Hill-Rom Century: troubleshooting and error codes
The ServiceForge knowledge base holds 43 documented symptoms for the Hill-Rom Century, with 129 ranked causes, 197 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Head up function does not work high voltage
Head up function is inoperative when activated from left or right siderail controls
Ranked causes
- 1. Missing or low voltage at head motor control pin (common)
- 2. Head motor not receiving power (common)
- 3. Defective head motor (occasional)
- 4. Transformer primary or secondary voltage failure (occasional)
- 5. Transformer circuit breaker tripped (occasional)
- 6. Secondary fuse blown in transformer box (occasional)
- 7. Primary fuse blown in transformer box (occasional)
- 8. Low voltage from control board to head motor (occasional)
- 9. Faulty cable from control board to head motor (occasional)
- 10. Defective control board (occasional)
- 11. Open circuit in siderail cable at test port (occasional)
- 12. Defective test port P200 (rare)
- 13. Open circuit in siderail cable from test port to siderail (occasional)
- 14. Head up button switch open or defective (occasional)
- 15. Defective head up button switch (occasional)
Verify control signal voltage at P200
- Test point
- P200 connector, pin 9 (head up) to pin 21 (VSS)
- Expected
- 12V DC (+3V), +3V
- Tools
- VOM set to DC volts
12 more checks with pass and fail actions, 16 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Knee up function does not work high voltage
Knee up function fails to operate when activated from left or right siderail; bed does not raise knee section.
Ranked causes
- 1. No voltage at P200 knee up control pin (pin 7) (common)
- 2. No voltage delivered to knee motor after control signal (common)
- 3. Knee motor defective or damaged (occasional)
- 4. Transformer output voltage not present (circuit breaker tripped) (occasional)
- 5. 12V transformer secondary fuse blown (outlet side) (occasional)
- 6. Primary transformer fuse blown (inlet power side) (occasional)
- 7. No voltage at motor control connector (P3) (occasional)
- 8. Control board to motor cable broken or disconnected (occasional)
- 9. Control board defective (occasional)
- 10. Siderail switch not making contact (open circuit) (occasional)
- 11. Test port connector defective (rare)
- 12. Cable between test port and siderail broken or disconnected (occasional)
- 13. Siderail knee up switch defective (open) (occasional)
- 14. Siderail knee up switch not closing when pressed (occasional)
Verify knee up function failure mode
- Test point
- Either left or right siderail knee up button
- Expected
- Function should activate when button is pressed
13 more checks with pass and fail actions, 14 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed up function does not work high voltage
Bed does not move upward when bed up button activated from left or right siderail; bed down function may work normally
Ranked causes
- 1. Hilow drive at mechanical limit of movement (common)
- 2. Hilow limit switch defective or not clicking (common)
- 3. No 12V DC power to bed up function at P200 pin 11 (common)
- 4. Hilow motor not receiving 24V DC (common)
- 5. No 32V AC at transformer primary (P1 pin 3) (occasional)
- 6. No 12V AC at transformer secondary (P1 pin 2) (occasional)
- 7. Open fuse at transformer inlet (primary side) (occasional)
- 8. Hilow motor power cable voltage wrong or disconnected (occasional)
- 9. Control board output defective (occasional)
- 10. Bed up siderail cable open or disconnected at test port (occasional)
- 11. Bed up siderail switch defective (open at rest) (occasional)
- 12. Bed up siderail switch defective (high resistance when pressed) (rare)
Attempt bed down button operation
- Test point
- Bed control
- Expected
- Bed moves downward
15 more checks with pass and fail actions, 12 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed down function does not work high voltage
Bed does not move downward when bed down button activated from left or right siderail; may be limit switch, motor, control circuit, or wiring issue
Ranked causes
- 1. Hilow limit switch malfunction or not clicking (common)
- 2. Insufficient voltage to hilow motor control signal (common)
- 3. Hilow motor defective (occasional)
- 4. Power supply 12V rail not present (transformer secondary fault) (occasional)
- 5. Fuse blown on secondary (common) side of transformer (occasional)
- 6. Fuse blown on primary (inlet) side of transformer (occasional)
- 7. Insufficient voltage to hilow motor from power supply (occasional)
- 8. Control board defective (occasional)
- 9. Test port P200 connector fault or siderail cable open (rare)
- 10. Siderail cable between test port and control open (rare)
- 11. Siderail bed down switch open or stuck open (rare)
Listen for hilow limit switch clicking when rod moved manually
- Test point
- Hilow rod grasped and moved back-and-forth 3 inches
- Expected
- Hilow limit switch clicking on and off
- Tools
- Manual grasp
