Hill-Rom TranStar PC-100: troubleshooting and error codes
The ServiceForge knowledge base holds 103 documented symptoms for the Hill-Rom TranStar PC-100, with 161 ranked causes, 211 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Stretcher will not enter neutral position
Brake/steer pedal does not lock in neutral; stretcher continues to roll or pedal does not move fully; affects procedural and power stretchers differently
Ranked causes
- 1. Defective casters (common)
- 2. Brake/steer link misaligned (common)
- 3. Bent wireforms in brake activation (occasional)
- 4. Rocker link misaligned or setscrew loose (occasional)
- 5. Caster ratchet noise or defect (occasional)
- 6. Brake/steer pedal interference (rare)
Verify brake/steer pedal locks in neutral
- Test point
- Brake/steer pedal
- Expected
- Pedal locks into neutral position
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Brake does not lock in brake position patient safety
Stretcher brake pedal does not seat firmly; casters roll or rotate when brake engaged
Ranked causes
- 1. Brake/steer link misaligned or loose (common)
- 2. Hex rod does not rotate 45 degrees (common)
- 3. Caster brake shoes misadjusted (common)
- 4. Defective caster (occasional)
- 5. Mechanical interference with brake/steer mechanism (rare)
Brake/steer pedal seats firmly in brake position
- Test point
- Brake pedal
- Expected
- Pedal seats in brake position, all four casters do not roll or rotate
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electric back panel does not raise or lower electrically
Back panel will not respond to electrical control buttons; remains at fixed height
Ranked causes
- 1. Stretcher not connected to external power or power cord damaged (common)
- 2. Nurse control ON/OFF switch not on or LEDs not lit (common)
- 3. Patient control switch not responding (occasional)
- 4. Control board plugs not seated properly (occasional)
Move back panel to mid-height position to verify potentiometer limits not causing issue
- Test point
- Back section articulation
- Expected
- Back panel can be manually cranked to low to mid-height position
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electric thigh panel does not raise or lower electrically
Thigh/knee section will not respond to electrical control buttons; remains at fixed position
Ranked causes
- 1. Stretcher not connected to external power or power cord damaged (common)
- 2. Nurse control ON/OFF switch not on or LEDs not lit (common)
- 3. Patient control switch not responding (occasional)
- 4. Control board plugs not seated properly (occasional)
Verify stretcher connected to external power and inspect cord
- Test point
- Power supply connection and power cord
- Expected
- Stretcher connected to external power supply with undamaged cord
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
OB/GYN foot section does not release correctly
Foot section assembly cannot be released from stretcher or moved into examination position
Ranked causes
- 1. Another part of stretcher interferes with foot section (common)
- 2. Stretcher not in standard position (common)
- 3. Release handle or cable not functioning properly (occasional)
- 4. Defective foot section assembly (rare)
Verify no stretcher parts interfere with foot section
- Test point
- Foot section assembly and adjacent stretcher components
- Expected
- No interference between foot section and other stretcher parts
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Siderail malfunction
Excessive or insufficient force to swing siderail; siderail falls or does not lock in raised position; latch does not engage
Ranked causes
- 1. Swing arm pivot bolt torque incorrect (common)
- 2. Siderail pivot bolts loose (common)
- 3. Siderail latch defective (occasional)
- 4. Siderail interference (rare)
Verify swing arm pivot bolt torque
- Test point
- Four swing arm pivot bolts (A)
- Expected
- Force to swing siderail out from stored position is appropriate
- Tools
- Torque wrench
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Steering Plus fifth wheel does not drop into steer position
Stretcher steering system fails to activate; fifth wheel does not drop when steer position engaged
Ranked causes
- 1. Fifth wheel pivot arms or bracket improperly assembled (common)
- 2. Bent pivot arms or brackets (occasional)
- 3. Fifth wheel worn (occasional)
- 4. Mechanical interference with Steering Plus assembly (rare)
Brake/steer pedal locks in steer position
- Test point
- Brake/steer pedal
- Expected
- Pedal locks into steer position
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Stretcher will not lower
Stretcher stuck in raised position; does not respond to Trendelenburg or Reverse Trendelenburg pedal; hydraulic release plate not fully depressing release pin
Ranked causes
- 1. Release mechanism misadjusted (common)
- 2. Bent or missing wireforms (occasional)
- 3. Defective head hydraulic cylinder (occasional)
- 4. Defective foot hydraulic cylinder (occasional)
Verify release plate fully depresses release pin
- Test point
- Hydraulic release plate (D) and release pin (B) on cylinder (A)
- Expected
- Release plate fully depresses release pin when foot pedal pushed
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Articulating headrest gas spring malfunction patient safety
Gas spring releases and locks inconsistently or headrest binding when push handle release activated
Ranked causes
- 1. Gas spring requires adjustment (common)
- 2. Gas spring requires replacement (occasional)
- 3. Mechlok device or gas spring cable routed incorrectly after internal maintenance (occasional)
Verify gas spring releases and locks throughout full range
- Test point
- Articulating headrest movement
- Expected
- Full range of motion with no binding, push handle engages/disengages completely
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Manual back articulation not functioning
Crank handle does not elevate back section or back section will not hold position
Ranked causes
- 1. Crank handle not attached or not engaging shaft mechanism (common)
- 2. Another part of stretcher interferes with back screw assembly (occasional)
- 3. Defective back screw assembly (rare)
Verify crank handle is attached and engaging shaft mechanism
- Test point
- Crank handle connection to back screw assembly
- Expected
- Crank handle firmly attached and fully engaged with shaft mechanism
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Manual thigh articulation not functioning
Crank handle does not elevate thigh section or thigh section will not hold position
Ranked causes
- 1. Crank handle not attached or not engaging shaft mechanism (common)
- 2. Another part of stretcher interferes with knee screw assembly (occasional)
- 3. Defective knee screw assembly (rare)
Verify crank handle is attached and engaging shaft mechanism
- Test point
- Crank handle connection to knee screw assembly
- Expected
- Crank handle firmly attached and fully engaged with shaft mechanism
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
OB/GYN foot supports will not raise/lower or rotate correctly
Foot support assembly does not respond to release handle activation for raising, lowering, or rotating inward/outward
Ranked causes
- 1. Missing or loose shoulder bolts, locknuts, or gas spring spacer (common)
- 2. Defective foot support gas spring (common)
- 3. Defective foot support Mechlok device (occasional)
Verify shoulder bolts, locknuts, and gas spring spacer are in place
- Test point
- Foot support assembly (Figure 2-12)
- Expected
- All hardware present and properly seated
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Push handle malfunction
Push handle does not fall into position or lock into place when release handle is activated
Ranked causes
- 1. Another part of stretcher interferes with push handle assembly (common)
- 2. Push handle latch defective (common)
- 3. Push handle defective (occasional)
Verify no stretcher parts interfere with push handle
- Test point
- Push handle assembly and adjacent components
- Expected
- No interference preventing push handle movement
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reduced pedal pumping
Stretcher requires excessive pump cycles to reach full height; takes more than 28 cycles (24 for PC-100/PC-200) or head/foot ends rise unevenly
Ranked causes
- 1. Pump linkage shoulder bolt loose (common)
- 2. Pump pedal weldment interference (occasional)
- 3. Defective hydraulic cylinder (occasional)
Verify pump cycles raise stretcher to full height
- Test point
- Stretcher frame at high position
- Expected
- Rises to high position in 28 cycles (24 for PC-100/PC-200)
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Thigh panel will not raise/lower
Procedural stretcher thigh/knee panel stuck; crank handle detached or assembly interferes; knee screw assembly defective
Ranked causes
- 1. Crank handle detached (common)
- 2. Knee screw assembly interference (occasional)
- 3. Defective knee screw assembly (occasional)
Verify crank handle attached to knee screw assembly
- Test point
- Crank handle (B) on knee screw assembly (A)
- Expected
- Crank handle firmly attached
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Back panel does not raise/lower correctly
Procedural/fixed height/trauma/OB/GYN stretcher back panel stuck or does not move smoothly; gas spring weak or missing hardware
Ranked causes
- 1. Missing or damaged panel hardware (occasional)
- 2. Defective back panel gas spring (common)
Verify panel hardware present and secure
- Test point
- Back panel headed pins (A), rivets (B), push nuts (C), retaining rings (D)
- Expected
- All hardware present and secure
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Electric Stretcher Siderail pivot force caution
Siderail storage force requires adjustment after installation or maintenance. Manual procedure to set pivot bolt torque and cable tension.
Ranked causes
- 1. Pivot bolt torque out of specification (common)
- 2. CPR cable tension out of adjustment (common)
Measure force required to store siderails
- Test point
- Siderail storage operation
- Expected
- Force neither excessive nor too easy
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: PC-100/PC-200 Procedural Stretcher Siderail Latch installation caution
Proper torque specification for lower pivot bolt and latch shoulder bolt during siderail latch installation to ensure correct operation.
Ranked causes
- 1. Lower pivot bolt torque specification (common)
- 2. Latch shoulder bolt torque specification (common)
Torque lower pivot bolt to specification
- Test point
- Lower pivot bolt (G)
- Expected
- 140 + 10 in-lb (15.8 + 1.1 N·m), +10 in-lb
- Tools
- Torque wrench 0-250 in-lb
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: PC-300–PC-550 Procedural/Fixed Height/Trauma/Surgical/OB/GYN/Electric Stretcher Siderail Latch installation caution
Proper torque specifications for lower pivot bolt and latch shoulder bolt during siderail latch installation.
Ranked causes
- 1. Lower pivot bolt torque out of specification (common)
- 2. Latch shoulder bolt torque out of specification (common)
Torque lower pivot bolt to specification
- Test point
- Lower pivot bolt (C)
- Expected
- 140 + 10 in-lb (15.8 + 1.1 N·m), +10 in-lb
- Tools
- Torque wrench 0-250 in-lb
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Corner steer caster does not lock in steer position
Stretcher with corner steer option fails to lock designated caster parallel to bed side when steer engaged
Ranked causes
- 1. Corner steer caster defective (common)
- 2. Brake/steer link screw loose (PC-100/PC-200 only) (common)
Brake/steer pedal locks in steer position
- Test point
- Brake/steer pedal
- Expected
- Pedal locks into steer position
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of corner steer
One corner of stretcher does not respond to steering control; rocker arm may be loose or linkage interferes
Ranked causes
- 1. Rocker arm setscrew loose (common)
- 2. Steering linkage interference (occasional)
Verify rocker arm setscrew torque
- Test point
- Rocker arm setscrew (E)
- Expected
- 40 + 8 in-lb (4.6 + 0.9 N·m), +/- 8 in-lb
- Tools
- Torque wrench
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nurse control switch not responsive patient safety
Nurse control switch does not register input commands or individual button functions are inoperative
Ranked causes
- 1. Nurse control switch button stuck or defective (common)
- 2. Nurse control switch wiring disconnected or loose (occasional)
Verify nurse control switch wires are connected
- Test point
- Individual switch wires (C) at nurse control switch (E)
- Expected
- All wires connected per wire diagram, Figure 4-16
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient control switch not responsive patient safety
Patient control switch does not register input commands for siderail height adjustment
Ranked causes
- 1. Patient control switch adhesive failure or misalignment (occasional)
- 2. Patient control switch connector loose or disconnected (common)
Verify patient control switch connector is seated
- Test point
- Patient control switch connector
- Expected
- Connector fully seated and locked
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Siderail cable connector not properly seated
Wiring connector at patient control switch or board connections loose or disconnected, preventing siderail operation
Ranked causes
- 1. Left siderail connector not connected to P2 on control board (common)
- 2. Right siderail connector not connected to P3 on control board (common)
Verify left siderail connector seated at P2
- Test point
- P2 connector on control board (H)
- Expected
- Left siderail wiring assembly connector fully seated
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Siderail does not move smoothly or operate patient safety
Siderail cable assembly misalignment or improper pivot installation causes binding or failure to cycle between stored and up positions
Ranked causes
- 1. Lower pivot bolt not properly torqued or damaged threads (occasional)
- 2. Misaligned roller guide or wave washers (occasional)
Verify alignment of roller guide, wave washers, and upright tube
- Test point
- Bolt holes in bottom rail
- Expected
- All components aligned with bolt holes
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Siderails bind or do not lock properly patient safety
Siderails raise or lower with difficulty, excessive force required, or latch does not lock into position
Ranked causes
- 1. Siderail track misalignment or debris (common)
- 2. Latch mechanism failure (occasional)
Raise and lower siderails and verify latch operation
- Test point
- Siderails and latch mechanism
- Expected
- Siderails raise and lower without binding; excessive force not required; latch locks into position
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Steering function malfunction
Stretcher steering pedal does not operate correctly or steering position not achieved
Ranked causes
- 1. Steering Plus system malfunction (unknown)
- 2. Corner steer system malfunction (unknown)
Verify stretcher in steer position with Steering Plus system
- Test point
- Steering Plus pedal
- Expected
- Fifth wheel drops into position
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Stretcher does not lower evenly when release pedal is pressed caution
Head and foot ends lower at different rates or stretcher lowers unevenly from high position
Ranked causes
- 1. Hydraulic valve malfunction or uneven fluid flow (common)
- 2. Cylinder seal leakage or damage (occasional)
Verify even lowering of head and foot ends when release pedal pressed
- Test point
- Right-side release pedal and stretcher ends
- Expected
- Head and foot ends lower evenly without uneven settling
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Stretcher does not rise to high position with pump pedal caution
Stretcher requires more than specified pump strokes or does not reach high position when pump pedal is pressed
Ranked causes
- 1. Hydraulic pump failure or low fluid level (common)
- 2. Hydraulic hose or valve blockage or leak (occasional)
Count pump strokes required to reach high position with 40 pound weight at head end
- Test point
- Right-side pump pedal
- Expected
- No more than 28 full pumps (24 for PC-100/PC-200) required; stretcher rises evenly
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Surgical stretcher back section does not respond to release handle or raise correctly
Back section does not lock at 10-15 degree intervals, or does not rise when release handle is activated
Ranked causes
- 1. Back section gas spring requires adjustment (common)
- 2. Defective back section gas spring (occasional)
Test gas spring locking and release handle engagement throughout motion range
- Test point
- Back section gas spring and release handle
- Expected
- Gas spring releases and locks every 10-15 degrees throughout full range; release handle engages/disengages midway with no binding
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Trauma stretcher cassette lift handle malfunction
Cassette lift handle does not operate or is obstructed
Ranked causes
- 1. Cassette lift cable and spring not properly attached (common)
- 2. Cassette lift mechanism obstructed (common)
Verify cable and spring are properly attached
- Test point
- Cassette lift cable and spring assembly
- Expected
- Cable and spring firmly attached with vinyl cap on spring open end
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Trauma stretcher foot section malfunction
Foot section is not attached, obstructed, or unable to move freely
Ranked causes
- 1. Foot section not attached to upper frame weldment pivot brackets (common)
- 2. Foot section obstructed or obstruction on frame (common)
Verify foot section is attached to pivot brackets
- Test point
- Foot section attachment to upper frame weldment
- Expected
- Foot section firmly attached to upper frame weldment pivot brackets
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Articulating headrest does not move freely or binds patient safety
Articulating headrest section fails to raise/lower or rotate freely forward/backward and up/down, or exhibits binding along either axis.
Ranked causes
- 1. Headrest gas spring or articulation hinge misalignment or wear (unknown)
Test headrest raise and lower motion
- Test point
- Articulating headrest with thumb push levers
- Expected
- Headrest moves freely with no binding along both axes
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Articulating headrest does not release at correct time patient safety
Push handle release does not actuate articulating headrest smoothly; gas spring and Mechlok device do not release simultaneously
Ranked causes
- 1. Gas spring tension misaligned with Mechlok device (common)
Engage push handle release to actuate articulating headrest
- Test point
- Push handle release
- Expected
- Headrest articulates smoothly when handle engaged
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Back panel binding
Back panel does not move smoothly or evenly during raise and lower
Ranked causes
- 1. Back panel mechanism binding or malfunction (unknown)
Test back panel raise and lower operation
- Test point
- Back panel
- Expected
- Back panel moves evenly without binding
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Back panel does not move evenly or binds caution
Back panel raises or lowers with binding or uneven motion
Ranked causes
- 1. Back panel track misalignment or debris (common)
Raise and lower back panel and verify smooth motion
- Test point
- Back panel and guides
- Expected
- Back panel moves evenly without binding
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Back panel release handle not activating at correct pressure patient safety
Release handle activates too early, too late, or with incorrect pressure when squeezing; back panel does not lock or unlock smoothly
Ranked causes
- 1. Gas spring tension incorrect (common)
Raise back panel and apply downward pressure on side requiring adjustment
- Test point
- Back panel side
- Expected
- Back panel moves smoothly
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Back section does not rise to highest position or gas spring binds patient safety
Back section panel fails to reach highest position when raised via release handle, or gas spring exhibits binding or fails to release and lock throughout full range.
Ranked causes
- 1. Back section gas spring or release handle mechanism wear or misalignment (unknown)
Place back section at lowest position
- Test point
- Back section position
- Expected
- Back section at lowest position
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Brake does not hold stretcher in place patient safety
Casters roll or rotate when brake pedal is engaged in brake position
Ranked causes
- 1. Brake pads worn or damaged (common)
Verify brake pedal engagement in brake position
- Test point
- Brake pedal and casters
- Expected
- Pedal seats in position, all four casters do not roll or rotate
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Brake does not hold stretcher or casters roll in brake position patient safety
When brake pedal is engaged, stretcher does not remain stationary or one or more casters continue to roll or rotate.
Ranked causes
- 1. Brake pad wear or brake mechanism malfunction (common)
Engage brake pedal and verify stationary position
- Test point
- Brake pedal and all four casters
- Expected
- Brake pedal seats and all four casters do not roll or rotate
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Brake pads worn or unserviceable caution
Brake pads are not serviceable and require replacement
Ranked causes
- 1. Brake pad wear beyond serviceable limit (unknown)
Visually inspect all brake pads
- Test point
- All brake pads
- Expected
- All brake pads are serviceable
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Brake pedal does not engage patient safety
Brake pedal does not seat in brake position or casters roll or rotate when brake engaged
Ranked causes
- 1. Brake system malfunction (unknown)
Test brake pedal engagement at brake position
- Test point
- Brake pedal
- Expected
- Brake pedal seats in brake position; all four casters do not roll or rotate
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Brake pedal does not seat in brake position patient safety
Brake pedal does not fully seat and casters continue to roll or rotate when brake is engaged
Ranked causes
- 1. Brake mechanism failure or misadjustment (unknown)
Verify Initial Actions have been completed
- Test point
- Troubleshooting procedure step 1
- Expected
- Initial Actions completed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Electric Stretcher Head Drive Assembly removal and replacement high voltage
Complete removal and reinstallation procedure for head drive assembly including gearbox linkage, CPR cable, actuator pin, motor housing, and fastener torque application.
Ranked causes
- 1. Head drive assembly requires removal for service, repair, or replacement (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Electric Stretcher Knee Drive Assembly removal and replacement high voltage
Complete removal and reinstallation procedure for knee drive assembly including actuator pin, retaining rings, motor housing, and fastener torque application.
Ranked causes
- 1. Knee drive assembly requires removal for service, repair, or replacement (unknown)
Verify stretcher is unplugged from power source
- Test point
- Power cord
- Expected
- Disconnected from outlet
17 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Electric Stretcher Potentiometer
Back section does not lower to correct position; gap between back and head sections is outside specification
Ranked causes
- 1. Potentiometer can position out of adjustment (common)
Measure gap between back section and head section support bracket
- Test point
- bottom surface of back section to top of head section support bracket
- Expected
- 1/8" (3.2 mm) to 1/2" (12.7 mm), 3.2 to 12.7 mm
- Tools
- 6" straight edge
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Fixed height stretcher upper frame height adjustment patient safety
Adjust sleep surface height from 28" to 32" in 1" increments by repositioning height adjustment pins and securing with screws
Ranked causes
- 1. Height adjustment required during installation or maintenance (common)
Record current and determine target height configuration
- Test point
- Sleep surface height
- Expected
- 28" (71 cm), 29" (74 cm), 30" (76 cm), 31" (79 cm), or 32" (81 cm) inches, 1 inch increments
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: foot support gas spring installation and operation check caution
Installation and operational verification of foot support gas spring after replacement to ensure proper release lever function and locking
Ranked causes
- 1. Gas spring installed and requires operational verification (common)
Verify gas spring initial installation positioning
- Test point
- rod end hole and base weldment hole alignment
- Expected
- hole in rod end (N) aligns with hole in foot support base weldment (O)
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: PC-100/PC-200 Procedural Stretcher Siderail Tube installation caution
Proper torque specification for lower pivot bolt during siderail tube installation to ensure correct operation and avoid thread damage.
Ranked causes
- 1. Lower pivot bolt torque out of specification (common)
Torque lower pivot bolt to specification
- Test point
- Lower pivot bolt (D)
- Expected
- 140 + 10 in-lb (15.8 + 1.1 N·m), +10 in-lb
- Tools
- Torque wrench 0-250 in-lb
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Siderail pivot bolt torque
Torque specification and adjustment procedure for siderail swing pivot bolts to ensure proper siderail storage force and operation.
Ranked causes
- 1. Torque specification for siderail pivot bolts (common)
Torque swing pivot nuts to specification
- Test point
- Swing pivot nuts (C)
- Expected
- 40 + 6 in-lb, + 6 in-lb
- Tools
- Torque wrench 0-250 in-lb, ½" open end wrench, ratchet, ½" socket
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cassette lift malfunction
Cassette lift handle does not function properly when pulled and rotated
Ranked causes
- 1. Cassette lift mechanism malfunction (unknown)
Test cassette lift handle activation
- Test point
- Cassette lift handle
- Expected
- Cassette lift handle functions properly when pulled and rotated 180°
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Casters do not rotate freely in neutral position caution
Casters bind or do not roll freely when stretcher is in neutral position
Ranked causes
- 1. Pedal does not seat properly in neutral or caster binding (unknown)
Verify neutral position pedal seating and caster rotation
- Test point
- Neutral position pedal and all four casters
- Expected
- Pedal seats in neutral position; all four casters rotate and roll freely
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Corner steer caster does not lock properly caution
Patient left foot end caster does not lock parallel to bed side or other casters do not spin freely in steer position
Ranked causes
- 1. Caster locking mechanism failure or misalignment (unknown)
Verify corner steer caster positioning in steer position
- Test point
- Patient left foot end caster and other casters
- Expected
- Patient left foot end caster locks parallel to bed side; all other casters spin freely
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CPR release does not flatten back section patient safety
CPR release handle does not fully flatten back section to horizontal position
Ranked causes
- 1. CPR release mechanism binding or malfunction (unknown)
Pull and hold CPR release handle
- Test point
- CPR release handle
- Expected
- Back section goes into full flat position; slight downward hand pressure may be needed for empty back section
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electric stretcher left/right siderail cable assembly and upright tube replacement high voltage
Siderail cable assembly and upright tube requires removal and replacement on electric stretcher models
Ranked causes
- 1. Cable assembly or upright tube damaged or requires service (unknown)
16 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electric stretcher siderail removal high voltage
Procedural steps for removing siderails from electric stretchers, including cable disconnection, wire tie removal, and fastener disassembly; two cable assembly versions exist.
Ranked causes
- 1. Electric siderail maintenance or replacement required (occasional)
Power isolation before siderail work
- Test point
- Stretcher power connection
- Expected
- Unit unplugged from power source
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot section binding
Foot section does not move smoothly or evenly during raise and lower
Ranked causes
- 1. Foot section mechanism binding or malfunction (unknown)
Test foot section raise and lower operation
- Test point
- Foot section
- Expected
- Foot section moves evenly without binding
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot section does not release or store properly under gliding surface patient safety
Foot section fails to release when handle is squeezed, or cannot be easily stored under the gliding patient surface.
Ranked causes
- 1. Foot section release handle or storage latch malfunction (unknown)
Test foot section release and storage
- Test point
- Foot section release handle
- Expected
- Foot section releases and stores easily under gliding patient surface
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot support gas spring release lever not functioning properly caution
Foot support push release lever does not release at correct travel point or does not return to original position; may release too early or require excessive pressure
Ranked causes
- 1. Foot support gas spring requires adjustment (common)
Verify foot support push release lever engagement point
- Test point
- foot support push release lever
- Expected
- engages approximately half-way through full travel
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot supports do not move freely or bind along rotation axes patient safety
Foot supports fail to raise/lower or rotate smoothly in forward/backward and up/down directions, or gas spring fails to lock throughout range.
Ranked causes
- 1. Foot support gas spring or hinge mechanism wear or misalignment (unknown)
Test foot support raise and lower motion
- Test point
- Foot support assembly with release and push levers
- Expected
- Foot support moves freely with no binding along both axes
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Head section does not respond to nurse control switch
- Head section does not respond to patient control switch
- Hilow pump does not raise stretcher to high position
- Hilow release pedal does not lower stretcher evenly
- Knee Gatch assembly stiff or misaligned
- Knee Gatch does not move smoothly
- Knee Gatch does not raise or lower evenly
- Knee section does not respond to patient control switch
- Manual head section crank does not articulate back section
- Manual thigh section crank does not articulate knee section
- Neutral position malfunction
- OB/GYN stretcher patient grip handle replacement
- Push handle lock or deployment malfunction
- Push handles do not drop to stored position
- Push handles do not lock in raised position or drop position
- Push handles do not lock into raised position
- Replacement: Base shroud and bellows assembly
- Replacement: OB/GYN stretcher upper frame catch shroud
- Replacement: Procedural/Electric/Trauma/Surgical/OB/GYN stretcher upper frame
- Reverse Trendelenburg does not lower foot end smoothly
- Reverse Trendelenburg foot end does not lower smoothly
- Reverse Trendelenburg function malfunction
- Siderail binding or latch malfunction
- Siderail does not raise or lower smoothly
- Siderail end tube replacement
- Siderail removal and replacement
- Siderail transfer stop/common tube replacement
- Siderails bind, require excessive force, or latch does not lock
- Steering Plus system fails to engage
- Steering system does not engage or fifth wheel does not drop
- Stretcher does not lower evenly
- Stretcher does not lower evenly during release operation
- Stretcher does not move freely in neutral position
- Stretcher does not raise to high position
- Stretcher does not rise evenly during pump operation
- Stretcher neutral position pedal does not seat
- Surgical brake steer pedal does not rotate freely
- Surgical brake/steer pedal malfunction
- Surgical stretcher release handle activation incorrect
- Switch siderail to opposite side
- Trendelenburg does not lower head end smoothly
- Trendelenburg function malfunction
- Trendelenburg head end does not lower smoothly
43 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI