Hettich MIKRO 185: troubleshooting and error codes
The ServiceForge knowledge base holds 67 documented symptoms for the Hettich MIKRO 185, with 158 ranked causes, 122 diagnostic checks with test points and expected values, and 83 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
No display high voltage
Display is not illuminated. No output on screen.
Ranked causes
- 1. No mains voltage (common)
- 2. Mains input fuses defective (F1, F2) (common)
- 3. Mains switch defective (occasional)
- 4. Electronics A3 defective (rare)
- 5. Electronics A1 defective (rare)
- 6. Control panel A2 defective (rare)
Verify mains power at outlet
- Test point
- House electrical outlet
- Expected
- Mains voltage present
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
FC ERROR 61.153 caution
Imbalance system not in working order.
Ranked causes
- 1. No magnet holder fastened to motor for imbalance detection (common)
- 2. Magnet defective or missing from holder (common)
- 3. Magnet holder misaligned above hall sensor (occasional)
- 4. Electronics A1 defective (occasional)
- 5. Motor bearings so damaged magnet no longer above sensor (occasional)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
4 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
FC ERROR 61.19 patient safety
Frequency converter detects overspeed; braked rundown until rotor stops; device inoperable.
Ranked causes
- 1. Loose contact on plug S602 or S504 (common)
- 2. Insulation damaged on speed sensor B1 cable (occasional)
- 3. Loose contact on plug S601 or S501 (occasional)
- 4. Control panel A2 defective (occasional)
- 5. Electronics A1 defective (occasional)
Wait for rotor to stop completely and check lid locked symbol
- Test point
- Display and rotor
- Expected
- Rotor stationary and lid locked symbol showing
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
FC ERROR 61.20 patient safety
Speed measurement of frequency converter is faulty on electronics A1; braked rundown until rotor stops; device inoperable.
Ranked causes
- 1. Loose contact on plug S602 or S504 (common)
- 2. Insulation damaged on speed sensor B1 cable (occasional)
- 3. Loose contact on plug S601 or S501 (occasional)
- 4. Control panel A2 defective (occasional)
- 5. Electronics A1 defective (occasional)
Wait for rotor to stop and verify lid locked symbol
- Test point
- Display and rotor
- Expected
- Rotor stopped and lid locked symbol displayed
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
TACHO ERROR 1 patient safety
During centrifugation, control panel A2 loses speedometer pulses from speed sensor B1.
Ranked causes
- 1. Speed sensor B1 defective or loose contact (common)
- 2. Loose contact on plug S602 or S504 (common)
- 3. Loose contact on plug S601 or S501 (common)
- 4. Control panel A2 defective (occasional)
- 5. Electronics A1 defective (occasional)
Wait for rotor to stop and check lid lock symbol
- Test point
- Display
- Expected
- Lid locked symbol displayed
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
TACHO ERROR 2 patient safety
After starting centrifugation, control panel A2 receives no speedometer pulses from speed sensor B1.
Ranked causes
- 1. Speed sensor B1 defective or loose contact (common)
- 2. Loose contact on plug S602 or S504 (common)
- 3. Loose contact on plug S601 or S501 (common)
- 4. Control panel A2 defective (occasional)
- 5. Electronics A1 defective (occasional)
Wait for rotor to stop and check lid lock symbol
- Test point
- Display
- Expected
- Lid locked symbol displayed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
COM ERROR 31 caution
Error in communication between control panel A2 and frequency converter on electronics A1 via serial interface.
Ranked causes
- 1. Loose contact on plug S601 or S501 (common)
- 2. Overtemperature switch in motor M1 open (occasional)
- 3. Electronics A1 defective (occasional)
- 4. Control panel A2 defective (occasional)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
COM ERROR 36 caution
Error in communication between control panel A2 and frequency converter on electronics A1 via serial interface; overtemperature switch in motor may be open.
Ranked causes
- 1. Loose contact on plug S601 or S501 (common)
- 2. Overtemperature switch in motor M1 open (occasional)
- 3. Electronics A1 defective (occasional)
- 4. Control panel A2 defective (occasional)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
FC ERROR 61.21 patient safety
Frequency converter detects underspeed; unbraked rundown until rotor stops; device inoperable.
Ranked causes
- 1. Motor sluggish due to bearing damage (common)
- 2. Motor winding short-circuit defective (common)
- 3. Electronics A1 defective (occasional)
- 4. Loose contact on plug S602 or S504 (occasional)
Wait for rotor to stop and check lid locked symbol
- Test point
- Display and rotor
- Expected
- Rotor completely stopped and symbol showing
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
FC ERROR 61.23 patient safety
During centrifugation, speed measurement on electronics A1 detects impermissible speed deviation.
Ranked causes
- 1. Speed sensor B1 defective or loose contact (common)
- 2. Loose contact on plug S602 or S504 (common)
- 3. Loose contact on plug S601 or S501 (common)
- 4. Electronics A1 defective (occasional)
Wait for rotor to stop and check lid lock symbol
- Test point
- Display
- Expected
- Lid locked symbol displayed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
FC ERROR 61.5 caution
Overtemperature in motor M1; device can no longer be operated.
Ranked causes
- 1. Overtemperature switch in motor opens due to excessive motor temperature (common)
- 2. Loose contact on plug S101 / electronics A1 (occasional)
- 3. Electronics A1 defective (occasional)
- 4. Motor defective (occasional)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
IMBALANCE patient safety
Impermissible imbalance detected; braked rundown until rotor stops.
Ranked causes
- 1. Rotor unevenly loaded (common)
- 2. Three rubber/metal bearings on motor damaged (occasional)
- 3. Imbalance switch-off maladjusted (occasional)
- 4. Electronics A1 defective (rare)
Wait for rotor to stop and verify lid locked symbol
- Test point
- Display and rotor
- Expected
- Rotor completely stopped and symbol displayed
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No mains voltage present at device high voltage
Device does not power up; no status LED indication
Ranked causes
- 1. Installation fuses blown (common)
- 2. Centrifuge mains input fuses blown (common)
- 3. No power at connecting cable or plug (occasional)
- 4. Electronics A3 failure (MIKRO 185) (rare)
Verify installation fuses intact
- Test point
- Building installation fuse box
- Expected
- Fuses intact and not blown
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
OVER SPEED 5 patient safety
Overspeed detected by speed measurement function on control panel A2; unbraked rundown until rotor stops.
Ranked causes
- 1. Loose contact on plug S602 or S504 (common)
- 2. Loose contact on plug S601 or S501 (common)
- 3. Electronics A1 defective (occasional)
- 4. Control panel A2 defective (occasional)
Wait for rotor to stop
- Test point
- Rotor and display
- Expected
- Rotor completely stopped
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
TACHO ERR 61.22 patient safety
During centrifugation, electronics A1 loses speedometer pulses from speed sensor B1.
Ranked causes
- 1. Speed sensor B1 defective or loose contact (common)
- 2. Loose contact on plug S602 or S504 (common)
- 3. Loose contact on plug S601 or S501 (common)
- 4. Electronics A1 defective (occasional)
Wait for rotor to stop and check lid lock symbol
- Test point
- Display
- Expected
- Lid locked symbol displayed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
UNDER SPEED 13 patient safety
Underspeed detected by speed measurement function on control panel A2; unbraked rundown until rotor stops.
Ranked causes
- 1. Loose contact on plug S602 or S504 (common)
- 2. Loose contact on plug S601 or S501 (common)
- 3. Electronics A1 defective (occasional)
- 4. Control panel A2 defective (occasional)
Wait for rotor to stop completely
- Test point
- Rotor
- Expected
- No rotation
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
COM ERROR 34 caution
Error in communication between control panel A2 and frequency converter on electronics A1 via serial interface.
Ranked causes
- 1. Loose contact on plug S601 or S501 (common)
- 2. Electronics A1 defective (occasional)
- 3. Control panel A2 defective (occasional)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
FC ERROR 60 caution
Enable signal not correctly transmitted to frequency converter on electronics A1; device can no longer be operated.
Ranked causes
- 1. Loose contact on plug S601 or S501 (common)
- 2. Control panel A2 defective (occasional)
- 3. Electronics A1 defective (occasional)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
FC ERROR 61.1 high voltage
Frequency converter error: Undervoltage in intermediate circuit; unbraked rundown until rotor stops; device can no longer be operated.
Ranked causes
- 1. Mains voltage too low (common)
- 2. Electronics A3 defective (occasional)
- 3. Electronics A1 defective (occasional)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
FC ERROR 61.16 caution
Enable signal not correctly transmitted to frequency converter on electronics A1; device can no longer be operated.
Ranked causes
- 1. Loose contact on plug S601 or S501 (common)
- 2. Control panel A2 defective (occasional)
- 3. Electronics A1 defective (occasional)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
FC ERROR 61.17 caution
Error in communication between control panel A2 and frequency converter on electronics A1 via serial interface.
Ranked causes
- 1. Loose contact on plug S601 or S501 (common)
- 2. Electronics A1 defective (occasional)
- 3. Control panel A2 defective (occasional)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
FC ERROR 61.4 caution
Overtemperature in frequency converter on electronics A1; unbraked rundown until rotor stops; device inoperable.
Ranked causes
- 1. Heat dissipation from frequency converter to housing insufficient (common)
- 2. Full-load operation at ambient temperature > 45°C (common)
- 3. Electronics A1 defective (occasional)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
MAINS INTER 11 caution
Mains interruption detected during centrifugation run; cannot be cleared with MAINS RESET.
Ranked causes
- 1. Mains voltage failure (common)
- 2. Loose contact in electrical lines (occasional)
- 3. Loose contact on plug S601 or S501 (occasional)
Wait for rotor to stop completely
- Test point
- Rotor
- Expected
- No rotation
MAINS INTERRUPT caution
Mains interruption occurs during centrifugation run.
Ranked causes
- 1. Mains voltage failure (common)
- 2. Loose contact in electrical lines (occasional)
- 3. Loose contact on plug S601 or S501 (occasional)
Wait for rotor to stop
- Test point
- Rotor
- Expected
- Rotor completely stopped
Motor does not start or run high voltage
Motor fails to operate; rotor does not spin up during centrifugation cycle initiation
Ranked causes
- 1. Installation fuses blown or open (common)
- 2. Centrifuge mains input fuses open (common)
- 3. No mains voltage at electronics (occasional)
Verify installation facility fuses are intact
- Test point
- Facility main electrical panel
- Expected
- Fuses intact, no open circuits
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
FC ERROR 61.10 high voltage
Internal error of frequency converter on electronics A1; device can no longer be operated. Possible short-circuit in motor.
Ranked causes
- 1. Short-circuit in motor (common)
- 2. Electronics A1 defective (occasional)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
FC ERROR 61.13 high voltage
Internal error of frequency converter on electronics A1; device can no longer be operated. Possible short-circuit in motor.
Ranked causes
- 1. Short-circuit in motor (common)
- 2. Electronics A1 defective (occasional)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
FC ERROR 61.2 high voltage
Frequency converter error: Overvoltage in intermediate circuit (> 410 V DC); unbraked rundown until rotor stops; device inoperable.
Ranked causes
- 1. Mains voltage too high (common)
- 2. Electronics A1 defective (occasional)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
FC ERROR 61.9 high voltage
Internal error of frequency converter on electronics A1; device can no longer be operated. Possible short-circuit in motor.
Ranked causes
- 1. Short-circuit in motor (common)
- 2. Electronics A1 defective (occasional)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
LED status indication not working or flashing error
Yellow status LED on centrifuge does not light during standby or run, or flashes continuously indicating unspecified error condition
Ranked causes
- 1. Specific error code active (see chapter 5) (common)
- 2. LED bulb or indicator circuit fault (occasional)
Identify displayed error code or flash pattern
- Test point
- Centrifuge front panel yellow LED
- Expected
- Steady light (normal operation) or specific flash pattern (error)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lid does not open after centrifugation patient safety
Motor-driven lid lock remains engaged and lid cannot be manually opened after rotor has stopped
Ranked causes
- 1. Rotor still rotating (standstill not detected) (occasional)
- 2. Lid lock microswitch failure or misalignment (rare)
Confirm rotor has reached complete standstill
- Test point
- Rotor visual inspection or tachometer reading
- Expected
- Rotor not rotating; 0 rpm
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lid does not open when button pressed patient safety
Lid lock motor does not respond to open button command
Ranked causes
- 1. Rotor still spinning (common)
- 2. Lid lock microswitch failed or misaligned (occasional)
Verify rotor standstill
- Test point
- Speed sensor B1 signal or electronics A1 display
- Expected
- Rotor speed = 0 rpm or standstill confirmed by electronics
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
LID ERROR 4.1 patient safety
Lid is not locked; lid lock is not actuated or jams during locking.
Ranked causes
- 1. Lid lock A4 defective (common)
- 2. Electronics A1 defective (occasional)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
LID ERROR 4.10 caution
Unknown lid lock type.
Ranked causes
- 1. Lid lock A4 defective (common)
- 2. Control panel A2 defective (occasional)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
LID ERROR 4.126 patient safety
Microswitch in lid lock opens during centrifugation run.
Ranked causes
- 1. Emergency unlocking performed during centrifugation (common)
- 2. Lid lock A4 defective (occasional)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
LID ERROR 4.127 patient safety
Lid lock not actuated or jams during locking.
Ranked causes
- 1. Lid lock A4 defective (common)
- 2. Electronics A1 defective (occasional)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
LID ERROR 4.5 patient safety
Microswitch in lid lock opens during centrifugation run.
Ranked causes
- 1. Emergency unlocking performed during centrifugation (common)
- 2. Lid lock A4 defective (occasional)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
LID ERROR 4.8 patient safety
Microswitch in lid lock opens during centrifugation run.
Ranked causes
- 1. Emergency unlocking performed during centrifugation (common)
- 2. Lid lock A4 defective (occasional)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
Motor stops unexpectedly during centrifugation
Running motor suddenly decelerates and stops without operator input
Ranked causes
- 1. Excessive rotor imbalance detected (common)
- 2. Motor overtemperature detected (occasional)
Observe centrifugation cycle and listen for abnormal vibration
- Test point
- Rotor and motor
- Expected
- Smooth quiet rotation without excessive vibration
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rotor imbalance detected during run caution
Motor switches off mid-centrifugation; rotor decelerates to stop due to detected imbalance exceeding permissible limits
Ranked causes
- 1. Rotor load unbalanced or improperly seated (common)
- 2. Imbalance detection sensor (Hall sensor) malfunction or magnet misalignment (rare)
Verify rotor load is balanced and properly seated
- Test point
- Rotor bucket positions and sample tube arrangement
- Expected
- Load symmetric; all positions filled with equal mass or balancing counterweights
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Status LED not illuminated
Yellow LED does not light up during standby or centrifugation run
Ranked causes
- 1. LED defective or not installed (occasional)
- 2. No power to LED circuit (common)
Verify mains power present at electronics
- Test point
- Electronics module power input
- Expected
- Mains voltage present
- Tools
- Multimeter
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on high voltage
No LED indication, no response to power switch
Ranked causes
- 1. Installation fuses blown or facility mains unavailable (common)
- 2. Centrifuge mains fuses open (common)
Verify mains availability at facility
- Test point
- Building electrical panel
- Expected
- Facility fuses intact, mains available
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
VERSION ERROR 12 caution
Centrifuge model and type saved in control panel A2 do not match electronics A1.
Ranked causes
- 1. Incorrect centrifuge model or type in control panel A2 (common)
- 2. Control panel A2 defective (rare)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Functional check after repair high voltage
Verification of centrifuge operation following service work including key/display function, rotor performance, temperature control, imbalance detection, power consumption and electrical safety
Ranked causes
- 1. Post-repair verification required (common)
Function of keys and display
- Test point
- Control panel
- Expected
- Keys responsive, display shows correct values during operation
7 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Centrifugation will not start
Motor does not start when run button pressed despite mains power present
Ranked causes
- 1. Lid not closed or lid lock not engaged (common)
Verify lid is fully closed and locked
- Test point
- Lid lock microswitch or electronics A1 lid status input
- Expected
- Lid lock engaged; microswitch closed; electronics confirm locked state
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CRC ERROR 27.1 caution
ROM read error; device can no longer be operated.
Ranked causes
- 1. Control panel A2 defective (common)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
CTRL ERROR 25.1 caution
EEPROM read or memory error; device can no longer be operated.
Ranked causes
- 1. Control panel A2 defective (common)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
CTRL ERROR 25.2 caution
EEPROM read or memory error; device can no longer be operated.
Ranked causes
- 1. Control panel A2 defective (common)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
FC ERROR 61.128 caution
Internal error of frequency converter.
Ranked causes
- 1. Electronics A1 defective (common)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
FC ERROR 61.129 caution
Internal error of frequency converter.
Ranked causes
- 1. Electronics A1 defective (common)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
FC ERROR 61.131 caution
Internal error of frequency converter.
Ranked causes
- 1. Electronics A1 defective (common)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
FC ERROR 61.132 caution
Internal error of frequency converter.
Ranked causes
- 1. Electronics A1 defective (common)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
FC ERROR 61.140 caution
Internal error of frequency converter.
Ranked causes
- 1. Electronics A1 defective (common)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
FC ERROR 61.141 caution
Internal error of frequency converter.
Ranked causes
- 1. Electronics A1 defective (common)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
FC ERROR 61.142 caution
Internal error of frequency converter.
Ranked causes
- 1. Electronics A1 defective (common)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
FC ERROR 61.18 caution
Frequency converter receives incorrect direction of rotation command; device can no longer be operated.
Ranked causes
- 1. Control panel A2 defective (common)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared and correct rotation command sent
FC ERROR 61.64 caution
Internal error of frequency converter.
Ranked causes
- 1. Electronics A1 defective (common)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
Imbalance detected during centrifugation patient safety
Motor switches off and rotor brakes to stop when excessive vibration is detected during run-up, centrifuging or braking
Ranked causes
- 1. Excessive rotor imbalance during operation (common)
Verify hall sensor on electronics A1 is functional
- Test point
- Hall sensor on electronics A1
- Expected
- Sensor detects magnet in motor holder
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
LED indication abnormal
LED flashes or does not respond as expected during operation or standby
Ranked causes
- 1. Error condition detected by electronics (common)
Observe LED flash pattern
- Test point
- Yellow LED on electronics A3
- Expected
- Steady light in standby/run, specific flash pattern for errors
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
LID ERROR 4.2 patient safety
The lid lock is not actuated.
Ranked causes
- 1. Electronics A1 defective (common)
Perform MAINS RESET
- Test point
- Control panel
- Expected
- Error cleared
More symptoms
- LID ERROR 4.3
- LID ERROR 4.4
- LID ERROR 4.6
- LID ERROR 4.7
- LID ERROR 4.9
- Motor overheats and shuts down
- Status LED flashing with error indication
7 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
COM ERROR 31 | Communication error between control panel A2 and frequency converter on electronics A1. | Serial interface communication fails. | error |
COM ERROR 34 | Incorrect data transmission between control panel and frequency converter | Data corruption or parity errors during communication | error |
COM ERROR 36 | No response from frequency converter to control panel | Frequency converter fails to acknowledge commands | error |
COM ERROR 36 | Communication error between control panel A2 and frequency converter; may involve motor overtemperature switch. | Serial interface communication fails; overtemperature switch may be open. | error |
CRC ERROR 27.1 | Control panel ROM error | ROM checksum verification failure | error |
CRC ERROR 27.1 | ROM read error in control panel; device inoperable. | Control panel ROM memory cannot be read or is corrupted. | critical |
CTRL ERROR 25.1 | EEPROM read or memory error in control panel; device inoperable. | Control panel EEPROM fails or cannot be read. | critical |
CTRL ERROR 25.1 | Control panel EEPROM error | EEPROM data corruption or read failure detected | error |
CTRL ERROR 25.2 | Control panel EEPROM error | EEPROM data corruption or read failure detected | error |
CTRL ERROR 25.2 | EEPROM read or memory error in control panel; device inoperable. | Control panel EEPROM memory failure. | critical |
FC ERROR 60 | Enable signal not correctly transmitted to frequency converter on electronics A1; device inoperable. | Enable signal transmission to frequency converter fails. | critical |
FC ERROR 61.1 | Frequency converter error: Undervoltage in intermediate circuit; unbraked rundown; device inoperable. | Intermediate circuit voltage drops below safe operating threshold. | critical |
FC ERROR 61.10 | Internal error of frequency converter on electronics A1; possible motor short-circuit; device inoperable. | Frequency converter detects internal fault. | critical |
FC ERROR 61.10 | Frequency converter overcurrent condition | Actual current through frequency converter exceeds rated maximum | critical |
FC ERROR 61.128 | Internal error of frequency converter | Converter detects unrecoverable internal fault | error |
FC ERROR 61.129 | Internal error of frequency converter | Converter detects unrecoverable internal fault | error |
FC ERROR 61.13 | Frequency converter overcurrent condition | Actual current through frequency converter exceeds rated maximum | critical |
FC ERROR 61.13 | Internal error of frequency converter on electronics A1; possible motor short-circuit; device inoperable. | Frequency converter internal fault detected. | critical |
FC ERROR 61.131 | Internal error of frequency converter | Converter detects unrecoverable internal fault | error |
FC ERROR 61.132 | Internal error of frequency converter | Converter detects unrecoverable internal fault | error |
FC ERROR 61.140 | Internal error of frequency converter | Converter detects unrecoverable internal fault | error |
FC ERROR 61.141 | Internal error of frequency converter | Converter detects unrecoverable internal fault | error |
FC ERROR 61.142 | Internal error of frequency converter | Converter detects unrecoverable internal fault | error |
FC ERROR 61.153 | Imbalance system not in working order; cannot detect imbalance. | Magnet, holder, or sensor alignment problem with imbalance detection circuit. | warning |
FC ERROR 61.153 | Imbalance detection system faulty in frequency converter | Imbalance sensor signal missing or malfunctioning | warning |
FC ERROR 61.16 | Enable signal not correctly transmitted to frequency converter on electronics A1; device inoperable. | Enable signal to frequency converter not transmitted properly. | critical |
FC ERROR 61.17 | No communication between control panel and frequency converter | Serial link dropout during operation | error |
FC ERROR 61.18 | Frequency converter receives incorrect direction of rotation command; device inoperable. | Direction of rotation signal is incorrect or invalid. | error |
FC ERROR 61.18 | Incorrect rotation direction command received by frequency converter | Frequency converter interprets wrong direction in command message | warning |
FC ERROR 61.19 | Overspeed detected by frequency converter | Measured speed exceeds programmed maximum limit | critical |
FC ERROR 61.19 | Frequency converter detects overspeed; braked rundown until rotor stops; device inoperable. | Rotor speed exceeds safe maximum during operation. | critical |
FC ERROR 61.2 | Frequency converter error: Overvoltage in intermediate circuit (> 410 V DC); unbraked rundown; device inoperable. | Intermediate circuit voltage exceeds 410 V DC. | critical |
FC ERROR 61.20 | Speed measurement of frequency converter is faulty on electronics A1; braked rundown; device inoperable. | Speed measurement from sensor fails or is inaccurate. | critical |
FC ERROR 61.21 | Underspeed detected by frequency converter | Measured speed falls below programmed minimum threshold | error |
FC ERROR 61.21 | Frequency converter detects underspeed; unbraked rundown until rotor stops; device inoperable. | Rotor speed falls below minimum safe operating speed. | critical |
FC ERROR 61.22 | Frequency converter not receiving speed pulses | Tachometer signal absent or too weak to detect | error |
FC ERROR 61.23 | Speed measurement faulty | Speed calculation error or sensor signal invalid | error |
FC ERROR 61.23 | Speed measurement on electronics A1 detects impermissible speed deviation during centrifugation. | Measured speed varies beyond allowable limits during operation. | error |
FC ERROR 61.4 | Frequency converter overtemperature | Internal temperature sensor exceeds safe limit in frequency converter | error |
FC ERROR 61.4 | Overtemperature in frequency converter on electronics A1; unbraked rundown; device inoperable. | Frequency converter internal temperature exceeds safe limit. | critical |
FC ERROR 61.5 | Overtemperature in motor M1; device can no longer be operated. | Motor temperature exceeds safe operating limit; overtemperature switch opens. | critical |
FC ERROR 61.64 | Internal error of frequency converter | Converter detects unrecoverable internal fault | error |
FC ERROR 61.9 | Frequency converter overcurrent condition | Actual current through frequency converter exceeds rated maximum | critical |
FC ERROR 61.9 | Internal error of frequency converter on electronics A1; possible motor short-circuit; device inoperable. | Frequency converter detects internal fault or motor winding short-circuit. | critical |
IMBALANCE | Impermissible imbalance detected during operation; braked rundown until rotor stops. | Rotor vibration or imbalance exceeds safe operating threshold. | error |
IMBALANCE | Imbalance detected on motor axis | Vibration sensor detects unbalanced load or mechanical imbalance | warning |
LID ERROR 4.1 | Lid lock error | Lid locking mechanism failure | warning |
LID ERROR 4.1 | Lid is not locked; lid lock not actuated or jams during locking. | Lid lock mechanism fails to engage or becomes stuck during locking attempt. | error |
LID ERROR 4.10 | Lid lock error | Lid locking mechanism failure | warning |
LID ERROR 4.10 | Unknown lid lock type detected by system. | Control panel unable to identify lid lock type. | warning |
LID ERROR 4.126 | Microswitch in lid lock opens during centrifugation run. | Microswitch opens unexpectedly during operation. | error |
LID ERROR 4.126 | Lid lock error | Lid locking mechanism failure | warning |
LID ERROR 4.127 | Lid lock not actuated or jams during locking. | Lid lock fails to operate or becomes stuck. | error |
LID ERROR 4.127 | Lid lock error | Lid locking mechanism failure | warning |
LID ERROR 4.2 | Lid lock error | Lid locking mechanism failure | warning |
LID ERROR 4.3 | Lid lock error | Lid locking mechanism failure | warning |
LID ERROR 4.4 | Lid lock error | Lid locking mechanism failure | warning |
LID ERROR 4.5 | Lid lock error | Lid locking mechanism failure | warning |
LID ERROR 4.5 | Microswitch in lid lock opens during centrifugation run. | Microswitch opens unexpectedly during active centrifugation. | error |
LID ERROR 4.6 | Lid lock jams during unlocking. | Lid lock becomes stuck when attempting to unlock. | error |
LID ERROR 4.6 | Lid lock error | Lid locking mechanism failure | warning |
LID ERROR 4.7 | Lid lock error | Lid locking mechanism failure | warning |
LID ERROR 4.7 | Lid lock jams during unlocking. | Lid lock mechanism stuck during unlock operation. | error |
LID ERROR 4.8 | Microswitch in lid lock opens during centrifugation run. | Microswitch opens during active centrifugation. | error |
LID ERROR 4.8 | Lid lock error | Lid locking mechanism failure | warning |
LID ERROR 4.9 | Lid lock error | Lid locking mechanism failure | warning |
LID ERROR 4.9 | Lid lock jams during unlocking. | Lid lock fails to unlock smoothly. | error |
MAINS INTER 11 | Mains interruption detected; cannot be cleared with MAINS RESET. | Power loss during centrifugation; error persists until manual intervention. | critical |
MAINS INTER 11 | Mains power interruption during centrifugation | Loss of main power during operation | warning |
MAINS INTERRUPT | Mains power interruption detected during centrifugation run. | Power supply loss occurs during active centrifugation. | critical |
No display | Display does not show any output | Unknown condition | error |
No display | Display is not illuminated; no visual output on control panel. | Power applied but display remains blank. | critical |
OVER SPEED 5 | Overspeed condition | Centrifuge speed exceeds safe limit | critical |
OVER SPEED 5 | Overspeed detected by speed measurement function on control panel A2; unbraked rundown occurs. | Measured speed exceeds maximum safe operating speed. | critical |
OVER SPEED 5 | Centrifuge speed exceeds maximum safe limit | Actual speed exceeds programmed maximum | critical |
TACHO ERR 61.22 | Electronics A1 loses speedometer pulses from speed sensor B1 during centrifugation. | Speed signal lost during centrifuge operation. | error |
TACHO ERROR 1 | Control panel loses speedometer pulses during centrifugation from speed sensor B1. | Speed signal lost during active centrifuge run. | error |
TACHO ERROR 2 | Control panel receives no speedometer pulses after centrifuge start. | No speed signal detected immediately after starting centrifugation. | error |
UNDER SPEED 13 | Centrifuge fails to reach minimum operating speed | Actual speed below programmed minimum after acceleration period | error |
UNDER SPEED 13 | Underspeed condition | Centrifuge speed falls below minimum during operation | error |
UNDER SPEED 13 | Underspeed detected by speed measurement function on control panel A2; unbraked rundown occurs. | Measured speed falls below minimum required operating speed. | critical |
VERSION ERROR 12 | Incorrect centrifuge model recognized by control system | Model mismatch during system initialization | error |
VERSION ERROR 12 | Centrifuge model and type in control panel A2 do not match electronics A1. | System detects model/type mismatch at startup or during operation. | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI