Hill-Rom P-8000: troubleshooting and error codes

The ServiceForge knowledge base holds 314 documented symptoms for the Hill-Rom P-8000, with 505 ranked causes, 792 diagnostic checks with test points and expected values, and 72 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

System fault detected

DCB fault indicator shows a fault condition. Multiple fault groups (general and battery charging) can occur independently, with priority from bottom to top when multiple faults present.

Ranked causes

  1. 1. Drive motor thermal shutdown (unknown)
  2. 2. Enable switch fault (unknown)
  3. 3. Invalid steer switch (unknown)
  4. 4. Engage limit switch fault (unknown)
  5. 5. Engage fault (unknown)
  6. 6. DCB to reed switch board loopback error (unknown)
  7. 7. Engage motor thermal fault (unknown)
  8. 8. Disengage fault (unknown)
  9. 9. Gas gauge communication fault (unknown)
  10. 10. EEPROM fault (unknown)

10 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Electric stretcher back section does not raise or lower electrically high voltage

Electric stretcher back section will not move up or down when electrical controls are used

Ranked causes

  1. 1. Stretcher not powered or at potentiometer limit position (common)
  2. 2. Power supply not connected or power cord damaged (common)
  3. 3. Head ON/OFF control not in ON position or indicators not lit (occasional)
  4. 4. Head drive or control board plugs disconnected (occasional)
First check

Reposition back section to mid-height using manual crank

Test point
Back section position
Expected
Back section moved away from potentiometer limit positions

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Electric stretcher back section does not raise or lower electrically high voltage

Back section fails to move up or down when electric controls are operated

Ranked causes

  1. 1. Stretcher not connected to external power supply or power cord damaged (common)
  2. 2. Head ON/OFF control not in ON position or indicators not lit (common)
  3. 3. Head drive not operating in both directions (occasional)
  4. 4. Control board plugs not connected correctly (occasional)
First check

Position crank handle to move back section to low to mid-height

Test point
Back section position
Expected
Back section moved off potentiometer limits to mid-range position

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Knee section does not raise or lower electrically high voltage

Electric knee articulation fails to respond to control inputs on siderails or nurse control

Ranked causes

  1. 1. Power supply disconnected or power cord damaged (common)
  2. 2. Head ON/OFF control not in ON position or indicators not lit (common)
  3. 3. Knee drive not operating in both directions (occasional)
  4. 4. Control board plugs not connected correctly (occasional)
First check

Verify external power supply connection and power cord condition

Test point
Power connection point
Expected
Power cord intact, stretcher connected to external supply

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

OB/GYN foot section does not release correctly caution

Foot section assembly cannot be released from examination position; may be stuck or release mechanism inoperative

Ranked causes

  1. 1. Mechanical interference from other stretcher components (occasional)
  2. 2. Patient platform not in standard position (common)
  3. 3. Release handle or cable assembly misaligned or damaged (occasional)
  4. 4. Foot section assembly worn or damaged (rare)
First check

Check for mechanical interference with foot section

Test point
Foot section assembly and surrounding components
Expected
No obstruction or mechanical interference

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

OB/GYN stretcher foot section does not release correctly caution

Foot section assembly fails to release from patient platform or operates incorrectly.

Ranked causes

  1. 1. Stretcher part interferes with foot section assembly (occasional)
  2. 2. Patient platform not in standard position (occasional)
  3. 3. Release handle or cable mechanism malfunction (common)
  4. 4. Defective foot section assembly (rare)
First check

Check for interference with foot section

Test point
Foot section assembly and surrounding stretcher parts
Expected
No parts interfere with foot section assembly

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

OB/GYN stretcher foot supports do not raise, lower, or turn correctly caution

Foot supports malfunction in raising, lowering, rotation inward or outward

Ranked causes

  1. 1. Missing or loose shoulder bolts, locknuts, or gas spring spacer (common)
  2. 2. Defective foot support gas spring (common)
  3. 3. Foot support does not lock at position or rises without button engaged (occasional)
  4. 4. Defective foot support Mechlok device preventing rotation (occasional)
First check

Inspect hardware installation on foot supports

Test point
Foot support assembly (see figure 2-8)
Expected
Shoulder bolts (A), locknuts (B), and gas spring spacer (C) all installed

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Weight displayed is incorrect patient safety

Scale displays weight value that does not match actual weight on stretcher

Ranked causes

  1. 1. Packing materials or shims not removed (common)
  2. 2. Interference between upper and lower frames (occasional)
  3. 3. Load beam cables connected in wrong order (occasional)
  4. 4. Scale requires calibration or calibration lost (occasional)
First check

Test with calibration weight under ideal conditions

Test point
Scale display
Expected
Weight reading matches 250 lb (113 kg) standard
Tools
250 lb calibration weight

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Weight shown on scale display is not correct

Scale reading is inaccurate after placing calibration weight on stretcher

Ranked causes

  1. 1. Packing materials or shims not removed (common)
  2. 2. Interference between upper and lower frames (occasional)
  3. 3. Load beam cables connected in wrong order (occasional)
  4. 4. Scale requires calibration (occasional)
First check

Apply 250 lb calibration weight and verify reading

Test point
Scale display
Expected
Correct weight shown lb

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Weight shown on scale display is not correct patient safety

Scale displays inaccurate weight readings when patient or calibration weight is placed on stretcher

Ranked causes

  1. 1. Packing materials or shims not removed from stretcher (common)
  2. 2. Interference between upper and lower frames (occasional)
  3. 3. Load beam cables not connected in correct order (occasional)
  4. 4. Scale requires calibration (occasional)
First check

Place 250 lb calibration weight on stretcher and check reading

Test point
Scale display
Expected
Weight shown is correct lb

5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Back section does not articulate manually with crank handle caution

Manual back articulation fails to raise or lower using crank handle mechanism

Ranked causes

  1. 1. Crank handle not attached or not engaging shaft mechanism (common)
  2. 2. Another part of stretcher interferes with back screw assembly (occasional)
  3. 3. Back screw assembly worn or damaged causing loss of position hold (occasional)
First check

Verify crank handle attachment and shaft engagement

Test point
Back screw assembly
Expected
Crank handle firmly attached and engaging shaft mechanism

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Back section does not articulate manually with crank handle caution

Manual crank handle fails to raise or lower back section; may stick, not engage, or section moves but won't stay in position

Ranked causes

  1. 1. Crank handle not attached or not engaging shaft mechanism (common)
  2. 2. Obstruction preventing back screw assembly movement (occasional)
  3. 3. Back screw assembly worn or damaged; does not hold position (occasional)
First check

Verify crank handle attachment and shaft engagement

Test point
Back screw assembly shaft mechanism connection
Expected
Crank handle fully attached and firmly engaged to shaft

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Load beam scale caution

Scale requires calibration after load beam replacement or P.C. board replacement to ensure accurate weight readings

Ranked causes

  1. 1. Load beam replaced (common)
  2. 2. Scale P.C. board replaced (common)
  3. 3. Scale enclosure replaced (occasional)
First check

Raise stretcher to highest position and verify sleep deck is flat

Test point
Sleep deck
Expected
Flat

19 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Controller error: SPD limit pot fault

DS1 LED flashing SPD limit pot fault code pattern on DS14 indicator during speed control operation

Ranked causes

  1. 1. Controller-DCB or DCB-junction board cable fault (common)
  2. 2. Controller fault (occasional)
  3. 3. Junction board fault (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Controller error: SPD limit pot fault

DS1 LED flashing SPD limit pot fault code pattern on DS14 indicator during speed control operation

Ranked causes

  1. 1. Controller-DCB or DCB-junction board cable fault (common)
  2. 2. Controller fault (occasional)
  3. 3. Junction board fault (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Controller error: thermal cutback

DS1 LED flashing thermal cutback code pattern on DS14 indicator during operation

Ranked causes

  1. 1. Operating temperature outside normal range (common)
  2. 2. Mechanical binding or item stuck in drive train or belt (common)
  3. 3. Controller fault (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Controller error: thermal cutback

DS1 LED flashing thermal cutback code pattern on DS14 indicator during operation

Ranked causes

  1. 1. Operating temperature outside normal range (common)
  2. 2. Mechanical binding or item stuck in drive train or belt (common)
  3. 3. Controller fault (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Controller error: throttle fault 1

DS1 LED flashing throttle fault 1 code pattern on DS14 indicator during handle operation

Ranked causes

  1. 1. Incorrect throttle voltage at controller (common)
  2. 2. Junction board, controller-DCB cable, or DCB-junction board cable fault (common)
  3. 3. Controller fault (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Controller error: throttle fault 1

DS1 LED flashing throttle fault 1 code pattern on DS14 indicator during handle operation

Ranked causes

  1. 1. Incorrect throttle voltage at controller (common)
  2. 2. Junction board, controller-DCB cable, or DCB-junction board cable fault (common)
  3. 3. Controller fault (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Current sense fault detected

Controller signals fault in the current sensing circuit

Ranked causes

  1. 1. Disconnected or damaged motor cables (common)
  2. 2. Incorrect battery voltage (occasional)
  3. 3. Defective controller (occasional)
First check

Inspect motor cable connections

Test point
Motor cable connectors
Expected
All cables firmly seated and undamaged

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Current sense fault detected

Controller signals fault in the current sensing circuit

Ranked causes

  1. 1. Disconnected or damaged motor cables (common)
  2. 2. Incorrect battery voltage (occasional)
  3. 3. Defective controller (occasional)
First check

Inspect motor cable connections

Test point
Motor cable connectors
Expected
All cables firmly seated and undamaged

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Electric stretcher back section manual articulation not working

Back section does not articulate smoothly with crank handle or fails to hold position

Ranked causes

  1. 1. Crank handle not attached or not engaging shaft (common)
  2. 2. Stretcher interference with back screw assembly (occasional)
  3. 3. Back screw assembly worn or failed (occasional)
First check

Verify crank handle attachment and engagement with shaft

Test point
Crank handle and back screw assembly shaft
Expected
Crank handle attached and engaging shaft mechanism

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Electric stretcher knee section does not raise or lower electrically high voltage

Knee section articulation fails to operate with electrical controls on electric stretcher

Ranked causes

  1. 1. Power supply disconnected or Head control not in ON position (common)
  2. 2. Knee Up/Down control not responding on siderails (occasional)
  3. 3. Control board plug connections loose or disconnected (occasional)
First check

Verify power supply connection and Head control status

Test point
Power cord and Head ON/OFF control and indicators
Expected
Stretcher connected to external power with intact cord; Head ON/OFF in ON; indicators lit

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Electric stretcher knee section manual articulation not working

Knee section does not articulate smoothly with crank handle or fails to hold position

Ranked causes

  1. 1. Crank handle not attached or not engaging shaft (common)
  2. 2. Stretcher interference with knee screw assembly (occasional)
  3. 3. Knee screw assembly worn or failed (occasional)
First check

Verify crank handle attachment and engagement with shaft

Test point
Crank handle and knee screw assembly shaft
Expected
Crank handle attached and engaging shaft mechanism

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Enable switch not responding

Handle enable switches do not activate stretcher movement

Ranked causes

  1. 1. Enable switch cable disconnected from junction board (common)
  2. 2. Enable switch or cable defective (common)
  3. 3. Junction board or connector fault (occasional)
First check

Verify enable switch connections at junction board

Test point
P2 and P5 connectors
Expected
Both connectors firmly seated

5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Enable switch not responding

Handle enable switches do not activate stretcher movement

Ranked causes

  1. 1. Enable switch cable disconnected from junction board (common)
  2. 2. Enable switch or cable defective (common)
  3. 3. Junction board or connector fault (occasional)
First check

Verify enable switch connections at junction board

Test point
P2 and P5 connectors
Expected
Both connectors firmly seated

5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Knee section does not articulate manually with crank handle caution

Manual crank handle fails to raise or lower knee section; may not engage or section moves but won't hold position

Ranked causes

  1. 1. Crank handle not attached or not engaging shaft mechanism (common)
  2. 2. Obstruction preventing knee screw assembly movement (occasional)
  3. 3. Knee screw assembly worn or damaged; does not hold position (occasional)
First check

Verify crank handle attachment and shaft engagement

Test point
Knee screw assembly shaft mechanism connection
Expected
Crank handle fully attached and firmly engaged to shaft

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Knee section does not articulate manually with crank handle caution

Manual knee articulation fails to raise or lower using crank handle mechanism

Ranked causes

  1. 1. Crank handle not attached or not engaging shaft mechanism (common)
  2. 2. Another part of stretcher interferes with knee screw assembly (occasional)
  3. 3. Knee screw assembly worn or damaged causing loss of position hold (occasional)
First check

Verify crank handle attachment and shaft engagement

Test point
Knee screw assembly
Expected
Crank handle firmly attached and engaging shaft mechanism

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Knee section does not raise or lower

Procedural or surgical stretcher knee section will not move up or down

Ranked causes

  1. 1. Crank handle not attached to knee screw assembly (common)
  2. 2. Interference between stretcher parts and knee screw assembly (occasional)
  3. 3. Knee screw assembly damaged or worn (occasional)
First check

Verify crank handle attachment

Test point
Knee screw assembly
Expected
Crank handle attached

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Knee section does not raise or lower electrically high voltage

Electric knee articulation fails to respond to control inputs on siderails; stretcher powered and controls appear functional

Ranked causes

  1. 1. Power supply disconnected or damaged (common)
  2. 2. Head ON/OFF control not in ON position or indicators not lit (common)
  3. 3. Control plugs loose or disconnected on control board (occasional)
First check

Verify external power connection and cord integrity

Test point
Power inlet connector
Expected
Power cord connected without visible damage

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Knee section will not raise or lower

Procedural or surgical stretcher knee section fails to raise or lower with crank handle operation

Ranked causes

  1. 1. Crank handle not attached to knee screw assembly (common)
  2. 2. Obstruction interfering with knee screw assembly (occasional)
  3. 3. Knee screw assembly defective (occasional)
First check

Verify crank handle is attached to knee screw assembly

Test point
Crank handle and knee screw assembly per figure 2-6 on page 2-23
Expected
Crank handle attached to knee screw assembly

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Main contactor fault detected

Controller signals fault in the main contactor or control circuit

Ranked causes

  1. 1. Disconnected or damaged motor cables (common)
  2. 2. Incorrect battery voltage (occasional)
  3. 3. Defective controller (occasional)
First check

Inspect motor cable connections

Test point
Motor cable connectors
Expected
All cables firmly seated and undamaged

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Main contactor fault detected

Controller signals fault in the main contactor or control circuit

Ranked causes

  1. 1. Disconnected or damaged motor cables (common)
  2. 2. Incorrect battery voltage (occasional)
  3. 3. Defective controller (occasional)
First check

Inspect motor cable connections

Test point
Motor cable connectors
Expected
All cables firmly seated and undamaged

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Main OFF fault detected

Controller signals a fault in the main OFF circuit or power control

Ranked causes

  1. 1. Disconnected or damaged motor cables (common)
  2. 2. Incorrect battery voltage (occasional)
  3. 3. Defective controller (occasional)
First check

Inspect motor cable connections

Test point
Motor cable connectors
Expected
All cables firmly seated and undamaged

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Main OFF fault detected

Controller signals a fault in the main OFF circuit or power control

Ranked causes

  1. 1. Disconnected or damaged motor cables (common)
  2. 2. Incorrect battery voltage (occasional)
  3. 3. Defective controller (occasional)
First check

Inspect motor cable connections

Test point
Motor cable connectors
Expected
All cables firmly seated and undamaged

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Main ON fault detected

Controller signals fault in the main ON circuit or power enable

Ranked causes

  1. 1. Disconnected or damaged motor cables (common)
  2. 2. Incorrect battery voltage (occasional)
  3. 3. Defective controller (occasional)
First check

Inspect motor cable connections

Test point
Motor cable connectors
Expected
All cables firmly seated and undamaged

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Main ON fault detected

Controller signals fault in the main ON circuit or power enable

Ranked causes

  1. 1. Disconnected or damaged motor cables (common)
  2. 2. Incorrect battery voltage (occasional)
  3. 3. Defective controller (occasional)
First check

Inspect motor cable connections

Test point
Motor cable connectors
Expected
All cables firmly seated and undamaged

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

OB/GYN foot supports malfunction during articulation caution

Foot supports fail to raise, lower, or turn inward/outward properly; may be weak, stuck, or loose

Ranked causes

  1. 1. Missing or loose shoulder bolts, locknuts, or gas spring spacer (common)
  2. 2. Defective foot support gas spring; insufficient lift assist (common)
  3. 3. Defective Mechlok device; foot support won't lock at intermediate positions (occasional)
First check

Verify presence and tightness of shoulder bolts, locknuts, spacer

Test point
Foot support fasteners (see Figure 2-8)
Expected
All shoulder bolts (A), locknuts (B), gas spring spacer (C) present and secure

4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

OB/GYN stretcher foot supports do not raise, lower, turn inward or outward correctly caution

Foot support assembly malfunction; does not respond to release button or fails to lock into position.

Ranked causes

  1. 1. Missing shoulder bolts, locknuts, or gas spring spacer (occasional)
  2. 2. Defective foot support gas spring (common)
  3. 3. Defective foot support Mechlok device (occasional)
First check

Verify hardware installation

Test point
Foot support assembly
Expected
Shoulder bolts, locknuts, gas spring spacer all present

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power section fault detected high voltage

Controller signals fault in the power supply or output stage

Ranked causes

  1. 1. Disconnected or damaged motor cables (common)
  2. 2. Incorrect battery voltage (occasional)
  3. 3. Defective controller (occasional)
First check

Inspect motor cable connections

Test point
Motor cable connectors
Expected
All cables firmly seated and undamaged

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power section fault detected high voltage

Controller signals fault in the power supply or output stage

Ranked causes

  1. 1. Disconnected or damaged motor cables (common)
  2. 2. Incorrect battery voltage (occasional)
  3. 3. Defective controller (occasional)
First check

Inspect motor cable connections

Test point
Motor cable connectors
Expected
All cables firmly seated and undamaged

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Procedural or surgical stretcher knee section will not raise or lower

Knee section fails to move up or down when crank handle is operated

Ranked causes

  1. 1. Crank handle not attached to knee screw assembly (common)
  2. 2. Part of stretcher interferes with knee screw assembly (occasional)
  3. 3. Knee screw assembly defective (occasional)
First check

Verify crank handle is attached to knee screw assembly

Test point
Knee screw assembly
Expected
Crank handle firmly attached

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Push handle malfunction

Push handle does not fall into position or lock properly; may interfere with other stretcher components.

Ranked causes

  1. 1. Interference from another stretcher component (common)
  2. 2. Push handle latch mechanism defective (occasional)
  3. 3. Push handle assembly damaged or bent (occasional)
First check

Verify no interference with push handle assembly

Test point
Push handle assembly path
Expected
Clear movement without obstruction from other components

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Scale display does not come on

Scale display screen fails to light up or show any indication

Ranked causes

  1. 1. Battery voltage low (common)
  2. 2. Display P.C. board defective (occasional)
  3. 3. Load beam cable or load beam damaged (occasional)
First check

Measure battery voltage at connector P4

Test point
Connector P4
Expected
Voltage greater than 3.0 VDC

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Scale display does not come on

Scale display screen is not illuminated or powered

Ranked causes

  1. 1. Battery voltage insufficient (≤3.0 VDC) (common)
  2. 2. Display P.C. board defective (occasional)
  3. 3. Load beam cable damaged or disconnected (occasional)
First check

Check battery voltage at connector P4

Test point
Connector P4
Expected
Voltage greater than 3.0 VDC, >
Tools
Multimeter or voltmeter

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Steering Plus steering system does not operate correctly patient safety

Brake/steer pedal locked; fifth wheel drops into position; pivot arms or brackets bent or damaged

Ranked causes

  1. 1. Fifth wheel pivot arms or pivot bracket incorrect assembly (occasional)
  2. 2. Pivot arms or pivot brackets bent (occasional)
  3. 3. Fifth wheel worn or defective (occasional)
First check

Fifth wheel pivot arms and pivot bracket correct assembly

Test point
Fifth wheel assembly
Expected
Pivot arms and brackets correctly assembled

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Steering Plus system does not operate correctly

Fifth wheel fails to drop into position, pivot arms or brackets are bent, or steering system is inoperative

Ranked causes

  1. 1. Fifth wheel pivot arms or brackets assembled incorrectly (common)
  2. 2. Pivot arms or brackets bent (occasional)
  3. 3. Fifth wheel defective (occasional)
First check

Verify brake/steer pedal locks into steer position

Test point
Brake/steer pedal
Expected
Pedal locks into steer position

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Steering Plus system does not operate correctly

Fifth wheel system fails to drop into position or steer properly; pivot arms or brackets may be bent or misassembled.

Ranked causes

  1. 1. Fifth wheel pivot arms or brackets misaligned or bent (occasional)
  2. 2. Fifth wheel assembly damaged or inoperative (occasional)
  3. 3. Fifth wheel assembly failure (rare)
First check

Verify brake/steer pedal locks into steer position

Test point
Brake/steer pedal
Expected
Pedal fully engages steer position

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Stretcher will not go into neutral position patient safety

Brake/steer pedal cannot be returned to neutral; stretcher does not remain stationary.

Ranked causes

  1. 1. Brake/steer link misaligned or screw loose at head or foot end (common)
  2. 2. Defective caster wheels (occasional)
  3. 3. Hex rod not turning from neutral position (occasional)
First check

Verify brake/steer pedal enters neutral position

Test point
Brake/steer pedal
Expected
Pedal goes fully into neutral position

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Stretcher will not go into neutral position

Brake/steer pedal cannot achieve neutral position or stretcher does not roll freely

Ranked causes

  1. 1. Brake/steer link (A) misaligned or screw (A) loose (common)
  2. 2. Defective casters (common)
  3. 3. Hex rod (C) misalignment with brake/steer link (occasional)
First check

Verify brake/steer pedal achieves neutral position

Test point
Brake/steer pedal
Expected
Pedal goes into neutral position

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Stretcher will not lower patient safety

Lowering function inoperative; release mechanism or Trendelenburg/Reverse Trendelenburg pedals not functioning.

Ranked causes

  1. 1. Release mechanism misadjusted (common)
  2. 2. Head end hydraulic cylinder defective (occasional)
  3. 3. Foot end hydraulic cylinder defective (occasional)
First check

Verify release mechanism operation

Test point
Hydraulic release plate (A) and release pin (B)
Expected
Release plate fully lowers release pin when Hilow pedal pressed

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Stretcher will not lower patient safety

Hydraulic release plate does not fully lower release pin; head or foot end does not lower when Trendelenburg or Reverse Trendelenburg pedal pressed

Ranked causes

  1. 1. Release mechanism not adjusted correctly (occasional)
  2. 2. Defective hydraulic cylinder at head end (occasional)
  3. 3. Defective hydraulic cylinder at foot end (occasional)
First check

Hydraulic release plate fully lowers release pin when Hilow pedal pressed

Test point
Release plate and pin
Expected
Release plate fully lowers release pin on cylinder

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Surgical stretcher articulating headrest push handle release/gas spring malfunction

Articulating headrest push handle does not engage/disengage completely or gas spring does not function smoothly

Ranked causes

  1. 1. Gas spring requires adjustment (common)
  2. 2. Gas spring worn or failed (occasional)
  3. 3. Cable routing incorrect after internal maintenance (rare)
First check

Verify gas spring and push handle operation throughout range

Test point
Gas spring and push handle
Expected
Gas spring releases and locks throughout full range; push handle engages and disengages completely; handle returns without binding

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Throttle signal out of range

Handle throttle output voltage not within expected range

Ranked causes

  1. 1. Insufficient supply voltage to junction board (common)
  2. 2. Damaged or defective handle strain element (common)
  3. 3. Potentiometer R1 out of adjustment (occasional)
First check

Measure supply voltage at P6.8

Test point
P6.8
Expected
more than 22 V, >22 V
Tools
DMM

7 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Throttle signal out of range

Handle throttle output voltage not within expected range

Ranked causes

  1. 1. Insufficient supply voltage to junction board (common)
  2. 2. Damaged or defective handle strain element (common)
  3. 3. Potentiometer R1 out of adjustment (occasional)
First check

Measure supply voltage at P6.8

Test point
P6.8
Expected
more than 22 V, >22 V
Tools
DMM

7 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Weight display flashing patient safety

Scale display shows weight value that flashes or blinks on and off

Ranked causes

  1. 1. Weight exceeds scale maximum of 699.9 lb (317.5 kg) (common)
  2. 2. Interference between upper and lower frames (occasional)
  3. 3. Scale requires calibration or scale P.C. board defective (occasional)
First check

Verify total weight on stretcher does not exceed limit

Test point
Scale display and load
Expected
Weight less than 699.9 lb (317.5 kg)
Tools
Scale display

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Weight shown on scale display is flashing

Scale display shows flashing weight reading

Ranked causes

  1. 1. Weight exceeds 699.9 lb maximum capacity (common)
  2. 2. Interference between upper and lower frames (occasional)
  3. 3. Scale PCB board failure (occasional)
First check

Verify weight on stretcher

Test point
Scale display during operation
Expected
Weight not exceeding 699.9 lb

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Active Brake system cable adjustment caution

Cable assembly slack or tension requires adjustment to ensure correct brake operation

Ranked causes

  1. 1. Cable assembly slack after replacement (common)
  2. 2. Cable assembly kinked during installation (occasional)
First check

Position brake/steer pedals in neutral

Test point
Pedal position
Expected
Pedals centered in neutral detent position

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Active Brake system cable assembly replacement caution

Cable assembly requires removal and reinstallation with proper routing and adjustment

Ranked causes

  1. 1. Cable assembly worn or damaged (occasional)
  2. 2. Cable routing incorrect after installation (occasional)
First check

Raise stretcher to high position

Test point
Stretcher height
Expected
Stretcher fully raised; stable on high position

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Active Brake System—Reduced Braking Ability patient safety

Stretcher brakes do not provide full braking force when Active Brake lever is engaged.

Ranked causes

  1. 1. Cable broken or damaged (common)
  2. 2. Cable misalignment or improper tension (occasional)
First check

Squeeze the Active Brake lever and observe hex rod movement

Test point
Active Brake lever and hex rod
Expected
Hex rod turns when lever squeezed

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Active Brake System—Reduced Braking Ability
  • Articulating headrest handle/gas spring malfunction
  • Auto Contour feature does not raise or lower
  • Auto Contour feature with on/off controls does not raise or lower correctly
  • Auto Contour or BackSaver Fowler feature does not raise or lower
  • Auto Contour or BackSaver Fowler feature does not raise or lower correctly
  • Back section does not lower smoothly with Head Down control
  • Back section does not lower with crank handle
  • Back section does not lower with nurse Head Down control
  • Back section does not raise or lower correctly
  • Back section does not rise smoothly
  • Back section does not rise smoothly with Head Up control
  • Back section does not rise with crank handle
  • Back section does not rise with nurse Head Up control
  • Back section gas spring replacement and adjustment
  • Calibration: Drive wheel chain tension
  • Calibration: Scale display configuration
  • Caster brake does not hold or brakes too hard
  • Control board does not control electrical functions
  • Foot supports do not articulate freely
  • Hardware failsafe activated
  • Hardware failsafe activated
  • Headrest does not articulate freely
  • Hydraulic cylinder descent rate too fast or raises too slowly
  • Knee section does not lower smoothly with Knee Down control
  • Knee section does not rise smoothly with Knee Up control
  • Loss of corner steer
  • Loss of corner steer
  • Low battery voltage indicated
  • Low battery voltage indicated
  • Nurse control switch inoperative
  • Over voltage condition detected
  • Over voltage condition detected
  • Precharge fault detected
  • Precharge fault detected
  • Procedural stretcher Auto Contour feature with on/off controls does not raise or lower correctly
  • Procedural stretcher Auto Contour or BackSaver Fowler feature does not raise or lower correctly
  • Push handle malfunction
  • Reduced braking ability
  • Reduced pedal pumps
  • Reduced pedal pumps
  • Scale display flashes Hands Off, CALC, and ---
  • Scale display flashes Hands Off, CALC, then dashes
  • Scale display flashes Hands Off, then CALC, then ----
  • Scale display flickers or shows partial characters
  • Scale display flickers or shows partial characters
  • Siderail cable assembly loose or damaged
  • Siderail does not operate correctly
  • Surgical stretcher articulating headrest gas spring replacement and adjustment
  • Surgical stretcher articulating headrest push handle/gas spring malfunction
  • Surgical stretcher back section gas spring replacement and adjustment
  • Surgical stretcher back section handle/gas spring malfunction
  • Surgical stretcher back section release handle/gas spring malfunction
  • Surgical stretcher back section release handle/gas spring malfunction
  • Trauma stretcher cassette lift handle does not function
  • Trauma stretcher cassette lift handle malfunction
  • Trauma stretcher cassette lift handle malfunction
  • Weight shown on scale display is flashing
  • Active Brake System does not reduce braking
  • Articulating headrest does not move freely
  • Auto Contour feature does not operate correctly (S/N before K series)
  • Auto Contour Off feature does not operate correctly (S/N before K series)
  • Back section does not flatten with CPR release
  • Back section does not lock into position
  • Back section does not lower smoothly
  • Back section does not lower smoothly
  • Back section does not move smoothly
  • Back section does not raise or lower correctly
  • Back section does not rise smoothly
  • BackSaver Fowler feature does not operate correctly
  • Batteries require replacement in battery box
  • Battery current measurement error
  • Battery has zero drive capacity remaining
  • Battery over current condition
  • Battery over voltage condition
  • Battery replacement required
  • Battery shutdown condition
  • Battery temperature fault
  • Battery voltage measurement error
  • Brake does not lock in brake position
  • Brake OFF fault detected
  • Brake OFF fault detected
  • Brake ON fault detected
  • Brake ON fault detected
  • Brake/steer link misalignment
  • Brake/steer pedal misaligned or not centered
  • Calibration: Active Brake cable adjustment
  • Calibration: Articulating headrest gas spring adjustment
  • Calibration: Ball plunger adjustment for cassette support
  • Calibration: Chain replacement timer reset
  • Calibration: Chain replacement timer reset
  • Calibration: Chain tension adjustment
  • Calibration: CPR Release Cable Assembly
  • Calibration: Down force motor assembly replacement
  • Calibration: Drive motor replacement in power drive assembly
  • Calibration: Drive wheel chains and sprockets replacement
  • Calibration: Drive wheel replacement in power drive assembly
  • Calibration: Electric stretcher potentiometer
  • Calibration: Foot-end Trendelenburg pedal adjustment
  • Calibration: Gas spring replacement in power drive assembly
  • Calibration: Head-end hydraulic cylinder reinstallation
  • Calibration: Hydraulic cylinder release mechanism
  • Calibration: Hydraulic cylinder torque and air removal
  • Calibration: Neutral detent adjustment
  • Calibration: Potentiometer adjustment
  • Calibration: Potentiometer gap adjustment
  • Calibration: Power drive assembly reed switch board replacement
  • Calibration: Power drive assembly transistor replacement
  • Calibration: Scale zero and span
  • Calibration: Side pedal cylinder release adjustment
  • Calibration: Siderail latch pivot bolt torque
  • Calibration: Siderail pivot bolt torque (Transport P8005)
  • Calibration: Siderail tube pivot bolt torque (P8000/P8010/P8020/P8040/P8050)
  • Calibration: Steering Plus fifth wheel center steer alignment
  • Calibration: Upper Frame and Base Shroud
  • Cassette lift does not function
  • Cassette lift handles do not function
  • Cassette support weldment does not stay in stored position
  • Caster brake gap out of specification
  • Caster brake shoe gap incorrect
  • Caster brake/steer link misalignment
  • Control board replacement
  • Control board replacement needed
  • Corner steer system does not lock foot-end caster
  • CPR Release Cable Assembly adjustment required
  • CPR Release Cable Assembly requires replacement
  • CPR release does not flatten back section
  • CPR release does not flatten back section
  • Crank handle does not lock into position
  • Crank handle does not spin freely
  • Cylinder release mechanism plunger misalignment or incorrect torque
  • Display does not come on
  • Display flickers or shows partial characters
  • Display shows Hands Off, CALC, and four dashes sequence
  • Drive Control Board (DCB) requires replacement
  • EEPROM fault detected
  • EEPROM fault detected
  • Electric stretcher fuses blown or inoperative
  • Fifth wheel caster misalignment in Steering Plus system
  • Foot section does not release or store
  • Foot section does not release or store
  • Foot support does not move freely
  • Foot support gas spring does not engage/disengage
  • Foot support gas spring does not release
  • Foot support gas spring does not release or lock correctly
  • Foot support gas spring does not release or lock properly
  • Foot support gas spring installation incomplete
  • Foot supports do not move freely
  • Foot-end Trendelenburg pedal not engaging cylinder release pin
  • Fuse replacement
  • Headrest gas spring does not engage/disengage
  • Headrest handle does not engage/disengage
  • Hilow system does not raise stretcher smoothly
  • Hilow system down function does not work
  • HPD fault detected
  • HPD fault detected
  • Hydraulic cylinder rod stowage adjustment
  • Hydraulic cylinders raise stretcher too slowly
  • Hydraulic cylinders raise stretcher with excessive pump cycles
  • IV pole does not lock at different heights
  • Knee section does not lock into position
  • Knee section does not lower smoothly
  • Knee section does not lower smoothly
  • Knee section does not lower with crank handle
  • Knee section does not raise and lower smoothly
  • Knee section does not rise smoothly
  • Knee section does not rise smoothly
  • Knee section does not rise with crank handle
  • Label unreadable or not bonded
  • Manual back crank does not turn smoothly
  • Manual back section crank does not operate smoothly
  • Manual knee crank does not turn smoothly
  • Manual knee section crank does not operate smoothly
  • Motor Controller requires replacement
  • Neutral detent adjustment
  • Neutral position does not allow free caster movement
  • Nurse control switch replacement
  • Nurse control switch replacement needed
  • OB/GYN Foot Support Mechlok Device removal required
  • OB/GYN Upper Frame Catch Shroud removal required
  • OB/GYN Upper Frame Catch Shroud replacement required
  • PACU extender does not articulate smoothly
  • Patient control switch does not respond
  • Patient control switch replacement
  • Patient control switch replacement needed
  • Patient grip handle removal and replacement
  • Permanent IV pole does not raise or lower
  • Permanent IV pole does not raise or lower correctly
  • Permanent IV pole does not raise or lower correctly
  • Potentiometer removal and replacement
  • Potentiometer replacement and adjustment needed
  • Power cord assembly replacement
  • Power cord replacement in battery box system
  • PROC/Wiring fault detected
  • PROC/Wiring fault detected
  • Push handles do not drop to stored position
  • Push handles do not lock in raised position
  • Reduced braking ability
  • Removal and Replacement: CPR Release Cable Assembly
  • Removal and Replacement: OB/GYN Foot Support Gas Spring
  • Removal and Replacement: OB/GYN Foot Support Mechlok® Device
  • Removal and Replacement: OB/GYN Upper Frame Catch Shroud
  • Removal and Replacement: Upper Frame and Base Shroud
  • Removal: Load beam scale
  • Replace Electric (P8020) stretcher siderail cable assembly and upright tube
  • Replace OB/GYN (P8050) stretcher patient grip handle
  • Replace patient grip handle (P8050)
  • Replace siderail cable assembly and upright tube (P8020)
  • Replace siderail latch (P8000/P8010/P8020/P8040/P8050 stretchers)
  • Replace siderail latch assembly
  • Replace siderail tube
  • Replace siderail tube (P8000, P8010, P8020, P8040, P8050)
  • Replace siderail tubes (P8000/P8010/P8020/P8040/P8050 stretchers)
  • Replace Transport (P8005) stretcher siderail tube
  • Replacement: CPR release cable assembly (P8020)
  • Replacement: Drive wheel chains and sprockets
  • Replacement: Head drive assembly (P8020)
  • Replacement: Head-end hydraulic cylinder
  • Replacement: Knee drive assembly (P8020)
  • Replacement: Knee screw assembly (P8000/P8010)
  • Replacement: Load beam scale
  • Replacement: Upper Frame and Base Shroud
  • Reverse Trendelenburg function does not lower foot end
  • Scale batteries depleted or non-functional
  • Scale battery replacement
  • Scale display does not respond to weight
  • Scale display shows Hands Off, CALC, and dashes
  • Scale P.C. board removal and replacement
  • Scale weight display flashing
  • Scale weight displayed is flashing
  • Scale weight reading incorrect
  • Scale weight shown is incorrect
  • Service indicator flashing
  • Siderail cable assembly and upright tube removal
  • Siderail cable assembly or upright tube replacement needed
  • Siderail cable assembly replacement
  • Siderail does not operate correctly
  • Siderail does not raise or lock
  • Siderail latch does not lock
  • Siderail latch removal and replacement
  • Siderail tube removal and replacement
  • Steering Plus system fifth wheel does not drop
  • Stow-Away Push Handle removal and installation
  • Stretcher will not deploy the wheel
  • Stretcher will not drive
  • Stretcher will not stow the wheel
  • Throttle zero adjustment verification
  • Throttle zero adjustment verification
  • Trauma stretcher cassette lift does not function
  • Trendelenburg function does not lower head end
  • Upper Frame and Base Shroud removal required
  • Upper Frame and Base Shroud replacement required
  • Weight display is flashing
  • Weight shown is not correct

254 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
○ ○○Throttle Fault 1Incorrect voltage at controller throttle inputwarning
○ ○Thermal CutbackOperating temperature outside range or mechanical binding detectedwarning
○ ○○○SPD Limit Pot FaultSpeed limit potentiometer signal not reaching controllerwarning
○ ○○Throttle Fault 1Incorrect voltage at controller throttle inputwarning
○ ○Thermal CutbackOperating temperature outside range or mechanical binding detectedwarning
○ ○○○SPD Limit Pot FaultSpeed limit potentiometer signal not reaching controllerwarning
0Scale display errorScale system malfunction or sensor errorunknown
1Scale display errorScale system malfunction or sensor errorunknown
2Scale display errorScale system malfunction or sensor errorunknown
3Scale display errorScale system malfunction or sensor errorunknown
5Scale display errorScale system malfunction or sensor errorunknown
6Scale display errorScale system malfunction or sensor errorunknown
Alarm (load beam cable disconnection)Load beam cables not connected correctly to scale P.C. board portsDuring calibration if cables are improperly connected or seatedwarning
Alarm (load beam cables not connected correctly)Load beam cables disconnected or misconnected during scale replacementLoad beam cables not connected to correct ports on scale P.C. boarderror
Battery over current conditionCharging current exceeded maximum safe limitCharging current measurement exceeds over-current threshold during chargingerror
Battery over voltage conditionBattery voltage exceeded maximum safe charging voltage limitBattery voltage measurement exceeds over-voltage threshold during charging cycleerror
Battery shutdownSystem entered power-down mode due to prolonged inactivityNo enable switch presses and no steer switch transitions for 20 minutes of continuous inactivityinfo
Battery temperature faultBattery temperature measurement outside acceptable charging rangeBattery temperature sensor reading exceeds high limit or drops below low limit during chargingwarning
Current measurement errorCharging system unable to accurately measure battery charging currentCurrent measurement circuit failure detected during chargingerror
DCB to reed switch board loopback errorCommunication or signal loopback test between DCB and reed switch board failedSelf-test function during DCB initialization detects communication failureerror
Disengage faultDisengage mechanism failed to complete disengagement sequenceDisengagement sequence does not complete within expected time after being requestederror
Drive motor thermal shutdownDrive motor temperature exceeded safe operating limit and system shut down motor driveMotor temperature sensor reading exceeds threshold during drive operationerror
DS11=0NAC counts greater than threshold, normal conditionDrive capacity indicator normalinfo
DS11=1 (flashing at 1 Hz)NAC counts less than threshold, zero drive capacity remainingBattery charge depleted below safe operating levelwarning
DS2=0, DS5=0, DS4=1Battery temperature faultBattery temperature out of safe operating rangewarning
DS2=0, DS5=1, DS4=0Voltage measurement faultDCB cannot measure battery voltage correctlyerror
DS2=0, DS5=1, DS4=1Current measurement faultDCB cannot measure battery current correctlyerror
DS2=1, DS5=0, DS4=0Over voltage conditionBattery voltage exceeds safe operating limiterror
DS2=1, DS5=0, DS4=1Over current conditionBattery current draw exceeds safe limiterror
DS7=0, DS8=0, DS9=0, DS10=1Drive motor thermal shutdownDrive motor temperature exceeds safe operating limitwarning
DS7=0, DS8=0, DS9=1, DS10=0Enable switch faultEnable switch malfunction or communication losserror
DS7=0, DS8=0, DS9=1, DS10=1Invalid steer switch faultSteer switch sends invalid signalerror
DS7=0, DS8=1, DS9=0, DS10=0Engage limit switch faultEngage limit switch not responding correctlyerror
DS7=0, DS8=1, DS9=0, DS10=1Engage motor faultEngage motor malfunction or loss of driveerror
DS7=0, DS8=1, DS9=1, DS10=0DCB to reed switch board loopback failureCommunication loss between DCB and reed switch boarderror
DS7=0, DS8=1, DS9=1, DS10=1Engage motor thermal shutdownEngage motor temperature exceeds safe limitwarning
DS7=1, DS8=0, DS9=0, DS10=0Disengage faultDisengage mechanism malfunctionerror
DS7=1, DS8=0, DS9=0, DS10=1Gas gauge communication failureDCB cannot communicate with gas gaugeerror
DS7=1, DS8=0, DS9=1, DS10=0EEPROM faultEEPROM memory error in DCBcritical
EEPROM faultInternal EEPROM memory failure in DCBDCB self-test detects EEPROM read/write failure during initializationcritical
Enable switch faultEnable switch input to DCB not functioning or providing incorrect signalEnable switch not transitioning between states as expected during operationerror
Engage faultEngage mechanism failed to reach expected engagement stateEngagement sequence does not complete within expected timeerror
Engage limit switch faultEngage limit switch not functioning or providing incorrect signalEngage mechanism unable to detect proper engagement positionerror
Engage motor thermal faultEngage motor temperature exceeded safe operating limitEngage motor temperature sensor reading exceeds threshold during engagementerror
Err0Scale error code 0Unknown condition triggering scale errorerror
Err0Scale error code 0Unknown conditionunknown
Err1Scale error code 1Unknown conditionunknown
Err1Scale error code 1Unknown condition triggering scale errorerror
Err2Scale error code 2Unknown conditionunknown
Err2Scale error code 2Unknown condition triggering scale errorerror
Err3Scale error code 3Unknown conditionunknown
Err3Scale error code 3Unknown condition triggering scale errorerror
Err5Scale error code 5Unknown conditionunknown
Err5Scale error code 5Unknown condition triggering scale errorerror
Err6Scale error code 6Unknown condition triggering scale errorerror
Err6Scale error code 6Unknown conditionunknown
Error 6Scale display errorUnknown; section heading only, no detail providedunknown
Gas gauge communication faultGas gauge module not communicating with DCBDCB unable to establish or maintain communication with gas gauge during system initialization or operationerror
IntelliDrive Battery Temperature FaultBattery temperature out of safe operating rangeBattery temperature exceeds thresholdwarning
IntelliDrive DCB to Reed Switch Board Loopback FaultCommunication fault between Drive Control Board and Reed Switch BoardLoopback test failure or communication losserror
IntelliDrive Disengage FaultDisengage mechanism malfunction detectedDisengage system failure or control faulterror
IntelliDrive Drive Motor Thermal ShutdownDrive motor overheated and system shut downDrive motor temperature exceeded thermal limitswarning
IntelliDrive Enable Switch FaultEnable switch malfunction detectedEnable switch circuit fault or switch failureerror
IntelliDrive Engage FaultEngage mechanism malfunction detectedEngage system failure or control faulterror
IntelliDrive Engage Limit Switch FaultEngage limit switch malfunction detectedLimit switch circuit fault or switch failureerror
IntelliDrive Engage Motor Thermal FaultEngage motor overheatedEngage motor temperature exceeded thermal limitswarning
IntelliDrive Over Voltage, Over Current, or Charging FaultElectrical fault in power system or chargerOver voltage, over current, or charging system malfunctioncritical
IntelliDrive Stretcher Will Not DriveDrive motor will not operate or stretcher cannot moveDrive system failure or motor control faulterror
Invalid steer switchSteer switch signal invalid or contradictory state detectedSteer switch transitions detected when not expected or no transition when requirederror
NAC at thresholdBattery normalized ampere-hour capacity within normal operating rangeMonitored at each battery usage cycleinfo
NAC below thresholdBattery normalized ampere-hour capacity depleted; zero drive capacity remainingBattery usage depletes available energy below minimum thresholdwarning
Voltage measurement errorCharging system unable to accurately measure battery voltageVoltage measurement circuit failure detected during charging initializationerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.