13 more checks with pass and fail actions, 12 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Knee down function does not work patient safety
Knee down motion does not activate when switch on left or right siderail is pressed
Ranked causes
- 1. No 12V DC supply to knee down function at P200 (common)
- 2. Knee motor power cable disconnected or defective (occasional)
- 3. Knee motor failure (occasional)
- 4. 32V AC supply voltage absent at transformer primary P1 (occasional)
- 5. 12V AC supply voltage absent at transformer secondary P1 (occasional)
- 6. Primary inlet power fuse blown (occasional)
- 7. No 12V DC at motor control connector P3 (occasional)
- 8. Control board malfunction (rare)
- 9. Test port cable open circuit (occasional)
- 10. Knee down switch contact open in cable P17 or P18 (occasional)
- 11. Knee down switch contact open or high resistance in siderail (occasional)
Is knee down function inoperative from both left and right siderails
- Test point
- Both siderail switches
- Expected
- Function fails to activate from either siderail
16 more checks with pass and fail actions, 11 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head down function does not work high voltage
Head down function is inoperative when activated from either left or right siderail; bed does not lower from head end.
Ranked causes
- 1. Head motor failure (common)
- 2. No voltage at head motor control (common)
- 3. Motor power cable failure (occasional)
- 4. Power supply failure (32V AC rail) (occasional)
- 5. Power supply failure (12V AC rail) (occasional)
- 6. Primary power fuse failure (occasional)
- 7. Control board failure (rare)
- 8. Test port failure (rare)
- 9. Test port cable open (occasional)
- 10. Siderail switch failure (occasional)
Measure voltage at P200(A) control signal
- Test point
- P200(A) pin 21 (VSS) to pin 8 (head down)
- Expected
- 12V DC (+3V), +/-3V
- Tools
- VOM set to DC volts
10 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
None of the bed functions work high voltage
No response to any bed control inputs; bed does not respond to head, knee, hilow, or other function commands
Ranked causes
- 1. Lockouts in locked position (common)
- 2. Bed not plugged in to power source (common)
- 3. Fuses open or blown in transformer box (common)
- 4. Circuit breaker tripped in transformer box (common)
- 5. Logic power supply voltage out of range (occasional)
- 6. Motor power supply voltage out of range (occasional)
- 7. Supervisor relay not functioning on control board (occasional)
- 8. Motor failure (rare)
Ensure lockouts are in unlocked position
- Test point
- Lockout switches on bed
- Expected
- Lockouts in unlocked position
7 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Trendelenburg or Reverse Trendelenburg function does not work patient safety
Bed lowers completely but does not go into Trendelenburg or Reverse Trendelenburg when activated
Ranked causes
- 1. Trendelenburg or Reverse Trendelenburg rods are bent (common)
- 2. Trendelenburg or Reverse Trendelenburg linkages disconnected or broken (common)
- 3. Plated bracket wedged between torque tubes (occasional)
- 4. Trendelenburg and Reverse Trendelenburg switches binding (occasional)
Visually inspect Trendelenburg or Reverse Trendelenburg rods
- Test point
- Rods
- Expected
- Rods are straight
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed does not operate patient safety
Operator reports bed functions not responding to controls or demonstrating problems during operation
Ranked causes
- 1. Bed not plugged into power source (common)
- 2. Function lockouts are enabled (occasional)
- 3. Bed functions at movement limit (occasional)
Verify bed power connection
- Test point
- Power outlet and bed plug
- Expected
- Bed plugged into appropriate power source
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Siderail is hard to rotate up or down caution
Siderail requires excessive force to rotate to up or down position
Ranked causes
- 1. Siderail strap bent causing slide rods misalignment (common)
- 2. Pivot points on strap dry or contaminated (common)
- 3. Siderail mechanism requires greasing (common)
Examine siderail strap for bending
- Test point
- Siderail strap (B) and slide rods (C)
- Expected
- Slide rods parallel to each other
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head up function failure with left siderail control high voltage
Head up function non-responsive when activated from left siderail only or when activated from right siderail
Ranked causes
- 1. Open or high-resistance connection in left siderail cable (common)
- 2. Defective head up button switch in left siderail (common)
Check siderail cable resistance to test port
- Test point
- Cable P17 or P18, test for continuity and resistance
- Expected
- Open circuit (near infinite resistance)
- Tools
- VOM set to ohms
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Knee section does not lower high voltage
Knee section will not lower when head down switch is activated; knee up function works normally when knee down switch is activated.
Ranked causes
- 1. Automatic contour switch failure (common)
- 2. Control board failure (occasional)
Verify knee up function by activating knee down switch
- Test point
- Knee actuator
- Expected
- Knee up function works normally
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Knee section does not raise high voltage
Knee section will not raise when the head up switch is activated on beds with automatic contour
Ranked causes
- 1. Automatic contour switch failure (common)
- 2. Control board failure (occasional)
Verify knee up function by activating knee up switch
- Test point
- Knee up switch
- Expected
- Knee up function works normally
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Knee up function does not work from one siderail only high voltage
One siderail control (left or right) fails; the other operates normally.
Ranked causes
- 1. Siderail-specific cable open circuit (common)
- 2. Siderail switch inoperative or not closing (common)
Test specific siderail cable for open circuit
- Test point
- P17 or P18 cable (select which siderail is failing), pins 7 and 5
- Expected
- Open circuit (infinite ohms) when cable disconnected and switch not pressed
- Tools
- VOM set to ohms; bed unplugged, battery disconnected
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pendant control assembly removal and replacement high voltage
Procedure to remove and replace the pendant control assembly including the control cable, pendant holder, and related connections on the Century+ bed.
Ranked causes
- 1. Pendant control assembly requires service or replacement (unknown)
- 2. Pendant control assembly installed and tested (unknown)
Raise bed to high position using hilow function
- Test point
- Bed height control
- Expected
- Bed at high position
15 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery Backup Box replacement and installation high voltage
Procedure to remove and install the battery backup box, including routing power cords and securing components to the bed frame.
Ranked causes
- 1. Battery backup box removal and reinstallation required (unknown)
Raise bed to full up position using bed up button
- Test point
- Sleep surface
- Expected
- Bed at maximum height
20 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery backup cells replacement high voltage
Battery backup cells (D) in the battery box require removal and replacement to maintain backup power functionality.
Ranked causes
- 1. Battery cells degradation or failure (unknown)
19 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed does not enter Reverse Trendelenburg position patient safety
Bed fails to move to Reverse Trendelenburg position when control lever and bed up switch activated from high position
Ranked causes
- 1. Reverse Trendelenburg function malfunction (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed does not enter Trendelenburg position patient safety
Bed fails to move to Trendelenburg position when control lever and bed down switch activated from high position
Ranked causes
- 1. Trendelenburg function malfunction (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed does not lower when bed down switch activated patient safety
Bed fails to lower when bed down switch is pressed with hilow function activated
Ranked causes
- 1. Hilow function malfunction or motor failure (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed does not raise with hilow function patient safety
Bed fails to raise when bed up switch is activated on hilow function
Ranked causes
- 1. Hilow raise function malfunction (unknown)
Activate bed up switch for hilow function
- Test point
- Bed up control switch
- Expected
- Bed raises
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed height does not adjust patient safety
Hilow function inoperable; requires manual crank operation via IV rod
Ranked causes
- 1. Hilow motor failure or disconnection (unknown)
Verify bed is unplugged from power source
- Test point
- Power outlet
- Expected
- Bed unplugged
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed lowers 3 to 4 inches and stops when Trendelenburg activated patient safety
When Trendelenburg or Reverse Trendelenburg function is activated, bed only lowers 3-4 inches then stops
Ranked causes
- 1. Trendelenburg and Reverse Trendelenburg switches binding (occasional)
Activate Trendelenburg or Reverse Trendelenburg and observe if both functions activate simultaneously
- Test point
- Function activation
- Expected
- Only activated function engages
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Brake Cam Assembly patient safety
Removal, replacement, and adjustment of brake cam assembly on beds with central brake and steer mechanism. Involves removing plastic covers, relieving cable tension, lowering bed, disconnecting power and battery, laying bed on side, removing roll pin, and installing cams.
Ranked causes
- 1. Brake cam assembly requires service (removal/replacement) (unknown)
15 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Head up function high voltage
Verification and adjustment procedure for head up motor and control circuit operation
Ranked causes
- 1. Normal post-repair verification (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Hilow limit switch high voltage
Calibration procedure for hilow limit switch assembly
Ranked causes
- 1. No adjustment required (common)
Calibration: Pendant control assembly
Adjustment procedure for pendant control assembly after installation.
Ranked causes
- 1. No calibration or adjustment required (common)
Calibration: Trendelenburg and Reverse Trendelenburg indicators patient safety
Indicator knobs for Trendelenburg and Reverse Trendelenburg require adjustment to ensure proper readings for inches and degrees
Ranked causes
- 1. Indicator knobs out of calibration (common)
Raise bed to high position
- Test point
- hilow function
- Expected
- bed at high position
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Trendelenburg or Reverse Trendelenburg Assembly patient safety
Replacement and functional verification of Trendelenburg or Reverse Trendelenburg assembly on Century+ bed
Ranked causes
- 1. Assembly replacement or maintenance required (unknown)
Mechanical linkage engagement
- Test point
- Trendelenburg and Reverse Trendelenburg mechanism
- Expected
- Linkage engaging properly
1 more check with pass and fail actions, 12 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Casters slip after adjustment patient safety
Brakes slip or casters do not hold position properly after adjustment procedure completed
Ranked causes
- 1. Floor wax buildup on tire tread (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head and knee sections do not lower with head down switch patient safety
Head section and knee section fail to lower when head down switch is pressed with knee lockout on
Ranked causes
- 1. Head down function or automatic contour malfunction (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head and knee sections do not raise with head up switch patient safety
Head section and knee section fail to raise when head up switch is pressed with knee lockout on
Ranked causes
- 1. Head up function or automatic contour malfunction (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head section does not lower patient safety
Head section fails to lower when head down switch is activated
Ranked causes
- 1. Head down function malfunction (unknown)
Activate head down switch
- Test point
- Head down control switch
- Expected
- Head section lowers
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head section does not raise patient safety
Head section fails to raise when head up switch is activated
Ranked causes
- 1. Head up function malfunction (unknown)
Activate head up switch
- Test point
- Head up control switch
- Expected
- Head section raises
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head section does not raise or lower patient safety
Head section manually inoperable; requires manual crank operation via IV rod
Ranked causes
- 1. Motor failure or disconnection (unknown)
Verify bed is unplugged from power source
- Test point
- Power outlet
- Expected
- Bed unplugged
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hilow limit switch replacement and verification high voltage
Switch malfunction requiring removal and replacement of the hilow limit switch assembly in the control box
Ranked causes
- 1. Hilow limit switch failure or malfunction (unknown)
Verify hilow function operation
- Test point
- Bed hilow control
- Expected
- Bed raises and lowers smoothly without binding or stopping prematurely
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hilow motor installation high voltage
Procedure to install the hilow motor on Century bed including engagement with drive shaft and securing with fasteners.
Ranked causes
- 1. Motor requires installation (unknown)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hilow motor removal and replacement high voltage
Procedure to remove and replace the hilow motor on Century bed, including safety precautions for powered mechanisms and electrical hazards.
Ranked causes
- 1. Motor requires removal or replacement (unknown)
11 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Knee section does not raise or lower patient safety
Knee function inoperable; requires manual crank operation via IV rod
Ranked causes
- 1. Knee motor failure or disconnection (unknown)
Verify bed is unplugged from power source
- Test point
- Power outlet
- Expected
- Bed unplugged
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lockouts hanging up on bezel
Lockout and switch mechanisms may not move freely due to improper bezel adjustment
Ranked causes
- 1. Lockouts or switches not properly adjusted relative to bezel (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Motor overheating or stalling caution
Motor stops running due to thermal overload condition; automatic thermal reset required
Ranked causes
- 1. Motor thermal overload (common)
Verify motor is not running and is warm to touch
- Test point
- Motor housing
- Expected
- Motor stopped with heat present
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Siderail removal and replacement high voltage
Removal, replacement, and functional testing of siderails on Century+ bed
Ranked causes
- 1. Routine maintenance or repair (common)
Raise bed to high position
- Test point
- Hilow function control
- Expected
- Bed reaches full high position
17 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Siderail switch not functioning patient safety
Nurse call, lighting, or bed control switch on siderail does not operate when activated
Ranked causes
- 1. Defective switch requiring replacement (common)
Verify new switch matches old switch before installation
- Test point
- Switch housing (G)
- Expected
- New and old switches are identical
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.