Hettich EBA 280: troubleshooting and error codes
The ServiceForge knowledge base holds 9 documented symptoms for the Hettich EBA 280, with 28 ranked causes, 24 diagnostic checks with test points and expected values, and 87 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Unit has no display high voltage
No display appears on the control panel after power-on. Yellow LED on electronics may not be illuminated.
Ranked causes
- 1. No mains voltage supplied to unit (common)
- 2. Mains input fuses F1, F2 defective (common)
- 3. Mains switch in appliance plug defective (occasional)
- 4. Electronics A3 defective (occasional)
- 5. Electronics A1 defective (occasional)
- 6. Control panel A2 defective (rare)
Check building electrical installation
- Test point
- In-house fuses
- Expected
- Intact
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
OVER SPEED 5 – Overspeed detected patient safety
Speed measurement on control panel detects rotor exceeding maximum permitted speed. Rotor coasts down unbraked until standstill. No further operation possible until error clears.
Ranked causes
- 1. Loose contact on plug S505 at electronics A1 (common)
- 2. Loose contact on plug S501 at electronics A1 (common)
- 3. Electronics A1 defective (occasional)
- 4. Control panel A2 defective (occasional)
Perform mains reset after rotor stops
- Test point
- Mains switch
- Expected
- Rotor at standstill before reset
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
TACHO ERROR 1 during centrifugation run
Control panel loses tachometer pulses from speed sensor during centrifugation. Motor speed becomes undetectable. Error clears after rotor stops and mains reset is performed.
Ranked causes
- 1. Speed sensor B1 defective or loose contact (common)
- 2. Loose contact on plug S501 (common)
- 3. Control panel A2 defective (occasional)
- 4. Electronics A1 defective (occasional)
Perform mains reset
- Test point
- Mains switch
- Expected
- Rotor standstill and lid locked symbol displayed
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
TACHO ERROR 2 after start command
After start command, control panel receives no tachometer pulses from speed sensor. Motor does not indicate rotation. Error clears after rotor stops and mains reset is performed.
Ranked causes
- 1. Speed sensor B1 defective or loose contact (common)
- 2. Loose contact on plug S501 (common)
- 3. Control panel A2 defective (occasional)
- 4. Electronics A1 defective (occasional)
Perform mains reset
- Test point
- Mains switch
- Expected
- Rotor standstill and lid locked symbol displayed
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
ROTOR ERROR 10.2 – Invalid rotor code read patient safety
Unrecognized or invalid rotor code detected during startup. Rotor brakes immediately and coasts to standstill. No further operation possible. Error may indicate wrong rotor installed or motor direction incorrect. Error clears after mains reset.
Ranked causes
- 1. Magnet coding on rotor defective (common)
- 2. Motor rotating in wrong direction (occasional)
- 3. Electronics A1 defective (occasional)
Perform mains reset
- Test point
- Mains switch
- Expected
- Rotor at standstill, error remains
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
LID ERROR 4.1 – Lid not locked patient safety
Lid lock is not actuated or jams during locking operation. Lid remains unlocked and centrifugation cannot start. Error clears after mains reset.
Ranked causes
- 1. Lid lock A4 defective (common)
- 2. Electronics A1 defective (occasional)
Perform mains reset
- Test point
- Mains switch
- Expected
- Lid lock attempts to engage after reset
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
LID ERROR 4.5 – Microswitch opens during centrifugation patient safety
Microswitch in lid lock opens prematurely during centrifugation run, interrupting operation. Typically caused by manual emergency unlock during run or defective microswitch. Error clears after mains reset.
Ranked causes
- 1. Emergency unlock performed by hand during run (occasional)
- 2. Lid lock A4 defective (occasional)
Perform mains reset
- Test point
- Mains switch
- Expected
- Centrifugation resumes normally
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ROTOR ERROR 10.1 – Start code not recognized patient safety
Rotor code magnet start pulse not recognized during startup. Rotor brakes down to standstill. No further operation possible except lid unlock. Operator must determine rotor status before restart. Error clears after mains reset.
Ranked causes
- 1. Magnet coding on rotor defective (common)
- 2. Electronics A1 defective (occasional)
Perform mains reset
- Test point
- Mains switch
- Expected
- Rotor at standstill
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
LID ERROR 4.4 – Lid lock jams during locking patient safety
Lid lock mechanism encounters resistance or jam when attempting to lock. Lid lock stalls before fully engaging. Error clears after mains reset.
Ranked causes
- 1. Lid lock A4 defective (common)
Perform mains reset
- Test point
- Mains switch
- Expected
- Lid lock disengages smoothly
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
COM ERROR 31 | Frequency converter serial interface communication failure | No connection between frequency converter on electronics A1 and serial interface | error |
COM ERROR 31, COM ERROR 33, COM ERROR 34, COM ERROR 36, FC E | Communication error with frequency converter | Serial interface communication failure between control panel A2 and electronics A1 | error |
COM ERROR 33 | Communication error between control panel and frequency converter | No correct data transmission between control panel A2 and frequency converter on electronics A1 | error |
COM ERROR 34 | Communication error between control panel and frequency converter | No correct data transmission between control panel A2 and frequency converter on electronics A1 | error |
COM ERROR 36 | No response (NAK) from frequency converter to control panel | Frequency converter on electronics A1 does not acknowledge (sends NAK) command from control panel A2 | error |
CRC-ERROR 27 | CRC error after Power On | Cyclic redundancy check fails immediately after power-on | critical |
CRC ERROR 27.1 | CRC error at standstill | Cyclic redundancy check fails when rotor is at standstill | error |
CRC ERROR 27, CRC ERROR 27.1 | ROM read error | Program memory checksum failure | critical |
CRTL-ERROR 22 | I2C bus communication error | Communication error detected on I2C bus between control panel and electronics | error |
CRTL-ERROR 25.1 | EEPROM read error | Electronics A1 cannot read data from EEPROM memory | error |
CRTL-ERROR 25.3 | EEPROM communication error | Communication with EEPROM memory fails | error |
CRTL-ERROR 25.4 | EEPROM memory error | EEPROM memory reports error condition or data corruption detected | error |
CTRL ERROR 22 | Communication error between control panel and EEPROM | Control panel A2 cannot communicate with memory IC6 | error |
CTRL ERROR 25.1, CTRL ERROR 25.2, CTRL ERROR 25.3, CTRL ERRO | EEPROM read or memory error | Memory malfunction on control panel | critical |
CTRL ERROR 25.2 | EEPROM write error | Electronics A1 cannot write data to EEPROM memory | error |
FC ERROR 60 | Faulty enable signal to frequency converter | Frequency converter does not receive valid enable signal from electronics A1 | error |
FC ERROR 60, FC ERROR 61.16 | Enable signal transmission error to frequency converter | Frequency converter does not receive enable command | critical |
FC ERROR 61.1 | Intermediate circuit undervoltage | DC voltage in power supply intermediate circuit too low | critical |
FC ERROR 61.1 | Frequency converter undervoltage error | DC supply voltage to frequency converter drops below minimum operating threshold | error |
FC ERROR 61.10 | Frequency converter overcurrent error | Output current from frequency converter to motor exceeds maximum safe level | critical |
FC ERROR 61.128 | Internal error of frequency converter | Frequency converter internal diagnostic detects malfunction | error |
FC ERROR 61.129 | Internal error of frequency converter | Frequency converter internal diagnostic detects malfunction or error condition | error |
FC ERROR 61.13 | Frequency converter overcurrent error | Output current from frequency converter to motor exceeds maximum safe level during operation | critical |
FC ERROR 61.131 | Internal error of frequency converter | Frequency converter internal diagnostic detects fault condition | error |
FC ERROR 61.132 | Internal error of frequency converter | Frequency converter internal diagnostic detects malfunction | error |
FC ERROR 61.140 | Internal error of frequency converter | Frequency converter internal diagnostic detects error condition | error |
FC ERROR 61.141 | Internal error of frequency converter | Frequency converter internal diagnostic detects malfunction | error |
FC ERROR 61.142 | Internal error of frequency converter | Frequency converter internal diagnostic detects fault or malfunction | error |
FC ERROR 61.153 | Imbalance detection system not working | Imbalance sensor magnet or Hall sensor malfunction | warning |
FC ERROR 61.154 | Invalid machine version detected | Frequency converter detects incorrect or unrecognized centrifuge model code | error |
FC ERROR 61.16 | Faulty enable signal to frequency converter | Frequency converter does not receive valid enable signal from electronics A1 | error |
FC ERROR 61.17 | Communication error between control panel and frequency converter | No communication or lost communication between control panel A2 and frequency converter on electronics A1 | error |
FC ERROR 61.18 | Incorrect motor rotation direction | Frequency converter detects motor rotation in wrong direction (reverse of commanded) | error |
FC ERROR 61.19 | Frequency converter detects overspeed | Motor speed monitored by frequency converter exceeds maximum safe operating threshold | error |
FC ERROR 61.19, FC ERROR 61.20 | Overspeed or speed measurement fault | Rotor speed exceeds limit (61.19) or speed sensor malfunction (61.20) | critical |
FC ERROR 61.2 | Intermediate circuit overvoltage | DC voltage in power supply intermediate circuit exceeds limit | critical |
FC ERROR 61.2 | Frequency converter overvoltage error | DC supply voltage to frequency converter exceeds maximum operating threshold | error |
FC ERROR 61.20 | Faulty speed measurement by frequency converter | Speed measurement signal from frequency converter is invalid or out of range | error |
FC ERROR 61.21 | Frequency converter detects underspeed | Motor speed monitored by frequency converter falls below minimum operating threshold | error |
FC ERROR 61.21 | Motor underspeed detected by frequency converter | Motor cannot maintain programmed speed | critical |
FC ERROR 61.23 | Speed measurement faulty | During centrifugation run, speed measurement on electronics A1 detects impermissible speed deviation | error |
FC ERROR 61.4 | Frequency converter overtemperature error | Frequency converter internal temperature exceeds safe operating limit | error |
FC ERROR 61.5 | Motor overtemperature detected | Temperature sensor in motor M1 opens due to excessive heat | critical |
FC ERROR 61.64 | Internal error of frequency converter | Frequency converter internal diagnostic detects malfunction or software error | error |
FC ERROR 61.64, FC ERROR 61.128, FC ERROR 61.129, FC ERROR 6 | Internal frequency converter errors | Various internal faults detected by frequency converter | critical |
FC ERROR 61.9 | Frequency converter overcurrent error | Output current from frequency converter to motor exceeds maximum safe level | critical |
FC ERROR 61.9, FC ERROR 61.10, FC ERROR 61.13 | Internal frequency converter error or motor short circuit | Frequency converter detects fault condition or short in motor winding | critical |
FC VERSION ERROR | Wrong electronics A1 installed | Control panel detects electronics A1 model or version incompatible with centrifuge | error |
IMBALANCE | Rotor imbalance detected | Rotor vibration exceeds allowed threshold during run | warning |
KEYBOARD ERROR | Membrane keyboard malfunction | Keyboard not detected or unresponsive | warning |
KEYBOARD ERROR | No or poor keyboard contact; keyboard not connected | Control panel cannot communicate with keyboard interface or detects contact fault | warning |
LID ERROR 4.1 | Lid lock error – lid not locked | Lid lock is not actuated or jams during locking | warning |
LID ERROR 4.10 | Unknown lid lock type | Lid lock type not recognized by control panel during initialization | warning |
LID ERROR 4.126 | Microswitch opens during centrifugation run | Microswitch in lid lock opens prematurely during active centrifugation | warning |
LID ERROR 4.127 | Lid lock error – lid not locked | Lid lock is not actuated or jams during locking | warning |
LID ERROR 4.2 | Lid lock error – lid lock not actuated | Lid lock does not respond to actuation command | warning |
LID ERROR 4.3 | Lid lock error – lid lock not actuated | Lid lock does not respond to actuation command | warning |
LID ERROR 4.4 | Lid lock jams during locking | Lid lock encounters mechanical resistance or jam when attempting to lock | warning |
LID ERROR 4.5 | Microswitch opens during centrifugation run | Microswitch in lid lock opens prematurely during active centrifugation | warning |
LID ERROR 4.6 | Lid lock jams during unlocking | Lid lock encounters resistance or jam when attempting to unlock | warning |
LID ERROR 4.7 | Lid lock jams during unlocking | Lid lock encounters resistance or jam when attempting to unlock | warning |
LID ERROR 4.8 | Microswitch opens during centrifugation run | Microswitch in lid lock opens prematurely during active centrifugation | warning |
LID ERROR 4.9 | Lid lock jams during unlocking | Lid lock encounters resistance or jam when attempting to unlock | warning |
MAINS INTER 11 | Mains interruption during centrifugation run | AC mains power supply is lost or interrupted while rotor is spinning | warning |
MAINS INTERRUPT | Mains interruption during centrifugation run | AC mains power supply is lost or interrupted while rotor is spinning | warning |
MAINS INTERRUPT, MAINS INTER 11 | Mains power interruption during run | AC power loss during centrifugation | critical |
OVER SPEED 5 | Overspeed detected by control panel | Control panel speed measurement detects rotor exceeding maximum permitted speed | error |
ROTOR ERROR 10.1 | Start code not recognized; rotor coding start pulse lost | Rotor code magnet start pulse not detected during startup | error |
ROTOR ERROR 10.2 | Invalid rotor code detected during startup | During start-up, invalid or unrecognized rotor code is read in | error |
ROTOR ERROR 10.3 | Code magnet lost | Rotor code magnet becomes detached or displaced from holder during operation | critical |
ROTOR ERROR 10.3 | Error reading the rotor code | Rotor code not detected during operation | error |
ROTOR ERROR 10.4 | Error reading the rotor code track | Rotor code track not detected | error |
ROTOR ERROR 10.4 | Code track malfunction | Hall sensor cannot read rotor code track properly | error |
ROTOR ERROR 10.5 | Code track failed during centrifugation | Rotor code track becomes unreadable after centrifugation has started | critical |
ROTOR ERROR 10.6 | Rotor code detection timeout after start | Rotor code not detected within timeout period after startup | error |
RPM > ROTOR MAX | Speed exceeds rotor maximum rated speed | Activated program sets speed higher than maximum rated speed for installed rotor | warning |
RXXX WRONG ROTOR | Rotor not approved for this device | Incompatible rotor detected (xxx is rotor code) | warning |
TACHO ERR 61.22 | Frequency converter receives no speed pulses | During centrifugation run, electronics A1 frequency converter no longer receives tachometer pulses from speed sensor B1 | error |
TACHO ERROR 1 | Speed pulse failure during centrifugation run | Control panel loses tachometer pulses from speed sensor B1 after centrifugation has started | error |
TACHO ERROR 2 | No speed pulses after start command | After centrifugation start command, control panel A2 receives no tachometer pulses | error |
TACHO ERROR 96.1 | Speed difference exceeds 10% between frequency converter and control panel | Speed measurement from frequency converter differs from control panel reading by more than 10% | warning |
TACHO ERROR 96.1 | Speed measurement discrepancy between systems | Speed measurement from electronics A1 and control panel A2 differ by >10% | warning |
UNDER SPEED 13 | Rotor running below programmed speed | Speed measurement below setpoint | error |
UNDER SPEED 13 | Underspeed detected by control panel | Control panel speed measurement detects rotor running below required speed threshold | error |
VERSION ERROR 12 | Incorrect centrifuge model recognized | Control panel detects centrifuge model version mismatch or unrecognized model | error |
VERSION ERROR 12 | Control panel and electronics version mismatch | Centrifuge model/type saved on A2 does not match electronics A1 | error |
WATCHDOG RESET | Internal error of control panel watchdog timer | Control panel watchdog timer triggers reset due to internal malfunction or hang | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.