Getinge CS300: troubleshooting and error codes
The ServiceForge knowledge base holds 46 documented symptoms for the Getinge CS300, with 69 ranked causes, 82 diagnostic checks with test points and expected values, and 86 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Autofill failure - purge pressure did not reach target patient safety
IAB tubing and Safety Disk fail to reach target purge pressure within required time; fault codes 60, C0, D0, 1A0
Ranked causes
- 1. Slow purge, vacuum leak or insufficient vacuum (common)
- 2. Leak in IAB circuit or catheter (common)
- 3. Extreme water buildup in lines (occasional)
- 4. Clogged purge line filter (occasional)
- 5. K3/K5 solenoid valve failure (occasional)
Run Safety Disk Leak Test
- Test point
- Safety Disk test port
- Expected
- No leaks detected; purge reaches target pressure within time limit
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Autofill failure - cylinder did not empty patient safety
Helium volume cylinder fails to empty (close S2) within 10 seconds at beginning of Autofill; system logs fault code 10 or 140
Ranked causes
- 1. No IAB connected or disconnected (common)
- 2. Empty sensor (S2) failure or misalignment (common)
- 3. No vacuum or vacuum loss (occasional)
- 4. Pinched or obstructed tubing to volume cylinder (occasional)
Verify IAB connection
- Test point
- IAB connector and pneumatic ports
- Expected
- IAB firmly seated with no gaps; audible click on connection
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pneumatic performance - low average pressure patient safety
Average pressure below 300 mmHg during Pneumatic Performance Test; indicates leak, regulator malfunction, or pump/compressor issue
Ranked causes
- 1. Leak in pneumatic system (common)
- 2. 8 PSI regulator needs adjustment or replacement (occasional)
- 3. Clogged muffler restricting flow (occasional)
- 4. Pump motor requires rebuild (rare)
Pneumatic Performance Test - average pressure measurement
- Test point
- Safety Disk pressure circuit during 30-second 150 bpm pump cycle
- Expected
- Average pressure between 300 and 413 mmHg, 300–413 mmHg
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Autofill failure - leak or blood detected patient safety
Blood or leak detected during IAB purge cycle at various stages; fault codes with prefix C, E, 4, 8, 9 (e.g. CE0, 4E0, 82 0)
Ranked causes
- 1. Blood detected from shuttle pressure surveillance (C/E prefix) (common)
- 2. Leak in catheter or Autofill system (4 prefix) (common)
- 3. Blood detected optically (8/9 prefix) (occasional)
Run Safety Disk Leak Test
- Test point
- Safety Disk test port and all IAB connections
- Expected
- No leaks detected; pressure holds
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Autofill failure - cylinder did not reach full sensor patient safety
Helium volume cylinder fails to reach full sensor (close S1) within 10 seconds during Sensor IAB calibration; fault code 50
Ranked causes
- 1. Low helium pressure in tank (common)
- 2. Full sensor position incorrect or volume cylinder sluggish (occasional)
Check helium tank pressure
- Test point
- Helium tank gauge
- Expected
- Pressure above minimum operating level per system spec
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Atmospheric transducer gain and offset
Calibrate Atmospheric absolute pressure transducer gain at 200 mmHg and offset at atmospheric pressure
Ranked causes
- 1. Atmospheric transducer gain out of specification (unknown)
- 2. Atmospheric transducer offset out of specification (unknown)
Attach manometer to atmospheric transducer U54 using tubing adapter
- Test point
- Atmospheric transducer U54 port
- Expected
- Adapter and tubing installed securely
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Balloon (Shuttle) transducer gain and offset
Calibrate Shuttle absolute pressure transducer gain at 380 mmHg and offset at atmospheric pressure
Ranked causes
- 1. Shuttle transducer gain out of specification (unknown)
- 2. Shuttle transducer offset out of specification (unknown)
Remove tubing from IAB FILL PORT and connect precision manometer
- Test point
- IAB FILL PORT on Safety Disk
- Expected
- Manometer connected securely
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fiber Optic Signal Count or Level below acceptable threshold caution
Fiber Optic Test fails Signal Count (< 55) or Signal Level (< 153) criteria; indicates lamp ferrule contamination or module failure
Ranked causes
- 1. Lamp ferrule contamination or debris (common)
- 2. Fiber Optic module failure or degradation (occasional)
Fiber Optic Test - Signal Count parameter
- Test point
- Sensor Module signal count during 30-second pump cycle
- Expected
- Signal Count ≥ 55 count, ≥55
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
K5 solenoid leak after K3 deactivation patient safety
Vacuum leak detected when K5 activated in test sequence; possible hole in Safety Disk or leak through K3 solenoid
Ranked causes
- 1. Hole in Safety Disk membrane (common)
- 2. Leak through K3 solenoid (occasional)
Safety Disk Leak Test - K5 activation phase
- Test point
- Safety Disk pressure circuit
- Expected
- Pressure change ≤ ±4 mmHg (adult disk) or ≤ ±6 mmHg (adult chamber) or ≤ ±10 mmHg (pediatric chamber) over 2 minutes, ±4 mmHg adult, ±6 mmHg adult chamber, ±10 mmHg pediatric
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Motor speed calibration out of tolerance caution
Motor actual speed not within 3% of 1760 RPM after adjustment; circuit or motor defect suspected
Ranked causes
- 1. Motor Controller board malfunction (occasional)
- 2. DC motor defect or bearing wear (occasional)
Motor Calibration/Test - initial speed set to 1600 RPM
- Test point
- Motor Controller board (8 volt input)
- Expected
- Actual motor speed within 20% of 1600 RPM (1280–1920 RPM), ±20%
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Perform motor speed and temperature verification caution
Technician monitors DC motor/pump speed and internal unit temperature via Pneumatic System Test screen to verify motor functionality and cooling adequacy during extended operation.
Ranked causes
- 1. Motor speed below nominal or not running (rare)
- 2. Unit overheating due to ambient temperature or blocked venting (occasional)
Observe motor speed reading during normal operation
- Test point
- Pneumatic System Test screen, Motor Speed field
- Expected
- Approximately 1762 rpm
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pneumatic performance - inadequate vacuum patient safety
Average vacuum reading above -176 mmHg (less vacuum) at sea level; indicates pneumatic leak or compressor issue requiring rebuild
Ranked causes
- 1. Leak in pneumatic system (common)
- 2. Compressor requires rebuild (occasional)
Pneumatic Performance Test - average vacuum measurement
- Test point
- Safety Disk vacuum circuit during 30-second pump cycle
- Expected
- Minimum -176 mmHg at sea level (adjust per altitude table for higher elevations), ≥ -176 mmHg sea level
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Run pneumatic safety disk leak test patient safety
Technician performs automated leak detection test via Service Diagnostics to check Safety Disk, solenoid control lines, and shuttle/drive circuit integrity; system measures pressure decay over time.
Ranked causes
- 1. Leak in Safety Disk assembly or IAB circuit (occasional)
- 2. Leak in fill circuit or transducer line (rare)
Verify all solenoids deactivated at start
- Test point
- Pneumatic System Test screen
- Expected
- No solenoids highlighted in reverse graphics; system in safe state
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Safety Disk membrane leak suspected patient safety
Vacuum leak detected during pneumatic test; Safety Disk may have hole in membrane or leak through K5 solenoid after K3 activation
Ranked causes
- 1. Hole in Safety Disk membrane (common)
- 2. Leak through K5 solenoid (occasional)
Safety Disk Leak Test - K3 activation phase
- Test point
- Safety Disk pressure circuit
- Expected
- Pressure change ≤ ±4 mmHg (adult disk) or ≤ ±6 mmHg (adult chamber) or ≤ ±10 mmHg (pediatric chamber) over 2 minutes, ±4 mmHg adult, ±6 mmHg adult chamber, ±10 mmHg pediatric
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Safety Vent (K6A) obstruction or failure caution
Drive manifold pressure does not drop to 150 mmHg within 2 seconds during vent test; K6A solenoid may be stuck or vent tubing kinked
Ranked causes
- 1. K6A solenoid valve failure or obstruction (common)
- 2. Kinked or obstructed safety vent tubing (occasional)
Safety Vent Test
- Test point
- Drive manifold pressure after K6A activation
- Expected
- Pressure drops to approximately 150 mmHg above atmosphere in less than 2 seconds, <2 seconds
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Autofill failure - software error or invalid state patient safety
Software error or system did not advance to next pneumatic state; fault codes 40, 80, A0, 100, 120, 1D0
Ranked causes
- 1. Internal software or state machine error (rare)
Power cycle system and retest Autofill
- Test point
- System power and Autofill function
- Expected
- Autofill completes successfully without state machine error
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Autofill failure - transducer calibration issue patient safety
Transducers could not be zeroed or reached target atmospheric pressure at end of Autofill cycle; fault code 90 or 180
Ranked causes
- 1. Transducer calibration failure or sensor error (common)
Perform transducer atmospheric calibration
- Test point
- Transducer calibration menu or test port
- Expected
- All three transducers zero correctly at atmospheric pressure
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood Detect function test patient safety
Verify Blood Detect sensor responds when light path is blocked during autofill
Ranked causes
- 1. Blood Detect sensor malfunction (rare)
Initiate autofill with sensor light path blocked
- Test point
- K6A LED and display message
- Expected
- 'Blood Detected' message appears and K6A LED extinguishes
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Autofill fill and purge times
Verify Bimba fill time, Shuttle fill time, and shuttle purge time meet specifications
Ranked causes
- 1. Over-pressure or leaks in fill gas path (occasional)
Measure Bimba fill time in Autofill status box
- Test point
- Bimba fill time display
- Expected
- 2.5 to 6 seconds, ±3.5 seconds
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Autofill function test patient safety
Verify system completes autofill cycle with catheter extender and balloon attached
Ranked causes
- 1. Autofill circuit malfunction (rare)
Initiate autofill with balloon attached
- Test point
- IAB Catheter Port with 40 cc balloon
- Expected
- Autofill cycle completes without error
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Autofill volume
Verify and adjust autofill volume to achieve correct shuttle pressure reading
Ranked causes
- 1. Volume cylinder setting out of specification (common)
Connect 38.5 cc syringe and calibration chamber
- Test point
- IAB Catheter Extender Input on Safety Disk
- Expected
- Helium pressure X4 > 215 PSIG
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Battery runtime
Verify battery can sustain system operation for minimum specified duration from full charge
Ranked causes
- 1. Battery runtime insufficient (unknown)
Ensure batteries fully charged
- Test point
- System charging indicator
- Expected
- Steady BATTERY CHARGING indicator
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Battery voltage and charging high voltage
Verify battery voltage and charging current during AC mains operation and battery-only operation
Ranked causes
- 1. Battery voltage out of specification (unknown)
Verify BATTERY CHARGING indicator state
- Test point
- System display
- Expected
- Steady illumination
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Blood Back sensor voltage verification
Verify Blood Back sensor voltage is within correct range and configure switches if needed
Ranked causes
- 1. Blood Back sensor voltage out of specification (occasional)
Measure Blood Back sensor voltage
- Test point
- TP4 (+) and TP5 (-) on Solenoid Driver board P/N 0670-00-0639
- Expected
- 1 to 2.3 volts, ±0.65 volts
- Tools
- DVM
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Condensate removal module voltage high voltage
Verify voltage to condensate removal chiller block is within specification
Ranked causes
- 1. Condensate removal module voltage out of specification (rare)
Measure voltage to condensate removal chiller block during pumping
- Test point
- L6 (near JP3) and TP31
- Expected
- 2.8 volts V, ±0.1 V
- Tools
- DVM
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Drive transducer offset
Drive pressure does not read zero when vented to atmosphere
Ranked causes
- 1. Drive transducer offset misalignment (common)
Vent drive pressure and measure reading
- Test point
- Drive transducer
- Expected
- 0 mmHg
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Drive transducer pressure
Drive transducer reading does not match Shuttle transducer reading during calibration procedure
Ranked causes
- 1. Drive transducer gain misalignment (common)
Activate K6A and K7, measure drive transducer pressure
- Test point
- Drive transducer
- Expected
- ±2 mmHg of Shuttle transducer reading
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Drive transducer setup
Connect DVM to Drive transducer and initialize zero pressure reference in Service Diagnostics
Ranked causes
- 1. Drive transducer not configured for calibration (unknown)
Connect DVM to Drive transducer signal and ground
- Test point
- TP24 (DRIVE_PRES) to TP33 (GND_A)
- Expected
- DVM connected with stable reading
- Tools
- DVM
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Fiber Optic Test
Verify lamp function, output, PWM, signal count and signal level of sensor module
Ranked causes
- 1. Fiber optic sensor module malfunction (rare)
Run Fiber Optic Test and verify waveforms
- Test point
- Fiber Optic Sensor Module output
- Expected
- Two similar waveforms display during 30-second test
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Full sensor adjustment
Adjust full sensor position to detect end of autofill correctly
Ranked causes
- 1. Full sensor misaligned (occasional)
Position full sensor for LED transition
- Test point
- FULL LED on volume cylinder
- Expected
- LED turns on then off as sensor moves right
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Internal transducer offset check
Verify Shuttle (Balloon), Atmospheric, and Drive transducer offset values are within acceptable range
Ranked causes
- 1. Transducer offset out of specification (unknown)
Read Shuttle transducer offset from display
- Test point
- System display Pneumatic System Test screen
- Expected
- Between -4 mmHg and 4 mmHg
- Tools
- System display
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pneumatic system pressure verification
Atmospheric, Shuttle and Drive transducers not reading within tolerance after calibration
Ranked causes
- 1. Transducer readings out of specification (occasional)
Run Pneumatic System test and verify transducer readings
- Test point
- Atmospheric, Shuttle, Drive transducers
- Expected
- All readings within 2 mmHg of each other; offsets between -4 and +4 mmHg
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Power supply voltages high voltage
Verify AC power supply output voltages are within tolerance on Main Board and power supply assembly connectors
Ranked causes
- 1. Power supply out of specification (unknown)
Verify red LEDs CR4 through CR9 illuminated on Main Board
- Test point
- Main Board
- Expected
- All six LEDs illuminated
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pressure regulator and vacuum verification
Verify pressure regulator output and vacuum levels in drive pneumatics
Ranked causes
- 1. Pressure regulator out of specification or pump motor issue (occasional)
Measure vacuum with K6A and K7 active
- Test point
- X2-Drive-Pres
- Expected
- 0 to 125 mmHg
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Reference voltage check
Verify 5V reference voltage on Front End board
Ranked causes
- 1. Reference voltage out of tolerance (unknown)
Measure reference voltage at TP22 to GND_A
- Test point
- TP22 to TP33 (GND_A)
- Expected
- 5V ±0.1V, ±0.1V
- Tools
- DVM
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Safety Disk leak test
Procedure to verify Safety Disk and solenoid integrity for vacuum and pressure leaks
Ranked causes
- 1. Safety Disk membrane hole or solenoid leak (rare)
Run Safety Disk Leak Test with luer plug
- Test point
- Safety Disk IAB catheter port
- Expected
- Pressure change ≤±4 mmHg for all tests, ±4 mmHg
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Solenoid driver board blood detect circuitry
Verify blood detect sensor voltage and configure gain/offset settings
Ranked causes
- 1. Blood Detect sensor voltage out of range (occasional)
Measure Blood Detect voltage
- Test point
- TP2 (+) and TP5 (-) on Solenoid Driver board P/N 0670-00-0639
- Expected
- 2.5 to 5 volts, ±1.25 volts
- Tools
- DVM
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display test fails
Base Screen Memory or Window Screen Memory test shows error or displays error location message on display
Ranked causes
- 1. Display board or interconnecting cable defect (unknown)
Base Screen Memory test
- Test point
- Display screen top to bottom
- Expected
- Screen illuminates uniformly and displays 'Base Screen Memory OK' message
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Drive pneumatics leak detected
Drive pressure readings continue to decrease during calibration indicating pneumatic leak
Ranked causes
- 1. Leak in drive pneumatics (occasional)
Run K6, K6A, K7, K8 Leak Test
- Test point
- Drive pneumatics circuit
- Expected
- Test #1 ±45 mmHg, Test #2 ±65 mmHg, Test #3 ±20 mmHg
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fiber Optic lamp PWM, output, or signal parameters out of spec caution
Fiber Optic Test fails PWM, Lamps Output, or Lamps Output/PWM criteria; lamp function degraded
Ranked causes
- 1. Fiber Optic lamp degradation or failure (common)
Fiber Optic Test - PWM parameter
- Test point
- Sensor Module lamp PWM control signal
- Expected
- PWM value 1–69 PWM units, 1–69
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Helium tank pressure offset out of specification
Helium pressure circuit calibration offset exceeds acceptable limits when tank removed from pump
Ranked causes
- 1. Helium transducer or circuit calibration drift (occasional)
Helium Tank Calibration test
- Test point
- Helium pressure circuit offset
- Expected
- Offset within specification limits
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
K6/K6A/K7/K8 drive section leak patient safety
Pressure loss or gain detected in drive pneumatics during 60-second hold periods at atmosphere, 8 PSI, and vacuum
Ranked causes
- 1. Leak in drive pneumatic circuit (common)
K6/K6A/K7/K8 Leak Test - Test #1 (60 sec at atmosphere)
- Test point
- Drive pneumatic transducer X2
- Expected
- Differential pressure ≤ ±45 mmHg, ±45 mmHg
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keypad or switch does not respond
Key presses not registered on Key Depressed display area or keypad LEDs not lighting
Ranked causes
- 1. Front panel switch or keypad wiring defect (unknown)
Verify key response and LED illumination
- Test point
- Front panel keys and associated LEDs
- Expected
- Each key press displays in Key Depressed area and lights corresponding LED(s)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Modem test fails
Modem loopback test shows errors or FAIL status on modem test screen
Ranked causes
- 1. Internal modem defect or configuration mismatch (unknown)
Modem loopback test
- Test point
- Internal modem communication
- Expected
- Status shows TEST PASSED and error count is 000
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Perform autofill calibration test patient safety
Technician runs automated autofill calibration via Functional Tests menu to verify fill volume accuracy and transducer calibration; system measures transducer offset differences from drive transducer baseline.
Ranked causes
- 1. Autofill volume out of calibration (occasional)
Check transducer offset values after autofill calibration
- Test point
- Pneumatic System Test screen, pressure readings in brackets
- Expected
- Shuttle and atmospheric transducers offset within specification; helium transducer offset (xx) indicates out of spec
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
RS-232 port communication error
Transmitted and received data do not match or data missing on RS-232 test
Ranked causes
- 1. RS-232 connector pins not jumped or defective serial communication channel (common)
Jump RS-232 transmit and receive pins
- Test point
- Rear panel RS-232 connector pins 2 and 3
- Expected
- Pins 2 and 3 electrically connected via bus wire
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
01 | RTOS and virtual device manager faults | Real-time operating system or virtual device manager error occurs | warning |
02 | Set Model Type Fault | Model type configuration error detected | critical |
1 | Power up Watch-dog test failed | During power-up self-test | critical |
10 | Helium volume cylinder did not empty (close S2) within 10 seconds at beginning of Autofill | Fault occurs during initial Autofill phase when S2 sensor should close | error |
117 | Front End A/D reference failure at power up | During analog-to-digital converter test | error |
118 | Front End HC11 cannot communicate with Main HC11 | During inter-CPU handshake | critical |
119 | Front End CPU failure | During Front End operation | critical |
120 | Monitor/Keypad CPU failure | During Monitor/Keypad operation | critical |
122 | Keypad did not respond to poll | During keypad polling | error |
123 | Invalid error condition | During error state detection | unknown |
140 | Helium volume cylinder did not empty (close S2) within 10 seconds following dead volume calculation | Fault occurs after dead volume calculation when S2 should close | error |
150 | Helium volume cylinder did not empty (close S2) within 10 seconds following dead volume calculation with no IAB | Fault occurs after dead volume when no IAB connected | error |
2 | 68020 CPU not functioning correctly | During boot sequence | critical |
20 | Initial IABP purge failure, could not empty Safety Disk | Fault occurs at beginning of Autofill when purging Safety Disk | error |
21 | ROM checksum test failed | During power-up self-test | critical |
22 | System RAM address test failed | During memory test | critical |
23 | RAM walking ones test failed | During memory pattern test | critical |
24 | RAM pattern test failed | During memory diagnostic | critical |
25 | UART communication loop back test failed | During serial port self-test | error |
26 | NMI test failed | During interrupt test | error |
27 | Watch-dog test failed | During watchdog timer test | error |
28 | Start up test failure | During initial power-up | critical |
29 | IABP/Comm HC11 software flip-flop error at power up | During boot sequence | error |
3 | Boot checksum incorrect | During power-up verification | critical |
30 | Helium volume cylinder did not move (open S2 empty sensor) within 5 seconds | Fault occurs when S2 empty sensor should open to release pressure | error |
30 | Service diagnostics ROM failure | During diagnostics self-test | error |
32 | Power-up RAM test failure | During memory self-test | critical |
33 | NMI detected during power up | During startup sequence | error |
33 | Bad system NVRAM info record fault | Non-volatile RAM contains corrupt or invalid system information | error |
34 | No communication with 68020 | During CPU handshake | critical |
35 | No communication with 6809 | During CPU handshake | critical |
36 | Software trap detected during power up | During boot sequence | critical |
37 | Soft interrupt during power up | During startup | error |
38 | Prolonged inflation test failure | During pressure test | error |
39 | Comm HC11 failed to interrupt | During interrupt sequence | error |
4 | Power up RAM test failed | During memory self-test | critical |
40 | IABP 6809 Datasette not compatible with 68020 Datasette | During data verification | error |
40 | Software error or blood detected during delay state | Fault occurs when system does not advance from delay state | error |
41 | Monitor 68020 Datasette not compatible with 6809 Datasette | During data verification | error |
42 | IABP 6809 Datasette not compatible with Front End HC11 | During data verification | error |
43 | IABP 6809 Datasette not compatible with Comm HC11 | During data verification | error |
44 | IABP 6809 Datasette not compatible with Monitor/Keypad HC11 | During data verification | error |
45 | System failure | General system failure condition detected | critical |
46 | Monitor 68020 Datasette not compatible with Front End HC11 | During data verification | error |
47 | Monitor 68020 Datasette not compatible with Comm HC11 | During data verification | error |
48 | Monitor 68020 Datasette not compatible with Monitor/Keypad HC11 | During data verification | error |
4E0 | Leak detected during purge with blood back indication | Leak or blood detected after reaching first surveillance level with leak prefix | error |
5 | 68020 code transfer to DRAM did not verify correctly | During boot sequence | critical |
50 | Motor speed out of specification | During motor speed test | error |
50 | Helium volume cylinder did not reach full sensor (close S1) within 10 seconds during Sensor IAB calibration | Fault occurs during calibration phase when cylinder should reach S1 full position | error |
51 | Adjusted motor speed out of specification | After motor speed adjustment | error |
52 | Drive transducer offset failure | During transducer calibration | warning |
53 | Shuttle transducer offset failure | During transducer calibration | warning |
54 | ATM transducer calibration failure | During transducer calibration | warning |
55 | 68020/6809 Sync failure entering system configuration mode | During configuration mode entry | error |
56 | Excessive drive pressure | During pressure monitoring | warning |
57 | Safety Vent test failed | During safety vent verification | warning |
57 | Autofill failure with detailed failure mode in last two or three digits | Autofill process fails at various stages; specific cause indicated by fault code suffix | error |
58 | Power-up Vent test failed | During power-up self-test | warning |
6 | Interrupt fault, 68020 exception condition during boot up | During power-up sequence | critical |
60 | Final IABP purge failure, could not reach 150 mmHg absolute pressure within 14 seconds | Fault occurs during final purge phase of Autofill when target pressure not reached | error |
61 | Comm HC11 ROM test failed | During ROM self-test | error |
62 | Comm HC11 RAM address test failed | During memory test | error |
63 | Comm HC11 RAM walking ones test failed | During memory pattern test | error |
64 | Comm HC11 RAM pattern test failed | During memory diagnostic | error |
65 | Comm HC11 shared RAM address test failed | During shared memory test | error |
66 | Comm HC11 shared RAM walking ones test failed | During shared memory pattern test | error |
67 | Comm HC11 shared RAM pattern test failed | During shared memory diagnostic | error |
68 | Comm HC11 Watch-dog test failed | During watchdog timer test | error |
69 | Comm HC11 power up RAM test failure | During power-up memory test | critical |
70 | IAB tubing and Safety Disk did not fill with helium from volume cylinder within 10 seconds | Fault occurs when filling phase does not complete in time | error |
70 | Comm HC11 main failure | During Comm HC11 operation | critical |
820 | Autofill failure with blood detected optically | Blood detected by optical blood detection sensor during Autofill | error |
90 | Transducer atmospheric calibration failure | Fault occurs at end of Autofill cycle when transducers cannot zero | error |
91 | Front End HC11 ROM test failed | During ROM self-test | error |
92 | Front End HC11 RAM address test failed | During memory test | error |
93 | Front End HC11 RAM walking ones test failed | During memory pattern test | error |
94 | Front End HC11 RAM pattern test failed | During memory diagnostic | error |
98 | Front End HC11 Watch-dog test failed | During watchdog timer test | error |
B0 | Unexpected shuttle gas pressure rise detected during IAB purge cycle | Fault occurs when shuttle pressure increases unexpectedly during purge | warning |
C0 | 14 second IAB purge time-out at initial purge level prior to Safety Vent test | Fault occurs when initial purge phase exceeds 14 seconds | error |
CE0 | 14 second IAB purge time-out with blood back detected from shuttle pressure | Purge fails to complete within 14 seconds and blood is detected | error |
D0 | Failure to reach first surveillance purge level in 1 second or 14 second IABP purge time-out | Fault occurs when purge level not reached in time or exceeds time limit | error |
E0 | Leak or blood detected after reaching first surveillance purge level | Fault occurs when leak or blood detected during purge phase | warning |
F0 | Leak or blood detected after reaching second surveillance purge level | Fault occurs when leak or blood detected in later purge phase | warning |
Maintenance Required Code 5 | Transducer calibration out of specification; specifically helium pressure transducer (X4) offset error | Helium transducer offset (xx) displayed on Pneumatic System Test screen after autofill calibration | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.