properly with weight on bed
Ranked causes
- 1. Alarm Silence feature system malfunction (unknown)
Enable and Alarm Silence control activation
- Test point
- Alarm Silence indicator
- Expected
- Indicator comes on
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm Silence feature does not operate patient safety
Pressing Alarm Silence control during bed exit alarm does not stop sound or indicator.
Ranked causes
- 1. Alarm Silence control or circuit fault (unknown)
Cause bed exit alarm condition with weight on bed
- Test point
- Alarm and indicators
- Expected
- Alarm sounds
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery backup does not operate patient safety
Bed functions do not work when unplugged from AC power and running on battery backup
Ranked causes
- 1. Battery failure or discharge (common)
Test head and knee functions on battery power
- Test point
- Bed unplugged, head/knee controls operated
- Expected
- The bed works correctly on battery power for head and knee functions
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery backup does not operate bed functions patient safety
After unplugging bed and activating battery backup, functions do not work correctly
Ranked causes
- 1. Battery backup circuit failure (unknown)
Unplug bed and activate battery backup; operate functions 10-12
- Test point
- Battery backup system
- Expected
- Head, knee, foot functions work correctly on battery power
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery charging error with AC power on caution
Battery charging indicator off despite AC mains present
Ranked causes
- 1. Battery charger malfunction or battery failure (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery replacement high voltage
Batteries must be replaced in pairs; procedure involves disconnecting power supply, removing cover, disconnecting cables, removing bracket and batteries, checking vent holes for blockage.
Ranked causes
- 1. Batteries depleted or faulty (unknown)
Set brakes on bed
- Test point
- Brake control
- Expected
- Brakes engaged
11 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed does not enter Trendelenburg when CPR handle held patient safety
Continuing to hold CPR handle after head descent does not transition bed into Trendelenburg position
Ranked causes
- 1. CPR control or trendelenburg sequencing circuit failure (unknown)
Continue holding CPR handle after head descent completes
- Test point
- CPR handle held position
- Expected
- Bed transitions into Trendelenburg position
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed does not lower to low position patient safety
Pressing Bed Down control does not lower bed to low position or bed stops during movement; Bed Not Down indicator does not turn off at maximum low position
Ranked causes
- 1. Motor, control, or mechanical failure (unknown)
Press Bed Down control
- Test point
- Bed Down control
- Expected
- Bed lowers to low position without stopping
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed does not raise to high position patient safety
Pressing Bed Up control does not raise bed to high position or bed stops during movement
Ranked causes
- 1. Motor, control, or mechanical failure (unknown)
Press Bed Up control
- Test point
- Bed Up control
- Expected
- Bed rises to high position without stopping
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed exit alarm does not activate or indicator malfunction patient safety
Enable control active but bed exit mode control does not arm system or alarm fails when patient leaves bed.
Ranked causes
- 1. Bed exit alarm circuit, sensor, or control fault (unknown)
Place 50 lb weight on center of bed and press Enable control
- Test point
- Bed exit alarm Enable indicator
- Expected
- Indicator comes on
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed flat function does not work correctly patient safety
Pressing Bed Flat control does not return head, knee, foot sections to horizontal; on G model or newer, sleep deck and bed do not return to flat and level position
Ranked causes
- 1. Motor, control, or mechanical failure (unknown)
Press Bed Flat control
- Test point
- Bed Flat control
- Expected
- All sections gradually return to horizontal while control is pressed; G model beds return to flat and level
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed not down indicator flashing patient safety
System indicates bed is not fully lowered when it should be
Ranked causes
- 1. Position sensor fault or hilow not fully down (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed reverse trendelenburg function does not operate patient safety
Reverse Trendelenburg control pressed but sleep surface does not tilt to maximum reverse trendelenburg or produces abnormal noise
Ranked causes
- 1. Motor or drive mechanism failure (unknown)
Operate the Reverse Trendelenburg control
- Test point
- Bed at intermediate position
- Expected
- Sleep surface gradually tilts to maximum reverse trendelenburg without problem or abnormal noise
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed trendelenburg function does not operate patient safety
Trendelenburg control pressed but sleep surface does not tilt gradually to maximum trendelenburg angle or produces abnormal noise
Ranked causes
- 1. Motor or drive mechanism failure (unknown)
Operate the Trendelenburg control at intermediate bed height
- Test point
- Bed raised to intermediate position
- Expected
- Sleep surface gradually tilts to maximum trendelenburg without problem or abnormal noise
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed wheels do not lock when brake is set patient safety
Setting brake does not lock all four wheels; bed moves despite brake being engaged
Ranked causes
- 1. Brake mechanism failure or damaged caster (unknown)
Set all bed brakes
- Test point
- Brake control
- Expected
- All four wheels lock and prevent bed movement
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed will not drive
IntelliDrive powered mobility system fails to propel bed
Ranked causes
- 1. PACM board power loss or drive motor failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed will not drive with wheel down
Mobility system fails to move with transport wheel in contact with floor
Ranked causes
- 1. Wheel deployment issue or motor brake engaged (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Boost position function does not work correctly patient safety
For beds with Boost Position System option, pressing and holding Boost control does not move bed to middle height with head and knee sections lowering and bed at ~7° Trendelenburg
Ranked causes
- 1. Boost circuit or control system failure; lockout engaged (unknown)
Verify lockouts are disengaged
- Test point
- Head, knee, bed up/down lockout controls
- Expected
- All lockout indicators are off
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: AC power fuse replacement high voltage
Procedure to remove and replace AC power fuses in the power supply module
Ranked causes
- 1. AC power fuse requires replacement (unknown)
Set brakes
- Test point
- Bed brake control
- Expected
- Brakes engaged
9 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Disable AMT mode
Configure Air Control Board to non-AMT operation after replacement or per specification
Ranked causes
- 1. AMT factory default enabled; must be disabled for this bed configuration (common)
Enter AMT disable mode
- Test point
- Control pod: Turn R + Turn L + Max-Inflate + Normal
- Expected
- Single beep after ~5 seconds
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Drive Belt—IntelliDrive® Transport System caution
Procedure to remove and replace the drive belt on IntelliDrive® transport system. Includes disassembly of drive box, motor mounting, pulley plates, and belt removal/installation steps.
Ranked causes
- 1. Drive belt replacement procedure (unknown)
Set the brakes
- Test point
- Bed brake mechanism
- Expected
- Brakes engaged and locked
11 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Drive Motor IntelliDrive Transport System caution
Removal and replacement procedure for drive motor assembly in VersaCare hospital bed IntelliDrive transport system.
Ranked causes
- 1. Drive motor replacement required (unknown)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Calibration: Head Angle Sensor (G-J model beds)
- Calibration: Head angle sensor for G model or newer
- Calibration: Head Siderail P.C. Board
- Calibration: Patient weight offset
- Calibration: Scale
- Calibration: Seat Section Bladder Assembly replacement
- Calibration: Sensor Strip replacement
- Calibration: Transformer and Battery Fuse
- Calibration: Transport Handle IntelliDrive voltage adjustment
- Calibration: Turn Assist Bladder Assembly replacement
- Cannot view software version on optional pod
- Caster removal foot end
- Caster removal head end
- Caster replacement
- Chair position function does not work correctly
- Control lockout indicator does not illuminate
- Control Pod P.C. Board removal and replacement procedure
- Control unlock indicator does not turn off
- CPR cable adjustment
- CPR function does not continue into Trendelenburg position
- CPR handle function does not work correctly
- CPR handle quick release not functioning properly
- CRC error
- Damaged handle cable or connector
- Damaged handle strain element
- Diagnostic or battery charging errors present
- Drive Motor Removal Procedure
- Error detected in dynamic air module
- Foot section does not extend to full out position
- Foot section does not retract to full in position
- Head angle alarm does not sound when head exceeds 30°
- Head Angle Display does not show 0° in level position
- Head Angle Display does not show 62° at maximum head height
- Head Angle Display shows incorrect angle
- Head angle sensor error
- Head section bladder assembly replacement
- Head section does not deflate when squeezed
- Head section does not inflate properly
- Head section does not lower to low position
- Head section does not re-inflate when pressure released
- Head section does not rise smoothly after CPR
- Head section does not rise to high position
- Head section motor does not operate after CPR handle release
- Head zone pressure lower than seat zone in articulated position
- Head zone pressure not lower than seat zone
- Head, knee, leg, or foot motor unexpected motion
- Heartbeat error
- Hilow motor position rate high
- Hilow motor position rate low
- Hilow motor unexpected motion
- IntelliDrive battery check
- IntelliDrive bed will not drive
- IntelliDrive bed will not drive with wheel down
- IntelliDrive controller check
- IntelliDrive handle enable switch check
- IntelliDrive handle gauge check
- IntelliDrive junction board debugging
- IntelliDrive motor check
- IntelliDrive PACM board deployment check
- IntelliDrive PACM board drive check
- IntelliDrive PACM board power check
- IntelliDrive PACM to junction board cable check
- IntelliDrive steer switch check
- IntelliDrive throttle check
- IntelliDrive Transport System does not move
- IntelliDrive visual inspection
- IntelliDrive wheel will not stow
- Knee section does not lower to low position
- Knee section does not rise to high position
- Left Turn Assist bladder does not inflate or alarm does not sound
- Load Beam—J Models replacement
- Logic P.C. Board All Lockout LED error
- Logic P.C. Board Bootloader error
- Logic P.C. Board CPR Switch failure
- Logic P.C. Board Drive Current Too Low error
- Logic P.C. Board EEPROM failure
- Logic P.C. Board Foot Current Overload error
- Logic P.C. Board Foot, Knee, Leg Unexpected Motion error
- Logic P.C. Board Head Current Overload error
- Logic P.C. Board Head Unexpected Motion error
- Logic P.C. Board Head, Knee, Hilow Lockout LED error
- Logic P.C. Board Heartbeat Failure error
- Logic P.C. Board Hilow Foot Current Overload error
- Logic P.C. Board Hilow Foot Position Rate High error
- Logic P.C. Board Hilow Foot Position Rate Low error
- Logic P.C. Board Hilow Foot Unexpected Motion error
- Logic P.C. Board Hilow Head Current Overload error
- Logic P.C. Board Hilow Head Position Rate High error
- Logic P.C. Board Hilow Head Position Rate Low error
- Logic P.C. Board Hilow Head Unexpected Motion error
- Logic P.C. Board Knee Current Overload error
- Logic P.C. Board Leg Current Overload error
- Logic P.C. Board Logic Circuit Failure error
- Logic P.C. Board Patient Pendant Failure error
- Logic P.C. Board Supervisory Relay error
- Mattress disconnected
- Mattress three zones do not inflate to high pressure
- Missing 24 VDC input
- Motor controller removal and replacement procedure
- Motor controller replacement needed
- Motors do not operate
- Night light cannot be disabled
- No AC power
- No DC power
- No DC power available
- Not reaching foot zone pressure
- Not reaching head zone pressure
- Not reaching left turn assist pressure
- Not reaching right turn assist pressure
- Not reaching seat zone pressure
- Outlet cable routing verification
- Patient control backlight LEDs cannot be disabled
- Patient head down function does not work correctly
- Patient head up function does not work correctly
- Patient pendant malfunction
- Power Cord replacement
- Power supply board replacement
- Power supply PCB error codes
- PPM will not arm with scale display
- PPM will not arm without scale display
- Pressure Relief/Normal mode does not activate
- Preventive Maintenance Checklist
- Replace foot hilow motor
- Replacement: Air Compressor
- Replacement: Control Pod P.C. Board
- Replacement: Foot Pedal Assembly
- Replacement: Night Light
- Replacement: Patient Control P.C. Board
- Reverse Trendelenburg has problems or abnormal noise
- Right Turn Assist bladder does not inflate or alarm does not sound
- SafeView alerts flashing yellow and green
- SafeView Alerts lights do not operate correctly
- SafeView® Alert lights flash yellow and green
- Scale display does not zero or shows incorrect weight
- Scale display error Err0
- Scale display error Err1
- Scale display error Err2
- Scale display error Err3
- Scale error
- Seat section bladder assembly replacement
- Seat section bladder hose disconnection difficult
- Service Required indicator is on after hardware removal
- Service Required indicator solid
- SideCom communication system error
- SideCom communication system malfunction
- Siderail detection switch error with SideCom
- Siderail detection switch error without SideCom
- Siderail detection switch malfunction
- Siderail does not latch
- Siderails do not lock in raised position
- Steady alarm sounds when brake is not set
- Steer caster does not lock in lengthways position
- Transformer and Battery Fuse replacement
- Transport handle cables require voltage adjustment
- Transport handle replacement procedure
- Transportation mode error code displayed
- Treatment surface inflation not maximum
- Treatment surface zones do not inflate to high pressure
- Treatment surface zones not inflated in flat position
- Trendelenburg at intermediate height has problems or abnormal noise
- Trendelenburg function does not work
- Turn Assist bladder does not inflate or alarm fails
- Wheel will not stow
- WIU does not connect to facility
164 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
¤ ¤¤ | Throttle Fault 1 | Controller not receiving correct throttle voltage from handles | error |
¤¤¤ ¤¤¤¤¤ | HPD | High Power Detect circuit fault | critical |
¤¤¤ ¤¤¤¤ | Brake OFF Fault | Brake circuit fails to release | critical |
¤¤¤ ¤ | PROC/Wiring Fault | Processor or internal wiring failure | critical |
¤¤¤¤ ¤¤¤¤ | Power Section Fault | Power output stage failure | error |
¤¤¤¤ ¤¤ | HW Failsafe | Hardware failsafe triggered unexpectedly | error |
¤¤¤¤ ¤ | Current Sense Fault | Current sensor or measurement circuit failure | error |
¤¤ ¤¤¤¤ | Main ON Fault | Main contactor ON circuit failure | error |
¤¤ ¤¤¤ | Main Cont FLTS | Main contactor fault detected | error |
¤¤ ¤ | Main OFF Fault | Main contactor OFF circuit failure | error |
¤ ¤¤¤¤¤ | Overvoltage | Battery voltage exceeds maximum operating range | warning |
¤ ¤¤¤¤ | Low Battery Voltage | Battery voltage below acceptable operating level | warning |
¤ ¤¤¤ | SPD Limit Pot Fault | Speed limit potentiometer circuit failure in controller or junction board | error |
¤ ¤ | Thermal Cutback | Controller detected overheating condition | error |
0 | NO ERRORS | Normal operation state | info |
1 | STUCK SWITCH | Caregiver pedal stuck or not releasing | error |
10 | ENCLOSURE TEMPERATURE | Enclosure temperature out of range | error |
11 | SMBUS | SMBus communication failure | error |
1-2 | Mattress Disconnected | Mattress disconnected from bed control system | warning |
12 | LOW VCC | Supply voltage below minimum threshold | error |
1-3 | Head Angle Sensor Error | Head angle sensor malfunction | warning |
13 | LOW BATTERY VOLTAGE | Battery voltage below minimum level | warning |
14 | CHARGE TIME | Battery charging time exceeded limit | error |
15 | D TO A COUNT LIMIT | Digital to analog conversion count exceeded | error |
1-5 | Head FSR1 and FSR2 Error | Head force sensor 1 and 2 failure | warning |
16 | CHARGE CURRENT | Charge current out of specification | error |
18 | VOLT/TIME SLOPE | Voltage change rate abnormal | error |
19 | CURRENT/TIME SLOPE | Current change rate abnormal | error |
20 | HIGH BATTERY VOLTAGE | Battery voltage above maximum threshold | error |
21 | D TO A OUTPUT | Digital to analog converter output error | error |
22 | CURRENT MEASUREMENT DIFFERENCE | Measured current values diverge | error |
2-2 | Accumulator Pressure Error | Pneumatic accumulator pressure out of range | error |
2-3 | Error Detected In Dynamic Air Module | Air Module control failure | error |
23 | CURRENT CALCULATION | Current calculation algorithm error | error |
24 | EXTERNAL EEPROM BUS | External EEPROM communication failure | error |
25 | CHARGE TIME 2 | Secondary battery charging time limit exceeded | error |
2-6 | Missing 24 VDC Input | 24 VDC power supply not detected | critical |
3-1 | Not Reaching Head Zone Pressure | Head pressure zone unable to reach set pressure | error |
3-2 | Not Reaching Seat Zone Pressure | Seat pressure zone unable to reach set pressure | error |
3-3 | Not Reaching Foot Zone Pressure | Foot pressure zone unable to reach set pressure | error |
3-4 | Not Reaching Right Turn Assist Pressure | Right turn assist zone unable to reach set pressure | error |
3-5 | Not Reaching Left Turn Assist Pressure | Left turn assist zone unable to reach set pressure | error |
4-2 | Scale Error | Scale/weight measurement system malfunction | error |
4-3 | Scale Error | Scale/weight measurement system malfunction | error |
4-4 | Scale Error | Scale/weight measurement system malfunction | error |
ACB General | Air Control Board general fault or error condition | ACB malfunction detected | error |
ACB Heartbeat | Air Control Board heartbeat signal missing or failed | ACB communication loss or malfunction | critical |
ADC Error | Analog-to-digital converter malfunction | Blue DS29=0, Green DS3=0, Red DS4=1, Amber DS1=1, DS5=0 | error |
Air node heartbeat failure | SCB lost heartbeat from pneumatic control node | Red D47=0, Yellow D49=2, Green D46=2 | critical |
All Lockout LED Error | Bed lockout controls failed | All articulation joints locked | error |
Battery charging current over limit | Charger output current exceeds maximum safe value | Blue D108=2, Green D45=2, Red D41=1, Amber D44=0, Service LED=1 | error |
Battery charging current sensor fault | Current sensor or ADC for battery charging broken | Blue D108=2, Green D45=2, Red D41=1, Amber D44=1, Service LED=1 | error |
Battery voltage < 18 V and > 2 V | Battery voltage critically low but not completely dead | Blue D108=2, Green D45=0, Red D41=1, Amber D44=0, Service LED=1 | critical |
Battery voltage < 2 V | Battery voltage completely absent or severely disconnected | Blue D108=2, Green D45=0, Red D41=1, Amber D44=1, Service LED=1 | critical |
Battery voltage too high | Battery voltage exceeds safe charging range | Blue D108=2, Green D45=0, Red D41=1, Amber D44=2, Service LED=1 | warning |
Blower Current Error | Air circulation motor current out of range | Blue DS29=1, Green DS3=1, Red DS4=2, Amber DS1=1, DS5=0 | error |
Bootloader error | Bootloader failed | System startup failure | critical |
CAN Heartbeat Failure | ACB communication heartbeat lost on CAN bus | Blue DS29=1, Green DS3=0, Red DS4=1, Amber DS1=0, DS5=0 | critical |
Circuit failure | General electrical circuit malfunction detected | Blue D108=0, Green D45=2, Red D41=1, Amber D44=2, Service LED=1 | error |
Composer communication error | SCB cannot communicate with Composer facility network module | Red D47=1, Yellow D49=2, Green D46=0 | error |
CPR Switch Failure | CPR mode switch malfunction | CPR switch circuit failure | error |
CRC | CRC Checksum Error | Communication or data integrity failure | critical |
Deflate Head | Command to deflate head section active | Blue DS29=1, Green DS3=0, Red DS4=1, Amber DS1=2, DS5=0 | info |
Deflate Heel | Command to deflate heel section active | Blue DS29=1, Green DS3=1, Red DS4=1, Amber DS1=2, DS5=0 | info |
Deflate Left Turn Assist | Left turn assist section deflation active | Blue DS29=0, Green DS3=1, Red DS4=2, Amber DS1=1, DS5=0 | info |
Deflate Right Turn Assist | Right turn assist section deflation active | Blue DS29=0, Green DS3=1, Red DS4=2, Amber DS1=0, DS5=0 | info |
Deflate Seat | Command to deflate seat section active | Blue DS29=1, Green DS3=1, Red DS4=0, Amber DS1=2, DS5=0 | info |
Disconnected Mattress | Mattress unit not detected or disconnected | Blue DS29=0, Green DS3=0, Red DS4=1, Amber DS1=2, DS5=0 | error |
Drive Current Too Low | Motor drive current insufficient | Power supply or drive circuit failure | warning |
EEPROM access error | UCB cannot access EEPROM memory | Blue D108=0, Green D45=0, Red D41=2, Amber D44=2, Service LED=1 | error |
EEPROM data corrupt | UCB detected EEPROM data corruption on startup | Blue D108=0, Green D45=0, Red D41=0, Amber D44=1, Service LED=1 | critical |
EEPROM Error | Memory write/read or data corruption error | Blue DS29=0, Green DS3=0, Red DS4=1, Amber DS1=0, DS5=0 | error |
EEPROM Failure | EEPROM memory failure | Memory read/write error | error |
EEPROM read error | SCB memory read operation failed | Red D47=2, Yellow D49=2, Green D46=1 | error |
EEPROM read error for bed articulation data | UCB cannot read articulation motor configuration from EEPROM | Blue D108=0, Green D45=0, Red D41=2, Amber D44=1, Service LED=1 | error |
EEPROM soft error | Non-critical EEPROM error detected and corrected | Red D47=1, Yellow D49=1, Green D46=1 | warning |
EEPROM Software Diagnostics | Critical EEPROM or software error detected | Blue DS29=2, Green DS3=2, Red DS4=2, Amber DS1=2, DS5=0 | critical |
EEPROM write error | SCB memory write operation failed | Red D47=2, Yellow D49=2, Green D46=0 | error |
Err0 | Scale Error Err0 | Scale display shows Err0 code | error |
Err1 | Scale system error or calibration fault | Scale initialization or calibration failure | error |
Err2 | Scale system error or measurement fault | Scale measurement failure or invalid reading | error |
Err3 | Scale system error or communication fault | Scale communication loss or data error | error |
Err3 | Scale Error Err3 | Scale display shows Err3 code | error |
Err4 | Scale Error Err4 | Scale display shows Err4 code | error |
Err5 | Scale Error Err5 | Scale display shows Err5 code | error |
Flash: Brake OFF Fault | Brake OFF control circuit fault | Brake release control failure | error |
Flash: Brake ON Fault | Brake ON control circuit fault | Brake engagement control failure | error |
Flash: Current Sense Fault | Motor current sensing circuit fault | Current sensor failure or circuit open | error |
Flash: EEPROM Fault | EEPROM memory circuit fault | EEPROM read/write failure | error |
Flash: HPD | High Power Discharge or hardware protection device triggered | Excessive current or temperature detection | error |
Flash: HW Failsafe | Hardware failsafe activated | Hardware safety mechanism triggered | error |
Flash: Low Battery Voltage | Controller battery voltage below acceptable level | Battery voltage too low for controller operation | warning |
Flash: Main Cont FLTS | Main motor contactor fault | Main motor contactor not engaging or disengaging | error |
Flash: Main OFF Fault | Main motor OFF control circuit fault | Motor OFF control signal failure | error |
Flash: Main ON Fault | Main motor ON control circuit fault | Motor ON control signal failure | error |
Flash: Overvoltage | Controller battery voltage exceeds safe operating limit | Battery voltage too high for controller circuits | warning |
Flash: Power Section Fault | Power section circuit fault | Power delivery circuit malfunction | error |
Flash: Precharge Fault | Precharge circuit fault | Capacitor precharge circuit not functioning | error |
Flash: PROC/Wiring Fault | Processor or wiring circuit fault | Processor internal fault or wiring integrity failure | error |
Flash: SPD Limit Pot Fault | Speed limit potentiometer circuit fault | SPD Limit Pot circuit failure or cable fault | error |
Flash: Thermal Cutback | Incubator temperature outside safe operating range or drive train binding | Bed temperature outside 58-103°F or mechanical obstruction in drive | warning |
Flash: Throttle Fault 1 | Incorrect throttle voltage or signal from handle controls | Throttle input voltage outside expected range or cable/junction board fault | error |
Foot Current Overload | Foot motor current exceeded | Motor stall or mechanical jam | warning |
Foot Knee Leg Unexpected Motion | Foot, knee, or leg moving without command | Unintended motor actuation | warning |
Foot overcurrent | Foot articulation motor drawing excessive current | Blue D108=0, Green D45=1, Red D41=1, Amber D44=2, Service LED=0 | error |
Head Angle Error | Head articulation position sensor fault | Blue DS29=1, Green DS3=0, Red DS4=2, Amber DS1=0, DS5=0 | error |
Head Current Overload | Head motor current exceeded | Motor stall or mechanical jam | warning |
Head Knee Hilow Lockout LED Error | Multiple articulation lockout failure | Head, knee, or hilow joint stuck | warning |
Head overcurrent | Head articulation motor drawing excessive current | Blue D108=0, Green D45=1, Red D41=1, Amber D44=1, Service LED=0 | error |
Head Unexpected Motion | Head section moving without command | Unintended motor actuation | warning |
Heartbeat Failure | Network heartbeat signal lost | Communication link failure | error |
Heartbeat failure of the ACB | UCB lost heartbeat signal from air control board | Blue D108=0, Green D45=2, Red D41=1, Amber D44=0, Service LED=1 | critical |
Heartbeat failure of the SCB | UCB lost heartbeat signal from SideCom board | Blue D108=2, Green D45=1, Red D41=2, Amber D44=1, Service LED=1 | critical |
Hilo foot overcurrent | Foot section motor drawing excessive current | Blue D108=0, Green D45=1, Red D41=0, Amber D44=2, Service LED=0 | error |
Hilo head overcurrent | Head section motor drawing excessive current | Blue D108=0, Green D45=1, Red D41=0, Amber D44=1, Service LED=0 | error |
Hilow Foot Current Overload | Hilow foot motor current exceeded | Motor stall or mechanical jam | warning |
Hilow Foot Position Rate High | Hilow foot moving too fast | Position feedback anomaly | warning |
Hilow Foot Position Rate Low | Hilow foot moving too slowly | Position feedback anomaly | warning |
Hilow Foot Unexpected Motion | Foot hilow section moving unexpectedly | Unintended motor actuation | warning |
Hilow Head Current Overload | Hilow head motor current exceeded | Motor stall or mechanical jam | warning |
Hilow Head Position Rate High | Hilow head moving too fast | Position feedback anomaly | warning |
Hilow Head Position Rate Low | Hilow head moving too slowly | Position feedback anomaly | warning |
Hilow Head Unexpected Motion | Head hilow section moving unexpectedly | Unintended motor actuation | warning |
Inflate Head | Command to inflate head section active | Blue DS29=0, Green DS3=1, Red DS4=0, Amber DS1=2, DS5=0 | info |
Inflate Heel | Command to inflate heel section active | Blue DS29=1, Green DS3=0, Red DS4=0, Amber DS1=2, DS5=0 | info |
Inflate Left Turn Assist | Left turn assist section inflation active | Blue DS29=0, Green DS3=0, Red DS4=2, Amber DS1=1, DS5=0 | info |
Inflate Right Turn Assist | Right turn assist section inflation active | Blue DS29=0, Green DS3=0, Red DS4=2, Amber DS1=0, DS5=0 | info |
Inflate Seat | Command to inflate seat section active | Blue DS29=0, Green DS3=1, Red DS4=1, Amber DS1=2, DS5=0 | info |
Knee Current Overload | Knee motor current exceeded | Motor stall or mechanical jam | warning |
Knee overcurrent | Knee section motor drawing excessive current | Blue D108=1, Green D45=1, Red D41=0, Amber D44=1, Service LED=0 | error |
LCB SPI communication lost | UCB cannot communicate with lower control board via SPI | Blue D108=2, Green D45=2, Red D41=0, Amber D44=0, Service LED=1 | error |
Leg Current Overload | Leg motor current exceeded | Motor stall or mechanical jam | warning |
Leg overcurrent | Leg section motor drawing excessive current | Blue D108=0, Green D45=1, Red D41=2, Amber D44=0, Service LED=0 | error |
Logic Circuit Failure | Logic PCB circuit malfunction | Logic board component failure | error |
Motor H-Bridge FET short detection | Motor driver power transistor short circuit detected | Blue D108=1, Green D45=1, Red D41=2, Amber D44=1, Service LED=1 | error |
Motor overvoltage detection | Motor supply voltage exceeds safe limit during operation | Blue D108=1, Green D45=1, Red D41=2, Amber D44=2, Service LED=1 | error |
No error | Normal condition with weight display in 100 lb increments | Blue DS29=x, Green DS3=x, Red DS4=x, Amber DS1=x, DS5=2 | info |
Patient Pendant Failure | Patient pendant not communicating | Wireless link failure or dead battery | warning |
PPM will not arm no scale display | Patient Position Monitor cannot be armed and scale display is blank | PPM activation attempted with scale communication loss | error |
PPM will not arm with scale display | Patient Position Monitor system cannot be armed despite scale operational | PPM activation attempted with scale display present but error condition | warning |
Scale display shows blank or distorted message | Scale display shows no text or garbled/distorted message | Display initialization failure or data corruption | warning |
Scale HB Data | Scale sensor heartbeat data missing | Blue DS29=1, Green DS3=1, Red DS4=2, Amber DS1=0, DS5=0 | warning |
Scale Input Error | Weight sensor input out of range or disconnected | Blue DS29=1, Green DS3=0, Red DS4=2, Amber DS1=1, DS5=0 | error |
SCB General | Scale Control Board general fault or error condition | SCB malfunction detected | error |
SCB Heartbeat | Scale Control Board heartbeat signal missing or failed | SCB communication loss or malfunction | critical |
Solenoid Overvoltage | Solenoid valve drive voltage exceeds limit | Blue DS29=0, Green DS3=1, Red DS4=1, Amber DS1=1, DS5=0 | error |
Solenoid Undervoltage | Solenoid valve drive voltage too low | Blue DS29=1, Green DS3=2, Red DS4=0, Amber DS1=0, DS5=0 | error |
Solid | Service Required Indicator Solid | Service Required indicator continuously illuminated | critical |
Supervisory Overvoltage, Off | Supply voltage exceeds safe limit in off state | Blue DS29=0, Green DS3=1, Red DS4=0, Amber DS1=0, DS5=0 | warning |
Supervisory Overvoltage, On | Supply voltage exceeds safe limit during operation | Blue DS29=0, Green DS3=1, Red DS4=0, Amber DS1=1, DS5=0 | warning |
Supervisory Relay Error | Supervisory relay malfunction | Relay circuit failure | error |
Supervisory Undervoltage | Supply voltage drops below safe minimum | Blue DS29=0, Green DS3=1, Red DS4=1, Amber DS1=0, DS5=0 | error |
System CPR switch failure | Cardiopulmonary resuscitation switch not responding | Blue D108=0, Green D45=0, Red D41=1, Amber D44=2, Service LED=1 | error |
System pendant failure | Patient pendant wireless module not responding | Blue D108=0, Green D45=2, Red D41=1, Amber D44=1, Service LED=1 | error |
UCB Heartbeat | Universal Control Board heartbeat signal missing or failed | UCB communication loss or malfunction | critical |
UCB heartbeat failure | SCB lost communication with UCB | Red D47=2, Yellow D49=1, Green D46=0 | critical |
Valve Overcurrent, Head, Seat, Heel | Solenoid current exceeds limit for mattress sections | Blue DS29=1, Green DS3=1, Red DS4=0, Amber DS1=0, DS5=0 | error |
Valve Overcurrent, Turn Assist | Solenoid current exceeds limit for turn assist function | Blue DS29=1, Green DS3=1, Red DS4=1, Amber DS1=0, DS5=0 | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.