The ServiceForge knowledge base holds 76 documented symptoms for the Hettich Rotanta 46 RSC, with 209 ranked causes, 106 diagnostic checks with test points and expected values, and 105 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Imbalance on motor axle caution

Centrifuge slows down until lid release occurs; further cooling continues to SET temperature

Ranked causes

  1. 1. Weight difference in rotor components (common)
  2. 2. Supporting lugs not lubricated (occasional)
  3. 3. Wrong IMBALANCE MODE set (occasional)
  4. 4. Imbalance switch S1 not connected (occasional)
  5. 5. Imbalance switch S1 defective (occasional)
  6. 6. Loose contact in cable or plug (occasional)
  7. 7. Flat ribbon cable to control panel defective (occasional)
  8. 8. Control panel or supply board defective (occasional)

Hatch drive motor timeout closing hatch caution

Hatch drive motor activated but did not fully close hatch within 30 seconds. Hatch end switch must report 'closed' within timeout. Further operation is not possible.

Ranked causes

  1. 1. Hatch end switch (S7) defective (common)
  2. 2. Hatch drive motor (M6) defective (occasional)
  3. 3. Loose connection at hatch drive motor plug (M6) (occasional)
  4. 4. Fuse F1 on positioning board (A14) blown (occasional)
  5. 5. Defective positioning board (A14) (occasional)
  6. 6. Flat ribbon cable to control panel (A5) defective (occasional)
  7. 7. Defective control panel (A5) (occasional)
First check

Measure voltage at hatch end switch (S7)

Test point
Positioning board (A14) measuring point MP1
Expected
Voltage change when hatch closes

1 more check with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Hatch drive motor timeout opening hatch caution

Hatch drive motor activated but did not fully open hatch within 30 seconds. Hatch end switch must report 'opened' within timeout. Further operation is not possible.

Ranked causes

  1. 1. Hatch end switch (S6) defective (common)
  2. 2. Hatch drive motor (M6) defective (occasional)
  3. 3. Loose connection at hatch drive motor plug (M6) (occasional)
  4. 4. Fuse F1 on positioning board (A14) blown (occasional)
  5. 5. Defective positioning board (A14) (occasional)
  6. 6. Flat ribbon cable to control panel (A5) defective (occasional)
  7. 7. Defective control panel (A5) (occasional)
First check

Measure voltage at hatch end switch (S6)

Test point
Positioning board (A14) measuring point MP2
Expected
Voltage change when hatch opens

1 more check with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No speedometer pulses after start command caution

Rotor slows down and stops after start; DC braking for 30 sec then release lid; further cooling continues

Ranked causes

  1. 1. Start implemented without rotor (common)
  2. 2. Motor not connected (occasional)
  3. 3. Motor defective (occasional)
  4. 4. Speed sensor defective or loose contact in plug (common)
  5. 5. Flat ribbon cable to control panel or frequency converter defective (occasional)
  6. 6. No enable signal at frequency converter (occasional)
  7. 7. Supply board, control panel, or frequency converter defective (occasional)
First check

Verify rotor installed before start

Test point
Rotor presence
Expected
Rotor present

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Position 1-4 not reached within 80 seconds

Rotor turns and speedometer pulses occur, but control panel does not recognize desired position within 80 seconds. Positioning information not plausible on tracks A or B, or brake defective. Further operation is not possible.

Ranked causes

  1. 1. Defective incremental encoder (B7) (common)
  2. 2. Loose connection at incremental encoder plug (B7) (common)
  3. 3. Fuse F1 on positioning board (A14) blown (occasional)
  4. 4. Magnetic brake (Y2) defective (occasional)
  5. 5. Defective positioning board (A14) (occasional)
  6. 6. Defective electronic interface/positioning (A5.1) (occasional)
  7. 7. Defective control panel (A5) (occasional)
First check

Measure encoder signals with oscilloscope

Test point
Electronic interface/positioning (A5.1) at encoder plug X3: CH A (pin 4, green wire), CH B (pin 5, yellow wire), GND (pi
Expected
5 V signals shifted 90 degrees on Channels A and B
Tools
Oscilloscope with 2V/DIV sensitivity

1 more check with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Insufficient speed - motor slippage too large

Motor slippage exceeds control adjustment limit of 5%. Rotor speed (ACTUAL) is lower than SET speed minus 5%. Centrifuge slows down and cooling continues.

Ranked causes

  1. 1. Motor not running smoothly due to bearing damage (common)
  2. 2. Motor defective (occasional)
  3. 3. Motor has short-circuited coil (occasional)
  4. 4. Loose contact in electrical connections (occasional)
  5. 5. Frequency converter (A2) defective (occasional)
  6. 6. Enable signal to frequency converter interrupted during centrifugation (rare)
First check

Open lid and perform mains reset

Test point
Motor power and control panel
Expected
System reinitializes; error clears if motor operates normally

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Serial I/O - no connection between control panel and frequency converter

Control panel has no connection to frequency converter via serial interface. Centrifuge slows down and continues cooling.

Ranked causes

  1. 1. Flat ribbon cable to frequency converter (A2) defective (common)
  2. 2. No voltage at frequency converter (A2) (occasional)
  3. 3. Overtemperature switch (F2) on brake resistor opened or not connected (occasional)
  4. 4. Overtemperature fuse (F5) blown or not connected (external braking chopper only) (occasional)
  5. 5. Supply board, control panel, or frequency converter defective (occasional)
  6. 6. Plug X5 on frequency converter not plugged or mispositioned (occasional)
First check

Perform mains reset

Test point
Frequency converter serial connection
Expected
Communication restored; error clears

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Overspeed detected caution

Speed sensor detects speed 250 RPM higher than maximum rotor speed; centrifuge slows and stops; further cooling continues

Ranked causes

  1. 1. Speed sensor cable insulation defective (occasional)
  2. 2. Loose contact in speed sensor cable (occasional)
  3. 3. Speed sensor defective (common)
  4. 4. Flat ribbon cable to control panel defective (occasional)
  5. 5. Control panel, frequency converter, or supply board defective (occasional)

Overtemperature in condenser caution

Temperature in condenser exceeds 58 °C. Centrifuge slows down and cooling switches off.

Ranked causes

  1. 1. Condenser dirty (common)
  2. 2. Ventilator defective (occasional)
  3. 3. Cable of temperature sensor on condenser (T3) defective (occasional)
  4. 4. Loose contact in condenser sensor plug (occasional)
  5. 5. Supply board (A1) defective (occasional)
First check

Measure voltage at temperature sensor (T3) on supply board (A1) at plug X1 pins 1-2

Test point
Supply board (A1) plug X1 (with plug X1 inserted), pins 1 and 2
Expected
2.98 V at 25 °C or 3.31 V at 58 °C, +/- 20 mV

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Frequency converter overtemperature caution

Frequency converter switches off automatically, rotor slows down unbraked, further operation not possible, cooling continues until set value achieved.

Ranked causes

  1. 1. Insufficient heat transfer from frequency converter to housing (common)
  2. 2. Full load operation at high ambient temperature (occasional)
  3. 3. Flat ribbon cable defective (occasional)
  4. 4. Supply board (A1) or frequency converter (A2) defective (rare)
First check

Carry out MAINS RESET

Test point
Mains switch
Expected
Error code cleared

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Incremental encoder - positioning information missing on tracks A or B

Rotor turns and speedometer pulses occur, but control panel does not recognize encoder pulses within two rounds. Tracks -A- and/or -B- defective or Track N disadjusted. Further operation is not possible.

Ranked causes

  1. 1. Defective incremental encoder (B7) (common)
  2. 2. Loose connection at incremental encoder plug (B7) (common)
  3. 3. Defective electronic interface/positioning (A5.1) (occasional)
  4. 4. Defective control panel (A5) (occasional)
First check

Measure encoder signals with oscilloscope

Test point
Electronic interface/positioning (A5.1) at encoder plug X3: CH A (pin 4, green wire), CH B (pin 5, yellow wire), GND (pi
Expected
5 V signals shifted 90 degrees on Channels A and B
Tools
Oscilloscope with 2V/DIV sensitivity; time-base deflection depends on speed

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Incremental encoder - Track N (zero pulse) missing or defective

Rotor turns and speedometer pulses occur, but control panel does not recognize zero-pulses within two rounds. Track N defective or not connected. Further operation is not possible.

Ranked causes

  1. 1. Defective incremental encoder (B7) (common)
  2. 2. Loose connection at incremental encoder plug (B7) (common)
  3. 3. Defective electronic interface/positioning (A5.1) (occasional)
  4. 4. Defective control panel (A5) (occasional)
First check

Measure Track N signal with oscilloscope

Test point
Electronic interface/positioning (A5.1) at encoder plug X3: Track N (pin 6, pink wire), GND (pin 2, white wire)
Expected
Zero pulse appears 1 time per revolution at 5 V
Tools
Oscilloscope with 2V/DIV sensitivity

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Invalid rotor code read during startup

An invalid rotor code was detected during start-up. Centrifuge slows down and cooling continues until SET temperature is reached.

Ranked causes

  1. 1. Magnetic code on rotor defective (common)
  2. 2. Speedometer system defective (occasional)
  3. 3. Loose contact in speed sensor plug (occasional)
  4. 4. Motor rotating in wrong direction (rare)
First check

Open lid or perform mains reset

Test point
Control panel
Expected
Error clears and system is ready for restart

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Key-operated switch voltage out of range

Error 57 displayed. Further operation not possible.

Ranked causes

  1. 1. Key-operated switch board is defective (occasional)
  2. 2. Cable to key-operated switch board is defective (occasional)
  3. 3. Loose contact in wire layout (occasional)
  4. 4. Supply board A1 is defective (rare)
First check

Measure voltage at supply board A1 plug X27 pins 1-2 for each switch position

Test point
Supply board A1, plug X27, pin 1 - pin 2
Expected
Position 1 (LOCK1) ≈ 1.4 V, Position 2 (LOCK2) ≈ 2.1 V, Position 3 ≈ 3.3 V, Within 0.9 V - 3.8 V range

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Lid lock opened during centrifugation patient safety

Lid lock opens during centrifugation; centrifuge slows and stops; further cooling continues

Ranked causes

  1. 1. Lid lock Y1 defective; opens during centrifugation (common)
  2. 2. Loose contact in cable or plug (occasional)
  3. 3. Flat ribbon cable to control panel defective (occasional)
  4. 4. Control panel or supply board defective (occasional)

Motor overtemperature high voltage

Error 67 displayed. Overtemperature switch in motor opens due to high motor temperature. Takes up to 10 minutes to reset. Converter switches off, rotor slows unbraked, cooling continues.

Ranked causes

  1. 1. Motor operating temperature too high (common)
  2. 2. Plug X4 on frequency converter A2 not plugged or loose contact (occasional)
  3. 3. Flat ribbon cable is defective (occasional)
  4. 4. Supply board A1 or frequency converter A2 is defective (rare)
First check

Measure voltage at plug X4 on frequency converter A2

Test point
Frequency converter A2, plug X4
Expected
16 V DC

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Overcurrent in motor circuit high voltage

Error 63 displayed. Frequency converter switches off, rotor slows unbraked, cooling continues.

Ranked causes

  1. 1. Short circuit in the motor (occasional)
  2. 2. Motor impedance is too low (occasional)
  3. 3. Flat ribbon cable is defective (occasional)
  4. 4. Supply board A1 or frequency converter A2 is defective (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Serial I/O - frequency converter NAK signal

Frequency converter cannot acknowledge command and sends NAK (no acknowledge) signal to control panel. Centrifuge slows down and continues cooling.

Ranked causes

  1. 1. Flat ribbon cable to control panel (A5) defective (common)
  2. 2. Flat ribbon cable to frequency converter (A2) defective (common)
  3. 3. Frequency converter (A2) defective (occasional)
  4. 4. Supply board (A1) or control panel (A5) defective (occasional)
First check

Perform mains reset

Test point
Frequency converter response
Expected
Converter acknowledges commands; error clears

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Temperature sensor at bottom of chamber short or open circuit

Error 53 displayed. Centrifuge slows down, cooling system switched off.

Ranked causes

  1. 1. Temperature sensor T2 at bottom of chamber is defective (common)
  2. 2. Sensor cable is defective (occasional)
  3. 3. Loose contact in the plug (occasional)
  4. 4. Supply board A1 is defective (rare)
First check

Measure voltage at supply board A1 plug X2 pins 4-5

Test point
Supply board A1, plug X2, pin 4 - pin 5
Expected
2.98 V (± 20 mV) at 25 °C, ±20 mV

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Temperature sensor at top of chamber short or open circuit

Error 54 displayed. Centrifuge slows down, cooling system switched off.

Ranked causes

  1. 1. Temperature sensor T1 at top of chamber is defective (common)
  2. 2. Sensor cable is defective (occasional)
  3. 3. Loose contact in the plug (occasional)
  4. 4. Control panel A5 is defective (rare)
First check

Measure voltage at control panel A5 plug X2 pins 1-2

Test point
Control panel A5, plug X2, pin 1 - pin 2
Expected
2.50 V = -23 °C to 3.77 V = +104 °C (2.98 V ± 20 mV at 25 °C), ±20 mV

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Temperature sensor on condenser short or open circuit

Error 55 displayed. Centrifuge slows down, cooling system switched off.

Ranked causes

  1. 1. Temperature sensor T3 on condenser is defective (common)
  2. 2. Sensor cable is defective (occasional)
  3. 3. Loose contact in the plug (occasional)
  4. 4. Supply board A1 is defective (rare)
First check

Measure voltage at supply board A1 plug X1 pins 1-2

Test point
Supply board A1, plug X1, pin 1 - pin 2
Expected
2.98 V (± 20 mV) at 25 °C, ±20 mV

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

EEPROM in frequency converter defective caution

Frequency converter switches off automatically, rotor slows down unbraked, further operation not possible, cooling continues until set value achieved.

Ranked causes

  1. 1. Flat ribbon cable defective (occasional)
  2. 2. Frequency converter (A2) defective (occasional)
  3. 3. Supply board (A1) defective (occasional)
First check

Carry out MAINS RESET

Test point
Mains switch
Expected
Error code cleared

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Enable signal not transmitted to frequency converter high voltage

Error 60 displayed. Further operation not possible.

Ranked causes

  1. 1. Flat ribbon cable to frequency converter A2 is defective (occasional)
  2. 2. Flat ribbon cable to control panel A5 is defective (occasional)
  3. 3. Supply board A1 is defective (rare)
First check

Measure voltage at supply board A1 plug X7 pins 1-5

Test point
Supply board A1, plug X7, pin 1 - pin 5
Expected
Lid not locked: 0 V; Lid locked after start: 15 V - 20 V, typical 18 V

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Hatch does not open or close

Drive motor for hatch does not respond to key presses; hatch position not displayed or stuck

Ranked causes

  1. 1. End switch not actuated or not detected (common)
  2. 2. Incorrect last end position stored in control panel (occasional)
  3. 3. Centrifuge not at standstill (occasional)
First check

Verify end switch activation

Test point
Hatch end switches
Expected
Switch actuates when hatch fully open or fully closed

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Incorrect feedback - control panel input signal mismatch

Control panel does not recognize expected signal from components. Centrifuge stops with error message.

Ranked causes

  1. 1. Control panel (A5) defective (occasional)
  2. 2. Supply board (A1) defective (occasional)
  3. 3. Flat ribbon cable between supply board and control panel defective (common)
First check

Perform mains reset

Test point
Control and supply boards
Expected
System reinitializes and error clears

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Initialization displays VERSION - ERROR 12

VERSION - ERROR 12 appears on display after mains supply is switched on

Ranked causes

  1. 1. Coding strip jumpers do not match machine and cooling version (common)
  2. 2. S-control panel not adjusted to centrifuge (occasional)
  3. 3. Frequency converter not adjusted to centrifuge (occasional)
First check

Verify jumper positions on coding strip match machine and cooling version

Test point
Supply board A1 coding strip positions A.0, A.4, A.5
Expected
Jumper A.4 and A.5 plugged for machine version 3; jumper A.0 plugged for cooling version 1

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Mains interruption during centrifugation high voltage

Power failure occurred during centrifugation or mains restored during operation. Centrifuge slows down to allow lid opening, or if rotor has stopped, lid release activates.

Ranked causes

  1. 1. Power failure (common)
  2. 2. Loose contact in electrical connections (occasional)
  3. 3. Flat ribbon cable to control panel (A5) defective (occasional)
First check

Open lid and press START key

Test point
Control panel
Expected
System ready for new cycle or manual operation

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Motor thermal overload protection trip high voltage

Error 65 displayed. I*t switch-off for motor; motor current too high for extended period. Converter switches off, rotor slows unbraked, cooling continues.

Ranked causes

  1. 1. Motor impedance too low; winding resistances too low (occasional)
  2. 2. Flat ribbon cable is defective (occasional)
  3. 3. Supply board A1 or frequency converter A2 is defective (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Overtemperature in centrifuge chamber caution

Temperature in centrifuge chamber exceeds threshold. Centrifuge slows down and cooling switches off.

Ranked causes

  1. 1. Overtemperature switch cable defective (common)
  2. 2. Loose contact in overtemperature switch plug (occasional)
  3. 3. Supply board (A1) defective (occasional)
First check

Measure voltage at overtemperature switch on supply board (A1) at plug X2 pins 1-2

Test point
Supply board (A1) plug X2 (with plug X2 inserted), pins 1 and 2
Expected
0 V if switch closed (overtemperature condition) or 18 V if switch open (normal)

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Overvoltage in frequency converter intermediate circuit high voltage

Error 64 displayed. Intermediate circuit voltage > 410 V DC, typically during braking. Converter switches off, rotor slows unbraked, cooling continues.

Ranked causes

  1. 1. Braking chopper A3 is defective (occasional)
  2. 2. Flat ribbon cable is defective (occasional)
  3. 3. Supply board A1 or frequency converter A2 is defective (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Serial I/O - control panel receives incorrect data from frequency converter

Control panel does not receive correct data from frequency converter. Centrifuge slows down and continues cooling.

Ranked causes

  1. 1. Flat ribbon cable to control panel (A5) defective (common)
  2. 2. Flat ribbon cable to frequency converter (A2) defective (common)
  3. 3. Supply board (A1), control panel (A5), or frequency converter (A2) defective (occasional)
First check

Perform mains reset

Test point
Frequency converter data lines
Expected
Correct data received; error clears

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Serial I/O - supply board NAK signal

Supply board cannot acknowledge command and sends NAK (no acknowledge) signal to control panel. Centrifuge slows down.

Ranked causes

  1. 1. Flat ribbon cable to control panel (A5) defective (common)
  2. 2. Supply board (A1) defective (occasional)
  3. 3. Control panel (A5) or frequency converter (A2) defective (occasional)
First check

Perform mains reset

Test point
Supply board response
Expected
Board acknowledges commands; error clears

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Speedometer pulses break down during rotation caution

Rotor slows down and stops during centrifugation; DC braking for 30 sec after stop; further cooling continues until SET temperature

Ranked causes

  1. 1. Speed sensor defective or loose contact in plug (common)
  2. 2. Flat ribbon cable to control panel or frequency converter defective (occasional)
  3. 3. Supply board, control panel, or frequency converter defective (occasional)
First check

Measure speedometer pulses at supply board

Test point
Plug X5, pin 4 to pin 2
Expected
Present
Tools
Multimeter

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Supply voltage lower than 205 V AC high voltage

Error 62 displayed. Frequency converter switches off, rotor slows unbraked, cooling continues.

Ranked causes

  1. 1. Supply voltage is too low (common)
  2. 2. Flat ribbon cable is defective (occasional)
  3. 3. Supply board A1 or frequency converter A2 is defective (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Temperature deviation too high

Error 58 displayed. Centrifuge slows down, cooling system switched off.

Ranked causes

  1. 1. Temperature range is set too low (common)
  2. 2. Failure in the cooling system (occasional)
  3. 3. Supply board A1 is defective (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Temperature sensor offset compensation needed

Temperature readings from sensors (T1, T2, T3) are inaccurate and require calibration

Ranked causes

  1. 1. Temperature sensor replaced (occasional)
  2. 2. S-control panel EPROM replaced (occasional)
  3. 3. Supply board replaced (occasional)
First check

Measure actual temperature at T1 (top chamber sensor)

Test point
Temperature sensor T1 at top of centrifuge chamber
Expected
Actual temperature matches sensor display within tolerance
Tools
Thermometer or temperature measuring device

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Version mismatch in initialization

Differences detected in initialization between supply board jumper settings, control panel EPROM, or frequency converter. Further operation is not possible.

Ranked causes

  1. 1. Incorrect jumper assignment (A.0-A.7) on supply board coding strip (common)
  2. 2. Wrong EPROM plugged in control panel (occasional)
  3. 3. Supply board (A1) defective (occasional)
First check

Perform mains reset

Test point
Power supply and control panel
Expected
System initializes with correct version parameters

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Rotor position 1 programming and storage

Software teaching procedure to establish rotor position 1 (bucket 1 under hatch) after incremental encoder or motor replacement

Ranked causes

  1. 1. Incorrect rotor position 1 stored in control panel (common)
  2. 2. Position 1 outside valid programming range (occasional)
First check

Verify zero position reached (position N)

Test point
Control panel and rotor position
Expected
Control panel beeps when rotor arm approaches center position

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error in frequency converter computing section high voltage

Error 61 displayed. Frequency converter switches off, rotor slows unbraked, cooling continues.

Ranked causes

  1. 1. Flat ribbon cable is defective (occasional)
  2. 2. Supply board A1 or frequency converter A2 is defective (occasional)
First check

Check for red LED flashing on frequency converter A2

Test point
Frequency converter A2 red LED
Expected
LED flashing indicates error in frequency converter

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error in program flow: impermissible operating state caution

Further operation is not possible.

Ranked causes

  1. 1. Impermissible operating state determined (occasional)
  2. 2. S-control panel (A5) defective (common)
First check

Carry out MAINS RESET

Test point
Mains switch
Expected
Error code cleared

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error in program flow: Watchdog hardware error caution

Further operation is not possible.

Ranked causes

  1. 1. Incorrect program flow with program interruption (occasional)
  2. 2. S-control panel (A5) defective (common)
First check

Carry out MAINS RESET

Test point
Mains switch
Expected
Error code cleared

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Frequency converter thermal overload protection trip high voltage

Error 66 displayed. I*t switch-off of frequency converter. Converter switches off, rotor slows unbraked, cooling continues.

Ranked causes

  1. 1. Flat ribbon cable is defective (occasional)
  2. 2. Supply board A1 or frequency converter A2 is defective (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Lid lock message mismatch - control closed, supply open

Control panel indicates lid lock closed while supply board indicates lid lock open. Centrifuge slows down and cooling system switches off.

Ranked causes

  1. 1. Flat ribbon cable to control panel (A5) defective (common)
  2. 2. Supply board (A1) or control panel (A5) defective (occasional)
First check

Perform mains reset

Test point
Power supply
Expected
System reinitializes and error clears

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Lid lock message mismatch - control open, supply closed

Control panel indicates lid lock open while supply board indicates lid lock closed. Centrifuge slows down and continues cooling.

Ranked causes

  1. 1. Flat ribbon cable to control panel (A5) defective (common)
  2. 2. Control panel (A5) or supply board (A1) defective (occasional)
First check

Perform mains reset

Test point
Power supply
Expected
System reinitializes and error clears

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low battery voltage (BATT warning) caution

BATT displayed and flashes during standstill. Centrifuge can operate for at least two weeks after warning. If voltage continues to fall, RAM data may become invalid and CONTROL - ERROR 92 appears on next power-on.

Ranked causes

  1. 1. Battery not connected (common)
  2. 2. Battery voltage too low (< 2.2 V) (common)
First check

Measure battery voltage at processor connection

Test point
Battery terminals
Expected
≥ 2.2 V, >= 2.2 V
Tools
Voltmeter or DMM

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Memory difference between RAM and EEPROM caution

BATT flashes on display. Battery not connected or empty, or S-control panel defective.

Ranked causes

  1. 1. Battery not connected or empty (common)
  2. 2. S-control panel (A5) defective (occasional)
First check

Check if BATT flashes on display

Test point
Display
Expected
BATT flashing or not flashing

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Serial I/O - control panel receives incorrect data from components

Control panel does not receive correct data from supply board and frequency converter. Centrifuge slows down.

Ranked causes

  1. 1. Flat ribbon cable to control panel (A5) defective (common)
  2. 2. Supply board (A1), control panel (A5), or frequency converter (A2) defective (occasional)
First check

Perform mains reset

Test point
Data communication channels
Expected
Correct data received; error clears

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Serial I/O - control panel receives incorrect data from supply board

Control panel does not receive correct data from supply board. Centrifuge slows down.

Ranked causes

  1. 1. Flat ribbon cable to control panel (A5) defective (common)
  2. 2. Supply board (A1) or control panel (A5) defective (occasional)
First check

Perform mains reset

Test point
Supply board data lines
Expected
Correct data received; error clears

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Serial I/O - no connection between control panel and supply board

Control panel has no connection to supply board via serial interface. Centrifuge slows down.

Ranked causes

  1. 1. Wrong machine or cooling version set on supply board (A1) (occasional)
  2. 2. Supply board (A1) or control panel (A5) defective (occasional)
First check

Perform mains reset

Test point
Serial interface and jumper settings
Expected
Correct version recognized; communication established

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Serial I/O - no connection between control panel and supply board/frequency converter

Control panel has no connection to supply board and frequency converter via serial interface. Centrifuge slows down.

Ranked causes

  1. 1. Flat ribbon cable to control panel (A5) defective (common)
  2. 2. Supply board (A1) or control panel (A5) defective (occasional)
First check

Perform mains reset

Test point
Serial interface
Expected
Communication established; error clears

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Standstill message mismatch - control not standstill, supply standstill

Control panel indicates no standstill while supply board indicates standstill. Centrifuge performs braked slow-down and continues cooling.

Ranked causes

  1. 1. Flat ribbon cable to control panel (A5) defective (common)
  2. 2. Control panel (A5) or supply board (A1) defective (occasional)
First check

Perform mains reset

Test point
Power supply
Expected
System reinitializes and error clears

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Standstill message mismatch - control standstill, supply not

Control panel indicates standstill while supply board indicates no standstill. Centrifuge performs braked slow-down and continues cooling.

Ranked causes

  1. 1. Flat ribbon cable to control panel (A5) defective (common)
  2. 2. Control panel (A5) or supply board (A1) defective (occasional)
First check

Perform mains reset

Test point
Power supply
Expected
System reinitializes and error clears

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unspecified error (INDEFINITE) caution

Further operation is not possible.

Ranked causes

  1. 1. EPROM defective (occasional)
  2. 2. S-control panel (A5) defective (occasional)
First check

Carry out MAINS RESET

Test point
Mains switch
Expected
Error code cleared

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery not connected to S-control panel

BATT message displayed on screen

Ranked causes

  1. 1. S-control panel battery disconnected (common)
First check

Verify S-control panel battery connected

Test point
S-control panel battery connector
Expected
Battery connected and DIP switch contact 1 closed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Imbalance mode setup

Configuring imbalance detection mode on supply board and in program

Ranked causes

  1. 1. Imbalance mode not configured (rare)
First check

Verify jumper position on plug X8

Test point
Plug X8 pins 3-4
Expected
Jumper installed at factory setting position 3-4

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Incremental encoder zero position (Track N)

Adjustment of zero pulse position after encoder or motor replacement to ensure correct rotor positioning reference

Ranked causes

  1. 1. Incremental encoder not properly positioned after installation (common)
First check

Rotate rotor by hand to detect zero position

Test point
Rotor and encoder output
Expected
Control panel buzzer activates at zero position (Track N)

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Leading time

Setting pre-cooling activation time before end of centrifugation to prevent sample undercooling

Ranked causes

  1. 1. Leading time not configured for low-speed or extended run conditions (occasional)
First check

Display current leading time setting

Test point
Display screen after PROG key
Expected
LEADING TIME / S <value> (0 = off)

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Temperature average time

Setting delay before actual temperature display during centrifugation startup

Ranked causes

  1. 1. Temperature average time not set to desired value (rare)
First check

Check current temperature average time setting

Test point
Display screen after PROG key
Expected
TEMP.AVERAGE <value> shown (preset 30 seconds)

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Temperature offset correction

Correcting temperature sensor readings when sample temperature deviates from set value during centrifugation runs

Ranked causes

  1. 1. Temperature sensor reading incorrect (common)
First check

Measure actual temperature at each temperature sensor (T1, T2, T3) with external temperature measuring device

Test point
Temperature sensor location in centrifuge chamber
Expected
Temperature matches displayed set value

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CONTROL - ERROR 92 displayed during initialization

Short acoustic signal sounds and CONTROL - ERROR 92 appears on display after switching on mains during initialization

Ranked causes

  1. 1. Normal functional check during initialization (common)
First check

Wait for functional check completion

Test point
Display during initialization step 2
Expected
CONTROL - ERROR 92 appears, then after 30 seconds VERSION - ERROR 12 or normal initialization sequence follows

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Entered program speed exceeds rotor maximum
  • Error code °C / *ERROR 58 temperature deviation after set point reached
  • Error code °C / *ERROR 58 temperature deviation before set point reached
  • Error in EEPROM of S-control panel
  • Error in EPROM of S-control panel
  • Error in RAM of S-control panel
  • Functional check after repair
  • Hardware error in error management of S-control panel
  • Initialization mode not activated
  • Lid lock mode not set correctly
  • Motor slippage check
  • Optical interface not enabled
  • Overpressure in cooling system
  • Positioning mode setup and verification
  • Program interlock cannot be set via key-operated switch
  • Test positioning (POS-MENU 4)

16 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
01TACHO-ERROR: Speedometer pulses break down during rotationSpeed sensor pulses interrupted during centrifugationerror
02TACHO-ERROR: No speedometer pulses after start commandControl panel receives no speedometer pulses after centrifugation starterror
03IMBALANCE: Imbalance on motor axle detectedImbalance detected on rotor motor axlewarning
04CONTROL-ERROR: Lid lock opened during centrifugationLid lock opens during active centrifugationerror
05N > MAX: Excessive speed detected (250 RPM above rotor max)Speed sensor detects speed 250 RPM above maximum rotor speed (n-max)error
06CONTROL-ERROR: Lid lock error; conflicting signals (control panel locked vs supply board unlocked)Lid lock signal mismatch between control panel and supply boarderror
07CONTROL-ERROR: Lid lock error; conflicting signals (control panel unlocked vs supply board locked)Lid lock signal mismatch between control panel and supply boarderror
08CONTROL-ERROR: Standstill error; conflicting signals (control panel standstill vs supply board running)Standstill signal mismatch between control panel and supply boarderror
09CONTROL-ERROR: Standstill error; conflicting signals (control panel running vs supply board standstill)Standstill signal mismatch between control panel and supply boarderror
10ROTORCODE: Invalid rotor code detectedInvalid or unrecognized rotor codeerror
11MAINS INTERRUPT: Mains interruption detectedPower supply interrupted during operationerror
12VERSION-ERROR: Error in initializationInitialization process failureerror
13N < MIN: Speed error; slippage too greatDetected speed below minimum; excessive slippageerror
26CONTROL-ERROR: Incorrect feedback; control panel input does not recognize expected signalControl panel input signal mismatcherror
30SER I/O-ERROR: No connection between control panel and serial interfaceSerial communication failure between control panel and interfaceerror
31SER I/O-ERROR: No connection between frequency converter and serial interfaceSerial communication failure between frequency converter and interfaceerror
32SER I/O-ERROR: No connection between supply board and serial interfaceSerial communication failure between supply board and interfaceerror
33SER I/O-ERROR: Data incorrectly transmitted from componentryData transmission error from internal componentserror
34SER I/O-ERROR: Data incorrectly transmitted between control panel and frequency converterData transmission error in control-converter linkerror
35SER I/O-ERROR: Data incorrectly transmitted between control panel and supply boardData transmission error in control-board linkerror
36SER I/O-ERROR: No acknowledgement (NAK) from frequency converter to control panelFrequency converter fails to acknowledge control panel commanderror
37SER I/O-ERROR: No acknowledgement (NAK) from supply board to control panelSupply board fails to acknowledge control panel commanderror
40POS-ERROR: No positioning information (track A and B defective)Position tracking system failure; tracks A and B non-functionalerror
41POS-ERROR: No positioning information (track N zero-pulse defective)Zero-pulse positioning signal failureerror
42POS-ERROR: Hatch drive motor needs too much time to closeHatch closing motor speed or response abnormalwarning
43POS-ERROR: Hatch drive motor needs too much time to openHatch opening motor speed or response abnormalwarning
46POS-ERROR: Position 1 to 4 not reached within 80 secDrive motor unable to reach target position within timeouterror
50°C / *-ERROR: Overpressure in cooling system detectedCooling system pressure exceeds safe limiterror
51°C / *-ERROR: Overtemperature on condenserCondenser temperature exceeds safe limiterror
52°C / *-ERROR: Overtemperature in centrifuge chamberChamber temperature exceeds safe limiterror
53°C / *-ERROR: Temperature sensor at bottom of centrifuge chamber defectiveBottom temperature sensor failure or malfunctionerror
54°C / *-ERROR: Temperature sensor at top of centrifuge chamber defectiveTop temperature sensor failure or malfunctionerror
55°C / *-ERROR: Temperature sensor on condenser defectiveCondenser temperature sensor failure or malfunctionerror
57LOCK-ERROR: Program locking defectiveProgram lock mechanism failure or malfunctionerror
58°C / *-ERROR: Temperature deviation too highTemperature differs excessively from setpointwarning
60FU/CCI-ERROR: Faulty release signal to frequency converterControl signal to frequency converter not transmitted properlyerror
61FU/CCI-ERROR: Frequency converter error in computing sectionInternal computing/processing error in frequency convertererror
62FU/CCI-ERROR: Frequency converter error; undervoltageFrequency converter supply voltage below safe operating levelerror
63FU/CCI-ERROR: Frequency converter error; overcurrentFrequency converter output current exceeds safe limiterror
64FU/CCI-ERROR: Frequency converter error; overvoltageFrequency converter supply voltage exceeds safe limiterror
65FU/CCI-ERROR: Frequency converter error; I²t switch-off for motorMotor current-time protection activated in frequency convertererror
66FU/CCI-ERROR: Frequency converter error; I²t switch-off of frequency converterFrequency converter current-time protection activatederror
67FU/CCI-ERROR: Overtemperature in motor (external)Motor temperature exceeds safe operating limiterror
68FU/CCI-ERROR: Overtemperature in frequency converterFrequency converter internal temperature exceeds safe limiterror
69FU/CCI-ERROR: Frequency converter EEPROM errorFrequency converter EEPROM memory failure or corruptionerror
90CONTROL-ERROR: S-control panel error; EPROMControl panel EPROM memory failure or corruptionerror
91CONTROL-ERROR: S-control panel error; RAMControl panel RAM memory failure or corruptionerror
92CONTROL-ERROR: S-control panel error; EEPROM/RAM differenceDiscrepancy between control panel EEPROM and RAM contentserror
94CONTROL-ERROR: S-control panel error; WatchdogControl panel watchdog timer failure or reseterror
95CONTROL-ERROR: S-control panel error; inadmissible operative stateControl panel in invalid or undefined operating stateerror
96N > ROTOR-MAX: Nominal speed higher than permitted rotor speed or RCFProgrammed speed or RCF exceeds rotor specificationserror
97CONTROL-ERROR: S-control panel error; EEPROMControl panel EEPROM memory failure or corruptionerror
99CONTROL-ERROR: S-control panel error; hardware failure in error managementError management hardware failure in control panelerror
BATTBattery voltage is too lowBattery voltage measured by processor falls below 2.2 V, checked after each MAINS RESET and at 30-minute intervalswarning
°C / * - ERROR 50Overpressure in cooling systemOverpressure switch activated (applies only to units with switch; ROTANTA 46 RSC has pins 2 and 3 jumpered at X11)error
°C / * - ERROR 51Overtemperature in condenser (> 58 °C)Condenser temperature exceeds 58 °C thresholdwarning
°C / * - ERROR 52Overtemperature in centrifuge chamberChamber temperature exceeds threshold as detected by overtemperature switchwarning
°C / *ERROR 58Temperature deviation from set point; either not approaching during warm-up or exceeding tolerance during steady-state centrifugation1) Temperature moving away from set value during warm-up beyond E58 DIFF threshold; 2) Temperature deviates >120 sec from set value once reached, beyond ERROR TEMPERATUR toleranceerror
CONTROL - ERROR 06Lid lock message faulty - control panel indicates open, supply board indicates closedContradictory lid lock signal between control panel and supply boarderror
CONTROL - ERROR 07Lid lock message faulty - control panel indicates closed, supply board indicates openContradictory lid lock signal between control panel and supply boarderror
CONTROL - ERROR 08Standstill message faulty - control panel indicates standstill, supply board does notContradictory standstill signal between control panel and supply boarderror
CONTROL - ERROR 09Standstill message faulty - control panel indicates no standstill, supply board indicates standstillContradictory standstill signal between control panel and supply boarderror
CONTROL - ERROR 26Incorrect feedback signal to control panelControl panel input does not recognize expected signalerror
CONTROL - ERROR 90Error in the EPROM of the S-control panelEPROM integrity check failure on S-control panel (A5)critical
CONTROL - ERROR 91Error in the RAM of the S-control panelRAM read/write failure or RAM integrity check failure on S-control panel (A5)critical
CONTROL - ERROR 92Difference in the memory areas of RAM and EEPROMData mismatch detected between RAM and EEPROM during initialization or operationerror
CONTROL - ERROR 94Error in program flow: Watchdog hardware error in the S-control panelProgram execution interrupted or infinite loop detected by watchdog timer on S-control panelcritical
CONTROL - ERROR 95Error in program flow: impermissible operating stateOperating state determined to be invalid or unsafe on S-control panelcritical
CONTROL - ERROR 97Error in the EEPROM of the S-control panelEEPROM integrity check failure on S-control panel (A5)critical
CONTROL - ERROR 99Hardware error in the error management of the S-control panelFailure in error detection or reporting system on S-control panel (A5)critical
ERROR 53Temperature sensor at bottom of chamber has short circuit or open circuit.Voltage < 0.50 V (short circuit) or > 4.50 V (open circuit) at sensor.error
ERROR 54Temperature sensor at top of chamber has short circuit or open circuit.Voltage outside permissible range 2.50 V to 3.77 V.error
ERROR 55Temperature sensor on condenser has short circuit or open circuit.Voltage < 0.50 V (short circuit) or > 4.50 V (open circuit) at sensor.error
ERROR 57Voltage from key-operated switch outside permissible range.Analog voltage at plug X27 outside 0.9 V - 3.8 V range.critical
ERROR 58Temperature deviation too high.Measured temperature deviates excessively from set value.warning
ERROR 60Enable signal not transmitted correctly to frequency converter.Enable signal not reaching frequency converter after MAINS RESET.critical
ERROR 61Error in frequency converter computing section.Internal microprocessor error in frequency converter A2.critical
ERROR 62Supply voltage lower than 205 V AC.Mains supply voltage drops below 205 V AC.warning
ERROR 63Overcurrent in motor circuit.Motor current exceeds frequency converter limit.critical
ERROR 64Overvoltage in frequency converter intermediate circuit.DC voltage in intermediate circuit exceeds 410 V, typically during braking.critical
ERROR 65Motor thermal overload protection activated (I*t switch-off).Motor current too high for extended duration.warning
ERROR 66Frequency converter thermal overload protection activated (I*t switch-off).Frequency converter internal temperature or current too high for extended duration.warning
ERROR 67Motor overtemperature.Motor overtemperature switch opens due to high motor temperature.warning
FU / CCI - ERROR 68Overtemperature in the frequency converterFrequency converter temperature exceeds safe operating limiterror
FU / CCI - ERROR 69EEPROM in the frequency converter is defectiveEEPROM read/write failure detected during converter operationerror
INDEFINITEUnspecified error (not specified error)Unknown or unclassified error condition detectedcritical
MAINS INTERRUPT (11)Power loss during centrifugationElectrical power failure or restoration during operationerror
N < MIN 13Insufficient motor speed due to excessive slippage (>5%)Rotor speed lower than SET speed minus 5%error
N < MIN 13Rotor speed (ACTUAL speed) is lower than the permitted range of controlDuring centrifugation run when actual speed drops below minimum thresholdwarning
N > ROTOR - MAX 96Error in the entered program; speed or RCF exceeds rotor maximumUser attempts to set speed or RCF higher than rotor specification allowswarning
POS-ERROR 40Incremental encoder missing positioning info on tracks A or BDefective encoder tracks or missing encoder pulses within two rotor revolutionscritical
POS-ERROR 41Incremental encoder Track N (zero pulse) defective or missingZero pulse not recognized within two rotor revolutionscritical
POS-ERROR 42Hatch close motor timeout (exceeds 30 seconds)Hatch drive motor activated but end switch does not report closed within 30 secondscritical
POS-ERROR 43Hatch open motor timeout (exceeds 30 seconds)Hatch drive motor activated but end switch does not report opened within 30 secondscritical
POS-ERROR 46Rotor does not reach desired position within 80 secondsPositioning command not completed within timeout; encoder tracks defective or brake not functioningcritical
ROTOR CODE 10Invalid rotor code read during startupDefective rotor code, defective speedometer, loose sensor plug, or motor rotating backwardserror
SER I/O - ERROR 30No serial communication between control panel and supply board/frequency converterSerial interface communication failureerror
SER I/O - ERROR 31No serial communication between control panel and frequency converterSerial interface communication failure with frequency convertererror
SER I/O - ERROR 32No serial communication between control panel and supply board due to version mismatchWrong machine or cooling version configured on supply boarderror
SER I/O - ERROR 33Control panel receives corrupt data from componentsData corruption from supply board or frequency convertererror
SER I/O - ERROR 34Control panel receives corrupt data from frequency converterData corruption from frequency convertererror
SER I/O - ERROR 35Control panel receives corrupt data from supply boardData corruption from supply boarderror
SER I/O - ERROR 36Frequency converter cannot acknowledge command (NAK signal)Frequency converter communication failureerror
SER I/O - ERROR 37Supply board cannot acknowledge command (NAK signal)Supply board communication failureerror
VERSION - ERROR 12Version mismatch in supply board jumpers, control panel EPROM, or frequency converterIncorrect jumper settings or wrong EPROM componentcritical

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.