Hill-Rom Progressa: troubleshooting and error codes

The ServiceForge knowledge base holds 15 documented symptoms for the Hill-Rom Progressa, with 23 ranked causes, 51 diagnostic checks with test points and expected values, and 124 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Calibration: Scale EN45501 Class

Periodic calibration of bed scale system using 100 kg standard weight per EN45501 standard to ensure measurement accuracy and repeatability

Ranked causes

  1. 1. Scale requires calibration per EN45501 standard (common)
  2. 2. Scale verification and testing required (common)
  3. 3. Weighing performance verification required (common)
First check

Navigate to Home Screen and access Preferences

Test point
GCI screen
Expected
Preferences icon displayed

12 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Bed Position Sensor patient safety

Bed frame and chair exit position calibration using control panel lockout and beep feedback signals to establish minimum and exit reference positions.

Ranked causes

  1. 1. Bed frame calibration required at minimum position (common)
  2. 2. Chair exit position calibration required at 1.5" to 1.75" from floor (common)
First check

Verify bed frame is level at or near both ends of tube

Test point
Intermediate frame relative to ground
Expected
0° relative to ground

7 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Controller detects overvoltage caution

Controller reports overvoltage fault. Battery or AC supply voltage exceeds normal operating range, which could damage components.

Ranked causes

  1. 1. Excessive supply voltage (rare)
  2. 2. Controller regulator failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Enable switches not responding patient safety

Switches installed in handles at grip do not register depression when force applied. Bed should move forward/backward based on handle direction but does not respond.

Ranked causes

  1. 1. Switch or cable defective (common)
  2. 2. Loose or corroded connector pins at P2, P5, or P6 (occasional)
First check

Measure switch continuity

Test point
Switch cable end
Expected
Closed (continuity) only when switch depressed
Tools
Meter

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Motor wiring faults indicated caution

Controller displays wiring faults or power section faults. Motor may not receive correct control signals or power due to wiring issues, loose connectors, or shorted connections.

Ranked causes

  1. 1. Loose or damaged motor connector (common)
  2. 2. Controller malfunction (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Scale weighing procedure

Procedure to perform patient weighing using the bed scale system after calibration is complete

Ranked causes

  1. 1. Scale requires zeroing and verification before patient weighing (common)
  2. 2. Scale hardware requires post-use sealing (common)
First check

Access scale weighing function

Test point
GCI Home Screen Scale Icon
Expected
Scale function activates

4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Throttle signal out of range caution

Handle throttle does not produce correct voltage output proportional to applied force. Strain element may be damaged or output signal cannot be adjusted to specifications.

Ranked causes

  1. 1. Damaged strain element in handle (common)
  2. 2. Potentiometer R1 adjustment failed (occasional)
First check

Measure zero voltage output at P6.10

Test point
P6.10 (ground ref P6.4)
Expected
2.4-2.6 V DC, +/-0.1 V
Tools
DMM

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery voltage low or absent caution

Controller not receiving correct battery voltage. Battery may be discharged or power connections are loose. When plugged in, AC supply voltage not present or insufficient.

Ranked causes

  1. 1. Battery discharged or defective (common)
First check

Measure battery voltage at P6.8 (plugged in)

Test point
P6.8
Expected
>32 V
Tools
DMM

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: GCI Screen

Touchscreen calibration of the GCI (Graphical Control Interface) display to ensure accurate stylus or touch input response

Ranked causes

  1. 1. GCI screen requires calibration for accurate touch input (occasional)
First check

Verify bed power and height

Test point
Bed electrical connection and height control
Expected
Bed connected to mains, raised to comfortable height

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Mattress Transducer patient safety

Calibration of mattress pressure transducer sensors to ensure accurate surface pressure and patient detection readings

Ranked causes

  1. 1. Mattress transducer requires periodic calibration (common)
First check

Prepare bed for transducer calibration

Test point
Bed occupancy and mattress condition
Expected
Bed unoccupied, mattress clear of any weight or objects

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Footboard Detection Switch removal and replacement high voltage

Removal and replacement of footboard detection switch mounted in foot section cover with two-wire connector.

Ranked causes

  1. 1. Footboard detection switch requires service or replacement (unknown)
First check

Bed height adjustment for technician access

Test point
Bed height control
Expected
Bed raised to comfortable working height

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

GCI Base Board removal and replacement high voltage

Removal and replacement of GCI base board mounted in thigh section with cable disconnection required.

Ranked causes

  1. 1. GCI base board requires service or replacement (unknown)
First check

Power isolation for safety

Test point
Main power and auxiliary outlet
Expected
Bed and auxiliary outlet disconnected from power sources

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Head Section Articulation Mechanism removal and replacement high voltage

Removal and replacement procedure for head section articulation mechanism including gear housing and cassette assembly.

Ranked causes

  1. 1. Head section mechanism requires service or replacement (unknown)
First check

Set brakes and verify bed is safe for service

Test point
Bed brake status
Expected
Brakes set and locked

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Scale Board—NAWI Compliant removal and replacement high voltage

Removal and replacement of NAWI-compliant scale board (EN 45501) with load beam cables and interconnect cable disconnection required.

Ranked causes

  1. 1. Scale board requires service or replacement (unknown)
First check

Power isolation for safety

Test point
Main power and auxiliary outlet
Expected
Bed and auxiliary outlet disconnected from power sources

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Scale Verification Report completion

Completion and documentation of scale performance verification according to EN45501 standard requirements

Ranked causes

  1. 1. Scale verification documentation required per EN45501 (common)
First check

Record bed and environmental information

Test point
Scale Verification Report Form header
Expected
Bed model, serial number, date, time, inspector ID, temperature, humidity, latitude, altitude, gravity constant completed

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
1040Battery voltage lowBattery voltage less than 18Vwarning
1041Battery voltage critical lowBattery voltage less than 2Vcritical
1042Battery voltage highBattery voltage too highcritical
1043Battery charging current over limitCharging current exceeds maximum limitcritical
1044Battery charging current sensor brokenCurrent sensor or ADC malfunctioncritical
1045LCB SPI communication lostSPI communication failure between boardscritical
1046Charging time over limitCharging duration exceeds limitwarning
1047Battery voltage cannot increaseVoltage fails to rise during constant current chargingwarning
1048Temperature out of rangeTemperature outside 10C to 50C operational rangewarning
1049Bed in transportation modeTransportation mode is activeinfo
1080EEPROM writing errorEEPROM write operation failscritical
1081EEPROM read errorEEPROM read operation failscritical
1082EEPROM bed articulation data errorEEPROM read error for articulation datacritical
1083EEPROM critical read compare failedEEPROM data validation failscritical
10C0CPR switch failedCPR switch malfunctioncritical
10C1Pendant failedPendant communication or operation failurecritical
1100Lost CAN communicationCAN bus communication failurecritical
1120Internal errorInternal system errorcritical
1140Motor FET short detectionFET short circuit detectedcritical
1141Motor over voltage detectionMotor voltage exceeds safe levelcritical
1142HiLo head motor over currentHead motor draws excessive currentcritical
1143HiLo foot motor over currentFoot motor draws excessive currentcritical
1144Head motor over current condition detectedMotor current exceeds acceptable limitserror
1145Knee motor over current condition detectedMotor current exceeds acceptable limitserror
1146Foot motor over current condition detectedMotor current exceeds acceptable limitserror
1147Leg motor over current condition detectedMotor current exceeds acceptable limitserror
1148Head Hilo position error detectedHead hilo motor position feedback faulterror
1149Foot Hilo position error detectedFoot hilo motor position feedback faulterror
114AHead position error detectedHead motor position feedback faulterror
114CKnee hardware over current condition detectedHardware current exceeds acceptable limitserror
114DLeg hardware over current condition detectedHardware current exceeds acceptable limitserror
114EHead Hilo hardware over current condition detectedHardware current exceeds acceptable limitserror
114FFoot Hilo hardware over current condition detectedHardware current exceeds acceptable limitserror
1150Head hardware over current condition detectedHardware current exceeds acceptable limitserror
1151Foot hardware over current condition detectedHardware current exceeds acceptable limitserror
1153Knee position error detectedKnee motor position feedback faulterror
1154Foot position error detectedFoot motor position feedback faulterror
1155Flex a foot position error detectedFlexAFoot motor position feedback faulterror
1156Head angle position error detectedHead motor angle position feedback faulterror
1157Bed frame position calibration errorCalibration out of range or failedwarning
1158Bed chair exit position calibration errorChair exit calibration out of range or failedwarning
1180Scale calibration errorScale calibration out of range or failedwarning
1181Scale heartbeat missing for 10 secondsScale communication loss or heartbeat detection failureerror
1182Scale load beam data errorLoad beam signal or data corruption detectederror
1183Scale ADC error detectedAnalog-to-digital conversion failureerror
2080EEPROM writing error on SideCom boardMemory write operation failureerror
2081EEPROM read error on SideCom boardMemory read operation failureerror
2083EEPROM failed critical read compare testData verification check failed during critical readerror
20C4Composer communication error; HITS server communication issueCommunication failure with HITS server for Navicareerror
2100Lost SCB board heartbeat; communication failureSideCom board heartbeat signal losserror
2120Internal error on SideCom boardInternal processing or logic errorerror
3084Air control board EEPROM data corruptionEEPROM read/write failure or triplicate data mismatcherror
30C6MCM valve current exceeds maximumMeasured current above acceptable rangeerror
30C7MCM valve current below minimumMeasured current below acceptable rangeerror
3100Air control board lost CAN communicationCAN bus communication failureerror
3120Internal error on air control boardInternal processing or logic errorerror
31C012 volt supply voltage below acceptable minimumVoltage pulled under 11 +/- 1 voltserror
31C1Head bladder unable to inflate to setpointInflation not reaching target in 10 +/- 1 minuteserror
31C2Seat bladder unable to inflate to setpointInflation not reaching target in 10 +/- 1 minuteserror
31C3Foot bladder unable to inflate to setpointInflation not reaching target in 10 +/- 1 minuteserror
31C4Head bladder unable to deflate to setpointDeflation not reaching target in 10 +/- 1 minuteserror
31C5Seat bladder unable to deflate to setpointDeflation not reaching target in 10 +/- 1 minuteserror
31C6Foot bladder unable to deflate to setpointDeflation not reaching target in 10 +/- 1 minuteserror
31C7RTA or LTA bladder unable to inflate to setpointInflation not reaching target in 10 +/- 1 minuteserror
31C8AA bladder unable to inflate to setpointInflation not reaching target in 10 +/- 1 minuteserror
31C9RTA or LTA bladder unable to deflate to setpointDeflation not reaching target in 10 +/- 1 minuteserror
31CAAA bladder unable to deflate to setpointDeflation not reaching target in 10 +/- 1 minuteserror
31CB12 volt supply voltage exceeds acceptable maximumVoltage pulled over 14.5 +/- 1 voltserror
3200Supervisory line pulled above max when called to disableVoltage pulled over 1.5 +/- 1 volts when disablederror
3201Supervisory line pulled above max when called to enableVoltage pulled over 14.5 +/- 1 volts when enablederror
3202Supervisory line pulled below min when called to enableVoltage pulled under 11 +/- 1 volts when enablederror
3203MCM valve current exceeds maximum countMCM valve current counts exceed MCMSO currenterror
3204Excessive leak detected in support cushionExcessive leak algorithm true for head/seat/foot zoneerror
3205Support cushion valve current exceeds limitsMeasured valve current above acceptable maximumerror
3206Support cushion valve current below minimumMeasured valve current below acceptable minimumerror
3207Valve current TA exceeds maximumMeasured valve current above acceptable maximumerror
3208Support cushion valve current under limitValve current below minimum thresholderror
3209P&V valve current under minimumP&V valve current below P&V current minimum thresholderror
320BP&V valve current over maximumP&V valve current above P&V current maximum thresholderror
4100Lost GUI board heart beat on GCI baseboardGCI baseboard lost communication with GUI boarderror
9003Accelerometer not calibratedAccelerometer calibration missing or invaliderror
9080EEPROM writing error on OIML boardEEPROM write failureerror
9081EEPROM read error on OIML boardEEPROM read failureerror
9083EEPROM failed critical read compare on OIML boardEEPROM checksum or compare failederror
90CFOIML error affects PPM of UCBOIML error detected impacting pneumatic system operationerror
9100Lost OIML board heart beatOIML board communication losserror
9120Internal error on OIML boardInternal OIML board fault detectederror
9180Scale calibration error on standalone scaleScale calibration out of tolerance or corruptederror
9181OIML heartbeat missing for 10 seconds, combined scale diagnosticOIML communication loss for extended perioderror
9182Scale load beam data error on standalone scaleLoad beam data invalid or out of rangeerror
9183Scale ADC error on standalone scaleAnalog-to-digital converter error detectederror
9184Over-current error on standalone scaleOver-current condition detectederror
A080EEPROM writing error on pneumatic control boardEEPROM write failureerror
A081EEPROM read error on pneumatic control boardEEPROM read failureerror
A0C5Pulmonary blower sensor error or faultBlower error detected in pulmonary blower onlyerror
A0C8Pulmonary blower pressure under limitBlower pressure below minimum thresholderror
A0C9Pulmonary blower pressure over limitBlower pressure above maximum thresholderror
A0CAWoundcare blower current under limitBlower current below minimum for woundcare blowererror
A0CBWoundcare blower current over limitBlower current above maximum for woundcare blowererror
A0CCWoundcare blower voltage under limitBlower supply voltage below minimumerror
A0CDWoundcare blower voltage over limitBlower supply voltage above maximumerror
A0CECritical error reported from pneumatic boardCritical fault detected on pneumatic control boardcritical
A100Lost pneumatic board heart beatPneumatic control board communication losserror
A20ASwitching valve error on pneumatic boardSwitching valve fault detected from CAN index 0x2A01 bits 20-22 or 0x2A01.21error
A20CSwitching valve current over limit on pneumatic boardSwitching valve current above maximumerror
A20DSwitching valve current under limit on pneumatic boardSwitching valve current below minimumerror
A20EExternal EEPROM critical data corrupted on pneumatic boardCritical external EEPROM data corruption detectederror
Brake OFF FaultBrake off solenoid or circuit has faultBrake deactivation circuit malfunctionerror
Brake ON FaultBrake on solenoid or circuit has faultBrake activation circuit malfunctionerror
Current Sense FaultMotor current sense circuit malfunctionCurrent measurement circuit unable to read motor currenterror
EEPROM FaultNon-volatile memory circuit has fault or data corruptedEEPROM read/write failureerror
FLASH_SPD_POTIntelliDrive speed limit potentiometer faultSpeed limit pot input invalid or out of rangeerror
FLASH_THERMALIntelliDrive controller thermal cutbackExcessive controller temperature detectederror
FLASH_THROTTLE_1IntelliDrive throttle fault 1Throttle input voltage out of specificationerror
HPDHigh potential difference or overvoltage detectionVoltage threshold exceeded in power supplyerror
HW FailsafeHardware failsafe circuit activatedSafety circuit detected abnormal conditionerror
Low Battery VoltageBattery voltage dropped below acceptable operating levelBattery voltage at P6.8 <32 V when plugged; P4.8 <22 V when unpluggedcritical
Main Cont FLTSMain contactor fault detectedMain relay circuit malfunctionerror
Main OFF FaultMain power off relay or circuit has faultController unable to verify main power off conditionerror
Main ON FaultMain power on relay or circuit has faultController unable to verify main power on conditionerror
OvervoltageSupply voltage exceeds maximum safe operating levelBattery or AC voltage above specificationcritical
Power Section FaultMain power or distribution circuit has faultPower delivery circuit malfunctionerror
Precharge FaultPrecharge circuit fault detectedCharging circuit initialization failureerror
PROC/Wiring FaultProcessor or junction board wiring fault detectedCommunication or wiring circuit malfunctionerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI