Hill-Rom Retractable Bed: troubleshooting and error codes
The ServiceForge knowledge base holds 45 documented symptoms for the Hill-Rom Retractable Bed, with 141 ranked causes, 206 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Bed air surface does not operate in Comfort Mode patient safety
Air mattress fails to inflate or deflate when Comfort Mode is active or control switches are pressed; excessive airloss LED may be on at control console
Ranked causes
- 1. Excessive airloss detected by system (common)
- 2. Air system power switch off (common)
- 3. Defective firm switch (occasional)
- 4. Air system P.C. board defect (occasional)
- 5. Inflate manifold defective or hose issues (occasional)
- 6. Solenoid shorted or defective (occasional)
- 7. Sleep surface control board defect (occasional)
- 8. No AC voltage at solenoid J/P 29 (occasional)
Excessive airloss LED status at control console
- Test point
- Control console indicator
- Expected
- LED on or off
17 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed does not enter Reverse Trendelenburg
Bed will not go into Reverse Trendelenburg position when the switch is activated on P8500 models.
Ranked causes
- 1. Reverse Trendelenburg switch or cable faulty (common)
- 2. Bed down limit switch or cable faulty (common)
- 3. Latch limit switch or cable faulty (common)
- 4. Head down limit switch or cable faulty (occasional)
- 5. Knee down limit switch or cable faulty (occasional)
- 6. Control board defective (occasional)
- 7. Hilow motor or motor cable faulty (rare)
Measure voltage at Reverse Trendelenburg switch circuit
- Test point
- J/P 2 pin 4 to pin 11
- Expected
- approximately 12V DC
- Tools
- VOM (voltmeter)
6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed does not lower when hilow down switch activated
The bed will not lower when the hilow down switch is activated by the operator.
Ranked causes
- 1. Hilow down switch or cable defective (common)
- 2. Cable between J/P 200 and control board damaged (common)
- 3. Limit switch or cable to limit switch defective (occasional)
- 4. Lockout switch or cable to lockout switch defective (occasional)
- 5. Control board defective (occasional)
- 6. Cable between control board and J/P 11 damaged (occasional)
- 7. Hilow motor defective (rare)
Measure voltage at J/P 200 when hilow down switch activated
- Test point
- J/P 200 pin 21 (black) to pin 10 (red)
- Expected
- approximately 12V DC
- Tools
- VOM set to V DC
6 more checks with pass and fail actions, 11 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed does not raise when hilow up switch activated patient safety
Hilow up switch pressed but bed fails to raise
Ranked causes
- 1. Hilow up switch or cable faulty (common)
- 2. Cable between J/P 200 and control board faulty (common)
- 3. Limit switch or its cable faulty (occasional)
- 4. Lockout switch or its cable faulty (occasional)
- 5. Control board faulty (occasional)
- 6. Cable between control board and J/P 11 faulty (occasional)
- 7. Hilow motor faulty (rare)
Measure voltage at J/P 200 when hilow up switch activated
- Test point
- J/P 200 pin 21 (black probe) to pin 11 (red probe)
- Expected
- approximately 12V DC
- Tools
- VOM set to V DC
6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed will not enter Trendelenburg position patient safety
Trendelenburg switch activated but bed does not tilt head-down; bed remains in horizontal position
Ranked causes
- 1. Trendelenburg switch malfunction or cable defect (common)
- 2. Bed up limit switch malfunction or cable defect (common)
- 3. Hook limit switch malfunction or cable defect (common)
- 4. Head down limit switch malfunction or cable defect (occasional)
- 5. Knee down limit switch malfunction or cable defect (occasional)
- 6. Control board malfunction (occasional)
- 7. HiLow motor malfunction or cable defect (rare)
Measure voltage at Trendelenburg switch output
- Test point
- J/P 2, pin 4 (black) to pin 13 (red)
- Expected
- Approximately 12V DC when switch activated
6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head section will not lower
Head section does not descend when head down switch is activated
Ranked causes
- 1. Head down switch or control cable defective (common)
- 2. Cable between J/P 200 and control board open or damaged (occasional)
- 3. Limit switch or limit switch cable defective (occasional)
- 4. Lockout switch or lockout switch cable defective (occasional)
- 5. Control board defective (occasional)
- 6. Cable between control board and J/P 11 open or damaged (rare)
- 7. Head motor defective (rare)
Measure voltage at J/P 200 with head down switch activated
- Test point
- J/P 200 pin 21 (black) to pin 8 (red)
- Expected
- approximately 12V DC
- Tools
- VOM set to V DC
6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head section will not raise patient safety
Head up switch activated but head section does not move upward
Ranked causes
- 1. Defective head up switch or cable (common)
- 2. Defective cable between J/P 200 and control board (common)
- 3. Defective limit switch or cable to limit switch (occasional)
- 4. Defective lockout switch or cable to lockout switch (occasional)
- 5. Defective control board (occasional)
- 6. Defective cable between control board and J/P 1 (rare)
- 7. Defective head motor (rare)
Measure voltage at J/P200 pins 21 and 9 while head up switch is activated
- Test point
- J/P200 pin 21 (black probe) and pin 9 (red probe)
- Expected
- approximately 12V DC
- Tools
- VOM set to measure V DC
6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Night light does not work
Hospital bed night light fails to illuminate when activated
Ranked causes
- 1. Night light bulb burned out (common)
- 2. Night light switch is off (common)
- 3. Photocell defective or needs adjustment (occasional)
- 4. Night light sensor adjustment out of range (occasional)
- 5. No AC power at night light board (occasional)
- 6. Night light sensor or sensor cable defective (occasional)
- 7. Night light board defective (occasional)
Inspect bulb for signs of failure
- Test point
- Night light bulb
- Expected
- Bulb intact and not darkened; continuity present if continuity tested
5 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reverse Trendelenburg not working patient safety
Bed will not go into Reverse Trendelenburg when switch is activated on P8400 models
Ranked causes
- 1. Reverse Trendelenburg switch defective or cable disconnected (common)
- 2. Bed up limit switch defective or cable disconnected (occasional)
- 3. Latch limit switch defective or cable disconnected (occasional)
- 4. Head down limit switch defective or cable disconnected (occasional)
- 5. Knee down limit switch defective or cable disconnected (occasional)
- 6. Control board defective (occasional)
- 7. Hi-low motor or cable defective (rare)
Measure voltage at J/P2 between pin 4 and pin 11 while activating Reverse Trendelenburg switch
- Test point
- J/P2 pin 4 (black), pin 11 (red)
- Expected
- approximately 12V DC
5 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Knee section fails to lower
Knee section will not lower when the head down switch is activated
Ranked causes
- 1. Testport or cable between testport and control board faulty (common)
- 2. Knee limit switch or its cable faulty (common)
- 3. Automatic contour down limit switch or its cable faulty (occasional)
- 4. Lockout switch or its cable faulty (occasional)
- 5. Control board (A) faulty (occasional)
- 6. Knee motor faulty (rare)
Measure voltage at J/P3 pins 2 and 10 while head down switch activated
- Test point
- J/P3 pin 2 (black probe), pin 10 (red probe)
- Expected
- approximately 12V DC
- Tools
- VOM set to measure V DC
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Knee section fails to raise patient safety
Knee section will not raise when the head up switch is activated.
Ranked causes
- 1. Testport or cable between testport and control board faulty (common)
- 2. Knee limit switch or cable to limit switch faulty (common)
- 3. Automatic contour up limit switch or cable faulty (occasional)
- 4. Lockout switch or cable to lockout switch faulty (occasional)
- 5. Control board faulty (occasional)
- 6. Knee motor faulty (rare)
Measure voltage at J/P 3 with head up switch activated
- Test point
- J/P 3, pin 2 (black) to pin 9 (red)
- Expected
- approximately 12V DC
- Tools
- VOM
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Knee section will not lower
Knee section does not lower when knee down switch is activated.
Ranked causes
- 1. Knee down switch or cable failure (common)
- 2. Cable between J/P 200 and control board open or shorted (occasional)
- 3. Limit switch or cable failure (occasional)
- 4. Lockout switch or cable failure (occasional)
- 5. Control board failure (rare)
- 6. Knee motor failure (rare)
Measure voltage at J/P 200 when knee down switch activated
- Test point
- J/P 200, pin 21 (black) to pin 6 (red)
- Expected
- approximately 12V DC
- Tools
- VOM set to measure V DC
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Knee section will not raise
Knee section (A) does not raise when knee up switch is activated.
Ranked causes
- 1. Knee up switch or cable faulty (common)
- 2. Cable between J/P 200 and control board faulty (occasional)
- 3. Limit switch or cable to limit switch faulty (occasional)
- 4. Lockout switch or cable to lockout switch faulty (occasional)
- 5. Control board faulty (occasional)
- 6. Knee motor faulty (rare)
Measure voltage at J/P 200 with knee up switch activated
- Test point
- J/P 200 pin 21 (black) to pin 7 (red)
- Expected
- approximately 12V DC
- Tools
- VOM set to V DC
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed does not exit Reverse Trendelenburg when bed up switch activated patient safety
P8400 models only. Bed remains in Reverse Trendelenburg position despite activating the bed up switch.
Ranked causes
- 1. Bed up switch or cable between switch and control board faulty (common)
- 2. Bed up limit switch or cable between switch and control board faulty (common)
- 3. Latch limit switch or cable between switch and control board faulty (common)
- 4. Control board defective (occasional)
- 5. Hilow motor or cable between motor and control board faulty (occasional)
Measure voltage at J/P 2 with bed up switch activated
- Test point
- J/P 2 pin 4 (black probe) to pin 12 (red probe)
- Expected
- approximately 12V DC
- Tools
- VOM set to V DC
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed will not come out of Trendelenburg patient safety
Trendelenburg out switch does not function; bed remains in Trendelenburg position when switch activated
Ranked causes
- 1. Bed up switch or cable failure (common)
- 2. Bed up limit switch failure (common)
- 3. Hook limit switch failure (occasional)
- 4. Control board failure (occasional)
- 5. Hilow motor or motor cable failure (occasional)
Measure voltage at J/P 2 when bed up switch activated
- Test point
- J/P 2 pin 4 (black) to pin 12 (red)
- Expected
- approximately 12V DC
- Tools
- VOM
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Function not responding patient safety
A bed function (e.g., siderails, head/foot section) does not operate despite operator input; may occur at control panel, siderail controls, or attendant console
Ranked causes
- 1. Bed not connected to power supply (common)
- 2. Motor power disabled via control console switch (common)
- 3. Function lockout activated (common)
- 4. Function at limit position (occasional)
- 5. Problem isolated to specific area (control point failure) (occasional)
Verify bed is plugged into power source
- Test point
- Power cord to outlet
- Expected
- Bed connected to appropriate power outlet
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reverse Trendelenburg will not lower patient safety
Bed will not exit reverse Trendelenburg position when reverse Trendelenburg down switch activated on P8500 models
Ranked causes
- 1. Bed down switch failure (common)
- 2. Bed down limit switch failure (occasional)
- 3. Latch limit switch failure (occasional)
- 4. Control board failure (occasional)
- 5. Hilow motor or motor cable failure (rare)
Voltage at J/P 2 pins 4 and 10 with bed down switch activated
- Test point
- J/P 2 pins 4 and 10
- Expected
- Approximately 12V DC
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Adjust Trendelenburg engage/disengage switches patient safety
Switch positioning and click tolerance adjustment on P8500 model beds after replacement
Ranked causes
- 1. Switch requires initial positioning after installation (common)
- 2. Hook travel and hilow up limit require prior verification (common)
Verify hook travel and hilow up limit adjustment
- Test point
- Bed hilow mechanism and hook
- Expected
- Proper hook travel and hilow limit per section 4-86 specifications
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Brakes not functioning after caster work patient safety
Brake/steer pedal mechanism not engaging or disengaging properly
Ranked causes
- 1. Brake cams not installed in correct orientation (common)
- 2. Caster adjustment not completed correctly (common)
Visually verify brake cam orientation matches notes from removal
- Test point
- Brake cam assembly (C)
- Expected
- Cams oriented as documented before removal
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Head Drive Screw Assembly—Roller Bearing replacement
Inoperative roller bearing at head end or foot end of drive screw assembly on CPR-equipped 8400/8500 models
Ranked causes
- 1. Inoperative bearing at head end (unknown)
- 2. Inoperative bearing at foot end (unknown)
Confirm head drive screw assembly functions properly after replacement
- Test point
- Head drive screw assembly
- Expected
- Operates correctly without binding or noise
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hilow limit switch inoperative or damaged caution
Bed does not reach proper hilow height limits; switch fails to trigger at correct position; mechanical or electrical failure of hilow up, hilow down, or Trendelenburg limit switches
Ranked causes
- 1. Hilow up or hilow down limit switch defective (common)
- 2. Trendelenburg or Reverse Trendelenburg limit switch defective (occasional)
Visually identify which limit switch is inoperative
- Test point
- Hilow up, hilow down, or Trendelenburg switch location per Figure 4-26
- Expected
- Switch intact and properly connected
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Trendelenburg limit switch inoperative caution
Trendelenburg or Reverse Trendelenburg limit switch not functioning, requiring replacement on P8500 models
Ranked causes
- 1. Damaged Trendelenburg or Reverse Trendelenburg limit switch (unknown)
- 2. Damaged hilow up or hilow down limit switch (unknown)
Mark the relative position of the holder containing the damaged switch
- Test point
- Holder position
- Expected
- Position marked for correct limit adjustment during reassembly
10 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Adjust Reverse Trendelenburg disengage switch patient safety
Positioning adjustment for Reverse Trendelenburg disengage switch to click point and tolerance
Ranked causes
- 1. Reverse Trendelenburg disengage switch requires positioning after replacement (common)
Position Reverse Trendelenburg disengage switch
- Test point
- Reverse Trendelenburg disengage switch slider
- Expected
- Switch clicks at specified dimension per figure 4-33
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Brake light not functioning
Brake light on nurse control panel does not illuminate when brake is engaged (P8500 models only)
Ranked causes
- 1. Inoperative brake light switch (common)
Verify brake light operates after replacement
- Test point
- Nurse control panel brake light
- Expected
- Brake light illuminates when brake engaged
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Air P.C. Board replacement
Removal and replacement of air P.C. board in manifold control box assembly
Ranked causes
- 1. Air P.C. Board requires replacement (unknown)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Caster adjustment caution
Brake/steer and brake casters require adjustment for proper locking base operation and smooth swivel movement.
Ranked causes
- 1. Caster setscrews out of adjustment (common)
Remove plastic leg covers (A) over brake/steer and brake casters
- Test point
- Left foot end and right head end casters
- Expected
- Covers removed, casters and setscrews accessible
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Control Board Assembly removal high voltage
Step-by-step disassembly procedure to access and remove the control board assembly power supply board from P8400 model beds
Ranked causes
- 1. Control board assembly requires removal for service or replacement (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Control Board Assembly replacement high voltage
Procedure to remove and replace the control board assembly (power supply board) on P8400 model beds
Ranked causes
- 1. Control board assembly requires removal or replacement (unknown)
Remove footboard control panel from bed
- Test point
- Footboard control panel
- Expected
- Panel removed
9 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Head drive screw slippage correction patient safety
Head section of bed slips down when elevated due to excessive grease accumulation on head drive screw brake assembly
Ranked causes
- 1. Excessive grease on head drive screw brake assembly (common)
Remove roll pin (A) from head end of head drive screw assembly
- Test point
- Head end of head drive screw assembly
- Expected
- Roll pin removed
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Head limit switch adjustment
Head section does not stop at proper positions or motor does not shut off correctly when head reaches limits
Ranked causes
- 1. Head limit switch misaligned or incorrectly positioned (common)
Run knee section to high position
- Test point
- Knee section control
- Expected
- Knee section reaches high position
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Hilow drive screw assembly slippage correction caution
Excessive grease accumulates on hilow drive screw brake assembly, causing hilow section to slip down when bed is not in low position
Ranked causes
- 1. Excessive grease on hilow drive screw brake assembly (common)
Remove roll pin from foot end of hilow drive screw assembly
- Test point
- Roll pin (A) at foot end
- Expected
- Pin removed
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Knee limit switch
Adjustment of knee limit switch package after replacement to ensure proper head section function
Ranked causes
- 1. Knee limit switch replacement completed (common)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Knee limit switch adjustment patient safety
Knee release does not reset and latch properly before knee motor shuts off in downward mode
Ranked causes
- 1. Knee limit switch out of adjustment (common)
Loosen the two screws (A) on the left side of the bed
- Test point
- Left side of bed, screws (A)
- Expected
- Screws loosened
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Reverse Trendelenburg disengage switch patient safety
Set switch clearance to specification during replacement or adjustment on P8500 models
Ranked causes
- 1. Switch requires calibration after replacement (common)
Measure clearance at dimension B when switch clicks
- Test point
- Reverse Trendelenburg disengage switch pivot link
- Expected
- 0.343" (0.87122 cm) minimum, minimum
- Tools
- Ruler or calipers
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Reverse Trendelenburg engage switch patient safety
Set switch clearance to specification during replacement or adjustment on P8500 models
Ranked causes
- 1. Switch requires calibration after replacement (common)
Measure clearance at dimension A when switch clicks
- Test point
- Reverse Trendelenburg engage switch pivot link
- Expected
- 0.156" (0.39624 cm) maximum in, maximum
- Tools
- Ruler or calipers
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: SideCom Entertainment Control Assembly service
Removal, replacement, and adjustment of entertainment control switches, speakers, and volume controls in siderail assembly
Ranked causes
- 1. Defective switch, speaker, or volume control requiring replacement (unknown)
Unplug bed from power source
- Test point
- Power outlet
- Expected
- Bed unplugged
8 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Siderail lubrication
Preventive maintenance procedure to ensure smooth siderail operation through proper lubrication
Ranked causes
- 1. Scheduled maintenance or frozen condition (common)
Apply grease to all designated grease ports
- Test point
- Grease ports (A)
- Expected
- Grease applied to each port per maintenance schedule
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: System voltage check high voltage
Validation procedure for system power supplies and control circuit voltage outputs
Ranked causes
- 1. Incorrect measurement technique or instrument setting (common)
Verify VOM is set to DC voltage measurement mode before checking J/P 2
- Test point
- J/P 2
- Expected
- VOM selector on V DC
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Head limit switch down function does not shut off at correct position
Head motor continues to run or shuts off too early/late when head section lowers
Ranked causes
- 1. Head limit switch package positioned incorrectly (common)
Loosen head limit switch mounting screws/nuts (A)
- Test point
- Fasteners at position (A) on head limit switch
- Expected
- Screws/nuts loose enough to allow package sliding
- Tools
- Phillips head screwdriver or 3/8" nut driver
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reverse Trendelenburg diagnosis inconclusive
All troubleshooting steps completed but problem persists after hilow motor replacement.
Ranked causes
- 1. Unknown fault or multiple component failure (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reverse Trendelenburg limit switch replacement caution
Switch is inoperative and requires replacement on P8500 models
Ranked causes
- 1. Reverse Trendelenburg limit switch damaged or inoperative (unknown)
Mark relative position of holder before removal
- Test point
- Holder containing switch
- Expected
- Position marked for correct limit adjustment
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Siderail movement restricted or frozen patient safety
Siderail does not move smoothly when raising to up position or removing from bed; restricted movement at pivot points
Ranked causes
- 1. Insufficient lubrication at pivot points (common)
Position siderail and locate grease ports
- Test point
- Grease ports (A) as shown in figure 4-53
- Expected
- Ports visible and accessible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Siderail nurse call, lighting, or bed control switch assembly replacement
Defective switch in siderail assembly requires removal and replacement with new switch
Ranked causes
- 1. Defective switch in nurse call, lighting, or bed control assembly (unknown)
Verify replacement switch matches defective switch
- Test point
- Defective switch vs. new switch
- Expected
- Same switch type and specifications
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Troubleshooting still unresolved patient safety
After replacement of all identified components, issue persists
Ranked causes
- 1. Undiagnosed hardware fault requiring manufacturer support (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit voltage measurements out of specification high voltage
VCC or V+ voltage does not measure +12V DC at testport cable J/P 200
Ranked causes
- 1. Control board assembly failure (common)
Measure VCC voltage
- Test point
- J/P 200 pins 5 and 21
- Expected
- +12V DC
- Tools
- VOM set to volts DC
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI