Hamilton Medical HAMILTON-T1: troubleshooting and error codes
The ServiceForge knowledge base holds 128 documented symptoms for the Hamilton Medical HAMILTON-T1, with 181 ranked causes, 301 diagnostic checks with test points and expected values, and 446 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Sensor status shows error or invalid state
Sensor data displays abnormal status values (hex) indicating defect, missing, overflow, underflow, temperature, or invalid conditions
Ranked causes
- 1. Sensor defect status (0004) (rare)
- 2. Sensor missing or not running (0008) (occasional)
- 3. Autozero sequence in progress (0010) (occasional)
- 4. Measurement overflow (0020) or underflow (0040) (occasional)
- 5. Temperature error - undertemperature (0100) or overtemperature (0200) (rare)
- 6. Sensor disabled (0080) or invalid (0400) (rare)
- 7. Sampling line occluded (2000) (occasional)
Read sensor status hex value
- Test point
- Sensor 1/2/3 status field
- Expected
- 0003 or similar with only Initialized (0001) and Calibrated (0002) flags set
- Tools
- Service software display
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Service software startup and technical state management caution
Procedure for starting HAMILTON-T1 service software, managing hardware/software versions, updating technical state after component replacement, and resetting blower timer.
Ranked causes
- 1. Initial service software startup and self-test required (common)
- 2. Technical state information needs recording and verification (common)
- 3. Component replacement requires technical state modification and restart (common)
- 4. Technical state corruption or EEPROM read failure requires cleanup (occasional)
- 5. Export and import technical state to/from USB drive (occasional)
Verify service software startup sequence completes
- Test point
- Main service software screen (page 0000)
- Expected
- Service software main screen displayed with no startup errors
4 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 measurement not displaying or incorrect values patient safety
SpO2 sensor connected but values not shown in monitoring tab or dynamic lung view, or readings are invalid
Ranked causes
- 1. SpO2 communication board not installed or not detected (occasional)
- 2. SpO2 hardware option not activated in configuration (common)
- 3. Incorrect SpO2 sensor type configured (common)
- 4. SpO2 sensor not activated or not connected to board (occasional)
- 5. Finger probe not properly attached or insufficient signal (occasional)
Verify SpO2 communication board installed
- Test point
- System Info display
- Expected
- SpO2 communication board listed in installed hardware
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery storage conditions inadequate caution
Battery stored at incorrect temperature or humidity; risk of reduced shelf life and performance
Ranked causes
- 1. Storage temperature exceeds 25°C over 6-month recharge interval (occasional)
- 2. Battery stored in discharged state (rare)
- 3. Extended exposure to extreme temperatures (rare)
- 4. Adverse storage environment (vibration, moisture, corrosive gas, etc.) (rare)
Verify storage temperature
- Test point
- Storage location
- Expected
- Between 5°C and 21°C; do not exceed 25°C over 6-month intervals, 5-21°C optimal; max 25°C
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Proximal test failures
Rinse flow, proximal pressure, proximal flow, or rinse tank test shows NOT OK; abnormal bubble patterns or pressure readings detected
Ranked causes
- 1. Rinse flow obstruction or unequal flow distribution (unknown)
- 2. Proximal pressure sensor malfunction or calibration drift (unknown)
- 3. Proximal flow sensor fault (unknown)
- 4. Rinse tank leak or connection loose (unknown)
Observe bubble pattern in water tubes during rinse flow test
- Test point
- Water tube outlets
- Expected
- Bubbles appear at both outlets; number of bubbles approximately equal
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 measurement not displaying values patient safety
SpO2 sensor connected but no measurement values shown on display (HAMILTON-T1 SN > SN3000 only)
Ranked causes
- 1. SpO2 communication board not installed (common)
- 2. SpO2 hardware option not activated (common)
- 3. SpO2 sensor type not configured (common)
- 4. SpO2 sensor not connected or inactive (common)
Verify SpO2 communication board is installed
- Test point
- System > Info > Info Tab
- Expected
- SpO2 communication board listed
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit fails to complete startup or preoperational checks patient safety
Ventilator does not pass required preoperational tests before clinical use
Ranked causes
- 1. Tightness test failed or not completed (unknown)
- 2. Flow sensor calibration not completed or failed (unknown)
- 3. O2 cell calibration not completed or failed (unknown)
- 4. Alarm system tests failed or not completed (unknown)
Perform tightness test
- Test point
- Patient circuit and breathing system
- Expected
- No air leaks detected; system holds pressure
9 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery state of health assessment patient safety
Monitor battery state of health (SoH); replace when SoH <20%; consider replacement when <40% or cell voltage difference >100 mV; calibrate if error ≥5%.
Ranked causes
- 1. Battery SoH <20% (occasional)
- 2. Battery SoH 20-40% or cell voltage asymmetry (occasional)
- 3. Battery error ≥5% (rare)
Check battery state of health (SoH)
- Test point
- Battery data screen (Page 1500)
- Expected
- SoH percentage displayed
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Expiratory valve (adult/pediatric)
Adjust expiratory valve gain using external pressure gauge to ensure correct pressure reading and gas flow control during ventilation.
Ranked causes
- 1. Expiratory valve requires calibration after service or sensor drift (common)
- 2. Adult/pediatric expiratory valve requires separate calibration (common)
- 3. Calibration blocked by corrupted calibration file (rare)
Verify external pressure gauge reading during gain adjustment
- Test point
- Pressure connector 500300
- Expected
- 50 mbar +/- 0.5 mbar (51 cmH2O +/- 0.5 cmH2O), +/- 0.5 mbar
- Tools
- External pressure gauge
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Neonatal expiratory valve test fails patient safety
Component test for neonatal expiratory valve returns NOT OK result during automated testing in service software
Ranked causes
- 1. Pressure sensors Pexpvalve and Pvent_control not equal when proportional valve fully closed (common)
- 2. Expiratory proportional valve not opening fully or Pexpvalve not reading zero when valve opened (occasional)
- 3. Pressure test out of tolerance during automated measurement (occasional)
Compare pressure sensor readings
- Test point
- Pexpvalve and Pvent_control
- Expected
- Equal readings with patient system tightened and valve closed
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air entry filter obstruction or excessive pressure drop
Air entry (Pfilter) test shows NOT OK; filter pressure exceeds normal tolerance; airway entry test fails
Ranked causes
- 1. HEPA filter clogged or blocked with debris; high differential pressure (common)
- 2. NBC filter adapter (161787) installed with altered pressure tolerance (occasional)
Run air entry test with default HEPA filter installed
- Test point
- HEPA filter pressure (Pfilter)
- Expected
- Pfilter within defined tolerance; test displays OK result
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery replacement due to age
Battery must be replaced based on installation or manufacturing date with software version < 2.2.0
Ranked causes
- 1. Two years after installation date (software < 2.2.0) (common)
- 2. Three years after manufacturing date (software < 2.2.0) (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery showing marginal state of health or cell voltage asymmetry
SoH between 20-40% or cell voltage difference exceeds 100 mV across cells
Ranked causes
- 1. Battery state of health (SoH) 20-40% (occasional)
- 2. Cell voltages permanently asymmetrical (>100 mV difference) (occasional)
Check battery SoH percentage
- Test point
- Battery data page (Page No. 1500)
- Expected
- SoH < 40%
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery state of health low
Battery SoH <20% requires replacement; SoH <40% should be considered for replacement depending on software version
Ranked causes
- 1. Battery SoH < 20% (common)
- 2. Battery SoH < 40% (occasional)
Monitor battery state of health
- Test point
- Battery status indicator
- Expected
- SoH ≥ 40%
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Buzzer does not sound during power loss caution
No audible alarm when all power sources (AC, DC, batteries) are removed
Ranked causes
- 1. Supercapacitors (gold caps) not fully charged (common)
- 2. Control board defective or buzzer circuit faulty (occasional)
Charge supercapacitors before power loss test
- Test point
- Device power state
- Expected
- Device powered on in Stand-by or Service Mode for 20 minutes minimum
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Buzzer does not sound during power loss or sounds less than 2 minutes high voltage
Unit loses all AC, DC, and battery power; buzzer does not alarm or sounds for insufficient time
Ranked causes
- 1. Supercapacitors (gold caps) on control board not sufficiently charged (occasional)
- 2. Control board defective or buzzer circuit failure (rare)
Verify supercapacitor charge time before test
- Test point
- Device power state
- Expected
- Powered on for at least 20 minutes in Standby or Service Mode
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Flow sensor (adult/pediatric and neonatal)
Calibrate inspiratory and expiratory flow sensors by rotating sensor mechanically during automatic calibration to measure airflow accurately at different pressures.
Ranked causes
- 1. Flow sensor requires calibration with adult breathing circuit (common)
- 2. System overwrites calibration values when switching sensor type (occasional)
Confirm flow sensor rotation performed
- Test point
- Flow sensor position
- Expected
- Sensor mechanically rotated before calibration start
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: O2 cell (offset and gain) caution
Calibrate oxygen sensor offset using room air reference, then gain using high-pressure oxygen to ensure accurate O2 concentration measurement.
Ranked causes
- 1. O2 cell requires offset calibration using room air baseline (common)
- 2. O2 cell gain calibration follows offset calibration (common)
Monitor offset calibration result
- Test point
- O2 cell calibration display
- Expected
- OK
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Expiratory valve exhaust tubing disconnected or blocked patient safety
Expiratory proportional valve exhaust not connected to bleed port causing O2 enrichment inside ventilator
Ranked causes
- 1. Expiratory valve exhaust tubing disconnected from bleed port (common)
- 2. Expiratory valve tubing blocked or expiratory valve membrane defective (occasional)
Verify expiratory valve exhaust tubing is connected to bleed port
- Test point
- Bottom cover expiratory valve bleed port (outlet)
- Expected
- Tubing from expiratory proportional valve exhaust firmly connected
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Expiratory valve malfunction or obstruction patient safety
Exhalation obstructed warning or valve not operating; O2 enrichment occurring inside ventilator
Ranked causes
- 1. Expiratory valve bleed port tubing disconnected or blocked (common)
- 2. Expiratory valve membrane damaged or defective (occasional)
Verify expiratory valve bleed port tubing connected
- Test point
- Expiratory valve bleed port at bottom cover outlet
- Expected
- Tubing from proportional valve exhaust connected to outlet
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fan not producing proper suction or airflow caution
Fan inlet does not produce sufficient suction to hold paper or airflow is insufficient
Ranked causes
- 1. Fan not properly assembled or obstructed (common)
- 2. Fan motor defective or not operating (occasional)
Inspect fan assembly and inlet for obstructions
- Test point
- Fan housing and inlet opening
- Expected
- No visible obstructions or debris; fan properly seated in housing
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leak test not passing patient safety
System does not maintain pressure, or tightness test shows leakage when obstruction valve is tested or during pressurization cycle.
Ranked causes
- 1. Obstruction valve not functioning or leaking (common)
- 2. Patient breathing circuit or filter not properly sealed (common)
Verify breathing circuit and filter assembly
- Test point
- Patient connection port, expiratory connection port, flow sensor ports
- Expected
- All connections seated, no visual gaps or damage
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nebulizer system pressure loss or leak caution
Nebulizer tubing connection or internal valve leaking; pressure does not hold or rises unexpectedly
Ranked causes
- 1. Nebulizer tubing not properly connected (common)
- 2. Nebulizer valve internal leak or O2 mixer seal failure (occasional)
Connect hand pump to nebulizer port
- Test point
- Nebulizer port connection
- Expected
- Hand pump and tube properly connected
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nebulizer tubing or valve leaking caution
Pressure loss detected in nebulizer circuit or backpressure indicates internal valve leakage
Ranked causes
- 1. Nebulizer tubing loose or improperly connected (common)
- 2. Nebulizer valve defective or O2 mixer internal leak (occasional)
Test nebulizer tubing connection tightness
- Test point
- Nebulizer port and tubing connection (Di=4 mm/L=150 mm)
- Expected
- 60 mbar (56-64 mbar) pressure sustained without loss
- Tools
- Hand pump (500330), pressure indicator
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Neonatal expiratory valve test verification
Expiratory valve pressure test fails, or pressure sensor readings do not match expected values when valve is closed or opened.
Ranked causes
- 1. Pressure sensor Pexpvalve or Pvent_control drift or malfunction (occasional)
- 2. Expiratory proportional valve solenoid not responding or valve stuck (occasional)
Select neonatal mode in expiratory valve test window
- Test point
- Mode selector icon in window header
- Expected
- Neonatal mode active
7 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nurse call relay not responding to alarms caution
Alarm condition does not trigger proper relay contact closure/opening or I:E timing relay not switching
Ranked causes
- 1. Nurse call connector not properly connected (common)
- 2. Relay contact failure or stuck relay (occasional)
Connect nurse call connector
- Test point
- Communication board nurse call port
- Expected
- Connector or cable properly seated
5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure sensor assembly removal and installation caution
Procedure for removing and installing the pressure sensor assembly on HAMILTON-T1 ventilators across three serial number ranges (SN >= SN3000, SN2292-2999, and < SN2292), including tube disconnection and reconnection steps
Ranked causes
- 1. Maintenance or replacement of pressure sensor assembly required (unknown)
- 2. Reinstallation of pressure sensor assembly (unknown)
Disconnect tube 2 and 3 from check valve assembly (SN >= SN3000)
- Test point
- Check valve assembly (5)
- Expected
- Tubes disconnected
7 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
RS-232 communication interface not functioning
Ventilator does not communicate with patient monitoring system or Datalogger software
Ranked causes
- 1. RS-232 protocol not configured correctly (common)
- 2. Communication cable not connected or faulty (common)
Verify RS-232 protocol configuration
- Test point
- Utilities > Configuration > General > RS-232 Protocol
- Expected
- HAMILTON protocol selected
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
RS-232 communication interface not working
Communication with external patient monitoring system or Datalogger software fails or no signal detected
Ranked causes
- 1. RS-232 protocol not configured correctly (common)
- 2. Communication cable disconnected or defective (occasional)
Configure RS-232 protocol
- Test point
- Configuration menu RS-232 Protocol setting
- Expected
- HAMILTON protocol selected
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System tightness test failed
System pressure not maintained or drops beyond acceptable level indicating internal or circuit leakage
Ranked causes
- 1. Leak in system or breathing circuit tubing (common)
- 2. Obstruction valve malfunction (SN3000 and above) (occasional)
Run tightness test with test configuration 4 installed
- Test point
- System pressure maintenance
- Expected
- OK result displayed after test completion
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Technical state corrupt or failed
Self-test fails and displays 'Check hardware compatibility' or 'Technical state failed' message. Part causing failure highlighted in red in HW version tab.
Ranked causes
- 1. Incompatible hardware component combination installed (occasional)
- 2. Control board replaced or EEPROM communication interrupted (occasional)
Examine HW version tab for highlighted parts in red
- Test point
- HW version display
- Expected
- No parts highlighted in red; all components show normal status
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air entry test with obstruction
Filter obstruction test to verify system response to blocked air entry
Ranked causes
- 1. Filter obstruction or air entry blockage (unknown)
Place obstruction (plastic foil/paper) on air entry with small hole
- Test point
- Air entry obstruction
- Expected
- Obstruction properly positioned with 120 x 60 mm dimensions and small hole
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm lamp or sound not generating correctly
Alarm system not producing correct lamp illumination or sound output for alarm priorities
Ranked causes
- 1. Alarm system hardware or firmware malfunction (unknown)
Activate high-priority alarm and verify lamp and sound
- Test point
- Alarm lamp, alarm sound
- Expected
- Correct lamp illumination and sound output for high priority
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm monitor test 2 watchdog does not activate caution
Alarm LED fails to blink or buzzer fails to sound when watchdog activated in alarm monitor test 2
Ranked causes
- 1. Watchdog circuit malfunction or alarm output failure (occasional)
Verify watchdog activates alarm LED and buzzer
- Test point
- Watchdog circuit and alarm outputs
- Expected
- Alarm LED blinking and buzzer sounding
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm silence key does not respond at step 3 caution
Pressing alarm silence key fails to turn on alarm silence LED at step 3
Ranked causes
- 1. Alarm silence key or switch circuit failure (occasional)
Verify alarm silence key activates LED
- Test point
- Alarm silence key and LED
- Expected
- LED ON after key press
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm silence LED does not blink during initialization caution
Alarm silence LED fails to blink at step 1 of alarm monitor test 1
Ranked causes
- 1. Alarm LED circuit malfunction (occasional)
Verify alarm silence LED blinks at step 1
- Test point
- Alarm silence LED
- Expected
- LED blinking
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm silence LED does not turn off at step 4 caution
Alarm silence LED remains on instead of turning off at step 4 of alarm monitor test 1
Ranked causes
- 1. Alarm LED control circuit malfunction (occasional)
Verify alarm silence LED is OFF at step 4
- Test point
- Alarm silence LED
- Expected
- LED OFF
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm silence LED does not turn on at step 2 caution
Alarm silence LED fails to illuminate solid at step 2 of alarm monitor test 1
Ranked causes
- 1. Alarm LED circuit malfunction (occasional)
Verify alarm silence LED is ON at step 2
- Test point
- Alarm silence LED
- Expected
- LED ON
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm silence LED turns on at step 5 caution
Alarm silence LED turns on when it should remain off after key press at step 5
Ranked causes
- 1. Alarm silence key or control logic malfunction (occasional)
Verify alarm silence LED stays OFF at step 5
- Test point
- Alarm silence LED
- Expected
- LED stays OFF
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm test failure patient safety
Correct alarm lamp does not illuminate, alarm sound is not generated, alarm priority is incorrect, or loudness adjustment does not affect alarm output.
Ranked causes
- 1. Alarm system not responding to test command (occasional)
Activate High alarm
- Test point
- Alr.High button
- Expected
- High priority alarm lamp illuminates, alarm sound generates
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ambient state test does not trigger alarm patient safety
Alarm silence LED does not turn on and buzzer does not sound during ambient state test at step 8
Ranked causes
- 1. Ambient sensor or alarm circuit malfunction (occasional)
Verify alarm silence LED and buzzer activate in ambient state
- Test point
- Ambient sensor and alarm circuit
- Expected
- Alarm silence LED ON and buzzer sounds
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery calibration error 5% or greater
Battery calibration error meets or exceeds 5%; calibration cycle required
Ranked causes
- 1. Battery calibration drift accumulated over time (occasional)
Measure battery calibration error
- Test point
- Battery management system calibration status
- Expected
- Error < 5%, Within 5%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery calibration error at or above 5%
Battery charger reports calibration error of 5% or greater, indicating battery requires calibration
Ranked causes
- 1. Battery charge measurement drift or calibration drift (occasional)
Check battery calibration error on charger
- Test point
- Charger error display
- Expected
- Error < 5%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery calibration maintenance due
Annual preventive maintenance calibration cycle recommended for active battery
Ranked causes
- 1. Normal maintenance interval reached during active battery use (common)
Verify charger model and revision
- Test point
- Charger label
- Expected
- Model 369104 revision 07 or higher
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery calibration required
Battery error ≥ 5% indicates calibration is needed to maintain accurate battery status monitoring
Ranked causes
- 1. Battery error ≥ 5% (unknown)
Check battery error percentage
- Test point
- Battery status display
- Expected
- Error ≥ 5%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery disposed improperly patient safety
Battery disposed by incineration or burning; risk of explosion and environmental hazard
Ranked causes
- 1. Battery incinerated or burned at end of life (rare)
Verify disposal method planned
- Test point
- Disposal procedure documentation
- Expected
- Disposed per local regulations, not incinerated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery error exceeds 5%
Battery Error parameter on Battery data page shows value of 5% or greater
Ranked causes
- 1. Battery error ≥ 5% (occasional)
Check battery Error percentage
- Test point
- Battery data page (Page No. 1500) - Error field
- Expected
- Error < 5%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery State of Health below 20%
Battery SoH indicator shows value less than 20%, indicating battery requires replacement
Ranked causes
- 1. Battery age or degradation (common)
Check battery State of Health indicator
- Test point
- Battery status display
- Expected
- SoH reading
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery State of Health between 20-40%
Battery SoH indicator shows 20-40%; battery replacement should be considered
Ranked causes
- 1. Battery approaching end of service life (occasional)
Check battery State of Health indicator
- Test point
- Battery management system display
- Expected
- SoH value between 20-40%
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery State of Health between 20% and 40%
Battery SoH indicator shows value between 20% and 40%, indicating battery replacement should be considered
Ranked causes
- 1. Battery age or partial degradation (common)
Check battery State of Health indicator
- Test point
- Battery status display
- Expected
- SoH reading between 20% and 40%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery storage conditions maintenance caution
Preventive measures for battery preservation during storage to maintain battery life and readiness
Ranked causes
- 1. Improper storage conditions reduce battery shelf life (common)
Verify storage temperature
- Test point
- Storage environment
- Expected
- 5°C to 21°C; not exceeding 25°C over six-month periods
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blower replacement based on timer
Replace blower module when blower timer reaches 100% or predicted to reach 100% within 6 months; update serial and revision numbers.
Ranked causes
- 1. Blower timer at or approaching 100% (common)
Monitor blower timer percentage
- Test point
- Blower timer tab (Page 1201)
- Expected
- Blower timer value displayed in percentage
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blower service required alarm displayed
Alarm message 'Blower service required' appears on display when blower operating time reaches 100%
Ranked causes
- 1. Blower operating time limit reached 100% (common)
Check blower timer percentage on display
- Test point
- Blower timer tab (Page No. 1201)
- Expected
- ≥100% indicates replacement required
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blower timer reset required
After blower module replacement, blower timer must be reset to track service life of new blower module correctly.
Ranked causes
- 1. Blower module replaced with new serial number (common)
Access Modify tab and select blower module component
- Test point
- Technical state Modify screen
- Expected
- Blower module component selectable and editable
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Alarm monitor 1 test caution
Test alarm monitoring system components in 10 steps including ambient state, fan failure alarm, audio pause button and buzzer; requires manual confirmation at each step
Ranked causes
- 1. Alarm monitoring system component verification required (common)
Alarm monitoring system components operational
- Test point
- Electronics tab Alarm Mon.1 window
- Expected
- Test completed status displayed in Info window
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Alarm monitor 2 watchdog test caution
Activate software watchdog; verify alarm LED blinks and buzzer sounds; requires manual confirmation and ventilator reboot afterward
Ranked causes
- 1. Software watchdog and alarm circuit verification required (common)
Alarm LED blinks during watchdog activation
- Test point
- Alarm LED indicator
- Expected
- LED blinks when watchdog activated
- Tools
- Visual inspection
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Alarm system test
Test speaker, yellow lamp, red lamp, audio pause LED and loudness levels; confirms alarm audible output and visual indicators
Ranked causes
- 1. Alarm system component verification required (common)
Speaker audible test
- Test point
- Loudspeaker component
- Expected
- Audible alarm sound confirmed by technician
- Tools
- Ear inspection
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Autozero sequence test
Test pressure and flow sensors (Paw, Qaw, Pvent_monitor, Pvent_control) during autozero operation; requires bacteria/HME filter, breathing circuit, and blocked flow sensor end
Ranked causes
- 1. Sensor autozero functionality verification (common)
Autozero test configuration TC4 setup
- Test point
- Sensor circuit pneumatic connections
- Expected
- All components connected per TC4 diagram; flow sensor end blocked
- Tools
- Visual inspection against diagram
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Battery fuel gauge recalibration caution
Battery fuel gauge accuracy degrades over time; red LED flashes when battery inserted into charger indicating need for recalibration. Recalibration cycle includes discharge, calibration charge, calibration discharge, and regular charge phases.
Ranked causes
- 1. Battery fuel gauge accuracy degradation due to charge cycles without rest periods (common)
Observe red LED indicator on battery charger
- Test point
- Battery charger front panel LED
- Expected
- Red LED flashing
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Binary valves autozero test
Test auto zero valves (Pvent_monitor, Pflowsensor) during operation and autozero sequence; setup requires bacteria/HME filter, adult breathing circuit, flow sensor tubes and ET-tube
Ranked causes
- 1. Binary valve autozero functionality verification (common)
Binary valve test setup configuration TC3
- Test point
- Pneumatic circuit test points
- Expected
- All components connected per TC3 diagram; test ready to start
- Tools
- Visual inspection against diagram
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Calibration: Blower flow test
- Calibration: Blower pressure test
- Calibration: Default settings reset
- Calibration: Expiratory proportional valve (Adult/Ped.)
- Calibration: Expiratory valve
- Calibration: Expiratory valve (Neonatal)
- Calibration: Expiratory valve offset and gain display
- Calibration: Expiratory valve test
- Calibration: Flow sensor
- Calibration: Flow sensor (Adult/Ped.)
- Calibration: Flow sensor values display (inspiratory/expiratory)
- Calibration: Inspiratory tube resistance (Rinsp) values display
- Calibration: Neonatal expiratory valve offset voltage
- Calibration: Neonatal flow sensor
- Calibration: Neonatal flow sensor offset and linearity
- Calibration: O2 cell offset and gain values display
- Calibration: O2 sensor
- Calibration: Pressure sensor adjustment
- Calibration: Pressure sensor values display
- Calibration: Pressure sensors
- Calibration: software update completion
- Calibration: SpO2 sensor measurement
- Calibration: Touchscreen
- Calibration: User interface test
- Check valve assembly tightness or obstruction valve failure (SN >= SN3000)
- Check valve backflow detected
- Check valve flap visually damaged or misaligned
- Check valve one-way function failure (SN < SN3000)
- Check valve tightness fails (SN >= 3000)
- Display brightness does not switch between day and night modes
- Distorted sound in alarm loudness range
- Fan failure alarm does not trigger
- Fan not providing adequate airflow
- Filter pressure sensor reads incorrectly
- Flow sensor values out of tolerance
- Flow test verification
- Hardkeys do not light when pressed
- Nebulizer tube installation
- Nebulizer valve not operating
- O2 mixer sensor values out of tolerance
- O2 mixer test verification
- Obstruction valve fails to release pressure (SN >= 3000)
- P&T knob does not respond or rotate incorrectly
- Pressure test values out of tolerance
- Print screen function
- Proximal flow out of tolerance
- Proximal pressure out of tolerance
- Real time clock (RTC) requires adjustment
- Real-time clock (RTC) date/time adjustment
- Red lamp does not blink (Control Board Rev. >= 09)
- Red lamp does not illuminate (Control Board Rev. < 09)
- Reset to default calibration settings
- Rinse flow tank pressure does not hold
- Rinse flow unequal between outlets
- Sensor status interpretation needed
- Service date reset after maintenance
- Service timer reset
- Software update required after service procedures
- Speaker does not produce audible sound
- Touchscreen calibration required
- Unit does not power on from AC mains
- Unit does not switch to battery power
- Unit does not switch to DC power
- Unit does not transition from battery to DC power
- Unit does not transition from DC to battery power
- Unit does not transition to battery power
- Ventilator unused for extended period
- Yellow alarm lamp does not illuminate
68 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
100000 | Patient alarms - conditions affecting patient safety during ventilation | Patient-related adverse conditions detected by the ventilator | critical |
100000 | Patient alarms related to pressure, flow, volume, apnea | Detected condition in patient breathing parameters | unknown |
132001 | unknown | ||
132002 | Check flow sensor tubing | Flow sensor tubing issue detected | warning |
132003 | Check flow sensor | Flow sensor malfunction detected | warning |
132004 | Check flow sensor | Flow sensor missing or disconnected | warning |
132005 | O2 sensor defective | O2 sensor malfunction detected | error |
132006 | O2 sensor n/a | O2 sensor not available or disconnected | warning |
132007 | O2 sensor not system compatible | Incompatible O2 sensor installed | error |
132008 | O2 sensor calibration needed | O2 sensor requires calibration | warning |
132010 | CO2 sensor disconnected | CO2 sensor not connected | warning |
132011 | CO2 sensor defect | CO2 sensor malfunction detected | error |
132012 | CO2 sensor over temperature | CO2 sensor temperature exceeds limit | error |
132013 | CO2 sensor warmup | CO2 sensor warming up | info |
132015 | Flow sensor calibration needed | Flow sensor requires calibration | warning |
132016 | Check CO2 airway adapter | CO2 airway adapter issue detected | warning |
132017 | Check CO2 sampling line | CO2 sampling line issue detected | warning |
132018 | SpO2: patient disconnected | Patient SpO2 probe disconnected | critical |
132019 | SpO2: probe n/a | SpO2 probe not available or disconnected | warning |
132020 | SpO2: low perfusion index | SpO2 perfusion index below threshold | warning |
132021 | SpO2: light interference | Ambient light interfering with SpO2 sensor | warning |
132022 | SpO2: sensor error | SpO2 sensor malfunction detected | error |
132023 | SpO2: adapter n/a | SpO2 adapter not available or disconnected | warning |
132024 | Circuit calibration needed | Ventilator circuit requires calibration | warning |
132025 | SpO2: poor signal | SpO2 signal quality too low | warning |
132026 | Check flow sensor tubing | Flow sensor tubing disconnected | warning |
132027 | Check flow sensor tubing | Pdiff sensor disconnected | warning |
132028 | Flow sensor calibration needed | Flow sensor requires calibration (low priority) | warning |
133001 | unknown | ||
133002 | Inspiratory volume limitation | Tidal volume exceeds set limit | warning |
133003 | High pressure | Airway pressure exceeds limit | error |
133004 | Check for blockage | High pressure detected at patient interface | error |
133005 | Check for blockage | Medium pressure detected at patient interface | warning |
141001 | Pressure limitation | Pressure limiting function active | info |
141002 | Vt high | Tidal volume exceeds upper alarm threshold | warning |
141003 | Vt low | Tidal volume falls below lower alarm threshold | warning |
141004 | High minute volume | Minute ventilation exceeds alarm threshold | warning |
141005 | Low minute volume | Minute ventilation falls below alarm threshold | warning |
141006 | High frequency | Respiratory frequency exceeds alarm threshold | warning |
141007 | Low frequency | Respiratory frequency falls below alarm threshold | warning |
141008 | High oxygen | Oxygen concentration exceeds alarm threshold | warning |
141009 | Low oxygen | Oxygen concentration falls below alarm threshold | error |
141010 | Apnea | Patient apnea detected | critical |
141011 | Apnea ventilation ended | Apnea ventilation mode terminated | warning |
141012 | Apnea ventilation | Apnea backup ventilation activated | error |
141013 | High pressure during sigh | High pressure measured during sigh breath | error |
141014 | Flip the flow sensor | Flow sensor orientation incorrect | error |
141016 | Disconnection on ventilator side | Ventilator side circuit disconnected | critical |
141017 | Disconnection on patient side | Patient side circuit disconnected | critical |
141018 | Exhalation obstructed | Exhalation flow blocked or obstructed | error |
141019 | IRV | Inverse ratio ventilation mode active | info |
141020 | ASV: Cannot meet target | ASV unable to reach target ventilation | warning |
141021 | Pressure limit has changed | ASV pressure limit adjusted automatically | info |
141023 | External flow sensor failed | External flow sensor malfunction detected | error |
141024 | Low pressure | Airway pressure falls below lower threshold | warning |
141025 | PetCO2 high | End-tidal CO2 exceeds alarm threshold | warning |
141026 | PetCO2 low | End-tidal CO2 falls below alarm threshold | warning |
141027 | Suctioning maneuver | Suctioning procedure initiated | info |
141028 | Performance limited by high altitude | Ventilator at high altitude operation limit | warning |
141029 | High SpO2 | Oxygen saturation exceeds alarm threshold | warning |
141030 | Low SpO2 | Oxygen saturation falls below alarm threshold (medium) | warning |
141031 | High pulse | Pulse rate exceeds alarm threshold | warning |
141032 | Low pulse | Pulse rate falls below alarm threshold | warning |
141033 | High PI | Perfusion index exceeds alarm threshold | warning |
141034 | Low PI | Perfusion index falls below alarm threshold | warning |
141035 | High SpHb | Hemoglobin saturation exceeds alarm threshold | warning |
141036 | Low SpHb | Hemoglobin saturation falls below alarm threshold | warning |
141037 | High SpCO | Carboxyhemoglobin exceeds alarm threshold | warning |
141038 | Low SpCO | Carboxyhemoglobin falls below alarm threshold | info |
141039 | High SpMet | Methemoglobin exceeds alarm threshold | warning |
141040 | Low SpMet | Methemoglobin falls below alarm threshold | info |
141041 | Low SpO2 | Oxygen saturation falls below alarm threshold (high) | critical |
141043 | Ventilation adjustment OFF | Automatic ventilation adjustment disabled | info |
141044 | Oxygenation adjustment OFF | Automatic oxygenation adjustment disabled | info |
141045 | Ventilation Controller at limit | Ventilation control algorithm at upper limit | warning |
141046 | Oxygenation Controller at limit | Oxygenation control algorithm at upper limit | warning |
141047 | High flow | Gas flow exceeds alarm threshold in nCPAP mode | warning |
141048 | Obstruction | Circuit obstruction detected | error |
141049 | High flow | Gas flow exceeds alarm threshold | warning |
141050 | Sensor simulation active | Ventilator in sensor simulation mode | info |
141051 | High PEEP | PEEP exceeds alarm threshold | warning |
141052 | Loss of PEEP | PEEP falls below set value | error |
141053 | FiO2 set to 100% due to low SpO2 | Oxygen concentration automatically set to maximum | critical |
141054 | Recruitment in progress | Lung recruitment maneuver active | info |
141055 | Oxygen control limit exceeded | Oxygen delivery exceeds control algorithm limit | warning |
141056 | SBT aborted | Spontaneous breathing trial terminated | warning |
141057 | SBT successfully fulfilled | Spontaneous breathing trial completed successfully | info |
141058 | High PVI | Pleth variability index exceeds alarm threshold | warning |
141059 | Low PVI | Pleth variability index falls below alarm threshold | info |
141060 | Check flow sensor for water | Water detected in flow sensor circuit | warning |
141061 | Check flow sensor for water | Water detected in flow sensor circuit (high priority) | error |
141062 | SpeakValve ON | Speaking valve compatibility mode active | info |
141063 | SpeakValve OFF | Speaking valve compatibility mode inactive | info |
141064 | Vt low | Tidal volume falls below lower alarm threshold (high priority) | error |
141065 | Check for blockage | Patient interface blockage suspected | error |
141066 | Maximum leak compensation | Leak compensation at maximum limit | error |
141067 | Oxygen adjustment OFF | Closed-loop FiO2 adjustment disabled | info |
141068 | Oxygen controller at limit | Closed-loop FiO2 controller at upper limit | warning |
141069 | Oxygen msg limit exceeded | Closed-loop FiO2 oxygen message limit exceeded | warning |
141070 | CPR ON | Cardiopulmonary resuscitation mode active | critical |
141071 | PetCO2 target range changed | PetCO2 target range modified | info |
141072 | Check Plimit | Pressure limit requires attention | warning |
200000 | Patient alarms related to temperature, battery, valves, sensors | Detected condition in temperature, battery status, valve operation, or sensor readings | unknown |
200000 | Technical events - non-critical system events or warnings | Technical system events detected during operation | warning |
231001 | Technical event: Pressure controller pressure low | Pressure controller detects low pressure condition | error |
231002 | Technical event: Pressure controller pressure high | Pressure controller detects high pressure condition | error |
231003 | Technical event: Flow controller flow low | Flow controller detects low flow condition | error |
231004 | Technical event: Flow controller flow high | Flow controller detects high flow condition | error |
231005 | Technical event: Inspiration valve leak | Inspiration valve detects leak condition | error |
231006 | Oxygen supply failed | Oxygen flow below required level | error |
231007 | Technical event: O2 controller flow high | O2 controller detects high flow condition | error |
231008 | Technical event: O2 valve leak | O2 valve detects leak condition | error |
231009 | Technical event: Blower controller speed low | Blower controller detects low speed condition | error |
231010 | Technical event: Blower controller speed high | Blower controller detects high speed condition | error |
231011 | Vent outlet temperature high | Temperature sensor detects high outlet temperature | warning |
231012 | Technical event: Qvent flow sensor defect | Qvent flow sensor malfunction detected | error |
231013 | Technical event: QO2 flow sensor defect | QO2 flow sensor malfunction detected | error |
231014 | Technical event: Ambient valve error | Ambient valve malfunction detected | error |
231015 | Autozero Pvent control OK | Autozero procedure completed successfully | info |
231016 | Autozero Pvent control out of range | Autozero calibration failed | warning |
231017 | Blower service required | Blower maintenance interval reached or malfunction detected | warning |
231018 | Technical event: Self-test with error | Power-on self-test or periodic self-test failed | error |
231019 | Technical event: Inspiration valve sensor defect | Inspiration valve sensor malfunction detected | error |
231020 | Technical event: Regulator data log on | Data logging feature activated | info |
231021 | O2 presence check failed | O2 detection circuit detects absence of oxygen | warning |
231022 | Technical event: Expiration valve pressure sensor defect | Expiration valve pressure sensor malfunction detected | error |
231023 | Technical event: Invalid flow sensor | Flow sensor configuration or detection failed | error |
231024 | Expiration valve cover adult | Adult expiration valve cover detected | info |
231025 | Expiration valve cover neo | Neonatal expiration valve cover detected | info |
231026 | Expiration valve cover invalid | Unknown or invalid expiration valve cover detected | error |
231027 | Wrong expiratory valve | Expiration valve mismatch detected | error |
231028 | Pressure controller flow high | Pressure controller detects excessive flow | warning |
231029 | Autozero Pexpiration valve OK | Expiration valve autozero completed successfully | info |
231030 | Autozero Pexpiration valve error | Expiration valve autozero failed | warning |
231031 | Expiration valve cover universal | Universal expiration valve cover detected | info |
231032 | Technical event: Expiration valve disconnected | Expiration valve connection lost or missing | error |
231033 | Technical event: Blower controller speed limited | Blower speed limited by controller | warning |
231034 | Tube resonance measurement OK | Tube resonance measurement completed successfully | info |
231035 | Tube resonance measurement error | Tube resonance measurement failed | warning |
231036 | Technical event: Pvent pressure sensor defect | Pvent pressure sensor malfunction detected | error |
231040 | Calibrate Pvent control OK | Pvent pressure control calibration completed successfully | info |
231041 | Calibrate Pvent control error | Pvent pressure control calibration failed | error |
231044 | Calibrate expiration valve OK | Expiration valve calibration completed successfully | info |
231045 | Calibrate expiration valve error | Expiration valve calibration failed | error |
231046 | Calibrate inspiration/expiration valve OK | Inspiration/expiration valve calibration completed successfully | info |
231047 | Calibrate inspiration/expiration valve error | Inspiration/expiration valve calibration failed | error |
231099 | Communication check | Inter-processor or system communication check | info |
232001 | Pressure sensor tolerance | Pressure sensor reading out of specification | warning |
232002 | Technical event: Pvent monitor sensor defect | Pvent monitor sensor malfunction detected | error |
232003 | Technical event: Paw sensor defect | Airway pressure sensor malfunction detected | error |
232004 | Pressure not released | System pressure not relieved as expected | error |
232005 | Technical event: Blower hot | Blower temperature exceeds limit | warning |
232006 | Technical event: Blower temperature sensor defect | Blower temperature sensor malfunction detected | error |
232007 | Check flow sensor tubing | Flow sensor tubing malfunction or blockage detected | warning |
232008 | Technical event: Pambient sensor defect | Ambient pressure sensor malfunction detected | error |
232009 | Pvent autozero valve error | Pvent autozero valve malfunction detected | warning |
232010 | Pvent autozero valve missing | Pvent autozero valve not detected | error |
232011 | Proximal autozero valve error | Proximal autozero valve malfunction detected | warning |
232012 | Proximal autozero valve missing | Proximal autozero valve not detected | error |
232013 | Distal autozero valve error | Distal autozero valve malfunction detected | warning |
232014 | Distal autozero valve missing | Distal autozero valve not detected | error |
232015 | Proximal/distal autozero valve missing | Proximal or distal autozero valve not detected | error |
232016 | Proximal/distal autozero valve error | Proximal or distal autozero valve malfunction detected | warning |
232017 | Autozero Pvent monitor OK | Pvent monitor autozero completed successfully | info |
232018 | Autozero Pvent monitor out of range | Pvent monitor autozero failed | warning |
232019 | Autozero Pvent control OK | Pvent control autozero completed successfully | info |
232020 | Autozero Pvent control out of range | Pvent control autozero failed | warning |
232022 | Autozero Paw out of range | Airway pressure autozero failed | warning |
232024 | Autozero Qaw out of range | Airway flow autozero failed | warning |
232025 | Autozero Paw/Qaw OK | Airway pressure and flow autozero completed successfully | info |
232026 | Autozero Paw/Qaw out of range | Airway pressure and flow autozero failed | warning |
232027 | Device temperature high | Internal device temperature exceeds safe operating limit | warning |
232028 | Technical event: Filter pressure high | Filter inlet pressure exceeds limit | warning |
232029 | Technical event: Tinst sensor defect | Temperature sensor malfunction detected | error |
232030 | Technical event: Clock error | Real-time clock malfunction detected | warning |
232031 | FiO2 calibration error | FiO2 sensor calibration failed | warning |
232032 | FiO2 calibration OK | FiO2 sensor calibration completed successfully | info |
232034 | Replace O2 sensor | O2 sensor defect or end-of-life detected | error |
232035 | Technical event: Pfilter sensor defect | Filter pressure sensor malfunction detected | error |
232036 | CO2 calibration error | CO2 sensor calibration failed | warning |
232037 | CO2 calibration OK | CO2 sensor calibration completed successfully | info |
232038 | CO2 calibration needed | CO2 sensor requires periodic recalibration | warning |
232040 | Quick disconnection detected | Gas supply line disconnection detected | info |
232041 | CO2 calibration start error | CO2 calibration procedure failed to start | warning |
232042 | CO2 calibration start OK | CO2 calibration procedure started successfully | info |
232043 | Quick reconnection detected | Gas supply line reconnected | info |
232044 | SpO2 sensor error | SpO2 sensor malfunction detected | warning |
232045 | Technical event: SpO2 sensor settings error | SpO2 sensor configuration invalid or missing | error |
232046 | Autozero expiration valve OK | Expiration valve autozero completed successfully | info |
232047 | Autozero expiration valve error | Expiration valve autozero failed | warning |
232048 | Autozero proximal valve leak | Proximal valve leak detected during autozero | warning |
232049 | Autozero distal valve leak | Distal valve leak detected during autozero | warning |
232050 | Pvent monitor calibration error | Pvent monitor calibration failed | warning |
232051 | Pvent monitor calibration OK | Pvent monitor calibration completed successfully | info |
232052 | Paw calibration error | Airway pressure calibration failed | warning |
232053 | Paw calibration OK | Airway pressure calibration completed successfully | info |
232054 | Qaw calibration error | Airway flow calibration failed | warning |
232055 | Qaw calibration OK | Airway flow calibration completed successfully | info |
232056 | Technical event: Pambient/Pfilter mismatch | Ambient and filter pressure sensor values inconsistent | warning |
232057 | Technical event: SpO2 sensor change failed | SpO2 sensor replacement or configuration change failed | error |
232058 | Paux sensor defect | Auxiliary pressure sensor malfunction detected | error |
232059 | Paux calibration error | Auxiliary pressure sensor calibration failed | warning |
232060 | Paux calibration OK | Auxiliary pressure sensor calibration completed successfully | info |
232061 | FiO2 calibration start error | FiO2 calibration procedure failed to start | warning |
232062 | FiO2 calibration start OK | FiO2 calibration procedure started successfully | info |
232063 | Qaw and tube calibration at flow timed out | Flow and tube calibration procedure exceeded time limit | warning |
232064 | Qaw and tube calibration at flow passed | Flow and tube calibration procedure completed successfully | info |
232065 | Qaw and tube calibration at flow failed | Flow and tube calibration procedure failed | error |
232099 | Communication check | Inter-processor or system communication check | info |
233001 | Technical event: Autozero Pvent monitor fail | Pvent monitor autozero procedure failed | error |
233002 | Technical event: Autozero Pvent control fail | Pvent control autozero procedure failed | error |
233008 | Aerogen nebulizer disconnected | Aerogen nebulizer is physically disconnected from the ventilator | warning |
234001 | Loudspeaker defective | Speaker hardware failure detected during startup or operation | error |
241001 | Replace HEPA filter | HEPA filter maintenance interval exceeded or filter clogged | warning |
241002 | Release valve defective | Tightness test detects release valve malfunction | error |
243004 | Buzzer defective | Buzzer test fails at startup or buzzer produces no sound | error |
243007 | Realtime clock failure | RTC chip failure or battery depleted during power loss | error |
243010 | Panel connection lost | Communication link between main board and control panel fails | error |
244001 | Loss of external power | Mains power supply disconnected or failed | warning |
244002 | Battery 1: Calibration required | Battery 1 capacity measurement indicates need for recalibration | warning |
244003 | Battery 2: Calibration required | Battery 2 capacity measurement indicates need for recalibration | warning |
244004 | Battery 1: Temperature high | Battery 1 temperature exceeds safe operating range during charge/discharge | warning |
244005 | Battery 2: Temperature high | Battery 2 temperature exceeds safe operating range during charge/discharge | warning |
244006 | Battery low | Battery charge level drops below high-priority threshold during operation | warning |
244007 | Battery power loss | Battery becomes disconnected or fails suddenly during operation | critical |
244008 | Battery 1: Wrong battery | Battery 1 type or model does not match system requirements | error |
244009 | Battery 2: Wrong battery | Battery 2 type or model does not match system requirements | error |
244010 | Battery low | Battery charge level drops below low-priority threshold during operation | warning |
244011 | Battery communication error | Battery management system loses communication with battery module | error |
244012 | Battery 1: Defective | Battery 1 self-test fails or capacity measurement indicates failure | error |
244013 | Battery 2: Defective | Battery 2 self-test fails or capacity measurement indicates failure | error |
244015 | Battery low | Battery charge level drops below medium-priority threshold during operation | warning |
244016 | Battery 1: Replacement required | Battery 1 has reached end of service life or unreliable performance detected | warning |
244017 | Battery 2: Replacement required | Battery 2 has reached end of service life or unreliable performance detected | warning |
244022 | Continue charging battery | Battery charging in progress, normal state during AC power connected | info |
246001 | Preventive maintenance required | Equipment operating hours or calendar interval exceeds maintenance schedule | warning |
246003 | Fan failure | Cooling fan stops or runs abnormally, temperature control compromised | error |
246016 | Touch not functional | Touch screen detects no response to finger contact or self-test fails | error |
246020 | Loudspeaker defective | Speaker hardware fails during HMI self-test or audio output ceases | error |
249001 | Options not found | System cannot read or locate option configuration file at startup | error |
249012 | Technical event | Last setting config file error | unknown |
249013 | Technical event | Research config file error | unknown |
249020 | Invalid communication board | Option board not recognized | error |
249021 | Check hardware compatibility | Hardware compatibility issue detected | warning |
250001 | Check humidifier communication | Humidifier communication failure | error |
250002 | Humidifier water low | Water level below threshold | warning |
250003 | Humidifier chamber temp low | Chamber temperature below setpoint | warning |
250004 | Humidifier Y-piece temp low | Y-piece temperature below setpoint | warning |
250005 | Humidifier check right tube | Right heating tube fault detected | warning |
250006 | Humidifier check left tube | Left heating tube fault detected | warning |
250007 | Humidifier check chamber | Chamber component fault detected | error |
250008 | Check humidifier | General humidifier fault | error |
250009 | Humidifier error | Unspecified humidifier error | error |
250010 | Humidifier water high | Water level above threshold | unknown |
250011 | Humidifier chamber temp high | Chamber temperature above setpoint | warning |
250012 | Humidifier Y-piece temp high | Y-piece temperature above setpoint | warning |
250013 | Humidifier tilt | Humidifier tilted or misaligned | unknown |
250014 | Check humidifier | Humidifier fault detected | error |
250015 | Check breathing circuit limbs | Breathing circuit limb fault detected | warning |
251001 | Check IntelliCuff communication | IntelliCuff communication loss | error |
251002 | Check IntelliCuff | IntelliCuff low fault | warning |
251003 | Cannot turn off IntelliCuff | IntelliCuff power-off failure | error |
251004 | Check IntelliCuff | IntelliCuff battery low | warning |
251005 | Cuff pressure high | Endotracheal cuff pressure exceeds limit | warning |
251006 | Cuff deflated | Cuff deflation detected | warning |
251007 | Cuff leak | Cuff leak detected | warning |
251008 | Check IntelliCuff | IntelliCuff medium fault | warning |
251009 | Cuff leak | IntelliCuff disconnection detected | warning |
251010 | Check IntelliCuff | IntelliCuff error | error |
251011 | Check IntelliCuff | IntelliCuff battery critical | critical |
251012 | Check IntelliCuff | IntelliCuff high fault | error |
251013 | Check IntelliCuff state | IntelliCuff state error | error |
252001 | Check TeslaSpy communication | TeslaSpy communication loss | error |
252002 | Reduce MRI scanner proximity | High magnetic field proximity detected | warning |
252003 | Service required | Magnetic field too high detected | error |
252004 | TeslaSpy defective | TeslaSpy unit fault detected | error |
252005 | TeslaSpy service required | TeslaSpy service condition detected | error |
255001 | Technical event | Percentage minimum volume high test failed | unknown |
255002 | Technical event | Percentage minimum volume high limit incorrect | unknown |
255003 | Technical event | Percentage minimum volume low test failed | unknown |
255004 | Technical event | Percentage minimum volume low limit incorrect | unknown |
255005 | Technical event | PEEP high limit incorrect | unknown |
255006 | Technical event | PEEP high test failed | unknown |
255007 | Technical event | PEEP low limit incorrect | unknown |
255008 | Technical event | PEEP low test failed | unknown |
255009 | Technical event | Oxygen low test failed | unknown |
255010 | Technical event | Oxygen high test failed | unknown |
255011 | Technical event | Percentage minimum volume positive change failed | unknown |
255012 | Technical event | Percentage minimum volume negative change failed | unknown |
255013 | Technical event | PEEP positive change test failed | unknown |
255014 | Technical event | PEEP negative change test failed | unknown |
255015 | Technical event | Oxygen positive change test failed | unknown |
255016 | Technical event | Oxygen negative change test failed | unknown |
255017 | Technical event | Recruitment maneuver runtime exceeded | unknown |
255018 | Technical event | Simulation supervision test failed | unknown |
255019 | Technical event | Percentage minimum volume high fine test failed | unknown |
255020 | Technical event | Percentage minimum volume high fine limit wrong | unknown |
255021 | Technical event | Percentage minimum volume low fine test failed | unknown |
255022 | Technical event | Percentage minimum volume low fine limit wrong | unknown |
255023 | Oscillation Oxygen | Oxygen parameter oscillation detected in closed-loop | warning |
255024 | Oscillation PEEP/CPAP | PEEP parameter oscillation detected in closed-loop | warning |
255025 | Oscillation %MinVol | Percentage minimum volume oscillation in closed-loop | warning |
255026 | Technical event | Data mismatch detected | unknown |
255027 | Technical event | PetCO2 value frozen in closed-loop | unknown |
255028 | Technical event | SpO2 value frozen in closed-loop | unknown |
255029 | Technical event | Spontaneous breathing frequency frozen | unknown |
255030 | Technical event | Closed-loop FiO2 oxygen low test failed | unknown |
255031 | Technical event | Closed-loop FiO2 oxygen high test failed | unknown |
255032 | Technical event | Closed-loop FiO2 oxygen positive change failed | unknown |
255033 | Technical event | Closed-loop FiO2 oxygen negative change failed | unknown |
255034 | Technical event | Closed-loop FiO2 simulation supervision failed | unknown |
255035 | Technical event | Closed-loop FiO2 SpO2 value frozen | unknown |
281001 | Firmware mismatch | PLD firmware version incompatible | error |
281002 | JTAG not working | JTAG programming interface not responding | error |
281003 | Check hardware compatibility | Hardware compatibility issue detected | warning |
283001 | Technical event | Ventilator GUI startup failed | unknown |
283003 | Language not loaded | Language file failed to load | warning |
283005 | Settings file error | Setup configuration file error | error |
283008 | Check settings | Settings configuration error detected | warning |
283009 | Self test failed | Power-on self test failure | error |
283010 | No ventilation after power-fail | Ventilation not resumed after power loss | critical |
283011 | Check IntelliCuff state | IntelliCuff mode error | error |
283012 | Panel settings file error | Panel setup configuration file error | error |
284001 | Startup failed | Service GUI startup failure | error |
284002 | Technical event | Service alarm high priority | unknown |
284003 | Technical event | Service alarm medium priority | unknown |
284004 | Technical event | Service alarm low priority | unknown |
284005 | Flow sensor calibration needed | Flow sensor requires recalibration | warning |
284006 | Restart device to enable Tests/Calib | Device restart required for tests or calibration | info |
285001 | Technical event | Alarm lamp error indicator defect | unknown |
285002 | Technical event | Alarm lamp warning indicator defect | unknown |
285003 | Technical event | Backlight defect detected | unknown |
285004 | Function key not operational | Function key malfunction detected | warning |
285005 | Touch not functional | Touchscreen malfunction detected | error |
291001 | Remote communication error (COM1) | COM1 port communication error | error |
291002 | Remote communication error (COM2) | COM2 port communication error | error |
291003 | Remote communication timeout (COM1) | COM1 communication timeout | warning |
300000 | Technical fault - ventilator enters safety mode | Fatal technical failure detected during operation | critical |
332001 | Safety ventilation activated due to flow sensor error | Breath management module detects flow sensor malfunction | critical |
341008 | Safety ventilation activated due to ventilation flow sensor defect | Ventilation monitoring controller detects flow sensor failure | critical |
341009 | Safety ventilation activated due to pressure sensor defect | Ventilation monitoring controller detects pressure sensor failure | critical |
346054 | Safety ventilation activated due to detected safety failure | Low-level system detects critical safety condition | critical |
383008 | Technical fault due to basic startup failure | Ventilator GUI fails basic startup sequence | error |
383009 | Technical fault due to screen failure | Ventilator GUI screen or display initialization fails | error |
383012 | Technical fault due to self-test failure | Ventilator GUI self-test sequence fails during startup | error |
384001 | Technical fault due to basic startup failure in settings GUI | Settings GUI fails basic startup sequence | error |
400000 | Technical fault - ventilator enters ambient state | Serious technical fault risking compromise of safe ventilation | critical |
431001 | Blower fault | Blower hardware failure or malfunction | critical |
431002 | Technical fault: Blower disconnected | Blower connection lost or unplugged | error |
431004 | Technical fault: Inspiration valve over-current | Inspiration valve drawing excessive current | error |
431006 | Technical fault: Expiration valve over-current | Expiration valve drawing excessive current | error |
431007 | Technical fault: Ventilation output over-temperature | Outlet temperature exceeds safe limit | error |
431008 | Technical fault: Ventilation flow sensor error | Flow sensor malfunction or measurement failure | error |
431009 | Technical fault: Oxygen flow sensor error | O2 flow sensor malfunction or measurement failure | error |
431010 | Technical fault: Control REGSPI timeout | Control register SPI communication timeout | error |
431011 | Technical fault: Monitor REGSPI timeout | Monitor register SPI communication timeout | error |
431012 | Technical fault: Flow sensor measurement imprecise | Flow sensor output out of expected range or unstable | warning |
431013 | Technical fault: Calibration read failed | Unable to read calibration data from memory | error |
431014 | Technical fault: Inspiration valve sensor defect | Inspiration valve position sensor failure | error |
431015 | Technical fault: Emergency off failed | Emergency stop function did not execute properly | critical |
431017 | Technical fault: Inspiration valve disconnected | Inspiration valve connector unplugged or lost connection | error |
432001 | Technical fault: Blower over-temperature | Blower motor temperature exceeds safe operating limit | error |
432002 | Technical fault: Instrument over-temperature | Internal instrument temperature exceeds safe limit | error |
432003 | Technical fault: Pressure not released to ambient | System unable to release pressure to atmosphere during ambient mode | error |
433001 | Technical fault: Breath monitoring tick timeout | Breath monitoring process did not respond within timeout period | error |
433002 | Technical fault: Control tick timing error | Control loop timing violation or delay exceeds tolerance | error |
443001 | Technical fault: Watchdog failed LLS | Low-level system watchdog timer failed to reset | critical |
444001 | Battery totally discharged | Battery charge level reached zero | critical |
444004 | Technical fault: Voltage out of tolerance | Power supply voltage outside acceptable operating range | error |
444005 | Technical fault: Shutdown failed | System unable to complete normal shutdown sequence | error |
446001 | Technical fault: CPU temperature critical | Processor temperature exceeds maximum safe operating temperature | critical |
446002 | Technical fault: Safety failed | Safety system self-test or monitoring failed | critical |
446003 | Technical fault: Watchdog failed GD | Gas delivery module watchdog timeout occurred | critical |
446004 | Technical fault: Watchdog failed GD valve regulator | Gas delivery valve regulator watchdog timeout | critical |
446005 | Technical fault: Watchdog failed GD blower regulator | Gas delivery blower regulator watchdog timeout | critical |
446006 | Technical fault: Watchdog failed GD monitoring | Gas delivery monitoring watchdog timeout | critical |
446007 | Technical fault: Watchdog failed GD valve log | Gas delivery valve logging watchdog timeout | error |
446008 | Technical fault: Watchdog failed GD blower log | Gas delivery blower logging watchdog timeout | error |
446009 | Technical fault: Watchdog failed GD MF measure | Gas delivery measurement function watchdog timeout | error |
446010 | Technical fault: Watchdog failed GD GPIO fault input | Gas delivery GPIO fault monitoring watchdog timeout | error |
446011 | Technical fault: Watchdog failed BM | Breath monitoring module watchdog timeout | critical |
446012 | Technical fault: Watchdog failed BM gas delivery communication | Breath monitoring to gas delivery communication watchdog timeout | critical |
446013 | Technical fault: Watchdog failed BM hardware control | Breath monitoring hardware control watchdog timeout | critical |
446014 | Technical fault: Watchdog failed BM BPG client receiver | Breath monitoring BPG client receiver watchdog timeout | critical |
446015 | Technical fault: Watchdog failed BM BPG server receiver | Breath monitoring BPG server receiver watchdog timeout | critical |
446016 | Technical fault: Watchdog failed BM breath data calculation | Breath data calculation watchdog timeout | critical |
446017 | Technical fault: Watchdog failed BM MF measure | Breath monitoring measurement function watchdog timeout | error |
446018 | Technical fault: Watchdog failed BPG | Breath pattern generation module watchdog timeout | critical |
446019 | Technical fault: Watchdog failed BPG GD client receiver | BPG gas delivery client receiver watchdog timeout | critical |
446020 | Technical fault: Watchdog failed BPG MF measure | BPG measurement function watchdog timeout | error |
446021 | Technical fault: Exception happened | Unhandled software exception occurred in system | critical |
446022 | Technical fault: Voltage error | Internal voltage supply anomaly detected | critical |
446023 | Technical fault: Watchdog failed LLS workload | Low-level system workload monitoring watchdog timeout | critical |
446024 | Technical fault: Am 3V3 error | 3.3V analog supply failure detected | critical |
446025 | Technical fault: Am ADC error | Analog-to-digital converter failure detected | error |
446026 | Technical fault: Am Vref error | Analog reference voltage failure detected | error |
446028 | Technical fault: Clock error | System clock or timing reference failure | critical |
446029 | Technical fault: Ambient failure detected | Ambient monitoring system detected a critical failure | error |
446030 | Technical fault: Am watchdog occurred | Analog module watchdog timeout occurred | error |
446031 | Technical fault: Restart ventilation failed | System unable to restart ventilation after ambient state | critical |
446032 | Technical fault: Patient vent overpressure | Patient ventilation outlet pressure exceeds maximum safe limit | critical |
481001 | Technical fault: First run migration error | Startup configuration migration failed | error |
481002 | Technical fault: CRC error | Data integrity check (CRC) failed on stored data | error |
481003 | Unknown part number | System encountered unrecognized part or configuration ID | error |
481004 | Technical state failed | System technical state verification failed at startup | error |
481005 | Technical fault: Synchronization error | System module synchronization failed during startup | error |
481006 | Technical fault: Reconnect startup type error | Startup type or mode recovery failed | error |
481007 | Technical fault: Reconnect accept error | System reconnection handshake failed during startup | error |
483001 | Technical fault: Ventilator GUI startup basic startup failure | Ventilator GUI failed basic startup checks | error |
483002 | Technical fault: Ventilator GUI startup screen failure | Ventilator GUI display screen failed to initialize | error |
483003 | Technical fault: Ventilator GUI startup load GUI basics failure | Ventilator GUI basic components failed to load | error |
483004 | Technical fault: Ventilator GUI startup self-test prolog failure | Ventilator GUI self-test prolog phase failed | error |
483005 | Technical fault: Ventilator GUI startup self-test failure | Ventilator GUI self-test execution failed | error |
483006 | Technical fault: Ventilator GUI startup self-test epilog failure | Ventilator GUI self-test epilog phase failed | error |
483007 | Technical fault: Ventilator GUI startup loading bitmaps failure | Ventilator GUI failed to load display bitmaps or graphics | error |
483008 | Technical fault: Ventilator GUI startup initialization failure | Ventilator GUI initialization sequence failed | error |
484001 | Technical fault: Setting GUI startup basic startup failure | Setting GUI failed basic startup checks | error |
484002 | Technical fault: Setting GUI startup screen failure | Setting GUI display screen failed to initialize | error |
484003 | Technical fault: Setting GUI startup load GUI basics failure | Setting GUI basic components failed to load | error |
484004 | Technical fault: Setting GUI startup self-test prolog failure | Setting GUI self-test prolog phase failed | error |
484005 | Technical fault: Setting GUI startup self-test failure | Setting GUI self-test execution failed | error |
484007 | Technical fault: Setting GUI startup loading bitmaps failure | Setting GUI failed to load display bitmaps or graphics | error |
484008 | Technical fault: Setting GUI startup initialization failure | Setting GUI initialization sequence failed | error |
485001 | Technical fault: Alarming GUI ambient failure detected | Alarming GUI detected system in ambient/unpowered state failure | error |
500000 | System failures invisible to user, internal use only, no user action required | Internal system error detected and managed by ventilator firmware | info |
500000 | Invisible technical fault - internal exception handling | Exception handling during ventilator operation | info |
700000 | Technical fault - ventilator enters safety mode | Fatal technical failure requiring patient safety intervention | critical |
743001 | Technical fault: Watchdog failed LLS | Panel module low-level system watchdog timeout | critical |
743003 | Technical fault: MMI connection lost | Man-machine interface connection lost or disconnected | error |
746001 | Technical fault: Watchdog failed ALR | Panel low-level system watchdog detected ALR module timeout | critical |
746002 | Technical fault: Watchdog failed VGUI | Panel low-level system watchdog detected ventilator GUI timeout | critical |
746003 | Technical fault: Watchdog failed VGUI mode control | Ventilator GUI mode control watchdog timeout on panel | critical |
746004 | Technical fault: Panel error failure detected | Panel hardware or communication error detected | error |
746005 | Technical fault: Exception happened | Unhandled exception in panel module occurred | critical |
746006 | Technical fault: Watchdog failed LLS workload | Panel low-level system workload monitoring watchdog timeout | critical |
746007 | Technical fault: Restart ventilation failed | Panel unable to restart ventilation after ambient state | critical |
783001 | Technical fault: Monitoring channel observation failed | Ventilator GUI monitoring channel failed to update or respond | error |
783002 | Technical fault: Trending channel observation failed | Ventilator GUI trending data channel failed to update | error |
783003 | Technical fault: VMC timeout | Ventilator mode control communication timeout occurred | error |
785001 | Technical fault: Alarming channel observation failed | Alarm channel failed to update or respond in alarming GUI | error |
785002 | Technical fault: Panel error failure detected | Panel error detected in alarming GUI module | error |
785003 | Technical fault: PM channel observation failed | Power management channel failed to update or respond | error |
Check hardware compatibility | Incompatible combination of hardware components detected during self-test | System detects installed hardware components that are not compatible with each other | critical |
C-SS-NNN | Alarm codes are structured with alarm class, subsystem, and ID number; C=class, SS=subsystem, NNN=ID | Various conditions trigger different alarm classes and severities based on patient safety impact | unknown |
Exhalation obstructed | Expiratory valve is blocked or obstruction detected in expiration pathway | Expiratory valve bleed port tubing disconnected, blocked, or valve membrane damaged; blockage applied during test | warning |
Technical state failed | Technical state cannot be read from EEPROM of a component or after control board replacement | Unit fails self-test due to corrupt or unreadable technical state | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI