Hamilton Medical C6: troubleshooting and error codes
The ServiceForge knowledge base holds 74 documented symptoms for the Hamilton Medical C6, with 187 ranked causes, 62 diagnostic checks with test points and expected values, and 442 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Expiratory valve test fails
Expiratory valve pneumatic test does not pass with NOT OK status message
Ranked causes
- 1. Ambient valve malfunction (common)
- 2. Pvent_control requires calibration (common)
- 3. Expiratory valve requires calibration (common)
- 4. Expiratory valve membrane degraded (common)
- 5. Expiratory valve cover malfunction (occasional)
- 6. Internal pneumatic leak (common)
- 7. Expiratory valve malfunction (occasional)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow sensor calibration fails
Flow sensor calibration cannot be completed
Ranked causes
- 1. Adult/pediatric flow sensor mismatch (common)
- 2. System leaks (common)
- 3. Faulty proximal flow sensor (occasional)
- 4. Ambient valve malfunction (occasional)
- 5. Expiratory membrane damaged (occasional)
- 6. Pressure sensor component failure (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air entry pressure exceeds limit with HEPA filter
Air entry pressure is greater than 2 mbar with HEPA filter installed
Ranked causes
- 1. HEPA filter clogged (common)
- 2. Pfilter sensor connection fault (common)
- 3. Pambient sensor reading incorrect (occasional)
- 4. Mainboard failure (occasional)
- 5. Blower malfunction (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blower flow test fails
Blower airflow test does not pass with NOT OK status message
Ranked causes
- 1. Internal pneumatic leak (common)
- 2. HEPA filter clogged (common)
- 3. Qvent sensor malfunction (occasional)
- 4. Blower module malfunction (occasional)
- 5. Ventilator mainboard failure (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blower pressure test fails
Blower pressure pneumatic test does not pass with NOT OK status message
Ranked causes
- 1. Internal pneumatic leak (common)
- 2. Ambient valve malfunction (common)
- 3. Expiratory valve membrane degraded (common)
- 4. Blower module malfunction (occasional)
- 5. Ventilator mainboard failure (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inspiratory valve test fails
Inspiratory valve pneumatic test does not pass with NOT OK status message
Ranked causes
- 1. Internal pneumatic leak (common)
- 2. Qvent sensor malfunction (common)
- 3. Ambient valve malfunction (common)
- 4. Inspiratory valve requires calibration (common)
- 5. Inspiratory valve cable disconnected (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure calibration fails or reaches incorrect value
Cannot reach 50 mbar (± 5 mbar) or Gain values exceed 1000 ± 50
Ranked causes
- 1. Blower flow issue (common)
- 2. Expiratory valve membrane damage (common)
- 3. System leaks (occasional)
- 4. Measurement device not zeroed or incorrect unit (occasional)
- 5. Pressure sensor malfunction (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure calibration target not reached
50 mbar (±5 mbar) cannot be achieved when Gain values exceed 1000±50
Ranked causes
- 1. Blower flow insufficient (common)
- 2. Expiratory valve membrane degraded (common)
- 3. System pneumatic leak (common)
- 4. Pressure gauge not zeroed or incorrect units (occasional)
- 5. Pressure sensor assembly malfunction (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exp. valve calibration fails or takes excessive time
Calibration takes >3 minutes or fails; pressure levels cannot be reached
Ranked causes
- 1. Expiratory membrane damaged or dirty (common)
- 2. Expiratory valve cable disconnected (common)
- 3. Faulty expiratory valve (occasional)
- 4. Expiratory valve cover failure (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Expiratory valve calibration takes >3 minutes or fails
Expiratory valve calibration procedure exceeds 3 minute duration or does not complete successfully
Ranked causes
- 1. Expiratory valve membrane defective (common)
- 2. Expiratory valve cable connection fault (common)
- 3. Expiratory valve malfunction (common)
- 4. Expiratory valve cover defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow system values out of tolerance
Flow measurement values are outside acceptable tolerance during system test
Ranked causes
- 1. Flow sensor requires calibration (common)
- 2. Flow sensor malfunction (common)
- 3. Blower malfunction (common)
- 4. Qvent sensor malfunction (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nebulizer valve test fails
Nebulizer valve pneumatic test does not pass with NOT OK status message
Ranked causes
- 1. High pressure oxygen supply disconnected (common)
- 2. Flow sensor requires calibration (common)
- 3. Qaw measurement abnormal (occasional)
- 4. O2 mixer assembly malfunction (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm lamp remains dark or displays incorrect color patient safety
Alarm lamp does not illuminate or displays wrong color (red, green, blue) during test
Ranked causes
- 1. Alarm lamp board connection fault (common)
- 2. Alarm lamp board defective (common)
- 3. Interaction panel board malfunction (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm monitor tests fail patient safety
One or more alarm monitoring tests do not pass
Ranked causes
- 1. Outdated software version (common)
- 2. Ventilator mainboard failure (common)
- 3. Persistent hardware issue (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm system loudspeaker does not sound patient safety
Loudspeaker does not produce alarm sounds or audible alerts
Ranked causes
- 1. Loudspeaker incorrectly connected (common)
- 2. Loudspeaker defective (common)
- 3. Interaction panel board malfunction (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio Pause key illumination does not toggle
Audio Pause key illumination does not turn on or off
Ranked causes
- 1. Alarm lamp board connection fault (common)
- 2. Alarm lamp board failure (common)
- 3. Interaction panel board malfunction (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Backup alarm lamp not blinking patient safety
Backup alarm lamp does not blink or flash during alarm state
Ranked causes
- 1. Alarm lamp board connection fault (common)
- 2. Alarm lamp board failure (common)
- 3. Interaction panel board malfunction (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery state of health below threshold
Battery SoH is less than 20% or cell voltages are asymmetrical (>100 mV difference), or battery error is ≥5%
Ranked causes
- 1. Battery state of health (SoH) < 20% (common)
- 2. Cell voltages asymmetrical (>100 mV difference) (occasional)
- 3. Battery error ≥ 5% (occasional)
Check state of health (SoH) of the battery
- Test point
- Battery management system
- Expected
- SoH < 20% requires replacement; SoH 20-40% optional replacement
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery state of health degraded patient safety
Battery SoH monitoring and replacement criteria during maintenance
Ranked causes
- 1. State of health (SoH) less than 20% (unknown)
- 2. Cell voltages asymmetrical (difference > 100 mV) (unknown)
- 3. Battery calibration error ≥ 5% (unknown)
Check state of health (SoH) of the battery
- Test point
- Battery monitoring system
- Expected
- ≥ 20% for retention, ≥ 40% for optional replacement consideration
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Binary valve test fails
Binary or autozero valve pneumatic test does not pass
Ranked causes
- 1. Autozero valves incorrectly connected (common)
- 2. FFC cable improperly connected (common)
- 3. Pressure sensor assembly failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Expiratory valve cannot reach certain pressure levels
Several or certain pressure levels are not attainable during expiratory valve calibration
Ranked causes
- 1. Expiratory valve membrane defective (common)
- 2. Expiratory valve cable disconnected (common)
- 3. Expiratory valve hardware failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
HAMILTON-C6 logo illumination does not toggle
HAMILTON-C6 logo illumination does not turn on or off
Ranked causes
- 1. Key panel connection fault (common)
- 2. Key panel defective (common)
- 3. Interaction panel board malfunction (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
HAMILTON-H900 humidifier not communicating with C6 ventilator
HAMILTON-H900 symbol does not appear on bottom left of display after power-on and cable connection
Ranked causes
- 1. AC power cable not connected to dedicated socket on ventilator rear (common)
- 2. Communication cable not connected to COM 3 port (common)
- 3. Ventilator not in Application mode (occasional)
Verify AC power cable is plugged into dedicated socket on ventilator rear
- Test point
- Dedicated power socket on ventilator rear
- Expected
- Cable securely connected
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Insp. valve calibration cannot be performed
Calibration fails to execute or Qvent does not reach target value
Ranked causes
- 1. Insp. valve not properly connected (common)
- 2. Faulty insp. valve (occasional)
- 3. Qvent sensor failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inspiratory valve calibration cannot be performed
Calibration procedure fails to execute on the inspiratory valve circuit
Ranked causes
- 1. Inspiratory valve improperly connected (common)
- 2. Inspiratory valve malfunction (common)
- 3. Qvent sensor failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
IntelliCuff communication not established
IntelliCuff symbol missing from ventilator display or communication error message shown after USB connection
Ranked causes
- 1. USB cable not connected correctly (common)
- 2. IntelliCuff Control software option not installed (occasional)
- 3. Ventilator not in Application mode (occasional)
Verify ventilator is in Application mode
- Test point
- Ventilator display
- Expected
- Application mode active
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
IntelliCuff pressure loss during checks patient safety
Pressure reading drops below expected 60 mbar when measured with hand pump at IntelliCuff port or after cuff connection
Ranked causes
- 1. Incorrect tube connection at IntelliCuff port (common)
- 2. IntelliCuff pressure tube worn or damaged (occasional)
- 3. Check valve malfunction on rear ventilator unit (occasional)
Apply pressure with hand pump to IntelliCuff port
- Test point
- IntelliCuff port on front of ventilator unit
- Expected
- 60 mbar
- Tools
- Hand pump PN 500330, tube 4 mm internal diameter
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loudspeaker sounds distorted or too quiet patient safety
Alarm sounds are distorted, muffled, or insufficient volume
Ranked causes
- 1. Loudspeaker improperly connected (common)
- 2. Loudspeaker defective (common)
- 3. Interaction panel board failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 input test fails
Oxygen input pneumatic test does not pass with NOT OK status message
Ranked causes
- 1. High pressure oxygen supply disconnected (common)
- 2. O2 proportional valve cable connection fault (common)
- 3. O2 mixer assembly malfunction (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 sensor value out of range
Displayed oxygen sensor value is outside acceptable range
Ranked causes
- 1. O2 sensor requires recalibration (common)
- 2. O2 sensor malfunction (common)
- 3. O2 mixer assembly connection fault (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure system values out of tolerance
Pressure measurement values are outside acceptable tolerance during system test
Ranked causes
- 1. System pneumatic leak (common)
- 2. Expiratory valve malfunction (common)
- 3. Pressure sensor malfunction (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Qvent does not reach target value
Flow measurement does not achieve expected calibration target during expiration monitoring
Ranked causes
- 1. Inspiratory valve connection fault (common)
- 2. Inspiratory valve defective (common)
- 3. Qvent sensor malfunction (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rear and/or bottom fan test fails
Rear or bottom ventilation fan does not pass diagnostic test
Ranked causes
- 1. Fan connection fault (common)
- 2. Fan malfunction (common)
- 3. Ventilator mainboard failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
RS-232 communication interface not responding
Patient monitoring system cannot communicate with ventilator via RS-232 interface; no data signal detected
Ranked causes
- 1. RS-232 interface not activated (common)
- 2. Communication port configured to wrong protocol (common)
- 3. Defective RS-232 cable or connector (occasional)
Verify RS-232 interface is activated
- Test point
- Device configuration menu
- Expected
- RS-232 interface enabled status
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
S logo illumination does not toggle
S logo illumination does not turn on or off
Ranked causes
- 1. Key panel incorrectly connected (common)
- 2. Key panel defective (common)
- 3. Interaction panel board malfunction (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 measurement not functional patient safety
SpO2 sensor communication board not responding or sensor reading not available in system display.
Ranked causes
- 1. SpO2 communication board not installed (common)
- 2. SpO2 sensor not activated in hardware options (common)
- 3. Incorrect sensor adapter type selected (occasional)
Verify SpO2 communication board is installed
- Test point
- System Info tab
- Expected
- SpO2 communication board information displayed
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tightness system test fails
Pneumatic tightness or leak test does not pass
Ranked causes
- 1. System pneumatic leak (common)
- 2. Expiratory valve membrane degraded (common)
- 3. Pneumatic component defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarming system test shows incorrect priorities patient safety
Alarm priority sequence or hierarchy is incorrect during system test
Ranked causes
- 1. Outdated software version (common)
- 3. Persistent priority issue (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ambient valve test fails
Ambient valve pneumatic test does not pass with NOT OK status message
Ranked causes
- 1. Ambient valve improperly installed (common)
- 2. Ambient valve malfunction (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery calibration required message displayed patient safety
Device displays 'Battery calibration required' message on screen
Ranked causes
- 1. Battery capacity reduced by more than 25% (common)
- 2. Battery requires calibration cycle (occasional)
Verify battery capacity via Service Software
- Test point
- Service Software display (SW ≥ 1.0.0)
- Expected
- Full charge capacity (fc.) value shown; capacity should be >= 80% of designed capacity (4000 mAh for HAMILTON-C6), >= 80% of design capacity
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bottom fan airflow insufficient caution
Paper does not stick to bottom fan inlet; airflow not being drawn into ventilator correctly
Ranked causes
- 1. Bottom fan motor failure or blockage (occasional)
- 2. Fan motor not powered (occasional)
Visually inspect bottom fan inlet for blockage
- Test point
- Bottom fan inlet
- Expected
- Clear of obstructions, no dust accumulation
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Buzzer does not sound during power loss patient safety
Alarm buzzer fails to activate or sounds for less than two minutes when power is lost
Ranked causes
- 1. Super capacitors (gold caps) not fully charged (common)
- 2. Mainboard failure or defective gold capacitors (rare)
Verify device has been powered for at least 20 minutes
- Test point
- Device operation time
- Expected
- 20 minutes minimum operating time in standby or service mode
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device does not switch to battery power patient safety
Ventilator does not transition to battery operation when mains power is disconnected
Ranked causes
- 1. Battery not installed or discharged (common)
- 2. Battery charging circuit failure (occasional)
Verify internal batteries are installed
- Test point
- Battery compartment
- Expected
- Batteries present and seated
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Expiratory valve heater element not warming caution
Expiratory valve internal surface does not heat up during operation; heater element not functioning
Ranked causes
- 1. Heater element defective (occasional)
- 2. Heater not powered during HiFlow mode (common)
Verify device is powered from AC mains
- Test point
- AC power inlet
- Expected
- Device operating on AC power
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nurse call relay not responding patient safety
Nurse call alarm relays do not change state during alarm activation or ventilation phases; interface not functional
Ranked causes
- 1. Nurse call connector not properly seated (common)
- 2. Relay contacts defective or stuck (occasional)
Verify nurse call connector is installed
- Test point
- Communication board connector
- Expected
- PN 160166 connector fully seated
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 mixer system values out of tolerance
Oxygen mixer measurement values are outside acceptable tolerance during system test
Ranked causes
- 1. High pressure oxygen supply disconnected (common)
- 2. Qvent sensor malfunction (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 sensor calibration fails
Gain calibration cannot be completed for oxygen sensor
Ranked causes
- 1. O2 sensor cable disconnected or damaged (common)
- 2. Faulty O2 sensor (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 sensor gain calibration fails
Oxygen sensor gain calibration procedure does not complete or fails validation
Ranked causes
- 1. O2 sensor cable disconnected (common)
- 2. O2 sensor malfunction (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Proximal flow test fails
Proximal flow measurement test does not pass with NOT OK status message
Ranked causes
- 1. Qvent sensor malfunction (common)
- 2. Pressure sensor assembly failure (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Proximal pressure test fails
Proximal pressure measurement test does not pass with NOT OK status message
Ranked causes
- 1. Autozero valve malfunction (common)
- 2. Pressure sensor assembly failure (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Proximal test flow through sensors not equal
Flow through both flow sensor connectors are not equal during rinse flow test
Ranked causes
- 1. Pressure sensor assembly malfunction (common)
- 2. Rinse flow assembly malfunction (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rear fan airflow insufficient caution
Paper does not stick to rear fan inlet; airflow not being drawn into ventilator correctly
Ranked causes
- 1. Rear fan motor failure or blockage (occasional)
- 2. Fan motor not powered (occasional)
Visually inspect rear fan inlet for blockage
- Test point
- Rear fan inlet
- Expected
- Clear of obstructions, no dust accumulation
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Status indicator LEDs not appearing as required
One or more LEDs on the status indicator board do not display
Ranked causes
- 1. Status indicator board connection fault (common)
- 2. Ventilator mainboard failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm loudness not increasing
Alarm volume does not increase when loudness setting is adjusted to maximum
Ranked causes
- 1. Alarm system malfunction (unknown)
Verify device is connected to mains AC power
- Test point
- AC power inlet
- Expected
- Device powered from mains
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ambient test fails
Ambient sensor or environmental test does not pass
Ranked causes
- 1. Ventilator mainboard failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Autozero test fails
Autozero valve pneumatic test does not pass with NOT OK status
Ranked causes
- 1. Pressure sensor assembly failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery capacity degradation at end-of-life patient safety
Reduced battery capacity leads to unreliable fuel gauge information and risk of unexpected ventilation stop during power failure
Ranked causes
- 1. Battery end-of-life degradation (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery capacity reduced below threshold patient safety
Device displays 'Battery calibration required' message; capacity loss exceeds 25% of designed capacity
Ranked causes
- 1. Battery aged past 80% capacity threshold (common)
Check battery capacity in Service Software
- Test point
- Service Software display
- Expected
- Capacity at or above 4000 mAh (80% of 5000 mAh design capacity), ≥4000 mAh
- Tools
- Service Software SW ≥ 1.0.0
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery cell voltages asymmetrical patient safety
Cell voltage difference exceeds 100 mV, indicating unbalanced or failing battery cells.
Ranked causes
- 1. Battery cells unbalanced or failing (occasional)
Read cell voltages from battery data
- Test point
- Battery data via service software
- Expected
- Cell voltage difference ≤ 100 mV, ±100 mV
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery condition maintenance check
Routine assessment of battery health and proper storage conditions during preventive maintenance
Ranked causes
- 1. Annual maintenance cycle requirement (common)
Verify battery manufacture date and capacity history
- Test point
- Battery connector (top) and Service Software
- Expected
- Manufacture date visible; capacity trend stable or improving; cycle count documented
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Battery error calibration required
- Battery error reading exceeds tolerance
- Battery state of health low
- Binary valve test shows short circuit
- Calibration: Battery charger maintenance cycle
- Calibration: Battery fuel gauge
- Excess pressure in breathing circuit during inspiration
- Exhalation obstructed alarm triggered
- Front panel keys or LEDs do not function
- Knob range does not function
- QO2 value out of tolerance
- Reset calibration to default settings
- Weak batteries causing incorrect fuel gauge or unexpected ventilation stop
- Weak batteries detected
14 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
132001 | Low airway pressure | Airway pressure falls below set threshold | warning |
132002 | Check flow sensor tubing | Flow sensor tubing issue detected | warning |
132003 | Check flow sensor | Flow sensor malfunction detected | warning |
132004 | Check flow sensor | Flow sensor missing or not detected | error |
132005 | O2 sensor defective | O2 sensor malfunction detected | error |
132006 | O2 sensor missing | O2 sensor not detected | error |
132007 | O2 sensor not system compatible | Incompatible O2 sensor detected | error |
132008 | O2 sensor calibration needed | O2 sensor requires calibration | warning |
132010 | CO2 sensor disconnected | CO2 sensor connection lost | error |
132011 | CO2 sensor defect | CO2 sensor malfunction detected | error |
132012 | CO2 sensor over temperature | CO2 sensor temperature exceeds limit | error |
132013 | CO2 sensor warmup | CO2 sensor warming up after power-on or reset | info |
132014 | Wrong flow sensor | Incompatible or incorrect flow sensor detected | error |
132015 | Flow sensor calibration needed | Flow sensor requires calibration | warning |
132016 | Check CO2 airway adapter | CO2 airway adapter issue detected | warning |
132017 | Check CO2 sampling line | CO2 sampling line problem detected | warning |
132018 | SpO2: patient disconnected | SpO2 sensor detects patient disconnection | warning |
132019 | SpO2: probe missing | SpO2 probe not detected | error |
132020 | SpO2: low perfusion index | Blood perfusion at sensor site too low | warning |
132021 | SpO2: light interference | Ambient light interferes with SpO2 reading | warning |
132022 | SpO2: sensor error | SpO2 sensor malfunction detected | error |
132023 | SpO2: adapter missing | SpO2 adapter not detected or disconnected | error |
132024 | Circuit calibration needed | Patient circuit requires calibration | warning |
132025 | SpO2: poor signal | SpO2 signal quality insufficient | warning |
132026 | Check flow sensor tubing | Flow sensor tubing disconnected or problem detected | warning |
132027 | Check flow sensor tubing | Pressure differential sensor tubing disconnected | warning |
132028 | Flow sensor calibration needed | Flow sensor requires low-priority calibration | warning |
133001 | Apnea detected | Patient apnea detected (no breathing) | critical |
133002 | Inspiratory volume limitation | Tidal volume exceeds set limit | warning |
133003 | High pressure | Airway pressure exceeds alarm threshold | error |
133004 | Check patient interface | Problem with patient interface detected | warning |
141001 | Pressure limitation | Ventilation reaches pressure limit during inspiration | warning |
141002 | Vt high | Tidal volume exceeds high threshold | warning |
141003 | Vt low | Tidal volume falls below low threshold | warning |
141004 | High minute volume | Total minute ventilation exceeds high limit | warning |
141005 | Low minute volume | Total minute ventilation falls below low limit | warning |
141006 | High frequency | Respiratory frequency exceeds high limit | warning |
141007 | Low frequency | Respiratory frequency falls below low limit | warning |
141008 | High oxygen | O2 concentration exceeds high limit | warning |
141009 | Low oxygen | O2 concentration falls below low limit | error |
141010 | Apnea | Patient apnea detected (no breathing) | critical |
141011 | Apnea ventilation ended | Apnea ventilation backup support ended | info |
141012 | Apnea ventilation | Backup ventilation activated due to apnea | warning |
141013 | High pressure during sigh | Airway pressure exceeds limit during sigh breath | warning |
141014 | Turn the flow sensor | Flow sensor orientation incorrect | warning |
141016 | Disconnection on ventilator side | Ventilator detects disconnection at ventilator-side circuit | error |
141017 | Disconnection on patient side | Ventilator detects disconnection at patient interface | error |
141018 | Exhalation obstructed | Exhalation pathway blocked or obstructed | error |
141019 | IRV | Inspiratory reserve volume alarm activated | unknown |
141020 | ASV: Cannot meet target | Adaptive Support Ventilation unable to achieve target | warning |
141021 | Pressure limit has changed | Pressure limit parameter modified during ASV operation | info |
141023 | External flow sensor failed | External flow sensor malfunction detected | error |
141024 | Low pressure | Airway pressure falls below low threshold | error |
141025 | PetCO2 high | End-tidal CO2 exceeds high limit | warning |
141026 | PetCO2 low | End-tidal CO2 falls below low limit | warning |
141027 | Suctioning maneuver | Suctioning maneuver activated or in progress | info |
141028 | Performance limited by high altitude | Ventilator performance affected by high altitude | warning |
141029 | High SpO2 | Oxygen saturation exceeds high limit | warning |
141030 | Low SpO2 | Oxygen saturation falls below low limit (medium priority) | warning |
141031 | High pulse | Heart rate exceeds high limit | warning |
141032 | Low pulse | Heart rate falls below low limit | warning |
141033 | High PI | Perfusion index exceeds high limit | warning |
141034 | Low PI | Perfusion index falls below low limit | warning |
141035 | High SpHb | Hemoglobin saturation exceeds high limit | warning |
141036 | Low SpHb | Hemoglobin saturation falls below low limit | warning |
141037 | High SpCO | Carboxyhemoglobin exceeds high limit | error |
141038 | Low SpCO | Carboxyhemoglobin falls below low limit | info |
141039 | High SpMet | Methemoglobin exceeds high limit | warning |
141040 | Low SpMet | Methemoglobin falls below low limit | info |
141041 | Low SpO2 | Oxygen saturation falls below low limit (high priority) | critical |
141043 | Ventilation adjustment OFF | Ventilation adjustment function disabled | info |
141044 | Oxygenation adjustment OFF | Oxygenation adjustment function disabled | info |
141045 | Ventilation Controller at limit | Ventilation controller reached upper limit | warning |
141046 | Oxygenation Controller at limit | Oxygenation controller reached upper limit | warning |
141047 | High flow | Flow rate exceeds high limit in nCPAP mode | warning |
141048 | Obstruction | Circuit obstruction detected | error |
141049 | High flow | Flow rate exceeds high limit | warning |
141050 | Sensor simulation active | Sensor simulation mode is active during operation | critical |
141051 | High PEEP | Positive end-expiratory pressure exceeds high limit | warning |
141052 | Loss of PEEP | PEEP falls below set threshold or is lost | error |
141053 | FiO2 set to 100% due to low SpO2 | Oxygen concentration automatically set to maximum due to hypoxia | critical |
141054 | Recruitment in progress | Recruitment maneuver running | info |
141055 | Oxygen control limit exceeded | Oxygen concentration exceeded configured limit | warning |
141056 | SBT aborted | Spontaneous breathing trial did not complete | warning |
141057 | SBT successfully fulfilled | Spontaneous breathing trial completed successfully | info |
141058 | High PVI | Pleth Variability Index above threshold | warning |
141059 | Low PVI | Pleth Variability Index below threshold | warning |
141060 | Check flow sensor for water | Water detected in flow sensor | warning |
141061 | Check flow sensor for water (high priority) | Water in flow sensor requiring immediate attention | warning |
141062 | SpeakValve ON | Speaking valve activated | info |
141063 | SpeakValve OFF | Speaking valve deactivated | info |
141064 | Vt low (high priority) | Tidal volume below alarm limit | warning |
141065 | Check patient interface | Patient interface connectivity or configuration error | warning |
141066 | Maximum leak compensation | Leak compensation reaches maximum threshold | warning |
231001 | Technical event: pressure controller pressure low | Inspiratory pressure below normal operating range | error |
231002 | Technical event: pressure controller pressure high | Inspiratory pressure above normal operating range | error |
231003 | Technical event: flow controller flow low | Measured flow below expected range | error |
231004 | Technical event: flow controller flow high | Measured flow exceeds normal range | error |
231005 | Technical event: inspiration valve leak | Inspiratory valve not sealing properly | error |
231006 | Oxygen supply failed | Oxygen flow from controller too low | critical |
231007 | Technical event: O2 controller flow high | Oxygen flow exceeds normal range | error |
231008 | Technical event: O2 valve leak | Oxygen valve not sealing | error |
231009 | Technical event: blower controller speed low | Blower rotation speed below normal | error |
231010 | Technical event: blower controller speed high | Blower rotation speed exceeds normal range | error |
231011 | Vent outlet temperature high | Exhaust temperature exceeds safe limit | warning |
231012 | Technical event: qvent flow sensor defect | Main flow sensor malfunction detected | error |
231013 | Technical event: qO2 flow sensor defect | Oxygen flow sensor malfunction | error |
231014 | Technical event: ambient valve error | Ambient pressure relief valve malfunction | error |
231017 | Blower service required | Blower maintenance interval reached or motor degradation detected | warning |
231018 | Technical event: self-test with error | Internal self-test detected abnormality | error |
231019 | Technical event: inspiration valve sensor defect | Inspiratory valve position sensor failure | error |
231020 | Technical event: regulator data log on | Regulator data logging active for diagnostics | info |
231022 | Technical event: pExp valve sensor defect | Expiratory pressure valve sensor malfunction | error |
231023 | Technical event: invalid flow sensor | Flow sensor type not recognized or unsupported | error |
231027 | Wrong expiratory valve | Expiratory valve type does not match ventilator configuration | error |
231032 | Technical event: expiration valve disconnected | Expiratory valve not properly connected | error |
231033 | Technical event: blower controller speed limited | Blower speed reduced below normal operating range | warning |
231036 | Technical event: pvent pressure sensor defect | Inspiratory pressure sensor malfunction | error |
232002 | Technical event: pvent monitor sensor defect | Pressure monitoring sensor malfunction | error |
232003 | Technical event: Paw sensor defect | Airway pressure sensor malfunction | error |
232004 | Pressure not released | System unable to vent pressurized circuit | warning |
232005 | Technical event: blower hot | Blower motor temperature elevated | warning |
232006 | Technical event: blower temperature sensor defect | Temperature sensor malfunction on blower | error |
232007 | Check flow sensor tubing | Airway flow sensor tubing obstruction or disconnection | warning |
232008 | Technical event: pambient sensor defect | Ambient pressure sensor malfunction | error |
232028 | Technical event: pFilter pressure high | Pressure drop across inlet filter exceeds limit | error |
232029 | Technical event: tinst sensor defect | Internal temperature sensor malfunction | error |
232030 | Technical event: clock error | Internal real-time clock malfunction | error |
232034 | Replace O2 sensor | Oxygen cell depleted or defective | warning |
232035 | Technical event: pfilter sensor defect | Filter pressure sensor malfunction | error |
232038 | CO2 calibration needed | Carbon dioxide sensor requires calibration | warning |
232045 | Technical event: SpO2 sensor settings error | Pulse oximetry sensor configuration error | error |
232056 | Technical event: pAmbient pFilter mismatch | Ambient and filter pressure sensors disagree | error |
232057 | Technical event: SpO2 sensor change failed | Pulse oximetry sensor replacement not recognized | error |
233008 | Aerogen nebulizer disconnected | Aerogen nebulizer becomes disconnected from ventilator | warning |
234001 | Loudspeaker defective | Loudspeaker failure detected during operation or startup | warning |
241001 | Replace HEPA filter | HEPA filter has reached end of service life | warning |
241002 | Release valve defective | Tightness test detects release valve failure | error |
243004 | Buzzer defective | Buzzer failure detected at startup | warning |
243007 | Realtime clock failure | Real-time clock (RTC) has been reset or failed | error |
243010 | Panel connection lost | Communication lost with control panel | error |
244001 | Loss of external power | Mains power supply disconnected or failed | critical |
244002 | Battery 1: Calibration required | Battery 1 requires calibration | warning |
244003 | Battery 2: Calibration required | Battery 2 requires calibration | warning |
244004 | Battery 1: Temperature high | Battery 1 temperature exceeds safe operating range | warning |
244005 | Battery 2: Temperature high | Battery 2 temperature exceeds safe operating range | warning |
244006 | Battery low | Battery charge level falls below high priority threshold | warning |
244007 | Battery power loss | Battery power supply has failed completely | critical |
244008 | Battery 1: Wrong battery | Incompatible or incorrect battery installed in slot 1 | error |
244009 | Battery 2: Wrong battery | Incompatible or incorrect battery installed in slot 2 | error |
244010 | Battery low | Battery charge level falls below low priority threshold | warning |
244011 | Battery communication error | Battery management bus communication failure | error |
244012 | Battery 1: Defective | Battery 1 hardware failure detected | error |
244013 | Battery 2: Defective | Battery 2 hardware failure detected | error |
244015 | Battery low | Battery charge level falls below medium priority threshold | warning |
244016 | Battery 1: Replacement required | Battery 1 has reached end of service life | warning |
244017 | Battery 2: Replacement required | Battery 2 has reached end of service life | warning |
244019 | Technical event: 28V_IP voltage out of tolerance | 28V auxiliary voltage outside specified limits | error |
244022 | Continue charging battery | Battery charging in progress; normal operation | info |
246001 | Preventive maintenance required | Service interval reached | warning |
246003 | Fan failure | Cooling fan hardware failure detected | error |
246016 | Touch not functional | Touch screen or touch panel failed during POST | error |
246020 | Loudspeaker defective | HMI loudspeaker hardware failure | warning |
246040 | Ventilator overpressure self-test failed | Self-test execution during startup or diagnostics | error |
246041 | Alarm LED error | LED circuit failure or lamp defect | error |
249001 | Options not found | Configuration file missing or unreadable on startup | error |
249002 | Option file write error | Failed to write option file to storage | error |
249003 | Option file version mismatch | Option file version incompatible with current firmware | error |
249004 | Default option file generated | Option file missing or unreadable, defaults applied | info |
249010 | Device configuration file error | Device configuration file read/write failure | error |
249011 | Setup configuration file error | Setup configuration file read/write failure | error |
249012 | Last setting configuration file error | Last settings file read/write failure | error |
249013 | Research configuration file error | Research configuration file read/write failure | error |
249020 | Invalid communication board | Unrecognized or incompatible option board detected | error |
249021 | Check hardware compatibility | Installed hardware incompatible with firmware or configuration | warning |
250001 | Check humidifier communication | Humidifier interface communication failure | error |
250002 | Humidifier water low | Water level below minimum threshold | warning |
250003 | Humidifier chamber temperature low | Chamber temperature below set point | warning |
250004 | Humidifier Y-piece temperature low | Y-piece temperature below set point | warning |
250005 | Humidifier check right tube | Right tube temperature or connection issue detected | warning |
250006 | Humidifier check left tube | Left tube temperature or connection issue detected | warning |
250007 | Humidifier check chamber | Chamber malfunction or connection error detected | warning |
250008 | Check humidifier | General humidifier malfunction detected | warning |
250009 | Humidifier error | Humidifier subsystem critical error | error |
250010 | Humidifier water high | Water level above maximum threshold | warning |
250011 | Humidifier chamber temperature high | Chamber temperature above set point | warning |
250012 | Humidifier Y-piece temperature high | Y-piece temperature above set point | warning |
250013 | Humidifier tilt | Humidifier orientation/tilt detected abnormal | critical |
250014 | Check humidifier | Humidifier malfunction or high-level fault detected | error |
251001 | Check IntelliCuff communication | IntelliCuff interface communication failure | error |
251002 | Check IntelliCuff | IntelliCuff battery low or malfunction | warning |
251003 | Cannot turn off IntelliCuff | IntelliCuff power-off command failed | error |
251004 | Check IntelliCuff | IntelliCuff battery level critically low | warning |
251005 | Cuff pressure high | Measured cuff pressure exceeds safe limit | critical |
251006 | Cuff deflated | Cuff remained deflated for extended period | warning |
251007 | Cuff leak | Cuff pressure drop indicating air leak | error |
251008 | Check IntelliCuff | IntelliCuff malfunction at medium severity level | warning |
251009 | Cuff leak | IntelliCuff disconnection or interface failure detected | error |
251010 | Check IntelliCuff | IntelliCuff critical error condition | error |
251011 | Check IntelliCuff | IntelliCuff battery at critical level | critical |
251012 | Check IntelliCuff | IntelliCuff malfunction at high severity level | error |
251013 | Check IntelliCuff state | IntelliCuff state invalid or unexpected | warning |
255001 | Percentage minimum volume high limit test failed | Self-test of min volume limit failed during startup | error |
255002 | Percentage minimum volume high limit wrong | Configured min volume limit exceeds acceptable range | warning |
255003 | Percentage minimum volume low limit test failed | Self-test of min volume limit failed during startup | error |
255004 | Percentage minimum volume low limit wrong | Configured min volume limit below acceptable range | warning |
255005 | PEEP high limit wrong | Configured PEEP high limit exceeds acceptable range | warning |
255006 | PEEP high limit test failed | Self-test of PEEP high limit failed | error |
255007 | PEEP low limit wrong | Configured PEEP low limit below acceptable range | warning |
255008 | PEEP low limit test failed | Self-test of PEEP low limit failed | error |
255009 | Oxygen low test failed | Self-test of oxygen system failed | error |
255010 | Oxygen high test failed | Self-test of oxygen system upper limit failed | error |
255011 | Percentage minimum volume positive change test failed | Self-test of min volume increase failed | error |
255012 | Percentage minimum volume negative change test failed | Self-test of min volume decrease failed | error |
255013 | PEEP positive change test failed | Self-test of PEEP increase failed | error |
255014 | PEEP negative change test failed | Self-test of PEEP decrease failed | error |
255015 | Oxygen positive change test failed | Self-test of oxygen increase failed | error |
255016 | Oxygen negative change test failed | Self-test of oxygen decrease failed | error |
255017 | Recruitment running time exceeded | Recruitment maneuver running time exceeded limit | warning |
255018 | Simulation supervise test failed | Internal simulation/test supervision failed | error |
255019 | Percentage minimum volume high fine adjustment test failed | Self-test of min volume fine limit failed | error |
255020 | Percentage minimum volume high fine limit wrong | Configured min volume fine limit exceeds range | warning |
255021 | Percentage minimum volume low fine adjustment test failed | Self-test of min volume fine limit failed | error |
255022 | Percentage minimum volume low fine limit wrong | Configured min volume fine limit below range | warning |
255023 | Oscillation oxygen | Oxygen parameter oscillating during operation | warning |
255024 | Oscillation PEEP/CPAP | PEEP/CPAP pressure oscillating during operation | warning |
255025 | Oscillation percentage minimum volume | Minimum volume parameter oscillating during operation | warning |
255026 | Data mismatch | Internal data consistency error detected | error |
255027 | PetCO2 frozen test failed | PetCO2 measurement frozen or not updating | error |
255028 | SpO2 frozen test failed | SpO2 measurement frozen or not updating | error |
255029 | Spontaneous frequency frozen test failed | Spontaneous breathing rate frozen or not updating | error |
283001 | Startup failed | Display GUI startup failed during boot | error |
283003 | Language not loaded | Language file failed to load on startup | error |
283005 | Settings file error | Display settings file read/write failure | error |
283008 | Check settings | Invalid or suspicious settings detected | warning |
283009 | Self test failed | Display self-test failed during startup | error |
283010 | No ventilation after power-fail | Ventilation did not resume after power failure recovery | critical |
283011 | Check IntelliCuff state | IntelliCuff state invalid on display startup | error |
283012 | Panel settings file error | Panel hardware settings file read/write failure | error |
284002 | Technical event alarm service high | Service alarm high priority triggered | critical |
284003 | Technical event alarm service medium | Service alarm medium priority triggered | warning |
284004 | Technical event alarm service low | Service alarm low priority triggered | info |
284005 | Flow sensor calibration needed | Flow sensor calibration overdue or required | warning |
284006 | Restart device to enable tests/calibration | Device restart required to access test/calibration mode | info |
285001 | Alarm lamp error defect | Error alarm lamp hardware failure detected | error |
285002 | Alarm lamp warning defect | Warning alarm lamp hardware failure detected | error |
285003 | Backlight defect | Display backlight hardware failure detected | error |
285004 | Function key not operational | Hardware key/button malfunction detected | error |
285005 | Touch not functional | Touch screen interface malfunction detected | error |
291001 | Remote communication error (COM1) | COM1 port communication error detected | error |
291002 | Remote communication error (COM2) | COM2 port communication error detected | error |
291003 | Remote communication timeout (COM1) | COM1 communication timeout no response received | error |
291004 | Remote communication timeout (COM2) | COM2 communication timeout no response received | error |
296001 | Remote control active | Remote control mode is currently active | info |
332001 | Flow sensor error | Safety mode technical failure detected | critical |
341001 | Breath settings not accepted | Settings rejected during mode configuration | error |
341002 | Returned breath settings incorrect | Mismatch between sent and confirmed settings | error |
341003 | Adaptive settings invalid | Adaptive control settings validation failure | warning |
341004 | Breath settings timeout | Settings not confirmed within timeout window | error |
341005 | Controller settings invalid | Controller parameter validation failed | error |
341006 | Returned alarm limits incorrect | Alarm limit confirmation mismatch | error |
341007 | Mode transition not defined | Requested ventilation mode transition unsupported | warning |
341008 | Ventilation flow sensor defect | Flow sensor failure detected | critical |
341009 | Ventilation pressure sensor defect | Pressure sensor failure detected | critical |
343001 | Communication timeout | Alarm module communication lost | error |
346002 | Watchdog failed: ALR | Low level supervisor watchdog timeout on alarm module | critical |
346003 | Watchdog failed: ALR measurement | Alarm module measurement watchdog timeout | critical |
346004 | Watchdog failed: ESL | Low level supervisor watchdog timeout on ESL | critical |
346005 | Watchdog failed: ESL MMI log | Watchdog timeout in MMI logging task | warning |
346006 | Watchdog failed: ESL control log | Watchdog timeout in control logging task | error |
346007 | Watchdog failed: ESL breath log | Watchdog timeout in breath monitoring log | error |
346008 | Watchdog failed: ESL event SDR reader | Watchdog timeout in event storage device reader | error |
346009 | Watchdog failed: ESL service SDR reader | Watchdog timeout in service storage device reader | warning |
346010 | Watchdog failed: ESL event SDR writer | Watchdog timeout in event storage device writer | error |
346011 | Watchdog failed: ESL service SDR writer | Watchdog timeout in service storage device writer | warning |
346012 | Watchdog failed: ESL measurement | Watchdog timeout in ESL measurement processing | critical |
346013 | Watchdog failed: GUIL touch | Watchdog timeout in touch screen interface | error |
346014 | Watchdog failed: GUIL PT knob | Watchdog timeout in pressure-time knob interface | error |
346015 | Watchdog failed: GUIL hardkeys | Watchdog timeout in hardkey input processing | error |
346016 | Watchdog failed: GUIL screenshot | Watchdog timeout in screenshot capture function | warning |
346017 | Watchdog failed: LM | Low level supervisor watchdog timeout on language module | error |
346019 | Watchdog failed: LLS hardware exception | Hardware exception in low level supervisor | critical |
346020 | Watchdog failed: LLS common server | Watchdog timeout in common service server | error |
346022 | Watchdog failed: LLS alarm status | Watchdog timeout in alarm status monitoring | error |
346023 | Watchdog failed: LLS measurement | Watchdog timeout in LLS measurement processing | critical |
346024 | Watchdog failed: PM | Watchdog timeout on power management module | critical |
346025 | Watchdog failed: PM measurement | Watchdog timeout in PM measurement function | error |
346026 | Watchdog failed: QSPI | Watchdog timeout on QSPI interface | critical |
346027 | Watchdog failed: QSPI MMI server | Watchdog timeout in QSPI MMI server | error |
346028 | Watchdog failed: QSPI BM server | Watchdog timeout in QSPI blower/motor server | error |
346029 | Watchdog failed: QSPI BPG server | Watchdog timeout in QSPI breath pattern generator server | error |
346030 | Watchdog failed: QSPI LLS server | Watchdog timeout in QSPI LLS server | critical |
346031 | Watchdog failed: QSPI interface | Watchdog timeout on QSPI serial interface | critical |
346032 | Watchdog failed: QSPI measurement | Watchdog timeout in QSPI measurement | error |
346033 | Watchdog failed: RTC real time clock | Watchdog timeout in real time clock | error |
346034 | Watchdog failed: RTC alarm clock | Watchdog timeout in alarm clock function | warning |
346035 | Watchdog failed: SC | Watchdog timeout on service controller | error |
346036 | Watchdog failed: SGUI | Watchdog timeout on service GUI | warning |
346037 | Watchdog failed: STU | Watchdog timeout on STU module | error |
346038 | Watchdog failed: VMC | Watchdog timeout on ventilation mode controller | critical |
346039 | Watchdog failed: VMC ventilation alarming | Watchdog timeout in VMC alarm task | error |
346040 | Watchdog failed: VMC ventilation control | Watchdog timeout in VMC control task | critical |
346041 | Watchdog failed: VMC ventilation monitoring | Watchdog timeout in VMC monitoring task | critical |
346042 | Watchdog failed: VMC measurement | Watchdog timeout in VMC measurement | error |
346043 | Watchdog failed: VGUI | Watchdog timeout on ventilation GUI | error |
346044 | Watchdog failed: VGUI mode control | Watchdog timeout in VGUI mode control task | error |
346045 | Watchdog failed: VGUI measurement | Watchdog timeout in VGUI measurement | error |
346046 | Watchdog failed: LLS EEPROM write | Watchdog timeout in EEPROM write operation | error |
346047 | Watchdog failed: BM alarming slow | Watchdog timeout in blower motor alarming | warning |
346048 | Watchdog failed: SND sound control | Watchdog timeout in sound control module | warning |
346049 | Watchdog failed: SND sound | Watchdog timeout in sound generation | warning |
346050 | Watchdog failed: VT | Watchdog timeout on ventilation timing module | critical |
346051 | Watchdog failed: CFG configuration | Watchdog timeout in configuration module | error |
346052 | Watchdog failed: EXM communication base | Watchdog timeout in external communication base | error |
346053 | Watchdog failed: IVS | Watchdog timeout on IVS module | error |
346054 | Safety failure detected | General safety system failure detected | critical |
346055 | Watchdog failed: PD black box logging | Watchdog timeout in black box data logging | warning |
346056 | Watchdog failed: EXM COM1 | Watchdog timeout on EXM communication port 1 | error |
346057 | Watchdog failed: EXM COM2 | Watchdog timeout on EXM communication port 2 | error |
383001 | Settings not accepted | Ventilation GUI settings rejected | warning |
383002 | Returned settings incorrect | Confirmation of GUI settings does not match sent values | error |
383003 | Settings validation error | GUI settings fail validation check | warning |
383004 | Monitoring channel observation failed | GUI monitoring channel lost or unresponsive | error |
383005 | VMC timeout | Communication timeout between GUI and VMC | error |
383006 | Returned device settings incorrect | Device settings confirmation mismatch | error |
383007 | Trending channel observation failed | GUI trending data channel lost | warning |
383008 | Startup basic startup failure | VGUI basic startup sequence failed | critical |
383009 | Startup screen failure | VGUI screen initialization failed | critical |
383010 | Startup load GUI basics failure | VGUI fails to load basic GUI resources | critical |
383011 | Startup selftest prolog failure | VGUI startup self-test initialization failed | error |
383012 | Startup selftest failure | VGUI startup self-test detected hardware fault | critical |
383013 | Startup selftest epilog failure | VGUI self-test cleanup phase failed | error |
383014 | Startup loading bitmaps failure | VGUI fails to load bitmap resources | error |
383015 | Startup initialization failure | VGUI initialization phase failed | critical |
384001 | Startup basic startup failure | SGUI basic startup sequence failed | critical |
384002 | Startup screen failure | SGUI screen initialization failed | critical |
384003 | Startup load GUI basics failure | SGUI fails to load basic GUI resources | critical |
384004 | Startup selftest prolog failure | SGUI startup self-test initialization failed | error |
384005 | Startup selftest failure | SGUI startup self-test detected hardware fault | critical |
384006 | Startup loading bitmaps failure | SGUI fails to load bitmap resources | error |
384007 | Startup initialization failure | SGUI initialization phase failed | critical |
385001 | Alarming channel observation failed | AGL alarming channel disconnected or lost | error |
385002 | Safety failure detected | AGL detected safety system fault | critical |
385003 | PM channel observation failed | AGL PM monitoring channel lost | error |
386001 | Bitmap not loaded | GUI layer fails to load bitmap resource | error |
431001 | Blower fault | Blower motor malfunction in ambient mode | critical |
431002 | Blower disconnected | Blower is not connected or powered | critical |
431004 | Inspiration valve overcurrent | Inspiration valve motor draws excessive current | error |
431006 | Expiration valve overcurrent | Expiration valve motor draws excessive current | error |
431007 | Vent output over temperature | Ventilation output exceeds safe temperature | critical |
431008 | Q-vent flow sensor error | Ventilation flow sensor malfunction | critical |
431009 | Q-O2 flow sensor error | Oxygen flow sensor malfunction | error |
431010 | Control REGSPI timeout | SPI communication timeout to control register | error |
431011 | Monitor REGSPI timeout | SPI communication timeout to monitor register | error |
431012 | Flow sensor measurement imprecise | Flow sensor reading variance exceeds tolerance | warning |
431013 | Calibration read failed | Stored calibration data cannot be retrieved | error |
431014 | Inspiration valve sensor defect | Inspiration valve position sensor malfunction | error |
431015 | Emergency off failed | Emergency shutdown did not execute successfully | critical |
431017 | Inspiration valve disconnected | Inspiration valve is not properly connected | critical |
432001 | Blower over temperature | Blower assembly exceeds safe operating temperature | critical |
432002 | Instrument over temperature | Ventilation instrument reaches excessive temperature | critical |
432003 | Pressure not released ambient | Pressure relief failed in ambient mode | error |
433001 | Breath monitoring tick timeout | Breath monitoring timing missed in BPG | error |
433002 | Control tick timing error | Control loop tick timing error in BPG | error |
443001 | Watchdog failed LLS | Ambient mode low level supervisor watchdog timeout | critical |
444001 | Battery totally discharged | Battery voltage depleted below minimum threshold | critical |
444004 | Voltage out of tolerance | Power supply voltage exceeds acceptable range | error |
444005 | Shutdown failed | Controlled shutdown sequence did not complete | error |
446001 | CPU temperature critical | CPU temperature exceeds critical limit | critical |
446002 | LLS safety subsystem failure | Safety monitoring system detects fault condition | critical |
446003 | LLS watchdog failed for gas delivery module | Watchdog timer timeout in gas delivery control | critical |
446004 | Watchdog failure for valve regulator in gas delivery | Valve regulator module timeout or lockup | critical |
446005 | Watchdog failure for blower regulator in gas delivery | Blower regulator module timeout or halted execution | critical |
446006 | Watchdog failure in gas delivery monitoring subsystem | Monitoring watchdog expired due to module unresponsiveness | critical |
446007 | Watchdog failure in valve control logging subsystem | Valve log watchdog timeout during logging operation | error |
446008 | Watchdog failure in blower control logging subsystem | Blower log watchdog timeout or logger module hang | error |
446009 | Watchdog failure in flow measurement module of gas delivery | Flow measurement processor timeout or unresponsive execution | critical |
446010 | GPIO fault input detection failed in gas delivery watchdog | GPIO input monitoring watchdog expired | error |
446011 | Breath management module watchdog failure | BM (Breath Management) watchdog timeout detected | critical |
446012 | Watchdog failure in breath management gas delivery communication | BM-GD communication timeout or message queue failure | critical |
446013 | Hardware control watchdog failure in breath management | Hardware control watchdog expired or unresponsive | critical |
446014 | Breath parameter generator BPG client receiver watchdog failed | BPG client watchdog timeout in breath management | error |
446015 | Breath parameter generator BPG server receiver watchdog failed | BPG server watchdog timeout or server process hang | error |
446016 | Breath data calculation watchdog failure in breath management | Breath calculation algorithm timeout or hang | critical |
446017 | Flow measurement watchdog failure in breath management | BM flow measurement module watchdog timeout | error |
446018 | Breath parameter generator module watchdog failure | BPG module watchdog timer expired | critical |
446019 | BPG gas delivery client receiver watchdog failed | BPG client communication watchdog timeout | error |
446020 | BPG flow measurement watchdog failure | BPG flow measurement subsystem timeout or hang | error |
446021 | Exception occurred in LLS software | Unhandled exception or critical software error detected | critical |
446022 | LLS voltage error detected | Supply voltage monitoring detected out-of-range condition | critical |
446023 | LLS workload watchdog timeout | LLS workload processing exceeded timeout threshold | warning |
446024 | LLS 3.3V supply error | 3.3V regulated supply out of specification | critical |
446025 | LLS analog-to-digital converter error | ADC calibration or conversion failure detected | error |
446026 | LLS reference voltage error | Vref supply out of range or unstable | error |
446028 | LLS clock error detected | System clock monitor detected oscillator or PLL failure | critical |
446029 | LLS ambient failure detected | Ambient monitor detected environmental fault condition | warning |
446030 | LLS watchdog occurred in analog module | Analog module watchdog timer timeout | error |
446031 | Restart ventilation failed in LLS | Ventilation restart command failed or timed out | critical |
446032 | Patient pressure overpressure condition detected | Patient line pressure exceeded safety threshold | critical |
481001 | STU first run migration error | System configuration migration failed on initial startup | error |
481002 | CRC error in system configuration | Configuration data CRC check failed during load | error |
481003 | Unknown part number detected | Unrecognized component part number in system configuration | warning |
481004 | Technical state failed | System state validation failed during initialization | error |
481005 | Synchronization error in system startup | Module synchronization timeout during system boot | error |
481006 | Reconnect startup type error | Reconnection startup mode not recognized or invalid | warning |
481007 | Reconnect accept error | System rejected reconnection attempt during startup | error |
483001 | VGUI startup basic startup failure | Ventilator GUI basic initialization failed during boot | error |
483002 | VGUI screen initialization failure | VGUI screen driver initialization failed | error |
483003 | VGUI basics loading failure | VGUI base framework failed to load | error |
483004 | VGUI self-test prolog failure | VGUI self-test initialization sequence failed | warning |
483005 | VGUI self-test failure | VGUI internal self-test detected failure condition | error |
483006 | VGUI self-test epilog failure | VGUI self-test completion sequence failed | warning |
483007 | VGUI bitmap loading failure | GUI bitmap or image resources failed to load | error |
483008 | VGUI initialization failure | VGUI full initialization sequence failed | error |
484001 | SGUI startup basic startup failure | Standby GUI basic initialization failed | error |
484002 | SGUI screen initialization failure | SGUI screen driver failed to initialize | error |
484003 | SGUI basics loading failure | SGUI framework failed to load during startup | error |
484004 | SGUI self-test prolog failure | SGUI self-test initialization sequence failed | warning |
484005 | SGUI self-test failure | SGUI self-test detected internal failure | error |
484007 | SGUI bitmap loading failure | SGUI image resources failed to load | error |
484008 | SGUI initialization failure | SGUI initialization sequence failed | error |
485001 | AGL ambient failure detected | Alarm GUI monitor detected environmental condition | warning |
743001 | Alarm controller watchdog failed for LLS | ALR module watchdog timeout detected | error |
743003 | ALR MMI connection lost | Alarm controller lost connection to man-machine interface | warning |
746001 | Patient panel LLS watchdog failure | LLS watchdog timeout detected by ALR module | error |
746002 | Patient panel VGUI watchdog failure | VGUI module watchdog timeout detected | error |
746003 | Patient panel VGUI mode control watchdog failure | VGUI mode control watchdog timeout | error |
746004 | Patient panel error failure detected | Panel error monitoring detected fault condition | error |
746005 | Patient panel LLS exception occurred | Unhandled exception in patient panel LLS | error |
746006 | Patient panel LLS workload watchdog failure | Patient panel workload processing timeout | warning |
746007 | Patient panel ventilation restart failed | Patient panel ventilation restart command failed | critical |
783001 | VGUI monitoring channel observation failed | VGUI monitoring signal lost or degraded | error |
783002 | VGUI trending channel observation failed | VGUI trend data acquisition lost or interrupted | error |
783003 | VGUI VMC timeout | VMC (Ventilator Management Controller) communication timeout | error |
785001 | AGL alarming channel observation failed | Alarm GUI alarm channel signal lost | error |
785002 | AGL panel error failure detected | AGL detected error in alarm panel interface | error |
785003 | AGL PM channel observation failed | AGL parameter monitoring channel observation lost | error |
Battery calibration required | Battery capacity has declined below acceptable threshold | Battery capacity loss exceeds 25% of design capacity | warning |
Battery calibration required | Battery capacity loss exceeds 25% of designed capacity; battery too weak for reliable operation | Battery capacity drops below 75% of designed capacity (below 3750 mAh for 5000 mAh design) | warning |
Red flashing LED | Fuel gauge calibration required | Battery fuel gauge needs recalibration before use | warning |
Red solid LED | Charger/calibrator error | Fault detected during charge or calibration cycle | error |
Weak battery | Battery capacity reduced; device may provide inaccurate fuel gauge or fail to maintain ventilation on AC power loss | Battery end-of-life condition; capacity severely depleted | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.