Hill-Rom CareAssist: troubleshooting and error codes
The ServiceForge knowledge base holds 22 documented symptoms for the Hill-Rom CareAssist, with 56 ranked causes, 99 diagnostic checks with test points and expected values, and 6 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
No functions work patient safety
Pendant and caregiver siderail controls do not respond; no bed movement or operations possible
Ranked causes
- 1. AC power fuses blown or interface PC board in-line fuse serviceable (common)
- 2. Cable disconnection on AC power or motor control PC board (common)
- 3. AC power cable not plugged in correctly (common)
- 4. Power outlet not functioning (occasional)
- 5. AC power cable damaged (occasional)
- 6. No AC voltage at transformer connection (occasional)
- 7. Interface PC board failure (occasional)
- 8. Patient pendant or siderail control failure (occasional)
Verify AC power fuses and interface PC board in-line cable fuse
- Test point
- AC power fuses, interface PC board in-line fuse
- Expected
- Serviceable (not blown)
7 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery backup not functioning caution
Bed loses power or does not operate when unplugged from AC power, or battery does not hold charge. Battery status indicator may not show charging or backup is unavailable.
Ranked causes
- 1. Battery expired or aged beyond 3 years (common)
- 2. Control unit malfunction or disconnected (occasional)
- 3. Battery disconnected or poorly connected (common)
- 4. Battery not charging properly or battery defective (occasional)
- 5. Power supply or motor control P.C. board failure (rare)
Inspect battery date and age
- Test point
- Battery date marking
- Expected
- Battery manufactured within 3 years of current date
4 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CPR malfunction patient safety
CPR (cardiopulmonary resuscitation) function not working; head section motor does not respond or release cable malfunction
Ranked causes
- 1. Damaged or missing mounting hardware on head section motor release mechanism (common)
- 2. Broken or damaged release cable (occasional)
- 3. Release cable adjustment incorrect (occasional)
- 4. Head section motor or release cable subassembly failure (rare)
Inspect head section motor release mechanism hardware
- Test point
- Release mechanism attachment points
- Expected
- All hardware secure and not damaged
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot section does not move or moves incorrectly caution
Foot section fails to extend or retract when controls activated, or moves intermittently. Knee or thigh section may be affected if motors are cross-connected.
Ranked causes
- 1. Control unit malfunction or disconnected (common)
- 2. Foot motor intermediate switch malfunction (occasional)
- 3. Foot motor intermediate switch open or short-circuited (occasional)
- 4. Power supply or motor control board failure (rare)
Verify at least one other function is working
- Test point
- Any bed function other than foot section
- Expected
- At least one other function operates normally
5 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SafeView alerts malfunction patient safety
SafeView alert lights flash, are on, or do not activate; yellow and green light indicators not working
Ranked causes
- 1. Both alerts flash yellow and green (common)
- 2. Only one alert flashes yellow and green (common)
- 3. Only one alert is on (common)
- 4. Alerts do not come on when bed plugged in (common)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed brake detection malfunction with AC power connected patient safety
Bed does not respond to brake detection when AC power is connected. Detection system fails or gives false indication of brake status. In-room alarm or alert may display.
Ranked causes
- 1. Brake detection switch connector disconnected or misaligned (common)
- 2. Brake detection switch malfunction or defective (occasional)
- 3. Foot column cable misconnection or poor connection (occasional)
Verify hilow function works
- Test point
- Bed height adjustment
- Expected
- Bed raises and lowers normally
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed connected to AC power, brakes not applied detection malfunction caution
Brake detection switch does not respond correctly when bed is powered and brakes applied
Ranked causes
- 1. Foot column cable connections or brake detection switch continuity fault (common)
- 2. Motor control board or power supply connector fault (occasional)
- 3. Power supply or motor control board failure (rare)
Verify foot column cable connections and brake detection switch
- Test point
- Terminals 1 and 6 of foot column, 8-pin connector in central rail
- Expected
- Connections secure and switch continuity correct
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head section does not move caution
Head section fails to raise or lower when controls are activated. Head Up or Head Down control buttons have no effect on head section position.
Ranked causes
- 1. Head section control unit malfunction or disconnected (common)
- 2. Head section motor malfunction or disconnected (occasional)
- 3. Power supply malfunction or failure (occasional)
Verify at least one other function is working
- Test point
- Any bed function other than head section
- Expected
- At least one other function (hi-lo, foot, etc.) operates normally
4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Knee section malfunction patient safety
Knee section does not respond to controls or moves incorrectly
Ranked causes
- 1. Faulty knee section control (common)
- 2. Faulty knee section motor (common)
- 3. Power supply or motor control PC board failure (occasional)
Verify at least one other function operates correctly
- Test point
- Other bed functions
- Expected
- At least one function works
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lockout malfunction patient safety
After pressing lockout button, no indicators flash
Ranked causes
- 1. Siderail cable connection incorrect in interface box or power supply (common)
- 2. Faulty siderail interface PC board (occasional)
- 3. Faulty caregiver siderail control panel (occasional)
Verify at least one other function operates correctly
- Test point
- Other bed functions
- Expected
- At least one function works
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Braking malfunction patient safety
Bed brakes do not hold when applied; casters do not lock when brake/steer bar is in brake position
Ranked causes
- 1. Damaged, loose or missing caster attachment screws or supports (common)
- 2. Caster lock mechanism failure (occasional)
Inspect caster attachment hardware
- Test point
- Caster attachment screws and supports
- Expected
- All screws secure and supports not damaged or missing
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hilow malfunction patient safety
Bed hilow section does not operate when controls are actuated
Ranked causes
- 1. Faulty hilow column motor or connector loose (common)
- 2. Power supply or motor control PC board failure (occasional)
Verify at least one other function is working
- Test point
- Other bed functions
- Expected
- At least one function operates
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Steering malfunction caution
Bed steering does not work correctly; foot left caster does not lock in lengthways direction when brake/steer bar is in steer position
Ranked causes
- 1. Damaged, loose or missing caster attachment screws or supports (common)
- 2. Caster steering mechanism failure (occasional)
Inspect caster attachment hardware
- Test point
- Caster attachment screws and supports
- Expected
- All screws secure and supports not damaged or missing
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Trendelenburg/Reverse Trendelenburg function does not work caution
Bed does not tilt head down or feet down when Trendelenburg/Reverse Trendelenburg controls are activated. Bed remains in flat position.
Ranked causes
- 1. Caregiver control unit malfunction or disconnected (common)
- 2. Power supply or motor control board failure (occasional)
Verify hilow function works
- Test point
- Bed height adjustment
- Expected
- Bed raises and lowers normally
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery replacement on A Model beds caution
Battery backup system requires polarities respected during installation to avoid power supply P.C. board damage
Ranked causes
- 1. Battery polarities reversed during installation (common)
Verify correct polarity during battery installation
- Test point
- Battery connection points
- Expected
- Red wire connected last with spark observed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: SafeView Alerts light assembly configuration
Configuration procedure to set which siderails must be up for lights to display green.
Ranked causes
- 1. Initial system configuration needed after installation (common)
Position siderails for desired configuration
- Test point
- Two head-end siderails and optionally one or both foot-end siderails
- Expected
- All configured siderails in up position
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Scale patient safety
Bed scale requires calibration to ensure accurate weight measurement. Performed after load beam replacement or scale board service.
Ranked causes
- 1. Load beam replaced or scale P.C. board changed (common)
Verify sleep deck is flat
- Test point
- Bed surface
- Expected
- Flat position
30 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CPR cable removal and replacement patient safety
CPR cable system requires correct routing and tension adjustment to prevent patient injury
Ranked causes
- 1. CPR cable incorrect routing or tension (common)
Route CPR cable through head section guide correctly
- Test point
- Head section guide bracket, handle connections
- Expected
- Right side cable to right handle, left side to left handle
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head section motor removal and replacement high voltage
Motor replacement procedure requiring polarity reversal, connector disconnection, and cable adjustment
Ranked causes
- 1. Head section motor defective (unknown)
Set brakes and raise sleep surface to highest
- Test point
- Bed controls
- Expected
- Sleep surface fully raised, brakes engaged
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Knee section motor removal and replacement high voltage
Motor replacement procedure on knee section with power supply access dependent on A or B model
Ranked causes
- 1. Knee section motor defective (unknown)
Set brakes and raise sleep surface to highest
- Test point
- Bed controls
- Expected
- Sleep surface fully raised, brakes engaged
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SafeView Alerts lights flash yellow and green after installation
Lights display incorrect color pattern (yellow and green flashing) after new light assembly installation, indicating a technical problem.
Ranked causes
- 1. Incorrect part installation or cable connection (common)
Verify light assembly bracket alignment and screw installation
- Test point
- Light assembly bracket (H) screw holes to bed end
- Expected
- Screw holes aligned and all three screws (I) installed
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scale P.C. board replaced patient safety
Replaced scale P.C. board requires calibration. Scale display will show kilograms until calibrated and converted to pounds.
Ranked causes
- 1. Scale P.C. board changed (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Scale Error 2 | Scale error condition on C model and newer beds | Scale measurement or calibration failure | unknown |
Scale Error 3 | Scale error condition on C model and newer beds | Scale measurement or calibration failure | unknown |
Scale Error 4 | Scale error condition on C model and newer beds | Scale measurement or calibration failure | unknown |
Scale Error 5 | Scale error condition on C model and newer beds | Scale measurement or calibration failure | unknown |
Service Required LED Flashes | Service required indicator active on C model and newer beds | System condition requiring technician service | warning |
Siderail Detection Switch Malfunction | Siderail detection switch failure on F model beds | Siderail switch not detecting position or stuck | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI