Hamilton Medical HAMILTON-T1: troubleshooting and error codes

The ServiceForge knowledge base holds 128 documented symptoms for the Hamilton Medical HAMILTON-T1, with 181 ranked causes, 301 diagnostic checks with test points and expected values, and 446 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Sensor status shows error or invalid state

Sensor data displays abnormal status values (hex) indicating defect, missing, overflow, underflow, temperature, or invalid conditions

Ranked causes

  1. 1. Sensor defect status (0004) (rare)
  2. 2. Sensor missing or not running (0008) (occasional)
  3. 3. Autozero sequence in progress (0010) (occasional)
  4. 4. Measurement overflow (0020) or underflow (0040) (occasional)
  5. 5. Temperature error - undertemperature (0100) or overtemperature (0200) (rare)
  6. 6. Sensor disabled (0080) or invalid (0400) (rare)
  7. 7. Sampling line occluded (2000) (occasional)
First check

Read sensor status hex value

Test point
Sensor 1/2/3 status field
Expected
0003 or similar with only Initialized (0001) and Calibrated (0002) flags set
Tools
Service software display

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Service software startup and technical state management caution

Procedure for starting HAMILTON-T1 service software, managing hardware/software versions, updating technical state after component replacement, and resetting blower timer.

Ranked causes

  1. 1. Initial service software startup and self-test required (common)
  2. 2. Technical state information needs recording and verification (common)
  3. 3. Component replacement requires technical state modification and restart (common)
  4. 4. Technical state corruption or EEPROM read failure requires cleanup (occasional)
  5. 5. Export and import technical state to/from USB drive (occasional)
First check

Verify service software startup sequence completes

Test point
Main service software screen (page 0000)
Expected
Service software main screen displayed with no startup errors

4 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 measurement not displaying or incorrect values patient safety

SpO2 sensor connected but values not shown in monitoring tab or dynamic lung view, or readings are invalid

Ranked causes

  1. 1. SpO2 communication board not installed or not detected (occasional)
  2. 2. SpO2 hardware option not activated in configuration (common)
  3. 3. Incorrect SpO2 sensor type configured (common)
  4. 4. SpO2 sensor not activated or not connected to board (occasional)
  5. 5. Finger probe not properly attached or insufficient signal (occasional)
First check

Verify SpO2 communication board installed

Test point
System Info display
Expected
SpO2 communication board listed in installed hardware

5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery storage conditions inadequate caution

Battery stored at incorrect temperature or humidity; risk of reduced shelf life and performance

Ranked causes

  1. 1. Storage temperature exceeds 25°C over 6-month recharge interval (occasional)
  2. 2. Battery stored in discharged state (rare)
  3. 3. Extended exposure to extreme temperatures (rare)
  4. 4. Adverse storage environment (vibration, moisture, corrosive gas, etc.) (rare)
First check

Verify storage temperature

Test point
Storage location
Expected
Between 5°C and 21°C; do not exceed 25°C over 6-month intervals, 5-21°C optimal; max 25°C

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Proximal test failures

Rinse flow, proximal pressure, proximal flow, or rinse tank test shows NOT OK; abnormal bubble patterns or pressure readings detected

Ranked causes

  1. 1. Rinse flow obstruction or unequal flow distribution (unknown)
  2. 2. Proximal pressure sensor malfunction or calibration drift (unknown)
  3. 3. Proximal flow sensor fault (unknown)
  4. 4. Rinse tank leak or connection loose (unknown)
First check

Observe bubble pattern in water tubes during rinse flow test

Test point
Water tube outlets
Expected
Bubbles appear at both outlets; number of bubbles approximately equal

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 measurement not displaying values patient safety

SpO2 sensor connected but no measurement values shown on display (HAMILTON-T1 SN > SN3000 only)

Ranked causes

  1. 1. SpO2 communication board not installed (common)
  2. 2. SpO2 hardware option not activated (common)
  3. 3. SpO2 sensor type not configured (common)
  4. 4. SpO2 sensor not connected or inactive (common)
First check

Verify SpO2 communication board is installed

Test point
System > Info > Info Tab
Expected
SpO2 communication board listed

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit fails to complete startup or preoperational checks patient safety

Ventilator does not pass required preoperational tests before clinical use

Ranked causes

  1. 1. Tightness test failed or not completed (unknown)
  2. 2. Flow sensor calibration not completed or failed (unknown)
  3. 3. O2 cell calibration not completed or failed (unknown)
  4. 4. Alarm system tests failed or not completed (unknown)
First check

Perform tightness test

Test point
Patient circuit and breathing system
Expected
No air leaks detected; system holds pressure

9 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery state of health assessment patient safety

Monitor battery state of health (SoH); replace when SoH <20%; consider replacement when <40% or cell voltage difference >100 mV; calibrate if error ≥5%.

Ranked causes

  1. 1. Battery SoH <20% (occasional)
  2. 2. Battery SoH 20-40% or cell voltage asymmetry (occasional)
  3. 3. Battery error ≥5% (rare)
First check

Check battery state of health (SoH)

Test point
Battery data screen (Page 1500)
Expected
SoH percentage displayed

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Expiratory valve (adult/pediatric)

Adjust expiratory valve gain using external pressure gauge to ensure correct pressure reading and gas flow control during ventilation.

Ranked causes

  1. 1. Expiratory valve requires calibration after service or sensor drift (common)
  2. 2. Adult/pediatric expiratory valve requires separate calibration (common)
  3. 3. Calibration blocked by corrupted calibration file (rare)
First check

Verify external pressure gauge reading during gain adjustment

Test point
Pressure connector 500300
Expected
50 mbar +/- 0.5 mbar (51 cmH2O +/- 0.5 cmH2O), +/- 0.5 mbar
Tools
External pressure gauge

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Neonatal expiratory valve test fails patient safety

Component test for neonatal expiratory valve returns NOT OK result during automated testing in service software

Ranked causes

  1. 1. Pressure sensors Pexpvalve and Pvent_control not equal when proportional valve fully closed (common)
  2. 2. Expiratory proportional valve not opening fully or Pexpvalve not reading zero when valve opened (occasional)
  3. 3. Pressure test out of tolerance during automated measurement (occasional)
First check

Compare pressure sensor readings

Test point
Pexpvalve and Pvent_control
Expected
Equal readings with patient system tightened and valve closed

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air entry filter obstruction or excessive pressure drop

Air entry (Pfilter) test shows NOT OK; filter pressure exceeds normal tolerance; airway entry test fails

Ranked causes

  1. 1. HEPA filter clogged or blocked with debris; high differential pressure (common)
  2. 2. NBC filter adapter (161787) installed with altered pressure tolerance (occasional)
First check

Run air entry test with default HEPA filter installed

Test point
HEPA filter pressure (Pfilter)
Expected
Pfilter within defined tolerance; test displays OK result

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery replacement due to age

Battery must be replaced based on installation or manufacturing date with software version < 2.2.0

Ranked causes

  1. 1. Two years after installation date (software < 2.2.0) (common)
  2. 2. Three years after manufacturing date (software < 2.2.0) (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery showing marginal state of health or cell voltage asymmetry

SoH between 20-40% or cell voltage difference exceeds 100 mV across cells

Ranked causes

  1. 1. Battery state of health (SoH) 20-40% (occasional)
  2. 2. Cell voltages permanently asymmetrical (>100 mV difference) (occasional)
First check

Check battery SoH percentage

Test point
Battery data page (Page No. 1500)
Expected
SoH < 40%

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery state of health low

Battery SoH <20% requires replacement; SoH <40% should be considered for replacement depending on software version

Ranked causes

  1. 1. Battery SoH < 20% (common)
  2. 2. Battery SoH < 40% (occasional)
First check

Monitor battery state of health

Test point
Battery status indicator
Expected
SoH ≥ 40%

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Buzzer does not sound during power loss caution

No audible alarm when all power sources (AC, DC, batteries) are removed

Ranked causes

  1. 1. Supercapacitors (gold caps) not fully charged (common)
  2. 2. Control board defective or buzzer circuit faulty (occasional)
First check

Charge supercapacitors before power loss test

Test point
Device power state
Expected
Device powered on in Stand-by or Service Mode for 20 minutes minimum

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Buzzer does not sound during power loss or sounds less than 2 minutes high voltage

Unit loses all AC, DC, and battery power; buzzer does not alarm or sounds for insufficient time

Ranked causes

  1. 1. Supercapacitors (gold caps) on control board not sufficiently charged (occasional)
  2. 2. Control board defective or buzzer circuit failure (rare)
First check

Verify supercapacitor charge time before test

Test point
Device power state
Expected
Powered on for at least 20 minutes in Standby or Service Mode

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Flow sensor (adult/pediatric and neonatal)

Calibrate inspiratory and expiratory flow sensors by rotating sensor mechanically during automatic calibration to measure airflow accurately at different pressures.

Ranked causes

  1. 1. Flow sensor requires calibration with adult breathing circuit (common)
  2. 2. System overwrites calibration values when switching sensor type (occasional)
First check

Confirm flow sensor rotation performed

Test point
Flow sensor position
Expected
Sensor mechanically rotated before calibration start

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: O2 cell (offset and gain) caution

Calibrate oxygen sensor offset using room air reference, then gain using high-pressure oxygen to ensure accurate O2 concentration measurement.

Ranked causes

  1. 1. O2 cell requires offset calibration using room air baseline (common)
  2. 2. O2 cell gain calibration follows offset calibration (common)
First check

Monitor offset calibration result

Test point
O2 cell calibration display
Expected
OK

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Expiratory valve exhaust tubing disconnected or blocked patient safety

Expiratory proportional valve exhaust not connected to bleed port causing O2 enrichment inside ventilator

Ranked causes

  1. 1. Expiratory valve exhaust tubing disconnected from bleed port (common)
  2. 2. Expiratory valve tubing blocked or expiratory valve membrane defective (occasional)
First check

Verify expiratory valve exhaust tubing is connected to bleed port

Test point
Bottom cover expiratory valve bleed port (outlet)
Expected
Tubing from expiratory proportional valve exhaust firmly connected

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Expiratory valve malfunction or obstruction patient safety

Exhalation obstructed warning or valve not operating; O2 enrichment occurring inside ventilator

Ranked causes

  1. 1. Expiratory valve bleed port tubing disconnected or blocked (common)
  2. 2. Expiratory valve membrane damaged or defective (occasional)
First check

Verify expiratory valve bleed port tubing connected

Test point
Expiratory valve bleed port at bottom cover outlet
Expected
Tubing from proportional valve exhaust connected to outlet

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fan not producing proper suction or airflow caution

Fan inlet does not produce sufficient suction to hold paper or airflow is insufficient

Ranked causes

  1. 1. Fan not properly assembled or obstructed (common)
  2. 2. Fan motor defective or not operating (occasional)
First check

Inspect fan assembly and inlet for obstructions

Test point
Fan housing and inlet opening
Expected
No visible obstructions or debris; fan properly seated in housing

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Leak test not passing patient safety

System does not maintain pressure, or tightness test shows leakage when obstruction valve is tested or during pressurization cycle.

Ranked causes

  1. 1. Obstruction valve not functioning or leaking (common)
  2. 2. Patient breathing circuit or filter not properly sealed (common)
First check

Verify breathing circuit and filter assembly

Test point
Patient connection port, expiratory connection port, flow sensor ports
Expected
All connections seated, no visual gaps or damage

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Nebulizer system pressure loss or leak caution

Nebulizer tubing connection or internal valve leaking; pressure does not hold or rises unexpectedly

Ranked causes

  1. 1. Nebulizer tubing not properly connected (common)
  2. 2. Nebulizer valve internal leak or O2 mixer seal failure (occasional)
First check

Connect hand pump to nebulizer port

Test point
Nebulizer port connection
Expected
Hand pump and tube properly connected

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Nebulizer tubing or valve leaking caution

Pressure loss detected in nebulizer circuit or backpressure indicates internal valve leakage

Ranked causes

  1. 1. Nebulizer tubing loose or improperly connected (common)
  2. 2. Nebulizer valve defective or O2 mixer internal leak (occasional)
First check

Test nebulizer tubing connection tightness

Test point
Nebulizer port and tubing connection (Di=4 mm/L=150 mm)
Expected
60 mbar (56-64 mbar) pressure sustained without loss
Tools
Hand pump (500330), pressure indicator

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Neonatal expiratory valve test verification

Expiratory valve pressure test fails, or pressure sensor readings do not match expected values when valve is closed or opened.

Ranked causes

  1. 1. Pressure sensor Pexpvalve or Pvent_control drift or malfunction (occasional)
  2. 2. Expiratory proportional valve solenoid not responding or valve stuck (occasional)
First check

Select neonatal mode in expiratory valve test window

Test point
Mode selector icon in window header
Expected
Neonatal mode active

7 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Nurse call relay not responding to alarms caution

Alarm condition does not trigger proper relay contact closure/opening or I:E timing relay not switching

Ranked causes

  1. 1. Nurse call connector not properly connected (common)
  2. 2. Relay contact failure or stuck relay (occasional)
First check

Connect nurse call connector

Test point
Communication board nurse call port
Expected
Connector or cable properly seated

5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pressure sensor assembly removal and installation caution

Procedure for removing and installing the pressure sensor assembly on HAMILTON-T1 ventilators across three serial number ranges (SN >= SN3000, SN2292-2999, and < SN2292), including tube disconnection and reconnection steps

Ranked causes

  1. 1. Maintenance or replacement of pressure sensor assembly required (unknown)
  2. 2. Reinstallation of pressure sensor assembly (unknown)
First check

Disconnect tube 2 and 3 from check valve assembly (SN >= SN3000)

Test point
Check valve assembly (5)
Expected
Tubes disconnected

7 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

RS-232 communication interface not functioning

Ventilator does not communicate with patient monitoring system or Datalogger software

Ranked causes

  1. 1. RS-232 protocol not configured correctly (common)
  2. 2. Communication cable not connected or faulty (common)
First check

Verify RS-232 protocol configuration

Test point
Utilities > Configuration > General > RS-232 Protocol
Expected
HAMILTON protocol selected

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

RS-232 communication interface not working

Communication with external patient monitoring system or Datalogger software fails or no signal detected

Ranked causes

  1. 1. RS-232 protocol not configured correctly (common)
  2. 2. Communication cable disconnected or defective (occasional)
First check

Configure RS-232 protocol

Test point
Configuration menu RS-232 Protocol setting
Expected
HAMILTON protocol selected

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System tightness test failed

System pressure not maintained or drops beyond acceptable level indicating internal or circuit leakage

Ranked causes

  1. 1. Leak in system or breathing circuit tubing (common)
  2. 2. Obstruction valve malfunction (SN3000 and above) (occasional)
First check

Run tightness test with test configuration 4 installed

Test point
System pressure maintenance
Expected
OK result displayed after test completion

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Technical state corrupt or failed

Self-test fails and displays 'Check hardware compatibility' or 'Technical state failed' message. Part causing failure highlighted in red in HW version tab.

Ranked causes

  1. 1. Incompatible hardware component combination installed (occasional)
  2. 2. Control board replaced or EEPROM communication interrupted (occasional)
First check

Examine HW version tab for highlighted parts in red

Test point
HW version display
Expected
No parts highlighted in red; all components show normal status

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air entry test with obstruction

Filter obstruction test to verify system response to blocked air entry

Ranked causes

  1. 1. Filter obstruction or air entry blockage (unknown)
First check

Place obstruction (plastic foil/paper) on air entry with small hole

Test point
Air entry obstruction
Expected
Obstruction properly positioned with 120 x 60 mm dimensions and small hole

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm lamp or sound not generating correctly

Alarm system not producing correct lamp illumination or sound output for alarm priorities

Ranked causes

  1. 1. Alarm system hardware or firmware malfunction (unknown)
First check

Activate high-priority alarm and verify lamp and sound

Test point
Alarm lamp, alarm sound
Expected
Correct lamp illumination and sound output for high priority

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm monitor test 2 watchdog does not activate caution

Alarm LED fails to blink or buzzer fails to sound when watchdog activated in alarm monitor test 2

Ranked causes

  1. 1. Watchdog circuit malfunction or alarm output failure (occasional)
First check

Verify watchdog activates alarm LED and buzzer

Test point
Watchdog circuit and alarm outputs
Expected
Alarm LED blinking and buzzer sounding

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm silence key does not respond at step 3 caution

Pressing alarm silence key fails to turn on alarm silence LED at step 3

Ranked causes

  1. 1. Alarm silence key or switch circuit failure (occasional)
First check

Verify alarm silence key activates LED

Test point
Alarm silence key and LED
Expected
LED ON after key press

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm silence LED does not blink during initialization caution

Alarm silence LED fails to blink at step 1 of alarm monitor test 1

Ranked causes

  1. 1. Alarm LED circuit malfunction (occasional)
First check

Verify alarm silence LED blinks at step 1

Test point
Alarm silence LED
Expected
LED blinking

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm silence LED does not turn off at step 4 caution

Alarm silence LED remains on instead of turning off at step 4 of alarm monitor test 1

Ranked causes

  1. 1. Alarm LED control circuit malfunction (occasional)
First check

Verify alarm silence LED is OFF at step 4

Test point
Alarm silence LED
Expected
LED OFF

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm silence LED does not turn on at step 2 caution

Alarm silence LED fails to illuminate solid at step 2 of alarm monitor test 1

Ranked causes

  1. 1. Alarm LED circuit malfunction (occasional)
First check

Verify alarm silence LED is ON at step 2

Test point
Alarm silence LED
Expected
LED ON

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm silence LED turns on at step 5 caution

Alarm silence LED turns on when it should remain off after key press at step 5

Ranked causes

  1. 1. Alarm silence key or control logic malfunction (occasional)
First check

Verify alarm silence LED stays OFF at step 5

Test point
Alarm silence LED
Expected
LED stays OFF

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm test failure patient safety

Correct alarm lamp does not illuminate, alarm sound is not generated, alarm priority is incorrect, or loudness adjustment does not affect alarm output.

Ranked causes

  1. 1. Alarm system not responding to test command (occasional)
First check

Activate High alarm

Test point
Alr.High button
Expected
High priority alarm lamp illuminates, alarm sound generates

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ambient state test does not trigger alarm patient safety

Alarm silence LED does not turn on and buzzer does not sound during ambient state test at step 8

Ranked causes

  1. 1. Ambient sensor or alarm circuit malfunction (occasional)
First check

Verify alarm silence LED and buzzer activate in ambient state

Test point
Ambient sensor and alarm circuit
Expected
Alarm silence LED ON and buzzer sounds

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery calibration error 5% or greater

Battery calibration error meets or exceeds 5%; calibration cycle required

Ranked causes

  1. 1. Battery calibration drift accumulated over time (occasional)
First check

Measure battery calibration error

Test point
Battery management system calibration status
Expected
Error < 5%, Within 5%

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery calibration error at or above 5%

Battery charger reports calibration error of 5% or greater, indicating battery requires calibration

Ranked causes

  1. 1. Battery charge measurement drift or calibration drift (occasional)
First check

Check battery calibration error on charger

Test point
Charger error display
Expected
Error < 5%

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery calibration maintenance due

Annual preventive maintenance calibration cycle recommended for active battery

Ranked causes

  1. 1. Normal maintenance interval reached during active battery use (common)
First check

Verify charger model and revision

Test point
Charger label
Expected
Model 369104 revision 07 or higher

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery calibration required

Battery error ≥ 5% indicates calibration is needed to maintain accurate battery status monitoring

Ranked causes

  1. 1. Battery error ≥ 5% (unknown)
First check

Check battery error percentage

Test point
Battery status display
Expected
Error ≥ 5%

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery disposed improperly patient safety

Battery disposed by incineration or burning; risk of explosion and environmental hazard

Ranked causes

  1. 1. Battery incinerated or burned at end of life (rare)
First check

Verify disposal method planned

Test point
Disposal procedure documentation
Expected
Disposed per local regulations, not incinerated

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery error exceeds 5%

Battery Error parameter on Battery data page shows value of 5% or greater

Ranked causes

  1. 1. Battery error ≥ 5% (occasional)
First check

Check battery Error percentage

Test point
Battery data page (Page No. 1500) - Error field
Expected
Error < 5%

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery State of Health below 20%

Battery SoH indicator shows value less than 20%, indicating battery requires replacement

Ranked causes

  1. 1. Battery age or degradation (common)
First check

Check battery State of Health indicator

Test point
Battery status display
Expected
SoH reading

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery State of Health between 20-40%

Battery SoH indicator shows 20-40%; battery replacement should be considered

Ranked causes

  1. 1. Battery approaching end of service life (occasional)
First check

Check battery State of Health indicator

Test point
Battery management system display
Expected
SoH value between 20-40%

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery State of Health between 20% and 40%

Battery SoH indicator shows value between 20% and 40%, indicating battery replacement should be considered

Ranked causes

  1. 1. Battery age or partial degradation (common)
First check

Check battery State of Health indicator

Test point
Battery status display
Expected
SoH reading between 20% and 40%

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery storage conditions maintenance caution

Preventive measures for battery preservation during storage to maintain battery life and readiness

Ranked causes

  1. 1. Improper storage conditions reduce battery shelf life (common)
First check

Verify storage temperature

Test point
Storage environment
Expected
5°C to 21°C; not exceeding 25°C over six-month periods

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Blower replacement based on timer

Replace blower module when blower timer reaches 100% or predicted to reach 100% within 6 months; update serial and revision numbers.

Ranked causes

  1. 1. Blower timer at or approaching 100% (common)
First check

Monitor blower timer percentage

Test point
Blower timer tab (Page 1201)
Expected
Blower timer value displayed in percentage

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Blower service required alarm displayed

Alarm message 'Blower service required' appears on display when blower operating time reaches 100%

Ranked causes

  1. 1. Blower operating time limit reached 100% (common)
First check

Check blower timer percentage on display

Test point
Blower timer tab (Page No. 1201)
Expected
≥100% indicates replacement required

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Blower timer reset required

After blower module replacement, blower timer must be reset to track service life of new blower module correctly.

Ranked causes

  1. 1. Blower module replaced with new serial number (common)
First check

Access Modify tab and select blower module component

Test point
Technical state Modify screen
Expected
Blower module component selectable and editable

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Alarm monitor 1 test caution

Test alarm monitoring system components in 10 steps including ambient state, fan failure alarm, audio pause button and buzzer; requires manual confirmation at each step

Ranked causes

  1. 1. Alarm monitoring system component verification required (common)
First check

Alarm monitoring system components operational

Test point
Electronics tab Alarm Mon.1 window
Expected
Test completed status displayed in Info window

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Alarm monitor 2 watchdog test caution

Activate software watchdog; verify alarm LED blinks and buzzer sounds; requires manual confirmation and ventilator reboot afterward

Ranked causes

  1. 1. Software watchdog and alarm circuit verification required (common)
First check

Alarm LED blinks during watchdog activation

Test point
Alarm LED indicator
Expected
LED blinks when watchdog activated
Tools
Visual inspection

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Alarm system test

Test speaker, yellow lamp, red lamp, audio pause LED and loudness levels; confirms alarm audible output and visual indicators

Ranked causes

  1. 1. Alarm system component verification required (common)
First check

Speaker audible test

Test point
Loudspeaker component
Expected
Audible alarm sound confirmed by technician
Tools
Ear inspection

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Autozero sequence test

Test pressure and flow sensors (Paw, Qaw, Pvent_monitor, Pvent_control) during autozero operation; requires bacteria/HME filter, breathing circuit, and blocked flow sensor end

Ranked causes

  1. 1. Sensor autozero functionality verification (common)
First check

Autozero test configuration TC4 setup

Test point
Sensor circuit pneumatic connections
Expected
All components connected per TC4 diagram; flow sensor end blocked
Tools
Visual inspection against diagram

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Battery fuel gauge recalibration caution

Battery fuel gauge accuracy degrades over time; red LED flashes when battery inserted into charger indicating need for recalibration. Recalibration cycle includes discharge, calibration charge, calibration discharge, and regular charge phases.

Ranked causes

  1. 1. Battery fuel gauge accuracy degradation due to charge cycles without rest periods (common)
First check

Observe red LED indicator on battery charger

Test point
Battery charger front panel LED
Expected
Red LED flashing

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Binary valves autozero test

Test auto zero valves (Pvent_monitor, Pflowsensor) during operation and autozero sequence; setup requires bacteria/HME filter, adult breathing circuit, flow sensor tubes and ET-tube

Ranked causes

  1. 1. Binary valve autozero functionality verification (common)
First check

Binary valve test setup configuration TC3

Test point
Pneumatic circuit test points
Expected
All components connected per TC3 diagram; test ready to start
Tools
Visual inspection against diagram

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Calibration: Blower flow test
  • Calibration: Blower pressure test
  • Calibration: Default settings reset
  • Calibration: Expiratory proportional valve (Adult/Ped.)
  • Calibration: Expiratory valve
  • Calibration: Expiratory valve (Neonatal)
  • Calibration: Expiratory valve offset and gain display
  • Calibration: Expiratory valve test
  • Calibration: Flow sensor
  • Calibration: Flow sensor (Adult/Ped.)
  • Calibration: Flow sensor values display (inspiratory/expiratory)
  • Calibration: Inspiratory tube resistance (Rinsp) values display
  • Calibration: Neonatal expiratory valve offset voltage
  • Calibration: Neonatal flow sensor
  • Calibration: Neonatal flow sensor offset and linearity
  • Calibration: O2 cell offset and gain values display
  • Calibration: O2 sensor
  • Calibration: Pressure sensor adjustment
  • Calibration: Pressure sensor values display
  • Calibration: Pressure sensors
  • Calibration: software update completion
  • Calibration: SpO2 sensor measurement
  • Calibration: Touchscreen
  • Calibration: User interface test
  • Check valve assembly tightness or obstruction valve failure (SN >= SN3000)
  • Check valve backflow detected
  • Check valve flap visually damaged or misaligned
  • Check valve one-way function failure (SN < SN3000)
  • Check valve tightness fails (SN >= 3000)
  • Display brightness does not switch between day and night modes
  • Distorted sound in alarm loudness range
  • Fan failure alarm does not trigger
  • Fan not providing adequate airflow
  • Filter pressure sensor reads incorrectly
  • Flow sensor values out of tolerance
  • Flow test verification
  • Hardkeys do not light when pressed
  • Nebulizer tube installation
  • Nebulizer valve not operating
  • O2 mixer sensor values out of tolerance
  • O2 mixer test verification
  • Obstruction valve fails to release pressure (SN >= 3000)
  • P&T knob does not respond or rotate incorrectly
  • Pressure test values out of tolerance
  • Print screen function
  • Proximal flow out of tolerance
  • Proximal pressure out of tolerance
  • Real time clock (RTC) requires adjustment
  • Real-time clock (RTC) date/time adjustment
  • Red lamp does not blink (Control Board Rev. >= 09)
  • Red lamp does not illuminate (Control Board Rev. < 09)
  • Reset to default calibration settings
  • Rinse flow tank pressure does not hold
  • Rinse flow unequal between outlets
  • Sensor status interpretation needed
  • Service date reset after maintenance
  • Service timer reset
  • Software update required after service procedures
  • Speaker does not produce audible sound
  • Touchscreen calibration required
  • Unit does not power on from AC mains
  • Unit does not switch to battery power
  • Unit does not switch to DC power
  • Unit does not transition from battery to DC power
  • Unit does not transition from DC to battery power
  • Unit does not transition to battery power
  • Ventilator unused for extended period
  • Yellow alarm lamp does not illuminate

68 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
100000Patient alarms - conditions affecting patient safety during ventilationPatient-related adverse conditions detected by the ventilatorcritical
100000Patient alarms related to pressure, flow, volume, apneaDetected condition in patient breathing parametersunknown
132001unknown
132002Check flow sensor tubingFlow sensor tubing issue detectedwarning
132003Check flow sensorFlow sensor malfunction detectedwarning
132004Check flow sensorFlow sensor missing or disconnectedwarning
132005O2 sensor defectiveO2 sensor malfunction detectederror
132006O2 sensor n/aO2 sensor not available or disconnectedwarning
132007O2 sensor not system compatibleIncompatible O2 sensor installederror
132008O2 sensor calibration neededO2 sensor requires calibrationwarning
132010CO2 sensor disconnectedCO2 sensor not connectedwarning
132011CO2 sensor defectCO2 sensor malfunction detectederror
132012CO2 sensor over temperatureCO2 sensor temperature exceeds limiterror
132013CO2 sensor warmupCO2 sensor warming upinfo
132015Flow sensor calibration neededFlow sensor requires calibrationwarning
132016Check CO2 airway adapterCO2 airway adapter issue detectedwarning
132017Check CO2 sampling lineCO2 sampling line issue detectedwarning
132018SpO2: patient disconnectedPatient SpO2 probe disconnectedcritical
132019SpO2: probe n/aSpO2 probe not available or disconnectedwarning
132020SpO2: low perfusion indexSpO2 perfusion index below thresholdwarning
132021SpO2: light interferenceAmbient light interfering with SpO2 sensorwarning
132022SpO2: sensor errorSpO2 sensor malfunction detectederror
132023SpO2: adapter n/aSpO2 adapter not available or disconnectedwarning
132024Circuit calibration neededVentilator circuit requires calibrationwarning
132025SpO2: poor signalSpO2 signal quality too lowwarning
132026Check flow sensor tubingFlow sensor tubing disconnectedwarning
132027Check flow sensor tubingPdiff sensor disconnectedwarning
132028Flow sensor calibration neededFlow sensor requires calibration (low priority)warning
133001unknown
133002Inspiratory volume limitationTidal volume exceeds set limitwarning
133003High pressureAirway pressure exceeds limiterror
133004Check for blockageHigh pressure detected at patient interfaceerror
133005Check for blockageMedium pressure detected at patient interfacewarning
141001Pressure limitationPressure limiting function activeinfo
141002Vt highTidal volume exceeds upper alarm thresholdwarning
141003Vt lowTidal volume falls below lower alarm thresholdwarning
141004High minute volumeMinute ventilation exceeds alarm thresholdwarning
141005Low minute volumeMinute ventilation falls below alarm thresholdwarning
141006High frequencyRespiratory frequency exceeds alarm thresholdwarning
141007Low frequencyRespiratory frequency falls below alarm thresholdwarning
141008High oxygenOxygen concentration exceeds alarm thresholdwarning
141009Low oxygenOxygen concentration falls below alarm thresholderror
141010ApneaPatient apnea detectedcritical
141011Apnea ventilation endedApnea ventilation mode terminatedwarning
141012Apnea ventilationApnea backup ventilation activatederror
141013High pressure during sighHigh pressure measured during sigh breatherror
141014Flip the flow sensorFlow sensor orientation incorrecterror
141016Disconnection on ventilator sideVentilator side circuit disconnectedcritical
141017Disconnection on patient sidePatient side circuit disconnectedcritical
141018Exhalation obstructedExhalation flow blocked or obstructederror
141019IRVInverse ratio ventilation mode activeinfo
141020ASV: Cannot meet targetASV unable to reach target ventilationwarning
141021Pressure limit has changedASV pressure limit adjusted automaticallyinfo
141023External flow sensor failedExternal flow sensor malfunction detectederror
141024Low pressureAirway pressure falls below lower thresholdwarning
141025PetCO2 highEnd-tidal CO2 exceeds alarm thresholdwarning
141026PetCO2 lowEnd-tidal CO2 falls below alarm thresholdwarning
141027Suctioning maneuverSuctioning procedure initiatedinfo
141028Performance limited by high altitudeVentilator at high altitude operation limitwarning
141029High SpO2Oxygen saturation exceeds alarm thresholdwarning
141030Low SpO2Oxygen saturation falls below alarm threshold (medium)warning
141031High pulsePulse rate exceeds alarm thresholdwarning
141032Low pulsePulse rate falls below alarm thresholdwarning
141033High PIPerfusion index exceeds alarm thresholdwarning
141034Low PIPerfusion index falls below alarm thresholdwarning
141035High SpHbHemoglobin saturation exceeds alarm thresholdwarning
141036Low SpHbHemoglobin saturation falls below alarm thresholdwarning
141037High SpCOCarboxyhemoglobin exceeds alarm thresholdwarning
141038Low SpCOCarboxyhemoglobin falls below alarm thresholdinfo
141039High SpMetMethemoglobin exceeds alarm thresholdwarning
141040Low SpMetMethemoglobin falls below alarm thresholdinfo
141041Low SpO2Oxygen saturation falls below alarm threshold (high)critical
141043Ventilation adjustment OFFAutomatic ventilation adjustment disabledinfo
141044Oxygenation adjustment OFFAutomatic oxygenation adjustment disabledinfo
141045Ventilation Controller at limitVentilation control algorithm at upper limitwarning
141046Oxygenation Controller at limitOxygenation control algorithm at upper limitwarning
141047High flowGas flow exceeds alarm threshold in nCPAP modewarning
141048ObstructionCircuit obstruction detectederror
141049High flowGas flow exceeds alarm thresholdwarning
141050Sensor simulation activeVentilator in sensor simulation modeinfo
141051High PEEPPEEP exceeds alarm thresholdwarning
141052Loss of PEEPPEEP falls below set valueerror
141053FiO2 set to 100% due to low SpO2Oxygen concentration automatically set to maximumcritical
141054Recruitment in progressLung recruitment maneuver activeinfo
141055Oxygen control limit exceededOxygen delivery exceeds control algorithm limitwarning
141056SBT abortedSpontaneous breathing trial terminatedwarning
141057SBT successfully fulfilledSpontaneous breathing trial completed successfullyinfo
141058High PVIPleth variability index exceeds alarm thresholdwarning
141059Low PVIPleth variability index falls below alarm thresholdinfo
141060Check flow sensor for waterWater detected in flow sensor circuitwarning
141061Check flow sensor for waterWater detected in flow sensor circuit (high priority)error
141062SpeakValve ONSpeaking valve compatibility mode activeinfo
141063SpeakValve OFFSpeaking valve compatibility mode inactiveinfo
141064Vt lowTidal volume falls below lower alarm threshold (high priority)error
141065Check for blockagePatient interface blockage suspectederror
141066Maximum leak compensationLeak compensation at maximum limiterror
141067Oxygen adjustment OFFClosed-loop FiO2 adjustment disabledinfo
141068Oxygen controller at limitClosed-loop FiO2 controller at upper limitwarning
141069Oxygen msg limit exceededClosed-loop FiO2 oxygen message limit exceededwarning
141070CPR ONCardiopulmonary resuscitation mode activecritical
141071PetCO2 target range changedPetCO2 target range modifiedinfo
141072Check PlimitPressure limit requires attentionwarning
200000Patient alarms related to temperature, battery, valves, sensorsDetected condition in temperature, battery status, valve operation, or sensor readingsunknown
200000Technical events - non-critical system events or warningsTechnical system events detected during operationwarning
231001Technical event: Pressure controller pressure lowPressure controller detects low pressure conditionerror
231002Technical event: Pressure controller pressure highPressure controller detects high pressure conditionerror
231003Technical event: Flow controller flow lowFlow controller detects low flow conditionerror
231004Technical event: Flow controller flow highFlow controller detects high flow conditionerror
231005Technical event: Inspiration valve leakInspiration valve detects leak conditionerror
231006Oxygen supply failedOxygen flow below required levelerror
231007Technical event: O2 controller flow highO2 controller detects high flow conditionerror
231008Technical event: O2 valve leakO2 valve detects leak conditionerror
231009Technical event: Blower controller speed lowBlower controller detects low speed conditionerror
231010Technical event: Blower controller speed highBlower controller detects high speed conditionerror
231011Vent outlet temperature highTemperature sensor detects high outlet temperaturewarning
231012Technical event: Qvent flow sensor defectQvent flow sensor malfunction detectederror
231013Technical event: QO2 flow sensor defectQO2 flow sensor malfunction detectederror
231014Technical event: Ambient valve errorAmbient valve malfunction detectederror
231015Autozero Pvent control OKAutozero procedure completed successfullyinfo
231016Autozero Pvent control out of rangeAutozero calibration failedwarning
231017Blower service requiredBlower maintenance interval reached or malfunction detectedwarning
231018Technical event: Self-test with errorPower-on self-test or periodic self-test failederror
231019Technical event: Inspiration valve sensor defectInspiration valve sensor malfunction detectederror
231020Technical event: Regulator data log onData logging feature activatedinfo
231021O2 presence check failedO2 detection circuit detects absence of oxygenwarning
231022Technical event: Expiration valve pressure sensor defectExpiration valve pressure sensor malfunction detectederror
231023Technical event: Invalid flow sensorFlow sensor configuration or detection failederror
231024Expiration valve cover adultAdult expiration valve cover detectedinfo
231025Expiration valve cover neoNeonatal expiration valve cover detectedinfo
231026Expiration valve cover invalidUnknown or invalid expiration valve cover detectederror
231027Wrong expiratory valveExpiration valve mismatch detectederror
231028Pressure controller flow highPressure controller detects excessive flowwarning
231029Autozero Pexpiration valve OKExpiration valve autozero completed successfullyinfo
231030Autozero Pexpiration valve errorExpiration valve autozero failedwarning
231031Expiration valve cover universalUniversal expiration valve cover detectedinfo
231032Technical event: Expiration valve disconnectedExpiration valve connection lost or missingerror
231033Technical event: Blower controller speed limitedBlower speed limited by controllerwarning
231034Tube resonance measurement OKTube resonance measurement completed successfullyinfo
231035Tube resonance measurement errorTube resonance measurement failedwarning
231036Technical event: Pvent pressure sensor defectPvent pressure sensor malfunction detectederror
231040Calibrate Pvent control OKPvent pressure control calibration completed successfullyinfo
231041Calibrate Pvent control errorPvent pressure control calibration failederror
231044Calibrate expiration valve OKExpiration valve calibration completed successfullyinfo
231045Calibrate expiration valve errorExpiration valve calibration failederror
231046Calibrate inspiration/expiration valve OKInspiration/expiration valve calibration completed successfullyinfo
231047Calibrate inspiration/expiration valve errorInspiration/expiration valve calibration failederror
231099Communication checkInter-processor or system communication checkinfo
232001Pressure sensor tolerancePressure sensor reading out of specificationwarning
232002Technical event: Pvent monitor sensor defectPvent monitor sensor malfunction detectederror
232003Technical event: Paw sensor defectAirway pressure sensor malfunction detectederror
232004Pressure not releasedSystem pressure not relieved as expectederror
232005Technical event: Blower hotBlower temperature exceeds limitwarning
232006Technical event: Blower temperature sensor defectBlower temperature sensor malfunction detectederror
232007Check flow sensor tubingFlow sensor tubing malfunction or blockage detectedwarning
232008Technical event: Pambient sensor defectAmbient pressure sensor malfunction detectederror
232009Pvent autozero valve errorPvent autozero valve malfunction detectedwarning
232010Pvent autozero valve missingPvent autozero valve not detectederror
232011Proximal autozero valve errorProximal autozero valve malfunction detectedwarning
232012Proximal autozero valve missingProximal autozero valve not detectederror
232013Distal autozero valve errorDistal autozero valve malfunction detectedwarning
232014Distal autozero valve missingDistal autozero valve not detectederror
232015Proximal/distal autozero valve missingProximal or distal autozero valve not detectederror
232016Proximal/distal autozero valve errorProximal or distal autozero valve malfunction detectedwarning
232017Autozero Pvent monitor OKPvent monitor autozero completed successfullyinfo
232018Autozero Pvent monitor out of rangePvent monitor autozero failedwarning
232019Autozero Pvent control OKPvent control autozero completed successfullyinfo
232020Autozero Pvent control out of rangePvent control autozero failedwarning
232022Autozero Paw out of rangeAirway pressure autozero failedwarning
232024Autozero Qaw out of rangeAirway flow autozero failedwarning
232025Autozero Paw/Qaw OKAirway pressure and flow autozero completed successfullyinfo
232026Autozero Paw/Qaw out of rangeAirway pressure and flow autozero failedwarning
232027Device temperature highInternal device temperature exceeds safe operating limitwarning
232028Technical event: Filter pressure highFilter inlet pressure exceeds limitwarning
232029Technical event: Tinst sensor defectTemperature sensor malfunction detectederror
232030Technical event: Clock errorReal-time clock malfunction detectedwarning
232031FiO2 calibration errorFiO2 sensor calibration failedwarning
232032FiO2 calibration OKFiO2 sensor calibration completed successfullyinfo
232034Replace O2 sensorO2 sensor defect or end-of-life detectederror
232035Technical event: Pfilter sensor defectFilter pressure sensor malfunction detectederror
232036CO2 calibration errorCO2 sensor calibration failedwarning
232037CO2 calibration OKCO2 sensor calibration completed successfullyinfo
232038CO2 calibration neededCO2 sensor requires periodic recalibrationwarning
232040Quick disconnection detectedGas supply line disconnection detectedinfo
232041CO2 calibration start errorCO2 calibration procedure failed to startwarning
232042CO2 calibration start OKCO2 calibration procedure started successfullyinfo
232043Quick reconnection detectedGas supply line reconnectedinfo
232044SpO2 sensor errorSpO2 sensor malfunction detectedwarning
232045Technical event: SpO2 sensor settings errorSpO2 sensor configuration invalid or missingerror
232046Autozero expiration valve OKExpiration valve autozero completed successfullyinfo
232047Autozero expiration valve errorExpiration valve autozero failedwarning
232048Autozero proximal valve leakProximal valve leak detected during autozerowarning
232049Autozero distal valve leakDistal valve leak detected during autozerowarning
232050Pvent monitor calibration errorPvent monitor calibration failedwarning
232051Pvent monitor calibration OKPvent monitor calibration completed successfullyinfo
232052Paw calibration errorAirway pressure calibration failedwarning
232053Paw calibration OKAirway pressure calibration completed successfullyinfo
232054Qaw calibration errorAirway flow calibration failedwarning
232055Qaw calibration OKAirway flow calibration completed successfullyinfo
232056Technical event: Pambient/Pfilter mismatchAmbient and filter pressure sensor values inconsistentwarning
232057Technical event: SpO2 sensor change failedSpO2 sensor replacement or configuration change failederror
232058Paux sensor defectAuxiliary pressure sensor malfunction detectederror
232059Paux calibration errorAuxiliary pressure sensor calibration failedwarning
232060Paux calibration OKAuxiliary pressure sensor calibration completed successfullyinfo
232061FiO2 calibration start errorFiO2 calibration procedure failed to startwarning
232062FiO2 calibration start OKFiO2 calibration procedure started successfullyinfo
232063Qaw and tube calibration at flow timed outFlow and tube calibration procedure exceeded time limitwarning
232064Qaw and tube calibration at flow passedFlow and tube calibration procedure completed successfullyinfo
232065Qaw and tube calibration at flow failedFlow and tube calibration procedure failederror
232099Communication checkInter-processor or system communication checkinfo
233001Technical event: Autozero Pvent monitor failPvent monitor autozero procedure failederror
233002Technical event: Autozero Pvent control failPvent control autozero procedure failederror
233008Aerogen nebulizer disconnectedAerogen nebulizer is physically disconnected from the ventilatorwarning
234001Loudspeaker defectiveSpeaker hardware failure detected during startup or operationerror
241001Replace HEPA filterHEPA filter maintenance interval exceeded or filter cloggedwarning
241002Release valve defectiveTightness test detects release valve malfunctionerror
243004Buzzer defectiveBuzzer test fails at startup or buzzer produces no sounderror
243007Realtime clock failureRTC chip failure or battery depleted during power losserror
243010Panel connection lostCommunication link between main board and control panel failserror
244001Loss of external powerMains power supply disconnected or failedwarning
244002Battery 1: Calibration requiredBattery 1 capacity measurement indicates need for recalibrationwarning
244003Battery 2: Calibration requiredBattery 2 capacity measurement indicates need for recalibrationwarning
244004Battery 1: Temperature highBattery 1 temperature exceeds safe operating range during charge/dischargewarning
244005Battery 2: Temperature highBattery 2 temperature exceeds safe operating range during charge/dischargewarning
244006Battery lowBattery charge level drops below high-priority threshold during operationwarning
244007Battery power lossBattery becomes disconnected or fails suddenly during operationcritical
244008Battery 1: Wrong batteryBattery 1 type or model does not match system requirementserror
244009Battery 2: Wrong batteryBattery 2 type or model does not match system requirementserror
244010Battery lowBattery charge level drops below low-priority threshold during operationwarning
244011Battery communication errorBattery management system loses communication with battery moduleerror
244012Battery 1: DefectiveBattery 1 self-test fails or capacity measurement indicates failureerror
244013Battery 2: DefectiveBattery 2 self-test fails or capacity measurement indicates failureerror
244015Battery lowBattery charge level drops below medium-priority threshold during operationwarning
244016Battery 1: Replacement requiredBattery 1 has reached end of service life or unreliable performance detectedwarning
244017Battery 2: Replacement requiredBattery 2 has reached end of service life or unreliable performance detectedwarning
244022Continue charging batteryBattery charging in progress, normal state during AC power connectedinfo
246001Preventive maintenance requiredEquipment operating hours or calendar interval exceeds maintenance schedulewarning
246003Fan failureCooling fan stops or runs abnormally, temperature control compromisederror
246016Touch not functionalTouch screen detects no response to finger contact or self-test failserror
246020Loudspeaker defectiveSpeaker hardware fails during HMI self-test or audio output ceaseserror
249001Options not foundSystem cannot read or locate option configuration file at startuperror
249012Technical eventLast setting config file errorunknown
249013Technical eventResearch config file errorunknown
249020Invalid communication boardOption board not recognizederror
249021Check hardware compatibilityHardware compatibility issue detectedwarning
250001Check humidifier communicationHumidifier communication failureerror
250002Humidifier water lowWater level below thresholdwarning
250003Humidifier chamber temp lowChamber temperature below setpointwarning
250004Humidifier Y-piece temp lowY-piece temperature below setpointwarning
250005Humidifier check right tubeRight heating tube fault detectedwarning
250006Humidifier check left tubeLeft heating tube fault detectedwarning
250007Humidifier check chamberChamber component fault detectederror
250008Check humidifierGeneral humidifier faulterror
250009Humidifier errorUnspecified humidifier errorerror
250010Humidifier water highWater level above thresholdunknown
250011Humidifier chamber temp highChamber temperature above setpointwarning
250012Humidifier Y-piece temp highY-piece temperature above setpointwarning
250013Humidifier tiltHumidifier tilted or misalignedunknown
250014Check humidifierHumidifier fault detectederror
250015Check breathing circuit limbsBreathing circuit limb fault detectedwarning
251001Check IntelliCuff communicationIntelliCuff communication losserror
251002Check IntelliCuffIntelliCuff low faultwarning
251003Cannot turn off IntelliCuffIntelliCuff power-off failureerror
251004Check IntelliCuffIntelliCuff battery lowwarning
251005Cuff pressure highEndotracheal cuff pressure exceeds limitwarning
251006Cuff deflatedCuff deflation detectedwarning
251007Cuff leakCuff leak detectedwarning
251008Check IntelliCuffIntelliCuff medium faultwarning
251009Cuff leakIntelliCuff disconnection detectedwarning
251010Check IntelliCuffIntelliCuff errorerror
251011Check IntelliCuffIntelliCuff battery criticalcritical
251012Check IntelliCuffIntelliCuff high faulterror
251013Check IntelliCuff stateIntelliCuff state errorerror
252001Check TeslaSpy communicationTeslaSpy communication losserror
252002Reduce MRI scanner proximityHigh magnetic field proximity detectedwarning
252003Service requiredMagnetic field too high detectederror
252004TeslaSpy defectiveTeslaSpy unit fault detectederror
252005TeslaSpy service requiredTeslaSpy service condition detectederror
255001Technical eventPercentage minimum volume high test failedunknown
255002Technical eventPercentage minimum volume high limit incorrectunknown
255003Technical eventPercentage minimum volume low test failedunknown
255004Technical eventPercentage minimum volume low limit incorrectunknown
255005Technical eventPEEP high limit incorrectunknown
255006Technical eventPEEP high test failedunknown
255007Technical eventPEEP low limit incorrectunknown
255008Technical eventPEEP low test failedunknown
255009Technical eventOxygen low test failedunknown
255010Technical eventOxygen high test failedunknown
255011Technical eventPercentage minimum volume positive change failedunknown
255012Technical eventPercentage minimum volume negative change failedunknown
255013Technical eventPEEP positive change test failedunknown
255014Technical eventPEEP negative change test failedunknown
255015Technical eventOxygen positive change test failedunknown
255016Technical eventOxygen negative change test failedunknown
255017Technical eventRecruitment maneuver runtime exceededunknown
255018Technical eventSimulation supervision test failedunknown
255019Technical eventPercentage minimum volume high fine test failedunknown
255020Technical eventPercentage minimum volume high fine limit wrongunknown
255021Technical eventPercentage minimum volume low fine test failedunknown
255022Technical eventPercentage minimum volume low fine limit wrongunknown
255023Oscillation OxygenOxygen parameter oscillation detected in closed-loopwarning
255024Oscillation PEEP/CPAPPEEP parameter oscillation detected in closed-loopwarning
255025Oscillation %MinVolPercentage minimum volume oscillation in closed-loopwarning
255026Technical eventData mismatch detectedunknown
255027Technical eventPetCO2 value frozen in closed-loopunknown
255028Technical eventSpO2 value frozen in closed-loopunknown
255029Technical eventSpontaneous breathing frequency frozenunknown
255030Technical eventClosed-loop FiO2 oxygen low test failedunknown
255031Technical eventClosed-loop FiO2 oxygen high test failedunknown
255032Technical eventClosed-loop FiO2 oxygen positive change failedunknown
255033Technical eventClosed-loop FiO2 oxygen negative change failedunknown
255034Technical eventClosed-loop FiO2 simulation supervision failedunknown
255035Technical eventClosed-loop FiO2 SpO2 value frozenunknown
281001Firmware mismatchPLD firmware version incompatibleerror
281002JTAG not workingJTAG programming interface not respondingerror
281003Check hardware compatibilityHardware compatibility issue detectedwarning
283001Technical eventVentilator GUI startup failedunknown
283003Language not loadedLanguage file failed to loadwarning
283005Settings file errorSetup configuration file errorerror
283008Check settingsSettings configuration error detectedwarning
283009Self test failedPower-on self test failureerror
283010No ventilation after power-failVentilation not resumed after power losscritical
283011Check IntelliCuff stateIntelliCuff mode errorerror
283012Panel settings file errorPanel setup configuration file errorerror
284001Startup failedService GUI startup failureerror
284002Technical eventService alarm high priorityunknown
284003Technical eventService alarm medium priorityunknown
284004Technical eventService alarm low priorityunknown
284005Flow sensor calibration neededFlow sensor requires recalibrationwarning
284006Restart device to enable Tests/CalibDevice restart required for tests or calibrationinfo
285001Technical eventAlarm lamp error indicator defectunknown
285002Technical eventAlarm lamp warning indicator defectunknown
285003Technical eventBacklight defect detectedunknown
285004Function key not operationalFunction key malfunction detectedwarning
285005Touch not functionalTouchscreen malfunction detectederror
291001Remote communication error (COM1)COM1 port communication errorerror
291002Remote communication error (COM2)COM2 port communication errorerror
291003Remote communication timeout (COM1)COM1 communication timeoutwarning
300000Technical fault - ventilator enters safety modeFatal technical failure detected during operationcritical
332001Safety ventilation activated due to flow sensor errorBreath management module detects flow sensor malfunctioncritical
341008Safety ventilation activated due to ventilation flow sensor defectVentilation monitoring controller detects flow sensor failurecritical
341009Safety ventilation activated due to pressure sensor defectVentilation monitoring controller detects pressure sensor failurecritical
346054Safety ventilation activated due to detected safety failureLow-level system detects critical safety conditioncritical
383008Technical fault due to basic startup failureVentilator GUI fails basic startup sequenceerror
383009Technical fault due to screen failureVentilator GUI screen or display initialization failserror
383012Technical fault due to self-test failureVentilator GUI self-test sequence fails during startuperror
384001Technical fault due to basic startup failure in settings GUISettings GUI fails basic startup sequenceerror
400000Technical fault - ventilator enters ambient stateSerious technical fault risking compromise of safe ventilationcritical
431001Blower faultBlower hardware failure or malfunctioncritical
431002Technical fault: Blower disconnectedBlower connection lost or unpluggederror
431004Technical fault: Inspiration valve over-currentInspiration valve drawing excessive currenterror
431006Technical fault: Expiration valve over-currentExpiration valve drawing excessive currenterror
431007Technical fault: Ventilation output over-temperatureOutlet temperature exceeds safe limiterror
431008Technical fault: Ventilation flow sensor errorFlow sensor malfunction or measurement failureerror
431009Technical fault: Oxygen flow sensor errorO2 flow sensor malfunction or measurement failureerror
431010Technical fault: Control REGSPI timeoutControl register SPI communication timeouterror
431011Technical fault: Monitor REGSPI timeoutMonitor register SPI communication timeouterror
431012Technical fault: Flow sensor measurement impreciseFlow sensor output out of expected range or unstablewarning
431013Technical fault: Calibration read failedUnable to read calibration data from memoryerror
431014Technical fault: Inspiration valve sensor defectInspiration valve position sensor failureerror
431015Technical fault: Emergency off failedEmergency stop function did not execute properlycritical
431017Technical fault: Inspiration valve disconnectedInspiration valve connector unplugged or lost connectionerror
432001Technical fault: Blower over-temperatureBlower motor temperature exceeds safe operating limiterror
432002Technical fault: Instrument over-temperatureInternal instrument temperature exceeds safe limiterror
432003Technical fault: Pressure not released to ambientSystem unable to release pressure to atmosphere during ambient modeerror
433001Technical fault: Breath monitoring tick timeoutBreath monitoring process did not respond within timeout perioderror
433002Technical fault: Control tick timing errorControl loop timing violation or delay exceeds toleranceerror
443001Technical fault: Watchdog failed LLSLow-level system watchdog timer failed to resetcritical
444001Battery totally dischargedBattery charge level reached zerocritical
444004Technical fault: Voltage out of tolerancePower supply voltage outside acceptable operating rangeerror
444005Technical fault: Shutdown failedSystem unable to complete normal shutdown sequenceerror
446001Technical fault: CPU temperature criticalProcessor temperature exceeds maximum safe operating temperaturecritical
446002Technical fault: Safety failedSafety system self-test or monitoring failedcritical
446003Technical fault: Watchdog failed GDGas delivery module watchdog timeout occurredcritical
446004Technical fault: Watchdog failed GD valve regulatorGas delivery valve regulator watchdog timeoutcritical
446005Technical fault: Watchdog failed GD blower regulatorGas delivery blower regulator watchdog timeoutcritical
446006Technical fault: Watchdog failed GD monitoringGas delivery monitoring watchdog timeoutcritical
446007Technical fault: Watchdog failed GD valve logGas delivery valve logging watchdog timeouterror
446008Technical fault: Watchdog failed GD blower logGas delivery blower logging watchdog timeouterror
446009Technical fault: Watchdog failed GD MF measureGas delivery measurement function watchdog timeouterror
446010Technical fault: Watchdog failed GD GPIO fault inputGas delivery GPIO fault monitoring watchdog timeouterror
446011Technical fault: Watchdog failed BMBreath monitoring module watchdog timeoutcritical
446012Technical fault: Watchdog failed BM gas delivery communicationBreath monitoring to gas delivery communication watchdog timeoutcritical
446013Technical fault: Watchdog failed BM hardware controlBreath monitoring hardware control watchdog timeoutcritical
446014Technical fault: Watchdog failed BM BPG client receiverBreath monitoring BPG client receiver watchdog timeoutcritical
446015Technical fault: Watchdog failed BM BPG server receiverBreath monitoring BPG server receiver watchdog timeoutcritical
446016Technical fault: Watchdog failed BM breath data calculationBreath data calculation watchdog timeoutcritical
446017Technical fault: Watchdog failed BM MF measureBreath monitoring measurement function watchdog timeouterror
446018Technical fault: Watchdog failed BPGBreath pattern generation module watchdog timeoutcritical
446019Technical fault: Watchdog failed BPG GD client receiverBPG gas delivery client receiver watchdog timeoutcritical
446020Technical fault: Watchdog failed BPG MF measureBPG measurement function watchdog timeouterror
446021Technical fault: Exception happenedUnhandled software exception occurred in systemcritical
446022Technical fault: Voltage errorInternal voltage supply anomaly detectedcritical
446023Technical fault: Watchdog failed LLS workloadLow-level system workload monitoring watchdog timeoutcritical
446024Technical fault: Am 3V3 error3.3V analog supply failure detectedcritical
446025Technical fault: Am ADC errorAnalog-to-digital converter failure detectederror
446026Technical fault: Am Vref errorAnalog reference voltage failure detectederror
446028Technical fault: Clock errorSystem clock or timing reference failurecritical
446029Technical fault: Ambient failure detectedAmbient monitoring system detected a critical failureerror
446030Technical fault: Am watchdog occurredAnalog module watchdog timeout occurrederror
446031Technical fault: Restart ventilation failedSystem unable to restart ventilation after ambient statecritical
446032Technical fault: Patient vent overpressurePatient ventilation outlet pressure exceeds maximum safe limitcritical
481001Technical fault: First run migration errorStartup configuration migration failederror
481002Technical fault: CRC errorData integrity check (CRC) failed on stored dataerror
481003Unknown part numberSystem encountered unrecognized part or configuration IDerror
481004Technical state failedSystem technical state verification failed at startuperror
481005Technical fault: Synchronization errorSystem module synchronization failed during startuperror
481006Technical fault: Reconnect startup type errorStartup type or mode recovery failederror
481007Technical fault: Reconnect accept errorSystem reconnection handshake failed during startuperror
483001Technical fault: Ventilator GUI startup basic startup failureVentilator GUI failed basic startup checkserror
483002Technical fault: Ventilator GUI startup screen failureVentilator GUI display screen failed to initializeerror
483003Technical fault: Ventilator GUI startup load GUI basics failureVentilator GUI basic components failed to loaderror
483004Technical fault: Ventilator GUI startup self-test prolog failureVentilator GUI self-test prolog phase failederror
483005Technical fault: Ventilator GUI startup self-test failureVentilator GUI self-test execution failederror
483006Technical fault: Ventilator GUI startup self-test epilog failureVentilator GUI self-test epilog phase failederror
483007Technical fault: Ventilator GUI startup loading bitmaps failureVentilator GUI failed to load display bitmaps or graphicserror
483008Technical fault: Ventilator GUI startup initialization failureVentilator GUI initialization sequence failederror
484001Technical fault: Setting GUI startup basic startup failureSetting GUI failed basic startup checkserror
484002Technical fault: Setting GUI startup screen failureSetting GUI display screen failed to initializeerror
484003Technical fault: Setting GUI startup load GUI basics failureSetting GUI basic components failed to loaderror
484004Technical fault: Setting GUI startup self-test prolog failureSetting GUI self-test prolog phase failederror
484005Technical fault: Setting GUI startup self-test failureSetting GUI self-test execution failederror
484007Technical fault: Setting GUI startup loading bitmaps failureSetting GUI failed to load display bitmaps or graphicserror
484008Technical fault: Setting GUI startup initialization failureSetting GUI initialization sequence failederror
485001Technical fault: Alarming GUI ambient failure detectedAlarming GUI detected system in ambient/unpowered state failureerror
500000System failures invisible to user, internal use only, no user action requiredInternal system error detected and managed by ventilator firmwareinfo
500000Invisible technical fault - internal exception handlingException handling during ventilator operationinfo
700000Technical fault - ventilator enters safety modeFatal technical failure requiring patient safety interventioncritical
743001Technical fault: Watchdog failed LLSPanel module low-level system watchdog timeoutcritical
743003Technical fault: MMI connection lostMan-machine interface connection lost or disconnectederror
746001Technical fault: Watchdog failed ALRPanel low-level system watchdog detected ALR module timeoutcritical
746002Technical fault: Watchdog failed VGUIPanel low-level system watchdog detected ventilator GUI timeoutcritical
746003Technical fault: Watchdog failed VGUI mode controlVentilator GUI mode control watchdog timeout on panelcritical
746004Technical fault: Panel error failure detectedPanel hardware or communication error detectederror
746005Technical fault: Exception happenedUnhandled exception in panel module occurredcritical
746006Technical fault: Watchdog failed LLS workloadPanel low-level system workload monitoring watchdog timeoutcritical
746007Technical fault: Restart ventilation failedPanel unable to restart ventilation after ambient statecritical
783001Technical fault: Monitoring channel observation failedVentilator GUI monitoring channel failed to update or responderror
783002Technical fault: Trending channel observation failedVentilator GUI trending data channel failed to updateerror
783003Technical fault: VMC timeoutVentilator mode control communication timeout occurrederror
785001Technical fault: Alarming channel observation failedAlarm channel failed to update or respond in alarming GUIerror
785002Technical fault: Panel error failure detectedPanel error detected in alarming GUI moduleerror
785003Technical fault: PM channel observation failedPower management channel failed to update or responderror
Check hardware compatibilityIncompatible combination of hardware components detected during self-testSystem detects installed hardware components that are not compatible with each othercritical
C-SS-NNNAlarm codes are structured with alarm class, subsystem, and ID number; C=class, SS=subsystem, NNN=IDVarious conditions trigger different alarm classes and severities based on patient safety impactunknown
Exhalation obstructedExpiratory valve is blocked or obstruction detected in expiration pathwayExpiratory valve bleed port tubing disconnected, blocked, or valve membrane damaged; blockage applied during testwarning
Technical state failedTechnical state cannot be read from EEPROM of a component or after control board replacementUnit fails self-test due to corrupt or unreadable technical statecritical

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI