Haemonetics MCS+ LN 9000: troubleshooting and error codes
The ServiceForge knowledge base holds 122 documented symptoms for the Haemonetics MCS+ LN 9000, with 217 ranked causes, 244 diagnostic checks with test points and expected values, and 205 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Protocol door open
System detects protocol card door is open and displays error 262
Ranked causes
- 1. Protocol door not closed (common)
- 2. Door microswitch failure (occasional)
- 3. Microswitch cable fault (occasional)
- 4. Ribbon cable failure (occasional)
- 5. Power supply failure (rare)
- 6. Processor PCB defect (rare)
Visually confirm protocol door position
- Test point
- Protocol Card door
- Expected
- Door fully closed
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Return cycle not completed patient safety
System displays error 389 when return cycle incomplete; bowl not empty or BLAD air signal detected early
Ranked causes
- 1. Bowl not completely emptying, viscous RBCs from over-centrifugation (common)
- 2. BLAD malfunction (occasional)
- 3. BLAD electrical connection failure (occasional)
- 4. BLAD signal measurement out of specification (occasional)
- 5. ACAD failure (occasional)
- 6. Processor PCB failure (rare)
Review centrifugation settings for current procedure
- Test point
- Procedure parameters
- Expected
- Centrifugation settings normal for blood type
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low yield or poor product quality patient safety
Procedures result in low yield, WBC contamination, RBC contamination, or poor product volumes
Ranked causes
- 1. Suboptimal machine parameters or settings (common)
- 2. Rotor not properly seated (occasional)
- 3. Bowl optics or sensor signal degradation (occasional)
- 4. Optics aiming volume misalignment (occasional)
- 5. Bowl vibration issue (occasional)
Confirm specific complaint from customer
- Test point
- Patient/donor record
- Expected
- Document precount, draw rate, hematocrit, volume processed, flow performance, ACD used, disposable lot number
10 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blue valve detected out of position patient safety
Blue valve did not respond to command and remains in incorrect position
Ranked causes
- 1. Valve solenoid power or control signal is incorrect (common)
- 2. Valve is mechanically stuck or failed (common)
- 3. Valve Driver PCB or Processor PCB is defective (occasional)
- 4. Valve compressor pneumatic pressure is low (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Card disposable mismatch
System displays error 372 when card does not match disposable
Ranked causes
- 1. Incorrect card-disposable pairing (common)
- 2. Disposable ID sensor malfunction (occasional)
- 3. Cable fault PCB to top deck (occasional)
- 4. Disposable ID PCB failure (occasional)
Visually verify card and disposable compatibility
- Test point
- Card and disposable labels or markings
- Expected
- Card and disposable IDs match
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Centrifuge cover is not locked patient safety
Centrifuge cover latch switches indicate cover is open or unlocked at start or during draw cycle
Ranked causes
- 1. Centrifuge cover halves not completely closed (common)
- 2. Cover latch microswitch misalignment (common)
- 3. Electrical connections to cover switch signals are damaged (occasional)
- 4. Advanced Processor PCB is defective (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
DPM pressure out of range
System displays error 345 when DPM pressure measurement exceeds acceptable range
Ranked causes
- 1. DPM pressure at limits due to operational conditions (common)
- 2. DPM calibration error (occasional)
- 3. Safety system configuration issue (occasional)
- 4. DPM PCB or assembly failure (occasional)
Close needle clamp and operate blood pump
- Test point
- Needle clamp, Red valve, Blood pump
- Expected
- Needle clamp closed, Red valve open, pump rotates clockwise
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Interlock failure between SPM and DPM
System displays error 370 when interlock failure detected between SPM/DPM
Ranked causes
- 1. Electrical connection failure in DPM or SPM circuit (common)
- 2. Filter sleeve not properly installed (common)
- 3. DPM or SPM assembly defect (occasional)
- 4. Motor driver PCB failure (occasional)
Inspect DPM and SPM electrical connections
- Test point
- All connections per system schematic
- Expected
- All connections secure and undamaged
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No donor pressure change detected at DPM
System displays error 373 when no donor pressure change detected at DPM
Ranked causes
- 1. DPM tubing occlusion or clamp (common)
- 2. Wet DPM filter (common)
- 3. DPM calibration error (occasional)
- 4. DPM PCB or assembly failure (occasional)
Inspect DPM tubing for occlusions or clamps
- Test point
- DPM tubing line
- Expected
- Clear tubing with no blockages or clamps
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No pressure change detected at DPM
System displays error 343 when DPM cannot detect venous pressure change
Ranked causes
- 1. Donor tubing occlusion or clamp (common)
- 2. Wet DPM filter (common)
- 3. DPM calibration error (occasional)
- 4. DPM assembly failure (occasional)
Inspect donor tubing for occlusions or clamps
- Test point
- Donor tubing line
- Expected
- Clear tubing, no blockages or clamps present
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No pressure change detected at SPM
System displays error 344 when no pressure change detected at the SPM
Ranked causes
- 1. SPM tubing occlusion or clamp (common)
- 2. Wet SPM filter (common)
- 3. SPM calibration error (occasional)
- 4. SPM PCB or assembly failure (occasional)
Inspect SPM tubing for occlusions or clamps
- Test point
- SPM tubing line
- Expected
- Clear tubing with no blockages or clamps
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No system pressure change detected at SPM
System displays error 374 when no system pressure change detected at SPM
Ranked causes
- 1. SPM tubing occlusion or clamp (common)
- 2. Wet SPM filter (common)
- 3. SPM calibration error (occasional)
- 4. SPM PCB or assembly failure (occasional)
Inspect SPM tubing for occlusions or clamps
- Test point
- SPM tubing line
- Expected
- Clear tubing with no blockages or clamps
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Orange valve detected out of position patient safety
Orange valve did not respond to command and remains in incorrect position
Ranked causes
- 1. Valve solenoid power or control signal is incorrect (common)
- 2. Valve is mechanically stuck or failed (common)
- 3. Valve Driver PCB or Processor PCB is defective (occasional)
- 4. Valve compressor pneumatic pressure is low (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Prime flow control problem
System displays error 351 when no flow detected due to high pressure or vacuum at DPM
Ranked causes
- 1. High pressure or vacuum condition at DPM (common)
- 2. DPM calibration error (occasional)
- 3. Safety system values incorrect (occasional)
- 4. DPM PCB or assembly failure (occasional)
Adjust needle clamp and blood pump operation
- Test point
- Needle clamp, Red valve, Blood pump
- Expected
- Needle clamp closed, Red valve open, pump clockwise
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Purple valve detected out of position patient safety
Purple valve did not respond to command and remains in incorrect position
Ranked causes
- 1. Valve solenoid power or control signal is incorrect (common)
- 2. Valve is mechanically stuck or failed (common)
- 3. Valve Driver PCB or Processor PCB is defective (occasional)
- 4. Valve compressor pneumatic pressure is low (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
White valve detected out of position patient safety
White valve did not respond to command and remains in incorrect position
Ranked causes
- 1. Valve solenoid power or control signal is incorrect (common)
- 2. Valve is mechanically stuck or failed (common)
- 3. Valve Driver PCB or Processor PCB is defective (occasional)
- 4. Valve compressor pneumatic pressure is low (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Yellow valve detected out of position patient safety
Yellow valve did not respond to command and remains in incorrect position
Ranked causes
- 1. Valve solenoid power or control signal is incorrect (common)
- 2. Valve is mechanically stuck or failed (common)
- 3. Valve Driver PCB or Processor PCB is defective (occasional)
- 4. Valve compressor pneumatic pressure is low (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air detected by blood line air detector patient safety
System displays error 331 when BLAD detects air in blood line
Ranked causes
- 1. Air in blood line (common)
- 2. Air detector electrical connection failure (occasional)
- 3. Processor PCB failure (rare)
Run air detector diagnostic test
- Test point
- Diagnostics menu - Air Detector test
- Expected
- Test confirms air is detected when expected
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air removed from air detector patient safety
System displays error 340 when air detector detects fluid instead of air
Ranked causes
- 1. Fluid in air detector when air expected (common)
- 2. Air detector electrical connection issue (occasional)
- 3. Processor PCB fault (rare)
Run air detector diagnostic
- Test point
- Diagnostics Air Detector test
- Expected
- Test confirms fluid detected when air expected
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Date and time not set correctly
Machine clock shows incorrect date/time; affects data logging timestamp accuracy
Ranked causes
- 1. Machine relocated to different time zone (occasional)
- 2. Daylight Savings Time change occurred (occasional)
- 3. Initial machine calibration not performed (rare)
Verify current date setting
- Test point
- Machine Parameters > Time of Day > Current Date
- Expected
- Today's date in MM-DD-YY format
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECV limit reached patient safety
System displays error 393 when maximum Extra Corporeal Volume limit reached; complete return cycle required
Ranked causes
- 1. Low hematocrit in donor blood (common)
- 2. Blood efficiency affected by flow restriction (common)
- 3. Blood pump rotor defect (common)
Review donor hematocrit and blood parameters
- Test point
- Donor pre-procedure blood work
- Expected
- Hematocrit within acceptable range for procedure
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Expected fluid flow to weigher not detected
System displays error 379 when no change in bag weight detected during collection
Ranked causes
- 1. Weigher calibration or offset error (common)
- 2. Load cell assembly failure (offset > 200) (occasional)
- 3. Processor PCB failure (rare)
Check weigher OFFSET and calibration value
- Test point
- Utilities - Weigher calibration menu
- Expected
- OFFSET less than 200, calibration within specification
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Extra corporeal volume limit reached during draw cycle
ECV limit alarm displayed; operator must decide whether to continue Draw or start Return Cycle
Ranked causes
- 1. Donor blood hematocrit very low (common)
- 2. Blood flow restriction affecting efficiency (common)
- 3. Blood pump rotor malfunction (common)
Measure blood pump flow efficiency
- Test point
- Blood pump outlet
- Expected
- Within ±10% of set volume
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No flow from donor detected
System displays error 341 when pump speeds decelerate to zero due to low pressure measurement
Ranked causes
- 1. Low donor pressure or flow restriction (common)
- 2. DPM calibration error (occasional)
- 3. DPM assembly failure (occasional)
Check donor tubing for occlusions or restrictions
- Test point
- Donor tubing
- Expected
- No clamps, kinks, or blockages visible
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Weigher bag weight changed unexpectedly (operation)
System displays error 387 when unexpected weight change occurs; do not touch weigher during operation
Ranked causes
- 1. External vibration or operator interference (common)
- 2. Weigher calibration or offset error (occasional)
- 3. Load cell assembly failure (occasional)
Verify weigher is not disturbed during operation
- Test point
- Weigher platform
- Expected
- No external contacts, vibration stable
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Weigher bag weight changed unexpectedly (TPE)
System displays error 386 when large weight change detected on weigher for TPE procedure
Ranked causes
- 1. Weigher calibration or offset error (common)
- 2. Load cell assembly failure (offset > 200) (occasional)
- 3. Processor PCB failure (rare)
Check weigher OFFSET and calibration
- Test point
- Utilities - Weigher calibration menu
- Expected
- OFFSET less than 200, calibration within specification
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
A/D converter test failed
Analog to digital converter does not read expected voltage levels (20%, 50%, or 80% of max)
Ranked causes
- 1. A/D converter circuit malfunction (unknown)
- 2. DPM or SPM transducer output out of range (unknown)
Measure DPM and SPM signal levels at 20%, 50%, 80% of maximum
- Test point
- DPM and SPM signal outputs
- Expected
- Values must be within specified range for each test level
Air sensor test failed
System cannot detect air in sensor line; audible warning and error code displayed
Ranked causes
- 1. Air detector sensor malfunction (unknown)
- 2. Sensor line blockage or disconnection (unknown)
Verify air detector sensor functionality
- Test point
- ACAD, BLAD, DLAD1, DLAD2 sensors
- Expected
- Each sensor must sense air; test line turned on to ensure air can be sensed
Calibration: Bowl Optics Signal (seven valve machine)
Adjust optics beam aiming and gain to achieve specified peak voltage reading on rim of aiming fixture
Ranked causes
- 1. Optics beam not peaked or offset/gain out of specification (unknown)
- 2. Offset voltage out of specification (unknown)
Peak optics beam on rim of aiming fixture
- Test point
- Optics aiming fixture #ETE-5007 rim (nick-free section)
- Expected
- 2.500 +/- 0.5 volts V, +/- 0.5
- Tools
- Voltage Meter, Fluke 21 or equivalent, optics adjusting fixture #ETB-5266
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Power Supply high voltage
Verify DC output voltages are within specification on Advanced Personality PCB test points TP1–TP5 using ground reference TP6
Ranked causes
- 1. 5V supply out of specification (unknown)
- 2. Other supply voltages out of specification (unknown)
Measure 5V output
- Test point
- TP pin 1
- Expected
- 5.1 +0.1/-0 VDC, +0.1/-0
- Tools
- Voltage Meter, Fluke 21 or equivalent
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Centrifuge cover switch test failed
Cover switch transition or relay not responding correctly during manual test
Ranked causes
- 1. Cover switch not transitioning or stuck (unknown)
- 2. Centrifuge relay not responding (unknown)
Check initial cover switch state
- Test point
- Centrifuge cover switch
- Expected
- Switch state matches cover position (open or closed)
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Cuff transducer out of tolerance
Cuff pressure reading exceeds specification limits; may be installed during test startup
Ranked causes
- 1. Cuff connected during test (common)
- 2. Cuff transducer malfunction (rare)
Verify cuff is disconnected from rear panel
- Test point
- Cuff connector
- Expected
- Cuff must be disconnected
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Disposable ID and protocol do not match patient safety
System displays error message when loaded disposable ID does not match selected protocol
Ranked causes
- 1. Wrong disposable loaded for selected protocol (common)
- 2. Disposable ID reader failure (occasional)
Verify operator protocol selection
- Test point
- Display screen protocol indication
- Expected
- Protocol matches disposable being installed
5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
DPM or SPM connector test failed
Analog verification fails; connector may be unseated or filter not properly removed
Ranked causes
- 1. DPM or SPM filter installed when test commenced (common)
- 2. DPM or SPM connector disconnected or unseated (unknown)
Check DPM and SPM filters installed
- Test point
- DPM and SPM interlock sensors
- Expected
- Filters must not be installed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
DPM or SPM out of tolerance
Transducer offset readings exceed ambient pressure limits when filters not installed
Ranked causes
- 1. DPM or SPM filter installed when test commenced (common)
- 2. DPM or SPM transducer failure or drift (unknown)
Check DPM and SPM filters are not installed
- Test point
- DPM and SPM interlock
- Expected
- Filters must not be installed; interlock test passes
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error message displayed during operation
System displays error code on screen; operator can press NO to mute alarm for up to 30 seconds or HELP for troubleshooting hints
Ranked causes
- 1. Main computer detected correctable situation (common)
- 2. Safety system detected unsafe condition (occasional)
Identify source of error message
- Test point
- Screen display
- Expected
- Error code visible on display
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
High pressure at DPM detected patient safety
Donor pressure monitor reading exceeds acceptable limits; system safety response triggered
Ranked causes
- 1. Actual high pressure condition in donor line (common)
- 2. DPM transducer malfunction or false reading (occasional)
Verify DPM pressure reading is within acceptable maximum limits
- Test point
- DPM transducer output
- Expected
- Pressure at or below maximum specified limit
Incomplete cycle waste bag full
System displays error 391 for TPE when waste bag is full during collection
Ranked causes
- 1. Waste bag full during TPE procedure (common)
- 2. Weigher calibration error (occasional)
Verify waste bag is full and replace
- Test point
- Waste bag weight or visual inspection
- Expected
- Bag full and ready for disposal
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Line sensor voltage out of range
Sensor output voltage too low or too high; tubing may be installed or sensor needs cleaning
Ranked causes
- 1. Tubing installed during test (common)
- 2. Line sensor dirty or contaminated (occasional)
Verify no tubing installed on line sensor
- Test point
- Line sensor input
- Expected
- Tubing must be removed for test
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low pressure at DPM detected patient safety
Donor pressure monitor reading below acceptable limits; system safety response triggered
Ranked causes
- 1. Actual low pressure condition in donor line (common)
- 2. DPM transducer malfunction or false reading (occasional)
Verify DPM pressure reading is within acceptable minimum limits
- Test point
- DPM transducer output
- Expected
- Pressure at or above minimum specified limit
Protocol card not detected
System displays message to insert protocol card; card insertion fails to be recognized
Ranked causes
- 1. Protocol card not inserted (common)
- 2. Protocol card CRC mismatch (occasional)
Verify protocol card is inserted in socket
- Test point
- Protocol card socket
- Expected
- Card fully seated in socket with no gaps
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ROM CRC test fails
Bar graph displayed during ROM test, audible warning sounds, system halts with error code
Ranked causes
- 1. Program chip CRC mismatch (rare)
- 2. NVRAM chip corrupted or replaced (occasional)
Observe ROM CRC test progress bar and final result
- Test point
- System display screen
- Expected
- Bar graph completes; no error message; no audible warning
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Safety system test failed patient safety
Safety system cannot open relay or respond to simulated error conditions; pump control compromised
Ranked causes
- 1. Safety system board malfunction (unknown)
- 2. Safety relay stuck or not responding (unknown)
Verify safety system can open relay under simulated error conditions
- Test point
- Safety system relay
- Expected
- Relay must open (power disconnected from both pumps) when error simulated
System fails power-on self-test
System emits continuous tone, random key lights displayed, POST diagnostic fails
Ranked causes
- 1. RAM stuck bit detected (common)
- 2. CPU functional failure (occasional)
Listen for continuous tone during power-on
- Test point
- System speaker output
- Expected
- Single tone for 2 seconds or less if RAM test passes
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Valve transitioning test failed patient safety
Valve position feedback incorrect or current transition not detected; valve may be stuck
Ranked causes
- 1. Valve stuck or mechanical failure (unknown)
- 2. Current feedback circuit fault (unknown)
Transition each valve (red, blue, yellow, green, white, orange, purple) and verify position feedback
- Test point
- Valve position feedback signal
- Expected
- Position feedback must match commanded state for each valve
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Watchdog test failed patient safety
Watchdog clock monitoring not responding to CPU halt; safety system synchronization lost
Ranked causes
- 1. CPU to safety system watchdog clock fault (unknown)
- 2. Safety system relay not responding to watchdog fault (unknown)
Stop watchdog clock generated by CPU and verify safety system reaction
- Test point
- Watchdog clock and safety relay
- Expected
- Safety system must open relay and post fault in fault register when clock stopped
Weigher out of tolerance
Load cell or offset reading exceeds specification limits; weight may be on device
Ranked causes
- 1. Weight remaining on weigher during test (common)
- 2. Load cell malfunction or drift (rare)
Verify weigher is unloaded
- Test point
- Weigher load cell
- Expected
- No weight on weigher; offset within ±125 digital units from standard
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC Pump Direction Wrong patient safety
AC pump is turning in the wrong direction.
Ranked causes
- 1. Direction sensor or drive error (unknown)
Verify AC pump rotation direction
- Test point
- AC pump direction feedback
- Expected
- Correct direction per hardware sensor
- Tools
- System monitor
AC Pump Position Incorrect patient safety
AC pump is lagging or preceding desired position by 128 counts.
Ranked causes
- 1. Motor position tracking error (unknown)
Verify AC pump position feedback via Advanced Motor Driver PCB
- Test point
- Advanced Motor Driver PCB
- Expected
- Within 128 counts of desired position
- Tools
- System monitor
AC Pump Too Fast patient safety
AC pump speed is beyond +/- 5 cc/min from programmed value during operation.
Ranked causes
- 1. Pump speed deviation (unknown)
Verify AC pump speed against programmed value
- Test point
- AC pump
- Expected
- +/- 5 cc/min from programmed value, +/- 5
- Tools
- System monitor
AC Pump Too Slow patient safety
AC pump speed is beyond +/- 5 cc/min from programmed value during operation.
Ranked causes
- 1. Pump speed deviation (unknown)
Verify AC pump speed against programmed value
- Test point
- AC pump
- Expected
- +/- 5 cc/min from programmed value, +/- 5
- Tools
- System monitor
AC/Blood Pump Ratio High patient safety
Anticoagulant/blood pump ratio is beyond +/- 2 from programmed value.
Ranked causes
- 1. Pump ratio deviation (unknown)
Verify AC/Blood pump ratio against programmed value
- Test point
- AC and Blood pumps
- Expected
- +/- 2 from programmed value, +/- 2
- Tools
- System monitor
AC/Blood Pump Ratio Low patient safety
Anticoagulant/blood pump ratio is beyond +/- 2 from programmed value.
Ranked causes
- 1. Pump ratio deviation (unknown)
Verify AC/Blood pump ratio against programmed value
- Test point
- AC and Blood pumps
- Expected
- +/- 2 from programmed value, +/- 2
- Tools
- System monitor
Bar code reader not scanning correctly
Bar code reader unable to read codes or display input value
Ranked causes
- 1. Bar code reader not correctly installed (common)
View bar code reader input value
- Test point
- Bar Code Diagnostics menu
- Expected
- Input Value = XX.XX displayed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood Pump Direction Wrong patient safety
Blood pump is turning in the wrong direction.
Ranked causes
- 1. Direction sensor or drive error (unknown)
Verify blood pump rotation direction
- Test point
- Blood pump direction feedback
- Expected
- Correct direction per hardware sensor
- Tools
- System monitor
Blood Pump Position Incorrect patient safety
Blood pump is lagging or preceding desired position by 128 counts.
Ranked causes
- 1. Motor position tracking error (unknown)
Verify blood pump position feedback via Advanced Motor Driver PCB
- Test point
- Advanced Motor Driver PCB
- Expected
- Within 128 counts of desired position
- Tools
- System monitor
Blood Pump Too Fast patient safety
Blood pump speed is beyond +/- 5 cc/min from programmed value during operation.
Ranked causes
- 1. Pump speed deviation (unknown)
Verify blood pump speed against programmed value
- Test point
- Blood pump
- Expected
- +/- 5 cc/min from programmed value, +/- 5
- Tools
- System monitor
Blood Pump Too Slow patient safety
Blood pump speed is beyond +/- 5 cc/min from programmed value during operation.
Ranked causes
- 1. Pump speed deviation (unknown)
Verify blood pump speed against programmed value
- Test point
- Blood pump
- Expected
- +/- 5 cc/min from programmed value, +/- 5
- Tools
- System monitor
Blue Valve Out of Position patient safety
Blue valve is out of position during operation.
Ranked causes
- 1. Valve position sensor or drive failure (unknown)
Verify blue valve position feedback
- Test point
- Blue valve feedback
- Expected
- Correct open/closed state
- Tools
- System monitor
Calibration CRC test failed
Power-on self-test fails; system displays error and audible warning; operator prompted to check error
Ranked causes
- 1. Calibration parameters corrupted or incorrect CRC value (unknown)
Verify calibration CRC computation
- Test point
- Calibration variables storage space
- Expected
- CRC values must match stored value in non-volatile memory
More symptoms
- Calibration: A/D Cuff
- Calibration: Bowl Optics Final (seven and eight valve machines)
- Calibration: Data Card baud rate
- Calibration: Data Card serial port 1 output voltage
- Calibration: Functional Test
- Calibration: Line Sensor
- Centrifuge actual speed does not match defined speed setpoint
- Centrifuge bowl not seating properly on spindle
- Centrifuge cover does not latch or seal
- Centrifuge does not maintain vacuum
- Centrifuge Not Turning
- Centrifuge Speed Too Fast
- Centrifuge Speed Too Slow
- Centrifuge Vacuum Not Detected
- Centrifuge vacuum system obstructed
- Channel 1 state test failed
- Channel 1 time base test failed
- Channel 2 state test failed
- Channel 2 time base test failed
- Configuration CRC test failed
- Device Disposal
- Display shows incorrect pixels or is blank
- DPM Filter Disconnected
- DPM interlock switches not functioning
- DPM pressure sensor inaccurate
- Drip monitor counts inaccurate at high speed
- Drip monitor counts inaccurate at low speed
- Error code 28 - Fluid detected in centrifuge
- Error code 74 - Division by Zero
- Error code 78 - Overflow Error
- Error code 79 - Array Bounds Error
- Error code 80 - Illegal Opcode
- Green Valve Out of Position
- Incomplete cycle no product collected
- Infrared interface configuration incorrect
- Infrared interface P1005 jumper not set
- Keyboard key does not respond or LED does not light
- Machine requires recovery after shutdown
- Machine shutdown during procedure
- Network diagnostics request number not displaying
- Orange Valve Out of Position
- Pinch valve does not open or close
- Power supply output voltage out of specification
- Protocol card does not match disposable
- Protocol parameters reset
- Purple Valve Out of Position
- Red Valve Out of Position
- Safety system fault detected
- Safety system pressure out of specification
- SPM Filter Disconnected
- SPM interlock switches not functioning
- SPM pressure sensor inaccurate
- SW1 switch baud rate not set to 19200
- SW2 switch configuration incorrect
- System startup self-test in progress
- Weigher reading inaccurate
- White Pump Direction Wrong
- White Pump Position Incorrect
- White Pump Too Fast
- White Pump Too Slow
- White Valve Out of Position
- Yellow Valve Out of Position
62 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
0 | Time Out Error | Software watchdog timing constraint exceeded | critical |
0 | A CPU error caused the MCS+ to reset. | Computer encountered an error during operation | critical |
100 | Centrifuge safety switch not operational | Safety switch failure detected | error |
101 | Return cycle taking too long | Bowl emptying time exceeds expected duration | warning |
102 | Air detected earlier than expected | Bowl emptying time shorter than expected | warning |
103 | Low AC flow detected by drip monitor | AC flow rate below threshold | warning |
104 | High AC flow detected by drip monitor | AC flow rate above threshold | warning |
105 | No AC flow detected by drip monitor | AC flow not detected | error |
106 | No cuff pressure detected | Cuff pressure measurement failure | error |
107 | No air plasma interface detected | Air plasma interface detection failure | error |
16 | CPU error caused system reset | System problem during operation | critical |
17 | CPU error caused system reset | System problem during operation | critical |
18 | CPU error caused system reset | System problem during operation | critical |
18 | Fatal Error | Computer execution flow error | critical |
19 | Stack Full Error | Program stack overflow detected | critical |
19 | CPU error caused system reset | System problem during operation | critical |
20 | Stack Empty Error | Program stack underflow detected | critical |
21 | CPU error caused system reset | System problem during operation | critical |
22 | CPU error caused system reset | System problem during operation | critical |
22 | Error Reading Queue | Queue read operation failed | error |
223 | Centrifuge cover is not locked | Centrifuge cover latch microswitches indicate open or unlocked cover at start or during draw cycle | error |
224 | Purple Valve Out of Position | Purple valve not in correct position during operation | warning |
225 | Orange Valve Out of Position | Orange valve not in correct position during operation | warning |
226 | Blue Valve Out of Position | Blue valve not in correct position during operation | warning |
227 | Yellow Valve Out of Position | Yellow valve not in correct position during operation | warning |
228 | White Valve Out of Position | White valve not in correct position during operation | warning |
229 | Green valve detected out of position | Green valve malfunction or disconnection during operation | error |
23 | CPU error caused system reset | System problem during operation | critical |
23 | No Queue Found | Expected queue structure not located in memory | error |
24 | Error Writing to Queue | Queue write operation failed | error |
24 | CPU error caused system reset | System problem during operation | critical |
25 | Centrifuge stalled | Centrifuge encoder signals indicate stall | error |
252 | System CPU error encountered | CPU error related to pressure failure causing MCS+ reset | critical |
254 | Red valve detected out of position | Red (Donor) valve malfunction or disconnection | error |
255 | DPM filter has been disconnected | DPM filter connection loss detected | error |
256 | SPM filter has been disconnected | SPM filter connection loss detected | error |
257 | High pressure error at DPM | DPM pressure exceeded upper limit | error |
258 | Low pressure error at DPM | DPM pressure below lower limit | error |
259 | SPM high pressure error | SPM pressure exceeded upper limit | error |
26 | Centrifuge running too fast | Spindle speed 300 RPM greater than set speed | warning |
260 | SPM low pressure error | SPM pressure below lower limit | error |
261 | Draw not allowed with full bowl | Cycle interrupted with full bowl; reseparation cannot be achieved | error |
262 | Protocol card door is open | Protocol door opened during operation | warning |
263 | Centrifuge cover now locked | Centrifuge cover locking mechanism engaged | info |
264 | Protocol card door is closed | Protocol card door closed after being opened | info |
265 | Pump manifold not installed | System detects white pump manifold is not snapped in place | error |
27 | Centrifuge vacuum not detected | Vacuum level less than 9±1 psi during operation | warning |
28 | Fluid detected in centrifuge | Fluid detected in centrifuge well | critical |
29 | CPU error related to optics system | CPU error related to optics system caused reset | critical |
30 | CPU error related to air sensor | CPU error related to air sensor caused reset | critical |
305 | Stop key pressed | Operator presses STOP key | info |
31 | Centrifuge running too slow | Centrifuge encoder signals indicate spindle speed 300 RPM less than set speed | warning |
32 | CPU error related to A/D time-out | CPU error during A/D conversion caused reset | critical |
32 | AD Time Out Error | Analog-to-digital converter timeout during conversion | error |
33 | Sig Queue Full Error | Signal queue buffer overflow | error |
33 | CPU error related to queue | CPU error in queue processing caused reset | critical |
331 | Air detected by blood line air detector (BLAD) | BLAD detects air in blood line | error |
34 | CPU error related to ACK | CPU error related to ACK signal caused reset | critical |
34 | Ack Error | Acknowledgment signal not received or invalid | error |
340 | Air removed from air detector | Air detector detects fluid when air expected | error |
341 | No flow from donor detected | Pump speed decelerates to zero due to low pressure | warning |
343 | No pressure change detected at DPM | DPM cannot detect venous pressure change | warning |
344 | No pressure change detected at SPM | SPM cannot detect system pressure change | warning |
345 | DPM pressure out of range | DPM pressure measurement exceeds limits | warning |
35 | CPU error related to weigher | CPU error related to weigher system caused reset | critical |
351 | Prime flow control problem | No flow due to high pressure or vacuum at DPM | warning |
36 | CPU error related to line red | CPU error related to line red condition caused reset | critical |
37 | Air detected by system monitor | DLAD1 or DLAD2 air detector detects air | error |
370 | Interlock failure SPM/DPM | Interlock failure detected between SPM and DPM | error |
372 | Card/disposable mismatch | Card does not match the disposable | error |
373 | No donor pressure change detected at DPM | DPM cannot detect donor venous pressure change | warning |
374 | No system pressure change detected at SPM | SPM cannot detect system pressure change | warning |
379 | Expected fluid flow to weigher not detected | No change in bag weight detected during collection | warning |
38 | CPU error related to ATOD converter | CPU error in ATOD converter caused reset | critical |
380 | Expected fluid flow from weigher not detected | No change in bag weight detected during outflow | warning |
386 | Weigher bag weight changed (TPE) | Large weight change detected on weigher for TPE | warning |
387 | Weigher bag weight changed unexpectedly | Unexpected weight change during operation | warning |
388 | Procedure recovery | Power off for short time | info |
389 | Return cycle not completed | Return cycle incomplete; bowl not empty or BLAD signals early | error |
39 | Centrifuge running too slow | Centrifuge encoder signals indicate spindle speed 300 RPM less than set speed | warning |
390 | Incomplete cycle no product collected | Cycle interrupted early with no platelets collected | warning |
391 | Incomplete cycle waste bag full | Waste bag full during TPE procedure | warning |
393 | ECV maximum limit reached | Maximum Extra Corporeal Volume limit reached | warning |
394 | Extra Corporeal Volume limit reached | ECV limit threshold crossed during draw cycle | warning |
40 | AC pump running too slow | AC pump encoder signals indicate pump moving slower than set speed | warning |
400 | CPU test failure | CPU opcodes or registers fail functional test during power-on POST | critical |
402 | ROM CRC test failure | CRC 16 checksum of program code does not match stored value in ROM chip | critical |
404 | Protocol card CRC failure | CRC checksum of protocol card contents does not match value in last two bytes of address space | error |
405 | Calibration CRC test failed | CRC mismatch between computed and stored calibration values during power-on self-test | error |
406 | Configuration CRC test failed | CRC mismatch between computed and stored configuration values during power-on self-test | error |
407 | Protocol card does not match disposable | Installed protocol card incompatible with loaded disposable during power-on test | error |
408 | ACAD sensor test failed | Air detector (ACAD) cannot sense air during self-test | error |
409 | BLAD sensor test failed | Air detector (BLAD) cannot sense air during self-test | error |
41 | Blood pump running too slow | Blood pump encoder signals indicate pump moving slower than set speed | warning |
410 | DLAD1 sensor test failed | Air detector (DLAD1) cannot sense air during self-test | error |
411 | DLAD2 sensor test failed | Air detector (DLAD2) cannot sense air during self-test | error |
413 | Purple Valve self-test has failed | Purple valve self-test fails during system startup or diagnostics | error |
414 | Orange Valve self-test has failed | Orange valve self-test fails during system startup or diagnostics | error |
415 | Red Valve self-test has failed | Red valve self-test fails during system startup or diagnostics | error |
416 | Blue Valve self-test has failed | Blue valve self-test fails during system startup or diagnostics | error |
417 | Yellow Valve self-test has failed | Yellow valve self-test fails during system startup or diagnostics | error |
418 | White Valve self-test has failed | White valve self-test fails during system startup or diagnostics | error |
419 | Green Valve self-test has failed | Green valve self-test fails during system startup or diagnostics | error |
42 | White pump running too slow | White pump encoder signals indicate pump moving slower than set speed | warning |
420 | A/D Converter 20% self-test has failed | A/D Converter 20% self-test fails during system diagnostics | error |
421 | A/D Converter 50% self-test has failed | A/D Converter 50% self-test fails during system diagnostics | error |
422 | A/D Converter 80% self-test has failed | A/D Converter 80% self-test fails during system diagnostics | error |
423 | DPM connector test failed | DPM filter installed when analog verification test commenced or connector disconnected | error |
423 | DPM and SPM filters cannot be installed during self-tests | DPM filter detected as installed during system self-test initialization | error |
424 | DPM and SPM filters cannot be installed during self-tests | SPM filter detected as installed during system self-test initialization | error |
424 | SPM connector test failed | SPM filter installed when analog verification test commenced or connector disconnected | error |
425 | DPM filter needs to be installed | DPM filter not detected by system during self-test | warning |
426 | SPM filter needs to be installed | SPM filter not detected by system during self-test | warning |
427 | MCS+ could not detect DPM removal during connector test | DPM connector test fails, system cannot detect DPM removal | error |
427 | DPM connector test failed (alternate) | DPM signal verification failed during analog voltage testing | error |
428 | SPM connector test failed (alternate) | SPM signal verification failed during analog voltage testing | error |
428 | MCS+ could not detect SPM removal during connector test | SPM connector test fails, system cannot detect SPM removal | error |
429 | DPM offset (ambient pressure) is out of acceptable range (2.32 to 2.68 VDC) | DPM self-test detects offset outside specification during startup | error |
430 | SPM offset (ambient pressure) is out of acceptable range (2.32 to 2.68 VDC) | SPM self-test detects offset outside specification during startup | error |
431 | Weigher signal is out of allowed range | Weigher signal out of tolerance; Error 466 occurs first; if unresolved after 60 sec, Error 431 displays | error |
432 | Cuff pressure is out of allowed range | Cuff pressure self-test detects pressure outside specification | error |
433 | Safety system failed self-test | Safety system cannot open relay or does not respond to simulated error conditions during power-on test | critical |
434 | Safety system self-test has failed | Safety system self-test fails during system diagnostics | critical |
435 | Line Sensor output is too low | Line Sensor output voltage below specification during self-test | error |
436 | Line Sensor output is too high | Line Sensor output voltage above specification during self-test | error |
438 | Watchdog self-test has failed | Watchdog self-test fails during system diagnostics | critical |
439 | Low pressure at DPM | DPM transducer reading indicates pressure below minimum acceptable limit during power-on test | error |
439 | DPM signal is out of allowed range at low pressure | DPM signal detected outside acceptable range during operation | error |
44 | AC pump running too fast | AC pump encoder signals indicate pump moving faster than set speed | warning |
441 | DPM signal is out of the allowed range | Self test problem: high pressure error at the DPM | error |
441 | High pressure at DPM | DPM transducer reading indicates pressure above maximum acceptable limit during power-on test | error |
443 | State Test on Channel 1 has failed | Self test problem: state test on channel 1 failed | error |
444 | State Test on Channel 2 has failed | Self test problem: state test on channel 2 failed | error |
447 | Time Base Test on Channel 1 has failed | Self test problem: time base test on channel 1 failed | error |
448 | Time Base Test on Channel 2 has failed | Self test problem: time base test on channel 2 failed | error |
449 | Safety System self-test has failed | Self test problem: safety system failed the self test | critical |
45 | Blood pump running too fast | Blood pump encoder signals indicate pump moving faster than set speed | warning |
452 | Safety System self-test has failed | Self test problem: safety system failed the self test | critical |
453 | Safety System self-test has failed | Self test problem: safety system failed the self test | critical |
46 | White pump running too fast | White pump encoder signals indicate pump moving faster than set speed | warning |
461 | Centrifuge vacuum system is obstructed | Vacuum switch indicates blockage between switch and centrifuge spindle during system start-up vacuum test | error |
462 | Disposable ID reader has failed the self-test | Self test problem: disposable ID reader failed self test | error |
466 | Remove weight from weigher | Weigher detects weight (±125 digital units above standard offset) at start of weigher self-test | warning |
466 | Measured offset is +125 digital units from programmed offset | Weight detected on weigher when should be empty | warning |
467 | Excessive cuff pressure measured | MCS+ measures excessive cuff pressure | error |
467 | Disconnect cuff from rear panel | Cuff transducer reading indicates cuff connected at start of cuff self-test | warning |
468 | Line sensor is dirty or tubing not removed | Self test problem: line sensor not clean or tubing not removed | warning |
469 | Open the centrifuge cover latch | Centrifuge cover initially closed when cover switch test begins | info |
469 | Centrifuge cover switch must be tested for safe operation | Self test: centrifuge cover latch open when should be closed | warning |
470 | Centrifuge cover switch must be tested for safe operation | Self test: centrifuge cover latch closed when should be open | warning |
470 | Close the centrifuge cover latch | Centrifuge cover initially open when cover switch test begins | info |
471 | Protocol parameters reset | Protocol parameter CRC mismatch detected during power-on self-test | warning |
472 | BIOS CRC failure detected | Computer detects failure of BIOS CRC | critical |
475 | Calibration revision is incompatible | Self test problem: calibration revision incompatible | warning |
476 | Configuration revision is incompatible | Self test problem: configuration revision incompatible | warning |
478 | Centrifuge relay test failed | Self test problem: centrifuge relay test failed | error |
48 | AC pump running in wrong direction | AC pump encoder signals indicate pump moving in wrong direction | error |
49 | Blood pump running in wrong direction | Blood pump encoder signals indicate pump moving in wrong direction | error |
5 | Air detected by BLAD sensor | Air detected during operation | error |
5 | The BLAD detects air in the system. | Air detected by the BLAD air detector | error |
50 | White pump running in wrong direction | White pump encoder signals indicate pump moving in wrong direction | error |
52 | AC pump position not correct | AC pump position 128 counts ahead or lagging from set position | warning |
53 | Blood pump position not correct | Blood pump position 128 counts ahead or lagging from set position | warning |
54 | White pump position not correct | White pump position 128 counts ahead or lagging from set position | warning |
56 | AC/Blood pump ratio incorrect | AC or Blood pump ratio error detected during operation | warning |
57 | AC/Blood pump ratio incorrect | Occlusion at AC or blood pump during operation | error |
58 | Red valve detected out of position | Valve position sensing failure or valve mechanism malfunction | error |
59 | AC/Blood pump ratio too low | Pump speed or efficiency loss, encoder or motor malfunction | error |
6 | Air detected by DLAD1 sensor | Air detected during operation | error |
60 | AC/Blood pump ratio too high | Pump overspeeding or encoder reading error | error |
61 | CPU error with ATOD failure | Analog-to-digital converter malfunction causing CPU reset | critical |
62 | Donor ECV limit reached | Extra Corporeal Volume reaches 690 ml threshold | warning |
63 | Safety system error with Time Base Safety Fault | Time base reference failure in safety processor | critical |
64 | Safety system watchdog error | Watchdog timer expiration indicating system hang | critical |
65 | DPM/SPM connector fault | Electrical connection or interlock failure on DPM or SPM module | critical |
66 | Pressure fault causing system reset | Pressure spike or anomalous pressure reading in system | warning |
67 | Air detector circuit failure | Air detector sensor malfunction or circuit fault | critical |
68 | Blood pump feedback to safety board fault | Blood pump feedback signal abnormality to safety processor | critical |
69 | Safety system ECV limit warning | Extra Corporeal Volume reaches 690 ml threshold | warning |
7 | Air detected by DLAD2 sensor | Air detected during operation | error |
70 | Safety system ratio fault with pump occlusion | AC or blood pump occlusion causing abnormal pump ratio | error |
71 | Pump motor failure | Pump motor mechanical binding or electrical fault | critical |
72 | Prime error causing system reset | Air in system or improper priming of disposable tubing | critical |
73 | Power interruption caused system reset | AC power loss | error |
74 | CPU error - division by zero | CPU division by zero fault | error |
75 | CPU error - single step | CPU single step interrupt | error |
76 | CPU error - NMI | Non-maskable interrupt | error |
77 | CPU error - breakpoint | Breakpoint interrupt during execution | error |
78 | CPU error - overflow | Arithmetic overflow condition | error |
79 | CPU error - array bounds | Array bounds violation | error |
80 | CPU error - invalid opcode | Invalid processor instruction | error |
81 | CPU error - escape | Escape instruction error | error |
82 | CPU error - timer | Timer interrupt fault | error |
83 | CPU error - DMA | Direct memory access fault | error |
84 | CPU error - Air ISR | Air detector interrupt service routine fault | error |
85 | CPU error - INT ISR | Interrupt service routine fault | error |
86 | CPU error - double fault | Nested exception during exception handling | critical |
87 | CPU error - coprocessor overrun | Coprocessor task overrun | error |
88 | CPU error - invalid TSS | Invalid task state segment | error |
89 | CPU error - segment | Segment table error | error |
90 | CPU error - stack fault | Stack underflow or overflow | error |
91 | CPU error - general protection fault | Memory protection violation | error |
92 | CPU error - page fault | Virtual memory page access error | error |
93 | CPU error - coprocessor overrun | Coprocessor task overrun | error |
94 | CPU error - unexpected INT | Unexpected interrupt | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.