Hill-Rom Affinity Three: troubleshooting and error codes
The ServiceForge knowledge base holds 69 documented symptoms for the Hill-Rom Affinity Three, with 127 ranked causes, 184 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Battery back-up not providing power patient safety
Battery back-up system not functioning or unable to power bed during mains power loss
Ranked causes
- 1. Battery not properly charged (common)
- 2. Battery fuse open or failed (common)
- 3. Loose cable connections or wiring (occasional)
- 4. Battery charge circuit voltage out of range (occasional)
- 5. Transformer assembly failure (occasional)
- 6. Logic control PCB failure (occasional)
- 7. Battery cells failed or depleted (occasional)
Verify battery is charged
- Test point
- Battery assembly
- Expected
- Battery properly charged and providing power
6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed air surface malfunction patient safety
Bed air surface does not function properly; air inflate, deflate, or mattress functions fail
Ranked causes
- 1. Faulty siderail switch board (common)
- 2. Logic control board output failure (occasional)
- 3. Cable or connector failure between boards (occasional)
- 4. Air compressor failure (occasional)
- 5. Air system control board failure (occasional)
- 6. Manifold, hose, or bladder obstruction/leak (occasional)
Test individual air function switches on both siderails
- Test point
- Air surface control buttons
- Expected
- Only one switch non-functional or all switches in one siderail fail
9 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head motor does not raise patient safety
Head section fails to rise when head up switch activated on siderail
Ranked causes
- 1. Bed unplugged or lockout switch on (common)
- 2. Sheet or blanket caught in CPR release (occasional)
- 3. Head up switch on siderail faulty (occasional)
- 4. Logic control board faulty (occasional)
- 5. Head motor overheated or failed (occasional)
- 6. Logic control board voltage fault (rare)
Verify bed power and lockout status
- Test point
- Power plug and lockout switch
- Expected
- Bed plugged in, lockout switch off
6 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot section does not raise patient safety
Foot motor fails to lower foot section when foot down switch is activated on birthing bed
Ranked causes
- 1. Power supply issue or lockout switch on (common)
- 2. Faulty siderail switch control board (occasional)
- 3. Logic control board relay failure (occasional)
- 4. Foot motor thermal overload or failure (occasional)
- 5. Logic control board voltage output failure (occasional)
Verify power and lockout switch status
- Test point
- Power plug and lockout switch
- Expected
- Bed plugged in, lockout switch off
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head motor does not lower patient safety
Head section fails to lower when head down switch activated on siderail
Ranked causes
- 1. Bed unplugged or lockout switch on (common)
- 2. Head down switch on siderail faulty (occasional)
- 3. Logic control board faulty (occasional)
- 4. Head motor overheated or failed (occasional)
- 5. Logic control board voltage fault (rare)
Verify bed power and lockout status
- Test point
- Power plug and lockout switch
- Expected
- Bed plugged in, lockout switch off
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hilow down function does not lower bed
Bed does not lower when hilow down switch is activated on siderail control
Ranked causes
- 1. Bed not plugged in or lockout switch is on (common)
- 2. Faulty hilow down switch or switch board (common)
- 3. Logic control P.C. board relay failure (occasional)
- 4. Hilow motor failure or overheating (occasional)
- 5. Logic control P.C. board voltage output failure (rare)
Verify bed is plugged in and lockout switch is off
- Test point
- Power and lockout switch
- Expected
- Bed plugged in, lockout switch off
6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hilow up function does not raise bed
Bed does not raise when hilow up switch is activated on siderail control
Ranked causes
- 1. Bed not plugged in or lockout switch is on (common)
- 2. Faulty hilow up switch or switch board (common)
- 3. Logic control P.C. board relay failure (occasional)
- 4. Hilow motor failure or overheating (occasional)
- 5. Logic control P.C. board voltage output failure (rare)
Verify bed is plugged in and lockout switch is off
- Test point
- Power and lockout switch
- Expected
- Bed plugged in, lockout switch off
6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot drive operation problems
Foot motor runs but exhibits noise, drift, or other operational issues
Ranked causes
- 1. Physical damage to drive system (occasional)
- 2. Improperly installed clevis or cotter pins (occasional)
- 3. Foot motor damaged under load (occasional)
- 4. Foot motor drift or loss of position hold (occasional)
Inspect drive system for physical damage
- Test point
- Motor shaft, mounting, fasteners
- Expected
- No bent shaft, loose parts, or visible damage
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot motor does not raise patient safety
Foot section fails to rise when foot up switch activated on siderail
Ranked causes
- 1. Bed unplugged or lockout switch on (common)
- 2. Foot up switch on siderail faulty (occasional)
- 3. Logic control board faulty (occasional)
- 4. Foot motor overheated or failed (occasional)
Verify bed power and lockout status
- Test point
- Power plug and lockout switch
- Expected
- Bed plugged in, lockout switch off
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head motor runs but has operational problems caution
Motor is noisy, drifts, coasts, or shows other performance issues during head section raise/lower operation
Ranked causes
- 1. Physical damage to drive system (bent motor shaft, loose metal) (common)
- 2. Clevis pins or cotter pins not installed properly on head motor drive (common)
- 3. Head motor damaged under load (occasional)
- 4. Head motor unable to retain position or drifts when powered off (occasional)
Visually inspect drive system for physical damage
- Test point
- Head motor shaft and drive mechanism
- Expected
- No bent shaft, no loose metal pieces
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hilow motor runs but has operational problems caution
Motor is noisy, drifts, coasts, or shows other performance issues during raise/lower operation
Ranked causes
- 1. Physical damage to drive system (bent motor shaft, loose metal) (common)
- 2. Clevis pins or cotter pins not installed properly on hilow motor drive (common)
- 3. Hilow motor damaged under load (occasional)
- 4. Hilow motor unable to retain position or drifts when powered off (occasional)
Visually inspect drive system for physical damage
- Test point
- Motor shaft and drive mechanism
- Expected
- No bent shaft, no loose metal pieces
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Night light does not illuminate
Night light fails to turn on when ambient light dims
Ranked causes
- 1. Lamp bulb or fixture damage (common)
- 2. Photocell malfunction (occasional)
- 3. Loose or disconnected lamp cable (occasional)
- 4. Logic control board failure (rare)
Test night light function with dimmed ambient light
- Test point
- Night light and ambient light sensor
- Expected
- Night light comes on when ambient light dims
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Siderail does not lock in raised position patient safety
Siderail does not click into locked position when raised from stow position, or does not remain latched when locked
Ranked causes
- 1. Obstructions or missing hardware in siderail mechanism (common)
- 2. Obstruction or loose fasteners preventing latch engagement (common)
- 3. Obstruction in center arm, loose fasteners, missing spring, or binding (occasional)
- 4. Interference from another bed component (rare)
Raise siderail from stow position and verify locking
- Test point
- Siderail assembly
- Expected
- Siderail rises and clicks into locked position
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Brake caster does not lock or roll caution
Brake caster pedal locks but casters do not roll, or pedal does not lock when pressed
Ranked causes
- 1. Excessive caster wear (common)
- 2. Caster assembly misalignment (occasional)
- 3. Mechanical interference with brake (rare)
Test brake pedal lock
- Test point
- Brake caster pedal
- Expected
- Pedal locks in brake position when pressed
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CPR release does not flatten head section patient safety
CPR release handle pulled but head section does not go to full flat position
Ranked causes
- 1. CPR function requires adjustment (common)
- 2. Head drive motor assembly damaged or missing parts (occasional)
- 3. Interference from another bed component (rare)
Pull and hold CPR release handle
- Test point
- CPR release handle
- Expected
- Head section goes to full flat position
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Brake and steer caster adjustment needed
Caster does not lock properly or brakes slip; ratcheting or improper steer position detected
Ranked causes
- 1. Brake setscrew tension incorrect (common)
- 2. Wax buildup on caster tire (occasional)
Brake/steer pedal in neutral position
- Test point
- Brake/steer pedal
- Expected
- Neutral position
6 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Brake caster does not hold position
Brake caster tire rolls when brake is engaged or brake slips during normal operation
Ranked causes
- 1. Brake setscrew insufficient tension (common)
- 2. Wax on brake caster tire (occasional)
Brake setscrew adjustment - initial setup
- Test point
- Setscrew (D)
- Expected
- Light ratcheting then stops when adjusted
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Battery assembly replacement high voltage
Replace batteries that maintain bed function; located in electronics pan with cable connections to logic control board P5 and battery connection points
Ranked causes
- 1. Battery cable assembly requires reconnection to logic control board (common)
- 2. Battery connection points require reconnection (common)
Set brake/steer pedal to brake position
- Test point
- brake/steer pedal
- Expected
- in brake position
7 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot section not securely mounted during slide-off replacement patient safety
Foot section (slide-off type) may detach from bed after installation due to improper mounting on yoke slide brackets or latch mechanism
Ranked causes
- 1. Foot section channel slides not aligned with mounting brackets (common)
- 2. Latch mechanism not locked into channel slide (occasional)
Align foot section channel slides with mounting brackets before sliding
- Test point
- channel slides (F) and mounting brackets (G)
- Expected
- slides aligned with brackets
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air system lumbar section does not deflate caution
Lumbar section fails to deflate when back deflate button is pressed
Ranked causes
- 1. Air system fault (unknown)
Press back deflate button and activate bed air surface
- Test point
- lumbar section
- Expected
- Deflates
Air system lumbar section does not inflate caution
Lumbar section fails to inflate when back inflate button is pressed
Ranked causes
- 1. Air system fault (unknown)
Press back inflate button and activate bed air surface
- Test point
- lumbar section
- Expected
- Inflates
Air system seat section does not deflate caution
Seat section mattress fails to deflate when seat deflate button is pressed and held
Ranked causes
- 1. Air system fault (unknown)
Press and hold seat deflate button and activate bed air surface
- Test point
- seat section mattress
- Expected
- Deflates
Air system seat section does not inflate caution
Seat section mattress fails to inflate when seat inflate button is pressed
Ranked causes
- 1. Air system fault (unknown)
Press seat inflate button and activate bed air surface
- Test point
- seat section mattress
- Expected
- Inflates
Auxiliary outlet assembly replacement high voltage
Replacement of auxiliary outlet assembly including cable routing restoration and connector verification.
Ranked causes
- 1. Auxiliary outlet assembly requires replacement (unknown)
Note cable routing and orientation before removal
- Test point
- Auxiliary outlet cable assembly (C) routing under seat section
- Expected
- Cable path documented for reinstallation
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auxiliary transformer assembly replacement high voltage
Replacement of auxiliary transformer with electrical safety precautions and verification of proper installation.
Ranked causes
- 1. Auxiliary transformer assembly requires replacement (unknown)
Ensure bed positioned correctly before service
- Test point
- Brake/steer pedal, bed position, head and hilow functions
- Expected
- Brake engaged, out of Trend-Like position, head and bed raised to high
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery backup does not provide sufficient power patient safety
Battery backup fails to operate bed functions when unplugged from power source
Ranked causes
- 1. Battery backup fault or depletion (common)
Unplug bed from power source; activate bed functions; verify battery operates functions
- Test point
- battery backup during unplugged operation
- Expected
- Battery backup provides sufficient power to operate bed functions
Bed does not auto-rise from low position when foot lowered patient safety
Bed fails to automatically rise from low position when foot is lowered from low position
Ranked causes
- 1. Automatic rise mechanism fault (unknown)
Lower bed to low position; lower foot section with foot down switch; observe auto-rise
- Test point
- bed at low position with lowered foot
- Expected
- Bed automatically rises from low position
Bed does not lower with hilow down switch patient safety
Bed fails to lower when hilow down switch is momentarily activated
Ranked causes
- 1. Hilow motor or drive system fault (unknown)
Momentarily activate the hilow down switch
- Test point
- hilow down switch
- Expected
- Bed lowers
Bed does not rise with hilow up switch patient safety
Bed fails to rise when hilow up switch is momentarily activated
Ranked causes
- 1. Hilow motor or drive system fault (unknown)
Momentarily activate the hilow up switch
- Test point
- hilow up switch
- Expected
- Bed rises
Brake and steer functions do not operate properly caution
Brake/steer pedal functions fail to operate or position incorrectly
Ranked causes
- 1. Brake and steer function adjustment needed (common)
Place brake steer pedal into brake position and then steer position
- Test point
- brake steer pedal
- Expected
- Brake and steer functions operate properly
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Battery charge circuit adjustment high voltage
Beds with 65345 logic control and 67208 interface PCBs require adjustment of battery charge circuit voltage for optimal battery function. Measurement taken at potentiometer R2 on logic control PCB.
Ranked causes
- 1. Battery charge circuit voltage out of specification (common)
Measure voltage at positive side of D15 (test point K) with ground at negative side of C1 (test point J)
- Test point
- D15 positive side (K) to C1 negative side (J)
- Expected
- 27.6 to 28.0 V DC, +/- 0.2 V
- Tools
- VOM set to V DC
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Foot Limit Switch caution
Foot section does not reach correct minimum position or foot travel limits are incorrect. Verified by measuring gap between foot and hilow lift arms.
Ranked causes
- 1. Foot limit switch (LFD or LFU) out of adjustment (common)
Measure gap between foot and hilow lift arms at foot low position
- Test point
- Between foot section left-hand lift arm (B) and hilow left-hand lift arm (C)
- Expected
- Exactly ¼" (6.4 mm) in, Exact
- Tools
- Tape measure
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Hilow Limit Switch caution
Bed does not reach correct maximum raised height or hilow travel limits are incorrect. Verified by measuring distance from top center of main frame to floor.
Ranked causes
- 1. Hilow limit switch (LBU or LBD) out of adjustment (common)
Measure bed height at maximum raised position
- Test point
- Top center point of main frame at head end to floor
- Expected
- 31" ± ½" (787.4 ± 12.7 mm) in, ±½"
- Tools
- Tape measure
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Line voltage fuse replacement high voltage
Replace mains fuses protecting high voltage circuit; fuses located in fuse holder at head end of main frame and on air system control P.C. board
Ranked causes
- 1. Serviceable fuse required in fuse housing (common)
Set brake/steer pedal to brake position
- Test point
- brake/steer pedal
- Expected
- in brake position
9 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Caregiver siderail PC switch board malfunction high voltage
Switch board on outboard siderail not functioning or requires replacement
Ranked causes
- 1. Faulty PC switch board (unknown)
Brake/steer pedal in brake position
- Test point
- Brake/steer pedal
- Expected
- Set to brake position
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Caster assembly worn or damaged
Caster does not roll or swivel freely; requires replacement after removal
Ranked causes
- 1. Caster bearing worn or debris trapped (occasional)
Brake/steer pedal in brake position
- Test point
- Brake/steer pedal
- Expected
- Brake position
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Central braking system malfunction
Brake and steer pedal functions not working; system requires removal and replacement of brake mechanism
Ranked causes
- 1. Brake strip or rocker arm failure (rare)
Siderail secured before bed is turned
- Test point
- Siderail assembly
- Expected
- Secured below bed or removed
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CPR function does not lower head to flat in time patient safety
Head section fails to lower to flat position within 7 seconds when CPR release lever activated
Ranked causes
- 1. CPR release cables require adjustment (common)
Distribute 50 lb weight on head section; activate CPR release lever; measure time to flat
- Test point
- head section with 50 lb load
- Expected
- Lowers to flat position within approximately 7 seconds
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot motor does not lower patient safety
Foot section fails to lower when foot down switch activated on siderail
Ranked causes
- 1. Foot down switch or logic control board faulty (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot motor excessive noise or binding caution
Foot motor emits excessive noise or exhibits mechanical binding/grinding when activated
Ranked causes
- 1. Foot motor mechanical fault or obstruction (unknown)
Activate foot up or down switch and listen for noise
- Test point
- foot motor
- Expected
- Free of excessive noise or mechanical binding/grinding
Foot section does not lower with foot down switch patient safety
Foot section fails to lower when foot down switch is pressed and activated
Ranked causes
- 1. Foot motor or drive system fault (unknown)
Press foot down switch and activate foot section
- Test point
- foot down switch
- Expected
- Foot section lowers
Foot section does not rise with foot up switch patient safety
Foot section fails to rise when foot up switch is pressed and activated
Ranked causes
- 1. Foot motor or drive system fault (unknown)
Press foot up switch and activate foot section
- Test point
- foot up switch
- Expected
- Foot section rises
Foot section lower limit incorrect patient safety
Foot drive does not stop at correct low position limit; gap from hilow lift arm incorrect
Ranked causes
- 1. Foot limit switch requires adjustment (common)
Lower foot section to low position; measure gap to foot end hilow lift arm
- Test point
- foot lift arm above foot end hilow lift arm
- Expected
- 1/4" (6.4 mm) above foot end hilow lift arm inches, exact
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot section not securely mounted during lift-off replacement patient safety
Foot section (lift-off type) may detach from bed after installation due to improper mounting on yoke latches
Ranked causes
- 1. Foot section slides not aligned with receptacles (common)
Align foot section slides with receptacles before mounting
- Test point
- foot section slides (D) and receptacles (F)
- Expected
- slides aligned with receptacles
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot section stops at incorrect limit position patient safety
Foot section does not stop at correct high position limit or gap from seat incorrect
Ranked causes
- 1. Foot limit switch requires adjustment (common)
Raise foot section to high position using foot up switch; measure gap to seat section
- Test point
- top of foot section above seat section
- Expected
- Approximately 2-1/2" (6.35 cm) above seat section inches
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot support cable binds or does not move freely caution
Cable ends bind in release handle mounting holes, restricting smooth operation of foot support release handle
Ranked causes
- 1. Cable tensioning nuts (F) too tight (common)
Verify cable ends move freely in release handle mounting holes
- Test point
- cable ends in release handle weldment (G) mounting holes
- Expected
- ride freely without binding
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Head motor excessive noise or binding caution
Head motor emits excessive noise or exhibits mechanical binding/grinding when activated
Ranked causes
- 1. Head motor mechanical fault or obstruction (unknown)
Activate head up or down switch and listen for noise
- Test point
- head motor
- Expected
- Free of excessive noise or mechanical binding/grinding
Head section does not lower with head down switch patient safety
Head section fails to lower when head down switch is momentarily activated
Ranked causes
- 1. Head motor or drive system fault (unknown)
Momentarily activate the head down switch
- Test point
- head down switch
- Expected
- Head section lowers
Head section does not rise with head up switch patient safety
Head section fails to rise when head up switch is momentarily activated
Ranked causes
- 1. Head motor or drive system fault (unknown)
Momentarily activate the head up switch
- Test point
- head up switch
- Expected
- Head section rises
Head section drifts downward after raising patient safety
Head section drifts downward or coasts after being raised with head function
Ranked causes
- 1. Head motor or hold mechanism fault (unknown)
Raise head section to high position; observe drift
- Test point
- head section
- Expected
- Stops and does not drift downward or coast
Hilow bed height limit incorrect at low position patient safety
Bed does not stop at correct low position limit or height measurement outside specification
Ranked causes
- 1. Hilow down limit switch (LBD) requires adjustment (common)
Lower bed to low position; measure bed height from floor
- Test point
- bed height from floor
- Expected
- 16" ± ½" (41 cm ± 12.7 mm) inches, ± ½ inch
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hilow bed height limit incorrect at top position patient safety
Bed does not stop at correct high position limit or height measurement outside specification
Ranked causes
- 1. Hilow up limit switch (LBU) requires adjustment (common)
Raise bed to high position; measure from floor to top of mainframe
- Test point
- side of bed from floor to top of mainframe
- Expected
- 31" ± ½" (79 cm ± 12.7 mm) inches, ± ½ inch
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hilow does not raise bed when foot lowered patient safety
Hilow function fails to automatically raise bed from low position when foot section is lowered to lowest point
Ranked causes
- 1. Hilow or foot limit switch malfunction (common)
Test hilow function from siderails and pendant control
- Test point
- Hilow control switches
- Expected
- Hilow function works properly
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hilow motor excessive noise or binding caution
Hilow motor emits excessive noise or exhibits mechanical binding/grinding when activated
Ranked causes
- 1. Hilow motor mechanical fault or obstruction (unknown)
Activate hilow up and down switch and listen for noise
- Test point
- hilow motor
- Expected
- Free of excessive noise or mechanical binding/grinding
Hilow sleep surface drifts downward after raising patient safety
Sleep surface drifts downward or coasts after being raised with hilow function
Ranked causes
- 1. Hilow motor or hold mechanism fault (unknown)
Raise and lower sleep surface using hilow function; observe drift
- Test point
- hilow sleep surface
- Expected
- Surface stops and does not drift downward or coast
Labor grips do not latch or release properly patient safety
Labor grips fail to latch in full up position or do not release when release handle activated
Ranked causes
- 1. Faulty labor grips (common)
Operate labor grips on both sides; raise to full up position; activate release handle
- Test point
- labor grips on both sides
- Expected
- Grips latch in full up position; release when handle activated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lockout switch assembly replacement high voltage
Replacement of lockout switch assembly with electrical safety precautions and connector verification.
Ranked causes
- 1. Lockout switch assembly requires replacement (unknown)
Verify unit unplugged and lockout engaged
- Test point
- Power source, lockout control
- Expected
- Unit unplugged, lockout control engaged
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lockout switch does not control functions properly caution
Lockout switch fails to lock/unlock inboard/outboard controls in OFF/ON positions
Ranked causes
- 1. Faulty rocker switch (common)
Place lockout switch in OFF position (locked); verify controls locked; place in ON position (unlocked); verify controls enabled
- Test point
- lockout switch in both OFF and ON positions
- Expected
- Functions lock in OFF position; functions enable in ON position
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Logic control P.C. board connector wiring order incorrect high voltage
Connector wiring sequence needs reversal for 68121 board compatibility; red, black, red, black alternating pattern required.
Ranked causes
- 1. Original connector wiring order incompatible with 68121 logic control board (common)
Verify connector shell and backshell locked together
- Test point
- Locking tabs (D)
- Expected
- Tabs locked, holding connector housing firmly together
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Logic control P.C. board replacement high voltage
Replacement of logic control P.C. board assembly with proper connector connections and post-installation verification.
Ranked causes
- 1. Logic control P.C. board requires replacement (unknown)
Verify wiring connectors firmly seated on logic control board receptacles
- Test point
- All connectors on P.C. board
- Expected
- Connectors fully inserted and secured
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Patient siderail PC switch board malfunction
- Siderail assembly replacement
- Siderail does not latch or lower properly
- Siderail drops when pins removed
- Siderail functions do not operate properly
- Transformer assembly replacement
- Trend-Like position cannot be set
- Trend-Like position does not operate smoothly or sleep surface insecure
- Wiring diagram installation and documentation
9 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI