Hill-Rom Advanta 2: troubleshooting and error codes
The ServiceForge knowledge base holds 18 documented symptoms for the Hill-Rom Advanta 2, with 30 ranked causes, 43 diagnostic checks with test points and expected values, and 9 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
SafeView Alerts malfunction patient safety
Alerts flash yellow and green, flash only one color, remain on, or do not illuminate when bed is plugged in
Ranked causes
- 1. Problem with SideCom P.C. board, Siderail Interface P.C. board, control pod P.C. board, or HOB sensor P.C. board (common)
- 2. Alert cable loose or disconnected (common)
- 3. Alert cable damaged (occasional)
- 4. P.C. board within Alert assembly damaged (occasional)
- 5. Siderail Interface P.C. board problem (only one Alert on) (occasional)
- 6. Alerts deactivated (occasional)
- 7. Bed Exit System not activated (occasional)
- 8. Both Alert cables loose or disconnected (common)
- 9. Both Alert cables damaged (occasional)
9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed will not move or respond to handle input caution
IntelliDrive transport system not responding to control handle input. LED DS1 on PACM board may be off, flashing, or showing abnormal behavior.
Ranked causes
- 1. Controller not powered (common)
- 2. Enable switches not working (common)
- 3. Controller detects internal fault (occasional)
- 4. Throttle signal not working correctly (occasional)
- 5. Controller LED flashes indicating specific fault code (unknown)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Antistatic precautions during sensor replacement caution
Requirement to wear and properly ground antistatic wrist strap to prevent equipment damage
Ranked causes
- 1. Antistatic protection not used during component handling (common)
Verify antistatic wrist strap is worn
- Test point
- Technician wrist
- Expected
- Strap securely fastened
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auxiliary outlet continuity fault high voltage
Electrical continuity detected between outlet Line, Neutral, or Ground lines and bed frame
Ranked causes
- 1. Outlet assembly or wiring insulation compromised (rare)
Measure continuity from outlet Line to bed frame
- Test point
- Auxiliary outlet Line terminal to bed frame
- Expected
- No continuity (open circuit)
- Tools
- Multimeter
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auxiliary outlet power cord interference with bed frame caution
Outlet power cord contacts or interferes with bed frame during head, knee, or foot section movement
Ranked causes
- 1. Power cord routed incorrectly or not secured properly (common)
Lower head section and inspect for interference
- Test point
- outlet power cord (A) routing along intermediate frame
- Expected
- No contact or tension on power cord
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed Not Down indicator does not flash when obstacle detected patient safety
Obstacle Detect™ Sensor is squeezed but Bed Not Down indicator fails to flash and bed continues to descend or does not stop as expected
Ranked causes
- 1. Obstacle Detect™ Sensor malfunction or misalignment (unknown)
Verify indicator response when sensor squeezed during bed descent
- Test point
- Bed Not Down indicator and Obstacle Detect™ Sensor
- Expected
- Indicator flashes; bed stops then raises for 2 seconds
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed Not Down indicator flashes when obstacle detected patient safety
When Obstacle Detect™ Sensor is squeezed while pressing Bed Down control, the indicator flashes and bed does not descend or stops mid-motion then raises for 2 seconds
Ranked causes
- 1. Obstacle Detect™ Sensor activation (common)
Squeeze and hold Obstacle Detect™ Sensor on bed side
- Test point
- Obstacle Detect™ Sensor location
- Expected
- Bed Not Down indicator flashes; bed does not go down
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CPR cable tension patient safety
CPR release mechanism movement and tactile feedback during CPR handle operation must be adjusted so release engages at correct handle travel point
Ranked causes
- 1. CPR cable tension incorrect after motor replacement or normal wear (occasional)
Verify CPR release mechanism movement point during handle travel
- Test point
- CPR handle and head motor release mechanism
- Expected
- Release mechanism begins to move from neutral position when handle is approximately halfway through travel
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Head section motor elastic position caution
New head section motors have elastic band that indicates correct position of internal switches and end of travel stops. Elastic must be removed and motor position verified.
Ranked causes
- 1. New motor installed with elastic band still attached or incorrectly positioned (common)
Verify elastic band removed from motor body
- Test point
- Motor body, rod end
- Expected
- No elastic band wound around motor; motor rotates freely
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Sleep deck section installation patient safety
Foot, thigh, seat and head sections must be installed in correct sequence and position for proper function and patient safety
Ranked causes
- 1. Sections installed in incorrect sequence or head section not fully up during seat installation (occasional)
Verify head section position before seat section installation
- Test point
- Head section hinge points and full up position
- Expected
- Head section fully raised and secured
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fifth wheel does not lock in steer position
Fifth wheel fails to lock aligned with bed frame when brake/steer pedals placed in Steer position
Ranked causes
- 1. Control cable tension incorrect (common)
Verify fifth wheel operation in Neutral position
- Test point
- brake/steer pedals
- Expected
- 5th wheel turns freely
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Knee section motor replacement high voltage
Motor removal and installation procedure for Advanta 2 bed knee section actuation
Ranked causes
- 1. Knee section motor failure or malfunction requiring replacement (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Obstacle Detect sensor malfunction during bed lowering patient safety
Bed Not Down indicator does not flash or bed does not stop/raise when sensor is squeezed during descent; sensor fails to trigger protective response
Ranked causes
- 1. Obstacle Detect sensor not functioning or disconnected (unknown)
Squeeze Obstacle Detect sensor while bed is lowering
- Test point
- Obstacle Detect sensor on bed side
- Expected
- Bed Not Down indicator flashes, bed stops then raises for 2 seconds, and stops
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Obstacle Detect sensor malfunction during function check patient safety
Sensor does not trigger Bed Not Down indicator or bed control response when squeezed; bed does not stop or raise as expected during lowering
Ranked causes
- 1. Obstacle Detect sensor failure or obstruction (unknown)
Squeeze and hold Obstacle Detect sensor on bed side
- Test point
- Obstacle Detect Sensor
- Expected
- Bed Not Down indicator flashes
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Obstacle Detect™ Sensor malfunction patient safety
Sensor fails to prevent bed from lowering when obstructed, or bed does not respond correctly to sensor activation during lowering cycle
Ranked causes
- 1. Obstacle Detect™ Sensor not functioning (unknown)
Activate Obstacle Detect™ Sensor and press Bed Down control
- Test point
- Obstacle Detect™ Sensor
- Expected
- Bed Not Down indicator flashes, bed does not go down
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Obstacle Detect™ Sensor replacement caution
Sensor strip removal and reinstallation on Advanta 2 hospital bed
Ranked causes
- 1. Sensor strip requires replacement or maintenance (unknown)
Set bed brakes
- Test point
- Brake control
- Expected
- Brakes engaged
7 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SafeView lights flash yellow and green after installation
After installing new light assemblies and connecting to AC power, the indicator lights flash yellow and green instead of behaving normally
Ranked causes
- 1. Incorrect part installation or cable connection (common)
Verify light assembly cables properly connected to siderail interface P.C. board
- Test point
- Connectors P38 and P39
- Expected
- Cables fully seated with no gaps
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SafeView lights not responding to bed movements and siderail position changes
Lights do not flash yellow during bed raise/lower or siderail lowering, or do not stay green in safe condition
Ranked causes
- 1. Bed Exit Alarm System not activated (common)
Verify Bed Exit Alarm System is activated
- Test point
- Alarm System control
- Expected
- System activated and ready
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
CAL Hd | Head angle calibration error | Head angle calibration procedure failure | error |
Hd Err | Head angle display error | Head angle sensor or display malfunction | error |
Scale Error 0 | Scale error condition 0 | Unknown condition related to scale operation | error |
Scale Error 1 | Scale error condition 1 | Unknown condition related to scale operation | error |
Scale Error 2 | Scale error condition 2 | Unknown condition related to scale operation | error |
Scale Error 3 | Scale error condition 3 | Unknown condition related to scale operation | error |
Scale Error 4 | Scale error condition 4 | Unknown condition related to scale operation | error |
Scale Error 5 | Scale error condition 5 | Unknown condition related to scale operation | error |
Scale Error 9 | Scale error condition 9 | Unknown condition related to scale operation | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI