Hill-Rom Propaq Encore: troubleshooting and error codes
The ServiceForge knowledge base holds 95 documented symptoms for the Hill-Rom Propaq Encore, with 118 ranked causes, 194 diagnostic checks with test points and expected values, and 30 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Battery does not hold charge
Monitor battery voltage drops rapidly when AC/DC power disconnected; unable to maintain operation on battery alone
Ranked causes
- 1. Battery aged or defective (common)
- 2. Battery charging circuit fault (occasional)
Battery charge test after AC adapter recharge
- Test point
- Monitor: battery voltage after 8-12 hour charge period
- Expected
- Battery maintains voltage and monitor operates for specified runtime on battery alone
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Button does not respond or sticks caution
One or more physical buttons on monitor do not respond to press or remain stuck in pressed position
Ranked causes
- 1. Button mechanism stuck or debris accumulated (occasional)
- 2. Button switch contact failure (rare)
Button function response verification
- Test point
- All front-panel buttons: FREEZE/UNFREEZE, SETUP, ALARMS, SUSPEND, LIMITS, NEXT, MAIN MENU, NIBP START/STOP
- Expected
- Each button generates expected menu/display change per button sequence table
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Calibration: Propaq Encore monitor
Perform factory calibration when monitor performance is questioned or as indicated in Section 1 maintenance schedule. Respiration, temperature, printer, and SpO2 options do not require calibration. Sidestream CO2 has no user-performable calibration.
Ranked causes
- 1. Routine calibration required per maintenance schedule (common)
- 2. Mainstream CO2 circuitry calibration needed (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Recharger Supply high voltage
Adjustment of voltage regulators RP5, RP3, and RP1 on the recharger board to ensure correct battery charging voltage and overvoltage protection limits.
Ranked causes
- 1. Factory calibration or recalibration required (unknown)
- 2. RP3 misadjustment risk (rare)
Verify Encore power switch is off
- Test point
- Power switch
- Expected
- OFF
21 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 pump failure (sidestream)
PUMP FAILURE, SERVICE REQUIRED message displayed (SCO2); sidestream CO2 measurement unavailable
Ranked causes
- 1. Sidestream pneumatic tubing kinked (common)
- 2. Pump failure on SSP board (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 sensor temperature control failure (mainstream)
SENSOR FAILURE, HEATER message displayed; sensor temperature control hardware fault
Ranked causes
- 1. Sensor temperature control hardware failure (common)
- 2. MSP board fault (occasional)
Determine if sensor or board is faulty
CO2 waveform absent (mainstream)
LACK OF WAVEFORM, SENSOR DISABLED message displayed; channel disabled due to no airway adapter output
Ranked causes
- 1. Airway adapter (cuvette) contamination (common)
- 2. Sensor light source failure (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 waveform degraded (mainstream)
DEGRADED WAVEFORM, CHECK ADAPTER message displayed; airway adapter contaminated or light source failure
Ranked causes
- 1. Airway adapter (cuvette) contamination (common)
- 2. Sensor light source failure (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG lead disconnect not detected patient safety
Monitor fails to generate equipment alarm or FAULT message when single or multiple ECG leads are disconnected from simulator
Ranked causes
- 1. ECG lead detection circuit fault (rare)
- 2. Alarm software not functioning (occasional)
Single ECG lead disconnect alarm generation
- Test point
- Monitor display and alarm indicators; disconnect LA lead
- Expected
- EQUIPMENT ALERT / MULTIPLE message displayed; ECG FAULT indicated; ALARM(S) OFF light flashes
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
ECG signal low or missing on oscilloscope display
ECG output voltage is not within expected range during calibration
Ranked causes
- 1. RP6 potentiometer incorrectly adjusted (common)
- 2. U55 short not properly connected (occasional)
Verify RP6 baseline offset voltage
- Test point
- Analog board RP6, measured at Digital board TP4
- Expected
- 0V dc, ±0.1V
- Tools
- Oscilloscope (DC-coupled), reference to TP4
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG waveform amplitude not correct
Displayed ECG peak-to-peak amplitude deviates from expected 10±1 mm with normal sinus rhythm simulator input
Ranked causes
- 1. ECG channel gain or front-end amplifier fault (occasional)
- 2. Patient simulator ECG output incorrect (occasional)
Verify patient simulator ECG configuration
- Test point
- Patient simulator controls
- Expected
- ECG Size: 1 mV amplitude; ECG Rate: 80 bpm; Waveform: normal sinus rhythm
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Expansion module rear panel reassembly and closure caution
Procedures for reassembling expansion module rear panel with proper gasket placement, torque specifications, and spacer management.
Ranked causes
- 1. Expansion module rear panel reassembly (common)
- 2. Improper spacer placement or gasket positioning (occasional)
Verify all gaskets are positioned correctly
- Test point
- Rubber gaskets around rear panel perimeter
- Expected
- All gaskets in proper position shown in figure
4 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Expansion module rear panel removal and reinstallation caution
Procedures for opening and closing expansion module rear panel with gasket and connector care, including spacer management.
Ranked causes
- 1. Expansion module service or component replacement (unknown)
- 2. Gasket seal degradation or spacer loss (occasional)
Verify all four rear panel fasteners are removed
- Test point
- Rear panel mounting screws
- Expected
- Four screws removed; spacers retained if present
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Green LED charging indicator does not turn on when AC adapter connected
Green LED on monitor's right side panel fails to illuminate when AC power adapter is plugged in and switched on
Ranked causes
- 1. AC power adapter not switched on (common)
- 2. AC power adapter not supplying power (occasional)
Verify AC power adapter power switch is on
- Test point
- AC power adapter power switch
- Expected
- Switch in ON position
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heart rate display incorrect
Monitor heart rate display does not match simulator setting or is outside tolerance band
Ranked causes
- 1. ECG signal processing or rate calculation error (occasional)
- 2. Patient simulator ECG rate incorrect (occasional)
Patient simulator ECG rate setting
- Test point
- Patient simulator controls
- Expected
- ECG Rate: 80 bpm
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Left side panel requires replacement caution
Procedures for removing and installing left side panel assembly with NIBP cuff hose connector and tubing.
Ranked causes
- 1. Left side panel damage or wear (unknown)
- 2. Adhesive gasket deterioration (occasional)
Verify left side panel clamp is removed
- Test point
- Left side panel U-shaped clamp
- Expected
- Clamp fully removed from side panel
1 more check with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor does not charge from DC power supply caution
Green LED charging indicator does not illuminate when DC power supply is connected and turned on
Ranked causes
- 1. Incorrect polarity of DC power supply connection (common)
- 2. DC power supply voltage out of operating range (occasional)
Verify DC power supply polarity
- Test point
- DC input connector on monitor right side panel
- Expected
- RIBBED lead connected to (+); SMOOTH lead connected to (-)
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No audio tone with QRS event
Soft beep tone does not sound with each detected QRS complex during normal ECG operation
Ranked causes
- 1. Audio output circuit or speaker fault (occasional)
- 2. Audio alarms disabled in software (common)
Verify audible vital signs alarms enabled
- Test point
- Monitor menu: Audible vital signs alarm setting
- Expected
- Alarms enabled (ON)
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recharger board connector or cable failure
Recharger board cables P2, P3, P12, P14, P15 may be disconnected or faulty, preventing proper charging and NIBP function.
Ranked causes
- 1. Recharger board cables disconnected or loose (occasional)
- 2. Recharger board removal or installation error (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Respiration rate display incorrect
Monitor respiration rate display does not match simulator setting or is outside tolerance band
Ranked causes
- 1. Respiration signal processing or rate calculation error (occasional)
- 2. Patient simulator respiration rate incorrect (occasional)
Patient simulator respiration rate setting
- Test point
- Patient simulator controls
- Expected
- Resp Rate: 40 br/m
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Respiration waveform amplitude not correct
Displayed respiration peak-to-peak amplitude deviates from expected 12.5±1 mm
Ranked causes
- 1. Respiration channel gain or front-end amplifier fault (occasional)
- 2. Patient simulator respiration output incorrect (occasional)
Verify patient simulator respiration configuration
- Test point
- Patient simulator controls
- Expected
- Resp Lead I, Baseline 1 KΩ, Rate 40 br/m, Amplitude 1.0Ω
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Right side panel requires replacement caution
Procedures for removing and installing right side panel assembly which is sealed with adhesive gasket to rear chassis.
Ranked causes
- 1. Right side panel damage or wear (unknown)
- 2. Adhesive gasket deterioration or contamination (occasional)
Verify right side panel clamps are removed
- Test point
- Side panel clamps
- Expected
- Two clamps removed from right side
1 more check with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Side panel installation and sealing caution
Procedures for installing side panels with adhesive gasket and securing with clamps to ensure proper sealing of rear chassis.
Ranked causes
- 1. New side panel installation required (common)
- 2. Improper side panel seating or clamp fit (occasional)
Verify gasket material is completely removed from rear chassis
- Test point
- Rear chassis mounting surface
- Expected
- All old gasket material removed; clean surface visible
3 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Analog board replacement needed high voltage
Analog board requires removal and replacement due to malfunction
Ranked causes
- 1. Analog board malfunction (unknown)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Analog board requires replacement caution
Analog board malfunction detected; entire monitor must be returned to Welch Allyn for service.
Ranked causes
- 1. Analog board internal failure (unknown)
Verify cable to P6 routing
- Test point
- P6 connector on Analog board
- Expected
- Cable goes behind and below cable to P5, and over cable to P7
10 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery does not charge
Green LED charging indicator on monitor right side panel does not illuminate and battery does not charge when ac power adapter is connected
Ranked causes
- 1. F1 fuse open or failed (common)
Verify ac power adapter functioning properly
- Test point
- ac power adapter output
- Expected
- Adapter operating normally
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery does not retain charge
Battery pack fails to hold charge during normal operation.
Ranked causes
- 1. Battery pack degraded or defective (common)
Check battery pack capacity
- Test point
- battery pack
- Expected
- manufacturer specifications
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cables or hoses pinched during reassembly caution
Internal cables or hoses may be compressed or damaged when closing the monitor casing, potentially causing shorts or reduced airflow.
Ranked causes
- 1. Incorrect cable routing during reassembly (common)
Verify cable P6 routing relative to P5 and P7
- Test point
- Analog board connectors P5, P6, P7
- Expected
- P6 behind/below P5 and over P7
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Display voltage adjustment high voltage
High voltage supply on recharger board requires adjustment to match display module label specification or optimize display contrast and brightness.
Ranked causes
- 1. Display voltage requires calibration following installation or display module replacement (common)
Install display module and reconnect cables
- Test point
- Display module, front and rear chassis connectors
- Expected
- Display installed and all cables reconnected
5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Display voltage adjustment high voltage
Adjust monitor display's high voltage supply on recharger board after display replacement or if contrast between illuminated and non-illuminated pixels affects readability
Ranked causes
- 1. Display voltage requires calibration after display change or contrast degradation (common)
Install replacement display and reconnect all cables
- Test point
- Front and rear chassis connections
- Expected
- Display powered, equipment alert present
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: EMI null adjustment
Adjustment of oscilloscope probe positions and potentiometers to null AC mains and high-frequency noise on analog and digital boards
Ranked causes
- 1. AC mains and high-frequency noise present on analog and digital boards (common)
Connect oscilloscope reference lead and verify noise level
- Test point
- TP43 or TP1 on Analog board
- Expected
- ≤ 15mV peak-to-peak noise, ≤ 15mV
- Tools
- Oscilloscope
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Functional verification patient safety
Verify proper operation of Encore monitor and options following module-level repairs, calibration, or when questioning accuracy or safety of patient functions
Ranked causes
- 1. Software or hardware malfunction following repair or calibration (unknown)
Perform power-up self-test
- Test point
- Monitor startup
- Expected
- Software testing successful, then hardware tests initiated
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Functional verification patient safety
Verify proper operation of Encore monitor and options after module-level repairs, calibration, or when accuracy or safety is questioned
Ranked causes
- 1. Module-level repairs or calibration performed (common)
Perform power-up self-test
- Test point
- Monitor startup
- Expected
- Software integrity checked first, then hardware tests initiated
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Invasive Blood Pressure
Adjustment of IBP simulator inputs and potentiometer to calibrate pressure channel readings to 200 mmHg
Ranked causes
- 1. IBP channels require calibration to match simulator pressure readings (common)
Connect IBP simulator and set initial pressure
- Test point
- P1 connector on Encore left side panel
- Expected
- Simulator connected and set to 0 mmHg
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: MSP Board barometric pressure
MSP board (MCO2) barometric pressure/altimeter requires calibration adjustment to match local conditions. RP1 is a potentiometer for this adjustment.
Ranked causes
- 1. Barometric pressure reading out of specification (unknown)
Display barometric pressure via service menu
- Test point
- SETUP > MORE > MORE > SERVICE > YES > CO2 TEST > MAIN TST
- Expected
- Within 3 mmHg of local barometric pressure
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: MSP Board electromagnetic interference emissions
RP2 is a null adjustment potentiometer for reducing electromagnetic interference emissions. Adjustment procedure uses oscilloscope measurement at test point TP50.
Ranked causes
- 1. EMI emissions require nulling adjustment (unknown)
Connect oscilloscope probe to test point
- Test point
- TP50 on MSP board
- Expected
- Probe tip connected; ground tip open
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 sensor calibration error (mainstream)
SENSOR FAILURE, CALIBRATION ERROR message displayed; sensor malfunction
Ranked causes
- 1. Sensor failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 sensor calibration error (sidestream)
CALIBRATION ERROR-SERVICE REQUIRED message displayed (SCO2); bench measurement failure
Ranked causes
- 1. CO2 bench failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 sensor EEPROM failure (mainstream)
SENSOR FAILURE - EEPROM message displayed; CO2 channel disabled until sensor cable disconnected
Ranked causes
- 1. Sensor EEPROM failure; typically checksum error (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 sensor exhaust line occlusion (sidestream)
OCCLUSION-CHECK EXHAUST PORT/TUBING message displayed (SCO2); exhaust path blocked
Ranked causes
- 1. Exhaust port or tubing occlusion (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 sensor motor drive failure (mainstream)
SENSOR FAILURE, MOTOR DRIVE message displayed; mainstream sensor motor drive hardware failed
Ranked causes
- 1. Mainstream sensor motor drive hardware failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 sensor motor failure (sidestream)
MOTOR FAILURE-SERVICE REQUIRED message displayed (SCO2); bench motor drive failure
Ranked causes
- 1. Motor drive failure in CO2 bench (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 waveform absent (sidestream)
LACK OF WAVEFORM-SERVICE REQUIRED message displayed (SCO2); no CO2 measurement data
Ranked causes
- 1. CO2 measurement chamber contamination or light source failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 waveform degraded (sidestream)
DEGRADED WAVEFORM-SERVICE REQUIRED message displayed (SCO2); measurement chamber or light source failure
Ranked causes
- 1. CO2 measurement chamber contamination or light source failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
DEFIB SYNC output signal out of specification high voltage
DEFIB SYNC connector pulse does not meet voltage and width specifications
Ranked causes
- 1. DEFIB SYNC output circuit fault (unknown)
Measure DEFIB SYNC output pulse
- Test point
- DEFIB SYNC connector
- Expected
- 5 V p-p pulse with 100 ±5 ms width, ±5 ms width
- Tools
- Oscilloscope at 1 V/div
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Defibrillator sync markers not appearing on ECG high voltage
Sync markers do not display on ECG waveform during defibrillator sync operation
Ranked causes
- 1. Defibrillator sync setup or configuration error (occasional)
Verify sync markers displayed on ECG during defibrillator connection
- Test point
- Monitor ECG display
- Expected
- Sync markers visible on ECG waveform
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dielectric breakdown detected during high-voltage test high voltage
Tester indicates dielectric breakdown when 4000V ac rms is applied between ECG leads and DC input connector
Ranked causes
- 1. Internal sheet insulation barriers not properly installed (unknown)
Apply high-voltage test between all ECG leads and DC input connector
- Test point
- All ECG leads on patient cable connected together; DC input connector all pins connected together
- Expected
- No dielectric breakdown V ac rms, N/A
- Tools
- High-voltage tester
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Digital board malfunction patient safety
Monitor requires service for Digital board issues; board cannot be field-repaired
Ranked causes
- 1. Digital board failure (unknown)
Visual inspection of Digital board
- Test point
- Button contact surfaces on underside
- Expected
- Clean and oil-free surfaces
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display contrast inadequate high voltage
Lack of contrast between illuminated and non-illuminated pixels affects display readability following display module change.
Ranked causes
- 1. Display voltage not optimized for contrast after display module replacement (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display pixels missing during pixel test
Display test reveals missing or defective pixels on LCD screen
Ranked causes
- 1. LCD display defect (rare)
Run pixel test on display
- Test point
- Monitor display screen
- Expected
- No missing pixels visible during three CYCLE iterations
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG signal amplitude out of specification high voltage
ECG signal measured at X1000 connector does not meet amplitude specification
Ranked causes
- 1. ECG connector or signal path fault (unknown)
Measure ECG signal amplitude at X1000 connector
- Test point
- ECG X1000 connector
- Expected
- 1 V ±100 mV baseline to peak of R-wave, ±100 mV
- Tools
- Oscilloscope set to 0.2 s/div and 0.5 V/div
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
EMI null adjustment pre-check
Verification of noise level before proceeding with EMI null adjustment procedure
Ranked causes
- 1. Initial noise level must be verified to establish baseline (common)
Connect oscilloscope and measure peak-to-peak noise
- Test point
- TP43 or TP1 on Analog board
- Expected
- ≤ 15mV peak-to-peak, ≤ 15mV
- Tools
- Oscilloscope
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Equipment alarm does not occur when Nellcor sensor disconnected patient safety
No equipment alarm generated when SpO2 sensor disconnected from monitor
Ranked causes
- 1. SpO2 fault detection circuit malfunction (rare)
Confirm equipment alarm when sensor disconnected
- Test point
- Monitor display
- Expected
- Equipment alarm and acknowledgment prompt
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fasteners torqued incorrectly caution
Six screws securing front chassis to rear chassis are over-torqued or under-torqued, potentially causing casing stress, misalignment, or insufficient securement.
Ranked causes
- 1. Fasteners not torqued to specification (common)
Torque six casing screws to specification
- Test point
- All six fasteners securing front to rear chassis
- Expected
- 6.65 in-lbs to 7.35 in-lbs (0.75 N-m to 0.83 N-m), 6.65 to 7.35 in-lbs
- Tools
- Calibrated torque wrench
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Functional verification fails patient safety
Monitor does not pass functional verification or accuracy/safety is questioned
Ranked causes
- 1. Monitor out of specification (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Green LED charging indicator does not illuminate and battery does not charge
The green LED on the monitor's right side panel does not light up, and the monitor battery fails to charge even when the AC power adapter is functioning properly.
Ranked causes
- 1. Fuse F1 open or failed (common)
Visually inspect fuse F1 for continuity or visible damage
- Test point
- F1 fuse location
- Expected
- Fuse element intact and continuous
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
IBP mean pressure reading out of range at 0 mmHg patient safety
Mean pressure does not display as 0 mmHg ±1 mmHg when simulator set to 0 mmHg
Ranked causes
- 1. IBP transducer not properly zeroed (common)
Verify mean pressure at 0 mmHg after settling
- Test point
- IBP P1 mean pressure display
- Expected
- 0 mmHg ±1 mmHg, ±1 mmHg
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
IBP mean pressure reading out of range at 200 mmHg patient safety
Mean pressure does not display as 200 mmHg ±2 mmHg when simulator set to 200 mmHg
Ranked causes
- 1. IBP signal conditioning or transducer linearity fault (occasional)
Verify mean pressure at 200 mmHg after settling
- Test point
- IBP P1 mean pressure display
- Expected
- 200 mmHg ±2 mmHg, ±2 mmHg
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
IBP waveform noise exceeds specification patient safety
Invasive blood pressure waveform display shows noise level greater than 3 mm
Ranked causes
- 1. IBP transducer not zeroed or pressure signal noise (common)
Measure IBP waveform noise at 0 mmHg
- Test point
- IBP P1 waveform display
- Expected
- Noise level ≤3 mm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Locking clips not secured caution
P10 and P11 locking clips on the monitor chassis are not fully engaged or secured, potentially allowing the casing to separate during operation.
Ranked causes
- 1. Locking clips not fully engaged during reassembly (common)
Verify P10 locking clip is secure
- Test point
- Front chassis P10 locking clip
- Expected
- Clip fully engaged and locked
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Masimo SpO2 data not displaying
- Masimo SpO2 saturation reading out of specification
- Monitor displays error messages on power-up
- Monitor does not pass Functional Verification
- Monitor does not power on but battery charging indicator lights
- Monitor does not power on but battery charging indicator lights
- Monitor time or date display incorrect
- Nellcor SpO2 data not displaying
- Nellcor SpO2 saturation reading out of specification
- Neonatal cuff does not inflate to specification
- NIBP calibration reading out of specification at 250 mmHg
- NIBP leakage exceeds specification during pump test
- NIBP measurements inaccurate or system not functioning
- NIBP pressure readings exceed tolerance at calibration points
- NIBP readings out of specification at 0 mmHg
- NIBP system fault
- O-ring gasket pinched during SpO2 module closure
- P1 FAULT transducer not detected message not appearing
- Pacer indicator not displaying
- Pressure transducer disconnected
- Printer does not output data strip on snapshot
- Printer does not start continuous printing
- Printer does not stop when door opened
- Rear chassis O-ring gasket missing or displaced
- Reinstall MSP/SpO2 circuit boards
- Replace MSP/SpO2 circuit boards
- Replace SpO2 circuit boards
- Safety check: Leakage current and dielectric strength
- Safety precautions during calibration
- Settings and time lost after power-on
- SpO2 module does not open
- Suspected faulty cable
- Temperature channel out of calibration
- Temperature probe reading out of specification for YSI 400 series
- Temperature probe reading out of specification for YSI 700 series
35 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
ALARM light flashing | Active patient alarm condition detected and alarm tone sounding | Patient vital sign parameter exceeds high alarm limit or falls below low alarm limit | warning |
ALARM(S) OFF | One or more alarm threshold limits have been disabled or all vital signs alarms suspended | User sets alarm limit to OFF via menu; user suspends all audible alarms via SERVICE menu | warning |
CALIBRATION ERROR-SERVICE REQUIRED | CO2 bench failure has occurred | During CO2 measurement (SCO2) | error |
CALIBRATION ERROR, TEMP DISABLED | Temperature sensor out of calibration | Temperature circuitry fails calibration check | error |
CALIBRATION ERROR, TEMP DISABLED | Temperature circuitry out of calibration | Monitor detects temperature measurement out of calibration | error |
CUFF FAULT: NIBP DISABLED, SERVICE REQUIRED | Single fault in NIBP system | NIBP measurement system detects malfunction | error |
CUFF FAULT: NIBP DISABLED, SERVICE REQUIRED | Non-invasive blood pressure system malfunction | Single fault detected in NIBP circuitry | error |
DEGRADED WAVEFORM, CHECK ADAPTER - (MCO2) | CO2 waveform signal degraded or absent | Poor signal quality from sensor or adapter | warning |
DEGRADED WAVEFORM, CHECK ADAPTER - (MCO2) | CO2 waveform degraded or weak | Reduced signal quality from sensor | warning |
DEGRADED WAVEFORM-SERVICE REQUIRED | Possible contamination or light source failure in CO2 chamber or SSP Board failure | During CO2 measurement (SCO2) | warning |
ECG FAULT | Electrocardiogram signal acquisition or processing error | Loss of ECG signal from one or more leads; inability to detect R-wave or process ECG channel | error |
EQUIPMENT ALERT / MULTIPLE | Multiple equipment faults or lead disconnections detected simultaneously | Loss of two or more ECG leads or simultaneous ECG and RESP signal loss | error |
EQUIPMENT ALERT - MULTIPLE LEAD FAIL | Multiple ECG leads have failed or become disconnected | Detection of missing signals on two or more ECG lead channels | error |
LACK OF WAVEFORM, SENSOR DISABLED | Airway adapter (cuvette) contaminated or light source in sensor faulty | During CO2 measurement (MCO2) | error |
LACK OF WAVEFORM, SENSOR DISABLED - (MCO2) | No CO2 waveform detected, sensor disabled | No signal received from CO2 sensor | error |
LACK OF WAVEFORM-SERVICE REQUIRED | Possible contamination or light source failure in CO2 chamber or SSP Board failure | During CO2 measurement (SCO2) | error |
MOTOR FAILURE-SERVICE REQUIRED | Motor drive in CO2 bench has failed | During CO2 measurement (SCO2) | error |
OCCLUSION-CHECK EXHAUST PORT/TUBING | Exhaust port or tubing occluded; repeated errors indicate pump failure on SSP board | During CO2 measurement (SCO2) | warning |
PROGRAM FAULT: SETTINGS LOST, TIME/DAY RESET | Battery depleted; configuration memory lost | Battery voltage too low at power-up | warning |
PROGRAM FAULT: SETTINGS LOST, TIME/DAY RESET | Drained battery causing loss of stored settings | Monitor powered on with depleted backup battery | warning |
PUMP FAILURE, SERVICE REQUIRED | Sidestream pneumatic tubing kinked or pump failure on SSP board | During CO2 measurement (SCO2) | error |
RESP FAULT | Respiration signal acquisition or processing error | Loss of respiration impedance signal; inability to detect respiration rate or process RESP channel | error |
SENSOR FAILURE, CALIBRATION ERROR | Sensor failure has occurred | During CO2 measurement (MCO2) | error |
SENSOR FAILURE, CALIBRATION ERROR - (MCO2) | Sensor calibration failure | Sensor calibration error detected | error |
SENSOR FAILURE - EEPROM - (MCO2) | Sensor EEPROM failure detected (typically checksum error) | Sensor EEPROM checksum error or failure | error |
SENSOR FAILURE - EEPROM - (MCO2) | CO2 sensor EEPROM memory error | EEPROM checksum error detected | error |
SENSOR FAILURE, HEATER - (MCO2) | CO2 sensor temperature control hardware failure | Sensor temperature control circuit malfunction | error |
SENSOR FAILURE, HEATER - (MCO2) | Sensor temperature control hardware failure | Sensor heater control circuit malfunction | error |
SENSOR FAILURE, MOTOR DRIVE - (MCO2) | Mainstream sensor motor drive hardware failed | Motor drive circuit malfunction | error |
SENSOR FAILURE, MOTOR DRIVE - (MCO2) | CO2 sensor motor drive hardware failure | Mainstream sensor motor drive malfunction | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI