Gyrus ACMI G400 Workstation: troubleshooting and error codes

The ServiceForge knowledge base holds 6 documented symptoms for the Gyrus ACMI G400 Workstation, with 8 ranked causes, 10 diagnostic checks with test points and expected values, and 102 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

System reset triggered (Fault 100-10)

Watchdog timeout or software execution failure causes system reset; display shows 'SYSTEM RESET'

Ranked causes

  1. 1. Unstable mains supply voltage (common)
  2. 2. Software watchdog timeout due to internal timing issue (rare)
First check

Measure mains supply voltage

Test point
Wall outlet and G400 power input
Expected
Within specification and stable

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Video output not functioning

Generator video output to monitor does not display or shows incorrect OSD overlay

Ranked causes

  1. 1. Video OSD display disabled or misconfigured (common)
  2. 2. Video input/output BNC connectors disconnected or faulty (occasional)
First check

Navigate to Video OSD Select menu

Test point
G400 front panel Mode switch
Expected
'Video OSD Select' displayed after pressing Mode multiple times

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Mains electrical safety test failure high voltage

Electrical safety measurements do not comply with Class 1 type BF specifications per IEC60601-1-1

Ranked causes

  1. 1. Internal electrical insulation failure (unknown)
First check

Verify safety tester calibration

Test point
Safety tester documentation
Expected
Tester calibrated and approved for Class 1 type BF devices

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Non-volatile memory corrupt (Fault 100-12)

Settings memory corrupted or not initialised; display shows 'SETTINGS CORRUPT' and 'SETTINGS FLT!'

Ranked causes

  1. 1. Settings memory corruption or initialisation failure (rare)
First check

Perform manual reset via Mode key

Test point
G400 front panel Mode key
Expected
Fault symbol flashes after first press; clears after second press

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

RF output not within specification high voltage

Generator RF power output does not meet expected values during testing with Biomed Test Cable

Ranked causes

  1. 1. Internal RF generation circuit fault (unknown)
First check

Verify Biomed Test Cable connection

Test point
G400 front panel
Expected
Display reads 'BioMed Non-Human-Use' PK3 200W, DES 120W

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

ROM checksum failure (Fault 100-11)

Internal firmware ROM corruption detected; display shows 'INTERNAL FAILURE' and '!!!!'

Ranked causes

  1. 1. ROM memory corruption or programming error (rare)
First check

Power cycle and observe self-test

Test point
G400 front panel display
Expected
'ATTACH INSTRUMENT' symbol displays after self-test completes successfully

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
100-10System reset triggered by watchdog timeoutSoftware execution failure or watchdog reset timeoutwarning
100-11ROM checksum failure detectedFirmware memory integrity check fails on startupcritical
100-12Non-volatile settings memory corruptedSettings memory not properly initialised or corruptederror
100-17System reset triggered by software execution failureBackground software execution errorwarning
100-18Continuous test timing failureBackground timing test exceeds time limitcritical
100-19Background loop timing failureBackground task execution timing exceeds specificationcritical
100-20Status event queue overrunInternal status event queue buffer overflowcritical
100-21User event queue overrunInternal user event queue buffer overflowcritical
100-22VFD event queue overrunDisplay interface event queue buffer overflowcritical
100-23Waveform event queue overrunRF waveform event queue buffer overflowcritical
200-10PWM signal shorted highPWM signal stuck at high logic level during self-testcritical
200-11PWM signal shorted lowPWM signal stuck at low logic level during self-testcritical
200-14Energy signal stuck highEnergy feedback signal remains high during self-testcritical
200-18CUT/COAG signal stuck highMode selection signal remains high during self-testcritical
200-19CUT/COAG signal stuck lowMode selection signal remains low during self-testcritical
200-20CLAMP signal error in DAC outputDigital-to-analog converter output for clamp control failscritical
200-21BOOST signal error in DAC outputDigital-to-analog converter output for boost control failscritical
200-22PEAKSET signal error in DAC outputDigital-to-analog converter output for peak setting failscritical
200-24RF_DET signal stuck lowRF detection feedback signal remains low during self-testcritical
200-25OVERDOSE signal permanently ONOverdose protection signal stuck high during self-testcritical
200-26ENERGY signal stuck lowEnergy feedback signal remains low during self-testcritical
200-27PlasmaBlend Transformer feedback failureInternal system fault detected during POSTcritical
200-28G400 Workstation too cold, dependent on temperatureUnit startup when ambient temperature is below operational thresholdwarning
200-31CURRLIM signal error (DAC output)Internal POST check failurecritical
200-32VOLTLIM signal error (comparator input)Internal POST check failurecritical
200-33BUSVOLT signal error (analogue input)Internal POST check failurecritical
200-34Incorrect PK RF board installedSystem detects wrong RF board type during POSTcritical
200-35RF relay 1 (socket) not operatingInternal relay test failure during POSTcritical
200-36RF relay 2 (poles) not operatingInternal relay test failure during POSTcritical
200-42CPU POST Thermistor failureTemperature sensor read error during POSTcritical
200-43CPU POST failureProcessor self-test fails at startupcritical
200-44RAM test failure [POST check]Memory read/write test fails during startupcritical
200-45Crystal failure [POST check]System clock oscillator malfunction detectedcritical
200-47Analogue reference failure [POST check]Voltage reference circuit test failscritical
200-48CLAMP_SET failure [POST check]Clamp voltage setting circuit fails testcritical
200-49BOOST_SET failure [POST check]Boost voltage setting circuit fails testcritical
200-50CURRLIM_SET failure [POST check]Current limit setting circuit fails testcritical
200-51PEAK_SET failure [POST check]Peak current setting circuit fails testcritical
200-52COAG_CUT failure [POST check]Coagulation/cutting mode detection failscritical
200-53PK_CUT_COAG failure [POST check]Peak cut/coag detection failscritical
200-54PK_SKT_SET failure [POST check]Peak socket setting circuit fails testcritical
200-56NO_RF failure [POST check]RF output detection fails during self-testcritical
200-65RFBUS_VOLTS failure [POST check]RF bus voltage measurement failscritical
200-67PSU_STATUS failure [POST check]Power supply unit status detection failscritical
200-68Mains input failure [POST check]Mains power detection fails during startupcritical
200-69OVERDOSE failure [POST check]Overdose protection circuit test failscritical
200-70PWM low failure [POST check]Pulse width modulation circuit test failscritical
200-71ID CAL circuit failure [POST check]Identification/calibration circuit test failscritical
200-72BUSLIM failure [POST check]Bus current limit circuit test failscritical
200-73CURRLIM failure [POST check]Current limit feedback circuit test failscritical
200-74VOLTLIM failure [POST check]Voltage limit feedback circuit test failscritical
200-76VCC_ANA_DIV_2 failure [POST check]Divided analogue supply voltage out of rangecritical
200-7712V failure [POST check]12V supply voltage out of specificationcritical
200-780V failure [POST check]Ground reference measurement failscritical
200-79PK-RF Thermistor failure [POST check]RF board temperature sensor failscritical
200-80RF_VOLTAGE [POST check]RF output voltage measurement failscritical
200-81RF_CURRENT failure [POST check]RF output current measurement failscritical
200-82ANALOGUE_REF failureAnalogue reference voltage out of specificationcritical
200-83VCC_ANA_DIV_2 failureDivided analogue supply out of rangecritical
200-8412V failure12V supply voltage out of specificationcritical
200-850V failureGround reference voltage out of specificationcritical
200-86RF_DET stuck high failureRF detection signal remains high continuouslycritical
200-87PSU_STATUS failurePower supply status signal malfunctioncritical
200-88Mains input failureMains power detection circuit malfunctioncritical
200-89BUSLIM failureBus current limit circuit test failscritical
200-90CURRLIM_SET failureCurrent limit setting circuit failscritical
200-92Over active ID countExcessive ID (instrument identification) counts detectedcritical
200-93External Watchdog unable to reset CPUExternal watchdog timer fails to reset processorcritical
200-94Internal Watchdog unable to reset CPUInternal watchdog timer fails to reset processorcritical
300-10Internal overheatingTemperature exceeds thermal shutdown thresholdwarning
300-11Excess RF input voltage error (RFBUS > set)RF bus voltage exceeds set limiterror
300-12PK-RF board overheatingRF board temperature exceeds thermal thresholdwarning
300-14Accessory error: excessive RF output currentRF output current exceeds safe limit, output shortederror
300-19Two instruments of the same type attachedSystem detects duplicate instrument types connectedwarning
300-20Unsupported instrument typeSystem detects incompatible instrumentwarning
300-22Detected instrument with buildup of debrisInstrument electrode contamination detectedinfo
400-10Footswitch BLUE pedal stuckBlue footswitch button continuously depressed during startuperror
400-11Footswitch YELLOW pedal stuckYellow footswitch button continuously depressed during startuperror
400-12Footswitch MODE pedal stuckMode footswitch button continuously depressed during startuperror
400-14Instrument identification circuit faultHandpiece cable or instrument not detected correctlyerror
400-15Front panel Yellow UP button stuckYellow UP button continuously depressed during startuperror
400-16Front panel Yellow DOWN button stuckYellow DOWN button continuously depressed during startuperror
400-17Front panel Blue UP button stuckBlue UP button continuously depressed during startuperror
400-18Front panel Blue DOWN button stuckBlue DOWN button continuously depressed during startuperror
400-19Front panel MODE button stuckMODE button continuously depressed during startuperror
400-20Front panel STANDBY button stuckSTANDBY button continuously depressed during startuperror
400-22Handswitch Yellow button stuckYellow handswitch button continuously depressed during startuperror
400-23Handswitch Blue button stuckBlue handswitch button continuously depressed during startuperror
400-24Handswitch Mode button stuckMode handswitch button continuously depressed during startuperror
600-01POST Aux board ADC check failureAuxiliary board ADC does not respond in timecritical
600-02Aux board relay checks failureAuxiliary board relay malfunction detectedcritical
600-03Aux board relay 1 failRelay 1 on auxiliary board fails testcritical
600-04Aux board relay 2 failRelay 2 on auxiliary board fails testcritical
600-05Aux board relay 3 failRelay 3 on auxiliary board fails testcritical
600-06Aux board relay 4 failRelay 4 on auxiliary board fails testcritical
600-07Aux board relay 5 failRelay 5 on auxiliary board fails testcritical
600-08Aux board relay 6 failRelay 6 on auxiliary board fails testcritical
600-09Aux board relay 7 failRelay 7 on auxiliary board fails testcritical
600-10Aux board relay 8 failRelay 8 on auxiliary board fails testcritical
600-11Aux board +5vPC_A failAuxiliary board 5V supply voltage malfunctioncritical
600-12Aux board +V_OSET failAuxiliary board offset voltage supply malfunctioncritical
600-13Handswitching errorHandswitch control circuit malfunctioncritical

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.