Hill-Rom Connex 6000: troubleshooting and error codes
The ServiceForge knowledge base holds 137 documented symptoms for the Hill-Rom Connex 6000, with 278 ranked causes, 213 diagnostic checks with test points and expected values, and 16 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Device does not power up
Unit remains off; no display, LEDs, or startup indicators
Ranked causes
- 1. New battery installed (occasional)
- 2. AC power disconnected (common)
- 3. Power cord defective (occasional)
- 4. Battery discharged (common)
- 5. Power button defective (rare)
- 6. Internal power-flex cable connection faulty (occasional)
- 7. Power supply defective (occasional)
- 8. Battery defective (occasional)
- 9. Main board defective (rare)
Inspect power cord for damage
- Test point
- Power cord ends and body
- Expected
- No cuts, burns, or visible damage
5 more checks with pass and fail actions, 14 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer does not print
No output from thermal printer despite power and connection
Ranked causes
- 1. Reactive side of thermal paper faces away from print head (common)
- 2. Thermal paper is wet (occasional)
- 3. USB cable is defective (occasional)
- 4. Printer module not connected (occasional)
- 5. Printer module has no power (common)
- 6. Printer module does not function (occasional)
- 7. Power harness is defective (occasional)
- 8. Main board is defective (rare)
- 9. MCE printer fuse F5 is blown (occasional)
9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Braun temperature not functional
Braun temperature module has critical error; service required
Ranked causes
- 1. Communication timeout or BDD transfer delay (occasional)
- 2. POST failure (occasional)
- 3. Braun display shows all icons or blank (occasional)
- 4. Unsupported parameter or sensor definition error (rare)
- 5. WACP serialization failure (rare)
- 6. Battery voltage too high (rare)
- 7. BDD compatibility issue (rare)
- 8. Send message or bridge failure (rare)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touchscreen does not respond
Touch inputs not recognized; screen may appear functional but unresponsive to touch
Ranked causes
- 1. Software error (common)
- 2. Touchscreen out of calibration (occasional)
- 3. Touchscreen lock activated (occasional)
- 4. Internal connection faulty (occasional)
- 5. Display flex cable broken (rare)
- 6. Main board touchscreen controller faulty (rare)
- 7. Excessive force or prolonged rubbing damage (rare)
Reboot device
- Test point
- Power button
- Expected
- Device powers off and restarts; touchscreen responsive
3 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Attach SpO2 sensor to monitor
SpO2 sensor not detected; connection missing or sensor faulty
Ranked causes
- 1. SpO2 sensor not detected (common)
- 2. SpO2 sensor is faulty (occasional)
- 3. SpO2/SpHb sensor expired (occasional)
- 4. No SpO2 sensor connected (common)
- 5. Patient cable is faulty (occasional)
- 6. SpO2 module is faulty (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No audible alarm occurs
Audible alarm sound not produced during alarm conditions; volume may be set too low or audio disabled
Ranked causes
- 1. No alarm was triggered (common)
- 2. Alarm audio set to Off (occasional)
- 3. Alarm audio set too low (occasional)
- 4. Faulty connection to speaker (occasional)
- 5. Speaker is defective (rare)
- 6. Main board is defective (rare)
Verify alarm trigger and audio response
- Test point
- Device during alarm condition
- Expected
- Visual indicator in message status area and audible sound at startup
5 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode reader powers on but does not transfer data
Reader is powered but fails to communicate or transfer scanned data
Ranked causes
- 1. No barcode reader license installed (occasional)
- 2. Barcode reader not programmed for USB COM Port Emulation mode (occasional)
- 3. Barcode reader model not supported or PID incorrect (rare)
- 4. Barcode reader did not enumerate properly (occasional)
- 5. Barcode scanner not programmed to interpret barcode correctly (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device does not communicate via Ethernet
Device cannot connect to computer network via Ethernet; no network communication established
Ranked causes
- 1. Device not configured properly (occasional)
- 2. Communications board not receiving power (occasional)
- 3. Ethernet connection from main board is faulty (occasional)
- 4. Network Ethernet switches not set correctly (occasional)
- 5. Cable run to switch too long (rare)
Verify device Ethernet configuration
- Test point
- Device network settings and network administrator records
- Expected
- Device IP address, subnet mask, gateway configured correctly for network
4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display is blank when power is on
No image or text visible on screen; backlight may or may not be active
Ranked causes
- 1. Device in Power-Saver mode (common)
- 2. Device powered down after inactivity (common)
- 3. Internal connection faulty (occasional)
- 4. Cable damaged (occasional)
- 5. Display end of useful life (rare)
Attempt to wake display
- Test point
- Screen touch or power button
- Expected
- Display activates and shows content
2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No ECG pane displayed
ECG tile or pane not visible on monitor display
Ranked causes
- 1. ECG module not connected (common)
- 2. Faulty patient cable (occasional)
- 3. Faulty module cable (occasional)
- 4. CVMS software does not support ECG (rare)
- 5. Profile not set to Continuous Monitoring (common)
Inspect ECG module LED status
- Test point
- ECG module LED indicator
- Expected
- LED illuminated (green or amber), indicating power and communication
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpHb frame blank on display
Hemoglobin parameter tile shows no data
Ranked causes
- 1. UI license not installed (occasional)
- 2. Wrong sensor connected (occasional)
- 3. Sensor or cable expired (occasional)
- 4. Sensor or cable defective (occasional)
- 5. SpHb parameter not enabled in Masimo module (occasional)
Verify UI license status
- Test point
- Service tool license verification
- Expected
- SpHb license installed and active
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Braun temperature not detectable
Braun ThermoScan PRO unable to detect temperature; measurement failed or recalled from memory
Ranked causes
- 1. Temperature measurement was not taken (occasional)
- 2. Measurement recalled from memory (occasional)
- 3. Loose or broken USB cable (occasional)
- 4. Braun thermometer battery depleted (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG electrodes disconnected patient safety
One or more ECG leads or electrodes are not connected properly
Ranked causes
- 1. Device configured for 5 leads but using 3 leads (common)
- 2. Specific IEC lead disconnected (common)
- 3. Specific AHA lead disconnected (common)
- 4. One lead is disconnected (common)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG not functional patient safety
ECG module fails to operate; data transmission or connection error
Ranked causes
- 1. ECG module error, WACP error, or data failure (occasional)
- 2. ECG module has not transmitted data for 30+ seconds (occasional)
- 3. ECG module disconnected (common)
- 4. Unexpected alarm compatibility bits (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Light bar does not turn on
Alarm light bar not illuminating; no amber or red LED indication during alarm or startup
Ranked causes
- 1. No alarm was triggered (common)
- 2. Faulty connection to light bar (occasional)
- 3. Light-bar board is defective (rare)
- 4. Main board is defective (rare)
Verify light bar operation during startup
- Test point
- Light bar LEDs during device power-on
- Expected
- Light bar flashes (amber and/or red LEDs)
4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Radio does not connect to network
Wireless/radio connection cannot establish; device out of range or unable to find access point
Ranked causes
- 1. Device out of range of access point (common)
- 2. Device not configured properly (occasional)
- 3. Antenna is defective (rare)
- 4. Communications board not receiving power (occasional)
Check wireless signal strength
- Test point
- Device network status screen RSSI indicator
- Expected
- RSSI value indicates adequate signal strength (typically >-80 dBm)
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Retry temperature measurement
Temperature reading failed; often accompanies other temperature messages
Ranked causes
- 1. Probe heater or data error occurred (occasional)
- 2. User settings require adjustment (occasional)
- 3. Ambient temperature is out of range (occasional)
- 4. SureTemp temperature module is faulty (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
USB accessories do not communicate with monitor
USB peripherals (devices, thumb drives, sensors) not recognized by device; no power or enumeration
Ranked causes
- 1. Accessory is defective (occasional)
- 2. Communications board not receiving power (occasional)
- 3. One or more USB host connections defective (occasional)
- 4. USB connection from communications board to main board faulty (occasional)
Test with known good USB accessory
- Test point
- USB port on device
- Expected
- Accessory is recognized; power and enumeration present
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Weight does not appear in parameter frame
Weight parameter not displayed in manual parameter frame
Ranked causes
- 1. Weight not selected in advanced settings (common)
- 2. Weight scale not licensed (occasional)
- 3. Weight scale not connected (occasional)
- 4. Weight scale not configured (occasional)
Verify weight selected in advanced settings
- Test point
- Advanced settings > Parameters tab
- Expected
- Weight parameter selected (enabled)
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery does not charge or run time is low
Battery remains discharged after charging; device shuts off prematurely during use
Ranked causes
- 1. Battery defective (common)
- 2. Battery connector board defective (occasional)
- 3. Battery end of useful life (occasional)
Charge battery for 5 hours
- Test point
- Battery during charging from AC power
- Expected
- Battery icon shows full charge symbol
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
BMI not displayed
BMI parameter not appearing in parameter frame
Ranked causes
- 1. BMI parameter not selected (common)
- 2. Height or weight was changed (occasional)
- 3. Weight scale not connected (occasional)
Verify BMI parameter selected
- Test point
- Advanced settings > Parameters tab
- Expected
- BMI parameter selected (enabled)
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Braun Err error
Braun displays error; ambient temperature outside operating range or changes too rapidly
Ranked causes
- 1. Ambient temperature outside operating range or changing too rapidly (common)
- 2. Thermometer docked improperly (occasional)
- 3. Thermometer has dirty contacts (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Braun PRO 6000 measurements inaccurate patient safety
Braun thermometer readings deviate from expected values
Ranked causes
- 1. Probe lens displaced (occasional)
- 2. Use inconsistent with IFU (occasional)
- 3. Ingress of cleaning solutions corroded electronics (rare)
Inspect probe lens for displacement
- Test point
- Probe lens and bezel seam
- Expected
- Lens properly seated with no visible gap in seam
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot communicate through USB client connection
Service tool or PC unable to establish USB connection with device; communication fails
Ranked causes
- 1. Battery charge is low (occasional)
- 2. Communications board does not receive power (occasional)
- 3. USB client is defective (rare)
Verify battery charge level
- Test point
- Battery status indicator on display
- Expected
- Battery icon shows adequate charge (at least 50%)
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low SpO2 signal quality
Poor sensor signal due to sensor placement or defective equipment; check sensor
Ranked causes
- 1. Poor sensor placement on patient (common)
- 2. Patient cable or sensor defective (occasional)
- 3. SpO2 module is faulty (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP frame on display is blank
Non-invasive blood pressure module display or data frame shows no content or is empty
Ranked causes
- 1. USB cable is defective (occasional)
- 2. NIBP module not connected (occasional)
- 3. NIBP module not functional (occasional)
Inspect USB cable between NIBP module and main board
- Test point
- USB cable connector ends and body
- Expected
- No cuts, damage, or bent pins; connectors fully seated
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No ECG waveform displayed
ECG pane present but no waveform tracing visible
Ranked causes
- 1. Start button not activated (common)
- 2. ECG gain too high (occasional)
- 3. Patient type set to Neonate (occasional)
Verify Start button activated
- Test point
- ECG tile Start button state
- Expected
- Start button activated and waveform streaming engaged
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No LTA alarm with patient injury or death risk patient safety
LTA (Long-Term Arrhythmia) alarm not functioning despite critical patient events
Ranked causes
- 1. LTA license not installed (occasional)
- 2. Arrhythmia detection disabled (V-Tach, V-Fib, Asystole) (occasional)
- 3. Invalid data being processed (rare)
Verify LTA license installed
- Test point
- Service tool license list
- Expected
- LTA license present and active
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Noisy ECG waveform
ECG waveform displays excessive noise or artifact
Ranked causes
- 1. 50 or 60 Hz filter disabled (common)
- 2. Wrong mains filter frequency selected (common)
- 3. Poor lead placement (common)
Verify mains filter enabled
- Test point
- ECG filter settings menu
- Expected
- Mains filter (50 or 60 Hz) enabled
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Radio software upgrade failed
Radio software update was interrupted or failed; hardware issue possible
Ranked causes
- 1. Connection with host was broken (common)
- 2. Radio not provisioned correctly (occasional)
- 3. Hardware error during upgrade (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 frame blank on display
SpO2 parameter tile shows no data or reading
Ranked causes
- 1. USB cable defective (common)
- 2. SpO2 module not connected (common)
- 3. SpO2 module not functional (occasional)
Inspect USB cable for damage and verify connection
- Test point
- SpO2 module USB connector
- Expected
- Cable intact, firmly connected
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 sensor expires soon
SpO2 sensor nearing expiration date; replacement needed
Ranked causes
- 1. SpO2 sensor nearing expiration (common)
- 2. SpO2 cable not functioning properly (occasional)
- 3. SpO2 module is faulty (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature frame blank on display
Temperature parameter tile shows no data
Ranked causes
- 1. USB cable defective (common)
- 2. Temperature module not connected (common)
- 3. Temperature module not functional (occasional)
Inspect USB cable for damage
- Test point
- Temperature module USB connector
- Expected
- Cable intact, firmly connected
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature module connect probe error
Temperature probe disconnected or module cannot detect probe connection
Ranked causes
- 1. No temperature probe connected (common)
- 2. Temperature probe is faulty (occasional)
- 3. SureTemp temperature module is faulty (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature not functional
Temperature system has critical error; service required
Ranked causes
- 1. Internal error occurred (occasional)
- 2. USB cable is disconnected (common)
- 3. Battery is depleted or missing (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to access advanced settings or enter code
Technician cannot access Advanced Settings menu; code entry blocked or settings unavailable
Ranked causes
- 1. Patient monitoring active or simulated (occasional)
- 2. Parameter alarm is on (occasional)
- 3. Intervals turned on (occasional)
Verify no patient monitoring active
- Test point
- Display and monitoring status
- Expected
- Patient monitoring stopped; simulation mode off
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to establish network communications
Radio cannot associate with access point or authenticate after 30 seconds; network connectivity failure
Ranked causes
- 1. Unable to associate with access point (common)
- 2. Unable to authenticate after 30 seconds (common)
- 3. DHCP timeout (occasional)
Monitor association time and verify access point detected
- Test point
- radio association process
- Expected
- association within 30 seconds
2 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery absent or faulty patient safety
No battery installed in monitor or battery is faulty
Ranked causes
- 1. No battery installed (common)
- 2. Battery is faulty (common)
Verify battery is installed
- Test point
- battery compartment
- Expected
- battery present and seated
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Braun thermometer readings do not transfer
Temperature measurements not appearing on monitor display after taking with Braun device
Ranked causes
- 1. External USB cable unplugged (common)
- 2. USB bus communication failure (error #00000014) (occasional)
Verify external USB cable connected
- Test point
- USB cable from dock to monitor
- Expected
- Cable firmly connected at both ends
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Braun ThermoScan battery does not charge
Thermometer rechargeable battery pack unable to charge in dock
Ranked causes
- 1. Rechargeable battery pack no longer takes charge (common)
- 2. Primary AA batteries installed instead of rechargeable (occasional)
Inspect battery type installed
- Test point
- Thermometer battery compartment
- Expected
- Rechargeable battery pack installed (not primary AA batteries)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Communications module did not power on properly
Communications board not powering or not properly connected; high-priority alarm
Ranked causes
- 1. Communications board not connected properly (common)
- 2. Communications board malfunctioned (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Connect temperature probe
Temperature probe disconnected or faulty; cannot measure patient temperature
Ranked causes
- 1. No temperature probe connected (common)
- 2. Temperature probe is faulty (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Desired profile does not appear in Profiles tab
Custom or expected monitoring profile not visible in profile selection menu
Ranked causes
- 1. Profile license not installed (occasional)
- 2. Profile not enabled (occasional)
Check profile license status
- Test point
- Advanced Settings > Service > Licensing
- Expected
- Profile license box is checked
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display is dim
Screen displays content but appears darker than normal; low brightness
Ranked causes
- 1. Brightness setting too low (common)
- 2. Display reached end of useful life (rare)
Check brightness setting
- Test point
- Settings > Device > Defaults > Display brightness
- Expected
- Brightness set to appropriate level (typically medium to high)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dock LED green but battery low or depleted
Dock shows charging indicator but thermometer battery not holding charge
Ranked causes
- 1. Primary AA batteries installed instead of rechargeable (common)
- 2. Dock defective (occasional)
Verify battery type in thermometer
- Test point
- Battery compartment
- Expected
- Rechargeable battery pack (not primary AA)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Failed to authenticate to server
Device cannot authenticate to server; authentication failed
Ranked causes
- 1. Server certificate validation failed (common)
- 2. Server not configured for authentication (occasional)
Verify server certificate is valid and not expired
- Test point
- server certificate
- Expected
- certificate valid and within validity period
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Insert correct color-coded probe well
Temperature probe well missing or incorrect color-coded well not detected
Ranked causes
- 1. Probe well is missing (common)
- 2. SureTemp temperature module is faulty (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Network cable unplugged
Ethernet network connection lost or unavailable due to unplugged cable
Ranked causes
- 1. Network cable is unplugged (common)
- 2. Network connection broken elsewhere (occasional)
Verify network cable connection at device
- Test point
- network cable connector at monitor
- Expected
- cable fully seated and connected
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP not functional
NIBP measurement fails with critical error; service required
Ranked causes
- 1. Internal or messaging errors occurred (occasional)
- 2. Ambient temperature out of range (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP tubing kinked
External or internal tubing obstruction prevents proper pressure transmission
Ranked causes
- 1. NIBP tubing kinked on outside of device (common)
- 2. NIBP tubing kinked inside module (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No connection for data send
Monitor not configured to network; cannot send clinical data
Ranked causes
- 1. Monitor not network configured (common)
- 2. Network connectivity unavailable (common)
Verify network configuration for data send
- Test point
- Advanced settings > Data Management > Clinical Data
- Expected
- send enabled and configured
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No ECG tile or LTA display with alarms off patient safety
ECG tile or LTA tile not visible; alarms disabled in settings
Ranked causes
- 1. Alarms disabled in advanced settings (common)
- 2. No license for feature (occasional)
Verify alarm status in advanced settings
- Test point
- Advanced settings > Alarms menu
- Expected
- Alarms enabled (toggle on)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Replace temperature probe
Temperature probe failure or exhaustion; measurement cannot proceed
Ranked causes
- 1. Temperature probe is faulty (common)
- 2. SureTemp temperature module is faulty (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Searching for respiration patient safety
ECG module unable to measure patient respirations
Ranked causes
- 1. ECG unable to measure respirations (common)
- 2. Not using Lead II to view respiration (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Set date and time
Date or time is not set or not set properly on the monitor
Ranked causes
- 1. Date or time not set (common)
- 2. Date or time set incorrectly (common)
Verify date is configured
- Test point
- system date setting
- Expected
- current date set correctly
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Snapshot button disabled
Snapshot capture button unavailable or grayed out
Ranked causes
- 1. Alarming snapshot in progress (occasional)
- 2. Less than one minute elapsed between snapshots (common)
Check if snapshot in progress
- Test point
- Snapshot status display
- Expected
- No snapshot currently being captured
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 not functional
SpO2 measurement fails; sensor defective or module error
Ranked causes
- 1. SpO2 sensor is defective (occasional)
- 2. Internal SpO2 error occurred (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tube type does not match device configuration
Connected NIBP tubing type differs from configured settings; informational message only
Ranked causes
- 1. Tube type mismatch with NIBP configuration (occasional)
- 2. Single lumen switch stuck on NIBP connection (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to determine NIBP; check inflation settings
NIBP measurement fails due to incorrect inflation parameters
Ranked causes
- 1. Target pressure too low (occasional)
- 2. Too many inflation attempts (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Attempting to acquire ECG/Impedance respiration patient safety
ECG acquisition ongoing; checking electrode and cable connections
Ranked causes
- 1. Electrode or cable connection issue during acquisition (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Braun display blank after software upgrade
- Braun display blank or shows all icons
- Braun display locked and unresponsive
- Braun displays Alert icon with '1'
- Braun displays Alert icon with 'U'
- Braun displays Err error
- Braun displays POS error
- Braun POS error
- C/F button not functional
- Calibration: Chinese fonts and IME installation
- Calibration: etCO2
- Calibration: Firmware upgrade procedure
- Calibration: License activation procedure
- Cannot analyze ECG
- Cannot measure ECG
- Clinician ID required to save data
- Device shutdown not available
- ECG module unable to detect arrhythmias
- ECG module unable to detect pacemaker
- ECG tile disabled
- Excessive patient movement during NIBP
- External device not recognized
- Functional error during configuration
- Functional verification: Connex device configuration
- Functional verification: ECG module
- Functional verification: ECG simulator configuration
- Functional verification: Pacemaker detection
- Incompatible Hillrom device
- Incorrect NIBP cuff size
- Invalid radio configuration
- Lost connectivity with host
- Low battery 30 minutes or less
- Low battery 5 minutes or less
- Low perfusion
- Low SpHb signal quality
- Manual parameter frame not displayed
- Maximum patient records saved
- Module LED flashing green
- Module LED off
- Module LED steady green
- Module LED yellow
- NIBP air leak detected
- NIBP inflation too quick
- No alarming snapshot auto-printed
- No data saved
- No pacer or ECG tile shows pacer detection off
- Operating in battery mode
- Patient ID required to save data
- Patient ID required to send data
- Patient list is full
- Printed waveform too small
- Pulse timer button not functional
- Radio configuration certificate error
- Radio error; power down and restart
- Radio not functional
- Radio out of network range
- Respiration alarm limits changed
- Searching for SpO2
- SpHb frame unavailable
- SpO2 mode only
- Stop intervals to select new patient
- Temperature tile displays --
- Temperature tile displays ++
- Temperature time limit exceeded
- Tissue contact lost
- Unable to determine NIBP; check connections and limit patient movement
- Unable to establish network communications; radio out of range
- Unable to establish network communications; service needed
- Unable to identify clinician
- Unable to load configuration
- Unable to read configuration from USB
- Unable to retrieve patient list
- Unable to save configuration to USB
- Unexpected restart occurred
- USB accessory disconnected
- USB communication failure
- Weight scale not functional
77 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
#000000001 | Internal software error | Software detected internal error condition | error |
#000000002 | Unclassified hardware error | Hardware detected generic error condition | error |
#000000003 | Graphics RAM POST failure | Graphics RAM power-on self-test failed | error |
#000000004 | System RAM POST failure | System RAM power-on self-test failed | error |
#000000005 | Watchdog POST failure | Watchdog timer self-test failed | error |
#000000006 | FLASH initiation failed | Flash memory initialization failed | error |
#000000007 | Display system error | Display controller or video system failed | error |
#000000008 | Real-time clock error | Real-time clock circuit failed | error |
#000000009 | Audio system error | Audio circuit or speaker failed | error |
#000000010 | Ethernet system error | Ethernet controller or network interface failed | error |
#000000011 | Touchscreen controller failed | Touchscreen controller circuit failed | error |
#000000012 | SMBUS errors detected | Five or more SMBUS errors within 1-minute period | error |
#000000013 | Communications module or main board failed | Communications hardware or main board failed | critical |
#000000014 | Main board USB hub failure | USB hub on main board failed | error |
#000000015 | Software watchdog timer reset | Software watchdog timer reset occurred | error |
NAK: FmFLASH/GnSTATUS/1 | EarlySense module software error preventing downgrade | Attempt to downgrade EarlySense module firmware after upgrade | error |