Hamilton Medical C6: troubleshooting and error codes

The ServiceForge knowledge base holds 74 documented symptoms for the Hamilton Medical C6, with 187 ranked causes, 62 diagnostic checks with test points and expected values, and 442 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Expiratory valve test fails

Expiratory valve pneumatic test does not pass with NOT OK status message

Ranked causes

  1. 1. Ambient valve malfunction (common)
  2. 2. Pvent_control requires calibration (common)
  3. 3. Expiratory valve requires calibration (common)
  4. 4. Expiratory valve membrane degraded (common)
  5. 5. Expiratory valve cover malfunction (occasional)
  6. 6. Internal pneumatic leak (common)
  7. 7. Expiratory valve malfunction (occasional)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flow sensor calibration fails

Flow sensor calibration cannot be completed

Ranked causes

  1. 1. Adult/pediatric flow sensor mismatch (common)
  2. 2. System leaks (common)
  3. 3. Faulty proximal flow sensor (occasional)
  4. 4. Ambient valve malfunction (occasional)
  5. 5. Expiratory membrane damaged (occasional)
  6. 6. Pressure sensor component failure (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air entry pressure exceeds limit with HEPA filter

Air entry pressure is greater than 2 mbar with HEPA filter installed

Ranked causes

  1. 1. HEPA filter clogged (common)
  2. 2. Pfilter sensor connection fault (common)
  3. 3. Pambient sensor reading incorrect (occasional)
  4. 4. Mainboard failure (occasional)
  5. 5. Blower malfunction (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Blower flow test fails

Blower airflow test does not pass with NOT OK status message

Ranked causes

  1. 1. Internal pneumatic leak (common)
  2. 2. HEPA filter clogged (common)
  3. 3. Qvent sensor malfunction (occasional)
  4. 4. Blower module malfunction (occasional)
  5. 5. Ventilator mainboard failure (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Blower pressure test fails

Blower pressure pneumatic test does not pass with NOT OK status message

Ranked causes

  1. 1. Internal pneumatic leak (common)
  2. 2. Ambient valve malfunction (common)
  3. 3. Expiratory valve membrane degraded (common)
  4. 4. Blower module malfunction (occasional)
  5. 5. Ventilator mainboard failure (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inspiratory valve test fails

Inspiratory valve pneumatic test does not pass with NOT OK status message

Ranked causes

  1. 1. Internal pneumatic leak (common)
  2. 2. Qvent sensor malfunction (common)
  3. 3. Ambient valve malfunction (common)
  4. 4. Inspiratory valve requires calibration (common)
  5. 5. Inspiratory valve cable disconnected (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pressure calibration fails or reaches incorrect value

Cannot reach 50 mbar (± 5 mbar) or Gain values exceed 1000 ± 50

Ranked causes

  1. 1. Blower flow issue (common)
  2. 2. Expiratory valve membrane damage (common)
  3. 3. System leaks (occasional)
  4. 4. Measurement device not zeroed or incorrect unit (occasional)
  5. 5. Pressure sensor malfunction (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pressure calibration target not reached

50 mbar (±5 mbar) cannot be achieved when Gain values exceed 1000±50

Ranked causes

  1. 1. Blower flow insufficient (common)
  2. 2. Expiratory valve membrane degraded (common)
  3. 3. System pneumatic leak (common)
  4. 4. Pressure gauge not zeroed or incorrect units (occasional)
  5. 5. Pressure sensor assembly malfunction (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Exp. valve calibration fails or takes excessive time

Calibration takes >3 minutes or fails; pressure levels cannot be reached

Ranked causes

  1. 1. Expiratory membrane damaged or dirty (common)
  2. 2. Expiratory valve cable disconnected (common)
  3. 3. Faulty expiratory valve (occasional)
  4. 4. Expiratory valve cover failure (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Expiratory valve calibration takes >3 minutes or fails

Expiratory valve calibration procedure exceeds 3 minute duration or does not complete successfully

Ranked causes

  1. 1. Expiratory valve membrane defective (common)
  2. 2. Expiratory valve cable connection fault (common)
  3. 3. Expiratory valve malfunction (common)
  4. 4. Expiratory valve cover defective (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flow system values out of tolerance

Flow measurement values are outside acceptable tolerance during system test

Ranked causes

  1. 1. Flow sensor requires calibration (common)
  2. 2. Flow sensor malfunction (common)
  3. 3. Blower malfunction (common)
  4. 4. Qvent sensor malfunction (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Nebulizer valve test fails

Nebulizer valve pneumatic test does not pass with NOT OK status message

Ranked causes

  1. 1. High pressure oxygen supply disconnected (common)
  2. 2. Flow sensor requires calibration (common)
  3. 3. Qaw measurement abnormal (occasional)
  4. 4. O2 mixer assembly malfunction (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm lamp remains dark or displays incorrect color patient safety

Alarm lamp does not illuminate or displays wrong color (red, green, blue) during test

Ranked causes

  1. 1. Alarm lamp board connection fault (common)
  2. 2. Alarm lamp board defective (common)
  3. 3. Interaction panel board malfunction (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm monitor tests fail patient safety

One or more alarm monitoring tests do not pass

Ranked causes

  1. 1. Outdated software version (common)
  2. 2. Ventilator mainboard failure (common)
  3. 3. Persistent hardware issue (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm system loudspeaker does not sound patient safety

Loudspeaker does not produce alarm sounds or audible alerts

Ranked causes

  1. 1. Loudspeaker incorrectly connected (common)
  2. 2. Loudspeaker defective (common)
  3. 3. Interaction panel board malfunction (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audio Pause key illumination does not toggle

Audio Pause key illumination does not turn on or off

Ranked causes

  1. 1. Alarm lamp board connection fault (common)
  2. 2. Alarm lamp board failure (common)
  3. 3. Interaction panel board malfunction (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Backup alarm lamp not blinking patient safety

Backup alarm lamp does not blink or flash during alarm state

Ranked causes

  1. 1. Alarm lamp board connection fault (common)
  2. 2. Alarm lamp board failure (common)
  3. 3. Interaction panel board malfunction (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery state of health below threshold

Battery SoH is less than 20% or cell voltages are asymmetrical (>100 mV difference), or battery error is ≥5%

Ranked causes

  1. 1. Battery state of health (SoH) < 20% (common)
  2. 2. Cell voltages asymmetrical (>100 mV difference) (occasional)
  3. 3. Battery error ≥ 5% (occasional)
First check

Check state of health (SoH) of the battery

Test point
Battery management system
Expected
SoH < 20% requires replacement; SoH 20-40% optional replacement

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery state of health degraded patient safety

Battery SoH monitoring and replacement criteria during maintenance

Ranked causes

  1. 1. State of health (SoH) less than 20% (unknown)
  2. 2. Cell voltages asymmetrical (difference > 100 mV) (unknown)
  3. 3. Battery calibration error ≥ 5% (unknown)
First check

Check state of health (SoH) of the battery

Test point
Battery monitoring system
Expected
≥ 20% for retention, ≥ 40% for optional replacement consideration

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Binary valve test fails

Binary or autozero valve pneumatic test does not pass

Ranked causes

  1. 1. Autozero valves incorrectly connected (common)
  2. 2. FFC cable improperly connected (common)
  3. 3. Pressure sensor assembly failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Expiratory valve cannot reach certain pressure levels

Several or certain pressure levels are not attainable during expiratory valve calibration

Ranked causes

  1. 1. Expiratory valve membrane defective (common)
  2. 2. Expiratory valve cable disconnected (common)
  3. 3. Expiratory valve hardware failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

HAMILTON-C6 logo illumination does not toggle

HAMILTON-C6 logo illumination does not turn on or off

Ranked causes

  1. 1. Key panel connection fault (common)
  2. 2. Key panel defective (common)
  3. 3. Interaction panel board malfunction (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

HAMILTON-H900 humidifier not communicating with C6 ventilator

HAMILTON-H900 symbol does not appear on bottom left of display after power-on and cable connection

Ranked causes

  1. 1. AC power cable not connected to dedicated socket on ventilator rear (common)
  2. 2. Communication cable not connected to COM 3 port (common)
  3. 3. Ventilator not in Application mode (occasional)
First check

Verify AC power cable is plugged into dedicated socket on ventilator rear

Test point
Dedicated power socket on ventilator rear
Expected
Cable securely connected

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Insp. valve calibration cannot be performed

Calibration fails to execute or Qvent does not reach target value

Ranked causes

  1. 1. Insp. valve not properly connected (common)
  2. 2. Faulty insp. valve (occasional)
  3. 3. Qvent sensor failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inspiratory valve calibration cannot be performed

Calibration procedure fails to execute on the inspiratory valve circuit

Ranked causes

  1. 1. Inspiratory valve improperly connected (common)
  2. 2. Inspiratory valve malfunction (common)
  3. 3. Qvent sensor failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

IntelliCuff communication not established

IntelliCuff symbol missing from ventilator display or communication error message shown after USB connection

Ranked causes

  1. 1. USB cable not connected correctly (common)
  2. 2. IntelliCuff Control software option not installed (occasional)
  3. 3. Ventilator not in Application mode (occasional)
First check

Verify ventilator is in Application mode

Test point
Ventilator display
Expected
Application mode active

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

IntelliCuff pressure loss during checks patient safety

Pressure reading drops below expected 60 mbar when measured with hand pump at IntelliCuff port or after cuff connection

Ranked causes

  1. 1. Incorrect tube connection at IntelliCuff port (common)
  2. 2. IntelliCuff pressure tube worn or damaged (occasional)
  3. 3. Check valve malfunction on rear ventilator unit (occasional)
First check

Apply pressure with hand pump to IntelliCuff port

Test point
IntelliCuff port on front of ventilator unit
Expected
60 mbar
Tools
Hand pump PN 500330, tube 4 mm internal diameter

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loudspeaker sounds distorted or too quiet patient safety

Alarm sounds are distorted, muffled, or insufficient volume

Ranked causes

  1. 1. Loudspeaker improperly connected (common)
  2. 2. Loudspeaker defective (common)
  3. 3. Interaction panel board failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 input test fails

Oxygen input pneumatic test does not pass with NOT OK status message

Ranked causes

  1. 1. High pressure oxygen supply disconnected (common)
  2. 2. O2 proportional valve cable connection fault (common)
  3. 3. O2 mixer assembly malfunction (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 sensor value out of range

Displayed oxygen sensor value is outside acceptable range

Ranked causes

  1. 1. O2 sensor requires recalibration (common)
  2. 2. O2 sensor malfunction (common)
  3. 3. O2 mixer assembly connection fault (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pressure system values out of tolerance

Pressure measurement values are outside acceptable tolerance during system test

Ranked causes

  1. 1. System pneumatic leak (common)
  2. 2. Expiratory valve malfunction (common)
  3. 3. Pressure sensor malfunction (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Qvent does not reach target value

Flow measurement does not achieve expected calibration target during expiration monitoring

Ranked causes

  1. 1. Inspiratory valve connection fault (common)
  2. 2. Inspiratory valve defective (common)
  3. 3. Qvent sensor malfunction (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Rear and/or bottom fan test fails

Rear or bottom ventilation fan does not pass diagnostic test

Ranked causes

  1. 1. Fan connection fault (common)
  2. 2. Fan malfunction (common)
  3. 3. Ventilator mainboard failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

RS-232 communication interface not responding

Patient monitoring system cannot communicate with ventilator via RS-232 interface; no data signal detected

Ranked causes

  1. 1. RS-232 interface not activated (common)
  2. 2. Communication port configured to wrong protocol (common)
  3. 3. Defective RS-232 cable or connector (occasional)
First check

Verify RS-232 interface is activated

Test point
Device configuration menu
Expected
RS-232 interface enabled status

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

S logo illumination does not toggle

S logo illumination does not turn on or off

Ranked causes

  1. 1. Key panel incorrectly connected (common)
  2. 2. Key panel defective (common)
  3. 3. Interaction panel board malfunction (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 measurement not functional patient safety

SpO2 sensor communication board not responding or sensor reading not available in system display.

Ranked causes

  1. 1. SpO2 communication board not installed (common)
  2. 2. SpO2 sensor not activated in hardware options (common)
  3. 3. Incorrect sensor adapter type selected (occasional)
First check

Verify SpO2 communication board is installed

Test point
System Info tab
Expected
SpO2 communication board information displayed

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Tightness system test fails

Pneumatic tightness or leak test does not pass

Ranked causes

  1. 1. System pneumatic leak (common)
  2. 2. Expiratory valve membrane degraded (common)
  3. 3. Pneumatic component defective (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarming system test shows incorrect priorities patient safety

Alarm priority sequence or hierarchy is incorrect during system test

Ranked causes

  1. 1. Outdated software version (common)
  2. 3. Persistent priority issue (rare)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ambient valve test fails

Ambient valve pneumatic test does not pass with NOT OK status message

Ranked causes

  1. 1. Ambient valve improperly installed (common)
  2. 2. Ambient valve malfunction (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery calibration required message displayed patient safety

Device displays 'Battery calibration required' message on screen

Ranked causes

  1. 1. Battery capacity reduced by more than 25% (common)
  2. 2. Battery requires calibration cycle (occasional)
First check

Verify battery capacity via Service Software

Test point
Service Software display (SW ≥ 1.0.0)
Expected
Full charge capacity (fc.) value shown; capacity should be >= 80% of designed capacity (4000 mAh for HAMILTON-C6), >= 80% of design capacity

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bottom fan airflow insufficient caution

Paper does not stick to bottom fan inlet; airflow not being drawn into ventilator correctly

Ranked causes

  1. 1. Bottom fan motor failure or blockage (occasional)
  2. 2. Fan motor not powered (occasional)
First check

Visually inspect bottom fan inlet for blockage

Test point
Bottom fan inlet
Expected
Clear of obstructions, no dust accumulation

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Buzzer does not sound during power loss patient safety

Alarm buzzer fails to activate or sounds for less than two minutes when power is lost

Ranked causes

  1. 1. Super capacitors (gold caps) not fully charged (common)
  2. 2. Mainboard failure or defective gold capacitors (rare)
First check

Verify device has been powered for at least 20 minutes

Test point
Device operation time
Expected
20 minutes minimum operating time in standby or service mode

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Device does not switch to battery power patient safety

Ventilator does not transition to battery operation when mains power is disconnected

Ranked causes

  1. 1. Battery not installed or discharged (common)
  2. 2. Battery charging circuit failure (occasional)
First check

Verify internal batteries are installed

Test point
Battery compartment
Expected
Batteries present and seated

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Expiratory valve heater element not warming caution

Expiratory valve internal surface does not heat up during operation; heater element not functioning

Ranked causes

  1. 1. Heater element defective (occasional)
  2. 2. Heater not powered during HiFlow mode (common)
First check

Verify device is powered from AC mains

Test point
AC power inlet
Expected
Device operating on AC power

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Nurse call relay not responding patient safety

Nurse call alarm relays do not change state during alarm activation or ventilation phases; interface not functional

Ranked causes

  1. 1. Nurse call connector not properly seated (common)
  2. 2. Relay contacts defective or stuck (occasional)
First check

Verify nurse call connector is installed

Test point
Communication board connector
Expected
PN 160166 connector fully seated

4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 mixer system values out of tolerance

Oxygen mixer measurement values are outside acceptable tolerance during system test

Ranked causes

  1. 1. High pressure oxygen supply disconnected (common)
  2. 2. Qvent sensor malfunction (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 sensor calibration fails

Gain calibration cannot be completed for oxygen sensor

Ranked causes

  1. 1. O2 sensor cable disconnected or damaged (common)
  2. 2. Faulty O2 sensor (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 sensor gain calibration fails

Oxygen sensor gain calibration procedure does not complete or fails validation

Ranked causes

  1. 1. O2 sensor cable disconnected (common)
  2. 2. O2 sensor malfunction (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Proximal flow test fails

Proximal flow measurement test does not pass with NOT OK status message

Ranked causes

  1. 1. Qvent sensor malfunction (common)
  2. 2. Pressure sensor assembly failure (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Proximal pressure test fails

Proximal pressure measurement test does not pass with NOT OK status message

Ranked causes

  1. 1. Autozero valve malfunction (common)
  2. 2. Pressure sensor assembly failure (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Proximal test flow through sensors not equal

Flow through both flow sensor connectors are not equal during rinse flow test

Ranked causes

  1. 1. Pressure sensor assembly malfunction (common)
  2. 2. Rinse flow assembly malfunction (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Rear fan airflow insufficient caution

Paper does not stick to rear fan inlet; airflow not being drawn into ventilator correctly

Ranked causes

  1. 1. Rear fan motor failure or blockage (occasional)
  2. 2. Fan motor not powered (occasional)
First check

Visually inspect rear fan inlet for blockage

Test point
Rear fan inlet
Expected
Clear of obstructions, no dust accumulation

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Status indicator LEDs not appearing as required

One or more LEDs on the status indicator board do not display

Ranked causes

  1. 1. Status indicator board connection fault (common)
  2. 2. Ventilator mainboard failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm loudness not increasing

Alarm volume does not increase when loudness setting is adjusted to maximum

Ranked causes

  1. 1. Alarm system malfunction (unknown)
First check

Verify device is connected to mains AC power

Test point
AC power inlet
Expected
Device powered from mains

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ambient test fails

Ambient sensor or environmental test does not pass

Ranked causes

  1. 1. Ventilator mainboard failure (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Autozero test fails

Autozero valve pneumatic test does not pass with NOT OK status

Ranked causes

  1. 1. Pressure sensor assembly failure (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery capacity degradation at end-of-life patient safety

Reduced battery capacity leads to unreliable fuel gauge information and risk of unexpected ventilation stop during power failure

Ranked causes

  1. 1. Battery end-of-life degradation (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery capacity reduced below threshold patient safety

Device displays 'Battery calibration required' message; capacity loss exceeds 25% of designed capacity

Ranked causes

  1. 1. Battery aged past 80% capacity threshold (common)
First check

Check battery capacity in Service Software

Test point
Service Software display
Expected
Capacity at or above 4000 mAh (80% of 5000 mAh design capacity), ≥4000 mAh
Tools
Service Software SW ≥ 1.0.0

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery cell voltages asymmetrical patient safety

Cell voltage difference exceeds 100 mV, indicating unbalanced or failing battery cells.

Ranked causes

  1. 1. Battery cells unbalanced or failing (occasional)
First check

Read cell voltages from battery data

Test point
Battery data via service software
Expected
Cell voltage difference ≤ 100 mV, ±100 mV

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery condition maintenance check

Routine assessment of battery health and proper storage conditions during preventive maintenance

Ranked causes

  1. 1. Annual maintenance cycle requirement (common)
First check

Verify battery manufacture date and capacity history

Test point
Battery connector (top) and Service Software
Expected
Manufacture date visible; capacity trend stable or improving; cycle count documented

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Battery error calibration required
  • Battery error reading exceeds tolerance
  • Battery state of health low
  • Binary valve test shows short circuit
  • Calibration: Battery charger maintenance cycle
  • Calibration: Battery fuel gauge
  • Excess pressure in breathing circuit during inspiration
  • Exhalation obstructed alarm triggered
  • Front panel keys or LEDs do not function
  • Knob range does not function
  • QO2 value out of tolerance
  • Reset calibration to default settings
  • Weak batteries causing incorrect fuel gauge or unexpected ventilation stop
  • Weak batteries detected

14 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
132001Low airway pressureAirway pressure falls below set thresholdwarning
132002Check flow sensor tubingFlow sensor tubing issue detectedwarning
132003Check flow sensorFlow sensor malfunction detectedwarning
132004Check flow sensorFlow sensor missing or not detectederror
132005O2 sensor defectiveO2 sensor malfunction detectederror
132006O2 sensor missingO2 sensor not detectederror
132007O2 sensor not system compatibleIncompatible O2 sensor detectederror
132008O2 sensor calibration neededO2 sensor requires calibrationwarning
132010CO2 sensor disconnectedCO2 sensor connection losterror
132011CO2 sensor defectCO2 sensor malfunction detectederror
132012CO2 sensor over temperatureCO2 sensor temperature exceeds limiterror
132013CO2 sensor warmupCO2 sensor warming up after power-on or resetinfo
132014Wrong flow sensorIncompatible or incorrect flow sensor detectederror
132015Flow sensor calibration neededFlow sensor requires calibrationwarning
132016Check CO2 airway adapterCO2 airway adapter issue detectedwarning
132017Check CO2 sampling lineCO2 sampling line problem detectedwarning
132018SpO2: patient disconnectedSpO2 sensor detects patient disconnectionwarning
132019SpO2: probe missingSpO2 probe not detectederror
132020SpO2: low perfusion indexBlood perfusion at sensor site too lowwarning
132021SpO2: light interferenceAmbient light interferes with SpO2 readingwarning
132022SpO2: sensor errorSpO2 sensor malfunction detectederror
132023SpO2: adapter missingSpO2 adapter not detected or disconnectederror
132024Circuit calibration neededPatient circuit requires calibrationwarning
132025SpO2: poor signalSpO2 signal quality insufficientwarning
132026Check flow sensor tubingFlow sensor tubing disconnected or problem detectedwarning
132027Check flow sensor tubingPressure differential sensor tubing disconnectedwarning
132028Flow sensor calibration neededFlow sensor requires low-priority calibrationwarning
133001Apnea detectedPatient apnea detected (no breathing)critical
133002Inspiratory volume limitationTidal volume exceeds set limitwarning
133003High pressureAirway pressure exceeds alarm thresholderror
133004Check patient interfaceProblem with patient interface detectedwarning
141001Pressure limitationVentilation reaches pressure limit during inspirationwarning
141002Vt highTidal volume exceeds high thresholdwarning
141003Vt lowTidal volume falls below low thresholdwarning
141004High minute volumeTotal minute ventilation exceeds high limitwarning
141005Low minute volumeTotal minute ventilation falls below low limitwarning
141006High frequencyRespiratory frequency exceeds high limitwarning
141007Low frequencyRespiratory frequency falls below low limitwarning
141008High oxygenO2 concentration exceeds high limitwarning
141009Low oxygenO2 concentration falls below low limiterror
141010ApneaPatient apnea detected (no breathing)critical
141011Apnea ventilation endedApnea ventilation backup support endedinfo
141012Apnea ventilationBackup ventilation activated due to apneawarning
141013High pressure during sighAirway pressure exceeds limit during sigh breathwarning
141014Turn the flow sensorFlow sensor orientation incorrectwarning
141016Disconnection on ventilator sideVentilator detects disconnection at ventilator-side circuiterror
141017Disconnection on patient sideVentilator detects disconnection at patient interfaceerror
141018Exhalation obstructedExhalation pathway blocked or obstructederror
141019IRVInspiratory reserve volume alarm activatedunknown
141020ASV: Cannot meet targetAdaptive Support Ventilation unable to achieve targetwarning
141021Pressure limit has changedPressure limit parameter modified during ASV operationinfo
141023External flow sensor failedExternal flow sensor malfunction detectederror
141024Low pressureAirway pressure falls below low thresholderror
141025PetCO2 highEnd-tidal CO2 exceeds high limitwarning
141026PetCO2 lowEnd-tidal CO2 falls below low limitwarning
141027Suctioning maneuverSuctioning maneuver activated or in progressinfo
141028Performance limited by high altitudeVentilator performance affected by high altitudewarning
141029High SpO2Oxygen saturation exceeds high limitwarning
141030Low SpO2Oxygen saturation falls below low limit (medium priority)warning
141031High pulseHeart rate exceeds high limitwarning
141032Low pulseHeart rate falls below low limitwarning
141033High PIPerfusion index exceeds high limitwarning
141034Low PIPerfusion index falls below low limitwarning
141035High SpHbHemoglobin saturation exceeds high limitwarning
141036Low SpHbHemoglobin saturation falls below low limitwarning
141037High SpCOCarboxyhemoglobin exceeds high limiterror
141038Low SpCOCarboxyhemoglobin falls below low limitinfo
141039High SpMetMethemoglobin exceeds high limitwarning
141040Low SpMetMethemoglobin falls below low limitinfo
141041Low SpO2Oxygen saturation falls below low limit (high priority)critical
141043Ventilation adjustment OFFVentilation adjustment function disabledinfo
141044Oxygenation adjustment OFFOxygenation adjustment function disabledinfo
141045Ventilation Controller at limitVentilation controller reached upper limitwarning
141046Oxygenation Controller at limitOxygenation controller reached upper limitwarning
141047High flowFlow rate exceeds high limit in nCPAP modewarning
141048ObstructionCircuit obstruction detectederror
141049High flowFlow rate exceeds high limitwarning
141050Sensor simulation activeSensor simulation mode is active during operationcritical
141051High PEEPPositive end-expiratory pressure exceeds high limitwarning
141052Loss of PEEPPEEP falls below set threshold or is losterror
141053FiO2 set to 100% due to low SpO2Oxygen concentration automatically set to maximum due to hypoxiacritical
141054Recruitment in progressRecruitment maneuver runninginfo
141055Oxygen control limit exceededOxygen concentration exceeded configured limitwarning
141056SBT abortedSpontaneous breathing trial did not completewarning
141057SBT successfully fulfilledSpontaneous breathing trial completed successfullyinfo
141058High PVIPleth Variability Index above thresholdwarning
141059Low PVIPleth Variability Index below thresholdwarning
141060Check flow sensor for waterWater detected in flow sensorwarning
141061Check flow sensor for water (high priority)Water in flow sensor requiring immediate attentionwarning
141062SpeakValve ONSpeaking valve activatedinfo
141063SpeakValve OFFSpeaking valve deactivatedinfo
141064Vt low (high priority)Tidal volume below alarm limitwarning
141065Check patient interfacePatient interface connectivity or configuration errorwarning
141066Maximum leak compensationLeak compensation reaches maximum thresholdwarning
231001Technical event: pressure controller pressure lowInspiratory pressure below normal operating rangeerror
231002Technical event: pressure controller pressure highInspiratory pressure above normal operating rangeerror
231003Technical event: flow controller flow lowMeasured flow below expected rangeerror
231004Technical event: flow controller flow highMeasured flow exceeds normal rangeerror
231005Technical event: inspiration valve leakInspiratory valve not sealing properlyerror
231006Oxygen supply failedOxygen flow from controller too lowcritical
231007Technical event: O2 controller flow highOxygen flow exceeds normal rangeerror
231008Technical event: O2 valve leakOxygen valve not sealingerror
231009Technical event: blower controller speed lowBlower rotation speed below normalerror
231010Technical event: blower controller speed highBlower rotation speed exceeds normal rangeerror
231011Vent outlet temperature highExhaust temperature exceeds safe limitwarning
231012Technical event: qvent flow sensor defectMain flow sensor malfunction detectederror
231013Technical event: qO2 flow sensor defectOxygen flow sensor malfunctionerror
231014Technical event: ambient valve errorAmbient pressure relief valve malfunctionerror
231017Blower service requiredBlower maintenance interval reached or motor degradation detectedwarning
231018Technical event: self-test with errorInternal self-test detected abnormalityerror
231019Technical event: inspiration valve sensor defectInspiratory valve position sensor failureerror
231020Technical event: regulator data log onRegulator data logging active for diagnosticsinfo
231022Technical event: pExp valve sensor defectExpiratory pressure valve sensor malfunctionerror
231023Technical event: invalid flow sensorFlow sensor type not recognized or unsupportederror
231027Wrong expiratory valveExpiratory valve type does not match ventilator configurationerror
231032Technical event: expiration valve disconnectedExpiratory valve not properly connectederror
231033Technical event: blower controller speed limitedBlower speed reduced below normal operating rangewarning
231036Technical event: pvent pressure sensor defectInspiratory pressure sensor malfunctionerror
232002Technical event: pvent monitor sensor defectPressure monitoring sensor malfunctionerror
232003Technical event: Paw sensor defectAirway pressure sensor malfunctionerror
232004Pressure not releasedSystem unable to vent pressurized circuitwarning
232005Technical event: blower hotBlower motor temperature elevatedwarning
232006Technical event: blower temperature sensor defectTemperature sensor malfunction on blowererror
232007Check flow sensor tubingAirway flow sensor tubing obstruction or disconnectionwarning
232008Technical event: pambient sensor defectAmbient pressure sensor malfunctionerror
232028Technical event: pFilter pressure highPressure drop across inlet filter exceeds limiterror
232029Technical event: tinst sensor defectInternal temperature sensor malfunctionerror
232030Technical event: clock errorInternal real-time clock malfunctionerror
232034Replace O2 sensorOxygen cell depleted or defectivewarning
232035Technical event: pfilter sensor defectFilter pressure sensor malfunctionerror
232038CO2 calibration neededCarbon dioxide sensor requires calibrationwarning
232045Technical event: SpO2 sensor settings errorPulse oximetry sensor configuration errorerror
232056Technical event: pAmbient pFilter mismatchAmbient and filter pressure sensors disagreeerror
232057Technical event: SpO2 sensor change failedPulse oximetry sensor replacement not recognizederror
233008Aerogen nebulizer disconnectedAerogen nebulizer becomes disconnected from ventilatorwarning
234001Loudspeaker defectiveLoudspeaker failure detected during operation or startupwarning
241001Replace HEPA filterHEPA filter has reached end of service lifewarning
241002Release valve defectiveTightness test detects release valve failureerror
243004Buzzer defectiveBuzzer failure detected at startupwarning
243007Realtime clock failureReal-time clock (RTC) has been reset or failederror
243010Panel connection lostCommunication lost with control panelerror
244001Loss of external powerMains power supply disconnected or failedcritical
244002Battery 1: Calibration requiredBattery 1 requires calibrationwarning
244003Battery 2: Calibration requiredBattery 2 requires calibrationwarning
244004Battery 1: Temperature highBattery 1 temperature exceeds safe operating rangewarning
244005Battery 2: Temperature highBattery 2 temperature exceeds safe operating rangewarning
244006Battery lowBattery charge level falls below high priority thresholdwarning
244007Battery power lossBattery power supply has failed completelycritical
244008Battery 1: Wrong batteryIncompatible or incorrect battery installed in slot 1error
244009Battery 2: Wrong batteryIncompatible or incorrect battery installed in slot 2error
244010Battery lowBattery charge level falls below low priority thresholdwarning
244011Battery communication errorBattery management bus communication failureerror
244012Battery 1: DefectiveBattery 1 hardware failure detectederror
244013Battery 2: DefectiveBattery 2 hardware failure detectederror
244015Battery lowBattery charge level falls below medium priority thresholdwarning
244016Battery 1: Replacement requiredBattery 1 has reached end of service lifewarning
244017Battery 2: Replacement requiredBattery 2 has reached end of service lifewarning
244019Technical event: 28V_IP voltage out of tolerance28V auxiliary voltage outside specified limitserror
244022Continue charging batteryBattery charging in progress; normal operationinfo
246001Preventive maintenance requiredService interval reachedwarning
246003Fan failureCooling fan hardware failure detectederror
246016Touch not functionalTouch screen or touch panel failed during POSTerror
246020Loudspeaker defectiveHMI loudspeaker hardware failurewarning
246040Ventilator overpressure self-test failedSelf-test execution during startup or diagnosticserror
246041Alarm LED errorLED circuit failure or lamp defecterror
249001Options not foundConfiguration file missing or unreadable on startuperror
249002Option file write errorFailed to write option file to storageerror
249003Option file version mismatchOption file version incompatible with current firmwareerror
249004Default option file generatedOption file missing or unreadable, defaults appliedinfo
249010Device configuration file errorDevice configuration file read/write failureerror
249011Setup configuration file errorSetup configuration file read/write failureerror
249012Last setting configuration file errorLast settings file read/write failureerror
249013Research configuration file errorResearch configuration file read/write failureerror
249020Invalid communication boardUnrecognized or incompatible option board detectederror
249021Check hardware compatibilityInstalled hardware incompatible with firmware or configurationwarning
250001Check humidifier communicationHumidifier interface communication failureerror
250002Humidifier water lowWater level below minimum thresholdwarning
250003Humidifier chamber temperature lowChamber temperature below set pointwarning
250004Humidifier Y-piece temperature lowY-piece temperature below set pointwarning
250005Humidifier check right tubeRight tube temperature or connection issue detectedwarning
250006Humidifier check left tubeLeft tube temperature or connection issue detectedwarning
250007Humidifier check chamberChamber malfunction or connection error detectedwarning
250008Check humidifierGeneral humidifier malfunction detectedwarning
250009Humidifier errorHumidifier subsystem critical errorerror
250010Humidifier water highWater level above maximum thresholdwarning
250011Humidifier chamber temperature highChamber temperature above set pointwarning
250012Humidifier Y-piece temperature highY-piece temperature above set pointwarning
250013Humidifier tiltHumidifier orientation/tilt detected abnormalcritical
250014Check humidifierHumidifier malfunction or high-level fault detectederror
251001Check IntelliCuff communicationIntelliCuff interface communication failureerror
251002Check IntelliCuffIntelliCuff battery low or malfunctionwarning
251003Cannot turn off IntelliCuffIntelliCuff power-off command failederror
251004Check IntelliCuffIntelliCuff battery level critically lowwarning
251005Cuff pressure highMeasured cuff pressure exceeds safe limitcritical
251006Cuff deflatedCuff remained deflated for extended periodwarning
251007Cuff leakCuff pressure drop indicating air leakerror
251008Check IntelliCuffIntelliCuff malfunction at medium severity levelwarning
251009Cuff leakIntelliCuff disconnection or interface failure detectederror
251010Check IntelliCuffIntelliCuff critical error conditionerror
251011Check IntelliCuffIntelliCuff battery at critical levelcritical
251012Check IntelliCuffIntelliCuff malfunction at high severity levelerror
251013Check IntelliCuff stateIntelliCuff state invalid or unexpectedwarning
255001Percentage minimum volume high limit test failedSelf-test of min volume limit failed during startuperror
255002Percentage minimum volume high limit wrongConfigured min volume limit exceeds acceptable rangewarning
255003Percentage minimum volume low limit test failedSelf-test of min volume limit failed during startuperror
255004Percentage minimum volume low limit wrongConfigured min volume limit below acceptable rangewarning
255005PEEP high limit wrongConfigured PEEP high limit exceeds acceptable rangewarning
255006PEEP high limit test failedSelf-test of PEEP high limit failederror
255007PEEP low limit wrongConfigured PEEP low limit below acceptable rangewarning
255008PEEP low limit test failedSelf-test of PEEP low limit failederror
255009Oxygen low test failedSelf-test of oxygen system failederror
255010Oxygen high test failedSelf-test of oxygen system upper limit failederror
255011Percentage minimum volume positive change test failedSelf-test of min volume increase failederror
255012Percentage minimum volume negative change test failedSelf-test of min volume decrease failederror
255013PEEP positive change test failedSelf-test of PEEP increase failederror
255014PEEP negative change test failedSelf-test of PEEP decrease failederror
255015Oxygen positive change test failedSelf-test of oxygen increase failederror
255016Oxygen negative change test failedSelf-test of oxygen decrease failederror
255017Recruitment running time exceededRecruitment maneuver running time exceeded limitwarning
255018Simulation supervise test failedInternal simulation/test supervision failederror
255019Percentage minimum volume high fine adjustment test failedSelf-test of min volume fine limit failederror
255020Percentage minimum volume high fine limit wrongConfigured min volume fine limit exceeds rangewarning
255021Percentage minimum volume low fine adjustment test failedSelf-test of min volume fine limit failederror
255022Percentage minimum volume low fine limit wrongConfigured min volume fine limit below rangewarning
255023Oscillation oxygenOxygen parameter oscillating during operationwarning
255024Oscillation PEEP/CPAPPEEP/CPAP pressure oscillating during operationwarning
255025Oscillation percentage minimum volumeMinimum volume parameter oscillating during operationwarning
255026Data mismatchInternal data consistency error detectederror
255027PetCO2 frozen test failedPetCO2 measurement frozen or not updatingerror
255028SpO2 frozen test failedSpO2 measurement frozen or not updatingerror
255029Spontaneous frequency frozen test failedSpontaneous breathing rate frozen or not updatingerror
283001Startup failedDisplay GUI startup failed during booterror
283003Language not loadedLanguage file failed to load on startuperror
283005Settings file errorDisplay settings file read/write failureerror
283008Check settingsInvalid or suspicious settings detectedwarning
283009Self test failedDisplay self-test failed during startuperror
283010No ventilation after power-failVentilation did not resume after power failure recoverycritical
283011Check IntelliCuff stateIntelliCuff state invalid on display startuperror
283012Panel settings file errorPanel hardware settings file read/write failureerror
284002Technical event alarm service highService alarm high priority triggeredcritical
284003Technical event alarm service mediumService alarm medium priority triggeredwarning
284004Technical event alarm service lowService alarm low priority triggeredinfo
284005Flow sensor calibration neededFlow sensor calibration overdue or requiredwarning
284006Restart device to enable tests/calibrationDevice restart required to access test/calibration modeinfo
285001Alarm lamp error defectError alarm lamp hardware failure detectederror
285002Alarm lamp warning defectWarning alarm lamp hardware failure detectederror
285003Backlight defectDisplay backlight hardware failure detectederror
285004Function key not operationalHardware key/button malfunction detectederror
285005Touch not functionalTouch screen interface malfunction detectederror
291001Remote communication error (COM1)COM1 port communication error detectederror
291002Remote communication error (COM2)COM2 port communication error detectederror
291003Remote communication timeout (COM1)COM1 communication timeout no response receivederror
291004Remote communication timeout (COM2)COM2 communication timeout no response receivederror
296001Remote control activeRemote control mode is currently activeinfo
332001Flow sensor errorSafety mode technical failure detectedcritical
341001Breath settings not acceptedSettings rejected during mode configurationerror
341002Returned breath settings incorrectMismatch between sent and confirmed settingserror
341003Adaptive settings invalidAdaptive control settings validation failurewarning
341004Breath settings timeoutSettings not confirmed within timeout windowerror
341005Controller settings invalidController parameter validation failederror
341006Returned alarm limits incorrectAlarm limit confirmation mismatcherror
341007Mode transition not definedRequested ventilation mode transition unsupportedwarning
341008Ventilation flow sensor defectFlow sensor failure detectedcritical
341009Ventilation pressure sensor defectPressure sensor failure detectedcritical
343001Communication timeoutAlarm module communication losterror
346002Watchdog failed: ALRLow level supervisor watchdog timeout on alarm modulecritical
346003Watchdog failed: ALR measurementAlarm module measurement watchdog timeoutcritical
346004Watchdog failed: ESLLow level supervisor watchdog timeout on ESLcritical
346005Watchdog failed: ESL MMI logWatchdog timeout in MMI logging taskwarning
346006Watchdog failed: ESL control logWatchdog timeout in control logging taskerror
346007Watchdog failed: ESL breath logWatchdog timeout in breath monitoring logerror
346008Watchdog failed: ESL event SDR readerWatchdog timeout in event storage device readererror
346009Watchdog failed: ESL service SDR readerWatchdog timeout in service storage device readerwarning
346010Watchdog failed: ESL event SDR writerWatchdog timeout in event storage device writererror
346011Watchdog failed: ESL service SDR writerWatchdog timeout in service storage device writerwarning
346012Watchdog failed: ESL measurementWatchdog timeout in ESL measurement processingcritical
346013Watchdog failed: GUIL touchWatchdog timeout in touch screen interfaceerror
346014Watchdog failed: GUIL PT knobWatchdog timeout in pressure-time knob interfaceerror
346015Watchdog failed: GUIL hardkeysWatchdog timeout in hardkey input processingerror
346016Watchdog failed: GUIL screenshotWatchdog timeout in screenshot capture functionwarning
346017Watchdog failed: LMLow level supervisor watchdog timeout on language moduleerror
346019Watchdog failed: LLS hardware exceptionHardware exception in low level supervisorcritical
346020Watchdog failed: LLS common serverWatchdog timeout in common service servererror
346022Watchdog failed: LLS alarm statusWatchdog timeout in alarm status monitoringerror
346023Watchdog failed: LLS measurementWatchdog timeout in LLS measurement processingcritical
346024Watchdog failed: PMWatchdog timeout on power management modulecritical
346025Watchdog failed: PM measurementWatchdog timeout in PM measurement functionerror
346026Watchdog failed: QSPIWatchdog timeout on QSPI interfacecritical
346027Watchdog failed: QSPI MMI serverWatchdog timeout in QSPI MMI servererror
346028Watchdog failed: QSPI BM serverWatchdog timeout in QSPI blower/motor servererror
346029Watchdog failed: QSPI BPG serverWatchdog timeout in QSPI breath pattern generator servererror
346030Watchdog failed: QSPI LLS serverWatchdog timeout in QSPI LLS servercritical
346031Watchdog failed: QSPI interfaceWatchdog timeout on QSPI serial interfacecritical
346032Watchdog failed: QSPI measurementWatchdog timeout in QSPI measurementerror
346033Watchdog failed: RTC real time clockWatchdog timeout in real time clockerror
346034Watchdog failed: RTC alarm clockWatchdog timeout in alarm clock functionwarning
346035Watchdog failed: SCWatchdog timeout on service controllererror
346036Watchdog failed: SGUIWatchdog timeout on service GUIwarning
346037Watchdog failed: STUWatchdog timeout on STU moduleerror
346038Watchdog failed: VMCWatchdog timeout on ventilation mode controllercritical
346039Watchdog failed: VMC ventilation alarmingWatchdog timeout in VMC alarm taskerror
346040Watchdog failed: VMC ventilation controlWatchdog timeout in VMC control taskcritical
346041Watchdog failed: VMC ventilation monitoringWatchdog timeout in VMC monitoring taskcritical
346042Watchdog failed: VMC measurementWatchdog timeout in VMC measurementerror
346043Watchdog failed: VGUIWatchdog timeout on ventilation GUIerror
346044Watchdog failed: VGUI mode controlWatchdog timeout in VGUI mode control taskerror
346045Watchdog failed: VGUI measurementWatchdog timeout in VGUI measurementerror
346046Watchdog failed: LLS EEPROM writeWatchdog timeout in EEPROM write operationerror
346047Watchdog failed: BM alarming slowWatchdog timeout in blower motor alarmingwarning
346048Watchdog failed: SND sound controlWatchdog timeout in sound control modulewarning
346049Watchdog failed: SND soundWatchdog timeout in sound generationwarning
346050Watchdog failed: VTWatchdog timeout on ventilation timing modulecritical
346051Watchdog failed: CFG configurationWatchdog timeout in configuration moduleerror
346052Watchdog failed: EXM communication baseWatchdog timeout in external communication baseerror
346053Watchdog failed: IVSWatchdog timeout on IVS moduleerror
346054Safety failure detectedGeneral safety system failure detectedcritical
346055Watchdog failed: PD black box loggingWatchdog timeout in black box data loggingwarning
346056Watchdog failed: EXM COM1Watchdog timeout on EXM communication port 1error
346057Watchdog failed: EXM COM2Watchdog timeout on EXM communication port 2error
383001Settings not acceptedVentilation GUI settings rejectedwarning
383002Returned settings incorrectConfirmation of GUI settings does not match sent valueserror
383003Settings validation errorGUI settings fail validation checkwarning
383004Monitoring channel observation failedGUI monitoring channel lost or unresponsiveerror
383005VMC timeoutCommunication timeout between GUI and VMCerror
383006Returned device settings incorrectDevice settings confirmation mismatcherror
383007Trending channel observation failedGUI trending data channel lostwarning
383008Startup basic startup failureVGUI basic startup sequence failedcritical
383009Startup screen failureVGUI screen initialization failedcritical
383010Startup load GUI basics failureVGUI fails to load basic GUI resourcescritical
383011Startup selftest prolog failureVGUI startup self-test initialization failederror
383012Startup selftest failureVGUI startup self-test detected hardware faultcritical
383013Startup selftest epilog failureVGUI self-test cleanup phase failederror
383014Startup loading bitmaps failureVGUI fails to load bitmap resourceserror
383015Startup initialization failureVGUI initialization phase failedcritical
384001Startup basic startup failureSGUI basic startup sequence failedcritical
384002Startup screen failureSGUI screen initialization failedcritical
384003Startup load GUI basics failureSGUI fails to load basic GUI resourcescritical
384004Startup selftest prolog failureSGUI startup self-test initialization failederror
384005Startup selftest failureSGUI startup self-test detected hardware faultcritical
384006Startup loading bitmaps failureSGUI fails to load bitmap resourceserror
384007Startup initialization failureSGUI initialization phase failedcritical
385001Alarming channel observation failedAGL alarming channel disconnected or losterror
385002Safety failure detectedAGL detected safety system faultcritical
385003PM channel observation failedAGL PM monitoring channel losterror
386001Bitmap not loadedGUI layer fails to load bitmap resourceerror
431001Blower faultBlower motor malfunction in ambient modecritical
431002Blower disconnectedBlower is not connected or poweredcritical
431004Inspiration valve overcurrentInspiration valve motor draws excessive currenterror
431006Expiration valve overcurrentExpiration valve motor draws excessive currenterror
431007Vent output over temperatureVentilation output exceeds safe temperaturecritical
431008Q-vent flow sensor errorVentilation flow sensor malfunctioncritical
431009Q-O2 flow sensor errorOxygen flow sensor malfunctionerror
431010Control REGSPI timeoutSPI communication timeout to control registererror
431011Monitor REGSPI timeoutSPI communication timeout to monitor registererror
431012Flow sensor measurement impreciseFlow sensor reading variance exceeds tolerancewarning
431013Calibration read failedStored calibration data cannot be retrievederror
431014Inspiration valve sensor defectInspiration valve position sensor malfunctionerror
431015Emergency off failedEmergency shutdown did not execute successfullycritical
431017Inspiration valve disconnectedInspiration valve is not properly connectedcritical
432001Blower over temperatureBlower assembly exceeds safe operating temperaturecritical
432002Instrument over temperatureVentilation instrument reaches excessive temperaturecritical
432003Pressure not released ambientPressure relief failed in ambient modeerror
433001Breath monitoring tick timeoutBreath monitoring timing missed in BPGerror
433002Control tick timing errorControl loop tick timing error in BPGerror
443001Watchdog failed LLSAmbient mode low level supervisor watchdog timeoutcritical
444001Battery totally dischargedBattery voltage depleted below minimum thresholdcritical
444004Voltage out of tolerancePower supply voltage exceeds acceptable rangeerror
444005Shutdown failedControlled shutdown sequence did not completeerror
446001CPU temperature criticalCPU temperature exceeds critical limitcritical
446002LLS safety subsystem failureSafety monitoring system detects fault conditioncritical
446003LLS watchdog failed for gas delivery moduleWatchdog timer timeout in gas delivery controlcritical
446004Watchdog failure for valve regulator in gas deliveryValve regulator module timeout or lockupcritical
446005Watchdog failure for blower regulator in gas deliveryBlower regulator module timeout or halted executioncritical
446006Watchdog failure in gas delivery monitoring subsystemMonitoring watchdog expired due to module unresponsivenesscritical
446007Watchdog failure in valve control logging subsystemValve log watchdog timeout during logging operationerror
446008Watchdog failure in blower control logging subsystemBlower log watchdog timeout or logger module hangerror
446009Watchdog failure in flow measurement module of gas deliveryFlow measurement processor timeout or unresponsive executioncritical
446010GPIO fault input detection failed in gas delivery watchdogGPIO input monitoring watchdog expirederror
446011Breath management module watchdog failureBM (Breath Management) watchdog timeout detectedcritical
446012Watchdog failure in breath management gas delivery communicationBM-GD communication timeout or message queue failurecritical
446013Hardware control watchdog failure in breath managementHardware control watchdog expired or unresponsivecritical
446014Breath parameter generator BPG client receiver watchdog failedBPG client watchdog timeout in breath managementerror
446015Breath parameter generator BPG server receiver watchdog failedBPG server watchdog timeout or server process hangerror
446016Breath data calculation watchdog failure in breath managementBreath calculation algorithm timeout or hangcritical
446017Flow measurement watchdog failure in breath managementBM flow measurement module watchdog timeouterror
446018Breath parameter generator module watchdog failureBPG module watchdog timer expiredcritical
446019BPG gas delivery client receiver watchdog failedBPG client communication watchdog timeouterror
446020BPG flow measurement watchdog failureBPG flow measurement subsystem timeout or hangerror
446021Exception occurred in LLS softwareUnhandled exception or critical software error detectedcritical
446022LLS voltage error detectedSupply voltage monitoring detected out-of-range conditioncritical
446023LLS workload watchdog timeoutLLS workload processing exceeded timeout thresholdwarning
446024LLS 3.3V supply error3.3V regulated supply out of specificationcritical
446025LLS analog-to-digital converter errorADC calibration or conversion failure detectederror
446026LLS reference voltage errorVref supply out of range or unstableerror
446028LLS clock error detectedSystem clock monitor detected oscillator or PLL failurecritical
446029LLS ambient failure detectedAmbient monitor detected environmental fault conditionwarning
446030LLS watchdog occurred in analog moduleAnalog module watchdog timer timeouterror
446031Restart ventilation failed in LLSVentilation restart command failed or timed outcritical
446032Patient pressure overpressure condition detectedPatient line pressure exceeded safety thresholdcritical
481001STU first run migration errorSystem configuration migration failed on initial startuperror
481002CRC error in system configurationConfiguration data CRC check failed during loaderror
481003Unknown part number detectedUnrecognized component part number in system configurationwarning
481004Technical state failedSystem state validation failed during initializationerror
481005Synchronization error in system startupModule synchronization timeout during system booterror
481006Reconnect startup type errorReconnection startup mode not recognized or invalidwarning
481007Reconnect accept errorSystem rejected reconnection attempt during startuperror
483001VGUI startup basic startup failureVentilator GUI basic initialization failed during booterror
483002VGUI screen initialization failureVGUI screen driver initialization failederror
483003VGUI basics loading failureVGUI base framework failed to loaderror
483004VGUI self-test prolog failureVGUI self-test initialization sequence failedwarning
483005VGUI self-test failureVGUI internal self-test detected failure conditionerror
483006VGUI self-test epilog failureVGUI self-test completion sequence failedwarning
483007VGUI bitmap loading failureGUI bitmap or image resources failed to loaderror
483008VGUI initialization failureVGUI full initialization sequence failederror
484001SGUI startup basic startup failureStandby GUI basic initialization failederror
484002SGUI screen initialization failureSGUI screen driver failed to initializeerror
484003SGUI basics loading failureSGUI framework failed to load during startuperror
484004SGUI self-test prolog failureSGUI self-test initialization sequence failedwarning
484005SGUI self-test failureSGUI self-test detected internal failureerror
484007SGUI bitmap loading failureSGUI image resources failed to loaderror
484008SGUI initialization failureSGUI initialization sequence failederror
485001AGL ambient failure detectedAlarm GUI monitor detected environmental conditionwarning
743001Alarm controller watchdog failed for LLSALR module watchdog timeout detectederror
743003ALR MMI connection lostAlarm controller lost connection to man-machine interfacewarning
746001Patient panel LLS watchdog failureLLS watchdog timeout detected by ALR moduleerror
746002Patient panel VGUI watchdog failureVGUI module watchdog timeout detectederror
746003Patient panel VGUI mode control watchdog failureVGUI mode control watchdog timeouterror
746004Patient panel error failure detectedPanel error monitoring detected fault conditionerror
746005Patient panel LLS exception occurredUnhandled exception in patient panel LLSerror
746006Patient panel LLS workload watchdog failurePatient panel workload processing timeoutwarning
746007Patient panel ventilation restart failedPatient panel ventilation restart command failedcritical
783001VGUI monitoring channel observation failedVGUI monitoring signal lost or degradederror
783002VGUI trending channel observation failedVGUI trend data acquisition lost or interruptederror
783003VGUI VMC timeoutVMC (Ventilator Management Controller) communication timeouterror
785001AGL alarming channel observation failedAlarm GUI alarm channel signal losterror
785002AGL panel error failure detectedAGL detected error in alarm panel interfaceerror
785003AGL PM channel observation failedAGL parameter monitoring channel observation losterror
Battery calibration requiredBattery capacity has declined below acceptable thresholdBattery capacity loss exceeds 25% of design capacitywarning
Battery calibration requiredBattery capacity loss exceeds 25% of designed capacity; battery too weak for reliable operationBattery capacity drops below 75% of designed capacity (below 3750 mAh for 5000 mAh design)warning
Red flashing LEDFuel gauge calibration requiredBattery fuel gauge needs recalibration before usewarning
Red solid LEDCharger/calibrator errorFault detected during charge or calibration cycleerror
Weak batteryBattery capacity reduced; device may provide inaccurate fuel gauge or fail to maintain ventilation on AC power lossBattery end-of-life condition; capacity severely depletedcritical

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.